More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

Size: px
Start display at page:

Download "More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays"

Transcription

1 Digital Industrial Radiology and Computed Tomography (DIR 2015) June 2015, Belgium, Ghent - More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays David Allen FRY 1, Christopher Jason STULL 1, Brandon Michael LATTIMORE 1 1 Nondestructive Testing & Evaluation Group, Los Alamos National Laboratory; Los Alamos, New Mexico USA Phone: , dafry@lanl.gov Abstract Los Alamos National Laboratory has recently completed the qualification of a computed tomography process for the evaluation of small parts. As part of the qualification, ASTM International Digital Radiography (DR) and Computed Tomography (CT) standards were used to develop a process control program. ASTM E2737 and E2698 were used to create a series of baseline tests for the individual DR images that make up a CT data set. Spatial resolution, contrast sensitivity, detector offset, and lag are performed initially and then repeated on a weekly basis to ensure the digital detector array is operating within tolerance. ASTM E1695 spatial resolution and contrast sensitivity tests are used to assess CT baseline and stability. The complexity of these tests, in addition to the frequency at which they are to be executed, necessitated the development of software to automatically analyze the process control data. To this end, Graphical User Interfaces (GUIs) were written using the scripting language, Python. It is the intent of the authors to make this software available to the ASTM community at a later date. Keywords: Computed tomography, digital radiography, process control 1. Introduction Los Alamos National Laboratory (LANL) has been doing Research & Development on digital radiography (DR) and computed tomography (CT) for many years. Recently, DR/CT processes have needed to be qualified for production use. All system variables must be addressed, established, and documented. Part of the qualification is performing a system baseline characterization. Periodic stability tests comparing results to the baseline are then performed to ensure the system is operating within acceptable limits. LANL produces components for the United States government s National Nuclear Security Agency (NNSA). In the NNSA environment, the high level requirements document for DR/CT is RMI T097 Qualification of Digital Radiographic Imaging Techniques [1]. The high level requirements have been incorporated into a LANL Design Agency specification SS6K0461 General Specifications for Digital Radiography of Los Alamos National Laboratory-Designed Parts and Assemblies [2]. Both of these documents rely on implementation of ASTM International standards E1695 [3] and E2737 [4] to satisfy requirements. Table 1 compares the requirements and standards. We aim to make these tests as easy and quick as possible. Currently we have agreed with the Design Agency to perform weekly performance checks. For our 10X magnification CT system based on DDA technology, we perform both DDA and CT checks. We have developed automated and semi-automated methods for analysis of the performance test images.

2 2. Terminology CNR Contrast to Noise Ratio CR Computed Radiography CSa achievable Contrast Sensitivity (see ASTM E2597 [5]) DDA Digital Detector Array DDL Digital Driving Level ERF Edge Response Function LSF Line Spread Function MTF Modulation Transfer Function PV Pixel Value SRb Basic Spatial Resolution (see ASTM E2597) isrb - Interpolated Basic Spatial Resolution (see ASTM E2597) SNR Signal to Noise Ratio Table 1 Requirement T097 SS6K0461 ASTM/DDA Standard Phantom Spatial X X E2737 E2002 Gauge Resolution/ or line pair MTF/SRb gauge Contrast Sensitivity/ CNR Signal-to-Noise (SNR) X X E2737 Penetrameter/ Shim X X E2737 Penetrameter/ Shim ASTM/CT Standard E1695 E1695 Phantom Uniform Disk Uniform Disk Measure isee SRb Python GUI CT MTF Python GUI CNR Python GUI CT CDF Python GUI or isee SNR Lag X X E2737 None Python GUI Lag Dynamic Range X X E2737 Penetrameters/ Python GUI CNR Shims Dimensional Uniformity X X E2445 CR Phantom isee Distance measure Detector Degradation X X E2737 None Python GUI or isee Display Brightness X X E2737 SMPTE RP133 Calibrated light meter Display Contrast X X E2737 SMPTE RP133 Visual 0% DDL against 5% DDL and 95% DDL against 100% DDL Display Spatial Resolution X X E2737 SMPTE RP133 Visual alternating 1% & and 100% contrast lines

3 3. DDA Checks ASTM International, E2698 Standard Practice for Radiological Examination Using Digital Detector Arrays [6] calls for tests for monitoring the DDA performance over time shall be performed in accordance with ASTM International, E2737 Standard Practice for Digital Detector Array Performance Evaluation and Long-Term Stability. 3.1 Spatial Resolution An image of the ASTM E2002 Gauge [7] with both vertical and horizontal orientations is captured with the gauge directly on the DDA face. ISee! Professional software [8] contains a modulation calculator based on an averaged line profile through the E2002 gauge see Figures 1 and 2. The line pair gauge is captured and analyzed for the largest pair with < 20% modulation (SRb) or interpolated to find the exact 20% modulation frequency (isrb). Figure 1: ISee! averaged line profile through ASTM E2002 duplex wire gauge (horizontal profile shown)

4 Figure 2: ISee! measurement of modulation of a wire pair. Three measurement points are user selected (white dashed lines) and the modulation reads in the upper right corner (arrow). 3.2 Contrast Sensitivity and Dynamic Range Images of appropriate material and thickness hole-type penetrameters on a shims is captured with the same parameters as production imaging for the thinnest and thickest sections to be imaged. Our Python CNR routine is used to analyze the images. The user selects the Figure 3: CNR determined from signal and noise inside and outside a penetrameter hole penetrameter hole for which CNR is desired and the routine automatically calculates CNR based on the region between two regions of interest around the hole with sides 2X and 4X the hole diameter and the hole see Figure 3. From CNR, achievable Contrast Sensitivity (CSa) can be calculated. CNR and/or CSa is calculated for both the thinnest and thickest regions of a given part, or an acceptable range of part thickness is determined with minimum CNR/CSa.

5 3.3 Signal-to-Noise Ratio (SNR) A graph of SNR vs. Exposure is generated from flat field images, see Fig. 4. A quick check of just one point, such as 90% of saturation, on the line is sufficient for the periodic tests. The slope of the line also gives the ASTM E2737 Efficiency of the detector. SNR vs Exposure 1E33 Technique SNR = *sqrt(E) SNR SNR = *sqrt(E) SQRT(mGy/frame) 1 frame 8 frame Linear (1 frame) Linear (8 frame) Figure 4: SNR vs. square root of exposure 3.4 Lag A graph of signal vs. time is generated from a series of images before and after the XGD is shut down, see Fig. 5. The user selects a region of interest and the software builds a table of average signal vs. time for a sequence of images taken with the CT acquisition software with no object.

6 % of 1st frame PV Shutdown 1E33 Technique (3.0 fps) Time (seconds) 1E33 Shutdown Shutdown 1E33 Technique (3.0 fps) % of 1st frame y = x Shutdown Power (Shutdown) Frames after Transition Figure 5: Lag measurement 3.5 Dimensional Uniformity An ASTM E2445 [9] or similar phantom, see Fig. 6, is measured to check dimensional features are uniform. Capturing this phantom image obtains the E2002 gauges for the spatial resolution check.

7 3.6 Detector Degradation Figure 6: LANL CR Phantom used for dimensional uniformity An Offset image is analyzed for signal level and SNR, and visually for any pattern. An increase in average offset PV or decrease in SNR signals degradation. 4. CT Checks ASTM E1570 Standard Practice for Computed Tomographic (CT) Examination [10] calls for initial and periodic system performance measurement. These requirements can be satisfied using the methods of ASTM 1695 Standard Test Method for Measurement of Computed Tomography (CT) System Performance.

8 4.2 Contrast Resolution An appropriate material and diameter uniform disk is scanned with the same parameters as production imaging. For an individual reconstruction slice, extracted from this data set, the user selects a circular region within the disk of material large enough to supply a statistically significant sample of pixels, but small enough to avoid a common reconstruction artifact referred to as cupping. Cupping is an artifact arising primarily from beam hardening (but also from internal scatter), resulting in reduced attenuation values near the centers of uniform cross-sections. These reduce attenuation values imply that the density in the interior region is less than that of the exterior region which is, of course, not the case for uniform cross-sections. Figure 7: Example tile selection from reconstruction slice of uniform disk. After the circular region is selected, the software draws a square, inscribed within the circle is drawn (see Figure 7). That square is then progressively subdivided into square tiles, until the tile size (i.e. edge length of the tile, in pixels) is equal to one pixel. The image below presents an example of this operation for an edge length of the initial square equal to 256 pixels.

9 Figure 8: Example tiling scheme Having performed this operation, the following mathematical operations are conducted for each set of tiles of a given size (e.g. for all 64 pixel by 64 pixel tiles, then for all 32 pixel by 32 pixel tiles, and so on). Example output from the Python-based GUI is given in Figure 9. 1) Calculate the mean attenuation of each of the tiles. 2) Calculate the standard deviation of the values computed in 1), in order to obtain the standard error in the mean note that this operation yields on values for each set of tiles. 3) Express the standard error in the mean as a percentage of the mean of the values computed in 1) and multiply by 3, which corresponds to a 50% false-negative rate (i.e. the case of threshold detectability).

10 4.1 Spatial Resolution Figure 9: Contrast resolution plot from Python-based GUI As for the contrast resolution system test, the spatial resolution system test begins with a reconstruction slice of a disk of uniform material. This system test, however, is concerned with the edge response function, and so, the user draws two circles that bracket the edge of the reconstruction slice of the disk; this is illustrated in the figure below. The pixels that fall between those two circles are then sorted and binned, according to their distances from the center of mass of the disk. These binned pixels are then averaged to obtain a single value for each bin. The Edge Response Function (ERF) is then computed as a piecewise cubic polynomial fitted to the averages computed previously using a set (determined by the number of points used to fit the cubic polynomial) of consecutive bins; performing this fit for multiple sets of consecutive bins yields the piecewise component. It is clear that the calculation (or placement) of the center of mass of the disk is important, as asymmetries can cause substantially erroneous results. Therefore, despite its apparent trivial nature, it is worth pointing out specifically that care must be taken in calculating (or placing) the center of mass.

11 Note the asymmetry of the circular region due to misplacement of the center of mass. Figure 10: Center of mass calculation (or placement) error. Having computed the ERF, the Line Spread Function (LSF) and Modulation Transfer Function (MTF) are reasonably straightforward to obtain. The procedure to compute the LSF begins by fitting a piecewise cubic polynomial to the ERF, essentially smoothing the ERF. For each piecewise fit, the derivative is evaluated at the center of the piecewise window. The ensemble of these derivative evaluations is then normalized by the maximum to arrive at the LSF. Finally, the MTF is calculated simply as the amplitude of the Fourier Transform, normalized to unity at the zero frequency. Example output from the Python-based GUI is given in Figure 11.

12 Figure 11: Spatial resolution plots from Python-based GUI. 5. Display Checks All display checks are based on a display of the SMPTE RP133 test pattern (Fig. 9) [11] per the requirements of ASTM E2698: The image display shall meet the following requirements as a minimum. Alternate image displays or requirements may be used with Cognizant Engineering Oraganization approval. The minimum brightness as measured off the image display screen at maximum Digital Driving Level (DDL) shall be 250 cd/m 2. The minimum contrast as determined by the ratio of the screen brightness at the maximum DDL compared to the screen brightness at the minimum DDL shall be 250:1. The image display shall be capable of displaying linear patterns of alternating pixels at full contrast in both the horizontal and vertical directions without aliasing. The display shall be free of discernable geometric distortion. The display shall be free of screen flicker, characterized by high frequency fluctuation of high contrast image details.

13 The image display shall be capable of displaying a 5 % DDL block against a 0 % DDL background and simultaneously displaying a 95 % DDL block against a 100 %background in a manner clearly perceptible to the user. The monitor shall be capable of discriminating the horizontal and vertical low contrast (1 %) modulation patterns at the display center and each of the four corner locations. A calibrated brightness meter measures the 100% and 0% DDL blocks on the test pattern to get the maximum brightness and contrast ratio. 6. Future Work Figure 12: SMPTE RP133 Test Pattern [12] We intend to continue the automation process in order to increase the speed to the test process including generation of an automated report. Grouping the tests under one GUI will also help the user complete the tests. Computed radiography (CR) tests can also be added to the suite (see ASTM E2445). Many of the same or similar tests are shared between DDA and CR.

14 Acknowledgements This article is a work performed under the auspices of the United States Department of Energy. Los Alamos National Laboratory is operated by Los Alamos National Security, LLC for the National Nuclear Security Administration of the U.S. Department of Energy under contract DE-AC52-06NA Approved for public release (LA-UR ); distribution is unlimited. References 1. National Nuclear Security Agency, T097 Qualification of Digital Radiographic Imaging Techniques 2. Los Alamos National Laboratory, SS6K0461 General Specifications for Digital Radiography of Los Alamos National Laboratory Designed Parts and Assemblies 3. ASTM International, E1695 Standard Test Method for Measurement of Computed Tomography (CT) System Performance 4. ASTM International, E2737 Standard Practice for Digital Detector Array Performance Evaluation and Long-Term Stability 5. ASTM International, E2597 Standard Practice for Manufacturing Characterization of Digital Detector Arrays 6. ASTM International, E2698 Standard Practice for Radiological Examination Using Digital Detector Arrays 7. ASTM International, E2002 Standard Practice for Determining Total Image Unsharpness in Radiology 8. Vision in X Industrial Imaging, isee! Professional User Manual 9. ASTM International, E2445 Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems 10. ASTM International, E1570 Standard Practice for Computed Tomographic (CT) Examination 11. The Society of Motion Picture and Television Engineers, RP133 Specifications for Medical Diagnostic Imaging Test Pattern for Television Monitors and Hard-Copy Recording Cameras 12. Rich Franzen's PNG Gallery,

Performance Evaluation of Industrial Computed Radiography Image Display System

Performance Evaluation of Industrial Computed Radiography Image Display System Performance Evaluation of Industrial Computed Radiography Image Display System More info about this article: http://www.ndt.net/?id=21169 Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and

More information

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks

More information

Nuclear Associates and

Nuclear Associates and Nuclear Associates 76-907 and 76-908 AAPM MRI Phantoms Users Manual March 2005 Manual No. 38616 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK m RSC CHROMATOGRAPHY MONOGRAPHS Chromatographie Integration Methods Second Edition Norman Dyson Dyson Instruments Ltd., UK THE ROYAL SOCIETY OF CHEMISTRY Chapter 1 Measurements and Models The Basic Measurements

More information

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline Medical Display Performance and Evaluation Mike Silosky, MS University of Colorado, School of Medicine Dept. of Radiology 1 Objectives Review display function, QA metrics, procedures, and guidance provided

More information

APPLICATIONS OF DIGITAL IMAGE ENHANCEMENT TECHNIQUES FOR IMPROVED

APPLICATIONS OF DIGITAL IMAGE ENHANCEMENT TECHNIQUES FOR IMPROVED APPLICATIONS OF DIGITAL IMAGE ENHANCEMENT TECHNIQUES FOR IMPROVED ULTRASONIC IMAGING OF DEFECTS IN COMPOSITE MATERIALS Brian G. Frock and Richard W. Martin University of Dayton Research Institute Dayton,

More information

Analysis of WFS Measurements from first half of 2004

Analysis of WFS Measurements from first half of 2004 Analysis of WFS Measurements from first half of 24 (Report4) Graham Cox August 19, 24 1 Abstract Described in this report is the results of wavefront sensor measurements taken during the first seven months

More information

Python Quick-Look Utilities for Ground WFC3 Images

Python Quick-Look Utilities for Ground WFC3 Images Instrument Science Report WFC3 2008-002 Python Quick-Look Utilities for Ground WFC3 Images A.R. Martel January 25, 2008 ABSTRACT A Python module to process and manipulate ground WFC3 UVIS and IR images

More information

Technologist Quality Control Procedures

Technologist Quality Control Procedures Technologist Quality Control Procedures The specific procedures for the Technologist Quality Control Program are those specified in the most current ACR MRI QC Manual. WEEKLY MRI EQUIPMENT QUALITY CONTROL

More information

DCI Requirements Image - Dynamics

DCI Requirements Image - Dynamics DCI Requirements Image - Dynamics Matt Cowan Entertainment Technology Consultants www.etconsult.com Gamma 2.6 12 bit Luminance Coding Black level coding Post Production Implications Measurement Processes

More information

Module 3: Video Sampling Lecture 16: Sampling of video in two dimensions: Progressive vs Interlaced scans. The Lecture Contains:

Module 3: Video Sampling Lecture 16: Sampling of video in two dimensions: Progressive vs Interlaced scans. The Lecture Contains: The Lecture Contains: Sampling of Video Signals Choice of sampling rates Sampling a Video in Two Dimensions: Progressive vs. Interlaced Scans file:///d /...e%20(ganesh%20rana)/my%20course_ganesh%20rana/prof.%20sumana%20gupta/final%20dvsp/lecture16/16_1.htm[12/31/2015

More information

UNIVERSAL SPATIAL UP-SCALER WITH NONLINEAR EDGE ENHANCEMENT

UNIVERSAL SPATIAL UP-SCALER WITH NONLINEAR EDGE ENHANCEMENT UNIVERSAL SPATIAL UP-SCALER WITH NONLINEAR EDGE ENHANCEMENT Stefan Schiemenz, Christian Hentschel Brandenburg University of Technology, Cottbus, Germany ABSTRACT Spatial image resizing is an important

More information

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera PRELIMINARY POWER DETECTORS ENERGY DETECTORS MONITORS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS CMOS Beam Profiling Camera AVAILABLE MODEL Beamage 3.0 (⅔ in CMOS

More information

FEASIBILITY STUDY OF USING EFLAWS ON QUALIFICATION OF NUCLEAR SPENT FUEL DISPOSAL CANISTER INSPECTION

FEASIBILITY STUDY OF USING EFLAWS ON QUALIFICATION OF NUCLEAR SPENT FUEL DISPOSAL CANISTER INSPECTION FEASIBILITY STUDY OF USING EFLAWS ON QUALIFICATION OF NUCLEAR SPENT FUEL DISPOSAL CANISTER INSPECTION More info about this article: http://www.ndt.net/?id=22532 Iikka Virkkunen 1, Ulf Ronneteg 2, Göran

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

TechNote: MuraTool CA: 1 2/9/00. Figure 1: High contrast fringe ring mura on a microdisplay

TechNote: MuraTool CA: 1 2/9/00. Figure 1: High contrast fringe ring mura on a microdisplay Mura: The Japanese word for blemish has been widely adopted by the display industry to describe almost all irregular luminosity variation defects in liquid crystal displays. Mura defects are caused by

More information

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program 1 Contents 0.0 INTRODUCTION 4 0.1 Overview and Purpose 4 0.2 The Phantom 4 0.3 The Required Images 5 0.4 The Image

More information

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? White Paper Uniform Luminance Technology What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? Tom Kimpe Manager Technology & Innovation Group Barco Medical Imaging

More information

Automatic Defect Recognition in Industrial Applications

Automatic Defect Recognition in Industrial Applications Automatic Defect Recognition in Industrial Applications Klaus Bavendiek, Frank Herold, Uwe Heike YXLON International, Hamburg, Germany INDE 2007 YXLON. The reason why 1 Different Fields for Usage of ADR

More information

An Alternative Architecture for High Performance Display R. W. Corrigan, B. R. Lang, D.A. LeHoty, P.A. Alioshin Silicon Light Machines, Sunnyvale, CA

An Alternative Architecture for High Performance Display R. W. Corrigan, B. R. Lang, D.A. LeHoty, P.A. Alioshin Silicon Light Machines, Sunnyvale, CA R. W. Corrigan, B. R. Lang, D.A. LeHoty, P.A. Alioshin Silicon Light Machines, Sunnyvale, CA Abstract The Grating Light Valve (GLV ) technology is being used in an innovative system architecture to create

More information

Characterisation of the far field pattern for plastic optical fibres

Characterisation of the far field pattern for plastic optical fibres Characterisation of the far field pattern for plastic optical fibres M. A. Losada, J. Mateo, D. Espinosa, I. Garcés, J. Zubia* University of Zaragoza, Zaragoza (Spain) *University of Basque Country, Bilbao

More information

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology High Frequency X-Ray Technology Wide Screen Control with Touch-Technology The Q-VISION HF Series TM The Q-VISION Series TM integrates innovative high frequency design along with superb functionality, resulting

More information

Project Summary EPRI Program 1: Power Quality

Project Summary EPRI Program 1: Power Quality Project Summary EPRI Program 1: Power Quality April 2015 PQ Monitoring Evolving from Single-Site Investigations. to Wide-Area PQ Monitoring Applications DME w/pq 2 Equating to large amounts of PQ data

More information

Doubletalk Detection

Doubletalk Detection ELEN-E4810 Digital Signal Processing Fall 2004 Doubletalk Detection Adam Dolin David Klaver Abstract: When processing a particular voice signal it is often assumed that the signal contains only one speaker,

More information

3-D position sensitive CdZnTe gamma-ray spectrometers

3-D position sensitive CdZnTe gamma-ray spectrometers Nuclear Instruments and Methods in Physics Research A 422 (1999) 173 178 3-D position sensitive CdZnTe gamma-ray spectrometers Z. He *, W.Li, G.F. Knoll, D.K. Wehe, J. Berry, C.M. Stahle Department of

More information

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS Item Type text; Proceedings Authors Habibi, A. Publisher International Foundation for Telemetering Journal International Telemetering Conference Proceedings

More information

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study: Case Study: Scalable Edge Enhancement Introduction Edge enhancement is a post processing for displaying radiologic images on the monitor to achieve as good visual quality as the film printing does. Edges

More information

Audio Compression Technology for Voice Transmission

Audio Compression Technology for Voice Transmission Audio Compression Technology for Voice Transmission 1 SUBRATA SAHA, 2 VIKRAM REDDY 1 Department of Electrical and Computer Engineering 2 Department of Computer Science University of Manitoba Winnipeg,

More information

Guidelines for Assuring Softcopy Image Quality

Guidelines for Assuring Softcopy Image Quality Guidelines for Assuring Softcopy Image Quality What s inside? Quality Control Guidelines Softcopy QA testing and frequencies Danny Deroo Product and R&D Manager QA Products ABSTRACT To ensure diagnostic

More information

Ride the Lightning. A Review of Radiographic Physics

Ride the Lightning. A Review of Radiographic Physics Ride the Lightning A Review of Radiographic Physics Overview Basic principles of x-ray production Imaging equipment Quality Control Technical Factors Image QA Preparing for the registry The ARRT just wants

More information

Results of the June 2000 NICMOS+NCS EMI Test

Results of the June 2000 NICMOS+NCS EMI Test Results of the June 2 NICMOS+NCS EMI Test S. T. Holfeltz & Torsten Böker September 28, 2 ABSTRACT We summarize the findings of the NICMOS+NCS EMI Tests conducted at Goddard Space Flight Center in June

More information

Signal Stability Analyser

Signal Stability Analyser Signal Stability Analyser o Real Time Phase or Frequency Display o Real Time Data, Allan Variance and Phase Noise Plots o 1MHz to 65MHz medium resolution (12.5ps) o 5MHz and 10MHz high resolution (50fs)

More information

Perfecting the Package Bare and Overmolded Stacked Dies. Understanding Ultrasonic Technology for Advanced Package Inspection. A Sonix White Paper

Perfecting the Package Bare and Overmolded Stacked Dies. Understanding Ultrasonic Technology for Advanced Package Inspection. A Sonix White Paper Perfecting the Package Bare and Overmolded Stacked Dies Understanding Ultrasonic Technology for Advanced Package Inspection A Sonix White Paper Perfecting the Package Bare and Overmolded Stacked Dies Understanding

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

New appraoch for X-ray weld inspection of pipeline segments

New appraoch for X-ray weld inspection of pipeline segments New appraoch for X-ray weld inspection of pipeline segments Lennart Schulenburg VisiConsult X-ray Systems & Solutions GmbH 1 Overview Weld inspection in heavy industries ( Pipe and Tank ) Analogue Film

More information

Part 1: Introduction to Computer Graphics

Part 1: Introduction to Computer Graphics Part 1: Introduction to Computer Graphics 1. Define computer graphics? The branch of science and technology concerned with methods and techniques for converting data to or from visual presentation using

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

RADIOGRAPHIC PERFORMANCE OF CYGNUS 1 AND THE FEBETRON 705

RADIOGRAPHIC PERFORMANCE OF CYGNUS 1 AND THE FEBETRON 705 RADIOGRAPHIC PERFORMANCE OF CYGNUS 1 AND THE FEBETRON 705 E. Rose ξ, R. Carlson, J. Smith Los Alamos National Laboratory, PO Box 1663, Mail Stop P-947 Los Alamos, NM 87545, USA Abstract Spot sizes are

More information

Modulation transfer function of a liquid crystal spatial light modulator

Modulation transfer function of a liquid crystal spatial light modulator 1 November 1999 Ž. Optics Communications 170 1999 221 227 www.elsevier.comrlocateroptcom Modulation transfer function of a liquid crystal spatial light modulator Mei-Li Hsieh a, Ken Y. Hsu a,), Eung-Gi

More information

Connection for filtered air

Connection for filtered air BeamWatch Non-contact, Focus Spot Size and Position monitor for high power YAG, Diode and Fiber lasers Instantly measure focus spot size Dynamically measure focal plane location during start-up From 1kW

More information

The Syscal family of resistivity meters. Designed for the surveys you do.

The Syscal family of resistivity meters. Designed for the surveys you do. The Syscal family of resistivity meters. Designed for the surveys you do. Resistivity meters may conveniently be broken down into several categories according to their capabilities and applications. The

More information

ITS-I. Test station for evaluation of image quality of image intensifier tubes. Fig. 1. Photo of the ITS-I test station: a)photo, b)block diagram

ITS-I. Test station for evaluation of image quality of image intensifier tubes. Fig. 1. Photo of the ITS-I test station: a)photo, b)block diagram OS-1 stage Monitor S-I support VM-I microscope M-I microscope Control center Target projector OS-2 stage DC-I camera Tube holder P-I platform IM meter Target slider a) b) BASIC INFORMATION: LVS voltage

More information

COMPARED IMPROVEMENT BY TIME, SPACE AND FREQUENCY DATA PROCESSING OF THE PERFORMANCES OF IR CAMERAS. APPLICATION TO ELECTROMAGNETISM

COMPARED IMPROVEMENT BY TIME, SPACE AND FREQUENCY DATA PROCESSING OF THE PERFORMANCES OF IR CAMERAS. APPLICATION TO ELECTROMAGNETISM COMPARED IMPROVEMENT BY TIME, SPACE AND FREQUENCY DATA PROCESSING OF THE PERFORMANCES OF IR CAMERAS. APPLICATION TO ELECTROMAGNETISM P. Levesque 1, P.Brémond 2, J.-L. Lasserre 3, A. Paupert 2, D. L. Balageas

More information

DCI Memorandum Regarding Direct View Displays

DCI Memorandum Regarding Direct View Displays 1. Introduction DCI Memorandum Regarding Direct View Displays Approved 27 June 2018 Digital Cinema Initiatives, LLC, Member Representatives Committee Direct view displays provide the potential for an improved

More information

homework solutions for: Homework #4: Signal-to-Noise Ratio Estimation submitted to: Dr. Joseph Picone ECE 8993 Fundamentals of Speech Recognition

homework solutions for: Homework #4: Signal-to-Noise Ratio Estimation submitted to: Dr. Joseph Picone ECE 8993 Fundamentals of Speech Recognition INSTITUTE FOR SIGNAL AND INFORMATION PROCESSING homework solutions for: Homework #4: Signal-to-Noise Ratio Estimation submitted to: Dr. Joseph Picone ECE 8993 Fundamentals of Speech Recognition May 3,

More information

Blueline, Linefree, Accuracy Ratio, & Moving Absolute Mean Ratio Charts

Blueline, Linefree, Accuracy Ratio, & Moving Absolute Mean Ratio Charts INTRODUCTION This instruction manual describes for users of the Excel Standard Celeration Template(s) the features of each page or worksheet in the template, allowing the user to set up and generate charts

More information

Solution for Nonuniformities and Spatial Noise in Medical LCD Displays by Using Pixel-Based Correction

Solution for Nonuniformities and Spatial Noise in Medical LCD Displays by Using Pixel-Based Correction Solution for Nonuniformities and Spatial Noise in Medical LCD Displays by Using Pixel-Based Correction Tom Kimpe, Albert Xthona, Paul Matthijs, and Lode De Paepe Liquid crystal displays (LCD) are rapidly

More information

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017 Display Quality Assurance: Considerations When Establishing a Display QA Program Mike Silosky, M.S. 8/3/2017 Objectives and Outline Why, Who, What, When, Where? Discuss the resources that may be needed

More information

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL COMPOSITE VIDEO METER MODEL VLM- COMPOSITE VIDEO METER MODEL VLM- NTSC TECHNICAL INSTRUCTION MANUAL VLM- NTSC TECHNICAL INSTRUCTION MANUAL INTRODUCTION EASY-TO-USE VIDEO LEVEL METER... SIMULTANEOUS DISPLAY...

More information

Communication Theory and Engineering

Communication Theory and Engineering Communication Theory and Engineering Master's Degree in Electronic Engineering Sapienza University of Rome A.A. 2018-2019 Practice work 14 Image signals Example 1 Calculate the aspect ratio for an image

More information

CAPTURE CAPTURE. VERSiON 1.2. Specialists in Medical. Digital Imaging Solutions

CAPTURE CAPTURE. VERSiON 1.2. Specialists in Medical. Digital Imaging Solutions CAPTURE Add-on module for capturing images from analog video sources CAPTURE anything to DICOm VERSiON 1.2 Specialists in Medical Digital Imaging Solutions iq-capture Connect any medical device to your

More information

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography Alexia Valasi 1, Maria Argyrou 1, Ioannis Floros 1, Elisavet Molyvda 2, Anastasios Siountas 2, Stella Synefia 1, Ioannis

More information

Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope

Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH CERN BEAMS DEPARTMENT CERN-BE-2014-002 BI Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope M. Gasior; M. Krupa CERN Geneva/CH

More information

Module 1: Digital Video Signal Processing Lecture 3: Characterisation of Video raster, Parameters of Analog TV systems, Signal bandwidth

Module 1: Digital Video Signal Processing Lecture 3: Characterisation of Video raster, Parameters of Analog TV systems, Signal bandwidth The Lecture Contains: Analog Video Raster Interlaced Scan Characterization of a video Raster Analog Color TV systems Signal Bandwidth Digital Video Parameters of a digital video Pixel Aspect Ratio file:///d

More information

FFDM Quality Control in Canada - a Vendor Neutral Approach

FFDM Quality Control in Canada - a Vendor Neutral Approach FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011

More information

Color Nonuniformity in Projection-Based Displays: Analysis and Solutions

Color Nonuniformity in Projection-Based Displays: Analysis and Solutions IEEE TRANSACTIONS ON VISUALIZATION AND COMPUTER GRAPHICS, VOL. XX, NO. Y, MONTH 23 1 Color Nonuniformity in Projection-Based Displays: Analysis and Solutions Aditi Majumder, Rick Stevens Abstract Large-area

More information

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Introduction Active neurons communicate by action potential firing (spikes), accompanied

More information

IHE. Display Consistency Test Plan for Image Displays HIMMS and RSNA. Integrating the Healthcare Enterprise

IHE. Display Consistency Test Plan for Image Displays HIMMS and RSNA. Integrating the Healthcare Enterprise HIMMS and RSNA IHE Integrating the Healthcare Enterprise Display Consistency Test Plan for Displays 2001-05-01 Marco Eichelberg 1, Klaus Kleber 2, Jörg Riesmeier 1, Adapted for IHE Year 3 by David Maffitt

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

Y.XST225-VF. INTERCONTROLE Escoffier 1 XYLON MG225VF RX PDS

Y.XST225-VF. INTERCONTROLE Escoffier 1 XYLON MG225VF RX PDS Y.XST225-VF 1 XYLON MG225VF - 0509- RX0905007PDS 1 Table of contents Y.XST225-VF High detail visibility - examples Y.XST225-VF vs. conventional X-ray systems Y.XST225-VF vs. µ-focus X-ray systems Unique

More information

Mechanical aspects, FEA validation and geometry optimization

Mechanical aspects, FEA validation and geometry optimization RF Fingers for the new ESRF-EBS EBS storage ring The ESRF-EBS storage ring features new vacuum chamber profiles with reduced aperture. RF fingers are a key component to ensure good vacuum conditions and

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0 CHEM 411L Instrumental Analysis Laboratory Revision 2.0 Noise In this laboratory exercise we will determine the Signal-to-Noise (S/N) ratio for an IR spectrum of Air using a Thermo Nicolet Avatar 360 Fourier

More information

Understanding PQR, DMOS, and PSNR Measurements

Understanding PQR, DMOS, and PSNR Measurements Understanding PQR, DMOS, and PSNR Measurements Introduction Compression systems and other video processing devices impact picture quality in various ways. Consumers quality expectations continue to rise

More information

Beam test of the QMB6 calibration board and HBU0 prototype

Beam test of the QMB6 calibration board and HBU0 prototype Beam test of the QMB6 calibration board and HBU0 prototype J. Cvach 1, J. Kvasnička 1,2, I. Polák 1, J. Zálešák 1 May 23, 2011 Abstract We report about the performance of the HBU0 board and the optical

More information

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Julie Smyth & Philip Doyle Regional Medical Physics Service Outline Preamble Image Quality

More information

Selection Criteria for X-ray Inspection Systems for BGA and CSP Solder Joint Analysis

Selection Criteria for X-ray Inspection Systems for BGA and CSP Solder Joint Analysis Presented at Nepcon Shanghai 2003 Abstract Selection Criteria for X-ray Inspection Systems for BGA and CSP Solder Joint Analysis Dr. David Bernard, Dage Precision Industries, 158-29 Hua Shan Road, Feng

More information

An Overview of Video Coding Algorithms

An Overview of Video Coding Algorithms An Overview of Video Coding Algorithms Prof. Ja-Ling Wu Department of Computer Science and Information Engineering National Taiwan University Video coding can be viewed as image compression with a temporal

More information

FPA (Focal Plane Array) Characterization set up (CamIRa) Standard Operating Procedure

FPA (Focal Plane Array) Characterization set up (CamIRa) Standard Operating Procedure FPA (Focal Plane Array) Characterization set up (CamIRa) Standard Operating Procedure FACULTY IN-CHARGE Prof. Subhananda Chakrabarti (IITB) SYSTEM OWNER Hemant Ghadi (ghadihemant16@gmail.com) 05 July 2013

More information

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series Introduction System designers and device manufacturers so long have been using one set of instruments for creating digitally modulated

More information

Import and quantification of a micro titer plate image

Import and quantification of a micro titer plate image BioNumerics Tutorial: Import and quantification of a micro titer plate image 1 Aims BioNumerics can import character type data from TIFF images. This happens by quantification of the color intensity and/or

More information

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality.

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality. The Big Picture of Quality MRI Quality Assurance and ACR MRI Accreditation Program Chen Lin, PhD Indiana University School of Medicine & Clarian Health Partners Diagnosis accuracy Image quality Knowledge

More information

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities J L Poletti 1995/1 NRL Report GUIDELINES FOR QUALITY ASSURANCE IN RADIATION PROTECTION FOR

More information

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio Interface Practices Subcommittee SCTE STANDARD SCTE 119 2018 Measurement Procedure for Noise Power Ratio NOTICE The Society of Cable Telecommunications Engineers (SCTE) / International Society of Broadband

More information

Financial disclosure statement. Fluoroscopic Equipment Design: What s s Different with Flat Panel? Concept of flat panel imager

Financial disclosure statement. Fluoroscopic Equipment Design: What s s Different with Flat Panel? Concept of flat panel imager Fluoroscopic Equipment Design: What s s Different with Flat Panel? John Rowlands Financial disclosure statement Research supported by Anrad Corp Anrad Corp is a manufacturers of selenium based flat panel

More information

ENERGY STAR Program Requirements Product Specification for Televisions. Eligibility Criteria Version 5.3

ENERGY STAR Program Requirements Product Specification for Televisions. Eligibility Criteria Version 5.3 ENERGY STAR Program Requirements Product Specification for Televisions Eligibility Criteria Version 5.3 Following is the Version 5.3 ENERGY STAR Product Specification for Televisions. A product shall meet

More information

Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs. By: Jeff Smoot, CUI Inc

Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs. By: Jeff Smoot, CUI Inc Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs By: Jeff Smoot, CUI Inc Rotary encoders provide critical information about the position of motor shafts and thus also their

More information

Global Trade Medical Supplies

Global Trade Medical Supplies Features: Achieve Clarity True to the Source Data A medical monitor needs to be capable of high brightness in order to meet performance standards. However, in order to achieve high brightness in an LCD

More information

Multimedia Systems Video I (Basics of Analog and Digital Video) Mahdi Amiri April 2011 Sharif University of Technology

Multimedia Systems Video I (Basics of Analog and Digital Video) Mahdi Amiri April 2011 Sharif University of Technology Course Presentation Multimedia Systems Video I (Basics of Analog and Digital Video) Mahdi Amiri April 2011 Sharif University of Technology Video Visual Effect of Motion The visual effect of motion is due

More information

Focused Ion Beam System MI4050

Focused Ion Beam System MI4050 SCIENTIFIC INSTRUMENT NEWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Focused Ion Beam System MI4050 Yasushi Kuroda *1, Yoshihisa

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. Multichannel Scaler Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. The ideal solution for: Time-resolved single-photon counting Phosphorescence lifetime spectrometry Atmospheric and

More information

Micro-DCI 53ML5100 Manual Loader

Micro-DCI 53ML5100 Manual Loader Micro-DCI 53ML5100 Manual Loader Two process variable inputs Two manually controlled current outputs Multiple Display Formats: Dual Channel Manual Loader, Single Channel Manual Loader, Manual Loader with

More information

Release Notes for LAS AF version 1.8.0

Release Notes for LAS AF version 1.8.0 October 1 st, 2007 Release Notes for LAS AF version 1.8.0 1. General Information A new structure of the online help is being implemented. The focus is on the description of the dialogs of the LAS AF. Configuration

More information

TROUBLESHOOTING DIGITALLY MODULATED SIGNALS, PART 2 By RON HRANAC

TROUBLESHOOTING DIGITALLY MODULATED SIGNALS, PART 2 By RON HRANAC Originally appeared in the July 2006 issue of Communications Technology. TROUBLESHOOTING DIGITALLY MODULATED SIGNALS, PART 2 By RON HRANAC Digitally modulated signals are a fact of life in the modern cable

More information

The Measurement Tools and What They Do

The Measurement Tools and What They Do 2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

EDDY CURRENT IMAGE PROCESSING FOR CRACK SIZE CHARACTERIZATION

EDDY CURRENT IMAGE PROCESSING FOR CRACK SIZE CHARACTERIZATION EDDY CURRENT MAGE PROCESSNG FOR CRACK SZE CHARACTERZATON R.O. McCary General Electric Co., Corporate Research and Development P. 0. Box 8 Schenectady, N. Y. 12309 NTRODUCTON Estimation of crack length

More information

VeriLUM 5.2. Video Display Calibration And Conformance Tracking. IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD USA

VeriLUM 5.2. Video Display Calibration And Conformance Tracking. IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD USA VeriLUM 5.2 Video Display Calibration And Conformance Tracking IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD 20824 USA Voice: 240-395-1600 Fax: 240-395-1601 Web: www.image-smiths.com Technical Support

More information

Rounding Considerations SDTV-HDTV YCbCr Transforms 4:4:4 to 4:2:2 YCbCr Conversion

Rounding Considerations SDTV-HDTV YCbCr Transforms 4:4:4 to 4:2:2 YCbCr Conversion Digital it Video Processing 김태용 Contents Rounding Considerations SDTV-HDTV YCbCr Transforms 4:4:4 to 4:2:2 YCbCr Conversion Display Enhancement Video Mixing and Graphics Overlay Luma and Chroma Keying

More information

Colour Reproduction Performance of JPEG and JPEG2000 Codecs

Colour Reproduction Performance of JPEG and JPEG2000 Codecs Colour Reproduction Performance of JPEG and JPEG000 Codecs A. Punchihewa, D. G. Bailey, and R. M. Hodgson Institute of Information Sciences & Technology, Massey University, Palmerston North, New Zealand

More information

Sample Analysis Design. Element2 - Basic Software Concepts (cont d)

Sample Analysis Design. Element2 - Basic Software Concepts (cont d) Sample Analysis Design Element2 - Basic Software Concepts (cont d) Samples per Peak In order to establish a minimum level of precision, the ion signal (peak) must be measured several times during the scan

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

DRAFT. Proposal to modify International Standard IEC

DRAFT. Proposal to modify International Standard IEC Imaging & Color Science Research & Product Development 2528 Waunona Way, Madison, WI 53713 (608) 222-0378 www.lumita.com Proposal to modify International Standard IEC 61947-1 Electronic projection Measurement

More information

RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation

RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation The DLS-20 Hiden s 20mm Triple Filter Quadrupole By comparison, 6mm Triple Filter Quadrupole Quadrupole High resolution Quadrupoles

More information

DVG-5000 Motion Pattern Option

DVG-5000 Motion Pattern Option AccuPel DVG-5000 Documentation Motion Pattern Option Manual DVG-5000 Motion Pattern Option Motion Pattern Option for the AccuPel DVG-5000 Digital Video Calibration Generator USER MANUAL Version 1.00 2

More information

ZONE PLATE SIGNALS 525 Lines Standard M/NTSC

ZONE PLATE SIGNALS 525 Lines Standard M/NTSC Application Note ZONE PLATE SIGNALS 525 Lines Standard M/NTSC Products: CCVS+COMPONENT GENERATOR CCVS GENERATOR SAF SFF 7BM23_0E ZONE PLATE SIGNALS 525 lines M/NTSC Back in the early days of television

More information

Update on Antenna Elevation Pattern Estimation from Rain Forest Data

Update on Antenna Elevation Pattern Estimation from Rain Forest Data Update on Antenna Elevation Pattern Estimation from Rain Forest Data Manfred Zink ENVISAT Programme, ESA-ESTEC Keplerlaan 1, 2200 AG, Noordwijk The Netherlands Tel: +31 71565 3038, Fax: +31 71565 3191

More information

LeCroy Digital Oscilloscopes

LeCroy Digital Oscilloscopes LeCroy Digital Oscilloscopes Get the Complete Picture Quick Reference Guide QUICKSTART TO SIGNAL VIEWING Quickly display a signal View with Analog Persistence 1. Connect your signal. When you use a probe,

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information