AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)
|
|
- Dulcie Leonard
- 5 years ago
- Views:
Transcription
1 AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop icon 3. Select the desired mode of operation by choosing the appropriate workspace in the File/Open Workspace menu. The workspaces are found in the directory D:V614r1/Workspaces. The following workspaces are good starting points and can be used for most samples: Tapping Mode: smif_default_tapping_d.wks Contact Mode: smif_default_contact_d.wks 4. Click on Scan in the Workspace window to connect the software to the instrument 5. Mount the appropriate AFM tip into the cantilever holder. Be sure that the tip is in contact with the back end and left side of the groove per the following picture. TESP Tip properly positioned in cantilever holder prior to clamping TESP Tip properly positioned in cantilever holder after clamping You may use your own AFM tip, or purchase one of the following from SMIF: Tapping Mode: TESP silicon probe Contact Mode: ORC8-10 silicon nitride probe 6. Mount the cantilever holder onto the end of the scanner head a. Tighten the set screw to unclamp the scanner head from the dove tail and remove the scanner head from the cradle. Be sure not to pull on the cables that connect the scanner head to the electronics box b. Mount the cantilever holder by gently fitting it onto the pins at the end of the scanner head. DO NOT use excessive force as this may damage the scanner head c. Replace the scanner head in the cradle and loosen the set screw to clamp the head assembly in the dove tail 7. Align the laser a. Turn the illumination off by entering a 0 in the Illumination box of the Other controls menu in the Scan ParmList window. b. Place the blank side of a business card face up on the granite stage so that the laser is visible on the card c. Open the Meter window by clicking on Meter in the Workspace window Page 1 of 6 Revision 7
2 d. Using the two knobs at the top of the scanner head, follow the diagram below to adjust the laser position so that the laser is visually aligned with is the tip of the cantilever on the AFM tip. 1. Move the laser in the x direction to find AFM tip edge. Then move slightly back from the edge. 2. Move the laser in the y direction to find the cantilever. 3. Move the laser in the x direction to the right to find the cantilever edge. Now make fine adjustments to the x and y laser position knobs to maximize the signal sum in the meter window as shown below Red dot centered for detector alignment Signal sum maximized for laser alignment e. Remove the business card from the granite stage 8. Align the photodetector a. Use the two knobs on the left side of the scanner head to adjust the photodetector position 1. Coarse adjustment: Center the position of the red dot in the window on the scanner head 2. Fine adjustment: Center the position of the red dot in the cross-hairs of the meter display screen as shown above Tapping Mode: Set the Vertical Deflection to 0.0 volts ( 0.5V) Contact Mode: Set the Vertical Deflection to about 2.0 Volts b. Close the Meter window Page 2 of 6 Revision 7
3 9. Load sample a. Place sample on stage b. Select Navigate window and move the Z-position up using the Z-motor up arrow button in the Navigate window. (This is a precaution to prevent the sample from hitting the head assembly) c. Increase illumination from zero to 100 d. Move the sample under the probe using the X-Y stage movement arrows in the Navigate window. 10. Locate and focus on tip a. Click on Locate Tip in the Navigate window b. Zoom all the way out by clicking on Zoom Out c. Center the tip of the cantilever under the cross hairs using the two adjustment knobs located to the left of the optical objective on the microscope d. Focus on the tip end of the cantilever using the Optics arrows keys in the Navigate window e. Click on Zoom in to zoom all the way in or use the up arrow to zoom partially in on the tip and re-focus and re-center the tip if necessary. f. Click OK (The software will automatically change to the focus on sample surface menu) 11. Focus on the sample surface a. Zoom out to the maximum position to get the largest field of view possible b. Use the Z-motor down arrow button in the Navigate window to visually close the gap between the sample and the scanner head to about 2-3mm. c. Now look at image on screen and keep adjusting the Z-motor position until the sample surface is in focus. If the sample is smooth and has no surface features to focus on, then change the Focus On setting to Tip Reflection and adjust the Z-position until the reflection of the AFM tip is in focus. Page 3 of 6 Revision 7
4 d. Move to the desired measurement area of the sample under the cross hairs using the X-Y arrow keys in the Navigate window e. Zoom in on the area of interest and fine focus using the Z-position if necessary 12. Tune Cantilever (Tapping Mode AFM only) a. Click on the Tuning Fork Icon b. Set the Start and End frequency to bracket the resonant frequency value given on the box that the cantilever probe came from. (Typically 0kHz 500kHz for TESP silicon probes) c. Set the target amplitude to 2 volts d. Set the peak offset to 5% e. Click on Auto Tune button f. Note: If the tool does not auto tune properly and the probe tip is in good condition, try resetting the controller communication connection by selecting Tools/Reset Controller and then repeating the auto tune. g. When completed, click the Exit button 13. Set Initial Scan Parameters a. In the Scan Controls panel, enter the following settings: 1. Scan Size: Set to desired measurement area (< 70 m) 2. Aspect Ratio 1:1 3. X and Y offsets: 0 4. Scan angle: 0 5. Scan rate: Tapping mode: 1 Hz Contact mode: 2 Hz 6. Samples/line: 256, 512, or Lines: 256, 512, or Slow scan enabled 9. XY closed loop Lower setting: Lower resolution, faster speed Higher setting: Higher resolution, slower speed Off for scan size of < 1 m On for scan size of > 1 m b. For Tapping Mode AFM, in the Feedback Controls Panel, enter the following settings: 1. Integral Gain: Proportional Gain: Leave other settings alone c. For Contact Mode AFM, in the Feedback Controls Panel, enter the following settings: 1. Setpoint: 0 volts (2 volts difference from value set in Step 8) 2. Integral Gain: Proportional Gain: Close the instrument cover 15. Click on the Engage Icon Engage Withdraw Page 4 of 6 Revision 7
5 16. Optimize Scan Parameters a. Tapping Mode AFM 1. If the line in the Z-Center position moves far to the extended end (indicator turns red), the tip has false engaged. Disengage the probe tip and execute the engage command again 2. Check to see if Trace and Retrace are tracking each other well. If they are tracking, the lines should look the same, but they will not necessarily exactly overlap each other. 3. If they are not tracking well, adjust the Setpoint and/or Gains, to improve the tracking (See training notebook pages 36-41). It is often easier to disable the slow scan axis while making adjustments, so that the scanned feature stays in the field of view. Typical adjustments are as follows: Lower the setpoint to improve tracking between the trace and re-trace Lower the gains to reduce the noise b. Contact Mode AFM 1. Once engaged increase the Setpoint by using the right arrow key and observe the Z-Center Position. If the line in the Z-Center position moves far to the extended end (indicator turns red), the tip has false engaged. Disengage the probe tip, increase the setpoint by +2 volts and execute the engage command again. If the Z-Center position does not change greatly, you are on the sample surface and can go to the next step. 2. Check to see if Trace and Retrace are tracking each other well. If they are tracking, the lines should look the same, but they will not necessarily exactly overlap each other. 3. If they are not tracking well, adjust the Setpoint, Gains, and/or Scan Rate to improve the tracking (See training notebook pages and 36-41). It is often easier to disable the slow scan axis while making adjustments, so that the scanned feature stays in the field of view. 4. Remember that in Contact Mode, increasing the Setpoint increases the force on the sample. Minimizing the imaging force is recommended to prevent damage to the sample and the probe tip. 17. Set desired Scan Size, Scan Angle, and Offsets Page 5 of 6 Revision 7
6 Image Capturing 1. Choose Capture Filename from the RealTime pull down menu 2. Choose your directory to store the image, and enter a filename for the image. Note: directories must be created in Windows outside of the AFM software. Click OK. 3. Click on the camera icon to begin capturing an image. The image will begin capturing at the beginning of a scan frame. To move the probe to the beginning of a scan frame, click on the frame up or frame down icon 4. When capturing, the status bar at the bottom of the screen will indicate Capture: On, and when the capture is completed, it will indicate Capture: Done When you are finished: 1. Click on the Withdraw icon to stop scanning Engage Withdraw 2. Open the instrument cover 3. Select the Navigate window a. Raise the Probe head away from the sample using the Z-motor up arrow in the Navigate window b. Move the stage to the load position (completely out from under the scanner head) using the X-Y stage arrows in the Navigate window 4. Unload your sample 5. Remove the cantilever holder from the end of the scanner head and unmount your probe (Store your probe in your personal Gel-Pak storage box) 6. Close the instrument cover 7. If you wish to save changes to the workspace, Choose File/Save Workspace As and give the file a new workspace name. DO NOT save changes to the SMIF default workspaces 8. Exit Nanoscope software 9. Logout of the Log Usage system on the SMIF web site Page 6 of 6 Revision 7
SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University
SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize
More informationQuick Start Bruker Dimension Icon AFM
Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch
More informationStandard Operating Procedure of nanoir2-s
Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),
More informationAFM Standard Operating Procedure
2013 AFM Standard Operating Procedure Karen Gaskell, David Ramsdell Surface Analysis Centre Department of Chemistry and Biochemistry University of Maryland 1/1/2013 Content Page 1 Hardware 2 1.1 MultiMode
More informationFigure 1. MFP-3D software tray
Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure
More informationStandard AFM Modes User s Manual
Standard AFM Modes User s Manual Part #00-0018-01 Issued March 2014 2014 by Anasys Instruments Inc, 325 Chapala St, Santa Barbara, CA 93101 Page 1 of 29 Table of contents Chapter 1. AFM Theory 3 1.1 Detection
More informationUniversity of MN, Minnesota Nano Center Standard Operating Procedure
Equipment Name: Atomic Force Microscope Badger name: afm Revisionist Paul Kimani Model: Dimension 3000 Date: October 1, 2013 Location: Bay 1 A. Description The Dimension 3000 consists of a rigid stage
More informationMagnetic Force Microscope (MFM)
Magnetic Force Microscope (MFM) [Digital Instruments NanoScopeIII] magnetic and topographic resolution ~20 nm [at least several hours/sample (1 cm 2 polished single grains, thin sections, or thin films)]
More informationSTANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM
STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact,
More informationQuick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core
Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core 1 Turn On the laser power 2 Open enclosure: - lift the door latch and open the enclosure door. 3 2 1 1.Unlock scanner: Lift the lever to the
More informationTT AFM LongBeach Procedures and Protocols V2.1
TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn
More informationAsylum MFP-3D Standard Operating Procedures
Asylum MFP-3D Standard Operating Procedures Document Version 1.5, November 2013 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope
More informationDektak Step by Step Instructions:
Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample
More informationAsylum MFP-3D Standard Operating Procedures
Asylum MFP-3D Standard Operating Procedures Document Version 2, September 2017 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope
More informationOperation Procedure for Phillips XL30 ESEM
Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor
More informationB-AFM. v East 33rd St., Signal Hill, CA (888)
B-AFM The B-AFM is a basic AFM that provides routine scanning. Ideal for scientists and educators, the B-AFM is capable of creating high-resolution topography images of nanostructures in standard scanning
More informationTopic: Instructional David G. Thomas December 23, 2015
Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December
More informationScanning Probe Microscope Training. Wenhui Pang
Scanning Probe Microscope Training Wenhui Pang Background - Comparison of AFM with Other Imaging Modalities Optical Microscopy SEM TEM AFM Resolution XY 200 nm 2 nm 0.1 nm 1 nm Z 500 nm N/A N/A 0.1 nm
More informationINSTALATION PROCEDURE
INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items
More informationINSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE
INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE ? 8 LCD Screen? 10.4 LCD Screen LCD Zoom Microscope Instruction Manual Please read the Instruction Manual carefully before installation and keep it for future
More informationScanning Electron Microscopy (FEI Versa 3D Dual Beam)
Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques
More informationSC24 Magnetic Field Cancelling System
SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface
More information[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL]
[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL] VER: 2.0 TABLE OF CONTENT 1. INTRODUCTION...- 4-1.1 THEORY OF OPERATION...- 4-1.1.1 Principle of Atomic Force Microscope (AFM)... - 4-2. INSTALLATION &
More informationSC24 Magnetic Field Cancelling System
SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface
More informationKratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions
Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information
More informationSEM- EDS Instruction Manual
SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through
More informationSIMET AVIKO D INSTRUCTION MANUAL SORTING Solutions, Ltd.
SIMET AVIKO D INSTRUCTION MANUAL 1870 SORTING Solutions, Ltd. 1. TABLE OF CONTENTS 1. TABLE OF CONTENTS...1 2. INTRODUCTION...2 2.1. Application...2 2.2. Operating Conditions...2 2.3. Electro - Optical
More information12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492
12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor to automatically
More informationDo not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company.
PLEASE, READ Notice: Copyright: A delivery set of your device may differ from the set described in the given manual. For more detailed information refer to the specification of your contract. No part of
More informationME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition
ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods
More informationSX7. Saga 7" Super Bright HDMI/3G-SDI Field Monitor with 3D-LUTs. Quick Start Guide. What s Included CHECKED BY
SX7 Quick Start Guide Saga 7" Super Bright HDMI/3G-SDI Field Monitor with 3D-LUTs What s Included 1 x Saga X7 Monitor 1 x V-Mount Plate (Attached) 1 x Mini-XLR to P-TAP Cable 1 x Dual Sony L Battery Adapter
More informationApplication Note AN-708 Vibration Measurements with the Vibration Synchronization Module
Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished
More informationimso-104 Manual Revised August 5, 2011
imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel
More information9070 Smart Vibration Meter Instruction Manual
9070 Smart Vibration Meter Instruction Manual Overall machine and bearing conditions: vibration values are displayed with color coded alarm levels for ISO values and Bearing Damage (BDU). Easy vibration
More informationISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:
Compensation look-up-table (LUT) and How to Generate Contents: Description Background theory Basic LUT pg 2 Creating a LUT pg 3 Using the LUT pg 7 Comment pg 9 The compensation look-up-table (LUT) contains
More informationAX20. Atlas 19.5" 3G-SDI/HDMI Field and Studio Monitor with 3D LUTs & Scopes. Quick Start Guide. What s Included CHECKED BY
AX20 Quick Start Guide Atlas 19.5" 3G-SDI/HDMI Field and Studio Monitor with 3D LUTs & Scopes What s Included 1 x Atlas 19.5" Monitor 1 x AC Adapter 1 x Sunhood CHECKED BY AX20 FRONT 1920 x 1080 19.5 inch
More informationPreparing for remote data collection at NE-CAT
Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person
More informationDektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP
Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It
More informationF7000NV ROBOT VISION OPERATING MANUAL
Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting
More informationUSER MANUEL. SNIPE 2 Ref R13
USER MANUEL SNIPE 2 Ref. 0141317R13 Contents 1. General Information 1-1. Introduction 1-2. Proper use and operation 1-3. Safety notes......... 2 3 3 2. Contents 2-1. Accessory included 2-2. Name of parts......
More informationPSC300 Operation Manual
PSC300 Operation Manual Version 9.10 General information Prior to any attempt to operate this Columbia PSC 300, operator should read and understand the complete operation of the cubing system. It is very
More informationBitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.
BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features
More informationFluke 190-Series II Firmware Upgrade V11.44
Fluke 190-Series II Firmware Upgrade V11.44 Requirements 1. Fluke 190- Series II ScopeMeter with firmware prior to V11.44 2. Supported models are: 190-102, 190-104, 190-062, 190-202, 190-204, 190-502,
More informationPersonal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.
4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject
More informationFirst Question: Camera head. Lighting unit. Shooting stage
Elmo P30 Visualiser First Question: Q. Is everyone familiar with exactly what a visualiser is? A. A visualiser is effectively a camera on an arm, usually with a shooting stage and its own lighting source.
More informationData Collection Using APEX3. March 30, Chemical Crystallography Laboratory
Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas
More informationChapter 4. Dish Antenna Installation. Installing a DISH 500 Antenna. Finding the Satellites
These instructions guide you through the installation of a satellite system which includes your receiver (included with this manual), and a DISH Pro DISH 500 antenna system that can be identified by the
More informationThree Axis Digital Readout System
NEWALL MEASUREMENT SYSTEMS C80 Three Axis Digital Readout System CONTENTS 2 SPECIFICATIONS 3 CONNECTIONS 4 MOUNTING 4 Arm Mounting (Non-adjustable) 4 Arm Mounting (Adjustable) 5 Face Mounting (Adjustable)
More informationNanoGiant Oscilloscope/Function-Generator Program. Getting Started
Getting Started Page 1 of 17 NanoGiant Oscilloscope/Function-Generator Program Getting Started This NanoGiant Oscilloscope program gives you a small impression of the capabilities of the NanoGiant multi-purpose
More informationNanonis STM Simulator Tutorial
Nanonis STM Simulator Tutorial Software Version 4 Manual Version 4.0 Contents Introduction... 4 Minimum System Requirements and Installation... 5 Getting Started... 6 Session Directories... 6 Online Help...
More informationSpectrum Analyser Basics
Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,
More informationJS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA
JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA The JS007WQK is loaded with userfriendly features and is ideal for use in heavy duty vehicles.
More informationTechnical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework
Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning
More informationDektakXT Profilometer. Standard Operating Procedure
DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,
More informationimso-104 Manual Revised July 19, 2012
imso-104 Manual Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel 2.21 On / Off 2.22
More informationNOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.
Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only
More informationSpectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303)
Spectroscopy Module Vescent Photonics, Inc. www.vescentphotonics.com 4865 E. 41 st Ave Denver, CO 80216 Phone: (303)-296-6766 Fax: (303)-296-6783 General Warnings and Cautions The following general warnings
More informationVideo wall Precautions when moving and handling. Precaution. Please adhere to the warning signs printed on the box.
Video wall Precaution 4-1. Precautions when moving and handling Please adhere to the warning signs printed on the box. Model: 47WV30MS 47WV30BR 47WV30BS 47WV30-BAAM 47WV30-BAAL 47WV50BR 47WV50MS 47LV35A
More informationCNC Router Parts Plasma Software Setup and Usage Guide
Plasma Software Setup and Software Setup for the TMC3in1 Mach3 If your control PC is not currently set up with Mach3 software, follow the CNC Router Parts CNC Software Setup Guide to install the correct
More informationGEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor
DiSEqC 1.2 Motorized H-H H Motor GEOSATpro GS120 Compatible with DiSEqC 1.2 & USALS Receivers Adjustable Hardware Limiters for 140 Degree Coverage Goto X Preprogrammed for North American Satellites LED
More informationClassroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5
Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Lecture Capture Setup... 6 Pause and Resume... 6 Considerations... 6 Video Conferencing Setup... 7 Camera Control... 8 Preview
More informationXRD-DSC Sample Alignment (BB) Part
XRD-DSC Sample Alignment (BB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 5 2. Sample alignment sequence...9
More informationInstruction manual. KUZMA 4POINT 14 inch TONEARM Serial Number:
Instruction manual KUZMA 4POINT 14 inch TONEARM Serial Number:.. 2016-09 1 KUZMA LTD INSTRUCTION MANUAL FOR 4POINT 14 tonearm The 4POINT 14 tonearm is a very precisely engineered piece of equipment, however,
More informationMTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement
A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and
More informationJ.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL Stewartsigns.com
DataMax INDOOR LED MESSAGE CENTER OWNER S MANUAL QUICK START J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL 34232 800-237-3928 Stewartsigns.com J.M. Stewart Corporation Indoor LED Message
More informationUsing AFM Phase Lag Data to Identify Microconstituents with Varying Values of Elastic Modulus
Using Data to Identify Microconstituents with Varying Values of Elastic Modulus D.N. Leonard*, A.D. Batchelor**, P.E. Russell*,** *Dept. of Material Science and Engineering, Box 7531, North Carolina State
More informationHRXRD Analysis of Epitaxial Thin Films
HRXRD Analysis of Epitaxial Thin Films on the Rigaku Smartlab Multipurpose Diffractometer Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT For assistance in the X-ray lab, contact
More informationKVANT. Spectrum 1,6W RGB laser system. Product datasheet
KVANT Spectrum 1,6W RGB laser system Product datasheet SPECTRUM series professional full colour laser system 1. DESCRIPTION Kvant Spectrum is professional laser system with wide range of use, mainly for
More informationFOSC-600 C and D I N S T A L L A T I O N I N S T R U C T I O N
FOSC-600 C and D I N S T A L L A T I O N I N S T R U C T I O N In-line and butt version Cold applied re-usable fiber optic closure Contents 1 Introduction 1.1 Product description 1.2 Capacity 2 General
More informationnanoir2-s TM System User s Manual Part # Issued August 2016 Page 1 of 52
nanoir2-s TM System User s Manual Part #00-0026-03 Issued August 2016 2016 by Anasys Instruments Inc, 325 Chapala St., Santa Barbara, CA 93101 Page 1 of 52 Table of contents Anasys Instruments Contact
More informationSoftware manual. ChipScan-Scanner 3.0
Software manual ChipScan-Scanner 3.0 Copyright c Langer EMV-Technik GmbH 2016 Version 1.10 Notice No part of this manual may be reproduced in any form or by any means (including electronic storage and
More informationB2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006
B2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006 The B 2 Spice A/D software allows for the simulation of digital, analog, and hybrid circuits. CPE 169, however, is only concerned with the
More informationFully ly Automaticti. Motorised Satellite t TV System. User s manual. ver 3.0.
ver 3.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au
More informationJ R Sky, Inc. Cross-Modulation Distortion Analyzer
J R Sky, Inc. Cross-Modulation Distortion Analyzer J R Sky, Inc. 505 Evening Star Lane Bozeman, Montana 59715 USA Tel: +1.406-582-8154 email: nuoptix@jrsky.com web: www.jrsky.com revised: November 29,
More informationInstructions for Use: Video Inspection Scope with Display
Instructions for Use: Video Inspection Scope with Display Brand Name of Product Video Inspection Scope Generic Name of Product Video Inspection Scope Product Code Number(s) FIS-004 Intended Use For visually
More informationOperating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA
Operating Manual Automated Gear Apollo Design Technology, Inc. 4130 Fourier Drive Fort Wayne, IN 46818 USA PH: +01(260)497-9191 FX: +01(260)497-9192 www.apollodesign.net 11-25-09 5-6 POWERING UP THE RIGHT
More informationUser Manual VM700T Video Measurement Set Option 30 Component Measurements
User Manual VM700T Video Measurement Set Option 30 Component Measurements 070-9654-01 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - FAX 781.665.0780 - TestEquipmentDepot.com
More informationLUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY
Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518
More informationDave Jones Design Phone: (607) Lake St., Owego, NY USA
Manual v1.00a June 1, 2016 for firmware vers. 2.00 Dave Jones Design Phone: (607) 687-5740 34 Lake St., Owego, NY 13827 USA www.jonesvideo.com O Tool Plus - User Manual Main mode NOTE: New modules are
More informationUsing the Agilent for Single Crystal Work
Using the Agilent for Single Crystal Work Generally, the program to access the Agilent, CrysalisPro, will be open. If not, you can start the program using the shortcut on the desktop. There is no password
More information1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at
JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence
More informationPreface... Specification...14
Contents for V680L Operating Manual Preface......3 Important safety tips...3 Accessories of product...4 Icons of product...5 Installation support...7 Description for operating steps... 8 Description for
More informationWorking with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I
Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I Prepared by: Kyle Botteon Questions? kyle.botteon@psu.edu 2 Background Information Recall that oscilloscopes (scopes)
More informationCharacterization of Diamond Samples CHESS Run Summer 2013
Characterization of Diamond Samples CHESS Run Summer 2013 5/22 5/28/2013 Characterization of Diamond Samples CHESS Run Summer 2013 Useful information Goals for this run Conditions of the experiment 1.
More informationManual placement system MPL3100. for BGA, CSP and Fine-Pitch components
Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision
More informationSubj: General MTI2000 and Piezo Task Setup and Operation. High-Voltage Displacement Meter (HDVM) Configuration.
6 March 2013 Radiant Technologies, Inc. 2835D Pan American Freeway NE Albuquerque, NM 87107 Tel: 505-842-8007 Fax: 505-842-0366 e-mail: radiant@ferrodevices.com From: Scott P. Chapman Radiant Technologies,
More informationInstallation and User Guide 458/CTR8 8-Channel Ballast Controller Module
Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module Helvar Data is subject to change without notice. www.helvar.com i Contents Section Page Introduction 1 Installation 2 1. Attach
More informationBooya16 SDR Datasheet
Booya16 SDR Radio Receiver Description The Booya16 SDR radio receiver samples RF signals at 16MHz with 14 bits and streams the sampled signal into PC memory continuously in real time. The Booya software
More informationPlease feel free to download the Demo application software from analogarts.com to help you follow this seminar.
Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a
More informationHow to Optimize Ad-Detective
How to Optimize Ad-Detective Ad-Detective technology is based upon black level detection. There are several important criteria to consider: 1. Does the video have black frames to detect? Are there any
More informationQuick Start. RSHS1000 Series Handheld Digital Oscilloscope
Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products
More informationPreface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT.
Preface The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Please note: Some components described in this manual may be optional.
More informationIndiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope
Indiana Center for Biological Microscopy Zeiss LSM-510 510-UV Confocal Microscope Microscope and the Attached Accessories Transmission Detector Halogen Lamp House Condenser Eyepiece Stage Scanning and
More informationOM2000N INSTALLATION MANUAL
OM2000N INSTALLATION MANUAL 2 1 Figure A 1 2 Laser Beam Output Window Power Cable 821001342 (Rev. B) DESCRIPTION The OM2000N oscillating mirror is an accessory for the 2000N family laser scanners: DS2100N,
More information15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833
15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor
More information26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703
26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor to automatically
More informationElectrical connection
Color sensors Dimensioned drawing en 02-2013/01 50121262 068-14515 12mm 32mm 10-30 V DC 500 Hz Scanner for color detection Simultaneous selection of up to 3 colors Detection independent of distance Teach-in
More informationA-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0
A-ATF-200-11(1) PictureGear Pocket Operating Instructions Version 2.0 Introduction PictureGear Pocket What is PictureGear Pocket? What is PictureGear Pocket? PictureGear Pocket is a picture album application
More informationFully ly Automaticti. Motorised Satellite t TV System. User s manual REV
REV. 1.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au
More informationivw-ud322 / ivw-ud322f
ivw-ud322 / ivw-ud322f Video Wall Controller Supports 2 x 2, 2 x 1, 3 x 1, 1 x 3, 4 x 1 & 1 x 4 Video Wall Array User Manual Rev. 1.01 i Notice Thank you for choosing inds products! This user manual provides
More information