Photon detectors. J. Va vra SLAC

Size: px
Start display at page:

Download "Photon detectors. J. Va vra SLAC"

Transcription

1 Photon detectors J. Va vra SLAC

2 Content Comment on timing strategies Vacuum-based detectors: - Hamamatsu MaPMTs - Burle MCP-PMTs with 25 and 10 µm dia. holes Gaseous-based detectors: - Micromegas + MCP Future developments 12/4/05 J.Va'vra, Japan

3 What detector do we want? Present prototype: Future Fast Focusing DIRC: We want to measure x, y and TOP (timeof-propagation) for each photon. We need a single photon timing resolution at a level of ~ ps, to be able to perform the TOP measurement and correct the chromatic error contribution to the Cherenkov angle. We need to operate at 15kG for the Super B-factory, or even at higher field, if the device would find a use at ILC. We want to have a highly pixilated detector. We started with a square pixel size of ~6x6mm. Now we aim for a rectangular size of ~2x8mm. The detector should have a good aging performance. 12/4/05 J.Va'vra, Japan

4 Single photoelectron timing resolution at B = 0 kg

5 High Resolution Timing We have tried these timing techniques: - Leading edge discriminator + single TDC + ADC correction - Constant fraction discriminator (CFD) + single TDC - Two leading edge discriminators with two TDCs per channel Note: There is no evidence that one method is better than the others. We have chosen the CFD method for the Focusing DIRC prototype. But, in retrospect, I think that for alarge scale system, the double-threshold + two TDCs might be a better. Amplifier rise time must be comparable to the photon detector s rise time, and both have to be fast. Need to have expensive tools: - PiLas laser diode with 35ps FWHM timing capability - Fast SiPMT to verify its correct timing operation - 2D-scanning setup to measure a PMT response across its face 12/4/05 J.Va'vra, Japan

6 Speed of the amplifier & detector is essential for good timing From V. Radeka talk at RICH /4/05 J.Va'vra, Japan

7 Examples of two amplifiers Elantek amplifier: - Gain ~130x, MCP-PMT with 25µm holes connected - A ~ 5mV - (ds o /dt) t=0 ~ 0.3V/1ns - t ~ (5x10-3 /0.3)*1ns ~15-20ps Ortec VT-120A amplifier: - Gain ~200x, MCP-PMT with 10µm holes connected - A ~ 5mV - (ds o /dt) t=0 ~ 1.2V/1ns - t ~ (5x10-3 /1.2)*1.0ns ~ 4-5ps Both amplifiers will do the excellent job from noise point of view. However, the Ortec VT-120A amp is much better match for the speed of the MCP-PMT with 10 µm holes. 12/4/05 J.Va'vra, Japan

8 PiLas laser diode and fiber optics Achieved ~ 40-70ps with: - 635, 430 and 407nm wavelengths - 63µm dimeter multi-mode fiber - 5 & 10 m fiber lengths - 1-to-3 fiber splitter - Home-made alignment with the x&y small stage - Mylar attenuators to get single photons - CFD discriminator or TDC/ADC electronics 12/4/05 J.Va'vra, Japan

9 Use a SiPMT detector to verify that the PiLas laser diode SiPMT: CFD analog out, 1ns/div: Use this one in this test Detector: 100 µm dia. GaP SiPMT (APD) operating in a Geiger mode with active quenching. APD developed by Sopko & Prochazka, CVUT Prague. The authors quote this timing resolution: diode ~ (FWHM = 58/2.35) ~ 25 ps for the single photoelectron regime. Therefore, we expect: PiLas ~ sqrt( result2 - APD2 - electronics2 ) ~ sqrt( ) ~23 ps; PiLas data sheet quotes: (35/2.35) ~15ps) - a small inconsistency due to some systematic error ( PiLas power set to ~11% might be too low). Electronics chain in this test: SLAC CFD, 30mV threshold, CFD analog output to the LeCroy 2228ATDC (25ps/count). 12/4/05 J.Va'vra, Japan

10 Hamamatsu H-8500 Flat panel MaPMT Hamamatsu Co. data sheet + SLAC measurements + my interpretation 12/4/05 J.Va'vra, Japan

11 Burle MCP-PMT parameter list Burle Co. data sheet + SLAC measurements + my interpretation!! 12/4/05 J.Va'vra, Japan

12 Timing studies in MaPMT and MCP-PMT Hamamatsu Flat Panel H8500 PMT: Burle MCP-PMT: MaPMT #2 MCP-PMT #3 Double Gaussian fit Burle MCP-PMT #3 has a very long tail due to recoil electrons from the MCP top surface. The tail contains ~20% of all events!!! The MCP-to-cathode distance is 6-7mm. Electronics chain used in this test: Final SLAC amplifier, final SLAC CFD providing the analog output to LeCroy 2228A TDC (25ps/count). Light source: Use the 635nm PiLas laser diode in a single photoelectron mode. 12/4/05 J.Va'vra, Japan

13 Dependence on the MCP PMT design Old design ( ): MCP-to-Cathode distance = 6 mm New design ( ): MCP-to-Cathode distance = 0.75 mm MCP-PMT #3 MCP-PMT #16 Double Gaussian fits. The reduction of the MCP-to-Cathode distance to 0.75mm limits the rate of recoiling photoelectrons from the MCP surface, which reduces the tail in the timing spectrum. These electrons are, however, lost from the detection efficiency, but the spectrum is more Gaussian. Nevertheless, tails would complicate the analysis, and we prefer to cut them. Electronics chain used in this test: Final SLAC amplifier, final SLAC CFD, LeCroy 2228A TDC (25ps/count). Light source: PiLas laser diode in the single photoelectron mode (635nm). 12/4/05 J.Va'vra, Japan

14 Ideal goal: no tails in the distributions MCP-PMT #16 (64 pixels) Double Gaussian fits. No tail in this type of MCP-PMT. Some pixels are better than others. Not clear why. 12/4/05 J.Va'vra, Japan

15 However, the new tube is inefficient around the edges New design ( ): MCP-to-Cathode distance = 0.75 mm The efficiency drops to zero half way through all edge pads. This inefficiency is related to the electrostatic design near the edges. Perhaps, one can have a small light collector around the boundary 12/4/05 J.Va'vra, Japan

16 Compare timing distribution on two different pads with the Phillips 7186 TDC MCP-PMT #16, Pad 14: MCP-PMT #16, Pad 24: Single Gaussian fit to the timing distribution generated in each laser head location. Measure typically = 70-80ps in the central pad region, slightly worse near the boundary. Worse timing resolution around edges is due to the charge sharing, causing lower pulse height, and possibly a cross-talk from hits in neighboring pads. Electronics chain in this test: final SLAC amplifier, final SLAC 32-channel CFD, Phillips 7186 TDC (25ps/count). Detector in this test: MCP-PMT #16 with MCP-to-Cathode distance of 750µm, 8x8 pads, 2.6kV. Light source in this test: PiLas laser diode in the single photoelectron mode (635nm). 12/4/05 J.Va'vra, Japan

17 Single photoelectron timing resolution at B = 15 kg 12/4/05 J.Va'vra, Japan

18 Burle MCP-PMT with 10µm holes 4-pixel MCP-PMT P01 tube for the initial tests. PMT has two MCPs with 10 µm dia. holes Cathode-to-MCP distance ~6mm According to Burle, this particular 10µm MCP should produce a gain of ~10 6 at 2.2kV. Setup had a capability to measure sensitivity to angles in 5 o steps between the magnetic field and axis perpendicular to the face plate. 12/4/05 J.Va'vra, Japan

19 Choice of amplifier and timing results at B = 0 kg 500mV/div, 1ns/div, 2.2kV: Ortec VT-120A amplifier, gain of 200x, (ds o /dt) t=0 ~ 1.2V/1ns Philips CFD discriminator and LeCroy TDC with 25ps/count. Elantek 130x amplifier with 1.5ns risetime gives a smaller pulse height. The detector controls the choice of amplifier: If the amplifier is too slow compared to the detector, one reduces the maximum peak amplitude for a given gain. On the other hand, if the amplifier is much faster than the detector, one increases the noise. 12/4/05 J.Va'vra, Japan

20 Timing results at B = 15 kg 2.7kV Ortec VT-120A amp Initially, there was some confusion what the maximum allowed voltage. Burle initially thought that it is -2.4kV. After I have overvoltaged the tube to -2.7kV to get a decent timing result at 15kG, Burle corrected the max voltage value to kV. I could have gone higher. This means that it is possible to reach a resolution of ~50ps at 15kG. 12/4/05 J.Va'vra, Japan

21 Sensitivity to MCP voltage at B = 15kG Ortec VT-120A amp, -2.65kV, 50mV/div, 1ns/div: The necessary voltage to get a good timing resolution is kV. 12/4/05 J.Va'vra, Japan

22 Sensitivity to angular rotation at B = 15kG Ortec VT-120A amp, -2.65kV, 100mV/div, 1ns/div: The MCP can be tilted by 3-5 o before pulse height is affected. At 10 o, one sees a clear reduction of pulse height, but the tube can still be used. At 15 o and above, the response is killed entirely. 12/4/05 J.Va'vra, Japan

23 Single photoelectron spatial response at B = 0 kg 12/4/05 J.Va'vra, Japan

24 Scanning setup to measure the PMT spatial response x&y stage for the fiber final focus : Stepper motor moves the end of the fiber equipped with a lens, resulting in the spot size of ~150 µm. The linear motor is set typically to: x-step ~ 100µm & y-step ~ 1mm. Light source: - PiLas laser diode operating in single photoelectron mode & 430 nm (on loan) & 407 nm (now). - Fiber is 63µm dia. multi-mode fiber, equipped with lenses at both ends. Analysis: - A hit is accepted into the efficiency definition if it is within a time window, and it is on the same pad as the laser head is pointing to. - To get a relative efficiency we normalize to the 2 inch dia. Photonis XP 2262B PMT ( or the DIRC PMT, ETL 9125FLB17). - DAQ trigger rate: 20kHz.. 12/4/05 J.Va'vra, Japan

25 Resolution of the scanning system Hamamatsu Flat Panel H8500 MaPMT #2: Micro-structure of the dynode electrodes: Resolution: Clearly see the details of the dynode electrode structure. Spatial resolution of the system is less than 100 µm, for a step size of 25µm. Electronics chain used in this test: Final SLAC amplifier, LeCroy 4413 discriminators with 100mV threshold, LeCroy 3377 TDCs with 0.5ns/count. 12/4/05 J.Va'vra, Japan

26 An example of the relative response along a line scan across eight pads Hamamatsu Flat Panel H8500 PMT #2: Burle MCP-PMT #3: The Hamamatsu MaPMT uniformity is ~1:2.5 and the Burle MCP-PMT uniformity is ~1:1.5, in this example. Electronics chain used in this test: Final SLAC amplifier, LeCroy 4413 discriminators with 100mV threshold, LeCroy 3377 TDCs with 0.5ns/count. 12/4/05 J.Va'vra, Japan

27 Burle MCP-PMT #8 relative detection efficiency (Normalized to the Photonis XP 2262B PMT) 635nm: At 635nm, which is close to the end of the Bialkali Q.E. range, the relative efficiency scaling to the Photonis PMT is not very reliable. 430nm: At 430nm, the relative efficiency is 50-60% relative to the Photonis PMT, if we include the late arrivals. This is approximately expected based on the MCP design (to be compared with the geometrical MCP collection efficiency (cathode-to-top MCP) of 60-65%, shown on page 6). Electronics chain used in this test: Final SLAC amplifier, LeCroy 4413 discriminators with 100mV threshold, LeCroy 3377 TDCs with 0.5ns/count Light source: PiLas laser diodes operating in the single photoelectron mode (635nm & 430nm). 12/4/05 J.Va'vra, Japan

28 Relative response across the MCP-PMT face Burle MCP-PMT # nm: 430nm: Burle MCP-PMT # nm: 430nm: Burle MCP-PMT # nm: 430nm: Burle MCP-PMT # nm: 430nm: Typical relative efficiency is 50-60% of the 2 inch dia. Photonis XP 2262B PMT at 430nm. The efficiency drops to 30-50% around the edges at 430nm. 12/4/05 J.Va'vra, Japan

29 Relative response across the MaPMT face Burle MCP-PMT # nm: 430nm: Hamamatsu MaPMT #1 635 nm: 430nm: Hamamatsu MaPMT #2 635 nm: 430nm: Hamamatsu MaPMT #4 635 nm: 430nm: MCP-PMT #16 has large inefficiency around edges (it has the MCP-to-cathode distance of 0.75 mm). Hamamatsu Flat Panel MaPMT relative efficiency is 50-70% of the Photonis XP 2262B PMT at 430nm. The efficiency drops to 30-50% around the edges at 430nm. 12/4/05 J.Va'vra, Japan

30 Aging of MCP-PMT

31 Aging of MCP-PMT Aging due to damage of the photocathode by ion bombardment. Burle claims a ~50% loss after of ~10 C/25cm 2 area of MCP-PMT. Example: DIRC single photon background rate is: ~200 khz per 1 dia PMT at a luminosity of ~10 34 cm -2 sec -1. If I assume that ~1/3 comes from the bar, we run ~6 months/year, then after 10 years, I get about ~10 13 pe - /cm 2. This translates to ~ 1-2 C/25cm 2, if we would have the MCP-PMTs in the present DIRC. The rate is dominated by the LUMI-term, caused by the radiative Bhabhas striking beam components. Nobody knows how to scale things for the Super B factory with a luminosity of > cm -2 sec -1, however, it is clear that one has to pay attention to this problem. 12/4/05 J.Va'vra, Japan

32 Coherent resonance effects? (observed in the prototype) This is what may happen when one tries to be too fast 12/4/05 J.Va'vra, Japan

33 Coherent excitation resonance effects During the run we typically get 3-4 Cherenkov photons, which do not arrive at the same time, so we probably do not suffer from this problem. However, this needs to be fixed. The effect generated by a PiLas producing enough light that multiple pixels fire. At a power of 25% we get a 10% probability to get a hit, which means that something like 6-7 pixels fire per one PiLas trigger. The pulses arrive to the MCP-PMT within < 1 ns, and are capable to excite the standing resonance. 12/4/05 J.Va'vra, Japan

34 Coherent excitation resonance effects The effect does not exist with the Hamamatsu MaPMTs (the same amplifier, the same LV PS, the same grounding). 12/4/05 J.Va'vra, Japan

35 Future developments with the non-gaseous detectors 12/4/05 J.Va'vra, Japan

36 R&D related to Focusing DIRC Develop rectangular pads of 2mm x 8mm, or 3mm x 12mm in size. Suppress the timing tails by reducing the gap between the photocathode and MCP surface. Do more tests with 10µm dia. hole MCP-PMTs in the magnet and estimate better the max possible field. Test the timing with a gaseous MCP + Micromegas photon detector equipped with the Bialkali photocathode. SiPMTs? Rate and aging tests. 12/4/05 J.Va'vra, Japan

37 New 1024-pixel Burle MCP A proposal how to connect pads: Large rectangular pad: 2x8 little ones Small margin around boundary 1024 pixels (32 x 32 pattern) Small pixel size: ~1.4mm x 1.4mm Pitch: 1.6 mm 12/4/05 J.Va'vra, Japan

38 New 256-pixel Hamamatsu MaPMT H-9500 A proposal how to connect pads: Large rectangular pad: 1x4 little ones 256 pixels (16 x 16 pattern). Pixel size: 2.8 mm x 2.8 mm Pitch of 3.04 mm. Very neat connections 12/4/05 J.Va'vra, Japan

39 Open area Burle MCP Small margin around the boundary 10 & 25 µm MCP hole diameter 64 pixel devices Pad size: 6 mm x 6 mm. The MCP-PMT still has 6-7mm cathode-to-mcp distance, thus making a long tail in the timing distribution Can change the resistor chain. Will study if the tail can be supressed by a choice of the MCP operating voltages. Elantek amplifier may not work with a 10µm MCP-PMT. 12/4/05 J.Va'vra, Japan

40 Silicon PhotoMultiplier (SiPM) (R. Mirzoyan, Max-Planck Inst., IEEE 2005) 42 m SiPMT 1 mm 20 m 1 mm Each pixel = binary device 24*24=576 pixels SiPM = analogue detector Pixels of the SiPM Need a rectangular pixel of 2mm x 8 mm for the Focusing DIRC. Active area presently % Dark count rate/mm 2 : counts/sec at room temperature Single pixel recovery time ~1 sec. A high breakdown probability limits the photon efficiency to ~30% only. 12/4/05 J.Va'vra, Japan

41 Gaseous Micropattern detectors Can they play a role among the fast detectors? Yes, if one can demonstrate three things: (a) longevity of the Bialkali photocathode in the gas, (b) high gain operation, and (c) good timing resolution.

42 Modular setup to test various detector ideas Modular ring structure: Quadruple-GEM + pads: Geometries tested: - Quadruple-GEM + pads - Tripple MCP + pads - GEM + Micromegas + pads - MCP + Micromegas + pads MCP + Micromegas + pads: 12/4/05 J.Va'vra, Japan

43 Micromegas + MCP with a pad readout J.Va vra & T. Sumiyoshi, Nucl.Instr.&Meth. A, 435(2004)334. E Drift-1 E Drift-2 An example of running conditions: E Drift-1 ~350V/cm E MCP ~10kV/cm E Drift-2 ~1.25kV/cm E Micromegas ~50kV/cm Ave. total gain ~2x10 5 Gain distribution in final application: G Micromegas ~2x10 3, G MCP ~100 V Micromegas ~500V, dv MCP ~1200V Photocathode: Metal mesh + Xenon UV light Works well in the single electron mode 12/4/05 J.Va'vra, Japan

44 Mesh and MCP can be made very clean s.s. electro-mesh: MCP: 1000 lpi mesh density (lines per inch) A square hole dimension: ~17 x 17 µm 2 A sidewall width: ~9 µm Made by: BuckBee-Mears Co. A hole diameter: ~50 µm A sidewall width: ~12 µm Thickness: ~1mm Made by: Hamamatsu 12/4/05 J.Va'vra, Japan

45 A good single photoelectron response Vary Micromegas gain mainly: Vary only the MCP gain: Very stable operation even at very high gain in 89.1%He % ic4h10 gas. Observe a slight turnover in the pulse height spectrum. For comparison: Giomataris has observed a clear turnover with ~30% of ic 4 H 10 in the Micromegas alone: 70%He+30%iC 4 H 10 12/4/05 J.Va'vra, Japan

46 MCP with inclined holes + Micromegas J.Va vra & T. Sumiyoshi, NIM A, 435(2004)334 & RICH2004 MCP: 1 dia, 1mm thick, 50micron holes photon MCP with straight holes, B=0kG: MCP with inclined holes, B=15kG: I-Cathode, or I-anode [na] di-cathode di-anode B = 0 kg, V Micromegas = 400 V IBF ~10% Voltage across MCP [V] IBF ~0% IDEA: Block the ion backflow (IBF) by inclined MCP holes in a magnetic field IBF reduction by aligning the MCP holes with the electron s Lorenz angle. Electrons drift & amplify along the MCP hole; ions are caught on the MCP walls. The measured IBF with inclined holes is negligible (consistent with a pa noise). The measured IBF with MCP with the straight holes at a level of ~10%!! No data on electron collection eff. No charging effects observed, which would indicate that the electric field would align with the MCP hole direction. If that would happen, the idea would not work. 12/4/05 J.Va'vra, Japan

47 Inclined MCP holes In this test, use Hamamatsu MCP with a 50µm hole diameter and an angle of 6.5 o. The picture shows a cut through the MCP to verify the angle. The inclined holes are a standard MCP technology as all vacuum-bases MCP- PMTs use them to limit the ion damage of the photocathode 12/4/05 J.Va'vra, Japan

48 Lorenz angle calculation at 15 kg 15kG The MCP angle is fixed by a choice of gas and MCP gain. Use Magboltz program (version 7.1). 12/4/05 J.Va'vra, Japan

49 I-Cathode, or I-anode [na] Experimental setup in the magnet 15kG 90%Ar+10%CH 4 gas Mercury UV lamp MCP needs to be rotated to the optimum azimuth. Indeed, one measures nearly zero cathode backflow current, i.e., consistent with B = 15kG, MCP with 6.5 o hole angles, I-cathode 90%Ar+10%CH - Emcp = 11kV/cm 4 I-cathode - Emcp = 9kV/cm I-anode [na] - Emcp = 9kV/cm I-anode [na] - Emcp = 11kV/cm Arbitrary azimuthal angle [Degrees] a picoammeter noise) at the azimuth angle where the electron transfer is at maximum (aligned with the electron Lorenz angle). 12/4/05 J.Va'vra, Japan

50 Hamamatsu Bialkali GPM R&D work Sumiyoshi, Va vra, Tokanai & Hamamatsu Hamamatsu built a double-mesh Micromegas structure w bialkali pc. Works both in 90%Ar+10%CH 4 or 90%Ar+10%CF 4. No detorioration of the photocathode observed within 5 days Gain of ~6x10 3, limited by secondary effects. Not sufficient for singlephoton detection. Work with MCP & Bialkali photocathode is in progress. 12/4/05 J.Va'vra, Japan

51 Discussion of the gaseous detectors The micro-pattern gas detectors have good aging rate, and can handle high rates (ions travel a short distance). Gaseous detectors could work easily up to 60 kg. Vacuum MCP-PMT will not work much above B~15kG at present. Gaseous detectors can use rather large MCP hole diameter of ~50µm. One could presumably make a large size photon detector using a mosaic of MCPs. Higher geometrical efficiency compared to the vacuum-based MCP- PMTs, at least in principle. Vacuum MCP-PMT has ~50% geometrical efficiency at best. Timing: Giamataris has achieved ~300ps with a Micromegas covered with CsI with just a leading edge disriminator. Adding a MCP will make it worse. The question how much. Needs to be measured. We have invented a simple method to block the ion flow to the cathode. Needs to be studies in more detail using a good simulation code. 12/4/05 J.Va'vra, Japan

52 Conclusion A single photon timing resolution at a level of ~ ps is much closer to a reality compared to a situation when we started. But, much more has to be done. 12/4/05 J.Va'vra, Japan

53 Backup slides

54 Lorenz angle calculation at 15 kg Lorentz angle [Degrees] Lorentz angle [Degrees] B = 15kG, E vs. B angle: 90 o Electric field in MCP hole [kv/cm] B = 15kG, E MCP = 9kV/cm, E vs. B angle: 90 o Methane in Ar/CH4 mix [%] The MCP angle is fixed by a choice of gas and MCP gain. With 90%Ar+10%CH 4 gas & E = 9kV/cm & B = 15kG: v along_e = µm/ns v along_b = 4.21 µm/ns long_along_e ~ 106 µm 2 /ns transv_along_b ~ 245 µm 2 /ns Very high diffusion => expect losses along the MCP hole walls Use Magboltz program (version 7.1). Thanks to Steve Biagi for always making sure that (a) I do it right, and (b) use the latest version of the program. 12/4/05 J.Va'vra, Japan

55 Ion backflow at optimum azimuth is negligible 90%Ar + 10% CH 4 gas The magnitude of the ion backflow at optimum azimuth is zero, consistent with a picoammeter noise. 12/4/05 J.Va'vra, Japan

56 Hamamatsu work with the gaseous photodetectors with Bialkali photocathode So far, they built successfully a Double-mesh Micromegas. Works both in 90%Ar+10%CH 4 or 90%Ar+10%CF 4 gases. Gain of ~6x10 3 reached with a coarser mesh. Coarser mesh yields higher gain (Gain ~ 6x10 3 for 34µm pitch, and Gain ~2x10 3 for 25µm pitch). Not yet good enough for the single electron operation with a good timing resolution. The Micromegas+MCP with inclined holes will be done next. 12/4/05 J.Va'vra, Japan

57 Results with Double Micromegas and Bialkali photocathode in 90%Ar+10%CH 4 gas Cs 137 source, NaI(Tl) convertor, Double Micromegas, P-10 gas QE 90%Ar+10%CF 4 (works as P-10 gas) Serial No. ZX 978 Connect the NaI(Tl) crystal to the Double-mesh Micromegas photodetector operating in the P-10 gas, and with a Cs 137 source obtain the result shown above. QE of Bialkali photocathode in 90%Ar+10%CF 4 gas (the P-10 gas gives similar results): a) 20.8% in vacuum, b) 13% in the gas, c) 20.0% in vacuum again. Wavelength 12/4/05 J.Va'vra, Japan

58 Total gas gain in 94.5%He+5.5%CH 4 gas at 1 bar A factor of ~15 of gain increase every 100 Volts across either the Capillary or the Micromegas in this gas. Use a Mercury UV lamp to do this measurement. 12/4/05 J.Va'vra, Japan

59 Comments on the timing resolution Measurement to produce single electrons off the s.s. mesh using a PiLas laser diode (430nm) was not successful. So, I do not have a direct result, unfortunately. However, perhaps, one could argue theoretically as follows: - Let s assume that the MCP has an average gain of I will use this simple formula: t ~(1/ N) coll /v drift where N = 50, coll =1/ is mean free path ( is Townsend coeff.) and v drift is electron drift velocity in the Micromegas at ~50kV/cm. - Using the Magboltz-Monte program, one obtains t < 100ps for a 90%He+10%CH 4 gas. - However, in addition, there are avalanche fluctuations, which will make it worse. The MCP will also add tails to the timing distribution. 12/4/05 J.Va'vra, Japan

60 How quickly is the ion removed from the insulator? Current [pa] ( ~50 sec) Single-Capillary+Micromegas Quadruple GEM Expon. (Quadruple GEM) Expon. (Single-Capillary+Micromegas) ( ~85 sec) Time [sec] The remnant charge is removed from the insulators of the detector (Kapton or Glass) with a time constant ~85sec for Quadruple-GEM, vs. ~50sec for the Single Capillary + Micromegas detector. Use a Mercury UV lamp (detector draws ~ 350nA). At that point switch lamp off and measure a discharge time constant of the decaying photocurrent. 12/4/05 J.Va'vra, Japan

Spatial Response of Photon Detectors used in the Focusing DIRC prototype

Spatial Response of Photon Detectors used in the Focusing DIRC prototype Spatial Response of Photon Detectors used in the Focusing DIRC prototype C. Field, T. Hadig, David W.G.S. Leith, G. Mazaheri, B. Ratcliff, J. Schwiening, J. Uher, J. Va vra SLAC 11/26/04 Presented by J.

More information

Study of Timing and Efficiency Properties of Multi-Anode Photomultipliers

Study of Timing and Efficiency Properties of Multi-Anode Photomultipliers Study of Timing and Efficiency Properties of Multi-Anode Photomultipliers T. Hadig, C.R. Field, D.W.G.S. Leith, G. Mazaheri, B.N. Ratcliff, J. Schwiening, J. Uher, J. Va vra Stanford Linear Accelerator

More information

Focusing DIRC R&D. J. Va vra, SLAC

Focusing DIRC R&D. J. Va vra, SLAC Focusing DIRC R&D J. Va vra, Collaboration to develop the Focusing DIRC: I. Bedajanek, J. Benitez, M. Barnyakov, J. Coleman, C. Field, David W.G.S. Leith, G. Mazaheri, B. Ratcliff, J. Schwiening, K. Suzuki,

More information

Table. J. Va vra,

Table. J. Va vra, J. Va vra, 7.12.2006 Table - Charge distribution spread in anode plane - Size of MCP holes - MCP thickness - PC-MCP-IN and MCP-OUT-anode gaps - Pad size and the grid line width - Photocathode choice 1

More information

Lifetime of MCP-PMTs

Lifetime of MCP-PMTs Lifetime of MCP-PMTs, Alexander Britting, Wolfgang Eyrich, Fred Uhlig (Universität Erlangen-Nürnberg) Motivation A few pros and cons of MCP-PMTs Approaches to increase lifetime Results of aging tests Outlook

More information

Performance of the MCP-PMT for the Belle II TOP counter

Performance of the MCP-PMT for the Belle II TOP counter Performance of the MCP-PMT for the Belle II TOP counter Kodai Matsuoka (KMI, Nagoya Univ.) S. Hirose, T. Iijima, K. Inami, Y. Kato, Y. Maeda, R. Mizuno, Y. Sato, K. Suzuki (Nagoya Univ.) TOP (Time Of Propagation)

More information

Tests of Timing Properties of Silicon Photomultipliers

Tests of Timing Properties of Silicon Photomultipliers FERMILAB-PUB-10-052-PPD SLAC-PUB-14599 Tests of Timing Properties of Silicon Photomultipliers A. Ronzhin a, M. Albrow a, K. Byrum b, M. Demarteau a, S. Los a, E. May b, E. Ramberg a, J. Va vra d, A. Zatserklyaniy

More information

HAPD and Electronics Updates

HAPD and Electronics Updates S. Nishida KEK 3rd Open Meeting for Belle II Collaboration 1 Contents Frontend Electronics Neutron Irradiation News from Hamamtsu 2 144ch HAPD HAPD (Hybrid Avalanche Photo Detector) photon bi alkali photocathode

More information

Solid State Photon-Counters

Solid State Photon-Counters Solid State Photon-Counters GMAPD (Geiger Mode Avalanche PhotoDiode) SiPM (Silicon Photo-Multiplier) Single element Photon Counter Multi Pixel Photon Counter 1-cell n-cells charge = k charge = nk Giovanni

More information

MCP Upgrade: Transmission Line and Pore Importance

MCP Upgrade: Transmission Line and Pore Importance MCP Upgrade: Transmission Line and Pore Importance Tyler Natoli For the PSEC Timing Project Advisor: Henry Frisch June 3, 2009 Abstract In order to take advantage of all of the benefits of Multi-Channel

More information

The TORCH PMT: A close packing, multi-anode, long life MCP-PMT for Cherenkov applications

The TORCH PMT: A close packing, multi-anode, long life MCP-PMT for Cherenkov applications The TORCH PMT: A close packing, multi-anode, long life MCP-PMT for Cherenkov applications James Milnes Tom Conneely 1 page 1 Photek MCP-PMTs Photek currently manufacture the fastest PMTs in the world in

More information

Production and Development status of MPPC

Production and Development status of MPPC Production and Development status of MPPC Kazuhisa Yamamura 1 Solid State Division, Hamamatsu Photonics K.K. Hamamatsu-City, 435-8558 Japan iliation E-mail: yamamura@ssd.hpk.co.jp Kenichi Sato, Shogo Kamakura

More information

PID summary J. Va vra

PID summary J. Va vra PID summary J. Va vra SuperB collaboration meeting in London, 2011 Speakers Barrel FDIRC - Jerry Va vra: Update on FDIRC prototype - Christophe Beigbeder: Barrel electronics status - Jerry Va vra: Comment

More information

MCP Signal Extraction and Timing Studies. Kurtis Nishimura University of Hawaii LAPPD Collaboration Meeting June 11, 2010

MCP Signal Extraction and Timing Studies. Kurtis Nishimura University of Hawaii LAPPD Collaboration Meeting June 11, 2010 MCP Signal Extraction and Timing Studies Kurtis Nishimura University of Hawaii LAPPD Collaboration Meeting June 11, 2010 Outline Studying algorithms to process pulses from MCP devices. With the goal of

More information

SLAC Cosmic Ray Telescope Facility

SLAC Cosmic Ray Telescope Facility SLAC Cosmic Ray Telescope Facility SLAC-PUB-13873 January 8, 2010 J. Va vra SLAC National Accelerator Laboratory, CA, USA Abstract SLAC does not have a test beam for the HEP detector development at present.

More information

Updates on the Central TOF System for the CLAS12 detector

Updates on the Central TOF System for the CLAS12 detector Updates on the Central TOF System for the CLAS1 detector First measurements of the timing resolution of fine-mesh Hamamatsu R7761-70 photomultipliers Wooyoung Kim, Slava Kuznetsov, Andrey Ni, and the Nuclear

More information

The hybrid photon detectors for the LHCb-RICH counters

The hybrid photon detectors for the LHCb-RICH counters 7 th International Conference on Advanced Technology and Particle Physics The hybrid photon detectors for the LHCb-RICH counters Maria Girone, CERN and Imperial College on behalf of the LHCb-RICH group

More information

The Time-of-Flight Detector for the ALICE experiment

The Time-of-Flight Detector for the ALICE experiment ALICE-PUB-- The Time-of-Flight Detector for the ALICE experiment M.C.S. Williams for the ALICE collaboration EP Division, CERN, Geneva, Switzerland Abstract The Multigap Resistive Plate Chamber (MRPC)

More information

TORCH a large-area detector for high resolution time-of-flight

TORCH a large-area detector for high resolution time-of-flight TORCH a large-area detector for high resolution time-of-flight Roger Forty (CERN) on behalf of the TORCH collaboration 1. TORCH concept 2. Application in LHCb 3. R&D project 4. Test-beam studies TIPP 2017,

More information

UVscope an instrument for calibration support

UVscope an instrument for calibration support Universidade de São Paulo UVscope an instrument for calibration support Giovanni La Rosa for the CTA ASTRI Project INAF/IASF-Palermo, Italy This work was conducted in the context of the CTA ASTRI Project

More information

Photo Multipliers Tubes characterization for WA105 experiment. Chiara Lastoria TAE Benasque 07/09/2016

Photo Multipliers Tubes characterization for WA105 experiment. Chiara Lastoria TAE Benasque 07/09/2016 Photo Multipliers Tubes characterization for WA105 experiment Chiara Lastoria TAE Benasque 07/09/2016 Outline WA105 experiment Dual Phase technology and TPC photon detection Photo Multipliers Tubes working

More information

Large photocathode 20-inch PMT testing methods for the JUNO experiment

Large photocathode 20-inch PMT testing methods for the JUNO experiment Large photocathode 20-inch PMT testing methods for the JUNO experiment N. Anfimov a on behalf of the JUNO collaboration. a Joint Institute for Nuclear Research, 141980, 6 Joliot-Curie, Dubna, Russian Federation

More information

CAEN Tools for Discovery

CAEN Tools for Discovery Viareggio March 28, 2011 Introduction: what is the SiPM? The Silicon PhotoMultiplier (SiPM) consists of a high density (up to ~10 3 /mm 2 ) matrix of diodes connected in parallel on a common Si substrate.

More information

ILC Detector Work. Dan Peterson

ILC Detector Work. Dan Peterson ILC Detector Work Dan Peterson ** Cornell/Purdue TPC development program Large Detector Concept TPC Detector Response Simulation and Track Reconstruction World Wide Study Detector R&D Panel This project

More information

Photodetector Testing Facilities at Nevis Labs & Barnard College. Reshmi Mukherjee Barnard College, Columbia University

Photodetector Testing Facilities at Nevis Labs & Barnard College. Reshmi Mukherjee Barnard College, Columbia University Photodetector Testing Facilities at Nevis Labs & Barnard College Reshmi Mukherjee Barnard College, Columbia University First AGIS Collaboration Meeting, UCLA, June 26-27, 2008 M64 MAPMT Testing for Double

More information

THE TIMING COUNTER OF THE MEG EXPERIMENT: DESIGN AND COMMISSIONING (OR HOW TO BUILD YOUR OWN HIGH TIMING RESOLUTION DETECTOR )

THE TIMING COUNTER OF THE MEG EXPERIMENT: DESIGN AND COMMISSIONING (OR HOW TO BUILD YOUR OWN HIGH TIMING RESOLUTION DETECTOR ) THE TIMING COUNTER OF THE MEG EXPERIMENT: DESIGN AND COMMISSIONING (OR HOW TO BUILD YOUR OWN HIGH TIMING RESOLUTION DETECTOR ) S. DUSSONI FRONTIER DETECTOR FOR FRONTIER PHYSICS - LA BIODOLA 2009 Fastest

More information

Operation of CEBAF photoguns at average beam current > 1 ma

Operation of CEBAF photoguns at average beam current > 1 ma Operation of CEBAF photoguns at average beam current > 1 ma M. Poelker, J. Grames, P. Adderley, J. Brittian, J. Clark, J. Hansknecht, M. Stutzman Can we improve charge lifetime by merely increasing the

More information

Commissioning and Initial Performance of the Belle II itop PID Subdetector

Commissioning and Initial Performance of the Belle II itop PID Subdetector Commissioning and Initial Performance of the Belle II itop PID Subdetector Gary Varner University of Hawaii TIPP 2017 Beijing Upgrading PID Performance - PID (π/κ) detectors - Inside current calorimeter

More information

TPC R&D by LCTPC. Organisation, results, plans. Jan Timmermans NIKHEF & DESY(2009) On behalf of the LCTPC Collaboration TILC09, Tsukuba

TPC R&D by LCTPC. Organisation, results, plans. Jan Timmermans NIKHEF & DESY(2009) On behalf of the LCTPC Collaboration TILC09, Tsukuba TPC R&D by LCTPC Organisation, results, plans Jan Timmermans NIKHEF & DESY(2009) On behalf of the LCTPC Collaboration TILC09, Tsukuba 20 April, 2009 LCTPC Collaboration IIHE ULB-VUB Brussels 2 LCTPC Collaboration

More information

The 20 inch MCP-PMT R&D in China

The 20 inch MCP-PMT R&D in China The 20 inch MCP-PMT R&D in China Sen Qian,On Behalf of the Workgroup Institute of High energy Physics, Chinese Academy of Science qians@ihep.ac.cn Oct. 25. 2016 Outline 1. The JUNO and MCP-PMT; 2. The

More information

Front End Electronics

Front End Electronics CLAS12 Ring Imaging Cherenkov (RICH) Detector Mid-term Review Front End Electronics INFN - Ferrara Matteo Turisini 2015 October 13 th Overview Readout requirements Hardware design Electronics boards Integration

More information

Drift Tubes as Muon Detectors for ILC

Drift Tubes as Muon Detectors for ILC Drift Tubes as Muon Detectors for ILC Dmitri Denisov Fermilab Major specifications for muon detectors D0 muon system tracking detectors Advantages and disadvantages of drift chambers as muon detectors

More information

Review of High Quantum Efficiency Large Area Photomultiplier Tubes

Review of High Quantum Efficiency Large Area Photomultiplier Tubes Jinping Solar Neutrino Workshop, LBNL, June 2014 Review of High Quantum Efficiency Large Area Photomultiplier Tubes Jianglai Liu Shanghai Jiao Tong University 2014/6/10 Disclaimer: I am not personally

More information

Hamamatsu R1584 PMT Modifications

Hamamatsu R1584 PMT Modifications Hamamatsu R1584 PMT Modifications Wenliang Li, Garth Huber, Keith Wolbaum University of Regina, Regina, SK, S4S-0A2 Canada October 31, 2013 Abstract Four Hamamatsu H6528 Photomultiplier Tube (PMT) assemblies

More information

Scintillation Tile Hodoscope for the PANDA Barrel Time-Of-Flight Detector

Scintillation Tile Hodoscope for the PANDA Barrel Time-Of-Flight Detector Scintillation Tile Hodoscope for the PANDA Barrel Time-Of-Flight Detector William Nalti, Ken Suzuki, Stefan-Meyer-Institut, ÖAW on behalf of the PANDA/Barrel-TOF(SciTil) group 12.06.2018, ICASiPM2018 1

More information

Front End Electronics

Front End Electronics CLAS12 Ring Imaging Cherenkov (RICH) Detector Mid-term Review Front End Electronics INFN - Ferrara Matteo Turisini 2015 October 13 th Overview Readout requirements Hardware design Electronics boards Integration

More information

Review of photo-sensor R&D for future water Cherenkov detectors NNN10 Dec

Review of photo-sensor R&D for future water Cherenkov detectors NNN10 Dec Review of photo-sensor R&D for future water Cherenkov detectors NNN10 Dec 15 2010 Hiroyuki Sekiya ICRR, University of Tokyo Special Thanks T. Abe F. Tokanai, & T. Sumiyoshi Hamamatsu Photonics 1 Contents/Disclaimer

More information

TPC R&D at Cornell and Purdue

TPC R&D at Cornell and Purdue TPC R&D at Cornell and Purdue Cornell University Purdue University T. Anous K. Arndt R. S. Galik G. Bolla D. P. Peterson I. P. J. Shipsey J. Ledoux Further information available at the web sites: http://www.lepp.cornell.edu/~dpp/linear_collider/large_prototype.html

More information

A new Scintillating Fibre Tracker for LHCb experiment

A new Scintillating Fibre Tracker for LHCb experiment A new Scintillating Fibre Tracker for LHCb experiment Alexander Malinin, NRC Kurchatov Institute on behalf of the LHCb-SciFi-Collaboration Instrumentation for Colliding Beam Physics BINP, Novosibirsk,

More information

SPE analysis of high efficiency PMTs for the DEAP-3600 dark matter detector

SPE analysis of high efficiency PMTs for the DEAP-3600 dark matter detector Journal of Physics: Conference Series SPE analysis of high efficiency PMTs for the DEAP-36 dark matter detector To cite this article: Kevin Olsen et al 211 J. Phys.: Conf. Ser. 312 7215 View the article

More information

Liquid Xenon Scintillation Detector with UV-SiPM Readout for MEG Upgrade

Liquid Xenon Scintillation Detector with UV-SiPM Readout for MEG Upgrade Liquid Xenon Scintillation Detector with UV-SiPM Readout for MEG Upgrade W. Ootani on behalf of MEG collaboration (ICEPP, Univ. of Tokyo) 13th Topical Seminar on Innovative Particle and Radiation Detectors

More information

Studies of large dynamic range silicon photomultipliers for the CMS HCAL upgrade

Studies of large dynamic range silicon photomultipliers for the CMS HCAL upgrade Studies of large dynamic range silicon photomultipliers for the CMS HCAL upgrade Yuri Musienko* FNAL(USA) Arjan Heering University of Notre Dame (USA) For the CMS HCAL group *On leave from INR(Moscow)

More information

SciFi A Large Scintillating Fibre Tracker for LHCb

SciFi A Large Scintillating Fibre Tracker for LHCb SciFi A Large Scintillating Fibre Tracker for LHCb Roman Greim on behalf of the LHCb-SciFi-Collaboration 14th Topical Seminar on Innovative Particle Radiation Detectors, Siena 5th October 2016 I. Physikalisches

More information

Lifetime of MCP-PMTs

Lifetime of MCP-PMTs Lifetime of MCP-PMTs, Merlin Böhm, Alexander Britting, Wolfgang Eyrich, Markus Pfaffinger, Fred Uhlig (Universität Erlangen-Nürnberg) Motivation Approaches to increase lifetime Results of aging tests Outlook

More information

A prototype of fine granularity lead-scintillating fiber calorimeter with imaging read-out

A prototype of fine granularity lead-scintillating fiber calorimeter with imaging read-out A prototype of fine granularity lead-scintillating fiber calorimeter with imaging read-out P.Branchini, F.Ceradini, B.Di Micco, A. Passeri INFN Roma Tre and Dipartimento di Fisica Università Roma Tre and

More information

Psec-Resolution Time-of-Flight Detectors T979

Psec-Resolution Time-of-Flight Detectors T979 1 Psec-Resolution Time-of-Flight Detectors T979 Argonne, Chicago, Fermilab, Hawaii, Saclay/IRFU, SLAC Camden Ertley University of Chicago All Experimenters Meeting July 14, 2008 (Bastille Day!) T979 People/Institutions

More information

Spectroscopy on Thick HgI 2 Detectors: A Comparison Between Planar and Pixelated Electrodes

Spectroscopy on Thick HgI 2 Detectors: A Comparison Between Planar and Pixelated Electrodes 1220 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, OL. 50, NO. 4, AUGUST 2003 Spectroscopy on Thick HgI 2 Detectors: A Comparison Between Planar and Pixelated Electrodes James E. Baciak, Student Member, IEEE,

More information

Beam test of the QMB6 calibration board and HBU0 prototype

Beam test of the QMB6 calibration board and HBU0 prototype Beam test of the QMB6 calibration board and HBU0 prototype J. Cvach 1, J. Kvasnička 1,2, I. Polák 1, J. Zálešák 1 May 23, 2011 Abstract We report about the performance of the HBU0 board and the optical

More information

Concept and operation of the high resolution gaseous micro-pixel detector Gossip

Concept and operation of the high resolution gaseous micro-pixel detector Gossip Concept and operation of the high resolution gaseous micro-pixel detector Gossip Yevgen Bilevych 1,Victor Blanco Carballo 1, Maarten van Dijk 1, Martin Fransen 1, Harry van der Graaf 1, Fred Hartjes 1,

More information

arxiv: v1 [physics.ins-det] 1 Nov 2015

arxiv: v1 [physics.ins-det] 1 Nov 2015 DPF2015-288 November 3, 2015 The CMS Beam Halo Monitor Detector System arxiv:1511.00264v1 [physics.ins-det] 1 Nov 2015 Kelly Stifter On behalf of the CMS collaboration University of Minnesota, Minneapolis,

More information

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes: PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties Due Thursday, 2 Nov 2017 For this lab, you will explore the properties of the working HeNe laser. 1. Observation of higher-order modes: Realign

More information

Single Photoelectron timing resolution of SiPM

Single Photoelectron timing resolution of SiPM Research & Study Detector Group Single Photoelectron timing resolution of SiPM XVII SuperB Workshop - Kick Off meeting May 29 th - June 1 st 2011 Isola d Elba Véronique Puill, IN2P3-LAL -GRED C. Bazin,

More information

High ResolutionCross Strip Anodes for Photon Counting detectors

High ResolutionCross Strip Anodes for Photon Counting detectors High ResolutionCross Strip Anodes for Photon Counting detectors Oswald H.W. Siegmund, Anton S. Tremsin, Robert Abiad, J. Hull and John V. Vallerga Space Sciences Laboratory University of California Berkeley,

More information

ELECTRON OPTICS OF ST-X, ST-Y SERIES OF STREAK & FRAMING CAMERA TUBES

ELECTRON OPTICS OF ST-X, ST-Y SERIES OF STREAK & FRAMING CAMERA TUBES ELECTRON OPTICS OF ST-X, ST-Y SERIES OF STREAK & FRAMING CAMERA TUBES INTRODUCTION The basic electron optics of this range of streak tubes were designed by Ching Lai at the Lawrence Livermore National

More information

Reading a GEM with a VLSI pixel ASIC used as a direct charge collecting anode. R.Bellazzini - INFN Pisa. Vienna February

Reading a GEM with a VLSI pixel ASIC used as a direct charge collecting anode. R.Bellazzini - INFN Pisa. Vienna February Reading a GEM with a VLSI pixel ASIC used as a direct charge collecting anode Ronaldo Bellazzini INFN Pisa Vienna February 16-21 2004 The GEM amplifier The most interesting feature of the Gas Electron

More information

First evaluation of the prototype 19-modules camera for the Large Size Telescope of the CTA

First evaluation of the prototype 19-modules camera for the Large Size Telescope of the CTA First evaluation of the prototype 19-modules camera for the Large Size Telescope of the CTA Tsutomu Nagayoshi for the CTA-Japan Consortium Saitama Univ, Max-Planck-Institute for Physics 1 Cherenkov Telescope

More information

An extreme high resolution Timing Counter for the MEG Upgrade

An extreme high resolution Timing Counter for the MEG Upgrade An extreme high resolution Timing Counter for the MEG Upgrade M. De Gerone INFN Genova on behalf of the MEG collaboration 13th Topical Seminar on Innovative Particle and Radiation Detectors Siena, Oct.

More information

Large Area, High Speed Photo-detectors Readout

Large Area, High Speed Photo-detectors Readout Large Area, High Speed Photo-detectors Readout Jean-Francois Genat + On behalf and with the help of Herve Grabas +, Samuel Meehan +, Eric Oberla +, Fukun Tang +, Gary Varner ++, and Henry Frisch + + University

More information

A fast and precise COME & KISS* QDC and TDC for diamond detectors and further applications

A fast and precise COME & KISS* QDC and TDC for diamond detectors and further applications A fast and precise COME & KISS* QDC and TDC for diamond detectors and further applications 3 rd ADAMAS Collaboration Meeting (2014) Trento, Italy *use commercial elements and keep it small & simple + +

More information

Very High QE bialkali PMTs

Very High QE bialkali PMTs Very High QE bialkali PMTs Mª Victoria Fonseca University Complutense, Madrid, Spain How a classical PMT is operating photons Quantum Efficiency Quantum efficiency (QE) of a sensor QE = N(ph.e.) / N(photons)

More information

Time Resolution Improvement of an Electromagnetic Calorimeter Based on Lead Tungstate Crystals

Time Resolution Improvement of an Electromagnetic Calorimeter Based on Lead Tungstate Crystals Time Resolution Improvement of an Electromagnetic Calorimeter Based on Lead Tungstate Crystals M. Ippolitov 1 NRC Kurchatov Institute and NRNU MEPhI Kurchatov sq.1, 123182, Moscow, Russian Federation E-mail:

More information

Performance and Radioactivity Measurements of the PMTs for the LUX and LZ Dark Matter Experiments

Performance and Radioactivity Measurements of the PMTs for the LUX and LZ Dark Matter Experiments Performance and Radioactivity Measurements of the PMTs for the LUX and LZ Dark Matter Experiments Carlos Hernandez Faham Brown University Carlos Faham Brown University Particle Astrophysics Group, June

More information

The Scintillating Fibre Tracker for the LHCb Upgrade. DESY Joint Instrumentation Seminar

The Scintillating Fibre Tracker for the LHCb Upgrade. DESY Joint Instrumentation Seminar The Scintillating Fibre Tracker for the LHCb Upgrade DESY Joint Instrumentation Seminar Presented by Blake D. Leverington University of Heidelberg, DE on behalf of the LHCb SciFi Tracker group 1/45 Outline

More information

Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720

Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720 O.H.W. Siegmund* a, J.B. McPhate a, A.S. Tremsin a, S.R. Jelinsky a, J.V. Vallerga a, R. Hemphill a, H.J. Frisch b, J. Elam c, A. Mane c, and the LAPPD Collaboration c a Experimental Astrophysics Group,

More information

STUDIES OF ENHANCED EDGE EMISSION OF A LARGE AREA CATHODE *

STUDIES OF ENHANCED EDGE EMISSION OF A LARGE AREA CATHODE * STUDIES OF ENHANCED EDGE EMISSION OF A LARGE AREA CATHODE * F. Hegeler, M. Friedman, M.C. Myers, S.B. Swanekamp, and J.D. Sethian Plasma Physics Division, Code 6730 Naval Research Laboratory, Washington,

More information

PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND. Doug Roberts U of Maryland, College Park

PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND. Doug Roberts U of Maryland, College Park PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND Doug Roberts U of Maryland, College Park Overview We have developed a system for measuring and scanning phototubes for the FDIRC Based primarily on

More information

Progress Update FDC Prototype Test Stand Development Upcoming Work

Progress Update FDC Prototype Test Stand Development Upcoming Work Progress Update FDC Prototype Test Stand Development Upcoming Work Progress Update OU GlueX postdoc position filled. Simon Taylor joins our group July 1, 2004 Position funded jointly by Ohio University

More information

Silicon PhotoMultiplier Kits

Silicon PhotoMultiplier Kits Silicon PhotoMultiplier Kits Silicon PhotoMultipliers (SiPM) consist of a high density (up to ~ 10 3 /mm 2 ) matrix of photodiodes with a common output. Each diode is operated in a limited Geiger- Müller

More information

Prospect and Plan for IRS3B Readout

Prospect and Plan for IRS3B Readout Prospect and Plan for IRS3B Readout 1. Progress on Key Performance Parameters 2. Understanding limitations during LEPS operation 3. Carrier02 Rev. C (with O-E-M improvements) 4. Pre-production tasks/schedule

More information

The Cornell/Purdue TPC

The Cornell/Purdue TPC The Cornell/Purdue TPC Cornell University Purdue University D. P. Peterson G. Bolla L. Fields I. P. J. Shipsey R. S. Galik P. Onyisi Information available at the web site: http://w4.lns.cornell.edu/~dpp/tpc_test_lab_info.html

More information

PMT Gain & Resolution Measurements in High Magnetic Fields

PMT Gain & Resolution Measurements in High Magnetic Fields PMT Gain & Resolution Measurements in High Magnetic Fields Vincent Sulkosky University of Virginia August 11 th, 2015 SoLID EC Meeting High-B Sensor-Testing Facility 2 The facility was designed for the

More information

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES Hale R. Farley, Jeffrey L. Guttman, Razvan Chirita and Carmen D. Pâlsan Photon inc. 6860 Santa Teresa Blvd

More information

A very brief review of recent SiPM developments

A very brief review of recent SiPM developments A very brief review of recent SiPM developments, Distefano Garcia School of Physics & Center for Relativistic Astrophysics, Georgia Institute of Technology, 837 State Street NW, Atlanta, GA 30332-0430,

More information

New gas detectors for the PRISMA spectrometer focal plane

New gas detectors for the PRISMA spectrometer focal plane M. Labiche - STFC Daresbury Laboratory New gas detectors for the PRISMA spectrometer focal plane New PPAC (Legnaro Padova Bucharest Zagreb) & Large Secondary e - Detector (Se - D) (Manchester-Daresbury-Paisley-

More information

3-D position sensitive CdZnTe gamma-ray spectrometers

3-D position sensitive CdZnTe gamma-ray spectrometers Nuclear Instruments and Methods in Physics Research A 422 (1999) 173 178 3-D position sensitive CdZnTe gamma-ray spectrometers Z. He *, W.Li, G.F. Knoll, D.K. Wehe, J. Berry, C.M. Stahle Department of

More information

Cathode Studies at FLASH: CW and Pulsed QE measurements

Cathode Studies at FLASH: CW and Pulsed QE measurements Cathode Studies at FLASH: CW and Pulsed QE measurements L. Monaco, D. Sertore, P. Michelato S. Lederer, S. Schreiber Work supported by the European Community (contract number RII3-CT-2004-506008) 1/27

More information

PoS(PhotoDet 2012)018

PoS(PhotoDet 2012)018 Development of a scintillation counter with MPPC readout for the internal tagging system Hiroki KANDA, Yuma KASAI, Kazushige MAEDA, Takashi NISHIZAWA, and Fumiya YAMAMOTO Department of Physics, Tohoku

More information

with Low Cost and Low Material Budget

with Low Cost and Low Material Budget Gaseous Beam Position Detectors, with Low Cost and Low Material Budget Gyula Bencédi on behalf of the REGaRD group MTA KFKI RMKI, ELTE November 29, 2011, Outline Physics Motivation Newish MWPCs, the Close

More information

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator Stanley Humphries, Copyright 2012 Field Precision PO Box 13595, Albuquerque, NM 87192 U.S.A. Telephone: +1-505-220-3975

More information

A TARGET-based camera for CTA

A TARGET-based camera for CTA A TARGET-based camera for CTA TeV Array Readout with GSa/s sampling and Event Trigger (TARGET) chip: overview Custom-designed ASIC for CTA, developed in collaboration with Gary Varner (U Hawaii) Implementation:

More information

Imaging TOP (itop), Cosmic Ray Test Stand & PID Readout Update

Imaging TOP (itop), Cosmic Ray Test Stand & PID Readout Update Imaging TOP (itop), Cosmic Ray Test Stand & PID Readout Update Tom Browder, Herbert Hoedlmoser, Bryce Jacobsen, Jim Kennedy, KurtisNishimura, Marc Rosen, Larry Ruckman, Gary Varner Kurtis Nishimura SuperKEKB

More information

A Cylindrical GEM Detector with Analog Readout for the BESIII Experiment. Gianluigi Cibinetto (INFN Ferrara) on behalf of the BESIIICGEM consortium

A Cylindrical GEM Detector with Analog Readout for the BESIII Experiment. Gianluigi Cibinetto (INFN Ferrara) on behalf of the BESIIICGEM consortium A Cylindrical GEM Detector with Analog Readout for the BESIII Experiment Gianluigi Cibinetto (INFN Ferrara) on behalf of the BESIIICGEM consortium Outline The BESIII experiment the Inner tracker The BESIII

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

A Review of Tracking Sessions

A Review of Tracking Sessions A Review of Tracking Sessions Madhu S. Dixit TRIUMF & Carleton University Durham ECFA Workshop 1-4 September 2004 8 minutes time for this summary (allow 2 minutes for questions) 3 tracking sessions lasting

More information

This work was supported by FINEP (Research and Projects Financing) under contract

This work was supported by FINEP (Research and Projects Financing) under contract MODELING OF A GRIDDED ELECTRON GUN FOR TRAVELING WAVE TUBES C. C. Xavier and C. C. Motta Nuclear & Energetic Research Institute, São Paulo, SP, Brazil University of São Paulo, São Paulo, SP, Brazil Abstract

More information

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling Commissioning the TAMUTRAP RFQ cooler/buncher E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling In order to efficiently load ions into a Penning trap, the ion beam should be

More information

The field cage for a large TPC prototype

The field cage for a large TPC prototype EUDET The field cage for a large TPC prototype T.Behnke, L. Hallermann, P. Schade, R. Diener December 7, 2006 Abstract Within the EUDET Programme, the FLC TPC Group at DESY in collaboration with the Department

More information

Quick Report on Silicon G-APDs (a.k.a. Si-PM) studies. XIV SuperB General Meeting LNF - Frascati

Quick Report on Silicon G-APDs (a.k.a. Si-PM) studies. XIV SuperB General Meeting LNF - Frascati Quick Report on Silicon G-APDs (a.k.a. Si-PM) studies XIV SuperB General Meeting LNF - Frascati Report of the work done in Padova Dal Corso F., E.F., Simi G., Stroili R. University & INFN Padova Outline

More information

A flexible FPGA based QDC and TDC for the HADES and the CBM calorimeters TWEPP 2016, Karlsruhe HADES CBM

A flexible FPGA based QDC and TDC for the HADES and the CBM calorimeters TWEPP 2016, Karlsruhe HADES CBM A flexible FPGA based QDC and TDC for the HADES and the CBM calorimeters TWEPP 2016, Karlsruhe + + + = PaDiWa-AMPS front-end Adrian Rost for the HADES and CBM collaborations PMT Si-PM (MPPC) 27.09.2016

More information

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera PRELIMINARY POWER DETECTORS ENERGY DETECTORS MONITORS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS CMOS Beam Profiling Camera AVAILABLE MODEL Beamage 3.0 (⅔ in CMOS

More information

Duke University. Plasma Display Panel. A vanished technique

Duke University. Plasma Display Panel. A vanished technique Duke University Plasma Display Panel A vanished technique Yida Chen Dr. Hubert Bray Math 190s: Mathematics of the Universe 31 July 2017 Introduction With the establishment of the atomic theory, we begin

More information

CBF500 High resolution Streak camera

CBF500 High resolution Streak camera High resolution Streak camera Features 400 900 nm spectral sensitivity 5 ps impulse response 10 ps trigger jitter Trigger external or command 5 to 50 ns analysis duration 1024 x 1024, 12-bit readout camera

More information

Leica TCS CARS. Live Molecular Profiling Technical Documentation. Living up to Life

Leica TCS CARS. Live Molecular Profiling Technical Documentation. Living up to Life Leica TCS CARS Live Molecular Profiling Technical Documentation Living up to Life Microscopes Inverted Leica DMI6000 CS Microscope anti-vibration table Specification Vibration insulation Passive Z-drive

More information

OPTICAL POWER METER WITH SMART DETECTOR HEAD

OPTICAL POWER METER WITH SMART DETECTOR HEAD OPTICAL POWER METER WITH SMART DETECTOR HEAD Features Fast response (over 1000 readouts/s) Wavelengths: 440 to 900 nm for visible (VIS) and 800 to 1700 nm for infrared (IR) NIST traceable Built-in attenuator

More information

arxiv: v2 [physics.ins-det] 24 Mar 2015

arxiv: v2 [physics.ins-det] 24 Mar 2015 Investigation of Hamamatsu H8500 phototubes as single photon detectors. R.A. Montgomery a, M. Hoek b, V. Lucherini a, M. Mirazita a, A. Orlandi a, S. Anefalos Pereira a, S. Pisano a, P. Rossi a,c, A. Viticchiè

More information

Characterizing Transverse Beam Dynamics at the APS Storage Ring Using a Dual-Sweep Streak Camera

Characterizing Transverse Beam Dynamics at the APS Storage Ring Using a Dual-Sweep Streak Camera Characterizing Transverse Beam Dynamics at the APS Storage Ring Using a Dual-Sweep Streak Camera Bingxin Yang, Alex H. Lumpkin, Katherine Harkay, Louis Emery, Michael Borland, and Frank Lenkszus Advanced

More information

Sensors for precision timing HEP

Sensors for precision timing HEP Sensors for precision timing HEP Adi Bornheim For the Caltech Precision Timing group 2/10/2016 Adi Bornheim, Meeting with Hamamatsu 1 Introduction & Overview We develop detectors for high energy physics

More information

DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC*

DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC* DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC* D. Sprehn, E. Jongewaard, A. Haase, A. Jensen, D. Martin, SLAC National Accelerator Laboratory, Menlo Park, CA 94020, U.S.A. A. Burke, SAIC, San

More information

Development of an Abort Gap Monitor for High-Energy Proton Rings *

Development of an Abort Gap Monitor for High-Energy Proton Rings * Development of an Abort Gap Monitor for High-Energy Proton Rings * J.-F. Beche, J. Byrd, S. De Santis, P. Denes, M. Placidi, W. Turner, M. Zolotorev Lawrence Berkeley National Laboratory, Berkeley, USA

More information

Thyratrons. High Energy Switches. Features. Description

Thyratrons. High Energy Switches. Features. Description Thyratrons Lighting Imaging Telecom High Energy Switches D A T A S H E E T Description Thyratrons are fast acting high voltage switches suitable for a variety of applications including radar, laser and

More information