QPHY-USB3 USB3.0 Serial Data Operator s Manual

Size: px
Start display at page:

Download "QPHY-USB3 USB3.0 Serial Data Operator s Manual"

Transcription

1 QPHY-USB3 USB3.0 Serial Data Operator s Manual Revision A April, 2009 Relating to the Following Release Versions: Software Option Rev. 5.8 USB3 Script Rev. 1.0 Style Sheet Rev. 1.2

2 LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, Tel: (845) , Fax: (845) Internet: by LeCroy Corporation. All rights reserved. LeCroy and other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice. QPHY-USB3-OM-E Rev A Rev A

3 TABLE OF CONTENTS QPHY-USB3 Software Option INTRODUCTION... 5 Compatibility... 5 SETUP AND INSTALLATION... 6 Required Equipment Available from LeCroy... 6 QUALIPHY COMPLIANCE TEST PLATFORM... 7 Oscilloscope Option Key Installation... 9 Typical (Recommended) Configuration... 9 Remote (Network) Configuration... 9 Oscilloscope Selection... 9 Accessing the QPHY-USB3 Software using QualiPHY Customizing QualiPHY Creating Custom Configurations QPHY-USB3 Operation QPHY-USB3 Test Configurations All Tests Demo of All Tests Empty Template QPHY-USB3 Variables Channel File Demo Mode Emulate Channel Saved Waveform Path Save Individual Runs Test Mode Eye Diagram & Jitter Test Variables QPHY-USB3 Test Descriptions Deskew Test Eye Diagram & Jitter Test Test Spread Spectrum Test Test Differential Voltage Swing Test Test De-emphasis Ratio Test Compliance Pattern Definitions CALIBRATION PROCEDURES Cable Deskewing using the Fast Edge Output (WavePro 7 Zi and WaveMaster 8Zi Only) Cable Deskewing without using the Fast Edge Output QPHY-USB3-OM-E Rev A 3

4 FIGURES Figure 1. Report menu in QualiPHY General Setup... 7 Figure 2. The Test Report includes a summary table with links to the detailed test results... 8 Figure 3. QualiPHY main menu and compliance test Standard selection menu Figure 4. QualiPHY configuration selection menu Figure 5. QualiPHY test item selection menu Figure 6. Variable Setup and Limits Manager windows Figure 7. Start button Figure 8. Example of pop-up connection diagram and dialog box Figure 9 - S31 response of cable S-parameter file provided by the USB-IF Figure 10 - Oscilloscope Configuration after Eye Diagram Test Figure 11 - Oscilloscope Configuration after the Jitter Test Figure 12 - Oscilloscope Configuration after the Spread Spectrum Test Figure 13 - Oscilloscope screen after Differential Voltage Swing Test Figure 14 - Oscilloscope screen after the De-emphasis Ratio Test Figure 15 - Timebase Settings for Deskew with the Fast Edge Output Figure 16 - Channel Pre-Processing Settings for Deskew with the Fast Edge Output Figure 17 - Trigger Settings for Deskew with the Fast Edge Output Figure 18 - Measurement Settings for Deskew with the Fast Edge Output Figure 19 - Adjusted Timebase Settings for Deskew Figure 20 - Save Waveform Settings for Deskew with the Fast Edge Output Figure 21 - Final Screen with Channel 2 and 3 Deskewed Figure 22. The Skew parameter right side dialog, Skew clock 2 tab, showing default setup QPHY-USB3-OM-E Rev A

5 INTRODUCTION QPHY-USB3 Software Option QPHY-USB3 is a software package designed to capture, analyze, and report measurements in conformance with SuperSpeed USB electrical specification standard. A copy of the specification can be found at Compatibility QPHY-USB3 is a software option compatible with the following LeCroy X-Stream oscilloscopes: WM813Zi, SDA813Zi, DDA813Zi (4x40 GS/s, 2x80GS/s) and higher bandwidth equivalents. SDA (4x20 GS/s at 6 GHz bandwidth, 2x40 GS/s at 13 GHz bandwidth) The minimum recommended bandwidth for testing USB 3.0 devices and hosts is 13GHz. QPHY-USB3-OM-E Rev A 5

6 SETUP AND INSTALLATION The following test equipment is required to perform USB 3.0 tests. Required Equipment Available from LeCroy Real Time Digital Oscilloscope as listed in Compatibility section with firmware release or later, and software option package QPHY-USB3 QualiPHY Software (version x or later) installed on either the oscilloscope or a separate PC. 50Ω Coax Cable with SMA Male Connectors, qty = 2 These cables should be nominally the same length so that the attenuation characteristics and length are well matched. 6 QPHY-USB3-OM-E Rev A

7 QUALIPHY COMPLIANCE TEST PLATFORM QPHY-USB3 Software Option QualiPHY is LeCroy s unique compliance test framework which leads the user through the compliance tests. QualiPHY displays connection diagrams to ensure tests run properly, automates the oscilloscope setup, and generates full compliance reports. QualiPHY makes USB 3.0 compliance testing easy and fast. The QualiPHY software application automates the test and report generation. Figure 1. Report menu in QualiPHY General Setup QPHY-USB3-OM-E Rev A 7

8 See the QualiPHY Operator s Manual for more information on how to use the QualiPHY framework. Figure 2. The Test Report includes a summary table with links to the detailed test results 8 QPHY-USB3-OM-E Rev A

9 Oscilloscope Option Key Installation QPHY-USB3 Software Option An option key must be purchased to enable the QPHY-USB3 option. Call LeCroy Customer Support to place an order and receive the code. Enter the key and enable the purchased option as follows: 1. From the oscilloscope menu select Utilities Utilities Setup Select the Options tab and click the Add Key button. 3. Enter the Key Code using the on-screen keyboard. 4. Restart the oscilloscope to activate the option after installation. Typical (Recommended) Configuration QualiPHY software can be executed from the oscilloscope or a host computer. The first step is to install QualiPHY. Please refer to the QualiPHY Operator s Manual for installation instructions. LeCroy recommends running QualiPHY on an oscilloscope equipped with Dual Monitor Display capability (Option DMD-1 for oscilloscopes where this is not standard). This allows the waveform and measurements to be shown on the oscilloscope LCD display while the QualiPHY application and test results are displayed on a second monitor. By default, the oscilloscope appears as a local host when QualiPHY is executed in the oscilloscope. Follow the steps under Oscilloscope Selection (as follows) and check that the IP address is Remote (Network) Configuration It is also possible to install and run QualiPHY on a host computer, controlling the oscilloscope with a Network/LAN Connection. The oscilloscope must already be configured, and an IP address (fixed or network-assigned) must already be established. Oscilloscope Selection Set up the oscilloscope using QualiPHY over a LAN (Local Area Network) by doing the following: 1. Make sure the host computer is connected to the same LAN as the oscilloscope. If unsure, contact your system administrator. 2. From the oscilloscope menu, select Utilities Utilities Setup 3. Select the Remote tab. 4. Verify the oscilloscope has an IP address and the control is set to TCP/IP. 5. Run QualiPHY in the host computer and click the General Setup button. 6. Select the Connection tab. 7. Enter the IP address from step 4 (previous). 8. Click the Close button. QualiPHY is now ready to control the oscilloscope. QPHY-USB3-OM-E Rev A 9

10 QualiPHY tests the oscilloscope connection after clicking the Start button. The system prompts you if there is a connection problem. QualiPHY s Scope Selector function can also be used to verify the connection. Please refer to the QualiPHY Operator s Manual for explanations on how to use Scope Selector and other QualiPHY functions. Accessing the QPHY-USB3 Software using QualiPHY This topic provides a basic overview of QualiPHY s capabilities. Please refer to the QualiPHY Operator s Manual for detailed information. Access the QPHY-USB3 software using the following steps: 1. Wait for the oscilloscope to start and have its main application running. 2. Launch QualiPHY from the Analysis menu if installed on the oscilloscope or from the desktop icon if installed on a host computer. 3. From the QualiPHY main window (as follows), select Standard, then USB3.0 SS from the pop-up menu (if not already selected). If you check the Pause on Failure box (circled) QualiPHY prompts to retry the measure whenever a test fails. Figure 3. QualiPHY main menu and compliance test Standard selection menu 10 QPHY-USB3-OM-E Rev A

11 QPHY-USB3 Software Option 4. Click the Configuration button in the QualiPHY main menu: 5. Select a configuration from the pop-up menu: 6. Click Start. Figure 4. QualiPHY configuration selection menu 7. Follow the pop-up window prompts. QPHY-USB3-OM-E Rev A 11

12 Customizing QualiPHY The predefined configurations in the Configuration screen cannot be modified. However, you can create your own test configurations by copying one of the standard test configurations and making modifications. A description of the test is also shown in the description field when selected. Figure 5. QualiPHY test item selection menu 12 QPHY-USB3-OM-E Rev A

13 Creating Custom Configurations Beginning with any of the pre-loaded configurations, 2. Click on the Variable Setup tab to change the variables for your new configuration. QPHY-USB3 Software Option 1. Click on the Test Selector tab to change what tests you would like to be included in your new configuration. 3. Click on the Limits tab to change which limit set should be used for your new configuration See QualiPHY Manual for more information 4. Once a change has been made to any of these sections, the Save As button becomes clickable on the bottom of the dialog. 5. Clicking the Save As button will prompt you for a new configuration name and description. Note: If a Custom Configuration was used for the procedure, the Save button will also become clickable on the bottom of the dialog. Clicking this button will update the current configuration with new changes. 6. Once a custom configuration is defined, script variables and the test limits can be changed by using the Variable Setup and Limits Manager from the Edit/View Configuration window. QPHY-USB3-OM-E Rev A 13

14 Figure 6. Variable Setup and Limits Manager windows 14 QPHY-USB3-OM-E Rev A

15 QPHY-USB3 Operation QPHY-USB3 Software Option After pressing Start in the QualiPHY menu, the software instructs how to set up the test using pop-up connection diagrams and dialog boxes. QualiPHY also instructs how to properly configure the Product Under Test (PUT) to change test signal modes (when necessary). Figure 7. Start button Figure 8. Example of pop-up connection diagram and dialog box QPHY-USB3-OM-E Rev A 15

16 QPHY-USB3 Test Configurations Configurations include variable settings and limit sets as well, not just test selections. See the QPHY-USB3 Variables section for a description of each variable value and its default value. All Tests This configuration will run all of the tests and the channel deskew procedure. The limit set in use is Default. All of the variables are set to their default settings. Demo of All Tests This configuration will run all of the tests. This limit set in use is Default. All of the variables are set to their default setting except Test Mode is set to Use Sav ed Data and the Demo Mode is set to Yes. This configuration will run the tests on the waveforms that are stored in the directory specified by the Saved Waveform Path variable. Empty Template This configuration is intentionally left blank so it can easily be copied and configured to the user s custom configuration. The limit set in use is Default. All of the variables are set to their default settings. 16 QPHY-USB3-OM-E Rev A

17 QPHY-USB3 Software Option QPHY-USB3 Variables Channel File This variable specifies the name of a Touchstone format S-parameter file to be used for emulating a serial data channel. The USB 3.0 specification requires that the Eye Diagram and Jitter tests be performed at the far end of a serial data test channel. QPHY-USB3 utilizes the Eye Doctor II functionality to emulate the serial data channel based on the specified S-parameter file. The S-parameter file must be located in the D:\Applications\EyeDr folder on the oscilloscope. The S-parameter file provided by the USB-IF has the following S31 response: Figure 9 - S31 response of cable S-parameter file provided by the USB-IF Demo Mode This variable allows the user to run the tests in demonstration mode using saved waveforms. The waveforms must be located in the location specified by the Saved Waveform Path variable. When this variable is set to Yes the user will still be prompted with connection diagrams just as if the tests were being run on live data. Emulate Channel This variable allows the user to select whether they would like QPHY-USB3 to emulate the serial data channel. The USB 3.0 specification requires that the Eye Diagram and Jitter tests be performed at the far end of a serial data channel, but if a physical channel is used for testing there is no need to emulate the channel. Saved Waveform Path This variable specifies the directory for QPHY-USB3 to find the waveforms to test when the Test Mode variable is set to Use Saved Data. Save Individual Runs When this value is set to "Yes" waveforms will be saved in a separate folder for each time the test is run. For example: D:\Waveforms\USB3\[Device Under Test]\Run1 QPHY-USB3-OM-E Rev A 17

18 When this value is set to "No" waveforms will be overwritten on every run and saved in: D:\Waveforms\USB3\[Device Under Test] Test Mode This variable allows the user to select whether they would like to Acquire New Data or Use Saved Data. If Acquire New Data is selected, QPHY-USB3 will capture new waveforms to test. If Use Saved Data is selected, QPHY-USB3 will test the waveforms found in the directory specified by the Saved Waveform Path variable. Eye Diagram & Jitter Test Variables The variables in this section allow the user to decide whether or not they would like the oscilloscope to perform equalization as described by the specification. If yes, the user also has the ability to decide the location of the 2 poles, the location of the zero and the DC gain for the continuous time linear equalizer (CTLE). 18 QPHY-USB3-OM-E Rev A

19 QPHY-USB3 Software Option QPHY-USB3 Test Descriptions Deskew This procedure measures the skew between the two oscilloscope channels used and sets the deskew value appropriately. Test Eye Diagram & Jitter Test The purpose of this test is to verify that the eye height and jitter are within the specification limits. For compliance testing the measurements are made after a reference test channel and after applying the reference Continuous Time Linear Equalizer function. Eye height is measured using compliance pattern CP0. Jitter is measured using both compliance pattern CP0 and compliance pattern CP1 (see below for definition). Eye height is measured over 10 million consecutive unit intervals. Jitter is extrapolated to BER. Refer to section of the Universal Serial Bus 3.0 Specification, Revision 1.0 for more details. At the completion of the Eye Diagram Test the oscilloscope is in the following configuration: Figure 10 - Oscilloscope Configuration after Eye Diagram Test On this screen the EyeOpening is measured in parameter P1. P2 is used for a Histogram Max parameter of a vertical slice of the center of the eye diagram and P3 is used for a Histogram Min parameter of a vertical slice of the center of the eye diagram. The difference of P2 and P3 is the result for the EyeOpening measurement. QPHY-USB3-OM-E Rev A 19

20 At the completion of the Jitter Test the oscilloscope is in the following configuration: Figure 11 - Oscilloscope Configuration after the Jitter Test On the oscilloscope screen Tj (at BER = ), Rj(spectral) and Dj(spectral) are measured in the SDA Jitter section. Tj, Rj and Dj are converted into UI by dividing their value by 200ps. These results are the Tj and Rj values in mui. Also shown on screen are the RjBUjHistogram, Bathtub curve and NQFit lines. Test Spread Spectrum Test The purpose of this test is to measure the spread spectrum modulation frequency and deviation and verify that they are within the specification limits. The data is demodulated and a low-pass filter is then applied to create an SSCTrack which displays how the data rate varies over time. The carrier frequency of the SSC Track is set to the center frequency of the acquired SSC waveform. The SSC Tra ck has a low pass filter (LPF) with a cutoff frequency of 2 MHz. The output of the SSC Track is converted to parts per million (ppm) deviation from the nominal bitrate so that it becomes a track of the deviation relative to the nominal bitrate. Spread spectrum is measured using compliance pattern CP1 (see below for definition). At the completion of the Spread Spectrum Test the oscilloscope is in the following configuration: 20 QPHY-USB3-OM-E Rev A

21 QPHY-USB3 Software Option Figure 12 - Oscilloscope Configuration after the Spread Spectrum Test The oscilloscope screen is now displaying the SSC Track. There are also 3 parameters on the screen. P1 is the minimum deviation from the nominal bitrate. P2 is the maximum deviation from the nominal bitrate. The spread spectrum modulation is downward so the minimum deviation is at the top of the SSC Track and the maximum deviation is at the bottom. P3 is the modulation frequency of the SSC Track. Test Differential Voltage Swing Test The purpose of this test is to measure the differential voltage swing and verify that it is within the specification limits. The differential voltage swing is measured using compliance pattern CP8 (see below for definition). At the completion of the Differential Voltage Swing Test the oscilloscope is in the following configuration: QPHY-USB3-OM-E Rev A 21

22 Figure 13 - Oscilloscope screen after Differential Voltage Swing Test Shown on the screen are F1, F2, P1, P2 and P3. F1 is the difference of the two input channels (C2-C3). F2 is the histogram of that differential waveform. P1 is the mode of the histogram gated to the left half of the screen. This is equal to the base differential voltage. P2 is the mode of the histogram gated to the right half of the screen. This is equal to the top differential voltage. P3 is the difference between P2 and P1. This is equal to the differential voltage swing pk-pk. Test De-emphasis Ratio Test The purpose of this test is to measure the de-emphasis ratio and verify that it is within the specification limits. The differential voltage swing is measured using compliance pattern CP7 (see below for definition). At the completion of the De-emphasis Ratio Test the oscilloscope is in the following configuration: 22 QPHY-USB3-OM-E Rev A

23 QPHY-USB3 Software Option Figure 14 - Oscilloscope screen after the De-emphasis Ratio Test On the screen are F1, Z1 and P1. F1 is the difference between the two input channels (C2-C3). Z1 is the zoom of F1 showing the full swing transition bit to display the de-emphasis. P1 is the voltage transmitter de-emphasis ratio parameter. This parameter returns the ratio between peak to peak voltages of full swing and de-emphasized bits. QPHY-USB3-OM-E Rev A 23

24 Compliance Pattern Definitions The USB3 SuperSpeed compliance patterns are defined in section of the Universal Serial Bus 3.0 Specification, Revision 1.0 and reproduced below for convenience. Compliance Pattern Value Description CP0 D0.0 Scrambled A pseudo-random data pattern that is exactly the same as logical idle but does not include SKP sequences CP1 D10.2 Nyquist frequency CP2 D24.3 Nyquist/2 CP3 K28.5 COM pattern CP4 LFPS The low frequency periodic signaling pattern CP5 K28.7 With de-emphasis CP6 K28.7 Without de-emphasis CP s and 0 s With de-emphasis. Repeating s and then s CP s and 0 s Without de-emphasis. Repeating s and then s 24 QPHY-USB3-OM-E Rev A

25 QPHY-USB3 Software Option CALIBRATION PROCEDURES Cable Deskewing using the Fast Edge Output (WavePro 7 Zi and WaveMaster 8Zi only) The following procedure demonstrates how to deskew two oscilloscope channels and cables using the fast edge output, with no need for any T connector or adapters. This can be done once the temperature of the oscilloscope is stable. The oscilloscope must be warmed up for at least a half-hour before proceeding. This procedure should be run again if the temperature of the oscilloscope changes by more than a few degrees. For the purpose of this procedure, the two channel s being deskewed will be referred to as Channel X and Channel Y. The reference channel will be Channel X and the channel being deskewed will be Channel Y. 1. Begin by recalling the Default Oscilloscope Setup 2. Configure the oscilloscope as follows: Timebase i. Fixed Sample Rate ii. Set the Sample Rate to 40 GS/s iii. Set the Time/Division to 1 ns/div Figure 15 - Timebase Settings for Deskew with the Fast Edge Output Channels i. Turn on Channel X and Channel Y ii. Set V/div for Channel X and Channel Y to 100mV/div iii. Set the Averaging of Channel X and Channel Y to 500 sweeps iv. Set the Interpolation of Channel X and Channel Y to Sinx/x Figure 16 - Channel Pre-Processing Settings for Deskew with the Fast Edge Output Trigger i. Configure to Source to be FastEdge ii. Set the Slope to Positive QPHY-USB3-OM-E Rev A 25

26 Figure 17 - Trigger Settings for Deskew with the Fast Edge Output Parameter Measurements: i. Set the source for P1 to CX and the measure to Delay ii. Set the source for P2 to CY and the measure to Delay iii. Set the source for P3 to M1 and the measure to Delay Figure 18 - Measurement Settings for Deskew with the Fast Edge Output 3. Set the display to Single Grid Click Display -> Single Grid 4. Using the appropriate adapter, connect Channel X to the Fast Edge Output of the oscilloscope 5. Adjust the Trigger Delay so that the Channel X signal crosses at the center of the screen 6. Change the Timebase to 50 ps/div Figure 19 - Adjusted Timebase Settings for Deskew 7. Fine tune the Trigger Delay so that the Channel X signal crosses at the exact center of the screen. 8. Press the Clear Sweeps button on the front panel to reset the averaging 9. Allow multiple acquisitions to occur until the waveform is stable on the screen. 10. Save Channel X to M1 Click File -> Save Waveform Set Save To Memory Set the Source to CX Set the Destination to M1 Click Save Now 26 QPHY-USB3-OM-E Rev A

27 QPHY-USB3 Software Option Figure 20 - Save Waveform Settings for Deskew with the Fast Edge Output 11. Disconnect Channel X from the Fast Edge Output and connect Channel Y to the Fast Edge Output. 12. Press the Clear Sweeps button on the front panel to reset the averaging. 13. Allow multiple acquisitions to occur until the waveform is stable on the screen. 14. From the Channel Y menu, adjust the Deskew of Channel Y until Channel Y is directly over the M1 trace. 15. Ensure that P3 and P2 are reasonably close to the same value. (Typically < 5ps difference) Figure 21 - Final Screen with Channel 2 and 3 Deskewed QPHY-USB3-OM-E Rev A 27

28 Cable Deskewing without using the Fast Edge Output The following procedure demonstrates how to deskew two oscilloscope channels and cables using the differential data signal, with no need for any T connector or adapters. This can be done once the temperature of the oscilloscope is stable. The oscilloscope must be warmed up for at least a half-hour before proceeding. This procedure should be run again if the temperature of the oscilloscope changes by more than a few degrees. 1. Connect a differential data signal to C2 and C3 using two approximately matching cables. Set up the oscillosc ope to use the maximum sample rate (e.g. for WaveRunner 204Xi: two channel mode; on the Smart Memory menu set Fixed Sample Rate 10GS/s). Set the timebase for a few repetitions of the pattern (at least a few dozen edges). 2. On the C3 menu, check Invert. Now C2 and C3 should look the same. 3. Using the Measure Setup, set P1 to measure the Skew of C2, C3. Turn on Statistics (Measure menu). Write down the mean skew value after it stabilizes. This mean skew value is the addition of Data skew + cable skew + channel skew. 4. Swap the cable connections on the Data source side (on the test fixture), and then press the Clear Sweeps button on the oscilloscope (to clear the accumulated statistics; since we changed the input). 5. Write down the mean skew value after it stabilizes. This mean skew value is the addition of (-Data skew) + cable skew + channel skew. 6. Add the two mean skew values and divide the sum in half: [Data skew + cable skew + channel skew] + [ (-Data skew) + cable skew + channel skew] 2 7. The above formula simplifies to: [cable skew + channel skew] 8. Set the resulting value as the Deskew value in C2 menu. 9. Restore the cable connections to their Step 1 settings (previous). Press the Clear Sweeps button on the oscilloscope. The mean skew value should be approximately zero - that is the data skew. Typically, results are <1ps given a test fixture meant to minimize skew on the differential pair. 10. On the C3 menu, un-check the Invert checkbox and turn off the parameters. In the previous procedure, we used the default setup of the Skew parameter (which is detecting positive edges on both signals at 50%). We also inverted C3 in order to make C2 and C3 both have positive edges at the same time. 28 QPHY-USB3-OM-E Rev A

29 QPHY-USB3 Software Option Alternately, we clearly could have not inverted C3 and instead selected the Skew clock 2 tab in the P1 parameter menu and set the oscilloscope to look for negative edges on the second input (C3). However, it is somewhat agreed that the previous procedure looks much more aesthetically pleasing from the display as it shows C2 and C3 with the same polarity. Figure 22. The Skew parameter right side dialog, Skew clock 2 tab, showing default setup QPHY-USB3-OM-E Rev A 29

QPHY-HDMI2 Operator s Manual

QPHY-HDMI2 Operator s Manual QPHY-HDMI2 Operator s Manual Revision B November, 2017 Relating to: XStreamDSO Version 8.5.x.x QualiPHY Version 8.5.x.x 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax:

More information

Thank you for purchasing a LeCroy Zi Oscilloscope Synchronization ProBus Module (Zi 8CH SYNCH module).

Thank you for purchasing a LeCroy Zi Oscilloscope Synchronization ProBus Module (Zi 8CH SYNCH module). Introduction Instructions Thank you for purchasing a LeCroy Zi Oscilloscope Synchronization ProBus Module (Zi 8CH SYNCH module). The module is designed for use on 4 30 GHz WaveMaster 8 Zi oscilloscopes

More information

QPHY-DDR4. Instruction Manual

QPHY-DDR4. Instruction Manual QPHY-DDR4 DDR4 Serial Data Compliance Software Instruction Manual Revision C November, 2017 Relating to: XStreamDSO Version 8.5.x.x QualiPHY Version 8.5.x.x 700 Chestnut Ridge Road Chestnut Ridge, NY,

More information

The Measurement Tools and What They Do

The Measurement Tools and What They Do 2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying

More information

WAVEJET 300 SERIES OSCILLOSCOPES. New Cover to Come. Unmatched Performance, Portability, and Value

WAVEJET 300 SERIES OSCILLOSCOPES. New Cover to Come. Unmatched Performance, Portability, and Value WAVEJET 300 SERIES OSCILLOSCOPES New Cover to Come Unmatched Performance, Portability, and Value ALL THE TOOLS YOU NEED Automatic Measurements Save time making measurements on your signals by using the

More information

Portable Performance for Debug and Validation

Portable Performance for Debug and Validation WaveJet 300A Oscilloscopes 100 MHz 500 MHz Portable Performance for Debug and Validation A UNIQUE TOOLSET FOR PORTABLE OSCILLOSCOPES Key Features 100 MHz, 200 MHz, 350 MHz and 500 MHz bandwidths Sample

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 9500B Users Supplement Issue: 2 Part Number: 1625019 Issue Date: 9/06 Print Date: October 2005 Page Count: 6 Version 11 This supplement contains information necessary to ensure the accuracy

More information

Procedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement

Procedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement Procedures Guide Tektronix HDMI Sink Instruments Differential Impedance Measurement Rev. 1.1: October 13, 2010 2 Measurement Procedures Equipment Required Table 1 lists the equipment required to perform

More information

WAVEJET 300 SERIES OSCILLOSCOPES. Unmatched Performance, Portability, and Value

WAVEJET 300 SERIES OSCILLOSCOPES. Unmatched Performance, Portability, and Value WAVEJET 300 SERIES OSCILLOSCOPES Unmatched Performance, Portability, and Value 1 WAVEJET 300 SERIES Unique Capabilities in a Low Bandwidth Oscilloscope The WaveJet 300 Series features unmatched performance

More information

X-Stream DSO Version 3.6 Release Notes

X-Stream DSO Version 3.6 Release Notes X-Stream DSO Version 3.6 Release Notes probe calibration and deskew fixture This release contains support for LeCroy s new TF-DSQ option. The fixture enables DC gain, offset, and skew calibration at the

More information

Draft Baseline Proposal for CDAUI-8 Chipto-Module (C2M) Electrical Interface (NRZ)

Draft Baseline Proposal for CDAUI-8 Chipto-Module (C2M) Electrical Interface (NRZ) Draft Baseline Proposal for CDAUI-8 Chipto-Module (C2M) Electrical Interface (NRZ) Authors: Tom Palkert: MoSys Jeff Trombley, Haoli Qian: Credo Date: Dec. 4 2014 Presented: IEEE 802.3bs electrical interface

More information

Digital Storage Oscilloscopes 2550 Series

Digital Storage Oscilloscopes 2550 Series Data Sheet Digital Storage Oscilloscopes 2550 Series The 2550 series digital storage oscilloscopes provide high performance and value in 2-channel and 4-channel configurations. With bandwidth from 70 MHz

More information

Agilent N5431A XAUI Electrical Validation Application

Agilent N5431A XAUI Electrical Validation Application Agilent N5431A XAUI Electrical Validation Application Methods of Implementation s Agilent Technologies Notices Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by

More information

SDAIII-CompleteLinQ Multi-Lane Serial Data, Noise and Crosstalk Analysis

SDAIII-CompleteLinQ Multi-Lane Serial Data, Noise and Crosstalk Analysis SDAIII-CompleteLinQ Multi-Lane Serial Data, Noise and Crosstalk Analysis TOOLS TO MEET SERIAL DATA ANALYSIS CHALLENGES Key Features Most complete jitter decomposition, eye diagram and analysis tools Up

More information

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool For the SIA Applications of Propagation Delay & Skew tool Determine signal propagation delay time Detect skewing between channels on rising or falling edges Create histograms of different edge relationships

More information

Agilent N6467A BroadR-Reach Compliance Test Application. Methods of Implementation

Agilent N6467A BroadR-Reach Compliance Test Application. Methods of Implementation Agilent N6467A BroadR-Reach Compliance Test Application Methods of Implementation s1 Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any means (including

More information

LeCroy Digital Oscilloscopes

LeCroy Digital Oscilloscopes LeCroy Digital Oscilloscopes Get the Complete Picture Quick Reference Guide QUICKSTART TO SIGNAL VIEWING Quickly display a signal View with Analog Persistence 1. Connect your signal. When you use a probe,

More information

Expect to Make Waves.

Expect to Make Waves. Expect to Make Waves. The New Oscilloscope Large 10.4" LCD touch screen Long capture time Extensive communication capabilities www.lecroy.com The New Oscillos From its large 10.4" LCD touch screen to its

More information

CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock

CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock Products: CMU200 CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock This application explains the setup and procedure to measure the exact time relationship

More information

Oscilloscope Guide Tektronix TDS3034B & TDS3052B

Oscilloscope Guide Tektronix TDS3034B & TDS3052B Tektronix TDS3034B & TDS3052B Version 2008-Jan-1 Dept. of Electrical & Computer Engineering Portland State University Copyright 2008 Portland State University 1 Basic Information This guide provides basic

More information

Eye Doctor II Advanced Signal Integrity Tools

Eye Doctor II Advanced Signal Integrity Tools Eye Doctor II Advanced Signal Integrity Tools EYE DOCTOR II ADVANCED SIGNAL INTEGRITY TOOLS Key Features Eye Doctor II provides the channel emulation and de-embedding tools Adds precision to signal integrity

More information

Datasheet SHF A

Datasheet SHF A SHF Communication Technologies AG Wilhelm-von-Siemens-Str. 23D 12277 Berlin Germany Phone +49 30 772051-0 Fax ++49 30 7531078 E-Mail: sales@shf.de Web: http://www.shf.de Datasheet SHF 19120 A 2.85 GSa/s

More information

Tektronix Inc. DisplayPort Standard

Tektronix Inc. DisplayPort Standard DisplayPort Standard 06-12-2008 DisplayPort Standard Tektronix MOI for Sink Tests (AWG Jitter Generation using Direct Synthesis and calibration using Real Time DPO measurements for Sink Devices) DisplayPort

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

MS-32 OSCILLOSCOPE MIXED SIGNAL OPTION. Add 32 Digital Channels to a 4 Channel Oscilloscope

MS-32 OSCILLOSCOPE MIXED SIGNAL OPTION. Add 32 Digital Channels to a 4 Channel Oscilloscope MS-32 OSCILLOSCOPE MIXED SIGNAL OPTION Add 32 Digital Channels to a 4 Channel Oscilloscope 4 Analog + 32 Digital Channel Capability LeCroy introduces the first oscilloscope solution to combine 4 analog

More information

WAVEEXPERT SERIES OSCILLOSCOPES WE 9000 NRO 9000 SDA 100G. The World s Fastest Oscilloscope

WAVEEXPERT SERIES OSCILLOSCOPES WE 9000 NRO 9000 SDA 100G. The World s Fastest Oscilloscope WAVEEXPERT SERIES OSCILLOSCOPES WE 9000 NRO 9000 SDA 100G The World s Fastest Oscilloscope The Fastest Oscilloscope in the Marketplace The WaveExpert and SDA 100G are the first instruments to combine the

More information

FLEXRAY TRIGGER, DECODE, AND PHYSICAL LAYER TEST

FLEXRAY TRIGGER, DECODE, AND PHYSICAL LAYER TEST FLEXRAY TRIGGER, DECODE, AND PHYSICAL LAYER TEST Operator s Manual MARCH, 2008 LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY 10977 6499 Tel: (845) 578 6020, Fax: (845) 578-5985 Internet:

More information

Serial Decode I2C TEN MINUTE TUTORIAL. December 21, 2011

Serial Decode I2C TEN MINUTE TUTORIAL. December 21, 2011 Serial Decode I2C TEN MINUTE TUTORIAL December 21, 2011 Summary LeCroy oscilloscopes have the ability to trigger on and decode multiple serial data protocols. The decode in binary, hex, or ASCII format,

More information

Operating Instructions

Operating Instructions Operating Instructions HAEFELY TEST AG KIT Measurement Software Version 1.0 KIT / En Date Version Responsable Changes / Reasons February 2015 1.0 Initial version WARNING Introduction i Before operating

More information

Reference. TDS7000 Series Digital Phosphor Oscilloscopes

Reference. TDS7000 Series Digital Phosphor Oscilloscopes Reference TDS7000 Series Digital Phosphor Oscilloscopes 07-070-00 0707000 To Use the Front Panel You can use the dedicated, front-panel knobs and buttons to do the most common operations. Turn INTENSITY

More information

Boosting Performance Oscilloscope Versatility, Scalability

Boosting Performance Oscilloscope Versatility, Scalability Boosting Performance Oscilloscope Versatility, Scalability Rising data communication rates are driving the need for very high-bandwidth real-time oscilloscopes in the range of 60-70 GHz. These instruments

More information

Operator's Manual. MS-250 Mixed Signal Oscilloscope Option

Operator's Manual. MS-250 Mixed Signal Oscilloscope Option Operator's Manual MS-250 Mixed Signal Oscilloscope Option MS-250 Mixed Signal Oscilloscope Option Operator's Manual April, 2017 MS-250 Mixed Signal Oscilloscope Option Operator's Manual 2017 Teledyne

More information

Agilent MOI for HDMI 1.4b Cable Assembly Test Revision Jul 2012

Agilent MOI for HDMI 1.4b Cable Assembly Test Revision Jul 2012 Revision 1.11 19-Jul 2012 Agilent Method of Implementation (MOI) for HDMI 1.4b Cable Assembly Test Using Agilent E5071C ENA Network Analyzer Option TDR 1 Table of Contents 1. Modification Record... 4 2.

More information

Keysight N6467A/N6467B BroadR-Reach Compliance Test Application. Methods of Implementation

Keysight N6467A/N6467B BroadR-Reach Compliance Test Application. Methods of Implementation Keysight N6467A/N6467B BroadR-Reach Compliance Test Application Methods of Implementation Notices Keysight Technologies 2014-2017 No part of this manual may be reproduced in any form or by any means (including

More information

MIPI DigRF 3G and v4 Decode

MIPI DigRF 3G and v4 Decode MIPI DigRF 3G and v4 Decode Key Features DigRF v4 Decodes Low-speed (26 Mb/s) Medium-speed (1248 Mb/s) High-speed (1456 Mb/s) View an entire DigRF 3G burst using intuitive decoding and table display. Click

More information

Boonton 4540 Remote Operation Modes

Boonton 4540 Remote Operation Modes Application Note Boonton 4540 Remote Operation Modes Mazumder Alam Product Marketing Manager, Boonton Electronics Abstract Boonton 4540 series power meters are among the leading edge instruments for most

More information

Keysight Method of Implementation (MOI) for VESA DisplayPort (DP) Standard Version 1.3 Cable-Connector Compliance Tests Using E5071C ENA Option TDR

Keysight Method of Implementation (MOI) for VESA DisplayPort (DP) Standard Version 1.3 Cable-Connector Compliance Tests Using E5071C ENA Option TDR Revision 1.00 February 27, 2015 Keysight Method of Implementation (MOI) for VESA DisplayPort (DP) Standard Version 1.3 Cable-Connector Compliance Tests Using E5071C ENA Option TDR 1 Table of Contents 1.

More information

DDA-UG-E Rev E ISSUED: December 1999 ²

DDA-UG-E Rev E ISSUED: December 1999 ² 7LPHEDVH0RGHVDQG6HWXS 7LPHEDVH6DPSOLQJ0RGHV Depending on the timebase, you may choose from three sampling modes: Single-Shot, RIS (Random Interleaved Sampling), or Roll mode. Furthermore, for timebases

More information

PicoScope 9300 Series migration guide

PicoScope 9300 Series migration guide sampling oscilloscopes since 2009 The 9300 Series is a leading-edge product family resulting from a long program of product development. From late 2017, in the process of adding new 15 GHz and 25 GHz models,

More information

TekExpress Ethernet Tx Compliance Solution Printable Application Help

TekExpress Ethernet Tx Compliance Solution Printable Application Help TekExpress Ethernet Tx Compliance Solution Printable Application Help *P077125300* 077-1253-00 TekExpress Ethernet Tx Compliance Solution Printable Application Help www.tek.com 077-1253-00 Copyright Tektronix.

More information

Generation of Novel Waveforms Using PSPL Pulse Generators

Generation of Novel Waveforms Using PSPL Pulse Generators Generation of Novel Waveforms Using PSPL Pulse Generators James R. Andrews, Ph.D, IEEE Fellow & Bob McLaughlin PSPL Founder & former President (retired) PSPL Sales Engineer Picosecond Pulse Labs (PSPL)

More information

Electrical Sampling Modules Datasheet 80E11 80E11X1 80E10B 80E09B 80E08B 80E07B 80E04 80E03 80E03-NV

Electrical Sampling Modules Datasheet 80E11 80E11X1 80E10B 80E09B 80E08B 80E07B 80E04 80E03 80E03-NV Electrical Sampling Modules Datasheet 80E11 80E11X1 80E10B 80E09B 80E08B 80E07B 80E04 80E03 80E03-NV The DSA8300 Series Sampling Oscilloscope, when configured with one or more electrical sampling modules,

More information

立肯科技 LeColn Technology

立肯科技 LeColn Technology DisplayPort PHY Validation 立肯科技 LeColn Technology 1 DisplayPort Basics Maximum bit rate DP1.2b 1.62Gb/s( RBR = reduced bit rate) 2.7Gb/s( HBR = high bit rate) 5.4Gb/s( HBR2 =high bit rate 2) DP1.3/1.4

More information

De-embedding Gigaprobes Using Time Domain Gating with the LeCroy SPARQ

De-embedding Gigaprobes Using Time Domain Gating with the LeCroy SPARQ De-embedding Gigaprobes Using Time Domain Gating with the LeCroy SPARQ Dr. Alan Blankman, Product Manager Summary Differential S-parameters can be measured using the Gigaprobe DVT30-1mm differential TDR

More information

Receiver Testing to Third Generation Standards. Jim Dunford, October 2011

Receiver Testing to Third Generation Standards. Jim Dunford, October 2011 Receiver Testing to Third Generation Standards Jim Dunford, October 2011 Agenda 1.Introduction 2. Stressed Eye 3. System Aspects 4. Beyond Compliance 5. Resources 6. Receiver Test Demonstration PCI Express

More information

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a

More information

Fluke 190-Series II Firmware Upgrade V11.44

Fluke 190-Series II Firmware Upgrade V11.44 Fluke 190-Series II Firmware Upgrade V11.44 Requirements 1. Fluke 190- Series II ScopeMeter with firmware prior to V11.44 2. Supported models are: 190-102, 190-104, 190-062, 190-202, 190-204, 190-502,

More information

Dual Scope Synchronization

Dual Scope Synchronization Dual Scope Synchronization Application Note Introduction The Tektronix DPO/DSA/MSO70000 models above 12GHz in bandwidth provide 50 GS/s sampling rate on each of 4 channels simultaneously, or 100 GS/s sampling

More information

Prepare for Next Generation USB Technology Testing

Prepare for Next Generation USB Technology Testing Prepare for Next Generation USB Technology Testing Disclaimer The USB 3.1 compliance test requirements are not final therefore all opinions, judgments, recommendations, etc., that are presented herein

More information

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Products: R&S RTO2002 R&S RTO2004 R&S RTO2012 R&S RTO2014 R&S RTO2022 R&S RTO2024 R&S RTO2044 R&S RTO2064 This application

More information

M1 OSCILLOSCOPE TOOLS

M1 OSCILLOSCOPE TOOLS Calibrating a National Instruments 1 Digitizer System for use with M1 Oscilloscope Tools ASA Application Note 11-02 Introduction In ASA s experience of providing value-added functionality/software to oscilloscopes/digitizers

More information

Agilent E4887A HDMI TMDS Signal Generator Platform

Agilent E4887A HDMI TMDS Signal Generator Platform Agilent E4887A HDMI TMDS Signal Generator Platform Data Sheet Version 1.9 Preliminary E4887A- 007 E4887A- 037 E4887A- 003 Page Convenient Compliance Testing and Characterization of HDMI 1.3 Devices The

More information

Agilent N6465A emmc Compliance Test Application

Agilent N6465A emmc Compliance Test Application Agilent N6465A emmc Compliance Test Application Methods of Implementation Agilent Technologies Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any means

More information

Fast Ethernet Consortium Clause 25 PMD-EEE Conformance Test Suite v1.1 Report

Fast Ethernet Consortium Clause 25 PMD-EEE Conformance Test Suite v1.1 Report Fast Ethernet Consortium Clause 25 PMD-EEE Conformance Test Suite v1.1 Report UNH-IOL 121 Technology Drive, Suite 2 Durham, NH 03824 +1-603-862-0090 Consortium Manager: Peter Scruton pjs@iol.unh.edu +1-603-862-4534

More information

40G SWDM4 MSA Technical Specifications Optical Specifications

40G SWDM4 MSA Technical Specifications Optical Specifications 40G SWDM4 MSA Technical Specifications Specifications Participants Editor David Lewis, LUMENTUM The following companies were members of the SWDM MSA at the release of this specification: Company Commscope

More information

RIGOL. Data Sheet. DS1000B Series Digital Oscilloscopes DS1074B, DS1104B, DS1204B. Product Overview. Easy to Use Design. Applications.

RIGOL. Data Sheet. DS1000B Series Digital Oscilloscopes DS1074B, DS1104B, DS1204B. Product Overview. Easy to Use Design. Applications. RIGOL Data Sheet Product Overview DS1000B series oscilloscopes are designed with four analog channels and 1 external trigger channel, which can capture multi-channel signal simultaneously and meet industrial

More information

EAN-Performance and Latency

EAN-Performance and Latency EAN-Performance and Latency PN: EAN-Performance-and-Latency 6/4/2018 SightLine Applications, Inc. Contact: Web: sightlineapplications.com Sales: sales@sightlineapplications.com Support: support@sightlineapplications.com

More information

USB 3.1 ENGINEERING CHANGE NOTICE

USB 3.1 ENGINEERING CHANGE NOTICE Title: SSP System Jitter Budget Applied to: USB_3_1r1.0_07_31_2013 Brief description of the functional changes: Change to the 10Gbps system jitter budget. The change reduces the random jitter (RJ) budget

More information

Quick Reference Manual

Quick Reference Manual Quick Reference Manual V1.0 1 Contents 1.0 PRODUCT INTRODUCTION...3 2.0 SYSTEM REQUIREMENTS...5 3.0 INSTALLING PDF-D FLEXRAY PROTOCOL ANALYSIS SOFTWARE...5 4.0 CONNECTING TO AN OSCILLOSCOPE...6 5.0 CONFIGURE

More information

Techniques for Extending Real-Time Oscilloscope Bandwidth

Techniques for Extending Real-Time Oscilloscope Bandwidth Techniques for Extending Real-Time Oscilloscope Bandwidth Over the past decade, data communication rates have increased by a factor well over 10X. Data rates that were once 1Gb/sec and below are now routinely

More information

Agilent Parallel Bit Error Ratio Tester. System Setup Examples

Agilent Parallel Bit Error Ratio Tester. System Setup Examples Agilent 81250 Parallel Bit Error Ratio Tester System Setup Examples S1 Important Notice This document contains propriety information that is protected by copyright. All rights are reserved. Neither the

More information

32 G/64 Gbaud Multi Channel PAM4 BERT

32 G/64 Gbaud Multi Channel PAM4 BERT Product Introduction 32 G/64 Gbaud Multi Channel PAM4 BERT PAM4 PPG MU196020A PAM4 ED MU196040A Signal Quality Analyzer-R MP1900A Series Outline of MP1900A series PAM4 BERT Supports bit error rate measurements

More information

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment Integrated Component Options Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment PRELIMINARY INFORMATION SquareGENpro is the latest and most versatile of the frequency

More information

Measurement User Guide

Measurement User Guide N4906 91040 Measurement User Guide The Serial BERT offers several different kinds of advanced measurements for various purposes: DUT Output Timing/Jitter This type of measurement is used to measure the

More information

ASNT_PRBS20B_1 18Gbps PRBS7/15 Generator Featuring Jitter Insertion, Selectable Sync, and Output Amplitude Control

ASNT_PRBS20B_1 18Gbps PRBS7/15 Generator Featuring Jitter Insertion, Selectable Sync, and Output Amplitude Control ASNT_PRBS20B_1 18Gbps PRBS7/15 Generator Featuring Jitter Insertion, Selectable Sync, and Output Amplitude Control Broadband frequency range from 20Mbps 18.0Gbps Minimal insertion jitter Fast rise and

More information

SDS1000C Specifications

SDS1000C Specifications SDS1000C Specifications File Version V1.2 Siglent technology Co., Ltd CHARACTERISTIC: The highest Single real-time sampling rate can be up to 500MHzsa/s; Equivalent sampling rate is up to 10GSa/s. Memory

More information

Benefits of the R&S RTO Oscilloscope's Digital Trigger. <Application Note> Products: R&S RTO Digital Oscilloscope

Benefits of the R&S RTO Oscilloscope's Digital Trigger. <Application Note> Products: R&S RTO Digital Oscilloscope Benefits of the R&S RTO Oscilloscope's Digital Trigger Application Note Products: R&S RTO Digital Oscilloscope The trigger is a key element of an oscilloscope. It captures specific signal events for detailed

More information

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s ET 150 Scope Controls Learning Objectives In this lesson you will: learn the location and function of oscilloscope controls. see block diagrams of analog and digital oscilloscopes. see how different input

More information

WaveSurfer Xs-A and MXs-A Oscilloscopes

WaveSurfer Xs-A and MXs-A Oscilloscopes WaveSurfer Xs-A and MXs-A Oscilloscopes WAVESURFER XS A AND MXS A SERIES OSCILLOSCOPES Getting Started Manual MARCH, 2009 LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY 10977 6499 Tel: (845)

More information

PicoScope 6407 Digitizer

PicoScope 6407 Digitizer YE AR PicoScope 6407 Digitizer HIGH PERFORMANCE USB DIGITIZER Programmable and Powerful 1 GHz bandwidth 1 GS buffer size 5 GS/s real-time sampling Advanced digital triggers Built-in function generator

More information

PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX

PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX w w w. m e n t o r. c o m PCIe: Eye Diagram Analysis in HyperLynx PCI Express Tutorial This PCI Express tutorial will walk you through time-domain eye diagram analysis

More information

EDL8 Race Dash Manual Engine Management Systems

EDL8 Race Dash Manual Engine Management Systems Engine Management Systems EDL8 Race Dash Manual Engine Management Systems Page 1 EDL8 Race Dash Page 2 EMS Computers Pty Ltd Unit 9 / 171 Power St Glendenning NSW, 2761 Australia Phone.: +612 9675 1414

More information

R&S RT-ZF20 Power Deskew Fixture User Manual

R&S RT-ZF20 Power Deskew Fixture User Manual R&S RT-ZF20 Power Deskew Fixture User Manual (B00X2) User Manual Test & Measurement 1800.0040.02 04 This manual describes the following R&S RT-ZF models: R&S RT-ZF20 (1800.0004.01) 2016 Rohde & Schwarz

More information

DisplayPort Standard. Agilent, Inc. Draft2 January 14, 2013

DisplayPort Standard. Agilent, Inc. Draft2 January 14, 2013 Agilent, Inc. DisplayPort Standard Draft2 January 14, 2013 Agilent MOI for DisplayPort PHY CTS 1.2b Source Testing Using DSA90000A/90000X/90000Q Series Oscilloscopes with U7232B DisplayPort Compliance

More information

2 AORM Setup & View Wizard

2 AORM Setup & View Wizard 2 AORM Setup & View Wizard Advanced ORM Setup and View Wizard The Advanced ORM package provides a Setup and View wizard to simplify setup of the most common AORM parameters and processing functions. The

More information

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

NanoGiant Oscilloscope/Function-Generator Program. Getting Started Getting Started Page 1 of 17 NanoGiant Oscilloscope/Function-Generator Program Getting Started This NanoGiant Oscilloscope program gives you a small impression of the capabilities of the NanoGiant multi-purpose

More information

Advanced Troubleshooting with Oscilloscopes 9000 Scope Hands-on Labs

Advanced Troubleshooting with Oscilloscopes 9000 Scope Hands-on Labs Advanced Troubleshooting with Oscilloscopes 9000 Scope Hands-on Labs Page Lab 1: Scope-based Protocol Analysis 2 Lab 2: Measurements & Analysis 10 Lab 3: InfiniiScan Zone-qualified Triggering 19 Lab 4:

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

FCPM-6000RC. Mini-Circuits P.O. Box , Brooklyn, NY (718)

FCPM-6000RC. Mini-Circuits  P.O. Box , Brooklyn, NY (718) USB / Ethernet Integrated Frequency Counter & Power Meter 50Ω -30 dbm to +20 dbm, 1 MHz to 6000 MHz The Big Deal Automatically synchronized power & frequency measurements USB and Ethernet control Includes

More information

2016 RIGOL TECHNOLOGIES, INC.

2016 RIGOL TECHNOLOGIES, INC. RIGOL Data Sheet Product Overview DS1000B series oscilloscopes are designed with four analog channels and 1 external trigger channel, which can capture multi-channel signal simultaneously and meet industrial

More information

Agilent Technologies 54522A

Agilent Technologies 54522A Agilent Technologies 54522A Data Sheet Product Specifications General Specifications Maximum Sample Rate 54522A 2 GSa/s Number of Channels (all are simultaneous acquisition) 54522A: 2 Record Length 32,768

More information

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator 20 Channel Digital Delay Generator Features of the 745T-20C: 20 Independent delay channels - 100 ps resolution - 25 ps rms jitter - 10 second range Output pulse up to 6 V/50 Ω Independent trigger for every

More information

Broadcast Television Measurements

Broadcast Television Measurements Broadcast Television Measurements Data Sheet Broadcast Transmitter Testing with the Agilent 85724A and 8590E-Series Spectrum Analyzers RF and Video Measurements... at the Touch of a Button Installing,

More information

TDSDVI DVI Compliance Test Solution

TDSDVI DVI Compliance Test Solution Online Help TDSDVI DVI Compliance Test Solution 077-0022-04 Adapted from TDSDVI Compliance Test Solution Online Help www.tektronix.com Copyright Tektronix. All rights reserved. Licensed software products

More information

GFT Channel Digital Delay Generator

GFT Channel Digital Delay Generator Features 20 independent delay Channels 100 ps resolution 25 ps rms jitter 10 second range Output pulse up to 6 V/50 Ω Independent trigger for every channel Fours Triggers Three are repetitive from three

More information

B. The specified product shall be manufactured by a firm whose quality system is in compliance with the I.S./ISO 9001/EN 29001, QUALITY SYSTEM.

B. The specified product shall be manufactured by a firm whose quality system is in compliance with the I.S./ISO 9001/EN 29001, QUALITY SYSTEM. VideoJet 8000 8-Channel, MPEG-2 Encoder ARCHITECTURAL AND ENGINEERING SPECIFICATION Section 282313 Closed Circuit Video Surveillance Systems PART 2 PRODUCTS 2.01 MANUFACTURER A. Bosch Security Systems

More information

JD725A Cable and Antenna Analyzer - Dual Port

JD725A Cable and Antenna Analyzer - Dual Port COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS JD725A Cable and Antenna Analyzer - Dual Port Key Features Portable and lightweight handheld instrument Built-in wireless frequency bands as well as the most

More information

20 GHz bandwidth 17.5 ps rise time

20 GHz bandwidth 17.5 ps rise time PicoScope 9300 Series THE NEW FACE OF SAMPLING OSCILLOSCOPES 20 GHz bandwidth 17.5 ps rise time A LONG LIST OF FEATURES 1 MS/s sequential sampling - the industry s fastest NRZ and RZ eye plots and measurements

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

WAVERUNNER 104XI AND 204XI

WAVERUNNER 104XI AND 204XI WAVERUNNER 104XI AND 204XI Leading Features 1 and 2 GHz Bandwidths 5 GS/s Sample Rate (10 GS/s interleaved) 10 Mpts/Ch standard Large Display and Small (6" deep) footprint Most complete problem solving

More information

PicoScope 6407 Digitizer

PicoScope 6407 Digitizer YE AR HIGH PERFORMANCE USB DIGITIZER Programmable and Powerful 1 GHz bandwidth 1 GS buffer size 5 GS/s real-time sampling Advanced digital triggers Built-in function generator USB-connected Signals Analysis

More information

Choosing an Oscilloscope

Choosing an Oscilloscope Choosing an Oscilloscope By Alan Lowne CEO Saelig Company (www.saelig.com) Post comments on this article at www.nutsvolts.com/ magazine/article/october2016_choosing-oscilloscopes. All sorts of questions

More information

C8491 C8000 1/17. digital audio modular processing system. 3G/HD/SD-SDI DSP 4/8/16 audio channels. features. block diagram

C8491 C8000 1/17. digital audio modular processing system. 3G/HD/SD-SDI DSP 4/8/16 audio channels. features. block diagram features 4 / 8 / 16 channel LevelMagic2 SDI-DSP with level or loudness (ITU-BS.1770-1/ ITU-BS.1770-2, EBU R128) control 16 channel 3G/HD/SD-SDI de-embedder 16 in 16 de-embedder matrix 16 channel 3G/HD/SD-SDI

More information

Dual Link DVI Receiver Implementation

Dual Link DVI Receiver Implementation Dual Link DVI Receiver Implementation This application note describes some features of single link receivers that must be considered when using 2 devices for a dual link application. Specific characteristics

More information

SM02. High Definition Video Encoder and Pattern Generator. User Manual

SM02. High Definition Video Encoder and Pattern Generator. User Manual SM02 High Definition Video Encoder and Pattern Generator User Manual Revision 0.2 20 th May 2016 1 Contents Contents... 2 Tables... 2 Figures... 3 1. Introduction... 4 2. acvi Overview... 6 3. Connecting

More information

EEG A1452 SCTE-104 Inserter Frame Card

EEG A1452 SCTE-104 Inserter Frame Card EEG A1452 SCTE-104 Inserter Frame Card Product Manual EEG Enterprises, Inc. 586 Main Street Farmingdale, New York 11735 TEL: (516) 293-7472 FAX: (516) 293-7417 Copyright EEG Enterprises, Inc. 2017 All

More information

WaveSurfer MXs. The Specs You Want The Features You Need 200 MHz to 1 GHz

WaveSurfer MXs. The Specs You Want The Features You Need 200 MHz to 1 GHz WaveSurfer MXs The Specs You Want The Features You Need 200 MHz to 1 GHz The Essential Tools for Efficient Validation and Debug Leading Features Bandwidths from 200 MHz to 1 GHz Up to 5 GS/s Long Capture

More information

SAT IF distribution system

SAT IF distribution system 7. Technical specifications Type cs43 RF input frequency range pr. 50-350 MHz inputs number 4 level pr. 55...88 dbµv 60...93 dbµv symbol rate 3 45 Ms/s return loss/impedance > 0 db/75 Ω LNB powering/control

More information

Serial ATA International Organization

Serial ATA International Organization Serial ATA International Organization Version 1.0 September 27, 2007 Serial ATA Interoperability Program Revision 1.2 Tektronix MOI for RSG Tests (Using AWG7102 and CHS Frame Error Analyzer) This document

More information

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT /12/14 BIT 10 TO 105 MSPS ADC

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT /12/14 BIT 10 TO 105 MSPS ADC LTC2280, LTC2282, LTC2284, LTC2286, LTC2287, LTC2288 LTC2289, LTC2290, LTC2291, LTC2292, LTC2293, LTC2294, LTC2295, LTC2296, LTC2297, LTC2298 or LTC2299 DESCRIPTION Demonstration circuit 851 supports a

More information