STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

Size: px
Start display at page:

Download "STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM"

Transcription

1 STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure sample conductivity and magnetic property in nano-scale region and image samples in liquid environment. Location: White Hall Room B-20B Primary Staff Contact: Dr. Weiqiang Ding Office: ESB G75D (cell) The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. WARNING: Directly looking into the laser beam can cause permanent eye damage or blindness. Always use a piece of paper if you want to view the beam image. NOTE: The purpose of this manual is for general measurements with the AFM. For more advanced measurements or measurements with accessories, please contact MFCF staff members or refer to the Asylum MPF-3D operation manual, which is located on the computer workbench. START UP 1. Log in your session on the FOM. Sign in on the logbook located on the corner bench. 2. Check the AFM controller box (Figure 1) at the bottom of the computer cart to ensure that it is powered up (the green LED above the Power switch is ON). 3. Open the right side door of the AFM enclosure box. Power up the Herzan TS-150 isolation table under the AFM. Press E button to enable the vibration isolation. Figure 1. AFM controller front panel [1]. 1

2 4. Open the MFP-3D software by clicking on the Stable Asylum Research MFP3D icon on the computer desktop. Wait until MFP-3D Ready is shown at the bottom of the MFP-3D software tray (Figure 2). 5. In the Master Panel, click on the Main tab and check the Imaging Mode at lower section of the tab. Select AC Mode (tapping mode) if it is not already shown. Contact a SRF staff if you need help on other operation modes. Figure 2. MFP-3D software tray PROBE MOUNTING 1. Find the tip holder assembly on the loading pedestal (Figure 3a,b) and go to step 4. Otherwise, go to next step. (a) (b) (c) Figure 3. (a) AFM tip holder assembly; (b) holder assembly in the loading pedestal; (c) Removing tip holder assembly from AFM head [1]. 2. If you could not find holder assembly on the loading pedestal, it should be attached to the AFM head. Remove the AFM head from the sample stage and lay it upside down next to the sample stage (Do not stretch the cable attached to the AFM head). 3. Remove the tip holder assembly from the AFM head by pressing the round button on the side (Figure 3c) and mount it into the loading pedestal apparatus (Figure 4A). 4. Loosen the clamp screw (the middle one, Figure 4B) with a screwdriver. 5. Check the AFM probe against the light to ensure that the cantilever is not damaged. 2

3 6. Slide the AFM probe under the tongue clamp (Figure 4B). You may need to gently lift the tongue clamp with a pair of sharp tweezers to accommodate thicker probe chips. 7. Position the probe to the center of the clear trapezoidal shaped quartz optical window (Figure 4C). 8. Tighten the clamp screw with the screwdriver until finger tight. Do NOT over tighten the screw (Figure 4E). 9. Remove the tip holder assembly from the loading apparatus and put it back into the MFP-3D head (Figure 3c). Figure 4. Proper probe loading procedure [2]. LASER ALIGNMENT 1. Flip the AFM head to its normal position and settle it on the table surface. NOTE: do not stretch or twist the cable. 2. Turn on the laser by turning the key on the front of the AFM controller 90 o clockwise to ON position (Figure 1). 2. Place the Newport infrared sensor card under the tip holder. You should see a laser spot on the card if the laser is not being blocked by anything. 3. If you can t see the laser spot, turn the thumbwheel on the right side of the AF head (labeled LDX, Figure 5) counter clockwise to move the laser spot outward along the cantilever length until the 3

4 spot appears. If no spot can be seen on the IR card, charge the IR card for a few seconds under the fluorescent light. Figure 5. AFM head with various alignment thumbwheels labeled [2]. 4. Turn the laser adjustment thumbwheel on the right side (labeled LDX, Figure 5) clockwise until the laser spot disappears (moves back onto the AFM chip) (Figure 6 step #1). Note that the laser spot will also not be visible on the IR card if it is outside the transparent quartz window. 5. Turn the LDX thumbwheel counter clockwise until the laser spot re-appear (Figure 6 step #2). 6. Turn the laser adjustment thumbwheel on the back side of the AFM head (labeled LDY, Figure 6) (Figure 6 step #3) while monitoring the laser spot on the IR card. The laser spot intensity will be temporarily reduced when it crosses the cantilever beam and the SUM reading on the Signal and Deflection Meter (Figure 7) will increase significantly. Reverse the direction if the laser spot completely disappears (goes beyond the transparent window). NOTE: Keep turning the thumbwheel forward beyond the limit will damage the LDY thumbwheel. 7. Maximize the SUM reading by adjusting both LDX and LDY thumbwheels (Figure 6 step #4). Figure 6. Laser alignment pattern [2]. 4

5 PHOTODECTOR (PD) ALIGNMENT 1. Check the Deflection voltage signal in the Signal and Deflection (S&D) meter (Figure 7). It may show Red (positive) or Blue (negative) if the voltage is not zero. 2. If the Deflection value is negative (blue), turn the PD adjustment thumbwheel (Figure 5) counter clock-wise (CCW) to get to around 0 V. If the Deflection value is red (positive), turn the PD thumbwheel clockwise (CW) to get around 0 V. Figure 7. Signal and Deflection meter [2]. SAMPLE MOUNTING 1. Attach your sample to a piece of glass slide with sticky tab, super glue, epoxy or vacuum grease. Contact MFCF staff if you need to measure in fluids. 2. Mount the sample slide at the center of the sample stage. 3. Secure the sample slide to the stage with the two small magnet blocks on both sides (Figure 8). Figure 8. Sample stage and sample positioning [1]. AFM HEAD MOUNTING 1. Lift the AFM head with two hands. Place the two back legs into the divots on the AFM baseplates first (Figure 9A). 5

6 Figure 9. Mount the AFM head onto the base plate [2]. 2. Move thumbs under the front of the AFM head (Figure 9B). Slowly lower the AFM head down towards the stage while continually monitoring the tip-sample separation to avoid tip crash (Figure 9B). 3. If the tip appears to crash to sample, lift the head back and adjust the front leg height to increase the tip-sample separation. Repeat the process. 4. Adjust sample position with respect to the cantilever tip with the adjustment knobs on the right side of the base plate (Figure 8). 5. Power up the Herzan active vibration isolation system under the AFM stage. ALIGNMENT CHECK WITH VIDEO CAMERA 1. Turn on the light source for the video camera. The black switch knob is located on the right corner of the AFM computer cart. 2. Load the video camera window by clicking the camera icon at the bottom of the AFM software window. Select the S-video input in the Source pull-down menu in the video window. 3. Adjust the image brightness with the light source switch knob. 4. Adjust the two camera translation knobs at the back of the AFM head (Figure 5) to position the AFM cantilever probe in the middle of the video screen. 5. Focus the camera on the cantilever probe with the camera focus ring (Figure 5) at the back of the AFM head. Turn the ring clockwise to focus down and counter-clockwise to focus up. 6. Check and ensure that the laser spot is positioned on the cantilever beam. If not, adjust the LDX, LDY laser positioning thumbwheels to position the laser spot onto the cantilever beam and maximize the SUM reading. 7. Adjust the focus ring clockwise to focus on the sample surface. 8. Find the area of interest on sample surface by adjusting the scanner adjustment knobs on the right side of the sample stage (Figure 8). 9. Position the area of interest under the AFM probe. 6

7 FRQUENCY TUNING IN AC (TAPPING) MODE 1. In software program Master Panel, select the Tune tab (Figure 10). 2. Set Target Amplitude to 1.0 V. 3. Set Target Percent to 5.0 %. 4. The default auto tune frequency range (50 khz to 400 khz) covers most drive frequencies of common cantilevers. Adjust the frequency range if the expected cantilever resonance frequency is out of the range. 5. Click on Auto Tune button to start the frequency sweep. 6. A frequency/phase response curve will show up if auto tune finishes successfully (Figure 11a). The Drive Frequency will be automatically updated in the Master Panel. 7. Check the S&D meter reading (Figure 11b). The Amplitude value should be close to the set Target Amplitude of 1.0 V. (a) Figure 10. Auto tune tab [2]. (b) Figure 11. (a) Cantilever frequency auto tune result. (b) S&D meter after auto tune (before tip engagement) [2]. 8. If you encounter Auto Tune failure, the laser may not be properly aligned. Repeat the laser alignment process. 7

8 PROBE ENGAGEMENT There are two ways to engage the tip to sample surface: Hard Engage and Soft Engage. The hard engage is quick but may result in minor tip wear. In soft engage, the tip engages at the lowest force possible and is suite able for expensive, super sharp tip or functionalized tip. Figure 12. A qualitative conceptual amplitude voltage vs. distance for AC mode imaging [2]. HARD ENGAGEMENT 1. Click the Main tab in Master Panel (Figure 13). 2. Select a Set Point voltage that is around 20% off the free air amplitude voltage. (E.g., if the free air amplitude is 1.0 V, the Set Point voltage should be 0.8 V). 3. Choose an Integral Gain of 8 to Choose a Scan Rate of 1 Hz. 5. Click the Engage button on the S&D meter (Figure 11b). Monitor the Z Voltage reading, which should be 150 V. Figure 13. Main tab in Master 6. Slowly turn the thumbwheel on the front of the AFM head Control for scan parameters [2]. counterclockwise to lower the tip towards sample surface. When the Amplitude Voltage decrease to the selected Set Point voltage, there is a concomitant chime sound (if the speaker is on). 7. Further turn the thumbwheel and monitor the Z Voltage, which should decrease. Stop when the Z Voltage drops to 70V. 8. The AFM tip is now stationary on the surface (not scanning). SOFT ENGAGEMENT 1. Select a Set Point voltage that is around 5% off the free air amplitude voltage. (E.g., if the free air amplitude is 1.0V, the Set Point voltage should be 0.95 V). 8

9 2. Click Engage button on the S&D meter (Figure 11b). 3. Slowly turn the thumbwheel on the front of the AFM counterclockwise to lower the tip towards sample surface while monitoring the amplitude voltage on S&D meter. Stop when the amplitude voltage equals the Set Point voltage. 4. Further turn the thumbwheel to set the Z Voltage to 30-50V (halfway blue). 5. Activate the radio label for the Set Point voltage on the Main tab (See Figure 13). 6. Slowly turn the big Hamster wheel on the front of the controller (Figure 1) to increase the force (i.e., decrease the Set Point value). The Z voltage will move to more positive voltage value (more red). Continue turning the wheel until you notice that Z voltage will no longer move to more positive value. At this point, the tip is engaged on the surface at the lowest possible Set Point force. 7. The AFM tip is now stationary on the surface (not scanning). SCANNING 1. Click the Do Scan button, or Frame up or Frame Down button on the Main tab of the Mater Panel to start scanning (Figure 13). 2. Set proper channel for image observation: Ht (Height) Am (Amplitude), Ph (Phase), ZS (Z sensor). 3. Check the Trace and Retrace fast scan lines below each image channel to see if they follow each other well. 4. If you use the Soft Engage, decrease the Set Point voltage value to increase the force applied to the sample. NOTE: Do not set the Set Point voltage lower than 50% of the free air amplitude voltage. 5. Slightly increase the Drive Amplitude in the Main tab and monitor the improvement. 6. Adjust the Integral Gain in the Main tab to better track the surface. Increase the Integral Gain until high frequency noise shows up in the trace/retrace lines. Decrease the Integral Gain until the noise disappears. 7. Select a proper Scan Rate for image acquisition. After adjusting the Scan Rate, click Frame Up or Frame Down buttons to initiate the newly entered scan rate. SAVING IMAGES 1. Select proper Scan Points and Scan Lines in the Main tab of the Master Panel. Use 128x128 or 256x256 for rough scan, 512x512 for data acquisition. 2. Adjust Image Z scale: Click Auto check box in top left corner of the image channel tabs. Click the Fix button on the left of the Data Scale section. 3. Select Real Time Flattening by choose the Line or Masked Line in the Live Flatten pull-down menu. 9

10 4. Give a proper filename for you image in the Base Name section at the bottom of the Master Panel. 5. Click the Save Images checkbox. Use Browse to choose an image folder. 6. Check the Save Status section to see if it will save current scan. 7. If you want to save a screen shot of one of the windows, you can make it the front window, and go to menu File -> save Graphics Select the desirable format and size. IMAGE ANALYSIS 1. In top menu, select AFM Analysis > Browse Saved Data to locate the saved images. 2. Open the chosen image file by placing the cursor over the image thumbnail and double clicking the mouse button. The image will open in the Display Window. Flattening/PlaneFit 1. Click the M button on the upper right corner of the image display window to open the Modify Panel. 2. Select the Flatten Tab (Figure 14). Set a Flatten Order to 0 in the pull-down menu, and select the All radio button underneath. 3. Click the Flatten button and the image will be flattened. 4. Select the Planefit Tab (Figure 14). 5. Set the Planefit Order to 1 in the pull-down menu. Select the All radio button. 6. Click the XY button to planefit the image in the X-Y plane with respect to Z. Figure 14. (left) Flatten tab and (right) Planefit tab [2]. 10

11 Roughness 1. Click the A button on the top right corner of the Display window to open the Analyze Panel. 2. Open the Roughness tab on the Analyze Panel (Figure 15). 3. The data present in the Roughness tab is the most recent image. 4. Click the Export Roughness button at the bottom to create a text file of the data. 5. Refer to Section in Operation manual for roughness measurement on masked image. Section Analysis 1. Click the A button in the upper right corner of the Display window to open the Analyze Panel. 2. Open the Section tab in Analyze Panel. 3. Select Line from the Mode pull-down menu. 4. Click the Draw button. 5. Move cursor over to image and draw a line on the feature of interest. A section plot will appear. 6. If needed, place cursors on the curve to measure points on the line. Figure 15. Roughness tab. 7. If the curve is not leveled, perform a 1 st order XY plane fit. Refer to Section B in operation manual for leveling the line sections. 8. Click the Export Roughness button at the bottom to create a text file of the data. For other image analysis procedures, please refer to the Chapter 7 in the Procedural Operation Manualette located on the corner workbench. SHUT DOWN 1. Click the Stop!!! button in the Master Panel to withdraw the tip. The S&D meter Z Voltage will read 0.0 V. 2. Turn the front thumbwheel clockwise to retract the tip from sample surface. Turning the thumbwheel in the wrong direction will cause tip and sample damage. 3. Turn off the power of the Herzan TS-150 active vibration isolation system. 4. Turn off the laser by switching the key on the AFM controller counter clockwise to OFF position. 5. Turn off camera fiber optics light (switch on the AFM computer table). 6. Remove the AFM head from sample stage. Set it upside down on the table. NOTE: Do NOT stretch the cable. 11

12 7. Remove the tip holder assembly from the AFM head and place it into the loading apparatus. 8. Loosen the middle screw and remove AFM tip from the holder. 9. Tighten the middle screw until finger tight. Leave the holder assembly on the loading apparatus. 10. Remove sample from the sample stage. 11. Set the AFM head (without tip holder assembly) back onto the sample stage. 12. Sign out on the FOM and sign out on the log book. Report any problem in the log off window on the FOM and in the note section of the log book. 13. Clean up the work area before leaving the room. REFERENCES 1. MFP-3D TM Atomic Force Microscope Installation and Operation Manual, Asylum Research Inc., Dr. Ryan Fuierer, Procedural Operation Manualette, Beta Version 10, September

13 EMERGENCY PROCEDURES If a user ever has a problem or an uncertainty should ASK someone who knows and can help. There are no penalties for asking for help but there may be for not reporting damage to the equipment that may delay or prevent others from working. If, at any time, the user needs to contact someone for help, call or locate the following staff of the Materials Fabrication and Characterization Facility (MFCF): Weiqiang Ding Office: ESB G75D Phone: (304) cell Harley Hart Office: White Hall 409 Phone: (412) cell If no one is available and the AFM is not acting as expected, the user should do the following: Shut down the AFM controller Shut down the computer Then, if possible, the user should stay with the AFM while trying to contact the above individuals. If it becomes necessary to leave the instrument then the user should leave a large, legible note on both the AFM and at least one of the above individuals offices, stating: The problem, describing what happened and steps taken When it occurred date and time User name and phone number If a dangerous situation is evident (smoke, fire, sparks, etc.), the user should press the power button on the power strip beside the AFM optical table or unplug the power strip to turn OFF power to the entire AFM system and notify the proper emergency personnel. If turning off the power would be unsafe in the user s estimation, the user should leave the facility and contact emergency personnel immediately. 13

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 1.5, November 2013 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 2, September 2017 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core 1 Turn On the laser power 2 Open enclosure: - lift the door latch and open the enclosure door. 3 2 1 1.Unlock scanner: Lift the lever to the

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

Standard AFM Modes User s Manual

Standard AFM Modes User s Manual Standard AFM Modes User s Manual Part #00-0018-01 Issued March 2014 2014 by Anasys Instruments Inc, 325 Chapala St, Santa Barbara, CA 93101 Page 1 of 29 Table of contents Chapter 1. AFM Theory 3 1.1 Detection

More information

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment Name: Atomic Force Microscope Badger name: afm Revisionist Paul Kimani Model: Dimension 3000 Date: October 1, 2013 Location: Bay 1 A. Description The Dimension 3000 consists of a rigid stage

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

AFM Standard Operating Procedure

AFM Standard Operating Procedure 2013 AFM Standard Operating Procedure Karen Gaskell, David Ramsdell Surface Analysis Centre Department of Chemistry and Biochemistry University of Maryland 1/1/2013 Content Page 1 Hardware 2 1.1 MultiMode

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

Magnetic Force Microscope (MFM)

Magnetic Force Microscope (MFM) Magnetic Force Microscope (MFM) [Digital Instruments NanoScopeIII] magnetic and topographic resolution ~20 nm [at least several hours/sample (1 cm 2 polished single grains, thin sections, or thin films)]

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

K Service Source. Apple High-Res Monochrome Monitor

K Service Source. Apple High-Res Monochrome Monitor K Service Source Apple High-Res Monochrome Monitor K Service Source Specifications Apple High-Resolution Monochrome Monitor Specifications Characteristics - 1 Characteristics Picture Tube 12-in. diagonal

More information

K Service Source. Apple High-Res Monochrome Monitor

K Service Source. Apple High-Res Monochrome Monitor K Service Source Apple High-Res Monochrome Monitor K Service Source Specifications Apple High-Resolution Monochrome Monitor Specifications Characteristics - 1 Characteristics Picture Tube 12-in. diagonal

More information

Do not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company.

Do not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company. PLEASE, READ Notice: Copyright: A delivery set of your device may differ from the set described in the given manual. For more detailed information refer to the specification of your contract. No part of

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence

More information

HDCVI Camera User s Manual

HDCVI Camera User s Manual HDCVI Camera User s Manual Version 1.0.3 Table of Contents 1 General Introduction... 1 1.1 Overview... 1 1.2 Features... 1 2 Device Framework... 2 3 Installation... 5 Model A... 5 Model B1 and B2... 6

More information

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.1 Click on the OMNIC icon to open the software. 1.2 Check that a signal is being measured by opening the Experiment Setup menu under the Collect

More information

B-AFM. v East 33rd St., Signal Hill, CA (888)

B-AFM. v East 33rd St., Signal Hill, CA (888) B-AFM The B-AFM is a basic AFM that provides routine scanning. Ideal for scientists and educators, the B-AFM is capable of creating high-resolution topography images of nanostructures in standard scanning

More information

Inspire Station. Programming Guide. Software Version 3.0. Rev A

Inspire Station. Programming Guide. Software Version 3.0. Rev A Inspire Station Programming Guide Software Version 3.0 Rev A Copyright 2016 Electronic Theatre Controls, Inc. All rights reserved. Product information and specifications subject to change. Part Number:

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Toshiba TLP-B2 Training Module Introduction

Toshiba TLP-B2 Training Module Introduction Toshiba TLP-B2 Training Module Introduction Toshiba TLP-B2 Introduction Welcome! Welcome to the Toshiba TLP-B2 training module offered by UT Career Services! It is highly recommended that all faculty &

More information

Figure 1: AFM image of a Tip-check sample

Figure 1: AFM image of a Tip-check sample Atomic Force Microscopy Atomic force microscopy is a microscope technique that involves viewing samples with a resolution of under a fraction of a nanometer. The applications for Atomic Force Microscopy

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision

More information

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products

More information

HDCVI Camera User s Manual

HDCVI Camera User s Manual HDCVI Camera User s Manual Version 1.0.2 Table of Contents 1 General Introduction... 1 1.1 Overview... 1 1.2 Features... 1 2 Device Framework... 2 3 Installation... 5 3.1 Installation for Model A... 5

More information

9070 Smart Vibration Meter Instruction Manual

9070 Smart Vibration Meter Instruction Manual 9070 Smart Vibration Meter Instruction Manual Overall machine and bearing conditions: vibration values are displayed with color coded alarm levels for ISO values and Bearing Damage (BDU). Easy vibration

More information

Experiment 9A: Magnetism/The Oscilloscope

Experiment 9A: Magnetism/The Oscilloscope Experiment 9A: Magnetism/The Oscilloscope (This lab s "write up" is integrated into the answer sheet. You don't need to attach a separate one.) Part I: Magnetism and Coils A. Obtain a neodymium magnet

More information

ELECTRICAL ADJUSTMENT INSTRUCTIONS

ELECTRICAL ADJUSTMENT INSTRUCTIONS ELECTRICAL ADJUSTMENT INSTRUCTIONS General Note: "CBA" is abbreviation for "Circuit Board Assembly." NOTE: Electrical adjustments are required after replacing circuit components and certain mechanical

More information

Checking the fluids of the cart

Checking the fluids of the cart 1 Preparing FACSAria for sterile sorting Prepare and autoclave these fluids at least 2 days in advance: Fluids Remarks Prepare 5L (for 1 period) or 10L (for 2 Each stainless-steel-sheath-tank has a periods)

More information

PLL1920M LED LCD Monitor

PLL1920M LED LCD Monitor PLL1920M LED LCD Monitor USER'S GUIDE www.planar.com Content Operation Instructions...1 Safety Precautions...2 First Setup...3 Front View of the Product...4 Rear View of the Product...5 Installation...6

More information

Data Acquisition Using LabVIEW

Data Acquisition Using LabVIEW Experiment-0 Data Acquisition Using LabVIEW Introduction The objectives of this experiment are to become acquainted with using computer-conrolled instrumentation for data acquisition. LabVIEW, a program

More information

Approved by: / / R. Battaglia 12/16/2016

Approved by: / / R. Battaglia 12/16/2016 Fabrication Laboratory Revision: H Rev Date: 12/16/16 Approved by: Process Engineer / / R. Battaglia 12/16/2016 Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Varian

More information

Veeco Dektak 6M Profilometer

Veeco Dektak 6M Profilometer Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO) 2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency

More information

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents: Compensation look-up-table (LUT) and How to Generate Contents: Description Background theory Basic LUT pg 2 Creating a LUT pg 3 Using the LUT pg 7 Comment pg 9 The compensation look-up-table (LUT) contains

More information

VectorVGA Tempest User Manual

VectorVGA Tempest User Manual VectorVGA Tempest User Manual 2 Notice Regarding This Product WARNING! To install this product you should: Be familiar with safe handling procedures for electronic components. Be able to use hand tools

More information

Mini Micro Pulse Lidar System

Mini Micro Pulse Lidar System Mini Micro Pulse Lidar System MiniMPL-532-C Sensor Suite Operations Manual Version: June 2016 THIS PAGE INTENTIONALLY LEFT BLANK 2 Table of Contents MINIMPL SENSOR SUITE SYSTEM: RECORD OF PURCHASE... 4

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Mitsubishi WD57734 DLP Chip Replacement

Mitsubishi WD57734 DLP Chip Replacement Mitsubishi WD57734 DLP Chip Replacement Replace the DLP chip in your Mitsubishi WD57734. Written By: David Sylvester ifixit CC BY-NC-SA www.ifixit.com Page 1 of 10 INTRODUCTION This guide will detail the

More information

Practicum 3, Fall 2010

Practicum 3, Fall 2010 A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

ipad Air 2 Wi-Fi Display Assembly Replacement

ipad Air 2 Wi-Fi Display Assembly Replacement ipad Air 2 Wi-Fi Display Assembly Replacement Fix a cracked or faulty screen by replacing the display assembly in an ipad Air 2 Wi-Fi. Written By: Evan Noronha ifixit CC BY-NC-SA www.ifixit.com Page 1

More information

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Last updated: October 12, 2011 by Josh Kurutz THIS PAGE = QUICK START GUIDE 0) At the computer, make sure VNMRJ is

More information

HDCVI Eyeball Camera Lite

HDCVI Eyeball Camera Lite HDCVI Eyeball Camera Lite User s Manual V1.0.0 Foreword General This user s manual (hereinafter referred to be "the Manual") introduces the functions, installation, and operations of the camera. Safety

More information

Using AFM Phase Lag Data to Identify Microconstituents with Varying Values of Elastic Modulus

Using AFM Phase Lag Data to Identify Microconstituents with Varying Values of Elastic Modulus Using Data to Identify Microconstituents with Varying Values of Elastic Modulus D.N. Leonard*, A.D. Batchelor**, P.E. Russell*,** *Dept. of Material Science and Engineering, Box 7531, North Carolina State

More information

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished

More information

ipad Air 2 Wi-Fi Display Assembly Replacement

ipad Air 2 Wi-Fi Display Assembly Replacement ipad Air 2 Wi-Fi Display Assembly Replacement Fix a cracked or faulty screen by replacing the display assembly in an ipad Air 2 Wi-Fi. Geschreven door: Evan Noronha ifixit CC BY-NC-SA nl.ifixit.com Pagina

More information

HDCVI Camera User s Manual

HDCVI Camera User s Manual HDCVI Camera User s Manual Version 1.0.0 Table of Contents 1 General Introduction... 1 1.1 Overview... 1 1.2 Features... 1 2 Device Framework... 2 3 Device Cable Structure... 4 4 Device Installation...

More information

imso-104 Manual Revised August 5, 2011

imso-104 Manual Revised August 5, 2011 imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel

More information

Analysis and sorting of cells with FACSAria II flow cytometer Tiina Pessa-Morikawa / Revised

Analysis and sorting of cells with FACSAria II flow cytometer Tiina Pessa-Morikawa / Revised Analysis and sorting of cells with FACSAria II flow cytometer 10.6.2010 Tiina Pessa-Morikawa / Revised 16-03-17 This instruction refers to the operating manual (FAUG, i.e. FACSAria II Users Guide and page

More information

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors. Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only

More information

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis

More information

nanoir2-s TM System User s Manual Part # Issued August 2016 Page 1 of 52

nanoir2-s TM System User s Manual Part # Issued August 2016 Page 1 of 52 nanoir2-s TM System User s Manual Part #00-0026-03 Issued August 2016 2016 by Anasys Instruments Inc, 325 Chapala St., Santa Barbara, CA 93101 Page 1 of 52 Table of contents Anasys Instruments Contact

More information

USER MANUAL. 28" 4K Ultra HD Monitor L28TN4K

USER MANUAL. 28 4K Ultra HD Monitor L28TN4K USER MANUAL 28" 4K Ultra HD Monitor L28TN4K TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety Info &

More information

RIPPLE TANK - projection, with strobe & kit

RIPPLE TANK - projection, with strobe & kit RIPPLE TANK - projection, with strobe & kit DESCRIPTION: Cat: SW3440-001 projection type with stroboscope, rippler & kit. The images created by IEC Ripple Tank Projection type are viewed from the side

More information

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT.

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Preface The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Please note: Some components described in this manual may be optional.

More information

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303)

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303) Spectroscopy Module Vescent Photonics, Inc. www.vescentphotonics.com 4865 E. 41 st Ave Denver, CO 80216 Phone: (303)-296-6766 Fax: (303)-296-6783 General Warnings and Cautions The following general warnings

More information

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS 3B SCIENTIFIC PHYSICS Complete Fine Beam Tube System 1013843 Instruction sheet 10/15 SD/ALF If it is to be expected that safe operation is impossible (e.g., in case of visible damage), the apparatus is

More information

K Service Source. Macintosh Color Display

K Service Source. Macintosh Color Display K Service Source Macintosh Color Display K Service Source Specifications Macintosh Color Display Specifications Characteristics - 1 Characteristics Picture Tube 14-in. diagonal (11.5-in. viewable image)

More information

Troubleshooting the CTS 1100

Troubleshooting the CTS 1100 CHAPTER 7 Troubleshooting the CTS 1100 Revised: November 2011, Contents You may want to periodically test system components using the hardware and software tests available in the Cisco TelePresence System

More information

FACSAria I Standard Operation Protocol Basic Operation

FACSAria I Standard Operation Protocol Basic Operation FACSAria I Standard Operation Protocol Basic Operation 1. Checking Lasers Status a. Please check the ON / OFF of the lasers. Sufficient time (~30 minutes) need to be given to allow the laser(s) to warm

More information

Safety Precautions and Maintenance

Safety Precautions and Maintenance Safety Precautions and Maintenance 1. Unplug the monitor before cleaning it with a slightly damp cloth. 2. Wiping the screen with a dry cloth is recommendable, and only during the power has been switched

More information

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject

More information

KVANT. Spectrum 1,6W RGB laser system. Product datasheet

KVANT. Spectrum 1,6W RGB laser system. Product datasheet KVANT Spectrum 1,6W RGB laser system Product datasheet SPECTRUM series professional full colour laser system 1. DESCRIPTION Kvant Spectrum is professional laser system with wide range of use, mainly for

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

Operating Instructions

Operating Instructions Operating Instructions LCDRV700 Digital LCD Color Monitor Please read this manual thoroughly before operating the unit, and keep it for future reference. V1.0 Contents 1. Precautions 2. Features 1 3 3.

More information

4. Alignment and Adjustments

4. Alignment and Adjustments 4. Alignment and Adjustments 4-1 Preadjustment 4-1-1 Factory Mode 1. Do not attempt these adjustments in the Video Mode. 2. The Factory Mode adjustments are necessary when either the EEPROM (IC02) or the

More information

DX80 Touchscreen User Guide

DX80 Touchscreen User Guide DX80 Touchscreen User Guide 07Mar2018 Page 1 of 16 Table of Contents Table of Contents GENERAL USE.... 5 Site to Site Calls.... 6 Connecting to an MBT Site.... 6 Add a Site to Favorites.... 6 Connecting

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Dektak 3030 Stylus Profilometer Version 2011 May 16 I. Purpose This Standard Operating Procedure

More information

Quick Help Teaching Room Technology Support

Quick Help Teaching Room Technology Support Quick Help Teaching Room Technology Support Technical assistance is available. If you require assistance, please call Ext 6066 Quick Help Technology Overview INDEX INDEX Touch Screen Is not active 3 Technology

More information

Version 04_08 4/8/2008

Version 04_08 4/8/2008 ` 1 2 Installation and Operation Manual Version 04_08 4/8/2008 3 Table of Contents MFP-3D Introduction... 15 1.1 Overview...15 1.2 MFP-3D System Overview...15 1.3 MFP-3D Stand-Alone (MFP-3D-SA )...15 1.4

More information

Analyzing and Saving a Signal

Analyzing and Saving a Signal Analyzing and Saving a Signal Approximate Time You can complete this exercise in approximately 45 minutes. Background LabVIEW includes a set of Express VIs that help you analyze signals. This chapter teaches

More information

[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL]

[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL] [AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL] VER: 2.0 TABLE OF CONTENT 1. INTRODUCTION...- 4-1.1 THEORY OF OPERATION...- 4-1.1.1 Principle of Atomic Force Microscope (AFM)... - 4-2. INSTALLATION &

More information

FOCUS VOLTAGE ADJUSTMENT 1. Receive RETMA pattern signal. 2. Adjust the FOCUS VOLUME on the FBT and make the picture on the screen be finest.

FOCUS VOLTAGE ADJUSTMENT 1. Receive RETMA pattern signal. 2. Adjust the FOCUS VOLUME on the FBT and make the picture on the screen be finest. General Information Also Covers: DVT-1484D, DVT-2084D Ferguson FG 14 CB 12V, FG 20 CB 12V Goodmans TVC 1400 & TVC 14 VP Electrical Adjustments (TV) GENERAL INFORMATION All adjustments are throughly checked

More information

ACUBRITE 23 SS. Manual. Stainless Steel Chassis 23" LCD Display. Content

ACUBRITE 23 SS. Manual. Stainless Steel Chassis 23 LCD Display. Content ACUBRITE 23 SS Stainless Steel Chassis 23" LCD Display Manual Introduction... 2 Hardware Installation... 2 The Display Timing... 5 The Display Outline Dimensions... 6 The Display Controls... 7 The Screen

More information

Linkage 3.6. User s Guide

Linkage 3.6. User s Guide Linkage 3.6 User s Guide David Rector Friday, December 01, 2017 Table of Contents Table of Contents... 2 Release Notes (Recently New and Changed Stuff)... 3 Installation... 3 Running the Linkage Program...

More information

F7000NV ROBOT VISION OPERATING MANUAL

F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting

More information

SB60 airsound BASE USER S GUIDE. version 1.0

SB60 airsound BASE USER S GUIDE. version 1.0 SB60 airsound BASE USER S GUIDE version.0 CONTENTS WELCOME PACKAGE CONTENTS The AirSOUND BASE SB60 AirSOUND BASE Controls & Indicators GETING STARTED 4 Remote control Sound Source / Setting memory FRONT

More information

HDCVI Eyeball Camera Lite

HDCVI Eyeball Camera Lite HDCVI Eyeball Camera Lite User s Manual V1.0.0 1 Foreword General This user s manual (hereinafter referred to be "the Manual") introduces the functions, installation, and operations of the camera. Safety

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt

More information

VPM2. Operation Manual

VPM2. Operation Manual VPM2 Operation Manual Whip Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com LISTED Features The Whip VPM2 is designed

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA The JS007WQK is loaded with userfriendly features and is ideal for use in heavy duty vehicles.

More information

CN Remove the scanner assembly (X476 and X576 models) and all doors/covers.

CN Remove the scanner assembly (X476 and X576 models) and all doors/covers. CN598-67045 www.hp.com/support IMPORTANT: Ensure the product firmware is upgraded to at least version 1336MR before performing this repair procedure. If the firmware upgrade cannot be completed, contact

More information

System Requirements SA0314 Spectrum analyzer:

System Requirements SA0314 Spectrum analyzer: System Requirements SA0314 Spectrum analyzer: System requirements Windows XP, 7, Vista or 8: 1 GHz or faster 32-bit or 64-bit processor 1 GB RAM 10 MB hard disk space \ 1. Getting Started Insert DVD into

More information

STANDARD OPERATING PROCEDURE:

STANDARD OPERATING PROCEDURE: Q-EXACTIVE STANDARD OPERATING PROCEDURE: Purpose of this Instrument: Essential tool for molecular weight identification for small molecules, peptides, proteins, ect. Location: WVU Chemistry Research Laboratory

More information