AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

Size: px
Start display at page:

Download "AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)"

Transcription

1 AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop icon 3. Select the desired mode of operation by choosing the appropriate workspace in the File/Open Workspace menu. The workspaces are found in the directory D:V614r1/Workspaces. The following workspaces are good starting points and can be used for most samples: Tapping Mode: smif_default_tapping_d.wks Contact Mode: smif_default_contact_d.wks 4. Click on Scan in the Workspace window to connect the software to the instrument 5. Mount the appropriate AFM tip into the cantilever holder. Be sure that the tip is in contact with the back end and left side of the groove per the following picture. TESP Tip properly positioned in cantilever holder prior to clamping TESP Tip properly positioned in cantilever holder after clamping You may use your own AFM tip, or purchase one of the following from SMIF: Tapping Mode: TESP silicon probe Contact Mode: ORC8-10 silicon nitride probe 6. Mount the cantilever holder onto the end of the scanner head a. Tighten the set screw to unclamp the scanner head from the dove tail and remove the scanner head from the cradle. Be sure not to pull on the cables that connect the scanner head to the electronics box b. Mount the cantilever holder by gently fitting it onto the pins at the end of the scanner head. DO NOT use excessive force as this may damage the scanner head c. Replace the scanner head in the cradle and loosen the set screw to clamp the head assembly in the dove tail 7. Align the laser a. Turn the illumination off by entering a 0 in the Illumination box of the Other controls menu in the Scan ParmList window. b. Place the blank side of a business card face up on the granite stage so that the laser is visible on the card c. Open the Meter window by clicking on Meter in the Workspace window Page 1 of 6 Revision 7

2 d. Using the two knobs at the top of the scanner head, follow the diagram below to adjust the laser position so that the laser is visually aligned with is the tip of the cantilever on the AFM tip. 1. Move the laser in the x direction to find AFM tip edge. Then move slightly back from the edge. 2. Move the laser in the y direction to find the cantilever. 3. Move the laser in the x direction to the right to find the cantilever edge. Now make fine adjustments to the x and y laser position knobs to maximize the signal sum in the meter window as shown below Red dot centered for detector alignment Signal sum maximized for laser alignment e. Remove the business card from the granite stage 8. Align the photodetector a. Use the two knobs on the left side of the scanner head to adjust the photodetector position 1. Coarse adjustment: Center the position of the red dot in the window on the scanner head 2. Fine adjustment: Center the position of the red dot in the cross-hairs of the meter display screen as shown above Tapping Mode: Set the Vertical Deflection to 0.0 volts ( 0.5V) Contact Mode: Set the Vertical Deflection to about 2.0 Volts b. Close the Meter window Page 2 of 6 Revision 7

3 9. Load sample a. Place sample on stage b. Select Navigate window and move the Z-position up using the Z-motor up arrow button in the Navigate window. (This is a precaution to prevent the sample from hitting the head assembly) c. Increase illumination from zero to 100 d. Move the sample under the probe using the X-Y stage movement arrows in the Navigate window. 10. Locate and focus on tip a. Click on Locate Tip in the Navigate window b. Zoom all the way out by clicking on Zoom Out c. Center the tip of the cantilever under the cross hairs using the two adjustment knobs located to the left of the optical objective on the microscope d. Focus on the tip end of the cantilever using the Optics arrows keys in the Navigate window e. Click on Zoom in to zoom all the way in or use the up arrow to zoom partially in on the tip and re-focus and re-center the tip if necessary. f. Click OK (The software will automatically change to the focus on sample surface menu) 11. Focus on the sample surface a. Zoom out to the maximum position to get the largest field of view possible b. Use the Z-motor down arrow button in the Navigate window to visually close the gap between the sample and the scanner head to about 2-3mm. c. Now look at image on screen and keep adjusting the Z-motor position until the sample surface is in focus. If the sample is smooth and has no surface features to focus on, then change the Focus On setting to Tip Reflection and adjust the Z-position until the reflection of the AFM tip is in focus. Page 3 of 6 Revision 7

4 d. Move to the desired measurement area of the sample under the cross hairs using the X-Y arrow keys in the Navigate window e. Zoom in on the area of interest and fine focus using the Z-position if necessary 12. Tune Cantilever (Tapping Mode AFM only) a. Click on the Tuning Fork Icon b. Set the Start and End frequency to bracket the resonant frequency value given on the box that the cantilever probe came from. (Typically 0kHz 500kHz for TESP silicon probes) c. Set the target amplitude to 2 volts d. Set the peak offset to 5% e. Click on Auto Tune button f. Note: If the tool does not auto tune properly and the probe tip is in good condition, try resetting the controller communication connection by selecting Tools/Reset Controller and then repeating the auto tune. g. When completed, click the Exit button 13. Set Initial Scan Parameters a. In the Scan Controls panel, enter the following settings: 1. Scan Size: Set to desired measurement area (< 70 m) 2. Aspect Ratio 1:1 3. X and Y offsets: 0 4. Scan angle: 0 5. Scan rate: Tapping mode: 1 Hz Contact mode: 2 Hz 6. Samples/line: 256, 512, or Lines: 256, 512, or Slow scan enabled 9. XY closed loop Lower setting: Lower resolution, faster speed Higher setting: Higher resolution, slower speed Off for scan size of < 1 m On for scan size of > 1 m b. For Tapping Mode AFM, in the Feedback Controls Panel, enter the following settings: 1. Integral Gain: Proportional Gain: Leave other settings alone c. For Contact Mode AFM, in the Feedback Controls Panel, enter the following settings: 1. Setpoint: 0 volts (2 volts difference from value set in Step 8) 2. Integral Gain: Proportional Gain: Close the instrument cover 15. Click on the Engage Icon Engage Withdraw Page 4 of 6 Revision 7

5 16. Optimize Scan Parameters a. Tapping Mode AFM 1. If the line in the Z-Center position moves far to the extended end (indicator turns red), the tip has false engaged. Disengage the probe tip and execute the engage command again 2. Check to see if Trace and Retrace are tracking each other well. If they are tracking, the lines should look the same, but they will not necessarily exactly overlap each other. 3. If they are not tracking well, adjust the Setpoint and/or Gains, to improve the tracking (See training notebook pages 36-41). It is often easier to disable the slow scan axis while making adjustments, so that the scanned feature stays in the field of view. Typical adjustments are as follows: Lower the setpoint to improve tracking between the trace and re-trace Lower the gains to reduce the noise b. Contact Mode AFM 1. Once engaged increase the Setpoint by using the right arrow key and observe the Z-Center Position. If the line in the Z-Center position moves far to the extended end (indicator turns red), the tip has false engaged. Disengage the probe tip, increase the setpoint by +2 volts and execute the engage command again. If the Z-Center position does not change greatly, you are on the sample surface and can go to the next step. 2. Check to see if Trace and Retrace are tracking each other well. If they are tracking, the lines should look the same, but they will not necessarily exactly overlap each other. 3. If they are not tracking well, adjust the Setpoint, Gains, and/or Scan Rate to improve the tracking (See training notebook pages and 36-41). It is often easier to disable the slow scan axis while making adjustments, so that the scanned feature stays in the field of view. 4. Remember that in Contact Mode, increasing the Setpoint increases the force on the sample. Minimizing the imaging force is recommended to prevent damage to the sample and the probe tip. 17. Set desired Scan Size, Scan Angle, and Offsets Page 5 of 6 Revision 7

6 Image Capturing 1. Choose Capture Filename from the RealTime pull down menu 2. Choose your directory to store the image, and enter a filename for the image. Note: directories must be created in Windows outside of the AFM software. Click OK. 3. Click on the camera icon to begin capturing an image. The image will begin capturing at the beginning of a scan frame. To move the probe to the beginning of a scan frame, click on the frame up or frame down icon 4. When capturing, the status bar at the bottom of the screen will indicate Capture: On, and when the capture is completed, it will indicate Capture: Done When you are finished: 1. Click on the Withdraw icon to stop scanning Engage Withdraw 2. Open the instrument cover 3. Select the Navigate window a. Raise the Probe head away from the sample using the Z-motor up arrow in the Navigate window b. Move the stage to the load position (completely out from under the scanner head) using the X-Y stage arrows in the Navigate window 4. Unload your sample 5. Remove the cantilever holder from the end of the scanner head and unmount your probe (Store your probe in your personal Gel-Pak storage box) 6. Close the instrument cover 7. If you wish to save changes to the workspace, Choose File/Save Workspace As and give the file a new workspace name. DO NOT save changes to the SMIF default workspaces 8. Exit Nanoscope software 9. Logout of the Log Usage system on the SMIF web site Page 6 of 6 Revision 7

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

AFM Standard Operating Procedure

AFM Standard Operating Procedure 2013 AFM Standard Operating Procedure Karen Gaskell, David Ramsdell Surface Analysis Centre Department of Chemistry and Biochemistry University of Maryland 1/1/2013 Content Page 1 Hardware 2 1.1 MultiMode

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

Standard AFM Modes User s Manual

Standard AFM Modes User s Manual Standard AFM Modes User s Manual Part #00-0018-01 Issued March 2014 2014 by Anasys Instruments Inc, 325 Chapala St, Santa Barbara, CA 93101 Page 1 of 29 Table of contents Chapter 1. AFM Theory 3 1.1 Detection

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment Name: Atomic Force Microscope Badger name: afm Revisionist Paul Kimani Model: Dimension 3000 Date: October 1, 2013 Location: Bay 1 A. Description The Dimension 3000 consists of a rigid stage

More information

Magnetic Force Microscope (MFM)

Magnetic Force Microscope (MFM) Magnetic Force Microscope (MFM) [Digital Instruments NanoScopeIII] magnetic and topographic resolution ~20 nm [at least several hours/sample (1 cm 2 polished single grains, thin sections, or thin films)]

More information

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact,

More information

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core 1 Turn On the laser power 2 Open enclosure: - lift the door latch and open the enclosure door. 3 2 1 1.Unlock scanner: Lift the lever to the

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 1.5, November 2013 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 2, September 2017 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

B-AFM. v East 33rd St., Signal Hill, CA (888)

B-AFM. v East 33rd St., Signal Hill, CA (888) B-AFM The B-AFM is a basic AFM that provides routine scanning. Ideal for scientists and educators, the B-AFM is capable of creating high-resolution topography images of nanostructures in standard scanning

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

Scanning Probe Microscope Training. Wenhui Pang

Scanning Probe Microscope Training. Wenhui Pang Scanning Probe Microscope Training Wenhui Pang Background - Comparison of AFM with Other Imaging Modalities Optical Microscopy SEM TEM AFM Resolution XY 200 nm 2 nm 0.1 nm 1 nm Z 500 nm N/A N/A 0.1 nm

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE ? 8 LCD Screen? 10.4 LCD Screen LCD Zoom Microscope Instruction Manual Please read the Instruction Manual carefully before installation and keep it for future

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL]

[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL] [AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL] VER: 2.0 TABLE OF CONTENT 1. INTRODUCTION...- 4-1.1 THEORY OF OPERATION...- 4-1.1.1 Principle of Atomic Force Microscope (AFM)... - 4-2. INSTALLATION &

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

SIMET AVIKO D INSTRUCTION MANUAL SORTING Solutions, Ltd.

SIMET AVIKO D INSTRUCTION MANUAL SORTING Solutions, Ltd. SIMET AVIKO D INSTRUCTION MANUAL 1870 SORTING Solutions, Ltd. 1. TABLE OF CONTENTS 1. TABLE OF CONTENTS...1 2. INTRODUCTION...2 2.1. Application...2 2.2. Operating Conditions...2 2.3. Electro - Optical

More information

12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492

12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492 12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor to automatically

More information

Do not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company.

Do not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company. PLEASE, READ Notice: Copyright: A delivery set of your device may differ from the set described in the given manual. For more detailed information refer to the specification of your contract. No part of

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

SX7. Saga 7" Super Bright HDMI/3G-SDI Field Monitor with 3D-LUTs. Quick Start Guide. What s Included CHECKED BY

SX7. Saga 7 Super Bright HDMI/3G-SDI Field Monitor with 3D-LUTs. Quick Start Guide. What s Included CHECKED BY SX7 Quick Start Guide Saga 7" Super Bright HDMI/3G-SDI Field Monitor with 3D-LUTs What s Included 1 x Saga X7 Monitor 1 x V-Mount Plate (Attached) 1 x Mini-XLR to P-TAP Cable 1 x Dual Sony L Battery Adapter

More information

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished

More information

imso-104 Manual Revised August 5, 2011

imso-104 Manual Revised August 5, 2011 imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel

More information

9070 Smart Vibration Meter Instruction Manual

9070 Smart Vibration Meter Instruction Manual 9070 Smart Vibration Meter Instruction Manual Overall machine and bearing conditions: vibration values are displayed with color coded alarm levels for ISO values and Bearing Damage (BDU). Easy vibration

More information

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents: Compensation look-up-table (LUT) and How to Generate Contents: Description Background theory Basic LUT pg 2 Creating a LUT pg 3 Using the LUT pg 7 Comment pg 9 The compensation look-up-table (LUT) contains

More information

AX20. Atlas 19.5" 3G-SDI/HDMI Field and Studio Monitor with 3D LUTs & Scopes. Quick Start Guide. What s Included CHECKED BY

AX20. Atlas 19.5 3G-SDI/HDMI Field and Studio Monitor with 3D LUTs & Scopes. Quick Start Guide. What s Included CHECKED BY AX20 Quick Start Guide Atlas 19.5" 3G-SDI/HDMI Field and Studio Monitor with 3D LUTs & Scopes What s Included 1 x Atlas 19.5" Monitor 1 x AC Adapter 1 x Sunhood CHECKED BY AX20 FRONT 1920 x 1080 19.5 inch

More information

Preparing for remote data collection at NE-CAT

Preparing for remote data collection at NE-CAT Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person

More information

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It

More information

F7000NV ROBOT VISION OPERATING MANUAL

F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting

More information

USER MANUEL. SNIPE 2 Ref R13

USER MANUEL. SNIPE 2 Ref R13 USER MANUEL SNIPE 2 Ref. 0141317R13 Contents 1. General Information 1-1. Introduction 1-2. Proper use and operation 1-3. Safety notes......... 2 3 3 2. Contents 2-1. Accessory included 2-2. Name of parts......

More information

PSC300 Operation Manual

PSC300 Operation Manual PSC300 Operation Manual Version 9.10 General information Prior to any attempt to operate this Columbia PSC 300, operator should read and understand the complete operation of the cubing system. It is very

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

Fluke 190-Series II Firmware Upgrade V11.44

Fluke 190-Series II Firmware Upgrade V11.44 Fluke 190-Series II Firmware Upgrade V11.44 Requirements 1. Fluke 190- Series II ScopeMeter with firmware prior to V11.44 2. Supported models are: 190-102, 190-104, 190-062, 190-202, 190-204, 190-502,

More information

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject

More information

First Question: Camera head. Lighting unit. Shooting stage

First Question: Camera head. Lighting unit. Shooting stage Elmo P30 Visualiser First Question: Q. Is everyone familiar with exactly what a visualiser is? A. A visualiser is effectively a camera on an arm, usually with a shooting stage and its own lighting source.

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

Chapter 4. Dish Antenna Installation. Installing a DISH 500 Antenna. Finding the Satellites

Chapter 4. Dish Antenna Installation. Installing a DISH 500 Antenna. Finding the Satellites These instructions guide you through the installation of a satellite system which includes your receiver (included with this manual), and a DISH Pro DISH 500 antenna system that can be identified by the

More information

Three Axis Digital Readout System

Three Axis Digital Readout System NEWALL MEASUREMENT SYSTEMS C80 Three Axis Digital Readout System CONTENTS 2 SPECIFICATIONS 3 CONNECTIONS 4 MOUNTING 4 Arm Mounting (Non-adjustable) 4 Arm Mounting (Adjustable) 5 Face Mounting (Adjustable)

More information

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

NanoGiant Oscilloscope/Function-Generator Program. Getting Started Getting Started Page 1 of 17 NanoGiant Oscilloscope/Function-Generator Program Getting Started This NanoGiant Oscilloscope program gives you a small impression of the capabilities of the NanoGiant multi-purpose

More information

Nanonis STM Simulator Tutorial

Nanonis STM Simulator Tutorial Nanonis STM Simulator Tutorial Software Version 4 Manual Version 4.0 Contents Introduction... 4 Minimum System Requirements and Installation... 5 Getting Started... 6 Session Directories... 6 Online Help...

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA The JS007WQK is loaded with userfriendly features and is ideal for use in heavy duty vehicles.

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

imso-104 Manual Revised July 19, 2012

imso-104 Manual Revised July 19, 2012 imso-104 Manual Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel 2.21 On / Off 2.22

More information

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors. Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only

More information

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303)

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303) Spectroscopy Module Vescent Photonics, Inc. www.vescentphotonics.com 4865 E. 41 st Ave Denver, CO 80216 Phone: (303)-296-6766 Fax: (303)-296-6783 General Warnings and Cautions The following general warnings

More information

Video wall Precautions when moving and handling. Precaution. Please adhere to the warning signs printed on the box.

Video wall Precautions when moving and handling. Precaution. Please adhere to the warning signs printed on the box. Video wall Precaution 4-1. Precautions when moving and handling Please adhere to the warning signs printed on the box. Model: 47WV30MS 47WV30BR 47WV30BS 47WV30-BAAM 47WV30-BAAL 47WV50BR 47WV50MS 47LV35A

More information

CNC Router Parts Plasma Software Setup and Usage Guide

CNC Router Parts Plasma Software Setup and Usage Guide Plasma Software Setup and Software Setup for the TMC3in1 Mach3 If your control PC is not currently set up with Mach3 software, follow the CNC Router Parts CNC Software Setup Guide to install the correct

More information

GEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor

GEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor DiSEqC 1.2 Motorized H-H H Motor GEOSATpro GS120 Compatible with DiSEqC 1.2 & USALS Receivers Adjustable Hardware Limiters for 140 Degree Coverage Goto X Preprogrammed for North American Satellites LED

More information

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Lecture Capture Setup... 6 Pause and Resume... 6 Considerations... 6 Video Conferencing Setup... 7 Camera Control... 8 Preview

More information

XRD-DSC Sample Alignment (BB) Part

XRD-DSC Sample Alignment (BB) Part XRD-DSC Sample Alignment (BB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 5 2. Sample alignment sequence...9

More information

Instruction manual. KUZMA 4POINT 14 inch TONEARM Serial Number:

Instruction manual. KUZMA 4POINT 14 inch TONEARM Serial Number: Instruction manual KUZMA 4POINT 14 inch TONEARM Serial Number:.. 2016-09 1 KUZMA LTD INSTRUCTION MANUAL FOR 4POINT 14 tonearm The 4POINT 14 tonearm is a very precisely engineered piece of equipment, however,

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL Stewartsigns.com

J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL Stewartsigns.com DataMax INDOOR LED MESSAGE CENTER OWNER S MANUAL QUICK START J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL 34232 800-237-3928 Stewartsigns.com J.M. Stewart Corporation Indoor LED Message

More information

Using AFM Phase Lag Data to Identify Microconstituents with Varying Values of Elastic Modulus

Using AFM Phase Lag Data to Identify Microconstituents with Varying Values of Elastic Modulus Using Data to Identify Microconstituents with Varying Values of Elastic Modulus D.N. Leonard*, A.D. Batchelor**, P.E. Russell*,** *Dept. of Material Science and Engineering, Box 7531, North Carolina State

More information

HRXRD Analysis of Epitaxial Thin Films

HRXRD Analysis of Epitaxial Thin Films HRXRD Analysis of Epitaxial Thin Films on the Rigaku Smartlab Multipurpose Diffractometer Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT For assistance in the X-ray lab, contact

More information

KVANT. Spectrum 1,6W RGB laser system. Product datasheet

KVANT. Spectrum 1,6W RGB laser system. Product datasheet KVANT Spectrum 1,6W RGB laser system Product datasheet SPECTRUM series professional full colour laser system 1. DESCRIPTION Kvant Spectrum is professional laser system with wide range of use, mainly for

More information

FOSC-600 C and D I N S T A L L A T I O N I N S T R U C T I O N

FOSC-600 C and D I N S T A L L A T I O N I N S T R U C T I O N FOSC-600 C and D I N S T A L L A T I O N I N S T R U C T I O N In-line and butt version Cold applied re-usable fiber optic closure Contents 1 Introduction 1.1 Product description 1.2 Capacity 2 General

More information

nanoir2-s TM System User s Manual Part # Issued August 2016 Page 1 of 52

nanoir2-s TM System User s Manual Part # Issued August 2016 Page 1 of 52 nanoir2-s TM System User s Manual Part #00-0026-03 Issued August 2016 2016 by Anasys Instruments Inc, 325 Chapala St., Santa Barbara, CA 93101 Page 1 of 52 Table of contents Anasys Instruments Contact

More information

Software manual. ChipScan-Scanner 3.0

Software manual. ChipScan-Scanner 3.0 Software manual ChipScan-Scanner 3.0 Copyright c Langer EMV-Technik GmbH 2016 Version 1.10 Notice No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

B2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006

B2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006 B2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006 The B 2 Spice A/D software allows for the simulation of digital, analog, and hybrid circuits. CPE 169, however, is only concerned with the

More information

Fully ly Automaticti. Motorised Satellite t TV System. User s manual. ver 3.0.

Fully ly Automaticti. Motorised Satellite t TV System. User s manual. ver 3.0. ver 3.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au

More information

J R Sky, Inc. Cross-Modulation Distortion Analyzer

J R Sky, Inc. Cross-Modulation Distortion Analyzer J R Sky, Inc. Cross-Modulation Distortion Analyzer J R Sky, Inc. 505 Evening Star Lane Bozeman, Montana 59715 USA Tel: +1.406-582-8154 email: nuoptix@jrsky.com web: www.jrsky.com revised: November 29,

More information

Instructions for Use: Video Inspection Scope with Display

Instructions for Use: Video Inspection Scope with Display Instructions for Use: Video Inspection Scope with Display Brand Name of Product Video Inspection Scope Generic Name of Product Video Inspection Scope Product Code Number(s) FIS-004 Intended Use For visually

More information

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA Operating Manual Automated Gear Apollo Design Technology, Inc. 4130 Fourier Drive Fort Wayne, IN 46818 USA PH: +01(260)497-9191 FX: +01(260)497-9192 www.apollodesign.net 11-25-09 5-6 POWERING UP THE RIGHT

More information

User Manual VM700T Video Measurement Set Option 30 Component Measurements

User Manual VM700T Video Measurement Set Option 30 Component Measurements User Manual VM700T Video Measurement Set Option 30 Component Measurements 070-9654-01 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - FAX 781.665.0780 - TestEquipmentDepot.com

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

Dave Jones Design Phone: (607) Lake St., Owego, NY USA

Dave Jones Design Phone: (607) Lake St., Owego, NY USA Manual v1.00a June 1, 2016 for firmware vers. 2.00 Dave Jones Design Phone: (607) 687-5740 34 Lake St., Owego, NY 13827 USA www.jonesvideo.com O Tool Plus - User Manual Main mode NOTE: New modules are

More information

Using the Agilent for Single Crystal Work

Using the Agilent for Single Crystal Work Using the Agilent for Single Crystal Work Generally, the program to access the Agilent, CrysalisPro, will be open. If not, you can start the program using the shortcut on the desktop. There is no password

More information

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence

More information

Preface... Specification...14

Preface... Specification...14 Contents for V680L Operating Manual Preface......3 Important safety tips...3 Accessories of product...4 Icons of product...5 Installation support...7 Description for operating steps... 8 Description for

More information

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I Prepared by: Kyle Botteon Questions? kyle.botteon@psu.edu 2 Background Information Recall that oscilloscopes (scopes)

More information

Characterization of Diamond Samples CHESS Run Summer 2013

Characterization of Diamond Samples CHESS Run Summer 2013 Characterization of Diamond Samples CHESS Run Summer 2013 5/22 5/28/2013 Characterization of Diamond Samples CHESS Run Summer 2013 Useful information Goals for this run Conditions of the experiment 1.

More information

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision

More information

Subj: General MTI2000 and Piezo Task Setup and Operation. High-Voltage Displacement Meter (HDVM) Configuration.

Subj: General MTI2000 and Piezo Task Setup and Operation. High-Voltage Displacement Meter (HDVM) Configuration. 6 March 2013 Radiant Technologies, Inc. 2835D Pan American Freeway NE Albuquerque, NM 87107 Tel: 505-842-8007 Fax: 505-842-0366 e-mail: radiant@ferrodevices.com From: Scott P. Chapman Radiant Technologies,

More information

Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module

Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module Helvar Data is subject to change without notice. www.helvar.com i Contents Section Page Introduction 1 Installation 2 1. Attach

More information

Booya16 SDR Datasheet

Booya16 SDR Datasheet Booya16 SDR Radio Receiver Description The Booya16 SDR radio receiver samples RF signals at 16MHz with 14 bits and streams the sampled signal into PC memory continuously in real time. The Booya software

More information

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a

More information

How to Optimize Ad-Detective

How to Optimize Ad-Detective How to Optimize Ad-Detective Ad-Detective technology is based upon black level detection. There are several important criteria to consider: 1. Does the video have black frames to detect? Are there any

More information

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products

More information

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT.

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Preface The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT. Please note: Some components described in this manual may be optional.

More information

Indiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope

Indiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope Indiana Center for Biological Microscopy Zeiss LSM-510 510-UV Confocal Microscope Microscope and the Attached Accessories Transmission Detector Halogen Lamp House Condenser Eyepiece Stage Scanning and

More information

OM2000N INSTALLATION MANUAL

OM2000N INSTALLATION MANUAL OM2000N INSTALLATION MANUAL 2 1 Figure A 1 2 Laser Beam Output Window Power Cable 821001342 (Rev. B) DESCRIPTION The OM2000N oscillating mirror is an accessory for the 2000N family laser scanners: DS2100N,

More information

15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833

15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833 15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor

More information

26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703

26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703 26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor to automatically

More information

Electrical connection

Electrical connection Color sensors Dimensioned drawing en 02-2013/01 50121262 068-14515 12mm 32mm 10-30 V DC 500 Hz Scanner for color detection Simultaneous selection of up to 3 colors Detection independent of distance Teach-in

More information

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0 A-ATF-200-11(1) PictureGear Pocket Operating Instructions Version 2.0 Introduction PictureGear Pocket What is PictureGear Pocket? What is PictureGear Pocket? PictureGear Pocket is a picture album application

More information

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV REV. 1.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au

More information

ivw-ud322 / ivw-ud322f

ivw-ud322 / ivw-ud322f ivw-ud322 / ivw-ud322f Video Wall Controller Supports 2 x 2, 2 x 1, 3 x 1, 1 x 3, 4 x 1 & 1 x 4 Video Wall Array User Manual Rev. 1.01 i Notice Thank you for choosing inds products! This user manual provides

More information