Lesson 14 BIOFEEDBACK Relaxation and Arousal

Size: px
Start display at page:

Download "Lesson 14 BIOFEEDBACK Relaxation and Arousal"

Transcription

1 Physiology Lessons for use with the Biopac Student Lab Lesson 14 BIOFEEDBACK Relaxation and Arousal Manual Revision EDA/GSR Richard Pflanzer, Ph.D. Associate Professor Indiana University School of Medicine Purdue University School of Science J.C. Uyehara, Ph.D. Biologist BIOPAC Systems, Inc. William McMullen Vice President BIOPAC Systems, Inc. BIOPAC Systems, Inc. 42 Aero Camino, Goleta, CA (805) , Fax (805) Web: BIOPAC Systems, Inc.

2 Page 4 II. EXPERIMENTAL OBJECTIVES 1) Introduce the concept of biofeedback training for relaxation. 2) Measure levels of arousal via heart rate and electrodermal activity (EDA). Biopac Student Lab III. MATERIALS BIOPAC disposable vinyl electrodes (EL503), 3 electrodes per Subject Electrode lead (SS2L) BIOPAC EDA setup o Disposable Setup: EDA/GSR Lead (SS57L) and EDA Gelled Electrodes (EL507 x 2) o Reusable setup: EDA/GSR transducer (SS3L) and Electrode gel (GEL1) Computer system Biopac Student Lab 3.7 BIOPAC data acquisition unit (MP36, MP35, or MP30 with cable and power) IV. EXPERIMENTAL METHODS For further explanation, use the online support options under the Help Menu. A. SET UP FAST TRACK Set Up DETAILED EXPLANATION OF SET UP STEPS 1. Turn the computer ON. The desktop should appear on the monitor. If it does not appear, ask the laboratory instructor for assistance. 2. Make sure the BIOPAC MP3X unit is turned OFF. 3. Plug the equipment in as follows: EDA setup (SS3L or SS57L) CH 1 Electrode lead set (SS2L) CH 2 4. Turn the MP3X Data Acquisition Unit ON. EDA Setup to CH 1 BIOPAC MP3X unit SS3L shown or SS57L EDA/GSR Lead SS2L Electrode Lead Set 5. Select a Subject (who has not exercised within the last hour). Set Up continues Fig Subject must not have done any rigorous exercise within the last hour. For optimal adhesion, electrodes should be placed on the skin at least 5 minutes before the start of Calibration

3 Lesson 14: Biofeedback Page 5 6. Place three electrodes on the Subject (Fig. 14.2). one on right forearm (just above wrist) right forearm WHITE lead 7. Attach the electrode lead (SS2L) to the Subject following the color code (Fig. 14.3). When the electrode cable is connected properly, the LEAD II electrode configuration will be established. one on inside right leg (just above ankle bone) right leg one on inside left leg BLACK lead (just above ankle bone) (ground) Fig Fig left leg RED lead medial surface of right leg, just above the ankle bone medial surface of left leg, just above the ankle bone right anterior forearm at the wrist (same side of arm as the palm of hand). Each of the pinch connectors on the end of the electrode cable needs to be attached to a specific electrode. The electrode cables are each a different color, and you should follow Fig to ensure that you connect each cable to the proper electrode. The pinch connectors work like a small clothespin, but will only latch onto the nipple of the electrode from one side of the connector. 8. Setup the EDA. HINT For a good signal to be picked up, it helps if the subjects have a little sweat on their hands (not a lot, but enough so that their hands are not completely smooth or cold). If subjects wash their hands just prior to the recording or if they have been sitting in a cold room, then they must do something to activate the sweat glands before beginning calibration or recording. If subjects begin with colder hands, the scale will be diminished and the signal will be easily saturated once they warm up during the lesson. SS57L and EL507 Attach two EL507 electrodes to the subject s fingertips and connect the SS57L lead, as shown in Fig. 14.4a. Fig. 14.4a SS57L and EL507 Setup Set Up continues

4 Page 6 SS3LA and GEL1 Fill both cavities of the EDA/GSR transducer (SS3L/SS3LA) with gel and attach to the Subject (Fig. 14.4b). IMPORTANT You must fill both SS3L sensor cavities with electrode gel (GEL1) before attaching to the fingers. 9. Start the Biopac Student Lab Program. 10. Choose Lesson 14 (L14-Biofbk-1). 11. Type in your filename. 12. Click OK. You must fill the each cavity of the SS3L/SS3LA EDA/GSR transducer with electrode gel to obtain accurate recordings. The SS3L and SS3LA attach to the fingertips in an identical manner (Fig. 14.4b) and should be in place for at least five minutes prior to the start of recording. The SS3L/SS3LA is typically placed on the index and middle finger of the left hand. Position the transducer so that the sensor is on the bottom of your fingertip (the part without the fingernail) and wrap the Velcro tape around the finger so the transducer fits snugly but not so tight that blood circulation is cut off. It s a fine line between tight and too tight. Use a unique identifier. This ends the Set Up procedure. Sensors attach to bottom of fingertips Biopac Student Lab Velcro strap wrap around fingers Fig. 14.4b SS3L/SS3LA attachment and connection END OF SET UP

5 Lesson 14: Biofeedback Page 7 B. CALIBRATION The Calibration procedure establishes the hardware s internal parameters (such as gain, offset, and scaling) and is critical for optimum performance. Pay close attention to the Calibration procedure. FAST TRACK Calibration DETAILED EXPLANATION OF CALIBRATION STEPS 1. Subject should sit in a relaxed state, breathing normally. 2. Click Calibrate (then YES, then OK). When you click Calibrate, two dialog boxes will pop-up to make sure you are ready for the calibration procedure. Click YES for the first box and click OK for the second box. 3. After the beep, inhale and exhale deeply once. You should see three tracings on the screen: the EDA recording, the ECG recording, and the Heart Rate (ECG Rate). A beep should occur four seconds into the recording. The Subject should take one deep inhale and exhale, and then return to normal breathing. The program needs to see some variation in the EDA and BPM data. Note: The Subject should inhale and exhale without tensing the chest muscles too much, as this may cause excessive EMG artifact. 4. Wait for the Calibration to stop. The Calibration will run for 20 seconds and then stop automatically, so let it run its course. 5. Check your calibration data. If similar, proceed to the Data Recording section. If different, Redo Calibration. At the end of the 20-sec calibration recording, your screen should resemble Fig EDA/GSR Figure 14.5 Both recording channels should show some fluctuation. If there is not any fluctuation, then it is possible that a transducer is not connected properly and you must redo calibration. If the display of Heart Rate shows large fluctuations (20 BPM or more) during the interval from 4-20 seconds, then it is possible that excessive EMG artifact during inhale and exhale caused loss of BPM tracking and you should redo the calibration and have the Subject breathe less deeply. If you did not hear a beep during the calibration, then you should make sure your computer volume is turned up and redo calibration. Redo by clicking Redo Calibration and repeating the entire calibration sequence. END OF CALIBRATION

6 Page 8 Biopac Student Lab C. RECORDING LESSON DATA FAST TRACK Recording DETAILED EXPLANATION OF RECORDING STEPS 1. Prepare for the recording. In order to work efficiently, read this entire section so you will know what to do for each recording segment. This lesson requires that the Subject concentrate on the monitor display of Heart Rate and Arousal (EDA) and try to change the reading(s) without physical movement. Heart Rate and level of arousal (EDA) are independent, so the Subject may be able to change one or the other or both. Check the last line of the journal and note the total amount of time available for the recording. When you begin recording, your Heart Rate and your EDA (Fig. 14.6) will be plotted in a thermometer style bar chart. Fig Heart Rate and Arousal (EDA) The Heart Rate will be displayed to the left and will increase (rise) as the heart rate increases. Arousal (EDA) will be displayed on the right, and will increase (rise) as the body resistance decreases, indicating a higher level of arousal. The upper and lower range of each reading was set with your calibration data. Both readings should begin with values in approximately the center of the display. If your reading for either bar begins beyond the mid-range, you may want to consider redoing calibration. The bar chart display only works while data is being recorded. The display will change to a standard data plot when recording is complete. 2. Click Record. The recording will begin and will run continuously until you click Stop or the time indicated runs out. 3. For the first 90 seconds of the recording, Subject should try to relax so that both bars move down. 4. After 90 seconds, Subject should try to move the bars up by increasing arousal and Recorder should insert an event marker. Recording continues Hints for achieving relaxation: a) Relax your posture. b) Breathe in and out very slowly. c) Imagine yourself at a warm, relaxing seashore. d) Smile. e) Close your eyes. To move the bars up, you need to become more aroused. Recorder should insert an event marker 90 seconds into the recording to indicate that the Subject is now attempting arousal versus relaxation. To insert an event marker, press the F9 key once. (As you are viewing the bar chart display, you will not see the marker as it is entered. A marker can also be entered after the data is recorded.)

7 Lesson 14: Biofeedback Page 9 Insert event marker at 90 seconds. (No marker label is needed.) Hints for achieving arousal: a) Think of a stressful situation. b) Have the Director clap behind you. c) Make a scowl with your face. d) Hold your breath. 5. Click Stop. When you click Stop, the data window will change to display a standard three-channel recording with ECG, Heart Rate, and EDA. This data represents the entire recording. 6. Review the data on the screen. If correct, go to Step 7. If all went well, your data should look similar to Fig and you can proceed to Step 7. Note: The recording will vary greatly from person to person, and it is hard to generalize about the results. Hopefully, you were able to manipulate your physiological responses to some degree. This may take some practice, as many people are not used to thinking that they have some control over things like their heart rate and nervous system. If incorrect, click Redo. Fig Sample data The data would be incorrect if: a) The BPM data fluctuates excessively, indicating that an electrode may have peeled up. b) The EDA signal goes outside of its plotting region, indicating that one of the EDA transducers may not have had good contact. If incorrect, you should redo the recording by clicking Redo and repeating Steps 2-6. Note that once you press Redo, the data you have just recorded will be erased. 7. Click Done. A pop-up window with options will be generated. Make your choice, and continue as directed. 8. Remove the electrodes and transducers. If an SS3L transducer was used, clean the gel cavities. If choosing the Record from another Subject option: a) Attach the electrodes and transducer to the new Subject per Set Up Steps 5 and 6, and continue the entire lesson from Set Up Step 9. b) Each the new Subject will need to use a unique file name. Remove the electrode cable pinch connectors, and peel off the electrodes. Throw out the electrodes (BIOPAC electrodes are not reusable). Wash the electrode gel residue from the skin, using soap and water. The electrodes may leave a slight ring on the skin for a few hours. This is normal, and does not indicate that anything is wrong. END OF RECORDING If an SS3L transducer was used, use alcohol to clean out the electrode gel cavities.

8 Page 10 V. DATA ANALYSIS FAST TRACK Data Analysis DETAILED EXPLANATION OF DATA ANALYSIS STEPS Biopac Student Lab 1. Enter the Review Saved Data mode. Enter the Review Saved Data mode. Note Channel Number (CH) designation: Channel CH 2 CH 41 CH 42 Displays ECG Heart Rate EDA The data window should come up the same as Fig Fig Turn OFF Channel 2, ECG. To turn a channel OFF, click on the channel number box while holding down the Ctrl key. 3. Setup your display window for optimal viewing of the data interval when the Subject was trying to relax. 4. Set up the measurement boxes as follows: Channel CH 41 CH 42 Measurement value value Data Analysis continues Fig Channel 2 turned off This is the data that represents Subject trying to relax between Time 0 and the event marker you inserted at 90 seconds. The following tools help you adjust the data window: Autoscale horizontal Horizontal(Time) Scroll Bar Autoscale waveforms Vertical (Amplitude) Scroll Bar Zoom Tool Zoom Previous The measurement boxes are above the marker region in the data window. Each measurement has three sections: channel number, measurement type, and result. The first two sections are pull-down menus that are activated when you click on them. Brief definition of measurements: value: displays the amplitude value for the channel at the point selected by the I-beam cursor. If a single point is selected, the value is for that point, if an area is selected, the value is the endpoint of the selected area. Note: The selected area is the area selected by the I-beam tool (including the endpoints).

9 Lesson 14: Biofeedback Page Measure the EDA and heart rate values about 10 seconds into the recording. A It takes a few heart rate cycles for the BPM data to be correct, so measuring data 10 seconds into the record will be an accurate baseline. Simply clicking the cursor at Time = 10 seconds will update the value measurements. 6. Scroll to the point of maximal relaxation. Fig Use the horizontal scroll bar to move through the data. Select a point at the end of the first interval (before the marker) representing the point of maximal relaxation. 7. Take the measurements on this point as needed. A Fig Scroll to the point of maximal arousal. Scroll to the second interval (after the marker). Select a point at the end of the interval representing the point of maximal arousal. 9. Take the measurements on this point as needed. A 10. Save or print the data file. You may save the data to a drive, save notes that are in the journal, 11. Exit the program. END OF DATA ANALYSIS or print the data file. END OF LESSON 14 Complete the Lesson 14 Data Report that follows.

10 Page 12 L14: Biofeedback Data Report starts on next page.

11 Biopac Student Lab Page 13 BIOFEEDBACK Relaxation and Arousal DATA REPORT Student s Name: Lab Section: Date: I. Data and Calculations Subject Profile Name Height Age Weight Gender: Male / Female A. Measurement CH. # Heart Rate (BPM) CH 40 value Table 14.1 Baseline (10 sec.) Condition Maximal Relaxation Maximal Arousal EDA (ΔμMHO) CH 41 value II. Questions B. Based on the data from Table 14.1, did the effects of the parasympathetic nervous system change with biofeedback? Explain the physiological mechanisms causing the results. C. Describe a biofeedback program for stress management. Include details such as the physiological variable(s) you would measure, the transducers needed, and your criterion for a successful training program.

12 Page 14 L14: Biofeedback D. Name the branches of the autonomic nervous system and explain their function. E. Define Biofeedback and explain in general terms how it works. F. What change, if any, did your EDA recording show when you were aroused? Relaxed? G. Why is EDA a useful measure for biofeedback training? End of Lesson 14 Data Report

Lesson 10 Manual Revision BIOPAC Systems, Inc.

Lesson 10 Manual Revision BIOPAC Systems, Inc. Physiology Lessons for use with the Lesson 10 ELECTROOCULOGRAM (EOG) I Eye Movement Saccades and Fixation during Reading Manual Revision 3.7.3 061808 Vertical J.C. Uyehara, Ph.D. Biologist BIOPAC Systems,

More information

Lesson 9. Physiology Lessons for use with the Biopac Student Lab GALVANIC SKIN RESPONSE & THE POLYGRAPH

Lesson 9. Physiology Lessons for use with the Biopac Student Lab GALVANIC SKIN RESPONSE & THE POLYGRAPH Physiology Lessons for use with the Biopac Student Lab PC under Windows 95/98/NT 4.0/2000 or Macintosh Lesson 9 GALVANIC SKIN RESPONSE & THE POLYGRAPH Manual Revision 08232001.PL3.6.6-ML3.0.3 Richard Pflanzer,

More information

Biopac Student Lab Lesson 9 ELECTRODERMAL ACTIVITY & POLYGRAPH Analysis Procedure. Rev

Biopac Student Lab Lesson 9 ELECTRODERMAL ACTIVITY & POLYGRAPH Analysis Procedure. Rev 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 9 ELECTRODERMAL ACTIVITY & POLYGRAPH Analysis Procedure Rev. 12292017 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana

More information

Lesson 1 EMG 1 Electromyography: Motor Unit Recruitment

Lesson 1 EMG 1 Electromyography: Motor Unit Recruitment Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 1 EMG 1 Electromyography: Motor Unit Recruitment PC running Windows XP or Mac OS X 10.3-10.4 Lesson Revision 1.20.2006 BIOPAC Systems,

More information

Overview. Signal Averaged ECG

Overview. Signal Averaged ECG Updated 06.09.11 : Signal Averaged ECG Overview Signal Averaged ECG The Biopac Student Lab System can be used to amplify and enhance the ECG signal using a clinical diagnosis tool referred to as the Signal

More information

Exercise 1: Muscles in Face used for Smiling and Frowning Aim: To study the EMG activity in muscles of the face that work to smile or frown.

Exercise 1: Muscles in Face used for Smiling and Frowning Aim: To study the EMG activity in muscles of the face that work to smile or frown. Experiment HP-9: Facial Electromyograms (EMG) and Emotion Exercise 1: Muscles in Face used for Smiling and Frowning Aim: To study the EMG activity in muscles of the face that work to smile or frown. Procedure

More information

#PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis)

#PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis) BIOPAC Systems, Inc. 42 Aero Camino Goleta, Ca 93117 Ph (805)685-0066 Fax (805)685-0067 www.biopac.com info@biopac.com #PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis) The Biopac

More information

BASIC TUTORIAL. Jocelyn Mariah Kremer, Mike Mullins Documentation BIOPAC Systems, Inc. William McMullen Vice President BIOPAC Systems, Inc.

BASIC TUTORIAL. Jocelyn Mariah Kremer, Mike Mullins Documentation BIOPAC Systems, Inc. William McMullen Vice President BIOPAC Systems, Inc. BASIC TUTORIAL Windows 7, Vista, or XP Mac OS X 10.5-10.7 Jocelyn Mariah Kremer, Mike Mullins Documentation BIOPAC Systems, Inc. William McMullen Vice President BIOPAC Systems, Inc. BIOPAC Systems, Inc.

More information

Heart Rate Variability Preparing Data for Analysis Using AcqKnowledge

Heart Rate Variability Preparing Data for Analysis Using AcqKnowledge APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 01.06.2016 Application Note 233 Heart Rate Variability Preparing Data for Analysis

More information

Pseudorandom Stimuli Following Stimulus Presentation

Pseudorandom Stimuli Following Stimulus Presentation BIOPAC Systems, Inc. 42 Aero Camino Goleta, CA 93117 Ph (805) 685-0066 Fax (805) 685-0067 www.biopac.com info@biopac.com Application Note AS-222 05.06.05 Pseudorandom Stimuli Following Stimulus Presentation

More information

Electrocardiogram (ECG) Version Date: 2017-FEB-15 Document Effective Date: 2017-MAY-15

Electrocardiogram (ECG) Version Date: 2017-FEB-15 Document Effective Date: 2017-MAY-15 Title: Electrocardiogram (ECG) Version Date: 2017-FEB-15 Document Effective Date: 2017-MAY-15 Number: Data Collection Site (DCS) Version: 2.4 Number of Pages: SOP_DCS_0009 5 1.0 Purpose: The purpose of

More information

VivoSense. User Manual Galvanic Skin Response (GSR) Analysis Module. VivoSense, Inc. Newport Beach, CA, USA Tel. (858) , Fax.

VivoSense. User Manual Galvanic Skin Response (GSR) Analysis Module. VivoSense, Inc. Newport Beach, CA, USA Tel. (858) , Fax. VivoSense User Manual Galvanic Skin Response (GSR) Analysis VivoSense Version 3.1 VivoSense, Inc. Newport Beach, CA, USA Tel. (858) 876-8486, Fax. (248) 692-0980 Email: info@vivosense.com; Web: www.vivosense.com

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Game Show Physiology

Game Show Physiology iworx Physiology Lab Experiment Experiment HN-8: Game Show Physiology Game Show Physiology Note: The lab presented here is intended for evaluation purposes only. iworx users should refer to the User Area

More information

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read

More information

Biopac Student Lab PRO Manual. Professional Version 3.7.7

Biopac Student Lab PRO Manual. Professional Version 3.7.7 Biopac Student Lab PRO Manual Professional Version 3.7.7 Reference Manual for Biopac Student Lab PRO Software and MP36/35/30 or MP45 Hardware 42 Aero Camino, Goleta, CA 93117 (805) 685-0066, FAX (805)

More information

PRODUCT SHEET

PRODUCT SHEET PULSE OXIMETRY OXY100E Module (18-321 BPM) OXY200 Module (for veterinary use only, 18-450 BPM) TSD124 Series SPO2 Transducers for OXY100E TSD270 Series SPO2 Transducers for OXY200 These modules measure

More information

PRODUCT SHEET

PRODUCT SHEET PULSE OXIMETRY OXY100E Module (18-321 BPM) OXY200 Module (for veterinary use only, 18-450 BPM) TSD124 Series SPO2 Transducers for OXY100E or OXYSSH-SYS TSD270 Series SPO2 Transducers for OXY200 These modules

More information

User Guide EMG. This user guide has been created to educate and inform the reader about doing EMG measurements

User Guide EMG. This user guide has been created to educate and inform the reader about doing EMG measurements User Guide EMG This user guide has been created to educate and inform the reader about doing EMG measurements For more information about NeXus, our BioTrace+ software, please visit our website or contact

More information

PRODUCT SHEET

PRODUCT SHEET ERS100C EVOKED RESPONSE AMPLIFIER MODULE The evoked response amplifier module (ERS100C) is a single channel, high gain, extremely low noise, differential input, biopotential amplifier designed to accurately

More information

iworx Sample Lab Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach

iworx Sample Lab Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach Background Do you ever wonder why your stomach growls, that funny sound it makes when you are really hungry? Stomach growling is the

More information

Introduction: Overview. EECE 2510 Circuits and Signals: Biomedical Applications. ECG Circuit 2 Analog Filtering and A/D Conversion

Introduction: Overview. EECE 2510 Circuits and Signals: Biomedical Applications. ECG Circuit 2 Analog Filtering and A/D Conversion EECE 2510 Circuits and Signals: Biomedical Applications ECG Circuit 2 Analog Filtering and A/D Conversion Introduction: Now that you have your basic instrumentation amplifier circuit running, in Lab ECG1,

More information

iworx Sample Lab Experiment AM-9: Crayfish Gut Pharmacology

iworx Sample Lab Experiment AM-9: Crayfish Gut Pharmacology Experiment AM-9: Crayfish Gut Pharmacology The Dissection 1. In a dissecting pan, cover a crayfish with ice for 5 minutes. 2. With scissors, remove the entire crayfish abdomen (tail). 3. Place the crayfish

More information

Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach

Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach Background Do you ever wonder why your stomach growls, that funny sound it makes when you are really hungry? Stomach growling is the

More information

Experiment PP-1: Electroencephalogram (EEG) Activity

Experiment PP-1: Electroencephalogram (EEG) Activity Experiment PP-1: Electroencephalogram (EEG) Activity Exercise 1: Common EEG Artifacts Aim: To learn how to record an EEG and to become familiar with identifying EEG artifacts, especially those related

More information

BioGraph Infiniti Physiology Suite

BioGraph Infiniti Physiology Suite Thought Technology Ltd. 2180 Belgrave Avenue, Montreal, QC H4A 2L8 Canada Tel: (800) 361-3651 ٠ (514) 489-8251 Fax: (514) 489-8255 E-mail: mail@thoughttechnology.com Webpage: http://www.thoughttechnology.com

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

PRODUCT SHEET TRANSDUCER MODULE GSR100C

PRODUCT SHEET TRANSDUCER MODULE GSR100C TRANSDUCER MODULE GSR100C The GSR100C electrodermal activity amplifier module is a single-channel, high-gain, differential amplifier designed to measure skin conductance via the constant voltage technique.

More information

Common Spatial Patterns 3 class BCI V Copyright 2012 g.tec medical engineering GmbH

Common Spatial Patterns 3 class BCI V Copyright 2012 g.tec medical engineering GmbH g.tec medical engineering GmbH Sierningstrasse 14, A-4521 Schiedlberg Austria - Europe Tel.: (43)-7251-22240-0 Fax: (43)-7251-22240-39 office@gtec.at, http://www.gtec.at Common Spatial Patterns 3 class

More information

R Series Quick Reference Guide

R Series Quick Reference Guide R Series Quick Reference Guide R Series Quick Reference Guide This Quick Reference Guide is intended to provide easy access to most common uses of the ZOLL R Series. The user must read and completely

More information

Tinnitus can be helped. Let us help you.

Tinnitus can be helped. Let us help you. What a relief. Tinnitus can be helped. Let us help you. What is tinnitus? Around 250 million people worldwide suffer Tinnitus is the perception of sounds or noise within the ears with no external sound

More information

Common Spatial Patterns 2 class BCI V Copyright 2012 g.tec medical engineering GmbH

Common Spatial Patterns 2 class BCI V Copyright 2012 g.tec medical engineering GmbH g.tec medical engineering GmbH Sierningstrasse 14, A-4521 Schiedlberg Austria - Europe Tel.: (43)-7251-22240-0 Fax: (43)-7251-22240-39 office@gtec.at, http://www.gtec.at Common Spatial Patterns 2 class

More information

Experiment BI-1: ECG Signal Conditioning

Experiment BI-1: ECG Signal Conditioning Experiment BI-1: ECG Signal Conditioning Equipment Required PC Mac Computer IX-TA-220 data acquisition unit and power supply USB cale ire-b3g ECG cale and electrode lead wires C-DIN-BB: Din to Breadoard

More information

User Guide. qeeg NeXus-32. This user guide has been created to educate and inform the reader about doing qeeg measurements with the NeXus-32

User Guide. qeeg NeXus-32. This user guide has been created to educate and inform the reader about doing qeeg measurements with the NeXus-32 User Guide qeeg NeXus-32 This user guide has been created to educate and inform the reader about doing qeeg measurements with the NeXus-32 For more information about NeXus, our BioTrace+ software, please

More information

SedLine Sedation Monitor

SedLine Sedation Monitor SedLine Sedation Monitor Quick Reference Guide Not intended to replace the Operator s Manual. See the SedLine Sedation Monitor Operator s Manual for complete instructions, including warnings, indications

More information

Preparation of the participant. EOG, ECG, HPI coils : what, why and how

Preparation of the participant. EOG, ECG, HPI coils : what, why and how Preparation of the participant EOG, ECG, HPI coils : what, why and how 1 Introduction In this module you will learn why EEG, ECG and HPI coils are important and how to attach them to the participant. The

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

Environmental Controls Laboratory

Environmental Controls Laboratory (Electro-Oculography Application) Introduction Spinal cord injury, cerebral palsy, and stroke are some examples of clinical problems which can have a large effect on upper extremity motor control for afflicted

More information

PLEN Motion Editor Operation Procedures Temporary File

PLEN Motion Editor Operation Procedures Temporary File 1 PLEN Motion Editor Operation Procedures Temporary File 2 Table of Contents I. Introduction 3 1. Overview of MotionEditor 3 2. Operating Environment. 3 3. Installing the MotionEditor... 3 4. PLEN's Preparation

More information

MANUAL User manual Biofeedback-training esense-app Electrodes

MANUAL User manual Biofeedback-training esense-app Electrodes MANUAL User manual Biofeedback-training esense-app Electrodes Version 3.2 29.03.2018 Contents Contents... 1 Delivery contents... 3 Skin response in general... 4 Skin response and biofeedback training...

More information

iworx Sample Lab Experiment AN-13: Crayfish Motor Nerve

iworx Sample Lab Experiment AN-13: Crayfish Motor Nerve Experiment AN-13: Crayfish Motor Nerve Background The purpose of this experiment is to record the extracellular action potentials of crayfish motor axons. These spontaneously generated action potentials

More information

Activity P32: Variation of Light Intensity (Light Sensor)

Activity P32: Variation of Light Intensity (Light Sensor) Activity P32: Variation of Light Intensity (Light Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Illuminance P32 Vary Light.DS P54 Light Bulb Intensity P54_BULB.SWS Equipment Needed

More information

Mackie Control and Cubase SX/SL

Mackie Control and Cubase SX/SL Mackie Control and Cubase SX/SL - 1 - The information in this document is subject to change without notice and does not represent a commitment on the part of Steinberg Media Technologies AG. The software

More information

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes Welch Allyn CardioPerfect Workstation Tango+ Interface Notes To setup Tango+ with the CardioPerfect stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

Basketball Questions

Basketball Questions Basketball is an exciting and fast paced game. It requires shooting, passing, catching, bouncing skills often while jumping and running. Watch the following video clips and answer the following questions

More information

Pulse Biomedical QRS-Card Tango M2 Interface Notes

Pulse Biomedical QRS-Card Tango M2 Interface Notes Pulse Biomedical QRS-Card Tango M2 Interface Notes To setup Tango M2 with your stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232 Cable used to communicate

More information

Lab experience 1: Introduction to LabView

Lab experience 1: Introduction to LabView Lab experience 1: Introduction to LabView LabView is software for the real-time acquisition, processing and visualization of measured data. A LabView program is called a Virtual Instrument (VI) because

More information

Procedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement

Procedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement Procedures Guide Tektronix HDMI Sink Instruments Differential Impedance Measurement Rev. 1.1: October 13, 2010 2 Measurement Procedures Equipment Required Table 1 lists the equipment required to perform

More information

Activity P27: Speed of Sound in Air (Sound Sensor)

Activity P27: Speed of Sound in Air (Sound Sensor) Activity P27: Speed of Sound in Air (Sound Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Speed of sound P27 Speed of Sound 1.DS (See end of activity) (See end of activity) Equipment

More information

The EEGer 4.3 Tutorial

The EEGer 4.3 Tutorial The EEGer 4.3 Tutorial How to Conduct A Neurofeedback Session 1 Introduction 3 1.1 How To Use This Tutorial............................................................................ 3 1.2 What You Should

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

Analysis of AP/axon classes and PSP on the basis of AP amplitude

Analysis of AP/axon classes and PSP on the basis of AP amplitude Analysis of AP/axon classes and PSP on the basis of AP amplitude In this analysis manual, we aim to measure and analyze AP amplitudes recorded with a suction electrode and synaptic potentials recorded

More information

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 OBJECTIVE To become familiar with state-of-the-art digital data acquisition hardware and software. To explore common data acquisition

More information

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z MestReNova A quick Guide page 1 MNova is a program to analyze 1D- and 2D NMR data. Start of MNova Start All Programs Chemie NMR MNova The MNova Menu 1. 2. Create expanded regions Adjust signal intensity

More information

Classroom. Chapter 3: Lesson 17

Classroom. Chapter 3: Lesson 17 Classroom Chapter 3: Lesson 17 Adventus Incorporated, 2001 Chapter 3: Lines & Spaces Above Middle C Lesson 17 This lesson plan was written for use with Piano Suite Premier software, and is intended as

More information

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Products: R&S RTO2002 R&S RTO2004 R&S RTO2012 R&S RTO2014 R&S RTO2022 R&S RTO2024 R&S RTO2044 R&S RTO2064 This application

More information

Mortara X-Scribe Tango+ Interface Notes

Mortara X-Scribe Tango+ Interface Notes Mortara X-Scribe Tango+ Interface Notes To setup Tango+ with the X-Scribe stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232 Cable used to communicate

More information

Thought Technology Ltd Belgrave Avenue, Montreal, QC H4A 2L8 Canada

Thought Technology Ltd Belgrave Avenue, Montreal, QC H4A 2L8 Canada Thought Technology Ltd. 2180 Belgrave Avenue, Montreal, QC H4A 2L8 Canada Tel: (800) 361-3651 ٠ (514) 489-8251 Fax: (514) 489-8255 E-mail: _Hmail@thoughttechnology.com Webpage: _Hhttp://www.thoughttechnology.com

More information

NIHR Southampton Biomedical Research Centre

NIHR Southampton Biomedical Research Centre The NIHR Southampton (BRC) has a tight quality assurance system for the writing, reviewing and updating of Standard Operating Procedures. As such, version-controlled and QA authorised Standard Operating

More information

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment Integrated Component Options Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment PRELIMINARY INFORMATION SquareGENpro is the latest and most versatile of the frequency

More information

Re: ENSC 370 Project Physiological Signal Data Logger Functional Specifications

Re: ENSC 370 Project Physiological Signal Data Logger Functional Specifications School of Engineering Science Simon Fraser University V5A 1S6 versatile-innovations@sfu.ca February 12, 1999 Dr. Andrew Rawicz School of Engineering Science Simon Fraser University Burnaby, BC V5A 1S6

More information

Basic LabVIEW Programming Amit J Nimunkar, Sara Karle, Michele Lorenz, Emily Maslonkowski

Basic LabVIEW Programming Amit J Nimunkar, Sara Karle, Michele Lorenz, Emily Maslonkowski Introduction This lab familiarizes you with the software package LabVIEW from National Instruments for data acquisition and virtual instrumentation. The lab also introduces you to resistors, capacitors,

More information

GE CardioSoft (Version 6.01 or higher) Tango M2 Interface Notes

GE CardioSoft (Version 6.01 or higher) Tango M2 Interface Notes GE CardioSoft (Version 6.01 or higher) Tango M2 Interface Notes To setup Tango M2 with your CardioSoft (V6.01 or higher), simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

COPYRIGHT NOVEMBER-1998

COPYRIGHT NOVEMBER-1998 Application Notes: Interfacing AG-132 GPS with G-858 Magnetometer 25430-AM Rev.A Operation Manual COPYRIGHT NOVEMBER-1998 GEOMETRICS, INC. 2190 Fortune Drive, San Jose, Ca 95131 USA Phone: (408) 954-0522

More information

Quick Reference Manual

Quick Reference Manual Quick Reference Manual V1.0 1 Contents 1.0 PRODUCT INTRODUCTION...3 2.0 SYSTEM REQUIREMENTS...5 3.0 INSTALLING PDF-D FLEXRAY PROTOCOL ANALYSIS SOFTWARE...5 4.0 CONNECTING TO AN OSCILLOSCOPE...6 5.0 CONFIGURE

More information

Experiment P32: Sound Waves (Sound Sensor)

Experiment P32: Sound Waves (Sound Sensor) PASCO scientific Vol. 2 Physics Lab Manual P32-1 Experiment P32: (Sound Sensor) Concept Time SW Interface Macintosh file Windows file waves 45 m 700 P32 P32_SOUN.SWS EQUIPMENT NEEDED Interface musical

More information

SigPlay User s Guide

SigPlay User s Guide SigPlay User s Guide . . SigPlay32 User's Guide? Version 3.4 Copyright? 2001 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or

More information

Data Acquisition Using LabVIEW

Data Acquisition Using LabVIEW Experiment-0 Data Acquisition Using LabVIEW Introduction The objectives of this experiment are to become acquainted with using computer-conrolled instrumentation for data acquisition. LabVIEW, a program

More information

FOTS100 User Manual. BIOPAC Systems, Inc. Opsens Inc. 42 Aero Camino, Goleta, CA Tel (805) , Fax (805)

FOTS100 User Manual. BIOPAC Systems, Inc. Opsens Inc. 42 Aero Camino, Goleta, CA Tel (805) , Fax (805) FOTS100 User Manual BIOPAC Systems, Inc. 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066, Fax (805) 685-0067 WWW.BIOPAC.COM 1 WARRANTY All products manufactured by Opsens inc. are warranted to be free

More information

BCM Calibration for E Abstract

BCM Calibration for E Abstract Jefferson Lab E8-4 Analysis Report July 22 BCM Calibration for E8-4 Patricia Solvignon Jefferson Lab E-mail solvigno@jlab.org Abstract In this note, the calibration procedure of the Beam Current Monitors

More information

EAN-Performance and Latency

EAN-Performance and Latency EAN-Performance and Latency PN: EAN-Performance-and-Latency 6/4/2018 SightLine Applications, Inc. Contact: Web: sightlineapplications.com Sales: sales@sightlineapplications.com Support: support@sightlineapplications.com

More information

PhysioPilot C2 Plus Fourth-Generation Software APPLICATIONS Contents EMG FEEDBACK...2 APPENDIX ARCHIVING SESSION DATA... 23

PhysioPilot C2 Plus Fourth-Generation Software APPLICATIONS Contents EMG FEEDBACK...2 APPENDIX ARCHIVING SESSION DATA... 23 HOW TO RUN AN EMG BIOFEEDBACK SESSION PhysioPilot C2 Plus Fourth-Generation Software APPLICATIONS Contents EMG FEEDBACK...2 APPENDIX... 23 ARCHIVING SESSION DATA... 23 COMMENT : This section explains how

More information

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO) 2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency

More information

Cardioline Cube Stress Tango M2 Interface Notes

Cardioline Cube Stress Tango M2 Interface Notes Cardioline Cube Stress Tango M2 Interface Notes To setup Tango M2 with the Cube stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232 Cable used to communicate

More information

ASE 369 K Measurements and Instrumentation. LAB #9: Impulse-Force Hammer; Vibration of Beams

ASE 369 K Measurements and Instrumentation. LAB #9: Impulse-Force Hammer; Vibration of Beams ASE 369 K Measurements and Instrumentation LAB #9: Impulse-Force Hammer; Vibration of Beams Equipment: Dell Optiplex computer with National Instruments PCI-MIO-16E-4 data-acquisition board and the Virtual

More information

Schiller AT-10 plus Tango M2 Interface Notes

Schiller AT-10 plus Tango M2 Interface Notes Schiller AT-10 plus Tango M2 Interface Notes To setup Tango M2 with the AT-10 plus stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232 Cable used to

More information

Marquette Case 12, Case 15, Case 16, and Centra Stress Systems Tango M2 Interface Notes

Marquette Case 12, Case 15, Case 16, and Centra Stress Systems Tango M2 Interface Notes Marquette Case 12, Case 15, Case 16, and Centra Stress Systems Tango M2 Interface Notes To setup Tango M2 with your stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger

More information

USB Mini Spectrum Analyzer User s Guide TSA5G35

USB Mini Spectrum Analyzer User s Guide TSA5G35 USB Mini Spectrum Analyzer User s Guide TSA5G35 Triarchy Technologies, Corp. Page 1 of 21 USB Mini Spectrum Analyzer User s Guide Copyright Notice Copyright 2011 Triarchy Technologies, Corp. All rights

More information

MANUAL User manual Biofeedback-training esense-app Electrodes

MANUAL User manual Biofeedback-training esense-app Electrodes MANUAL User manual Biofeedback-training esense-app Electrodes Version 3.2.6 12th of January, 2019 Contents Contents... 1 Delivery contents... 4 Skin response in general... 5 Skin response and biofeedback

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

Brain-Computer Interface (BCI)

Brain-Computer Interface (BCI) Brain-Computer Interface (BCI) Christoph Guger, Günter Edlinger, g.tec Guger Technologies OEG Herbersteinstr. 60, 8020 Graz, Austria, guger@gtec.at This tutorial shows HOW-TO find and extract proper signal

More information

Calibrating the timecode signal input

Calibrating the timecode signal input Chapter 5 Calibrating the timecode signal input Computer hardware can introduce an offset between the timecode signal and the video signal, which causes the timecode and video to be offset when they are

More information

(B) EL BACK LIGHT Press button S1 for EL back light with 3 seconds auto light up. (C) BASIC OPERATION OF BUTTONS

(B) EL BACK LIGHT Press button S1 for EL back light with 3 seconds auto light up. (C) BASIC OPERATION OF BUTTONS (A) FEATURES PM200 is a multi-function watch designed to have Chronograph, Count Down Timer, Alarm and Heart Rate Monitor. The main features are listed as follow: 6 modes operation: Normal Time, Heart

More information

FOUR CHANNEL USB RECORDER PCRU01. User manual. Table of Contents

FOUR CHANNEL USB RECORDER PCRU01. User manual. Table of Contents FOUR CHANNEL USB RECORDER PCRU01 User manual Table of Contents Features 2 Specifications 2 hardware 2 software: 2 system requirements 2 Software installation 2 SAFETY and WARNINGS 3 Warranty 3 Connections

More information

LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution. A. Plotting a GM Plateau. This lab will have two sections, A and B.

LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution. A. Plotting a GM Plateau. This lab will have two sections, A and B. LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution This lab will have two sections, A and B. Students are supposed to write separate lab reports on section A and B, and submit the

More information

PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX

PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX w w w. m e n t o r. c o m PCIe: Eye Diagram Analysis in HyperLynx PCI Express Tutorial This PCI Express tutorial will walk you through time-domain eye diagram analysis

More information

Analyzing and Saving a Signal

Analyzing and Saving a Signal Analyzing and Saving a Signal Approximate Time You can complete this exercise in approximately 45 minutes. Background LabVIEW includes a set of Express VIs that help you analyze signals. This chapter teaches

More information

ISCEV SINGLE CHANNEL ERG PROTOCOL DESIGN

ISCEV SINGLE CHANNEL ERG PROTOCOL DESIGN ISCEV SINGLE CHANNEL ERG PROTOCOL DESIGN This spreadsheet has been created to help design a protocol before actually entering the parameters into the Espion software. It details all the protocol parameters

More information

Quick Start for TrueRTA (v3.5) on Windows XP (and earlier)

Quick Start for TrueRTA (v3.5) on Windows XP (and earlier) Skip directly to the section that covers your version of Windows (XP and earlier, Vista or Windows 7) Quick Start for TrueRTA (v3.5) on Windows XP (and earlier) Here are step-by-step instructions to get

More information

Tinnitus Management Strategies to help you conquer tinnitus like never before.

Tinnitus Management Strategies to help you conquer tinnitus like never before. Tame your tinnitus. Tinnitus Management Strategies to help you conquer tinnitus like never before. Around 250 million people worldwide suffer from tinnitus. What is tinnitus? Tinnitus is the perception

More information

Mortara XScribe Versions 3.xx and 5.xx Tango M2 Interface Notes

Mortara XScribe Versions 3.xx and 5.xx Tango M2 Interface Notes Mortara XScribe Versions 3.xx and 5.xx Tango M2 Interface Notes To setup Tango M2 with the XScribe cardiac stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables

More information

Bridges and Arches. Authors: André Holleman (Bonhoeffer college, teacher in research at the AMSTEL Institute) André Heck (AMSTEL Institute)

Bridges and Arches. Authors: André Holleman (Bonhoeffer college, teacher in research at the AMSTEL Institute) André Heck (AMSTEL Institute) Bridges and Arches Authors: André Holleman (Bonhoeffer college, teacher in research at the AMSTEL Institute) André Heck (AMSTEL Institute) A practical investigation task for pupils at upper secondary school

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

BrainMaster tm System Type 2E Module & BMT Software for Windows tm. Display Screens for Master.exe

BrainMaster tm System Type 2E Module & BMT Software for Windows tm. Display Screens for Master.exe BrainMaster tm System Type 2E Module & BMT Software for Windows tm Display Screens for Master.exe 1995-2004 BrainMaster Technologies, Inc., All Rights Reserved BrainMaster and From the Decade of the Brain

More information

The Effect of Music Tempo on the Psychophysiological Measures of Stress

The Effect of Music Tempo on the Psychophysiological Measures of Stress The Effect of Music Tempo on the Psychophysiological Measures of Stress Briana Brownlow 8 Abstract Music and its influence on stress have been researched extensively (Pelletier, 2004). Specifically, the

More information

Source/Receiver (SR) Setup

Source/Receiver (SR) Setup PS User Guide Series 2015 Source/Receiver (SR) Setup For 1-D and 2-D Vs Profiling Prepared By Choon B. Park, Ph.D. January 2015 Table of Contents Page 1. Overview 2 2. Source/Receiver (SR) Setup Main Menu

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

COSMED Quark CPET metabolic cart Tango M2 Interface Notes

COSMED Quark CPET metabolic cart Tango M2 Interface Notes COSMED Quark CPET metabolic cart Tango M2 Interface Notes To setup Tango M2 with the COSMED Quark CPET stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables

More information