Differential Scanning Calorimetry (DSC) Of Powder Coating Materials

Size: px
Start display at page:

Download "Differential Scanning Calorimetry (DSC) Of Powder Coating Materials"

Transcription

1 Differential Scanning Calorimetry (DSC) Of Powder Coating Materials Introduction This manual provides instructions to prepare samples for and subsequently perform a DSC (differential scanning calorimetry) test on thermoset powder coating materials. The DSC machine used in this instruction manual is the Perkin Elmer DSC-6. DSC tests are used to determine a number of temperature related physical properties of materials, such as Tg (glass transition temperature), degree of cure, and reaction onset temperature. Methods for the analysis of DSC results are beyond the scope of this set of instructions, and will be covered in a separate document. In order to perform the described task, minimal laboratory experience is necessary. The time to complete one full measurement cycle is approximately one half hour. Sections Sample Preparation. Loading the DSC. Sample DSC Run. Sample DSC Run Completion Equipment List 1. Perkin Elmer Sample pans and lids 2. Perkin Elmer Pan Press 3. Tweezers 4. Perkin Elmer DSC 6 Differential Scanning Calorimeter 5. Razor lades 6. Copy paper (Not shown)

2 Step 1 Sample Preparation Caution: Razor blades are dangerous. Take care to ensure you do not cut yourself during their use. Note: Do not touch the collected samples, sample pans or lids with your fingers. The DSC test is sensitive to contamination from skin oils and this could affect the accuracy of your results.. Scrape or cut shavings of the cured powder coating material onto a piece of clean paper, using a razor blade. Collect enough sample material to fill a sample pan.. Transfer the sample material onto a 3 x2 piece of paper with a fold lengthwise down the center as shown. Step 2 Sample Preparation Note: Use tweezers to handle sample pans and lids to prevent contamination from hand oils or foreign materials. Note: It is normal for small amounts of the sample to spill as it is transferred into aluminum sample pan.. dd sample material to the sample pan until it is level with the top edge of the pan.. Place a pan lid over the sample material in the pan. The lids are non-directional so it does not matter which side is up. C. Press the lid into pan using tweezers. C

3 Step 3 Sample Preparation Note: Use tweezers to handle prepared samples.. Using tweezers, place the prepared sample pan on the cylinder body at the base of the pan press.. Pull the cylinder body downward to expose the center anvil. The cylinder body is spring loaded in the up (closed) position. C. Slide the sample to a centered position over the anvil. D. Gently allow the cylinder body to rise back up into the closed position. E. The sample should now be positioned on top of the anvil, at the bottom of the anvil-cylinder assembly. Step 4 Sample Preparation Note: Use tweezers to handle prepared samples.. Push down on the hand pad at the top of the press; this drives a crimping piston into the anvil chamber and seals the pan. Press the pad downward until you hear a soft click, indicating the pan edge has been crimped over the lid.. Remove the sealed pan from the press by pulling the cylinder body downward to expose the sealed sample pan. Remove with tweezers. C. Place the sealed sample pan on a hard surface, like the press base, and poke a small hole in the top of the pan lid using the tip of the tweezers. The hole allows gasses to escape during the DSC process.

4 Step 5 Loading the DSC. Note: Do not remove the sample pan from the cavity on the left hand side of the furnace; this is a calibration standard and must be in place for the test to run correctly.. Pick up and remove the outer furnace access door, using your tweezers, and set it aside.. Remove the inner furnace access door with your tweezers and set it aside. C. Remove sample from previous DSC run from the cavity on the right hand side of the inner furnace. D. Place the sample you just prepared into the cavity on the right hand side of the inner furnace. The pan will center itself on the raised button in the center of the cavity. E. Replace inner furnace access doors. F. Replace outer furnace access door. Step 6 Sample DSC Run. Turn on the computer and the DSC machine. The on-off switch for the DSC machine is at the back of the instrument.. Wait for the computer to finish its start routine. fter startup, the screen will have a number of icons displayed on the left hand side. Double click the Pyris Manager icon.

5 Step 7 Sample DSC Run Note: When the software is opened, a menu tool bar will appear at the top of the desktop.. Open the instrument viewer screen by double clicking on the DSC 6 button located on the menu tool bar at the top of the desktop. Step 8 Sample DSC Run Note: The Instrument Viewer screen is a real time graphical display of the instrument. The graph shows endothermic heat flow vs elapsed time.. Open the Method Editor screen by single clicking on the page icon as shown; this allows you to input the run parameters.

6 Step 9 Sample DSC Run Note: The Method Editor Window will open. This window contains three tab selectable pages for user to interface with. These windows are used to program the test parameters. Note: Hitting the enter/return key on your computer keyboard enters any value or text you have typed, as does the Tab key and clicking on a different text block.. Click on the sample tab to ensure that it is selected and active.. Click in the white text block area labeled Sample ID: and type in the sample information: Part#, lot# sample#. Click in the Operator ID: text area and type in your initials. Step 10 Sample DSC Run. Click on the Initial State tab to activate this interface window.. Ensure that the Temperature value is set to C and that the Y Initial value is set to 20.00mW. These are preset default values but they can be adjusted if necessary by using the arrow buttons next to each entry area, or by clicking in the text block and typing in a value.

7 Step 11 Sample DSC Run. Click on the Program tab to activate this interface window.. Ensure that the Initial Temp value is set to C. The value you set in the initial temp text block will be displayed as the first step, highlighted in blue, in the program display area below the initial temp value. Step 12 Sample DSC Run. Click on the dd a step button located just the right of the program display area. This will button opens a pop-up window named Method Step Options.. Click on Temperature Scan from the list. C. Click on the OK button at bottom of pop-up screen to return to program screen. C

8 Step 13 Sample DSC Run Note: The Temperature Scan Window will open. This window allows you to set the conditions for the second step of the program, a scan, in which the temperature will be increased to the set point at the rate you have set.. Ensure that the To temperature option is set to C.. Ensure that the Rate option is set to C/min. Step 14 Sample DSC Run. Click on the End Condition button to open the End Condition interface window.. Set the Go To: temperature to C. The end condition value sets the temperature of the furnace at the completion of the program cycle. C. Click on the Done button.

9 Step 15 Sample DSC Run. Press the DSC start/stop run button to begin the program. Sample DSC Run Completion The DSC run will be complete after the equipment reaches the end condition set in Step 14. s stated in the introduction, analysis of the resulting data is beyond the scope of this set of instructions and will be addressed in a separate document. Concluding remark on DSC testing Differential Scanning Calorimeters work by comparing the ability of two separate samples, placed on two separate heating elements, to take on energy in the form of heat. The DSC measures and graphs the difference in energy needed to maintain both samples at the same temperature. The graph is then evaluated for inflection points, peaks and peak areas, to determine results.

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1 READ THIS FIRST Morphologi G3 Quick Start Guide MAN0412 Issue1.1 Malvern Instruments Ltd. 2008 Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Yellow Frog. Manual Version 1.1

Yellow Frog. Manual Version 1.1 Yellow Frog Manual Version 1.1 1 YellowFrog Contents PC Requirements...... 2 YellowFrog Power Meter Measurement.... 3 YellowFrog PC Software..... 3 Main Screen....... 4 Input Overload....... 5 Battery

More information

Inserting the batteries. Basic settings of the remote control

Inserting the batteries. Basic settings of the remote control Inserting the batteries Procedure prior to first use or when changing batteries Remove the back plate to expose the battery tray. Insert 2 x AA 1.5V alkaline batteries. Ensure the polarity of the batteries

More information

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.1 Click on the OMNIC icon to open the software. 1.2 Check that a signal is being measured by opening the Experiment Setup menu under the Collect

More information

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200 Manual for TV software TT-Viewer version 2.7.0 Figure: TT-budget S2-3200 Index Manual TT-Viewer 3 1. Starting TT-Viewer software 3 2. General settings 5 3. Assignment of hardware 6 3.1 Unicable 7 4. Renderer

More information

Activity P27: Speed of Sound in Air (Sound Sensor)

Activity P27: Speed of Sound in Air (Sound Sensor) Activity P27: Speed of Sound in Air (Sound Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Speed of sound P27 Speed of Sound 1.DS (See end of activity) (See end of activity) Equipment

More information

Data Acquisition Using LabVIEW

Data Acquisition Using LabVIEW Experiment-0 Data Acquisition Using LabVIEW Introduction The objectives of this experiment are to become acquainted with using computer-conrolled instrumentation for data acquisition. LabVIEW, a program

More information

GETTING STARTED... 2 ENVIRONMENT SCAN... 2

GETTING STARTED... 2 ENVIRONMENT SCAN... 2 Sputnik User Manual GETTING STARTED... 2 SPUTNIK OVERVIEW... 2 ENVIRONMENT SCAN... 2 SETTING UP SPUTNIK... 2 CONDUCTING AN ENVIRONMENT SCAN... 3 ENVIRONMENT SCAN RESULTS... 3 1 GETTING STARTED Sputnik

More information

VPM2. Operation Manual

VPM2. Operation Manual VPM2 Operation Manual Whip Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com LISTED Features The Whip VPM2 is designed

More information

Amcom Software RECEIVING AT ITEM LEVEL W/POCKET PC SCANNER DAILY PROCESS

Amcom Software RECEIVING AT ITEM LEVEL W/POCKET PC SCANNER DAILY PROCESS Software RECEIVING AT ITEM LEVEL W/POCKET PC SCANNER DAILY PROCESS CHECKLIST... 2 GENERAL SCREEN LAYOUT... 3 SCANNER CONVENTIONS... 3 XFER PRICEBK TO SCANNER... 4 ENTER RECEIVING W/SYMBOL SCANNER... 6

More information

VPM2. Operator's Manual

VPM2. Operator's Manual VPM2 Operator's Manual Whip Mix Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com Features The Whip Mix VPM2 is designed

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

PSC300 Operation Manual

PSC300 Operation Manual PSC300 Operation Manual Version 9.10 General information Prior to any attempt to operate this Columbia PSC 300, operator should read and understand the complete operation of the cubing system. It is very

More information

TA Instruments Cement Analysis Software Getting Started Guide

TA Instruments Cement Analysis Software Getting Started Guide TA Instruments Cement Analysis Software Getting Started Guide Overview Cement Analysis software from TA Instruments is a standalone software application for running cement setting experiments and analyzing

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE Thank you for purchasing a TSI AeroTrak Model 9110 Portable Airborne Particle Counter (particle counter). This guide will help you

More information

Desktop. Basic use of EndNote. Important start info 3 tips p. 1. Entering references manually p. 3

Desktop. Basic use of EndNote. Important start info 3 tips p. 1. Entering references manually p. 3 Basic use of EndNote Desktop Important start info 3 tips p. 1 Entering references manually p. 3 Import references from databases / search engines p. 4 Check for duplicates p. 5 Using EndNote with Word

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

JAMAR TRAX RD Detector Package Power Requirements Installation Setting Up The Unit

JAMAR TRAX RD Detector Package Power Requirements Installation Setting Up The Unit JAMAR TRAX RD The TRAX RD is an automatic traffic recorder designed and built by JAMAR Technologies, Inc. Since the unit is a Raw Data unit, it records a time stamp of every sensor hit that occurs during

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

Using the Agilent for Single Crystal Work

Using the Agilent for Single Crystal Work Using the Agilent for Single Crystal Work Generally, the program to access the Agilent, CrysalisPro, will be open. If not, you can start the program using the shortcut on the desktop. There is no password

More information

How to Optimize Ad-Detective

How to Optimize Ad-Detective How to Optimize Ad-Detective Ad-Detective technology is based upon black level detection. There are several important criteria to consider: 1. Does the video have black frames to detect? Are there any

More information

42 Freestanding Infrared Multi Touch Screen Kiosk User s Manual

42 Freestanding Infrared Multi Touch Screen Kiosk User s Manual 42 Freestanding Infrared Multi Touch Screen Kiosk User s Manual Manual Version L42HD-T2.2 Safety Instructions Please keep the display away from any heat sources such as radiators or direct sunlight. Place

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Video Conference Classroom Documentation

Video Conference Classroom Documentation Updated: 8/18/2017 Video Conference Classroom Documentation Contents About These Classrooms... 2 Where... 2 Podium Overview... 2 On Top of Podium... 2 Inside the Podium... 2 Equipment Information... 2

More information

A6 OPERATING INSTRUCTIONS

A6 OPERATING INSTRUCTIONS Amerec s A6 control for the AX steamer is a touch screen control intended to be mounted on a wall, generally in or near the steam bath. It may be mounted directly on the wall surface or, using an optional

More information

Getting started with

Getting started with Getting started with Electricity consumption monitoring single phase for homes and some smaller light commercial premises OVERVIEW: The OWL Intuition-e electricity monitoring system comprises of three

More information

LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution. A. Plotting a GM Plateau. This lab will have two sections, A and B.

LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution. A. Plotting a GM Plateau. This lab will have two sections, A and B. LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution This lab will have two sections, A and B. Students are supposed to write separate lab reports on section A and B, and submit the

More information

Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display.

Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display. Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display. Your Stack ST8961 VS module allows you to synchronize, overlay, and record data available on your Stack

More information

Calibrating and Profiling Your Monitor

Calibrating and Profiling Your Monitor Calibrating and Profiling Your Monitor For this module, you will need: Eye-One measurement device Counterweight (used for LCD screens only) New, modern displays are better First, you need to use a good

More information

HOW TO USE THE POLYCOM REMOTE CONTROL... 2 MAKING A CALL FROM THE ADDRESS BOOK... 3 ANSWERING A CALL... 4 HANGING UP A CALL... 4 REDIALING A CALL...

HOW TO USE THE POLYCOM REMOTE CONTROL... 2 MAKING A CALL FROM THE ADDRESS BOOK... 3 ANSWERING A CALL... 4 HANGING UP A CALL... 4 REDIALING A CALL... HOW TO USE THE POLYCOM REMOTE CONTROL... 2 MAKING A CALL FROM THE ADDRESS BOOK... 3 ANSWERING A CALL... 4 HANGING UP A CALL... 4 REDIALING A CALL... 4 CAMERA CONTROLS... 5 AUDIO CONTROLS... 5 VIDEO INPUT...

More information

Neutron Spectrometer Operation Manual

Neutron Spectrometer Operation Manual Neutron Spectrometer Operation Manual MIT Department of Physics (Dated: October 16, 2014) This document is for assisting in the understanding and accessing of the technical aspects of the neutron physics

More information

RQflex 10. Manual English

RQflex 10. Manual English RQflex 10 Manual English Contents 1 Introduction 3 2 Description of the instrument 5 display, keys 6 technical data 8 3 Setting up the instrument 9 insertion of batteries 10 setting the time and date 11

More information

UNITY 2 TM. Version 1.0. February 2008

UNITY 2 TM. Version 1.0. February 2008 UNITY 2 TM ULTRA 50:50 Operators Manual Version 1.0 February 2008 1. Loading tubes onto ULTRA 50:50...3 1.1. DiffLok analytical end caps for sample tubes...3 1.2. Capping tubes with DiffLok caps...3 1.3.

More information

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Please read the user manual before operation. PAGE 1 Content 1. Company Profile... 3 2. Product Introduction... 3 3. Operation Instruction...

More information

LedSet User s Manual V Official website: 1 /

LedSet User s Manual V Official website:   1 / LedSet User s Manual V2.6.1 1 / 42 20171123 Contents 1. Interface... 3 1.1. Option Menu... 4 1.1.1. Screen Configuration... 4 1.1.1.1. Instruction to Sender/ Receiver/ Display Connection... 4 1.1.1.2.

More information

NDT Supply.com 7952 Nieman Road Lenexa, KS USA

NDT Supply.com 7952 Nieman Road Lenexa, KS USA ETher ETherCheck Combined Eddy Current & Bond Testing Flaw Detector The ETherCheck is a combined Eddy Current and Bond Testing Flaw Detector which comes with a rich range of features offered by a best

More information

Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators. Technical Brief

Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators. Technical Brief Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators Technical Brief Quickly and easily apply a level correction table to compensate for external losses or power variations

More information

Class Notes for Cite While You Write Basics. EndNote Training

Class Notes for Cite While You Write Basics. EndNote Training Class Notes for Cite While You Write Basics EndNote Training EndNote X8 Class Notes for Cite While You Write Basics 1 January 3, 2017 Your EndNote data, both on the desktop and online, can be used in Microsoft

More information

Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual

Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual Instructions for use of Products AS9100, AS9101, AS9200, AS9201 TM526 Revised 2/19 Table of Contents 1 Description...2 2...3

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

There are three categories of unique transitions to choose from, all of which can be found on the Transitions tab:

There are three categories of unique transitions to choose from, all of which can be found on the Transitions tab: PowerPoint 2013 Applying Transitions Introduction If you've ever seen a PowerPoint presentation that had special effects between each slide, you've seen slide transitions. A transition can be as simple

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

Character Users Guide

Character Users Guide Cha r a c t e r Us e r sgui de Character Users Guide Metric Halo $Revision: 1619 $ Publication date $Date: 2012-02-10 20:41:00-0400 (Friday, 10 Feb 2012) $ Copyright 2011 Metric Halo Table of Contents

More information

Multi channel programmable Filter-Amplifier

Multi channel programmable Filter-Amplifier Operating instructions VS 21 Multi channel programmable Filter-Amplifier Operation LED Handset OK 41A For digital and analogue channels 6 terrestrial inputs: B I-II / B III / VHF-UHF and 3 UHF inputs over

More information

Florida State University Thayumanasamy Somasundaram

Florida State University Thayumanasamy Somasundaram 2016 Florida State University 2013-16 Thayumanasamy Somasundaram [PANALYTICAL X-PERT PRO POWDER DATA COLLECTION] A quick start-up procedure for collecting powder x-ray diffraction data from PANalytical

More information

CLIPSTER. 3D LUT File Generation with the Kodak Display Manager. Supplement

CLIPSTER. 3D LUT File Generation with the Kodak Display Manager. Supplement Supplement: CLIPSTER 3D LUT File Generation with the Kodak Display Manager (Version 1.0) CLIPSTER 3D LUT File Generation with the Kodak Display Manager Supplement Supplement for the CLIPSTER Documentation:

More information

when it comes to quality! BMR GmbH 1

when it comes to quality! BMR GmbH 1 when it comes to quality! BMR GmbH 1 2 DressView Dressing systems Issue June 2016 1 Key functions 2 2 Menu structure 3 2.1 Main-menu 4 2.2 Terminal-menu 5 2.2.1 Adjusting the rotational speed in Terminal-menu

More information

Instructions for DataMark PDF Temperature Logger

Instructions for DataMark PDF Temperature Logger Instructions for DataMark PDF Temperature Logger DataMark provides a convenient way to obtain high accuracy temperature data in transit. The unit is a self contained temperature logger that plugs directly

More information

HV/PHA Adjustment (PB) Part

HV/PHA Adjustment (PB) Part HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i

More information

XRD-DSC Sample Alignment (BB) Part

XRD-DSC Sample Alignment (BB) Part XRD-DSC Sample Alignment (BB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 5 2. Sample alignment sequence...9

More information

May 2006 Edition /A. Getting Started Guide for the VSX Series Version 8.5

May 2006 Edition /A. Getting Started Guide for the VSX Series Version 8.5 May 2006 Edition 3725-21286-008/A Getting Started Guide for the VSX Series Version 8.5 GETTING STARTED GUIDE FOR THE VSX SERIES Trademark Information Polycom, the Polycom logo design, and ViewStation are

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes Welch Allyn CardioPerfect Workstation Tango+ Interface Notes To setup Tango+ with the CardioPerfect stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232

More information

SIDRA INTERSECTION 8.0 UPDATE HISTORY

SIDRA INTERSECTION 8.0 UPDATE HISTORY Akcelik & Associates Pty Ltd PO Box 1075G, Greythorn, Vic 3104 AUSTRALIA ABN 79 088 889 687 For all technical support, sales support and general enquiries: support.sidrasolutions.com SIDRA INTERSECTION

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

OCT 15 Rev B

OCT 15 Rev B AMP+ HVA280-3pxm XE High-Voltage Plug Connector with 1-Stage Latching Application Specification 114-32125 15 OCT 15 Rev B NOTE All numerical values are in metric units [with U.S. customary units in brackets].

More information

Procedures for conducting User QA on the scanner

Procedures for conducting User QA on the scanner Procedures for conducting User QA on the scanner Sample setup: The User QA phantom is clearly labeled and is stored on one of the shelves to the side of the magnet. Note proper orientation of the bottle,

More information

Crimp & Cleave Termination Instructions for SEL ST Connectors

Crimp & Cleave Termination Instructions for SEL ST Connectors Your Optical Fiber Solutions Partner Crimp & Cleave Termination Instructions for SEL ST Connectors For Use With: ST Termination Kit (SEL, Part Number BT05402-01) 200 µm HCS Fiber-Optic Cable ST Crimp &

More information

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Last updated: October 12, 2011 by Josh Kurutz THIS PAGE = QUICK START GUIDE 0) At the computer, make sure VNMRJ is

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

Quick Help Teaching Room Technology Support

Quick Help Teaching Room Technology Support Quick Help Teaching Room Technology Support Technical assistance is available. If you require assistance, please call Ext 6066 Quick Help Technology Overview INDEX INDEX Touch Screen Is not active 3 Technology

More information

EndNote Tutorial Handout Table of Contents

EndNote Tutorial Handout Table of Contents EndNote Tutorial Handout Table of Contents What is EndNote?... 2 Getting Started... 2 Create a New Library... 2 EndNote Interface... 3 Set the Reference Style... 4 Adding References to Your Library...

More information

WAVES H-EQ HYBRID EQUALIZER USER GUIDE

WAVES H-EQ HYBRID EQUALIZER USER GUIDE WAVES H-EQ HYBRID EQUALIZER USER GUIDE TABLE OF CONTENTS CHAPTER 1 INTRODUCTION...3 1.1 WELCOME...3 1.2 PRODUCT OVERVIEW...3 1.3 CONCEPTS AND TERMINOLOGY...4 1.4 COMPONENTS...7 CHAPTER 2 QUICK START GUIDE...8

More information

Library ebooks and Your Sony ereader

Library ebooks and Your Sony ereader Library ebooks and Your Sony ereader When using your Sony ereader for the first time to obtain library ebooks and audiobooks, download the Overdrive Media Console, Adobe Digital Editions, and the Sony

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

Calibrating the timecode signal input

Calibrating the timecode signal input Chapter 5 Calibrating the timecode signal input Computer hardware can introduce an offset between the timecode signal and the video signal, which causes the timecode and video to be offset when they are

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

Make Music Cards. Choose instruments, add sounds, and press keys to play music. scratch.mit.edu. Set of 9 cards

Make Music Cards. Choose instruments, add sounds, and press keys to play music. scratch.mit.edu. Set of 9 cards Make Music Cards Choose instruments, add sounds, and press keys to play music. Set of 9 cards Make Music Cards Try these cards in any order: Play a Drum Make a Rhythm Animate a Drum Make a Melody Play

More information

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt

More information

Electronic Roll Feed (ERF) Registration System

Electronic Roll Feed (ERF) Registration System 3 Electronic Roll Feed (ERF) Registration System Information Folder 9.6 May 2014 Replaces IF 9.6 dated February 2013 Safety Blanking Die Guard: Fabricated guards enclose the blanking die to prevent accidental

More information

A BEGINNER'S GUIDE TO ENDNOTE ONLINE

A BEGINNER'S GUIDE TO ENDNOTE ONLINE A BEGINNER'S GUIDE TO ENDNOTE ONLINE EndNote Online is a free tool which can help you collect, share, and organise your references. This tutorial will teach you how to use EndNote Online by guiding you

More information

Setup Guide. CalMAN Client for SCRATCH. Rev. 1.1

Setup Guide. CalMAN Client for SCRATCH. Rev. 1.1 Setup Guide CalMAN Client for SCRATCH Rev. 1.1 Introduction CalMAN Required Software Version: CalMAN Display Calibration Software interfaces directly with ASSIMILATE SCRATCH software through the CalMAN

More information

TVS-2-TAIT VPU-15-TAIT

TVS-2-TAIT VPU-15-TAIT TVS-2-TAIT High Level Rolling Code Scrambler for the Tait Orca & TM-8000 Radios VPU-15-TAIT Voice Inversion Scrambler for the Tait Orca & TM-8000 Radios Manual Revision: 2018-03-05 Rev B Covers Software

More information

RADIO FREQUENCY SYSTEMS

RADIO FREQUENCY SYSTEMS RADIO FREQUENCY SYSTEMS Optimizer RT FAQ s Q. What information is require before running the software? The Serial Number of each ACU MUST be recorded with the Model number of the antenna that it is attached

More information

Tutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program

Tutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program Tutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program Step cycles are defined usually by choosing a recorded ENG waveform that shows long lasting, continuos, consistently

More information

Introduction to EndNote Online

Introduction to EndNote Online Introduction to EndNote Online Creating an EndNote Online account Go to EndNote Online. Click on the Access EndNote Online button and, if prompted, enter your Warwick username and password to confirm you

More information

Spooky2 Scalar Digitizer

Spooky2 Scalar Digitizer Spooky2 Scalar Digitizer INSTRUCTION November 2018 Spooky2 Team Brief Overview Biofeedback is a way of finding out what frequencies are best suited to your body. Spooky2 Scalar is the only Scalar device

More information

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English TOUCH SCREEN CONTROLLER for Air Conditioning Control System Model BMS-CT5120UL English Contents 1 Precautions for safety.................................................. 5 2 Main functions........................................................

More information

HD Digital Set-Top Box Quick Start Guide

HD Digital Set-Top Box Quick Start Guide HD Digital Set-Top Box Quick Start Guide Eagle Communications HD Digital Set-Top Box Important Safety Instructions WARNING TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE THIS PRODUCT TO RAIN

More information

User Manual OVP Raman

User Manual OVP Raman Version 6 User Manual OVP Raman 2006 BRUKER OPTIK GmbH, Rudolf-Plank-Straße 27, D-76275 Ettlingen, www.brukeroptics.com All rights reserved. No part of this manual may be reproduced or transmitted in any

More information

Casambi App FAQ. Version Casambi Technologies Oy.

Casambi App FAQ. Version Casambi Technologies Oy. Casambi App FAQ Version 1.3 30.9.2016 Casambi Technologies Oy 1 of 12 GENERAL 3 Q: What is Casambi app used for? 3 Q: Which mobile devices are supported? 3 Q: Where can I get the Casambi app? 3 FIRST TIME

More information

Precision DeEsser Users Guide

Precision DeEsser Users Guide Precision DeEsser Users Guide Metric Halo $Revision: 1670 $ Publication date $Date: 2012-05-01 13:50:00-0400 (Tue, 01 May 2012) $ Copyright 2012 Metric Halo. MH Production Bundle, ChannelStrip 3, Character,

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Cover Page for Lab Report Group Portion. Boundary Layer Measurements Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:

More information

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read

More information

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT X-Ray Reflectivity using the PANalytical X Pert Pro MPD Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray Modified for configuration used at University

More information

Video Wall Display User s Manual

Video Wall Display User s Manual Video Wall Display User s Manual Manual Version TL46-55H1.1 Our Full Product Range Page 1 Safety Instructions Please handle the display with extreme care, significant impact will damage the LCD panel.

More information

Operating Instructions for Scanmark insight 4ES Scanner with QuickScore

Operating Instructions for Scanmark insight 4ES Scanner with QuickScore Operating Instructions for Scanmark insight 4ES Scanner with QuickScore This document will guide you through the process of scanning and grading exams and printing their results. It has been written specifically

More information

Pulse Biomedical QRS-Card Tango M2 Interface Notes

Pulse Biomedical QRS-Card Tango M2 Interface Notes Pulse Biomedical QRS-Card Tango M2 Interface Notes To setup Tango M2 with your stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232 Cable used to communicate

More information

Welcome to the U-verse App

Welcome to the U-verse App iphone 2.5.3 Welcome to the U-verse App The U-verse app is an AT&T service that uses your iphone to provide a user interface for U-verse TV. Using Edge, 3G and WiFi technology, the U-verse app provides

More information

Digital Compass. Package Contents. Hardware Overview. Basic Operation. Article: CNS-DC3 Quick Guide v1.0

Digital Compass. Package Contents. Hardware Overview. Basic Operation. Article: CNS-DC3 Quick Guide v1.0 Digital Compass Article: CNS-DC3 Quick Guide v1.0 Package Contents 1. CANYON CNS-DC3 Main Unit 2. Documentation CD 3. User Quick Guide Hardware Overview [SR] Button [M] Button [L/R] Button [S/S] Button

More information

Lab Determining the Screen Resolution of a Computer

Lab Determining the Screen Resolution of a Computer Lab 1.3.3 Determining the Screen Resolution of a Computer Objectives Determine the current screen resolution of a PC monitor. Determine the maximum resolution for the highest color quality. Calculate the

More information

Analog Discovery Scope and Waveform Generator Edited 11/15/2016 by Eric Scotti & DGH

Analog Discovery Scope and Waveform Generator Edited 11/15/2016 by Eric Scotti & DGH Analog Discovery Scope and Waveform Generator Edited 11/15/2016 by Eric Scotti & DGH Specifications The Analog Discovery contains several devices but we will likely only use the 2 channel oscilloscope

More information

Syntor X Flash Memory Module Revision C

Syntor X Flash Memory Module Revision C Syntor X Flash Memory Module Revision C The PIEXX SynXFlash memory module, along with the supplied PC software, replaces the original SyntorX code plugs and allows you to easily set modify and update your

More information

GS122-2L. About the speakers:

GS122-2L. About the speakers: Dan Leighton DL Consulting Andrea Bell GS122-2L A growing number of utilities are adapting Autodesk Utility Design (AUD) as their primary design tool for electrical utilities. You will learn the basics

More information