1 Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it!

2 Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers, it is important to know the exact instantaneous distribution of the intensity across the laser beam, i.e. the laser's mode. During time and usage of the intensity. This has a big impact on the correct working of the laser beam. By analysing the laser beam, adjustments can be made to This results in: Less waste Higher production rates Faster set up time More consistency of processes

3 Our Laser Beam Analyser = your solu on! All beam analysing methods known to date are cumbersome and have one serious drawback: they do not allow measurement and continuous monitoring during an experiment or during material processing. Our offered Laser Beam Analyser (LBA) however, does make it possible to monitor the intensity distribution across the laser beam during an experiment or material processing, without noticeable loss of beam intensity (the intensity is reduced by mere 0.8 %). The compact system can be setup easy and quickly and is easy to operate. Connected to an oscilloscope or with our Laser Diagnostic Software on a pc, you are able to analyze the laser beam of your CO2 or YAG laser in an instant! With our LBA you can easily measure the laser during normal working

4 LBA Working Principle Driven by a synchronous motor, a needle with a high reflective surface rotates in a plane perpendicular to the beam. While passing across the beam it reflects a tiny fraction of the beam onto two pyroelectric detectors. The needle is moving at constant speed and the power distribution across the beam section, is represented by the amplitude of the amplified signals from the detectors. The LBA is designed for displaying simultaneous the intensity profiles in the x and y axes (see picture) on a dual trace oscilloscope. APPLICATIONS 1. Adjustment of CW-Lasers The LBA is mounted and centred on the beam in a convenient position while the oscilloscope is placed beside the mirror to be adjusted. During adjustment the variation of the intensity profile can be directly observed and the laser can easily be adjusted to get an optimum mode. 2. Intensity Profile Control in Material Processing The LBA is permanently installed in an easy accessible place in the beam path between the laser and the work station, the oscilloscope being installed at the work station. This arrangement allows control of the instantaneous mode and power during the working process. The beam intensity loss due to the LBA is only 0.8%. Permanent installation in the beam path does not impair production quality. When a laser is used in production - e.g. for steel cutting - a "good mode" is a prerequisite for a clean cut. The Laser Beam Analyser offers permanent control of the intensity profile. Further development toward fully automatic operation is feasible. The signal can be processed electronically and be used for automatic production control by comparison of actual to optimum parameters. There are some interesting aspects to lab applications. It is generally known that in metals processing a considerable portion of the radiation is reflected from the metal surface and reenters the laser via the focusing lens in the form of a parallel beam. This retro reflection increases the power density inside the laser and can lead to a substantial variation of the laser's mode. Direct observation of this phenomenon by means of the LBA allows the optimization of the process parameters.

5 3. Measurement of Beam Divergence As the rotating needle passes across the beam at constant speed there is a fixed relationship between passage time and beam width. Taking model LBA1 for an example, 400µs shown on the CRT screen are equivalent to a beam width of 10mm. By successively installing the Beam Analyser at different distances from the laser, the beam diameter (1/e) can be measured and divergence can easily be calculated from these values. 4. Adjustment of Pulsed-Lasers For pulsed lasers it is suggestible (even not indispensable) to employ a memory oscilloscope. When a laser pulse meets the needle, a very narrow "peak" will appear on the screen in a position corresponding to that of the needle at the pulse instant. As the time goes on, the screen will be filled with many such "peaks", whose enveloping curve represents the laser's mode. To avoid a long waiting time with low repetition rate lasers, a synchronization signal is supplied by the LBA to drive a laser pulse when, after every revolution, the needle is in a slightly different position. A reset signal is also provided to erase the screen after every scanning. Furthermore the facility of locking the needle across the beam, enables the observation of the pulse's time-shape, point-to-point in the beam cross section. 5. Measurements of Focus The needle's high reflective surface and its moving speed make the LBA suitable for measurement of the focus diameter. The unit is displaced along the beam until the minimum beam diameter has been found. This measurement allows to determine the power density at the focus point, as well as the influence of lenses of different focal lengths and shapes. Distorted lenses or mirrors can be spotted by comparative measurement. SPECIFICATIONS The LBA is available in three different sizes: LBA1 for beam diameter till 25mm; 50(60) Hz scanning rate LBA2 for beam diameter till 40mm; 50(60) Hz scanning rate LBA3 for beam diameter till 60mm; 50(60) Hz scanning rate The detectors window is available either for CO2 or YAG laser, as well as for both The detectors amplifier is switchable in 4 gain ranges ( ). The spatial resolution and the sensitivity depend linearly on the chosen gain. With gain 0.02 the following values are reached: spatial resolution: 4µm sensitivity: 0.3mV/mm 2 increasing the gain increases the sensitivity and reduces the resolution in the same factor. Output noise signal is always less than 5mV (rms) Please specify in your order required version and model, detector's window, supply voltage (220 or 110V) and frequency (50 or 60Hz). Dimension (mm) LBA1 : 187x89x185 / LBA2 293x130x287 / LBA3 360x150x355

6 LDS - Laser Diagnos c System To be able to influence the beam recursively or to get comparison values of the measurements, an evaluation system is needed which goes beyond the capacity of an oszillograph. The signals of an LBA can be made visible on a monitor with a PC (Windows based), a fast speed ADC-Plug-In board and the appropriate software (Laser Diagnostic System LDS). In this case the computer is able to do several operations as for example monitoring the beam An interesting application can be found for industrial production. The intensity distribution of the laser beam while correct configured will be stored. The actual signal is than during production constantly compared with the stored required information. The tolerance values for exceeding the correct values can be chosen. The deviation of the mode structure or intensity, that adversely affects the quality, is immediately detected and displayed on the computer audibly and visually. Since the signal height is directly proportional to the power, the highest power at optimal mode be adjusted fast and easy again. In addition all important laser beam parameters, like peak values, beam diameter at the point 1/e2 and the maximum value e/2 as well as the power/intensity are constantly displayed. For rapid pulsed lasers, the LDS averages up to 16 measurements. In this way, short -term variations which have in regard to the mode profile no importance, are reduced. A print of the display is possible at any time.

7 LDS - Laser Diagnos c System With a software extension, the distribution of the laser mode can be represented in 16 colours over the entire sectional view. Especially for long time testing, this view provides an excellent overview providing the user a quick and easy overview of the Mode structure. The 3-D representation of the laser diagnostic system provides beside the user friendliness urements. to also receive in a decreed manner a more in addition to user friendliness on account of the menu-driven functions via the possibility to information from the data supplied by the LBA. tensity with all needed information This so-called mountain view, provides at a glance an overview of the maximum power, the location of it in the beam as well as information about sub-maximums which appear for the first time in this view. From up to 440 measurements, at which the LBA is driven through the beam by means of a stepping-motor, a mountain view is built up with a resolution of 50µm. The built up picture can be viewed from all angles. As additional information is provided by means of so-called silhouettes, representing the course of the intensity maxima separately again. The LDS provides height line cuts which provide information about the beam diameter of a certain intensity. LDS and LBA are universal tools providing excellent laser beam analysis helping to optimise CO2 or YAG high performance laser usage.

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera PRELIMINARY POWER DETECTORS ENERGY DETECTORS MONITORS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS CMOS Beam Profiling Camera AVAILABLE MODEL Beamage 3.0 (⅔ in CMOS

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing ECNDT 2006 - Th.1.1.4 Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing R.H. PAWELLETZ, E. EUFRASIO, Vallourec & Mannesmann do Brazil, Belo Horizonte,

More information

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS FOCUS ON FINE SOLUTIONS THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS Welding lasers from ROFIN ROFIN s laser sources for welding satisfy all criteria for the optimized laser

More information

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO) 2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency

More information

Connection for filtered air

Connection for filtered air BeamWatch Non-contact, Focus Spot Size and Position monitor for high power YAG, Diode and Fiber lasers Instantly measure focus spot size Dynamically measure focal plane location during start-up From 1kW

More information

Screen investigations for low energetic electron beams at PITZ

Screen investigations for low energetic electron beams at PITZ 1 Screen investigations for low energetic electron beams at PITZ S. Rimjaem, J. Bähr, H.J. Grabosch, M. Groß Contents Review of PITZ setup Screens and beam profile monitors at PITZ Test results Summary

More information

UPU 5000 Inline Delamination Detection and Process Trending with the Ultrasonic Measuring System

UPU 5000 Inline Delamination Detection and Process Trending with the Ultrasonic Measuring System UPU 5000 Inline Delamination Detection and Process Trending with the Ultrasonic Measuring System Delamination Detection and Process Trending with the GreCon Ultrasonic Measuring System Exploit all reserves

More information

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis

More information

SC26 Magnetic Field Cancelling System

SC26 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC26 SC26 Magnetic Field Cancelling System Makes the ambient magnetic field OK for electron beam tools in 300 mm wafer fabs Real time, wideband cancelling from DC to > 9 khz fields

More information

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING R.H. Pawelletz, E. Eufrasio, Vallourec & Mannesmann do Brazil, Belo Horizonte, Brazil; B. M. Bisiaux,

More information

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator 20 Channel Digital Delay Generator Features of the 745T-20C: 20 Independent delay channels - 100 ps resolution - 25 ps rms jitter - 10 second range Output pulse up to 6 V/50 Ω Independent trigger for every

More information

M2-Measurement Report

M2-Measurement Report Fraunhofer Institute for Laser Technology ILT Steinbachstraße 15 52074 Aachen Tel. 0241 8906 0 www.ilt.fraunhofer.de Aachen, July 29 th 2016 201901 DIVERSE OE 131 Industriekleinprojekte Authors: Dr. rer.

More information

GFT Channel Digital Delay Generator

GFT Channel Digital Delay Generator Features 20 independent delay Channels 100 ps resolution 25 ps rms jitter 10 second range Output pulse up to 6 V/50 Ω Independent trigger for every channel Fours Triggers Three are repetitive from three

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

Lt DELTA USA, Inc

Lt DELTA USA, Inc Infrared LOOP SCANNER Rota-Sonde TS2006 Infrared - high sensitivity 480 F or 750 F Quick and easy commissioning Self-monitoring and alarm functions Lt 1037 1 Applications R o t a - S o n d e TS2 0 0 6

More information

4.9 BEAM BLANKING AND PULSING OPTIONS

4.9 BEAM BLANKING AND PULSING OPTIONS 4.9 BEAM BLANKING AND PULSING OPTIONS Beam Blanker BNC DESCRIPTION OF BLANKER CONTROLS Beam Blanker assembly Electron Gun Controls Blanker BNC: An input BNC on one of the 1⅓ CF flanges on the Flange Multiplexer

More information

Measurement of Microdisplays at NPL

Measurement of Microdisplays at NPL Conference on Microdisplays Measurement of Microdisplays at NPL Christine Wall, Dr Julie Taylor, Colin Campbell 14 th Sept 2001 Overview Displays measurement at NPL Why measure microdisplays? Measurement

More information

SHOWLINE SL NITRO 510 LED STROBE LUMINAIRE SPECIFICATIONS.

SHOWLINE SL NITRO 510 LED STROBE LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. SHOWLINE SL NITRO 510 LED STROBE LUMINAIRE SPECIFICATIONS. 1.) The luminaire shall be an LED strobe luminaire employing one-thousand, three-hundred and fifty (1350) white LED engines.

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE Instrument cathode-ray tube D14-363GY/123 INSTRUMENT CATHODE-RAY TUBE mono accelerator 14 cm diagonal rectangular flat face internal graticule low power quick heating cathode high brightness, long-life

More information

Elements of a Television System

Elements of a Television System 1 Elements of a Television System 1 Elements of a Television System The fundamental aim of a television system is to extend the sense of sight beyond its natural limits, along with the sound associated

More information

TV Synchronism Generation with PIC Microcontroller

TV Synchronism Generation with PIC Microcontroller TV Synchronism Generation with PIC Microcontroller With the widespread conversion of the TV transmission and coding standards, from the early analog (NTSC, PAL, SECAM) systems to the modern digital formats

More information

Ultrasonic Testing adapts to meet the needs of the Automotive Tube Industry

Ultrasonic Testing adapts to meet the needs of the Automotive Tube Industry Ultrasonic Testing adapts to meet the needs of the Automotive Tube Industry By Mark Palynchuk, Western Instruments Inc. Mill-Line Ultrasonic Testing (UT) has typically been limited to wall thicknesses

More information

4 MHz Lock-In Amplifier

4 MHz Lock-In Amplifier 4 MHz Lock-In Amplifier SR865A 4 MHz dual phase lock-in amplifier SR865A 4 MHz Lock-In Amplifier 1 mhz to 4 MHz frequency range Low-noise current and voltage inputs Touchscreen data display - large numeric

More information

A dedicated data acquisition system for ion velocity measurements of laser produced plasmas

A dedicated data acquisition system for ion velocity measurements of laser produced plasmas A dedicated data acquisition system for ion velocity measurements of laser produced plasmas N Sreedhar, S Nigam, Y B S R Prasad, V K Senecha & C P Navathe Laser Plasma Division, Centre for Advanced Technology,

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE INSTRUMENT CATHODE-RAY TUBE 14 cm diagonal rectangular flat face domed mesh post-deflection acceleration improved spot quality for character readout high precision by internal permanent magnetic correction

More information

SHOWLINE SL NITRO 510C LED STROBE LUMINAIRE SPECIFICATIONS.

SHOWLINE SL NITRO 510C LED STROBE LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. SHOWLINE SL NITRO 510C LED STROBE LUMINAIRE SPECIFICATIONS. The luminaire shall be an LED strobe luminaire employing five hundred and twenty eight (528) red, green, blue and white

More information

Application Note #63 Field Analyzers in EMC Radiated Immunity Testing

Application Note #63 Field Analyzers in EMC Radiated Immunity Testing Application Note #63 Field Analyzers in EMC Radiated Immunity Testing By Jason Galluppi, Supervisor Systems Control Software In radiated immunity testing, it is common practice to utilize a radio frequency

More information

In-process inspection: Inspector technology and concept

In-process inspection: Inspector technology and concept Inspector In-process inspection: Inspector technology and concept Need to inspect a part during production or the final result? The Inspector system provides a quick and efficient method to interface a

More information

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency CATHODE RAY OSCILLOSCOPE Basic block diagrams Principle of operation Measurement of voltage, current and frequency 103 INTRODUCTION: The cathode-ray oscilloscope (CRO) is a multipurpose display instrument

More information

SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS.

SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS. SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. 1.) The luminaire shall be a motorized colour mixing luminaire employing seven (7) homogenized red, green, blue, and white LED

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

The software concept. Try yourself and experience how your processes are significantly simplified. You need. weqube.

The software concept. Try yourself and experience how your processes are significantly simplified. You need. weqube. You need. weqube. weqube is the smart camera which combines numerous features on a powerful platform. Thanks to the intelligent, modular software concept weqube adjusts to your situation time and time

More information

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes: PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties Due Thursday, 2 Nov 2017 For this lab, you will explore the properties of the working HeNe laser. 1. Observation of higher-order modes: Realign

More information

The Cathode Ray Tube

The Cathode Ray Tube Lesson 2 The Cathode Ray Tube The Cathode Ray Oscilloscope Cathode Ray Oscilloscope Controls Uses of C.R.O. Electric Flux Electric Flux Through a Sphere Gauss s Law The Cathode Ray Tube Example 7 on an

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

CHAPTER 3 OSCILLOSCOPES AND SIGNAL GENERATOR

CHAPTER 3 OSCILLOSCOPES AND SIGNAL GENERATOR CHAPTER 3 OSCILLOSCOPES AND SIGNAL GENERATOR OSCILLOSCOPE 3.1 Introduction The cathode ray oscilloscope (CRO) provides a visual presentation of any waveform applied to the input terminal. The oscilloscope

More information

MCP Upgrade: Transmission Line and Pore Importance

MCP Upgrade: Transmission Line and Pore Importance MCP Upgrade: Transmission Line and Pore Importance Tyler Natoli For the PSEC Timing Project Advisor: Henry Frisch June 3, 2009 Abstract In order to take advantage of all of the benefits of Multi-Channel

More information

Product Brochure Version R&S RSC Step Attenuator Where precise signal levels count

Product Brochure Version R&S RSC Step Attenuator Where precise signal levels count Product Brochure Version 02.00 Step Attenuator Where precise signal levels count RSC_bro_en_5214-4413-12_v0200.indd 1 07.09.2018 10:36:40 Step Attenuator At a glance The is a switchable, mechanical step

More information

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL COMPOSITE VIDEO METER MODEL VLM- COMPOSITE VIDEO METER MODEL VLM- NTSC TECHNICAL INSTRUCTION MANUAL VLM- NTSC TECHNICAL INSTRUCTION MANUAL INTRODUCTION EASY-TO-USE VIDEO LEVEL METER... SIMULTANEOUS DISPLAY...

More information

FRONT-END AND READ-OUT ELECTRONICS FOR THE NUMEN FPD

FRONT-END AND READ-OUT ELECTRONICS FOR THE NUMEN FPD FRONT-END AND READ-OUT ELECTRONICS FOR THE NUMEN FPD D. LO PRESTI D. BONANNO, F. LONGHITANO, D. BONGIOVANNI, S. REITO INFN- SEZIONE DI CATANIA D. Lo Presti, NUMEN2015 LNS, 1-2 December 2015 1 OVERVIEW

More information

QUAD Camera Inputs with 8 display options (see pg 14-17) Handbook & Instructions. ...and more!

QUAD Camera Inputs with 8 display options (see pg 14-17) Handbook & Instructions. ...and more! QUAD040 Handbook & Instructions Useful compact quad unit that can be wall mounted or desk mounted. Multiple display options make it indispensable to the professional CCTV engineer. 4 Camera Inputs with

More information

PCM ENCODING PREPARATION... 2 PCM the PCM ENCODER module... 4

PCM ENCODING PREPARATION... 2 PCM the PCM ENCODER module... 4 PCM ENCODING PREPARATION... 2 PCM... 2 PCM encoding... 2 the PCM ENCODER module... 4 front panel features... 4 the TIMS PCM time frame... 5 pre-calculations... 5 EXPERIMENT... 5 patching up... 6 quantizing

More information

Light Emitting Diodes

Light Emitting Diodes By Kenneth A. Kuhn Jan. 10, 2001, rev. Feb. 3, 2008 Introduction This brief introduction and discussion of light emitting diode characteristics is adapted from a variety of manufacturer data sheets and

More information

Development of an Abort Gap Monitor for High-Energy Proton Rings *

Development of an Abort Gap Monitor for High-Energy Proton Rings * Development of an Abort Gap Monitor for High-Energy Proton Rings * J.-F. Beche, J. Byrd, S. De Santis, P. Denes, M. Placidi, W. Turner, M. Zolotorev Lawrence Berkeley National Laboratory, Berkeley, USA

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

SHOWLINE SL epar 180 LUMINAIRE SPECIFICATIONS.

SHOWLINE SL epar 180 LUMINAIRE SPECIFICATIONS. SHOWLINE SL epar 180 LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. 1.) The luminaire shall be a color mixing luminaire employing nineteen (19) homogenized red, green, blue, and white LED engines. The

More information

HV/PHA Adjustment (PB) Part

HV/PHA Adjustment (PB) Part HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i

More information

Scanning System S-2100

Scanning System S-2100 2D laser measurement system The fastest 2D laser measurement system in the world 119 m range Scan rate >1 million points/sec 360 vertical field of view System description The PENTAX Scanning System S-2100

More information

MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer

MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer Link Instruments Innovative Test & Measurement solutions since 1986 Store Support Oscilloscopes Logic Analyzers Pattern Generators Accessories MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer $ The

More information

OPTICAL POWER METER WITH SMART DETECTOR HEAD

OPTICAL POWER METER WITH SMART DETECTOR HEAD OPTICAL POWER METER WITH SMART DETECTOR HEAD Features Fast response (over 1000 readouts/s) Wavelengths: 440 to 900 nm for visible (VIS) and 800 to 1700 nm for infrared (IR) NIST traceable Built-in attenuator

More information

LD OEM/LD PDS/LD PeCo

LD OEM/LD PDS/LD PeCo LD OEM/LD PDS/LD PeCo Features LD OEM/PDS: 360 field of view LD OEM: the basic platform to LD PeCo: 90 field of view provide customized software 14,400 Hz scan rate programs on board and offers the Class

More information

Operating instructions Retro-reflective sensor. OJ50xx laser / / 2010

Operating instructions Retro-reflective sensor. OJ50xx laser / / 2010 Operating instructions Retro-reflective sensor OJ50xx laser 70481 / 00 05 / 010 Contents 1 Preliminary note3 1.1 Symbols used 3 Safety instructions 3 4 Installation 4 4.1 Installation of the supplied mounting

More information

SHOWLINE SL BAR 640 LINEAR WASH LUMINAIRE SPECIFICATIONS.

SHOWLINE SL BAR 640 LINEAR WASH LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. SHOWLINE SL BAR 640 LINEAR WASH LUMINAIRE SPECIFICATIONS. 1.) The luminaire shall be a color mixing luminaire employing twenty-four (24) red, green, blue, and white LED engines.

More information

Test of ScannerMAX Saturn 1 with 600Hz Sine-wave input, having an optical scan angle of 40 optical degrees peak to peak.

Test of ScannerMAX Saturn 1 with 600Hz Sine-wave input, having an optical scan angle of 40 optical degrees peak to peak. Test of ScannerMAX Saturn 1 with 600Hz Sine-wave input, having an optical scan angle of 40 optical degrees peak to peak. What follows are scope screen shots of a test of ScannerMAX Saturn 1B with our standard

More information

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. Multichannel Scaler Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. The ideal solution for: Time-resolved single-photon counting Phosphorescence lifetime spectrometry Atmospheric and

More information

IMPAC Infrared Thermometers

IMPAC Infrared Thermometers IMPAC Infrared Thermometers focusable optics for non-contact temperature measurements on metals, ceramics, graphite etc. between 300 and 3300 C IS 140 IGA 140 IS 140-PB IGA 140-PB Short response times

More information

-Technical Specifications-

-Technical Specifications- Annex I to Contract 108733 NL-Petten: the delivery, installation, warranty and maintenance of one (1) X-ray computed tomography system at the JRC-IET -Technical Specifications- INTRODUCTION In the 7th

More information

Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes

Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes RTO_app-bro_3607-2855-92_v0100.indd 1 Microvolt-level measurements with the R&S RTO Test & Measurement Application Brochure 01.00 Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes

More information

STX Stairs lighting controller.

STX Stairs lighting controller. Stairs lighting controller STX-1795 The STX-1795 controller serves for a dynamic control of the lighting of stairs. The lighting is switched on for consecutive steps, upwards or downwards, depending on

More information

Operating Instructions

Operating Instructions CNTX Contrast sensor Operating Instructions CAUTIONS AND WARNINGS SET-UP DISTANCE ADJUSTMENT: As a general rule, the sensor should be fixed at a 15 to 20 angle from directly perpendicular to the target

More information

Log-detector. Sweeper setup using oscilloscope as XY display

Log-detector. Sweeper setup using oscilloscope as XY display 2002/9/4 Version 1.2 XYdisp user manual. 1. Introduction. The XYdisp program is a tool for using an old DOS PC or laptop as XY display to show response curves measured by a sweeper log-detector combination.

More information

2 Types of films recommended for international exchange of television programmes

2 Types of films recommended for international exchange of television programmes Rec. ITU-R BR.265-8 1 RECOMMENDATION ITU-R BR.265-8* Rec. ITU-R BR.265-8 STANDARDS FOR THE INTERNATIONAL EXCHANGE OF PROGRAMMES ON FILM FOR TELEVISION USE (Question ITU-R 240/11) (1956-1959-1963-1966-1970-1974-1982-1986-1990-1992-1997)

More information

High Resolution Multicolor Contrast Scanner. Dimensioned drawing

High Resolution Multicolor Contrast Scanner. Dimensioned drawing Specifications and description KRTM 20 High Resolution Multicolor Contrast Scanner Dimensioned drawing en 01-2011/06 50116669 12mm 20mm 50mm 12-30 V DC 50 / 25 khz We reserve the right to make changes

More information

R&S RSC Step Attenuator Where precise signal levels count

R&S RSC Step Attenuator Where precise signal levels count Test & Measurement Product Brochure 01.00 Step Attenuator Where precise signal levels count Step Attenuator At a glance The is a switchable, mechanical step attenuator. It is available in various models

More information

HEAD. HEAD VISOR (Code 7500ff) Overview. Features. System for online localization of sound sources in real time

HEAD. HEAD VISOR (Code 7500ff) Overview. Features. System for online localization of sound sources in real time HEAD Ebertstraße 30a 52134 Herzogenrath Tel.: +49 2407 577-0 Fax: +49 2407 577-99 email: info@head-acoustics.de Web: www.head-acoustics.de Data Datenblatt Sheet HEAD VISOR (Code 7500ff) System for online

More information

onitoring Bearing Vibration with Seismic Transducers

onitoring Bearing Vibration with Seismic Transducers DEPARTMENTS onitoring Bearing Vibration M with Seismic Transducers Dr. Ryszard Nowicki Bently Nevada Asset Condition Monitoring Sales Application Engineer ryszard.nowicki@ge.com 7 8 O R B I T Vol.31 No.2

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

NONDESTRUCTIVE INSPECTION OF A COMPOSITE MATERIAL SAMPLE USING A LASER ULTRASONICS SYSTEM WITH A BEAM HOMOGENIZER

NONDESTRUCTIVE INSPECTION OF A COMPOSITE MATERIAL SAMPLE USING A LASER ULTRASONICS SYSTEM WITH A BEAM HOMOGENIZER NONDESTRUCTIVE INSPECTION OF A COMPOSITE MATERIAL SAMPLE USING A LASER ULTRASONICS SYSTEM WITH A BEAM HOMOGENIZER J. M. S. Sakamoto 1, 4, A. Baba 2, B. R. Tittmann 3, J. Mulry 3, M. Kropf, 3 and G. M.

More information

CATHODE-RAY OSCILLOSCOPE (CRO)

CATHODE-RAY OSCILLOSCOPE (CRO) CATHODE-RAY OSCILLOSCOPE (CRO) I N T R O D U C T I O N : The cathode-ray oscilloscope (CRO) is a multipurpose display instrument used for the observation, measurement, and analysis of waveforms by plotting

More information

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance The Linatron -M is a modular system. The control console, modulator, and RF unit are common to all model configurations. Only the X-ray head changes to match the application. The Linatron - M is designed

More information

Choosing an Oscilloscope

Choosing an Oscilloscope Choosing an Oscilloscope By Alan Lowne CEO Saelig Company (www.saelig.com) Post comments on this article at www.nutsvolts.com/ magazine/article/october2016_choosing-oscilloscopes. All sorts of questions

More information

COMPARED IMPROVEMENT BY TIME, SPACE AND FREQUENCY DATA PROCESSING OF THE PERFORMANCES OF IR CAMERAS. APPLICATION TO ELECTROMAGNETISM

COMPARED IMPROVEMENT BY TIME, SPACE AND FREQUENCY DATA PROCESSING OF THE PERFORMANCES OF IR CAMERAS. APPLICATION TO ELECTROMAGNETISM COMPARED IMPROVEMENT BY TIME, SPACE AND FREQUENCY DATA PROCESSING OF THE PERFORMANCES OF IR CAMERAS. APPLICATION TO ELECTROMAGNETISM P. Levesque 1, P.Brémond 2, J.-L. Lasserre 3, A. Paupert 2, D. L. Balageas

More information

CRT Dynamics. A report on the dynamical properties of CRT based visual displays

CRT Dynamics. A report on the dynamical properties of CRT based visual displays CRT Dynamics A report on the dynamical properties of CRT based visual displays Display-Metrology & Systems 2007 Introduction In 2004 more LCD-monitors have been sold as computer monitors than CRT-based

More information

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel)

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel) Digital Delay / Pulse Generator Digital delay and pulse generator (4-channel) Digital Delay/Pulse Generator Four independent delay channels Two fully defined pulse channels 5 ps delay resolution 50 ps

More information

Label Applicator HERMA 400

Label Applicator HERMA 400 Label Applicator HERMA 400 Info GB V6.0 / 30.3.07 [ Facts & Data ] R Constructional variant Right-hand or left-hand version, always flexible to use L Power supply / Line voltage Max. power consumption

More information

Y.XST225-VF. INTERCONTROLE Escoffier 1 XYLON MG225VF RX PDS

Y.XST225-VF. INTERCONTROLE Escoffier 1 XYLON MG225VF RX PDS Y.XST225-VF 1 XYLON MG225VF - 0509- RX0905007PDS 1 Table of contents Y.XST225-VF High detail visibility - examples Y.XST225-VF vs. conventional X-ray systems Y.XST225-VF vs. µ-focus X-ray systems Unique

More information

45 mm : all channels. Output indicator Yellow LED System status indicator Light source Infrared (880 nm) Opening angle +/ 4 Emission angle +/ 5

45 mm : all channels. Output indicator Yellow LED System status indicator Light source Infrared (880 nm) Opening angle +/ 4 Emission angle +/ 5 SPACEGUARD SERIES Description 1-12 metre sensing range 16 to 56 parallel scanning beams Active height of 0 mm to 2520 mm Detector length of 1928 mm to 2648 mm 3 different channel placement models Cable

More information

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Page 1 of 5 8500A The Giga-tronics 8500A Series Peak Power Meters THE ACCURACY STANDARD channel capability. For tests

More information

PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND. Doug Roberts U of Maryland, College Park

PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND. Doug Roberts U of Maryland, College Park PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND Doug Roberts U of Maryland, College Park Overview We have developed a system for measuring and scanning phototubes for the FDIRC Based primarily on

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

Electrical connection

Electrical connection Color sensors Dimensioned drawing en 02-2013/01 50121262 068-14515 12mm 32mm 10-30 V DC 500 Hz Scanner for color detection Simultaneous selection of up to 3 colors Detection independent of distance Teach-in

More information

DPD80 Infrared Datasheet

DPD80 Infrared Datasheet Data Sheet v1.4 DPD8 Infrared DPD8 Infrared Datasheet Resolved Inc. www.resolvedinstruments.com info@resolvedinstruments.com 217 Resolved Inc. All rights reserved. DPD8 Infrared General Description The

More information

Specifications. Mechanical Information. Mass (grams) Dimensions (mm) 15 x 75 Housing. Anodised Aluminium Isolated Body

Specifications. Mechanical Information. Mass (grams) Dimensions (mm) 15 x 75 Housing. Anodised Aluminium Isolated Body Beta TX Datasheet Beta-TX The Beta-TX is a complete self contained laser diode system which can operate in both CW and modulation modes. The Beta- TX features high speed modulation with a bandwidth of

More information

IS 140 IGA 140 IS 140-PB IGA 140-PB IS 140-PN IGA 140-PN IS 140-ET IGA 140-ET

IS 140 IGA 140 IS 140-PB IGA 140-PB IS 140-PN IGA 140-PN IS 140-ET IGA 140-ET IMPAC Infrared Temperature Sensors focusable optics for non-contact temperature measurements on metals, ceramics, graphite etc. between 220 and 3500 C IS 140 IGA 140 IS 140-PB IGA 140-PB IS 140-PN IGA

More information

PITZ Introduction to the Video System

PITZ Introduction to the Video System PITZ Introduction to the Video System Stefan Weiße DESY Zeuthen June 10, 2003 Agenda 1. Introduction to PITZ 2. Why a video system? 3. Schematic structure 4. Client/Server architecture 5. Hardware 6. Software

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

Precision testing methods of Event Timer A032-ET

Precision testing methods of Event Timer A032-ET Precision testing methods of Event Timer A032-ET Event Timer A032-ET provides extreme precision. Therefore exact determination of its characteristics in commonly accepted way is impossible or, at least,

More information

GFT Channel Slave Generator

GFT Channel Slave Generator GFT1018 8 Channel Slave Generator Features 8 independent delay channels 1 ps time resolution < 100 ps rms jitter for optical triggered delays 1 second range Electrical or optical output Three trigger modes

More information

4. ANALOG TV SIGNALS MEASUREMENT

4. ANALOG TV SIGNALS MEASUREMENT Goals of measurement 4. ANALOG TV SIGNALS MEASUREMENT 1) Measure the amplitudes of spectral components in the spectrum of frequency modulated signal of Δf = 50 khz and f mod = 10 khz (relatively to unmodulated

More information

Design Studies For The LCLS 120 Hz RF Gun Injector

Design Studies For The LCLS 120 Hz RF Gun Injector BNL-67922 Informal Report LCLS-TN-01-3 Design Studies For The LCLS 120 Hz RF Gun Injector X.J. Wang, M. Babzien, I. Ben-Zvi, X.Y. Chang, S. Pjerov, and M. Woodle National Synchrotron Light Source Brookhaven

More information

Contents. 1. System Description 3. Overview 3 Part Names 3 Operating Conditions 7 Start-up Procedure 7. 2.

Contents. 1. System Description 3. Overview 3 Part Names 3 Operating Conditions 7 Start-up Procedure 7. 2. Rigel 1550 Terahertz Spectrometer User Manual Contents info@tetechs.com 1. System Description 3 Overview 3 Part Names 3 Operating Conditions 7 Start-up Procedure 7 2. Safety 9 Laser Safety 9 Electrical

More information

SHOWLINE SL PUNCHLITE 220 LED LUMINAIRE SPECIFICATIONS.

SHOWLINE SL PUNCHLITE 220 LED LUMINAIRE SPECIFICATIONS. SHOWLINE SL PUNCHLITE 220 LED LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. 1.) The luminaire shall be a color mixing luminaire employing nine (19) homogenized red, green, blue, and white LED engines.

More information

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT.

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Preface The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Please note: Some components described in this manual may be optional.

More information

ECP1 MC / MM PHONO PREAMPLIFIER OWNERS MANUAL

ECP1 MC / MM PHONO PREAMPLIFIER OWNERS MANUAL ECP1 MC / MM PHONO PREAMPLIFIER OWNERS MANUAL WARNING: To reduce risk of fire or electric shock, do not expose this appliance to rain or moisture. Do not remove cover. No user serviceable parts inside.

More information

ISOMET. Compensation look-up-table (LUT) and Scan Uniformity

ISOMET. Compensation look-up-table (LUT) and Scan Uniformity Compensation look-up-table (LUT) and Scan Uniformity The compensation look-up-table (LUT) contains both phase and amplitude data. This is automatically applied to the Image data to maximize diffraction

More information

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Please read the user manual before operation. PAGE 1 Content 1. Company Profile... 3 2. Product Introduction... 3 3. Operation Instruction...

More information