1 The Big Picture of Quality MRI Quality Assurance and ACR MRI Accreditation Program Chen Lin, PhD Indiana University School of Medicine & Clarian Health Partners Diagnosis accuracy Image quality Knowledge and expertise of radiologists Effective communication Workflow and efficiency Safety Satisfaction of patients and referring physicians Basic Elements for Image Quality Other Contributions to Image Quality 2D SE 60 2D T1 FLAIR 2D FLASH 3D FLASH 3D MPRAGE IQ=4 IQ=6 IQ=4 IQ=5 IQ=5 Signal, Spatial Resolution, Contrast 1. Signal 2. Spatial Resolution 3. Contrast 4. Artifacts 5. Noise 6. Temporal resolution 7. Scan time 8. Temporal Stability or Reproducibility 9. MRS line-width and baseline Common MRI Artifacts Noise in MRI Magnetic field (B 0 ) 1. Main field inhomogeneity 2. Susceptibility artifacts RF Field (B 1 ) 1. Cross-talk (Imperfect slice profile) 2. Zipper (RF leak) 3. RF feed through 4. RF noise/interference 5. Inhomogeneity (intensity and suppression) Data Acquisition and Image Reconstruction 1. Aliasing 2. Truncation (Gibbs Ringing) 3. Blurring 4. Chemical shift 5. Partial volume Physiology / Patient 1. Bulk Motion 2. Flow Originated from the patient / object. Also introduced by the hardware components in the receiving chain/path. Thermal/random noise has the same characteristic over a wide frequency range, i.e. White noise Noise measurement: Standard deviation of pixel intensity in noise region. Signal Intensity Noise Level Gradients (G) 1. Ghosting 2. Distortion or Non-linearity rbw frequency 1

2 Signal to Noise Ratio (SNR) Signal: Average pixel intensity for a specific region/tissue SNR = Signal / Noise SNR = f(seq) B 0 PD x y z ( N PE NEX / RBW ) 1/2 Contrast to Noise Ratio (CNR) Contrast: Signal Tissue1 Signal Tissue2 CNR = Contrast/Noise Better measure of conspicuity SNR 8 SNR 32 = 0.5 SNR 8 SNR 128 = 0.25 SNR 8 Use consistent imaging parameters for SNR measurement and comparison. Artifact to Noise Ratio Artifact: Average of pixel intensity in an artifact region Major Influences to Image Quality 1. Scanner performance, imaging technique & protocol MRCP w. 3D SPACE & Var FA MRCP w. 3D SPACE & Const FA 2. Patient management 3. Exam planning (Slice position, orientation and coverage; Acquisition timing) 4. Image presentation (Film and Display Quality) Image Quality Assurance Scanner Performance Monitoring Acceptance Testing after Installation, Upgrade and Service Scheduled Daily/Weekly QA Scheduled Quarterly/Annually QA Test all coils, RF room integrity, PMU, Display, Injector, etc. Preventive Maintenance Testing for Advanced Applications Protocol Optimization and Standardization Tech Education & Training Image Quality/Artifacts Reporting and Troubleshooting (Continuous Improvement Process) 2/3/2006 ACR QA Program Personnel Qualification Preventive Maintenance and Safety Weekly QA by MR Technologists: Phantom scans, visual checks, film test Acceptance and Annual Testing by Physicist Re-accreditation Every Three Years: Attestation of weekly and annual QA Submit expanded phantom scans Submit clinical cases ACR Weekly QA Tests Visual Checks Patient table RF shield Patient monitoring Safety Equipment Film QA Base + Fog Speed Contrast 2

3 Features in ACR Phantom Appropriate size, shape, loading and relaxation time. Stable for over time An uniform region Grid for geometric accuracy Wedges for slice position measurement Ramps for slice thickness measurement Shallow holes for low contrast detectability measurement Small holes for spatial resolution measurement ACR Phantom Appropriate for most head coils Phantom Setup and slice positioning Straight and in the center of the coil. Use a customized holder or the same support. Line up slice #1 and #11 to the cross of wedges ACR Weekly QA Protocol CSE sequence, TR/TE = 500ms/20ms Acq. matrix = 256x256 and FOV = 250mm 11 slice of 5mm w. 100% gap What about other imaging options? None, except image intensity correction if phase array coil is used. (for accreditation) Normalize or Pre-scan Norm (Siemens) SCIC or PURE (GE) 1. Central Frequency (CF) Typically set to the resonance frequency of water protons. Proportional to the field strength (B 0 ). Search for the largest water signal using spectroscopy sequence with 10Hz increment. Affects the quality of fat saturation, the severity of off-resonance artifacts, etc. Finding Center Frequency Display as one of the calibration / pre-scan parameters. Can also be found in DICOM header: DICOM TAG (0018, 0084) Imaging Frequency ACR recommended criteria: < +/-1.5ppm different between weekly tests Possible cause of failure: Phantom not at iso-center Magnet drift RF (synthesizer) instability 2. Transmitter Gain/Voltage Reflects RF power required to achieve certain flip angle Depends on RF coil (the coupling), phantom (coil load), RF pulse type, etc. Should remain constant over time if nothing in the pulse sequence or hardware has changed. 3

4 184.6 mm Accurate Image 4/14/2009 Finding Transmitter Gain/Voltage Criteria and Causes of Failure GE: displayed on screen (db) Siemens: options >> adjustments Transmitter amplitude (temp) (V) Philips: under system performance parameters rf_act_drivescale Toshiba: acquisition window Center frequency (MHz) RF level Recommended criteria: 10% increase from baseline. Possible cause of failure: Incorrect positioning of phantom. Defect in RF transmitting coil (such as improper quadrature phase shift). RF (amplifier) instability. 3. Geometric Accuracy Gradient Nonlinearity Directly related to the image distortion, affect the accuracy of distance and area measurement. Especially important for applications such as volumetric imaging, image co-registration, imagebased stereotactic procedures. Frequency Linear Gradient Non-linear Gradient X (cm) Object Geometric Accuracy Measurement Criteria and Causes of Failure mm mm ACR Recommended Criteria: +/- 2mm in all directions Possible cause of failure: Gradient non-linearity and miss-calibration Field inhomogeneity and low receiver bandwidth Eddy currents Inappropriate phantom positioning Measure distances along main axes of phantom Compare with known values 4

5 4. Spatial Resolution Measurement Ability to detect small structures. Depends on FOV and acquisition matrix size But also depends on: pulse sequence type relaxation times k-space sampling filter settings etc. Spatial Resolution Measurement Resolution Grid Method UL LR Evaluate the separation of holes arranged in two square arrays. The resolution is the smallest size of holes can be separated. Prevent Partial Volume Error Resolution Holes Image Matrix Criteria and Causes of Failure ACR Recommended Criteria: <= 1.0 mm Possible cause of failure: Blurred by image filter. Interference from artifacts Poor SNR Increased eddy currents Poor gradient calibration Poor B 0 uniformity Reduced stability of system 5. Low Contrast Detectability (LCD) The ability to distinguish objects with subtle difference in signal intensity. CNR = S1 S2 /(St. Dev. of Noise) LCD: A visual test. Related to SNR, but more subjective. Imaging Slice LCD Measurement Holes of different size Plastic Sheet of different thickness 5

6 Total Number of Spokes 4/14/2009 The Relationship between SNR and LCD Signal-to-Noise Ratio Criteria and Cause of Failure ACR Recommend Criteria: More than 9 for 1.5T or lower More than 37 for 3.0T Reduction of 4 spokes from the baseline Possible Cause of Failure: Inaccurate slice position Poor SNR Artifacts Expanded Phantom Scans 1. ACR Sag Loc 2. ACR Ax T1 3. ACR Ax PD/T2 4. Site Brain Ax T1 5. Site Brain Ax T2 Weekly Phantom Scans 11 slices of 5 skip 5 Complete the Phantom Site Scanning Data Form (Large Phantom) online. Additional Phantom Tests Image Uniformity Use imaging intensity correction option with a phased array coil. Slice thickness and slice offset Be careful with your phantom position Ghosting artifact Checked in a recent PM or annual testing Ask your physicist to check the phantom images before ACR submission. ACR MRI Clinical Modules Head Spine MSK Body MRA Cardiac Choice of Basic and Specialty exams Each scanner must apply for all modules routinely performed ACR MRI Evaluation Categories Pulse Sequences and Image Contrast Spatial and Temporal Resolution Exam Identification: Missing Information Anatomic Coverage and Imaging Planes Artifacts Filming Technique (for hard copy film submissions only) Need good planning and team work. 6

7 ACR Accreditation Preparation ACR Online Submission Refill ACR phantom Perform PM and annual tests. Review your clinical protocols and make sure they meet ACR requirements (i.e. the resolution and scan time.) Register online and choose the type clinical cases carefully according to your patient population. Timeline for Clinical cases and phantom scans Phantom Scans 1 Notified by ACR Clinical cases (45 days) Phantom Scans 2 Material Due (45 days) Extension I (30 days) Extension II (30 days) Application Due (60 days) Tips and Suggestions Discuss with ACR if you are not clear about what they ask for. Acquisition matrix: frequency encoding and phase encoding used Reconstruction matrix: number of pixels in each dimension of the image Display matrix: number of pixels in each dimension of the post-processed image Tips and Suggestions QA for Advanced Applications Pay attention to anatomical coverage. Check image quality and present your best case. Submit all the required series including the localizer with cross-reference, but only what s required. Readable, but too much noise fmri QA Spectroscopy QA Flow Quantification QA Ask your radiologist to review the clinical images before ACR submission. 7

8 AU 4/14/2009 Challenges of fmri fmri QA Measurements Based on signal difference between rest and active states. Such difference (i.e. BOLD effect) is only a few %. While problems such as baseline drift can be corrected in postprocessing, it should still be kept to minimum Time Run extended fmri protocol with low and high excitation flip angles. Measure : SNR (RF check) SGR (Signal to Ghost Ratio) (Gradient & eddy current check) Signal intensity drifts and fluctuations. (System stability check) * A. Simmons, et al, MRM 41: (1999) fmri QA Example fmri QA Results of Two GE 3T Scanner Variation (%) Ghosting (%) Noise SNR SB w. PA < SB w. PA & IFIS SB w. BC LB w. PA LB w. BC < < < < SB=Short Bore; LB=Long Bore; PA=Phased Array; BC=Bird Cage Possible Cause of Failure Instability of RF amplifier and receiver Changes of eddy currents due to temperature change Summary Contributing factors in MR image quality. Principles behind the ACR weekly tests. ACR accreditation process. fmri QA 8

9 Thank You! 9

MR Accreditation Programs - E. Jackson

MR Accreditation Programs - E. Jackson MRI Accreditation Programs: An Overview of Each and Specifics of One Edward F. Jackson, PhD Department of Imaging Physics 1 Diagnostic - MRI Safety and Accreditation Educational Objectives At the conclusion

More information

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program 1 Contents 0.0 INTRODUCTION 4 0.1 Overview and Purpose 4 0.2 The Phantom 4 0.3 The Required Images 5 0.4 The Image

More information

Breast MR Imaging and Quality Control

Breast MR Imaging and Quality Control Breast MR Imaging and Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives 1. Provide an overview of breast MR imaging and MR-guided biopsy procedures. 2.

More information

Nuclear Associates and

Nuclear Associates and Nuclear Associates 76-907 and 76-908 AAPM MRI Phantoms Users Manual March 2005 Manual No. 38616 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia M R I Physics Course chapter 12 Artifacts and Suppression Techniques Artifacts

More information

Abstract. Learning Objectives 8/1/2017

Abstract. Learning Objectives 8/1/2017 SAM Practical Medical Physics TU-B-201-0 AAPM Annual Meeting 2017 1 Abstract This course will teach the participant to identify common artifacts found clinically in MR, DR, CT, PET, to determine the causes

More information

Procedure Manual for MRI of the Brain

Procedure Manual for MRI of the Brain Baxter Protocol 161003 SYN RC W H E R E S C I E N C E M E E T S S E R V I C E Baxter Protocol 161003 A Phase 3 Randomized, Double-Blind, Placebo-Controlled Study of the Safety and Effectiveness of Immune

More information

True comfort and flexibility with the power of 3T.

True comfort and flexibility with the power of 3T. True comfort and flexibility with the power of 3T. With a large 71 cm aperture and the quietest exams in the industry, the Vantage Titan 3T is the most comfortable 3T MRI system for all of your patients.

More information

Experiences in ACR MRI Accreditation Vendor Nuances That Every Clinical MRI Physicist Should Know

Experiences in ACR MRI Accreditation Vendor Nuances That Every Clinical MRI Physicist Should Know Experiences in ACR MRI Accreditation Vendor Nuances That Every Clinical MRI Physicist Should Know By Kathryn (Kat) W. Huff, M.S., DABR Prepared for The 2013 AAPM Spring Clinical Meeting Validating Me I

More information

Troubleshooting Guide. Prep, Scan Errors, and Artifacts

Troubleshooting Guide. Prep, Scan Errors, and Artifacts Troubleshooting Guide Prep, Scan Errors, and Artifacts Preparing for the Study Participant Compliance with MRI Scans Participant Prep Having your participant prepped will allow you to run your study with

More information

RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc

RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc Outline : Spine echo pulse sequence SE Fast spin echo pulse sequence FSE Inversion recovery pulse sequence IR Gradient pulse sequence GS Pulse

More information

Premium 1.5T MRI System

Premium 1.5T MRI System DISCLAIMER Some products and features described here are optional and not commercially available in all countries. We cannot guarantee that the system and all of options are available in all area due to

More information

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017 Display Quality Assurance: Considerations When Establishing a Display QA Program Mike Silosky, M.S. 8/3/2017 Objectives and Outline Why, Who, What, When, Where? Discuss the resources that may be needed

More information

Peacefully quiet. Remarkably fast.

Peacefully quiet. Remarkably fast. Peacefully quiet. Remarkably fast. 2 Excellent image quality Streamlined workflow Outstanding patient comfort Toshiba Medical s Vantage Galan 3T offers a transformational experience for you and your patients

More information

Philips Site Yearly Performance Evaluation Philips Achieva - Gibbons 1.5T 1-Jun-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Achieva - Gibbons 1.5T 1-Jun-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Achieva Gibbons.T Jun8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

Technologist Quality Control Procedures

Technologist Quality Control Procedures Technologist Quality Control Procedures The specific procedures for the Technologist Quality Control Program are those specified in the most current ACR MRI QC Manual. WEEKLY MRI EQUIPMENT QUALITY CONTROL

More information

EPI. Thanks to Samantha Holdsworth!

EPI. Thanks to Samantha Holdsworth! EPI Faster Cartesian approach Single-shot, Interleaved, segmented, half-k-space Delays, etc -> Phase corrections Flyback EPI GRASE Thanks to Samantha Holdsworth! 1 EPI: Speed vs Distortion Fast Spin Echo

More information

MSK Imaging Fundamentals

MSK Imaging Fundamentals MSK Imaging Fundamentals Goals Improve image quality Provide best possible product for our customers Patient Referring clinician Radiologist Reduce number of callback cases Goal reduction of 50% in 3 months

More information

Peacefully quiet. Remarkably fast.

Peacefully quiet. Remarkably fast. Peacefully quiet. Remarkably fast. Excellent image quality Streamlined workflow Outstanding patient comfort Canon Medical Systems Vantage Galan 3T offers a transformational experience for you and your

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

FFDM Quality Control in Canada - a Vendor Neutral Approach

FFDM Quality Control in Canada - a Vendor Neutral Approach FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011

More information

PATIENT POSITION IMAGING PARAMETERS

PATIENT POSITION IMAGING PARAMETERS 3 PLANE LOC Patient Entry Feet First Imaging Mode 2D Patient Position Prone Pulse Sequence Gradient Echo Coil Configuration 7breast both MRI Imaging Options Seq, Fast Plane 3-PLANE Acceleration Factor

More information

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline Medical Display Performance and Evaluation Mike Silosky, MS University of Colorado, School of Medicine Dept. of Radiology 1 Objectives Review display function, QA metrics, procedures, and guidance provided

More information

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. RSNA 2006 November 26 to December 1 Chicago Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. Once again, more than 60,000 participants (including professional

More information

Request for Proposals

Request for Proposals Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete

More information

Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1

Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1 Multi echo Multi slice (MEMS) High Performance fmri at CFMRI Table of Contents Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1 Introduction... 2 MEMS Protocols... 4 Run MEMS protocol...

More information

Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory

Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory Jacob Matthews 4/13/2012 Supervisor: Rhodri Cusack, PhD Assistance: Annika Linke,

More information

Procedures for conducting User QA on the scanner

Procedures for conducting User QA on the scanner Procedures for conducting User QA on the scanner Sample setup: The User QA phantom is clearly labeled and is stored on one of the shelves to the side of the magnet. Note proper orientation of the bottle,

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 1 TG270 Goals Provide an update

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 TG270 Goals Provide an update to

More information

CT Numbers: Think of a number, double it, add 20, divide by 4. Jane Edwards Royal Free Hospital, London

CT Numbers: Think of a number, double it, add 20, divide by 4. Jane Edwards Royal Free Hospital, London CT Numbers: Think of a number, double it, add 20, divide by 4 Jane Edwards Royal Free Hospital, London Background 2 different manufacturers CT scanners available on site: GE Lightspeed Plus; Philips Brilliance

More information

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

More Info at Open Access Database  Process Control for Computed Tomography using Digital Detector Arrays Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18082 Process Control for Computed

More information

2012 Computed Tomography

2012 Computed Tomography 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section

More information

2017 Computed Tomography

2017 Computed Tomography 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Qualified Medical Physicist s Section 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist

More information

NMR. picospin. Maintenance Guide

NMR. picospin. Maintenance Guide NMR picospin Maintenance Guide 269-302600 Revision A January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Thermo Fisher Scientific 5225 Verona

More information

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually Manufacturer s Recommendations for Alternate Dental CBCT QA Program VaTech: Model PAX i3d, i3d Green, i3d Smart Table 6 Medical Physicist s Computed Tomography QC Survey Required Test or Item Procedure

More information

Guidelines for Assuring Softcopy Image Quality

Guidelines for Assuring Softcopy Image Quality Guidelines for Assuring Softcopy Image Quality What s inside? Quality Control Guidelines Softcopy QA testing and frequencies Danny Deroo Product and R&D Manager QA Products ABSTRACT To ensure diagnostic

More information

DVM-3000 Series 12 Bit DIGITAL VIDEO, AUDIO and 8 CHANNEL BI-DIRECTIONAL DATA FIBER OPTIC MULTIPLEXER for SURVEILLANCE and TRANSPORTATION

DVM-3000 Series 12 Bit DIGITAL VIDEO, AUDIO and 8 CHANNEL BI-DIRECTIONAL DATA FIBER OPTIC MULTIPLEXER for SURVEILLANCE and TRANSPORTATION DVM-3000 Series 12 Bit DIGITAL VIDEO, AUDIO and 8 CHANNEL BI-DIRECTIONAL FIBER OPTIC MULTIPLEXER for SURVEILLANCE and TRANSPORTATION Exceeds RS-250C Short-haul and Broadcast Video specifications. 12 Bit

More information

Performance Evaluation of Industrial Computed Radiography Image Display System

Performance Evaluation of Industrial Computed Radiography Image Display System Performance Evaluation of Industrial Computed Radiography Image Display System More info about this article: http://www.ndt.net/?id=21169 Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and

More information

Digital BPMs and Orbit Feedback Systems

Digital BPMs and Orbit Feedback Systems Digital BPMs and Orbit Feedback Systems, M. Böge, M. Dehler, B. Keil, P. Pollet, V. Schlott Outline stability requirements at SLS storage ring digital beam position monitors (DBPM) SLS global fast orbit

More information

Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25

Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25 3Plane Loc SSFSE Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25 Save Series Scan After acquisition, scroll

More information

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview ACR Topics Requirements Today

More information

A COMPUTERIZED SYSTEM FOR THE ADVANCED INSPECTION OF REACTOR VESSEL STUDS AND NUTS BY COMBINED MULTI-FREQUENCY EDDY CURRENT AND ULTRASONIC TECHNIQUE

A COMPUTERIZED SYSTEM FOR THE ADVANCED INSPECTION OF REACTOR VESSEL STUDS AND NUTS BY COMBINED MULTI-FREQUENCY EDDY CURRENT AND ULTRASONIC TECHNIQUE More Info at Open Access Database www.ndt.net/?id=18566 A COMPUTERIZED SYSTEM FOR THE ADVANCED INSPECTION OF REACTOR VESSEL STUDS AND NUTS BY COMBINED MULTI-FREQUENCY EDDY CURRENT AND ULTRASONIC TECHNIQUE

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

Image quality in non-gated versus gated reconstruction of tongue motion using Magnetic Resonance Imaging:

Image quality in non-gated versus gated reconstruction of tongue motion using Magnetic Resonance Imaging: This talk was presented 26 June 2008, at the 22nd International Congress and Exhibition of Computer Assisted Radiology and Surgery, in Barcelona at the Hotel Constanza from June 25 to 28, 2008. See http://kochanski.org/gpk/papers/2008/carstalk.html

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

R&S EDS300 DME/Pulse Analyzer Specifications

R&S EDS300 DME/Pulse Analyzer Specifications R&S EDS300 DME/Pulse Analyzer Specifications year Data Sheet Version 04.01 CONTENTS Definitions... 3 Specifications... 4 Frequency... 4 Level... 4 DME signal analysis... 4 TACAN signal analysis (R&S EDS-K1

More information

Release Notes. Multi-Band EPI C2P. Release February 2013

Release Notes. Multi-Band EPI C2P. Release February 2013 Release Notes Multi-Band EPI C2P Release 008 13 February 2013 Installation 1. Restart the system (reboot host and MRIR) 2. Extract the.zip file to a temporary directory 3. Run the installer.bat file 4.

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

New Filling Pattern for SLS-FEMTO

New Filling Pattern for SLS-FEMTO SLS-TME-TA-2009-0317 July 14, 2009 New Filling Pattern for SLS-FEMTO Natalia Prado de Abreu, Paul Beaud, Gerhard Ingold and Andreas Streun Paul Scherrer Institut, CH-5232 Villigen PSI, Switzerland A new

More information

RF considerations for SwissFEL

RF considerations for SwissFEL RF considerations for H. Fitze in behalf of the PSI RF group Workshop on Compact X-Ray Free Electron Lasers 19.-21. July 2010, Shanghai Agenda Introduction RF-Gun Development C-band development Summary

More information

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications March 2006 Agilent multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical

More information

Understanding CT image quality

Understanding CT image quality IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Understanding CT image quality Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office Role of

More information

The FLASH objective: SASE between 60 and 13 nm

The FLASH objective: SASE between 60 and 13 nm Injector beam control studies winter 2006/07 talk from E. Vogel on work performed by W. Cichalewski, C. Gerth, W. Jalmuzna,W. Koprek, F. Löhl, D. Noelle, P. Pucyk, H. Schlarb, T. Traber, E. Vogel, FLASH

More information

PINMRF. Checkout Quiz - Varian Inova-300 Version

PINMRF. Checkout Quiz - Varian Inova-300 Version PINMRF Checkout Quiz - Varian Inova-300 Version Please carefully read every question and select the best answer(s). Some questions may have more than one correct answer. You must select all the correct

More information

User's Manual. Rev 1.0

User's Manual. Rev 1.0 User's Manual Rev 1.0 Digital TV sales have increased dramatically over the past few years while the sales of analog sets are declining precipitously. First quarter of 2005 has brought the greatest volume

More information

Super Idea for Ultrasonic Inspection

Super Idea for Ultrasonic Inspection Super Idea for Ultrasonic Inspection Ultrasound Inspection Ultrasound phased array (PA) and Time Of Flight Diffraction (TOFD), two of the new NDT technologies, have become one important development trend

More information

Role of Color in Telemedicine Applications. Elizabeth A. Krupinski, PhD

Role of Color in Telemedicine Applications. Elizabeth A. Krupinski, PhD Role of Color in Telemedicine Applications Elizabeth A. Krupinski, PhD Background Color displays common clinical practice Radiology growing acceptance & use Other ologies & telemed routinely used No validated

More information

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series Introduction System designers and device manufacturers so long have been using one set of instruments for creating digitally modulated

More information

Minimize your cost for Phased Array & TOFD

Minimize your cost for Phased Array & TOFD Minimize your cost for Phased Array & TOFD Latest ultrasonic flaw detector from SIUI, incorporates the latest advancements in Encoder In/Out UT/ TOFD Probe high-performance Phased Array and TOFD detection

More information

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? White Paper Uniform Luminance Technology What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? Tom Kimpe Manager Technology & Innovation Group Barco Medical Imaging

More information

Practicum 3, Fall 2010

Practicum 3, Fall 2010 A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization

More information

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018)

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018) The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018) Table of Contents ACR Mammography Accreditation... 1 The 1999 ACR Mammography

More information

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications Technical Specifications Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only for detailed Technical Specifications visit www.radimage.com Can I get just the right

More information

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications 100 Hz to 26.5 GHz The HP 71910A/P is a receiver for monitoring signals from 100 Hz to 26.5 GHz. It provides a cost effective combination

More information

TSG 90 PATHFINDER NTSC Signal Generator

TSG 90 PATHFINDER NTSC Signal Generator Service Manual TSG 90 PATHFINDER NTSC Signal Generator 070-8706-01 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless

More information

Detailed Design Report

Detailed Design Report Detailed Design Report Chapter 4 MAX IV Injector 4.6. Acceleration MAX IV Facility CHAPTER 4.6. ACCELERATION 1(10) 4.6. Acceleration 4.6. Acceleration...2 4.6.1. RF Units... 2 4.6.2. Accelerator Units...

More information

QRF5000 MDU ENCODER. Data Sheet

QRF5000 MDU ENCODER. Data Sheet Radiant Communications Corporation 5001 Hadley Road South Plainfield NJ 07080 Tel (908) 757-7444 Fax (908) 757-8666 WWW.RCCFIBER.COM QRF5000 MDU ENCODER Data Sheet Version 1.1 1 Caution Verify proper grounding

More information

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications Data Sheet Version 02.00 CONTENTS Definitions... 3 Specifications... 4 Level... 5 Result display... 6 Trigger... 7 Ordering information...

More information

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Julie Smyth & Philip Doyle Regional Medical Physics Service Outline Preamble Image Quality

More information

Monitor QA Management i model

Monitor QA Management i model Monitor QA Management i model 1/10 Monitor QA Management i model Table of Contents 1. Preface ------------------------------------------------------------------------------------------------------- 3 2.

More information

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Tube current modulation and dose reduction : How TCM works Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) This product is no longer carried in our catalog. AFG 2020 Characteristics Features Ordering Information Characteristics

More information

The Ultimate Long-term EEG Monitoring System

The Ultimate Long-term EEG Monitoring System TM The Ultimate Long-term EEG Monitoring System TM The Ultimate Long-term EEG Monitoring System The Ultimate Long-term EEG Monitoring System When the Epilepsy Monitoring Unit demands performance, Neuvo

More information

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks

More information

MultiMac SM. Eddy Current Instrument for Encircling Coil, Sector and Rotary Probe Testing of Tube, Bar, & Wire

MultiMac SM. Eddy Current Instrument for Encircling Coil, Sector and Rotary Probe Testing of Tube, Bar, & Wire MultiMac SM Eddy Current Instrument for Encircling Coil, Sector and Rotary Probe Testing of Tube, Bar, & Wire Features of the MultiMac SM Electronics Simultaneous Coil and/or Rotary Probe operation Differential

More information

LHC Beam Instrumentation Further Discussion

LHC Beam Instrumentation Further Discussion LHC Beam Instrumentation Further Discussion LHC Machine Advisory Committee 9 th December 2005 Rhodri Jones (CERN AB/BDI) Possible Discussion Topics Open Questions Tune measurement base band tune & 50Hz

More information

R&S ZVA110 Vector Network Analyzer Specifications

R&S ZVA110 Vector Network Analyzer Specifications ZVA110_dat-sw_en_5214-4813-22_cover.indd 1 Data Sheet 04.00 Test & Measurement R&S ZVA110 Vector Network Analyzer Specifications 15.11.2013 14:42:28 CONTENTS Definitions... 3 Specifications... 4 Overview...

More information

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display SR850 DSP Lock-In Amplifier 1 mhz to 102.4 khz frequency range >100 db dynamic reserve 0.001 degree phase resolution Time constants

More information

Video Signals and Circuits Part 2

Video Signals and Circuits Part 2 Video Signals and Circuits Part 2 Bill Sheets K2MQJ Rudy Graf KA2CWL In the first part of this article the basic signal structure of a TV signal was discussed, and how a color video signal is structured.

More information

2 MHz Lock-In Amplifier

2 MHz Lock-In Amplifier 2 MHz Lock-In Amplifier SR865 2 MHz dual phase lock-in amplifier SR865 2 MHz Lock-In Amplifier 1 mhz to 2 MHz frequency range Dual reference mode Low-noise current and voltage inputs Touchscreen data display

More information

Imaging diagnostico in Sanità Stato attuale e prospettive

Imaging diagnostico in Sanità Stato attuale e prospettive Imaging diagnostico in Sanità Stato attuale e prospettive Sandro Paini AMI & Oncology Pisa 20.12.2016 Fully digital SiPM (dsipm ) invented within Philips Research PMT APD(analog) SiPM(Analog) dsipm( Digital)

More information

EMI/EMC diagnostic and debugging

EMI/EMC diagnostic and debugging EMI/EMC diagnostic and debugging 1 Introduction to EMI The impact of Electromagnetism Even on a simple PCB circuit, Magnetic & Electric Field are generated as long as current passes through the conducting

More information

Perfecting the Package Bare and Overmolded Stacked Dies. Understanding Ultrasonic Technology for Advanced Package Inspection. A Sonix White Paper

Perfecting the Package Bare and Overmolded Stacked Dies. Understanding Ultrasonic Technology for Advanced Package Inspection. A Sonix White Paper Perfecting the Package Bare and Overmolded Stacked Dies Understanding Ultrasonic Technology for Advanced Package Inspection A Sonix White Paper Perfecting the Package Bare and Overmolded Stacked Dies Understanding

More information

News from Rohde&Schwarz Number 195 (2008/I)

News from Rohde&Schwarz Number 195 (2008/I) BROADCASTING TV analyzers 45120-2 48 R&S ETL TV Analyzer The all-purpose instrument for all major digital and analog TV standards Transmitter production, installation, and service require measuring equipment

More information

APPLICATION OF PHASED ARRAY ULTRASONIC TEST EQUIPMENT TO THE QUALIFICATION OF RAILWAY COMPONENTS

APPLICATION OF PHASED ARRAY ULTRASONIC TEST EQUIPMENT TO THE QUALIFICATION OF RAILWAY COMPONENTS APPLICATION OF PHASED ARRAY ULTRASONIC TEST EQUIPMENT TO THE QUALIFICATION OF RAILWAY COMPONENTS K C Arcus J Cookson P J Mutton SUMMARY Phased array ultrasonic testing is becoming common in a wide range

More information

Global Trade Medical Supplies

Global Trade Medical Supplies Features: Achieve Clarity True to the Source Data A medical monitor needs to be capable of high brightness in order to meet performance standards. However, in order to achieve high brightness in an LCD

More information

PAM-1840 Preamplifier Operation Manual

PAM-1840 Preamplifier Operation Manual PAM-1840 Preamplifier Operation Manual 1 TABLE OF CONTENTS INTRODUCTION 3 GENERAL INFORMATION 4 SPECIFICATIONS 4 OPERATING INSTRUCTIONS 5 MAINTENANCE 6 2 INTRODUCTION BEFORE APPLYING POWER Review this

More information

SM02. High Definition Video Encoder and Pattern Generator. User Manual

SM02. High Definition Video Encoder and Pattern Generator. User Manual SM02 High Definition Video Encoder and Pattern Generator User Manual Revision 0.2 20 th May 2016 1 Contents Contents... 2 Tables... 2 Figures... 3 1. Introduction... 4 2. acvi Overview... 6 3. Connecting

More information

GS Bloch Equations Simulator 1. GS Introduction to Medical Physics IV Exercise 1: Discrete Subjects

GS Bloch Equations Simulator 1. GS Introduction to Medical Physics IV Exercise 1: Discrete Subjects GS02-1193 Bloch Equations Simulator 1 GS02-1193 Introduction to Medical Physics IV Exercise 1: Discrete Subjects Once SpinWright is running, select the Subject tab. The GUI display toward the top of the

More information

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2. CMRRpack2 a collection of pulse sequences from the University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.63_package This document generated: Mon Dec 16 2013 15:55:46 Contents

More information

4 MHz Lock-In Amplifier

4 MHz Lock-In Amplifier 4 MHz Lock-In Amplifier SR865A 4 MHz dual phase lock-in amplifier SR865A 4 MHz Lock-In Amplifier 1 mhz to 4 MHz frequency range Low-noise current and voltage inputs Touchscreen data display - large numeric

More information

Achieving More Efficient Data Review with OpenLAB CDS

Achieving More Efficient Data Review with OpenLAB CDS Achieving More Efficient Data Review with OpenLAB CDS Technical Note Introduction Separations have enjoyed major advances that have significantly reduced chromatographic run times. Ultra high performance

More information

Prisma Optical Networks Ancillary Modules

Prisma Optical Networks Ancillary Modules Optoelectronics Prisma Optical Networks Ancillary Modules Description The Prisma platform is capable of utilizing a combination of modules which address a variety of revenue generating applications. The

More information

R&S FSW-K160RE 160 MHz Real-Time Measurement Application Specifications

R&S FSW-K160RE 160 MHz Real-Time Measurement Application Specifications FSW-K160RE_dat-sw_en_3607-1759-22_v0200_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S FSW-K160RE 160 MHz Real-Time Measurement Application Specifications 06.04.2016 17:16:27 CONTENTS Definitions...

More information

Keysight Technologies ad Integrated RF Test Solution

Keysight Technologies ad Integrated RF Test Solution Keysight Technologies 802.11ad Integrated RF Test Solution E7760A Wideband Transceiver M1650A mmwave Transceiver Data Sheet Introduction Design your 802.11ad device with confidence Evaluating devices at

More information

WHY CHOOSE HITACHI? * Based on Hitachi's factory shipment records, as of end of March, Others 743 EUROPE.

WHY CHOOSE HITACHI? * Based on Hitachi's factory shipment records, as of end of March, Others 743 EUROPE. WHY CHOOSE HITACHI? For more than 30 years, Hitachi has been leading the way in Open MRI. With more than 7,000 systems delivered worldwide*, Hitachi is at the forefront of MRI technology. * Based on Hitachi's

More information

Mammo Tomosynthesis 5MP

Mammo Tomosynthesis 5MP Mammo Tomosynthesis 5MP Display system for digital breast imaging Approved for TOMOSYNTHESIS MAMMOGRAPHY The standard of care For thousands of women every day, details make all the difference. This understanding

More information

Overview of All Pixel Circuits for Active Matrix Organic Light Emitting Diode (AMOLED)

Overview of All Pixel Circuits for Active Matrix Organic Light Emitting Diode (AMOLED) Chapter 2 Overview of All Pixel Circuits for Active Matrix Organic Light Emitting Diode (AMOLED) ---------------------------------------------------------------------------------------------------------------

More information

SHOWLINE SL NITRO 510 LED STROBE LUMINAIRE SPECIFICATIONS.

SHOWLINE SL NITRO 510 LED STROBE LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. SHOWLINE SL NITRO 510 LED STROBE LUMINAIRE SPECIFICATIONS. 1.) The luminaire shall be an LED strobe luminaire employing one-thousand, three-hundred and fifty (1350) white LED engines.

More information