FREQUENCY COUNTERS TO 18 GHZ USING THE DATUM FREQUENCY STANDARD

Size: px
Start display at page:

Download "FREQUENCY COUNTERS TO 18 GHZ USING THE DATUM FREQUENCY STANDARD"

Transcription

1 TECHNICAL MANUAL AF-166 INSTRUMENT CALIBRATION FREQUENCY COUNTERS TO 18 GHZ USING THE DATUM FREQUENCY STANDARD THIS PUBLICATION SUPERSEDES NAVAIR 17-20AF-166 DATED 1 FEBRUARY 2005 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR FICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO MEASUREMENT SCIENCE DEPARTMENT, CORONA DIVISION, NAVAL SURFACE WARFARE CENTER, P.O. BOX 5000, CORONA, CA AF-166 DESTRUCTION NOTICE - FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD THAT WILL PREVENT DISCLOSURE CONTENTS OR RECONSTRUCTION THE DOCUMENT. PUBLISHED BY DIRECTION COMMANDER NAVAL AIR SYSTEMS COMMAND 1 SEPTEMBER 2005

2 LIST EFFECTIVE PAGES The purpose of this list is to identify the pages in this document affected by any technical content changes made since the previous release of the document. NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical line in the outer margin of the page. Page Page Page Page No. Revision* No. Revision* No. Revision* No. Revision* Title to A 0 69 to to to i to to to A-1 to A to to to *Zero in this column indicates an original page. R in this column indicates a revised page. **All pages not listed are blank pages. A

3 TABLE CONTENTS Section Page 1 Introduction and Description Equipment Requirements Preliminary Operations Calibration Process Period and Time Interval Tests Direct Voltage Tests Peak Voltage Tests... 9 Navy Calibration Checklist Appendix A: Representative Test Instruments... A-1 ILLUSTRATIONS Figure Page 1 Frequency Accuracy and Sensitivity Test Configuration Period and Time Interval Test Configuration Time Base Accuracy Test Configuration Direct Voltage Test Configuration... 8 TABLES Table Page 1 Calibration Description Equipment Requirements... 2 i

4 SECTION 1 INTRODUCTION AND DESCRIPTION 1.1 This procedure describes the calibration of Frequency Counters to 18 GHz using the Datum Frequency Standard. The instrument being calibrated is referred to herein as the TI (Test Instrument). 1.2 All comments concerning this procedure should be directed to the Measurement Science Department, Corona Division, Naval Surface Warfare Center, P.O. Box 5000, Corona, CA This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration, whether specifically tested for or not, should be corrected. 1.4 This procedure tests the TIs time base to the manufacturer s 1 year specification. 1.5 A list of representative frequency counters is given in Appendix A. TI Characteristics Table 1. Calibration Description Performance Specifications Test Method Frequency sensitivity Frequency measurement Period/time interval measurement Time base oscillator accuracy Tolerance: Tested to the specified limits. A power meter is used to ensure the specified power level into the TI input for cases in which the TI is either out-of-tolerance or does not make a measurement Tolerance: Tested to the specified limits. Tested to accuracy specifications for the TI time base based on a 1 year aging rate, temperature, and line voltage variations. Most TI accuracy specifications are time base accuracy ±1 count Tolerance: Tested to specified limits. In many cases, this is a functional check of channel 2 (B channel) to ensure operation of channel 2 for counters that do not have frequency measurement capability on channel 2. The tolerances on the checklist are generally looser than the manufacturer specifications Tested to the TI accuracy specification based on a 1 year oscillator aging rate Tested with a signal source. Tested with a frequency standard and a signal source during the sensitivity test. Tested with a square wave generated by a function generator. Tested with a frequency standard and a frequency counter. DC voltage measurement Tolerance: Tested to the specified limits Tested with a DC source. Peak voltage test Tolerance: Tested to the specified limits Tested with a function generator. 1

5 SECTION 2 EQUIPMENT REQUIREMENTS NOTES Minimum use specifications are the principal parameters required for performance of the calibration, and are included to assist in the selection of alternate equipment, which may be used at the discretion of the using laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable equipment must bear evidence of current calibration. The instruments utilized in this procedure were selected from those known to be available at Navy calibration facilities and the listing by make or model number carries no implication of preference, recommendation, or approval for use by other agencies. It is recognized that equivalent equipment produced by other manufacturers may be capable of equally satisfactory performance in this procedure. Item Table 2. Equipment Requirements Minimum Use Specifications Calibration Equipment 2.1 Frequency standard Frequency uncertainty: < ±1 x 10-9 Datum with or without satellite lock 2.2 Signal source Function generator Waveforms: Sine, Square Wave Square wave rise time: <20 nsec Frequency range: 1 Hz to 10 MHz Amplitude uncertainty: ±1% External frequency reference: 10 MHz Square wave asymmetry: <1% at 1 khz Microwave synthesizer Frequency range: 10 MHz to 18 GHz Amplitude uncertainty: ±1 db maximum External frequency reference: 10 MHz 2.3 Frequency counter Frequency range: 1 to 10 MHz External frequency reference: 10 MHz Display: 12 digits Gate time: Selectable from 100 msec to 500 msec Hewlett-Packard 33120AOPT001 Hewlett-Packard 83630AOPT001, 008, H53 Hewlett-Packard 53132AOPT010, Ω feedthrough termination DC resistance: 50 Ω ±1 Ω Tektronix or equivalent 2.5 BNC cable Type RG-58 cable Local supply 2.6 Coaxial cable Frequency range: 10 MHz to 18 GHz Length: 3 ft. maximum Insertion loss: 10 to 500 MHz: 0.05 db 1 to 3 GHz: 0.5 db 4 to 12 GHz: 1.0 db 13 to 18 GHz: 1.2 db Connectors: Type N male to Type N male 2.7 Coaxial adapters Frequency range: 0 to 18 GHz Used to adapt the coaxial cable type N male connectors to the various TI input connectors M/A-COM or equivalent Local supply 2

6 Item Minimum Use Specifications Calibration Equipment 2.8 Coaxial attenuator set Frequency range: 10 Hz to 10 MHz Attenuation: 3 db and 6 db nom Weinschel AS Power meter Frequency range: 10 MHz to 18 GHz Hewlett-Packard 437B 2.10 Power sensor Power sensor #1 Power range: -10 to -25 dbm Hewlett-Packard 8481A Power sensor #2 Power range: -25 to -40 dbm Hewlett-Packard 8481D 2.11 DC source Output: 10 mv to 1000 V Uncertainty: ±0.003% iv ±40 µv Fluke 5700AAN SECTION 3 PRELIMINARY OPERATIONS 3.1 Set all auxiliary equipment controls as necessary to avoid damage to the equipment and so that dangerous voltages will not be present on the output terminals when power switches are turned on. 3.2 Frequency Standard Setup If the frequency standard (item 2.1) is being used in the mode where it is acquiring and using data from GPS satellites, the instrument must be in such a mode for a minimum of 24 hours. Ensure that the frequency standard front panel LOCKED and TRACKING lights are illuminated If the frequency standard (item 2.1) is not locked and tracking satellites, turn the instrument on and ensure it is in Flywheel Mode (FW Mode). This mode uses the internal rubidium frequency oscillator of the frequency standard. The warm-up time is 30 minutes. 3.3 Turn all other power switches on, and allow a sufficient warm-up time for the equipment (refer to the applicable checklist for the TI warm-up time). 3.4 Perform only those tests as indicated on the checklist for the TI. SECTION 4 CALIBRATION PROCESS NOTE Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding. 4.1 FREQUENCY ACCURACY AND SENSITIVITY TESTS Set all TI channels to DC coupling. If the TI does not have selectable coupling for a channel, disregard this step. 3

7 4.1.2 Set all TI input attenuators, if any, for all TI input channels to the minimum setting. This may be X1 for some counters Place the signal sources in their default state, as applicable. NOTES The function generator, item 2.1, will be used for frequencies from 0 Hz to 10 MHz. The microwave synthesizer, item 2.2.2, will be used for frequencies from 10 MHz to 18.0 GHz. Use the BNC cable with the function generator and use the coaxial cable with the microwave synthesizer. For the Hewlett-Packard 33120A function generator, cycle the power to set it in its default state. For the Hewlett-Packard 83630A microwave synthesizer, press PRESET, to set it in its default state Connect a 10 MHz PUT from the frequency standard to the External Frequency Reference Input of the signal source as shown by the solid line in Figure MHz (Rear) 2.1 FREQUENCY STANDARD NOTE DO NOT CONNECT EQUIPMENT SHOWN BY DASHED LINES UNTIL INSTRUCTED. EXTERNAL FREQUENCY REFERENCE IN (Rear) 2.2 SIGNAL SOURCE PUT ATTEN TERM INPUT TI Figure 1. Frequency Accuracy and Sensitivity Test Configuration NOTE Both the function generator and the microwave synthesizer have 50 Ω outputs. If the TI input channel is 50 Ω, a feedthrough termination should not be used. If the TI input impedance is 1 MΩ, a 50 Ω feedthrough termination is required. If the TI input channel is selectable, then set the input impedance to 50 Ω and do not use a feedthrough termination, unless the TI checklist says otherwise If required, connect a 50 Ω feedthrough termination as shown by the dashed lines in Figure Connect the signal source output to the TI input channel specified on the checklist, using the attenuator and/or the termination if instructed. Set the function generator for sine wave and set the microwave synthesizer for a CW (continuous wave), as applicable If the checklist gives specific TI control settings, set the TI controls as indicated. 4

8 NOTE A power meter is used to verify the power level into the TI for the cases in which the TI does not meet the sensitivity tolerance due to uncertainties associated with the input power level when using the microwave synthesizer. For example, a TI may have a -20 dbm sensitivity specification at 18 GHz. The microwave source is set for -20 dbm, but the source has a power level uncertainty and the microwave cable has insertion loss. Since the actual power level may be less than the value specified for the TI (-20 dbm in this example), a power meter is used to ensure the specified power level into the TI input for cases in which the TI is either out-of-tolerance or does not make a measurement Set the signal source frequency and output level controls for each of the settings indicated in column 2 of the checklist. Refer to the checklist and insert an attenuator by the dashed lines in Figure 1 if required on the TI checklist. At each test point, verify that the TI indication is within the tolerance limits specified on the checklist. If the TI is within tolerance, skip to step ; otherwise, proceed with step Some TIs have an adjustable sensitivity knob that may need to be adjusted to the maximum setting. NOTES It is not necessary that the TI display have exactly the same units or display as the values on the checklist. For example, the checklist may list 9 to 11 Hz for the TI tolerance, but the TI displays khz. The TI is within tolerance. The function generator is used for frequencies from 0 to 10 MHz. The microwave synthesizer is used for frequencies above 10 MHz. Use the appropriate signal source dependent on the test frequency indicated on the checklist Measure the output at the end of the microwave cable, at the TI interface, with the power meter and appropriate power sensor. Adjust the microwave synthesizer power as necessary to obtain the checklist power level specified in step 4.1.8, as measured on the power meter. With the power level now established using the power meter, verify that the TI indication is within the tolerance limits specified on the checklist for step Repeat steps through for each specified TI channel, TI setup instructions, and measurement points indicated on the checklist step If the TI is out-of-tolerance and does not have a time base output, maintenance is required. If the TI is out-of-tolerance and has a time base output, skip to Section 4.3 to test and adjust the TI time base. Return to the beginning of Section 4.1 after the adjustment. 4.2 PERIOD AND TIME INTERVAL TESTS If period and time interval tests are not required per the checklist, skip this section. NOTE The TI input channels may either be Channels A, B, C, etc or Channels 1, 2, 3, etc. The instructions are for Channels A, B, C, etc.; but are meant to be universal for counters with Channels 1, 2, 3, etc Connect a 10 MHz PUT from the frequency standard to the External Frequency Reference Input of the signal source as shown by the solid line in Figure 2. Only use a 50 Ω termination, shown by dashed lines, at the input of the TI if the TI input is high impedance (1 MΩ). 5

9 10 MHz (Rear) 2.1 FREQUENCY STANDARD EXTERNAL FREQUENCY REFERENCE IN (Rear) FUNCTION GENERATOR PUT 2.4 TERM TI INPUT, CH A Figure 2. Period and Time Interval Test Configuration Set the function generator as follows: Frequency: 1 khz Waveform: Square Wave Amplitude: 200 mv rms Connect the function generator output to the TI input channel specified on the checklist, using the termination if instructed Set the TI to measure Period on Channel A. (The TI checklist may instruct to measure the period on a different TI Channel, and may give specific TI control settings.) Verify that the TI indication is within the specified tolerance limits on the checklist. If the checklist does not have any more tests in this section, skip to Section Set the TI to the following settings for a time interval measurement. (The following settings are applicable for most TIs; however, the checklist may have specific instructions for the TI.) TIME INTERVAL MEASUREMENT A to B COMMON A Channel A and B to DC coupling SLOPE for Channel A to Positive (+) SLOPE for Channel B to Negative (-) Ensure that the TI indication is approximately 500 microseconds, and record this value. If the display is fluctuating, choose the lowest number. The sum of this value and the value in step will equal the period (1 msec) of the square wave. 6

10 4.2.9 Set the TI as follows (the following settings are applicable for most TIs; however, the checklist may have specific instructions for the TI): SLOPE for Channel A to Negative (-) SLOPE for Channel B to Positive (+) Ensure that the TI indication is approximately 500 microseconds, and record this value. If the display is fluctuating, choose the highest number. The sum of this value and the value in step will equal the period (1 msec) of the square wave Add the values from steps and and verify that the sum is within the specified tolerance limits. The tolerance on the checklist includes count error and trigger error for the time interval measurements. This is why the tolerance for the period value is different from the tolerance value for the period in step Disconnect all equipment. 4.3 TIME BASE OSCILLATOR ACCURACY TEST If the time base oscillator accuracy test is not shown on the checklist it is not required, skip this section Connect the equipment as shown in Figure MHz (REAR) 2.1 FREQUENCY STANDARD EXTERNAL FREQUENCY REFERENCE IN (REAR) 2.3 FREQUENCY COUNTER CHANNEL 1 FREQUENCY REFERENCE PUT (REAR) TI Figure 3. Time Base Accuracy Test Configuration Set the standard frequency counter as follows: Default state (Hewlett-Packard 53132A counter: Cycle the Power) Frequency Measurement, Channel 1 (Hewlett-Packard 53132A: already selected in default state) Select 50 Ω for Channel 1 7

11 Select external reference (Hewlett-Packard 53132A: automatically senses and uses an external reference) Gate Time 0.5 seconds (Hewlett-Packard counter: Press GATE&EXTARM, press GATE&EXTARM again when GATE:TIME is displayed, use UP ARROW to increase gate time to.500 seconds, press ENTER, press RUN.) Verify that the standard frequency counter indication is within the TI checklist tolerance values If the TI meets the requirement of step 4.3.4, skip to step Using an insulated tool, adjust the oscillator to the value on the TI checklist, if possible. The gate time of the standard frequency counter may be increased for added resolution, if necessary Allow 24 hours for the oscillator to stabilize. Repeat steps through until the oscillator is within the checklist tolerance If any adjustments are performed, repeat Section 4.1. Repeat Section 4.2 if required, per the checklist and then continue to Section Disconnect all equipment. 4.4 DIRECT VOLTAGE TESTS WARNING VOLTAGES HAZARDOUS TO LIFE MAY BE PRESENT. USE EXTREME CAUTION. NOTES Manually set the TI to the required voltage ranges to hold the measurement in the specified range. When measurements must be performed in the TI auto mode because the TI is not equipped with a manual range selector, the calibration test measurement values have been chosen to avoid any measurement problem which may arise as a result of the TI automatic down-ranging and up-ranging characteristics. Use only the function and ranges that are applicable to the TI being calibrated If the direct voltage test is not required per the checklist or the instrument does not have a DC voltage measurement capability, skip this section Connect the equipment as shown in Figure 4, observing proper polarity DC SOURCE TI Figure 4. Direct Voltage Test Configuration 8

12 4.4.3 Set the TI controls to measure direct voltage, positive polarity, on the range indicated on the checklist Set the dc source controls for each of the nominal values listed on the checklist and verify that the TI indication is within the tolerance limits listed Set the dc source controls for minimum output Reverse the input leads to the TI for negative polarity Set the dc source output controls to indicate the absolute value of the nominal voltage listed for the negative polarity test. Verify that the TI indication is within the tolerance limits. (The TI indication will be a negative value only for TIs having an autopolarity capability.) Set the dc source output controls for minimum output (standby) and reverse the input leads to the TI for positive polarity Set the dc source controls to each calibration point indicated for each TI range listed on the checklist. At each setting, verify that the indication is within the tolerances listed Unless other measurements are to be performed, set all power switches to off and disconnect the equipment. 4.5 PEAK VOLTAGE TESTS If the peak voltage test is not required per the checklist or the instrument does not have a peak voltage measurement capability, skip to step Connect the function generator output to the TI input channel specified on the TI checklist. NOTE The function generator has a 50 Ω output impedance. If the TI input channel is 50 Ω, a feedthrough termination should not be used. If the TI input impedance is 1 MΩ, a 50 Ω feedthrough termination is required. If the TI input channel is selectable, then set the input impedance to 50 Ω and do not use a feedthrough termination, unless the TI checklist says otherwise Set the function generator for a sine wave output Set the TI to measure peak voltage. (The TI checklist may give specific TI control settings.) Set the function generator frequency and output level controls for each of the settings indicated in column 2 of the checklist. At each test point, verify that the TI indication is within the tolerance limits specified on the checklist Repeat steps and for each specified TI channel, TI setup instructions, and measurement points indicated on the checklist step Unless other measurements are to be performed, turn all power switches to off or standby, and disconnect the equipment from the TI. (10 blank) 9

13 TEST INST (S) AN/USM-207A or AN/USM-207B Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) FREQ A INPUT -- ck ( ) NA -- (use 50 Ω term) GATE TIME 1 SEC -- ck ( ) NA mv rms 10 Hz 9 to 11 Hz mv rms 10 khz to khz mv rms 100 khz to khz mv rms 1 MHz to 1, khz mv rms 10 MHz to MHz s Time Base 10 MHz 9,999,994 to10,000,006 MHz Time Base Adjustment 10 MHz 9,999, to10,000, Hz (12 blank) 11

14 TEST INST (S) Anadex CF500R and CF501R Counter/Timers PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour INPUT (use 50 Ω term) -- ck ( ) NA SECONDS -- ck ( ) NA -- DISPLAY TIME MAX -- ck ( ) NA -- RESET -- ck ( ) NA mv rms with 6 db attenuator 20 Hz 19 to 21 Hz RESET -- ck ( ) NA mv rms with 6 db attenuator 10 khz 9,998 to 10,002 Hz RESET -- ck ( ) NA mv rms with 6 db attenuator 200 khz to khz Time Base 100 khz 99,989 to 100,011 Hz Time Base Adjustment 100 khz 99, to 100, Hz (14 blank) 13

15 TEST INST (S): Anadex CF601R Counter/Timers PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES Unmarked rear panel input (use 50 Ω term) -- ck ( ) NA SECONDS X1 -- ck ( ) NA -- DISPLAY TIME MAX -- ck ( ) NA -- INPUT MULTIPLIER X1 -- ck ( ) NA -- Rear Panel RATE & EXT Switch F -- ck ( ) NA -- Rear Panel MEMORY Switch Off -- ck ( ) NA -- RESET -- ck ( ) NA mv rms with 6 db attenuator 20 Hz 19 to 21 Hz RESET -- ck ( ) NA mv rms with 6 db attenuator 10 khz 9,998.0 to 10,002.0 Hz RESET -- ck ( ) NA mv with 6 db attenuator 200 khz 199,979.0 to 200,021.0 Hz (16 blank) 15

16 TEST INST (S) Anadex CF QA2 Time Base Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour INPUT 1 (rear panel BNC) -- ck ( ) NA SECONDS INPUT 1A -- ck ( ) NA SECONDS INPUT 1B -- ck ( ) NA SECONDS INPUT 2A -- ck ( ) NA SECONDS INPUT 2B -- ck ( ) NA -- MULT 1A: 1 -- ck ( ) NA -- MULT 1B: 1 -- ck ( ) NA -- DISPLAY TIME MAX -- ck ( ) NA -- INPUT 1A -- ck ( ) NA -- RESET -- ck ( ) NA mv rms with 6 db attenuator 20 Hz 19 to 21 Hz mv rms with 6 db attenuator 10 khz 9,998.0 to 10,002.0 Hz mv rms with 6 db attenuator 200 khz 199,989.0 to 200,011.0 Hz INPUT 1B -- ck ( ) NA mv rms with 6 db attenuator 200 khz 199,989.0 to 200,011.0 Hz INPUT 2 (rear panel BNC) -- ck ( ) NA -- INPUT 2A -- ck ( ) NA -- RESET -- ck ( ) NA mv rms with 6 db attenuator 20 Hz 19 to 21 Hz mv rms with 6 db attenuator 10 khz 9,998.0 to 10,002.0 Hz mv rms with 6 db attenuator 200 khz 199,989.0 to 200,011.0 Hz Page 1 of 2 17

17 TEST INST (S) Anadex CF QA2 Time Base Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) INPUT 2B -- ck ( ) NA mv rms with 6 db attenuator 200 khz 199,989.0 to 200,011.0 Hz Page 2 of 2 18

18 TEST INST (S) Tektronix DC501 Electronic Counter P/I With or Without OPT02 PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour INPUT (use 50 Ω term) -- ck ( ) NA DISPLAY TIME 1 SEC -- ck ( ) NA mv rms 20 Hz to khz mv rms 1 MHz to 1, khz dbm 100 MHz to MHz Time Base 1 MHz 999,879 to 1,000,121 Hz Time Base Adjustment 1 MHz 999, to 1,000, Hz (20 blank) 19

19 TEST INST (S) Tektronix DC501OPT01 Digital Counter P/I With or Without OPT02 (OPT01: 1 x 10-7 /month time base) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour INPUT (use 50 Ω term) -- ck ( ) NA DISPLAY TIME 1 SEC -- ck ( ) NA mv rms 20 Hz to khz mv rms 1 MHz to 1, khz dbm 100 MHz to MHz Time Base 1 MHz 999,999 to 1,000,001 Hz Time Base Adjustment 1 MHz 999, to 1,000, Hz (22 blank) 21

20 TEST INST (S) Tektronix DC5010 Prog Universal Counter/Timer and DC510 Counter P/I PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CH A (no 50 Ω term) -- ck ( ) NA TERM 50 Ω -- ck ( ) NA -- FREQ A -- ck ( ) NA mv rms 9.9 to 10.1 Hz 1 25 mv rms 1 MHz khz to 1.000,006 MHz dbm 300 MHz ,499 to ,501 MHz 4.2 Period and Time Interval Tests Channel A -- ck ( ) NA PER A -- ck ( ) NA Period A 1,000 µsec µsec to msec Do not select TIME A to B -- ck ( ) NA -- Do not select COMMON A -- ck ( ) NA -- CH B slope doesn t matter -- ck ( ) NA -- Set WIDTH A -- ck ( ) NA Width CH A (Slope: +, -) -- ck ( ) NA Record Width CH A (Slope: -, +) -- ck ( ) NA Record Sum of and ,000 µsec to 1,001.0 µsec (24 blank) 23

21 TEST INST (S) Tektronix DC503 Counter/Timer P/I PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CH A (use 50 Ω term) -- ck ( ) NA FREQ A GATE TIME 1s -- ck ( ) NA mv rms 10 Hz 9.9 to 10.1 Hz mv rms 1 MHz khz to 1.000,011 MHz dbm 100 MHz to MHz 4.2 Period and Time Interval Tests CHANNEL B -- ck ( ) NA PER B -- ck ( ) NA -- 1 µsec clock rate -- ck ( ) NA Period B 1,000 µsec 998 to 1,002 µsec (26 blank) 25

22 TEST INST (S) Tektronix DC503OPT01 Electronic Counter P/I (OPT01: Time base temperature stability: ±5 x 10-7 /month) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CH A (use 50 Ω term) -- ck ( ) NA FREQ A GATE TIME 1s -- ck ( ) NA mv rms 10 Hz 9.9 to 10.1 Hz mv rms 1 MHz khz to 1, MHz dbm 100 MHz to MHz 4.2 Period and Time Interval Tests Period Channel B -- ck ( ) NA PERIOD B -- ck ( ) NA -- 1 µsec clock rate -- ck ( ) NA Period B 1,000 µsec 998 to 1,002 µsec (28 blank) 27

23 TEST INST (S) Tektronix DC503A Counter/Timer P/I PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CH A (use 50 Ω term) -- ck ( ) NA mv rms 10 Hz 9.9 to 10.1 Hz 1 20 mv rms 1 MHz khz to 1.000,002 MHz dbm 125 MHz to MHz 4.2 Period and Time Interval Tests CHANNEL B -- ck ( ) NA PER B -- ck ( ) NA -- 1 µsec CLOCK RATE -- ck ( ) NA Period B 1,000 µsec 998 to 1,002 µsec (30 blank) 29

24 TEST INST (S) Tektronix DC504 and DC504OPT01 Counter/Timer P/I (OPT01: 1 x 10-7 /month time base) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour (DC504) INPUT (use 50 Ω term) -- ck ( ) NA DISPLAY TIME 1 SEC -- ck ( ) NA mv rms 20 Hz to khz mv rms 10 MHz to MHz dbm 50 MHz to MHz (DC504OPT01) INPUT (use 50 Ω term) -- ck ( ) NA DISPLAY TIME 1 SEC -- ck ( ) NA mv rms 20 Hz to khz mv rms 10 MHz to MHz dbm 50 MHz to MHz (32 blank) 31

25 TEST INST (S) Tektronix DC504A Counter/Timer P/I PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour MΩ INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ to 10 MHz -- ck ( ) NA -- TIME/AVGS AUTO -- ck ( ) NA mv rms 20 Hz to khz 1 30 mv rms 1 MHz to khz mv rms 10 MHz to khz FUNCTION FREQ to 100 MHz -- ck ( ) NA dbm 10 MHz 9, to 10,000.1 khz dbm 100 MHz 99,999 to 100,001 khz Time Base 10 MHz 9,999,990 to 10,000,010 Hz Time Base Adjustment 10 MHz 9,999,999.5 to 10,000,000.5 Hz (only if necessary) (34 blank) 33

26 TEST INST (S) Tektronix DC505A and DC505AOPT01 Electronic Counters P/I (OPT01: 1 x 10-7 /month time base) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CH A (use 50 Ω term) -- ck ( ) NA CH A (for DC505A) -- ck ( ) NA -- FREQ A -- ck ( ) NA mv rms 10 Hz 9.9 to 10.1 Hz 1 50 mv rms 1 MHz to 1, khz dbm 250 MHz to MHz CH A (for DC505AOPT01) -- ck ( ) NA FREQ A -- ck ( ) NA mv rms 10 Hz 9.9 to 10.1 Hz 1 50 mv rms 1 MHz to 1, khz dbm 250 MHz to MHz (36 blank) 35

27 TEST INST (S) Tektronix DC508A Counter/Timer PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour DIRECT INPUT (no 50 Ω term) -- ck ( ) NA Ω -- ck ( ) NA -- RESOLUTION 1 Hz -- ck ( ) NA mv rms 10 Hz 9 to 11 Hz dbm 50 MHz ,949 to ,051 MHz dbm 100 MHz ,899 to ,101 MHz PRESCALE INPUT (no 50 Ω term) -- ck ( ) NA -- DISPLAY TIME 1 Hz -- ck ( ) NA dbm 100 MHz ,899 to ,101 MHz dbm 500 MHz ,499 to ,501 MHz DISPLAY TIME 10 Hz dbm 1.3 GHz ,69 to ,31 MHz (38 blank) 37

28 TEST INST (S) Tektronix DC509 and DC509OPT01 Counter/Timers P/I PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CHANNEL A (50 Ω term) -- ck ( ) NA FREQUENCY A -- ck ( ) NA -- AVERAGES AUTO -- ck ( ) NA -- All Pushbuttons Out -- ck ( ) NA -- AUTO LEVEL -- ck ( ) NA mv rms 10 Hz 9.9 to 10.1 Hz 1 20 mv rms 1 MHz ,0 khz to 1.000,002,0 MHz dbm 135 MHz ,86 to ,15 MHz 4.2 Period and Time Interval Tests Channel A -- ck ( ) NA Channel A, Level: AUTO -- ck ( ) NA -- Channel B, Level: AUTO -- ck ( ) NA -- AVERAGES ck ( ) NA Period A 1,000 µsec µsec to msec Do not select TIME A to B -- ck ( ) NA -- Do not select COMMON A -- ck ( ) NA -- CH B slope doesn t matter -- ck ( ) NA -- Set WIDTH A -- ck ( ) NA Width Channel A (slope: +,-) -- NA Record Width Channel A (slope: -,+) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec (40 blank) 39

29 TEST INST (S) Philips PM6670 and PM6671 Counter/Timer PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 15 minutes A INPUT (use 50 Ω term) -- ck ( ) NA FREQ A -- ck ( ) NA -- MEASURING TIME 1s -- ck ( ) NA mv rms with 6 db attenuator 10 Hz 9.9 to 10.1 Hz dbm 60 MHz ,63 to ,37 MHz dbm 120 MHz ,2 to ,8 MHz 4.2 Period and Time Interval Tests PERIOD A -- ck ( ) NA Period A 1,000 µsec to 1,000.1 µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Hz Time Base 10 MHz 9,999,940 to 10,000, Time Base Adjustment 10 MHz 9,999, to 10,000, (42 blank) 41

30 TEST INST (S) Philips PM6672 Counter/Timer PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time A INPUT (use 50 Ω term) -- ck ( ) NA FREQ A or C -- ck ( ) NA -- MEASURING TIME 1s -- ck ( ) NA mv rms with 6 db attenuator 10 Hz 9.9 to 10.1 Hz dbm 60 MHz ,63 to ,37 Hz dbm 120 MHz ,2 to ,8 MHz C INPUT (no 50 Ω term) -- ck ( ) NA MHz dbm 70 MHz ,57 to , dbm 500 MHz to dbm 1 GHz to 1.000,007 MHz GHz 4.2 Period and Time Interval Tests PERIOD A -- ck ( ) NA Period A 1,000 µsec to 1,000.1 µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Hz Time Base 10 MHz 9,999,940 to 10,000, Time Base Adjustment 10 MHz 9,999, to 10,000, (44 blank) 43

31 TEST INST (S) Philips PM6680 Electronic Counter With or Without OPTPM9621 (OPTPM9621: 1.3 GHz Input C PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (no 50 Ω term) -- ck ( ) NA INPUT 50 Ω -- ck ( ) NA MEASURING TIME 1 s -- ck ( ) NA mv 10 Hz 9.9 to 10.1 Hz dbm 50 MHz MHz ,750 to , dbm 110 MHz ,450 to , dbm 220 MHz ,900 to , B INPUT (no 50 Ω term) -- ck ( ) NA -- Set to measure FREQ B -- ck ( ) NA mv 10 Hz 9.9 to 10.1 Hz dbm 50 MHz MHz ,750 to , dbm 160 MHz ,200 to ,800 C INPUT (OPTPM9621 only) -- ck ( ) NA Set to measure FREQ C -- ck ( ) NA MHz 70 dbm 70 MHz ,650 to , dbm 500 MHz ,500 to ,500 GHz 1.3 dbm 1.3 GHz ,499 to ,501 Page 1 of 2 45

32 TEST INST (S) Philips PM6680 Electronic Counter With or Without OPTPM9621 (OPTPM9621: 1.3 GHz Input C) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 4.2 Period and Time Interval Tests PERIOD A -- ck ( ) NA Period A 1,000 µsec to1,000.1 µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Time Base 10 MHz 9,999,950 to 10,000,050 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Page 2 of 2 46

33 TEST INST (S) Philips PM6680B Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (no 50 Ω term) -- ck ( ) NA INPUT 50 Ω -- ck ( ) NA -- MEASURING TIME 1 s -- ck ( ) NA mv 10 Hz 9.9 to 10.1 Hz dbm 50 MHz ,750 to , dbm 110 MHz ,450 to , dbm 220 MHz ,900 to , B INPUT (no 50 Ω term) -- ck ( ) NA -- Set to measure FREQ B -- ck ( ) NA mv 10 Hz 9.9 to 10.1 Hz dbm 50 MHz ,750 to , dbm 110 MHz ,450 to , dbm 220 MHz ,900 to ,100 MHz MHz 4.2 Period and Time Interval Tests PERIOD A -- ck ( ) NA Period A 1,000 µsec to 1,000.1 µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Hz Time Base 10 MHz 9,999,950 to 10,000, Time Base Adjustment 10 MHz 9,999, to 10,000, (48 blank) 47

34 TEST INST (S) Monsanto 100A Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 72 hours A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ -- ck ( ) NA -- TIME 1 SEC -- ck ( ) NA mv rms 10 Hz 9 to 11 Hz TIME 1 msec -- ck ( ) NA mv rms 5 MHz 4,998 to 5,002 khz mv rms 10 MHz 9,997 to 10,003 khz (50 blank) 49

35 TEST INST (S) Monsanto 100B Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ -- ck ( ) NA -- TIME 1 SEC -- ck ( ) NA mv rms 10 Hz 9 to 11 Hz TIME 1 msec -- ck ( ) NA mv rms 1 MHz to 1,000.1 khz dbm 40 MHz 39,998 to 40,002 khz (52 blank) 51

36 TEST INST (S) Fluke 1900A Multifunction Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ -- ck ( ) NA -- RESOLUTION AUTO -- ck ( ) NA mv rms 10 Hz 9 to 11 Hz mv rms 10 MHz to MHz dbm 80 MHz to MHz (54 blank) 53

37 TEST INST (S) Fluke 1910A Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ -- ck ( ) NA -- Rear panel CLOCK to INT -- ck ( ) NA -- TRIGGER LEVEL PRESET -- ck ( ) NA -- RESOLUTION AUTO -- ck ( ) NA mv rms 10 Hz to khz 1 15 mv rms 1 khz to khz mv rms 10 MHz 9, to 10, khz dbm 100 MHz to MHz dbm 125 MHz to MHz RESOLUTION 1 Hz -- ck ( ) NA dbm 10 MHz to khz OVERFLOW Period and Time Interval Tests A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION PER -- ck ( ) NA -- RESOLUTION AUTO Period 1,000 µsec to 1, µsec 1 If TI is out-of-tolerance, adjust the TI rear panel CLOCK ADJ for as close to a khz OVERFLOW indication as possible. (56 blank) 55

38 TEST INST (S) Fluke 1911A Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ -- ck ( ) NA -- Rear panel CLOCK to INT -- ck ( ) NA -- TRIGGER LEVEL PRESET -- ck ( ) NA -- RESOLUTION AUTO -- ck ( ) NA mv rms 10 Hz to khz 1 18 mv rms 1 khz to mv rms 10 MHz 9, to 10, dbm 100 MHz to MHz dbm 125 MHz to RESOLUTION 1 Hz -- ck ( ) NA dbm 10 MHz to khz OVERFLOW 1 B INPUT (no 50 Ω term) -- ck ( ) NA dbm 50 MHz to MHz 170 dbm 170 MHz to dbm 180 MHz to dbm 250 MHz to Period and Time Interval Tests A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION PER -- ck ( ) NA -- RESOLUTION AUTO -- ck ( ) NA Period 1,000 µsec to 1, µsec 1 If TI is out-of-tolerance, adjust the TI rear panel CLOCK ADJ for as close to a khz OVERFLOW indication as possible. (58 blank) 57

39 TEST INST (S) Fluke 1912A Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION FREQ -- ck ( ) NA -- Rear panel CLOCK to INT -- ck ( ) NA -- TRIGGER LEVEL PRESET -- ck ( ) NA -- RESOLUTION AUTO -- ck ( ) NA mv rms 10 Hz to khz 1 15 mv rms 1 khz to mv rms 10 MHz 9, to 10, dbm 100 MHz to MHz dbm 125 MHz to RESOLUTION 1 Hz -- ck ( ) NA dbm 10 MHz to khz OVERFLOW 1 B INPUT (no 50 Ω term) -- ck ( ) NA dbm 50 MHz to MHz dbm 170 MHz to dbm 180 MHz to dbm 520 MHz to Period and Time Interval Tests A INPUT (use 50 Ω term) -- ck ( ) NA FUNCTION PER -- ck ( ) NA -- RESOLUTION AUTO -- ck ( ) NA Period 1,000 µsec to 1, µsec 1 If TI is out-of-tolerance, adjust the TI rear panel CLOCK ADJ for as close to a khz OVERFLOW indication as possible. (60 blank) 59

40 TEST INST (S) Fluke 1920A Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour INPUT 5 Hz to 125 MHz (use 50 Ω term) -- ck ( ) NA FREQ 5 Hz to 125 MHz -- ck ( ) NA -- RESOLUTION 1 Hz -- ck ( ) NA mv rms 10 Hz to khz 1 15 mv rms 1 khz to khz mv rms 10 MHz 9.999,989 to ,011 MHz dbm 125 MHz ,874 to ,126 MHz FREQ 50 MHz to 520 MHz (no 50 Ω term) -- ck ( ) NA dbm 50 MHz ,949 to ,051 MHz dbm 260 MHz ,739 to ,261 MHz dbm 520 MHz ,479 to ,521 MHz Time Base 10 MHz 9,999,990 to 10,000,010 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (62 blank) 61

41 TEST INST (S) Fluke 1952A Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time NA 1 hour CH A (use 50 Ω term) -- ck ( ) NA MODE CONT -- ck ( ) NA -- RANGE GATE 1.0 S -- ck ( ) NA -- FUNCTION FREQ A -- ck ( ) NA -- TRIGGER LEVEL -- ck ( ) NA -- CH A PRESET -- ck ( ) NA mv rms 10 Hz to khz mv rms 100 khz to khz dbm 50 MHz 49, to 50, khz dbm 80 MHz 79,998,559 to 80,001,441 MHz 4.2 Period and Time Interval Tests CH A (use 50 Ω term) -- ck ( ) NA FUNCTION PERIOD A -- ck ( ) NA -- RANGE PERIOD AVGD ck ( ) NA Period A 1,000 msec to msec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 msec to msec Time Base 10 MHz 9,999,820 to 10,000,180 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (64 blank) 63

42 TEST INST (S) Fluke 1952B Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) CH A (use 50 Ω term) -- ck ( ) NA MODE CONT -- ck ( ) NA -- RANGE GATE 1.0 S -- ck ( ) NA -- FUNCTION FREQ A -- ck ( ) NA -- TRIGGER LEVEL CH A PRESET -- ck ( ) NA mv rms 10 Hz to khz mv rms 100 khz to khz dbm 50 MHz 49, to 50, khz dbm 80 MHz 79, to 80, khz 4.2 Period and Time Interval Tests CH A (use 50 Ω term) -- ck ( ) NA FUNCTION PERIOD A -- ck ( ) NA -- RANGE PERIOD ck ( ) NA Period A msec to msec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and msec to msec Time Base 10 MHz 9,999,940 to 10,000,060 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (66 blank) 65

43 TEST INST (S) Fluke 1953A Universal Counter/Timer With or Without OPT05 and OPT07 (OPT05: Time Base Multiplier) (OPT07: 520 MHz Channel C Input) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) CH A INPUT (50 Ω term) -- ck ( ) NA MODE CONT -- ck ( ) NA -- RANGE GATE 1.0 S -- ck ( ) NA -- FUNCTION FREQ A -- ck ( ) NA -- TRIGGER LEVEL CH A PRESET -- ck ( ) NA mv rms 10 Hz to khz mv rms 100 khz to khz dbm 70 MHz 69, to 70, khz dbm 125 MHz 124, to 125, khz OPT07 Channel C only CH C (no 50 Ω term) -- ck ( ) NA -- FUNCTION FREQ C -- ck ( ) NA dbm 50 MHz 49,999,949 to 50,000,051 MHz dbm 250 MHz 249,999,749 to 250,000,251 MHz dbm 520 MHz 519,999,479 to 520,000,521 MHz 4.2 Period and Time Interval Tests CH A (use 50 Ω term) -- ck ( ) NA FUNCTION PERIOD A -- ck ( ) NA -- RANGE PER AVGD ck ( ) NA Period A msec to msec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and msec to msec 9,999,990 to 10,000,010 Hz Time Base 10 MHz 9,999, to Time Base Adjustment 10 MHz 10,000, (only if step fails) (68 blank) 67

44 TEST INST (S) Fluke 1953AOPT20 Counter/Timer With or Without OPT05 and OPT07 (OPT05: Time Base Multiplier) (OPT07: 520 MHz Channel C) (OPT20: ±1.5 x 10-8 /month time base) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) CH A (use 50 Ω term) -- ck ( ) NA MODE CONT -- ck ( ) NA -- RANGE GATE 1.0 S -- ck ( ) NA -- FUNCTION FREQ A -- ck ( ) NA -- TRIGGER LEVEL CH A PRESET -- ck ( ) NA mv rms 10 Hz to khz mv rms 100 khz to khz dbm 70 MHz 69, to 70, khz dbm 125 MHz 124, to 125, khz OPT07 CHANNEL C only CH C (no 50 Ω term) -- ck ( ) NA -- FUNCTION FREQ C -- ck ( ) NA dbm 50 MHz 49,999,990 to 50,000,010 MHz dbm 250 MHz 249,999,954 to 250,000,046 MHz dbm 520 MHz 519,999,905 to 520,000,095 MHz 4.2 Period and Time Interval Tests CH A (use 50 Ω term) -- ck ( ) NA FUNCTION PERIOD A -- ck ( ) NA -- RANGE PER AVGD ck ( ) NA Period A µsec to msec Time Interval A to B (+, -) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and µsec to msec Page 1 of 2 69

45 TEST INST (S) Fluke 1953AOPT20 Counter/Timer With or Without OPT05 and OPT07 (OPT05: Time Base Multiplier) (OPT07: 520 MHz Channel C) (OPT20: ±1.5 x 10-8 /month time base) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) Time Base 10 MHz 9,999,997 to 10,000,003 Hz Time Base Adjustment 9,999, to 10,000, H Page 2 of 2 70

46 TEST INST (S) Racal-Dana 1991 Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time -- ck ( ) NA 72 hours A INPUT (no 50 Ω term) -- ck ( ) NA FREQ A -- ck ( ) NA mv rms 10 Hz 9.90 to Hz dbm 80 MHz ,199 to ,801 MHz dbm 160 MHz ,39 to ,61 MHz 4.2 Period and Time Interval Tests A INPUT (no 50 Ω term) -- ck ( ) NA Period A 1,000 µsec to 1, µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Time Base 10 MHz 9,999,899 to 10,000,101 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (72 blank) 71

47 TEST INST (S) Racal-Dana 1992 Electronic Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time -- ck ( ) NA 72 hours A INPUT (no 50 Ω term) -- ck ( ) NA FREQ A -- ck ( ) NA mv rms 10 Hz 9.90 to Hz dbm 80 MHz ,199 to ,801 MHz dbm 160 MHz ,39 to ,61 MHz C INPUT (no 50 Ω term) -- ck ( ) NA -- FREQ C -- ck ( ) NA dbm 40 MHz ,599 to ,401 MHz dbm 600 MHz ,99 to ,01 MHz dbm 1.3 GHz ,99 to ,01 MHz 4.2 Period and Time Interval Tests PERIOD A -- ck ( ) NA Period A 1,000 µsec to 1, µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Time Base 10 MHz 9,999,899 to 10,000,101 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (74 blank) 73

48 TEST INST (S) Racal-Dana 1992OPT04E Universal Counter With or Without OPT55, ETI (OPT04E: Oven Oscillator, <5 x /day) (OPT55: GPIB Interface) PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time -- ck ( ) NA 72 hours A INPUT (no 50 Ω term) -- ck ( ) NA FREQ A -- ck ( ) NA mv rms 10 Hz 9.90 to Hz dbm 80 MHz ,985 to ,015 MHz dbm 160 MHz ,96 to ,04 MHz C INPUT (no 50 Ω term) -- ck ( ) NA -- FREQ C -- ck ( ) NA -- MHz dbm 40 MHz ,992 to , dbm 600 MHz ,88 to , dbm 1.3 GHz 1, ,76 to 1, , Period and Time Interval Tests A INPUT (no 50 Ω term) -- ck ( ) NA Period A 1,000 µsec to 1, µsec Time Interval A to B (+,-) -- NA Record Time Interval A to B (-, +) -- NA Record Sum of and ,000 µsec to 1,001.0 µsec Time Base 10 MHz 9,999,997 to 10,000,003 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (76 blank) 75

49 TEST INST (S) Racal-Dana 1994 Universal Counter PROC. NA 17-20AF-166 MFR MODEL SER. CALIBRATION ERANCES (5) 3.3 Warm-up Time -- ck ( ) NA 72 hours A INPUT (select 50 Ω, no term required) -- ck ( ) NA mv rms 10 Hz 9.90 to Hz dbm 80 MHz ,039 to ,961 MHz dbm 160 MHz ,079 to ,921 MHz B INPUT (select 50 Ω, no term required) -- ck ( ) NA dbm 10 MHz 9.90 to Hz dbm 50 MHz ,399 to ,601 MHz dbm 100 MHz ,799 to ,201 MHz 4.2 Period and Time Interval Tests A INPUT (no 50 Ω term) -- ck ( ) NA Period A 1,000 µsec to 1, µsec Time Interval A to B (+,-) 500 µsec to µsec Time Interval A to B (-, +) 500 µsec to µsec Sum of and ,000 µsec to 1,001.0 µsec Time Base 10 MHz 9,999,879 to 10,000,121 Hz Time Base Adjustment 10 MHz 9,999, to 10,000, Hz (78 blank) 77

Agilent 5345A Universal Counter, 500 MHz

Agilent 5345A Universal Counter, 500 MHz Agilent 5345A Universal Counter, 500 MHz Data Sheet Product Specifications Input Specifications (pulse and CW mode) 5356C Frequency Range 1.5-40 GHz Sensitivity (0-50 deg. C): 0.4-1.5 GHz -- 1.5-12.4 GHz

More information

Assembly Level Service Guide

Assembly Level Service Guide Assembly Level Service Guide This guide describes how to service the Agilent 53150A, 53151A, and 53152A Microwave Frequency Counters. The information in this guide applies to instruments having the number

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 9500B Users Supplement Issue: 2 Part Number: 1625019 Issue Date: 9/06 Print Date: October 2005 Page Count: 6 Version 11 This supplement contains information necessary to ensure the accuracy

More information

Over 5000 VXI cards and mainframes in stock. 1000's of pieces of Test Equipment in stock. Looking for Test Equipment? Visit us on the web at www.recycledequipment.com Recycled Equipment buys, sells, and

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

Sources of Error in Time Interval Measurements

Sources of Error in Time Interval Measurements Sources of Error in Time Interval Measurements Application Note Some timer/counters available today offer resolution of below one nanosecond in their time interval measurements. Of course, high resolution

More information

PM 6680B / PM 6681 / PM 6681R

PM 6680B / PM 6681 / PM 6681R PM 6680B / PM 6681 / PM 6681R Timer / Counter / Analyzers Rubidium Frequency Reference / Counter / Calibrator Technical Data PM 6681: the highest performance timer/counter/ analyzer available The PM 6681

More information

Microwave Counter, Power Meter and DVM in One Portable Package

Microwave Counter, Power Meter and DVM in One Portable Package Agilent 53140 Series Microwave Counter, Power Meter and DVM in One Portable Package Product Overview Everything you need for the installation and maintenance of microwave links: A choice of frequency counter

More information

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel)

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel) Digital Delay / Pulse Generator Digital delay and pulse generator (4-channel) Digital Delay/Pulse Generator Four independent delay channels Two fully defined pulse channels 5 ps delay resolution 50 ps

More information

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02 Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02 A fully integrated high-performance cross-correlation signal source analyzer from 5 MHz to 33+ GHz Key Features Complete broadband

More information

R-1580A Microwave Downconverter. Product Brochure

R-1580A Microwave Downconverter. Product Brochure R-1580A Microwave Downconverter Product Brochure Jan 2018 Highlights The DSII Model R-1580A Microwave Downconverter extends the coverage of the R-1550A, or other DSII wide range receivers, to 22 GHz. The

More information

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications 100 Hz to 26.5 GHz The HP 71910A/P is a receiver for monitoring signals from 100 Hz to 26.5 GHz. It provides a cost effective combination

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) This product is no longer carried in our catalog. AFG 2020 Characteristics Features Ordering Information Characteristics

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AUTOMATIC VIDEO CORRECTOR TEKTRONIX, MODEL 1440 (NSN )

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AUTOMATIC VIDEO CORRECTOR TEKTRONIX, MODEL 1440 (NSN ) DEPARTMENT OF THE ARMY TECHNICAL BULLETIN TB 11-5820-861-35 CALIBRATION PROCEDURE FOR AUTOMATIC VIDEO CORRECTOR TEKTRONIX, MODEL 1440 (NSN 5820-00-570-1978) Headquarters, Department of the Army, Washington,

More information

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Page 1 of 5 8500A The Giga-tronics 8500A Series Peak Power Meters THE ACCURACY STANDARD channel capability. For tests

More information

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System 7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System A fully integrated high-performance cross-correlation signal source analyzer with platforms from 5MHz to 7GHz, 26GHz, and 40GHz Key

More information

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications Technical Reference TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications 070-8990-04 Please check for change information at the rear of this manual. Fifth

More information

Full-featured CW Microwave Counters for Field, Factory or Lab

Full-featured CW Microwave Counters for Field, Factory or Lab Full-featured CW Microwave Counters for Field, Factory or Lab Product Overview Agilent 53150A 20 GHz Counter Agilent 53151A 26.5 GHz Counter 46 GHz Counter High performance microwave counters: at home,

More information

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A Technical Reference TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A Digitizing Oscilloscopes Performance Verification and Specifications 070-9384-01 Copyright

More information

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix v1.2 Burlington County College INSTRUCTION GUIDE for the Hewlett Packard FUNCTION GENERATOR Model #33120A and Tektronix OSCILLOSCOPE Model #MSO2004B Summer 2014 Pg. 2 Scope-Gen Handout_pgs1-8_v1.2_SU14.doc

More information

2 MHz Lock-In Amplifier

2 MHz Lock-In Amplifier 2 MHz Lock-In Amplifier SR865 2 MHz dual phase lock-in amplifier SR865 2 MHz Lock-In Amplifier 1 mhz to 2 MHz frequency range Dual reference mode Low-noise current and voltage inputs Touchscreen data display

More information

Synthesized Clock Generator

Synthesized Clock Generator Synthesized Clock Generator CG635 DC to 2.05 GHz low-jitter clock generator Clocks from DC to 2.05 GHz Random jitter

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 744 Users Supplement Issue: 6 Part Number: 691287 Issue Date: 4/06 Print Date: September 1998 Page Count: 8 Revision/Date: 1, 2/99 This supplement contains information necessary to ensure

More information

Section one: General description

Section one: General description Section one: General description I Introduction This instrument is one of multifunctional and hi-precision frequency conters that measures frequency from 10Hz to 2700MHz It features eight digits, bright

More information

Oscilloscope Guide Tektronix TDS3034B & TDS3052B

Oscilloscope Guide Tektronix TDS3034B & TDS3052B Tektronix TDS3034B & TDS3052B Version 2008-Jan-1 Dept. of Electrical & Computer Engineering Portland State University Copyright 2008 Portland State University 1 Basic Information This guide provides basic

More information

!Ill ~ 168. Model490 Dual Input, Dual Trace Automatic Peak Power Meter

!Ill ~ 168. Model490 Dual Input, Dual Trace Automatic Peak Power Meter Model490 Dual Input, Dual Trace Automatic Peak Power Meter No other power meter can offer you these features: Help Mode: A Help Mode feature has been added to the Model 490 Automatic Peak Power Meter.

More information

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor LadyBug Technologies, LLC LB5908A True-RMS Power Sensor LB5908ARev8 LadyBug Technologies www.ladybug-tech.com Telephone: 707-546-1050 Page 1 LB5908A Data Sheet Key PowerSensor+ TM Specifications Frequency

More information

Microwave/Counter/Analyzer & Integrated Power Meter

Microwave/Counter/Analyzer & Integrated Power Meter Visit us at www.testequipmentdepot.com 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Microwave/Counter/Analyzer & Integrated Power Meter MCA3000 Series Data Sheet Available

More information

Fluke PM6690. Frequency Counter/Timer/Analyzer

Fluke PM6690. Frequency Counter/Timer/Analyzer Fluke PM6690 Frequency Counter/Timer/Analyzer High resolution and speed: - 12 digits/s frequency measurements - 100 ps single-shot time resolution - 0.001 phase resolution Frequency range: 300 MHz standard;

More information

PERFORMANCE SPECIFICATION SHEET

PERFORMANCE SPECIFICATION SHEET INCH-POUND 5 October 2016 SUPERSEDING w/amendment 2 July 2016 PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUG, ELECTRICAL, COAXIAL, RADIO FREQUENCY, SERIES SMA (CABLED) PIN CONTACT, RIGHT ANGLE, CLASS

More information

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I Prepared by: Kyle Botteon Questions? kyle.botteon@psu.edu 2 Background Information Recall that oscilloscopes (scopes)

More information

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630)

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630) Vorne Industries 87/719 Analog Input Module User's Manual 1445 Industrial Drive Itasca, IL 60143-1849 (630) 875-3600 Telefax (630) 875-3609 . 3 Chapter 1 Introduction... 1.1 Accessing Wiring Connections

More information

Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change.

Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change. Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change. Applications Academic/Education Design, development and manufacturing of passive and active

More information

SAFETY TERMS AND SYMBOLS

SAFETY TERMS AND SYMBOLS Contents Page Safety Terms and Symbols. 1 Introduction. 4 Preliminary Notes 5 Panel Description. 6 Operation Method... 10 Specifications 13 Maintenance. 16 Cleaning 16. Troubleshooting 16. Interchangeable

More information

Noise Detector ND-1 Operating Manual

Noise Detector ND-1 Operating Manual Noise Detector ND-1 Operating Manual SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2 Louisville, CO 80027 Phone: (303) 665-1852 Fax: (303) 604-6088 Table of Contents ND-1 Description...... 3 Safety and Preparation

More information

PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUGS, ELECTRICAL, COAXIAL, RADIO FREQUENCY (SERIES SMA (CABLED) - PLUG, PIN CONTACT, CLASS 2)

PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUGS, ELECTRICAL, COAXIAL, RADIO FREQUENCY (SERIES SMA (CABLED) - PLUG, PIN CONTACT, CLASS 2) INCH-POUND MIL-PRF-39012/55G 6 February 2008 SUPERSEDING MIL-PRF-39012/55G 6 January 2006 PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUGS, ELECTRICAL, COAXIAL, RADIO FREQUENCY (SERIES SMA (CABLED) -

More information

Frequency. Range. HP 83751A/B: 2 GHz to 20 GHz HP 83752A/B: 10 MHz to 20 GHz. Standard 10 MHz timebase: f10 ppm

Frequency. Range. HP 83751A/B: 2 GHz to 20 GHz HP 83752A/B: 10 MHz to 20 GHz. Standard 10 MHz timebase: f10 ppm Frequency Range HP 83751A/B: 2 GHz to 20 GHz HP 83752A/B: 10 MHz to 20 GHz Timebase Stability Standard 10 MHz timebase: f10 ppm High stability timebase (Option le5): Accuracy = Calibration &Aging rate

More information

INTRODUCTION This procedure should only be performed if the instrument fails to meet the Performance Check tests for Output Zero or Offset Accuracy

INTRODUCTION This procedure should only be performed if the instrument fails to meet the Performance Check tests for Output Zero or Offset Accuracy INTRODUCTION This procedure should only be performed if the instrument fails to meet the Performance Check tests for Output Zero or Offset Accuracy (steps A and B). Gain, which affects DC Accuracy, cannot

More information

USB-TG124A Tracking Generator User Manual

USB-TG124A Tracking Generator User Manual USB-TG124A Tracking Generator User Manual Signal Hound USB-TG124A User Manual 2017, Signal Hound, Inc. 35707 NE 86th Ave La Center, WA 98629 USA Phone 360.263.5006 Fax 360.263.5007 This information is

More information

R-1550A Tempest Wide Range Receiver

R-1550A Tempest Wide Range Receiver R-1550A Tempest Wide Range Receiver Product Brochure Version 0.2.00 April 2008 Dynamic Sciences International, Inc. R-1550A TEMPEST Wide Range Measurement Receiver Made specifically for TEMPEST testing

More information

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017 v3 INSTRUCTION GUIDE for the Keysight (formally Agilent; Hewlett-Packard) WAVEFORM GENERATOR Model #33509B 33500 Series and Tektronix OSCILLOSCOPE Model #MSO2004B Fall 2017 Pg. 2 Scope-Gen Handout_part

More information

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator 20 Channel Digital Delay Generator Features of the 745T-20C: 20 Independent delay channels - 100 ps resolution - 25 ps rms jitter - 10 second range Output pulse up to 6 V/50 Ω Independent trigger for every

More information

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment Integrated Component Options Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment PRELIMINARY INFORMATION SquareGENpro is the latest and most versatile of the frequency

More information

Agilent 8560 E-Series Spectrum Analyzers

Agilent 8560 E-Series Spectrum Analyzers Agilent 8560 E-Series Spectrum Analyzers Data Sheet 8560E 30 Hz to 2.9 GHz 8561E 30 Hz to 6.5 GHz 8562E 30 Hz to 13.2 GHz 8563E 30 Hz to 26.5 GHz 8564E 30 Hz to 40 GHz 8565E 30 Hz to 50 GHz 8565E SPECTRUM

More information

4.9 BEAM BLANKING AND PULSING OPTIONS

4.9 BEAM BLANKING AND PULSING OPTIONS 4.9 BEAM BLANKING AND PULSING OPTIONS Beam Blanker BNC DESCRIPTION OF BLANKER CONTROLS Beam Blanker assembly Electron Gun Controls Blanker BNC: An input BNC on one of the 1⅓ CF flanges on the Flange Multiplexer

More information

Keysight FieldFox Microwave Analyzers

Keysight FieldFox Microwave Analyzers Quick Reference Guide Contents Keysight FieldFox Microwave Analyzers Do you have everything?... 1 The Power Button and LED... 1 Battery Usage... 2 Measure Return Loss (CAT Mode)... 3 Measure 1-Port Cable

More information

SatLabs Recommendation for a Common Inter-Facility Link for DVB-RCS terminals

SatLabs Recommendation for a Common Inter-Facility Link for DVB-RCS terminals SatLabs Recommendation for a Common Inter-Facility Link for DVB-RCS terminals Version 1.6-06/01/2005 This document is the result of a cooperative effort undertaken by the SatLabs Group. Neither the SatLabs

More information

SELECTION GUIDE Series of RF and Universal Frequency Counter/Timers

SELECTION GUIDE Series of RF and Universal Frequency Counter/Timers SELECTION GUIDE 53200 Series of RF and Universal Frequency Counter/Timers With the Keysight Technologies, Inc. 53200 RF and Universal Frequency Counters/Timers, You Get: More bandwidth 350 MHz baseband

More information

Microwave/Counter/Analyzer with Integrated Power Meter MCA3000 Series Datasheet

Microwave/Counter/Analyzer with Integrated Power Meter MCA3000 Series Datasheet Microwave/Counter/Analyzer with Integrated Power Meter MCA3000 Series Datasheet With industry-leading frequency and time resolution, the MCA Series comes standard with internal memory, a fast data transfer

More information

MICROWAVE MEASURING INSTRUMENTS

MICROWAVE MEASURING INSTRUMENTS ICROWAVE SYSTE ANAYZER E453//, E538// 70 Hz band 70/140 Hz band Custom-made product OPTION The E453 and E538 series are used to measure the transmission line characteristics in the BB and IF bands in terrestrial

More information

SMPTE STANDARD Gb/s Signal/Data Serial Interface. Proposed SMPTE Standard for Television SMPTE 424M Date: < > TP Rev 0

SMPTE STANDARD Gb/s Signal/Data Serial Interface. Proposed SMPTE Standard for Television SMPTE 424M Date: < > TP Rev 0 Proposed SMPTE Standard for Television Date: TP Rev 0 SMPTE 424M-2005 SMPTE Technology Committee N 26 on File Management and Networking Technology SMPTE STANDARD- --- 3 Gb/s Signal/Data Serial

More information

R3267/3273 Spectrum Analyzers

R3267/3273 Spectrum Analyzers R3267/3273 Spectrum Analyzers For 3rd-Generation Mobile Communications Present Digital Communication standards (W-CDMA, PDC, PHS, IS-136, GSM, DECT, cdmaone ) R3267/3273 New communication technologies

More information

FRQM-2 Frequency Counter & RF Multimeter

FRQM-2 Frequency Counter & RF Multimeter FRQM-2 Frequency Counter & RF Multimeter Usage Instructions Firmware v2.09 Copyright 2007-2011 by ASPiSYS Ltd. Distributed by: ASPiSYS Ltd. P.O.Box 14386, Athens 11510 (http://www.aspisys.com) Tel. (+30)

More information

R&S FSMR Measuring Receiver Specifications

R&S FSMR Measuring Receiver Specifications Test & Measurement Data Sheet 09.00 R&S FSMR Measuring Receiver Specifications CONTENTS Specifications... 3 Frequency...3 Measuring receiver... 3 Frequency counter...3 RF power...4 RF level (tuned receiver)...5

More information

GFT Channel Slave Generator

GFT Channel Slave Generator GFT1018 8 Channel Slave Generator Features 8 independent delay channels 1 ps time resolution < 100 ps rms jitter for optical triggered delays 1 second range Electrical or optical output Three trigger modes

More information

PIM Master. Maintenance Manual

PIM Master. Maintenance Manual Maintenance Manual PIM Master MW8208A, MW8209A, MW8219A Passive InterModulation Analyzer MW8208A: 869 MHz to 894 MHz MW8209A: 925 MHz to 960 MHz MW8219A: 1930 MHz to 1930 MHz and 2110 MHz to 2155 MHz Anritsu

More information

USER MANUAL. Blackburst, Sync, Audio Tone Generator. For Models BSG-50, RM-50/BSG, SR-50/BSG. Doc Rev. F (C) Copyright 2014

USER MANUAL. Blackburst, Sync, Audio Tone Generator. For Models BSG-50, RM-50/BSG, SR-50/BSG. Doc Rev. F (C) Copyright 2014 HORITA BSG-50 Blackburst, Sync, Audio Tone Generator USER MANUAL For Models BSG-50, RM-50/BSG, SR-50/BSG Doc. 070450 Rev. F (C) Copyright 2014 P.O. Box 3993, Mission Viejo, CA 92690 (949) 489-0240 www.horita.com

More information

Analog Arts SA985 SA975 SA935 SA915 Product Specifications [1]

Analog Arts SA985 SA975 SA935 SA915 Product Specifications [1] www.analogarts.com Analog Arts SA985 SA975 SA935 SA915 Product Specifications [1] 1. These models consist of an oscilloscope, a spectrum analyzer, a data recorder, and a frequency & phase meter. Oscilloscope/

More information

LadyBug Technologies LLC Manual PowerSensor+ Field Certification Procedure

LadyBug Technologies LLC Manual PowerSensor+ Field Certification Procedure LadyBug Technologies LLC Manual PowerSensor+ Field Certification Procedure Procedure Applies to Following Power Sensors LB478A, LB479A, LB480A, LB559A, LB579A, LB589A, LB679A, LB680A Contents Purpose:...

More information

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & General Safety Summary. Version 1.03

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & General Safety Summary. Version 1.03 General Safety Summary General Safety Summary User Manual Digital Storage Oscilloscopes Models 2534, 2540 & 2542 Review the following safety precautions to avoid injury and prevent damage to this product

More information

Technical Data. HF Tuner WJ-9119 WATKINS-JOHNSON. Features

Technical Data. HF Tuner WJ-9119 WATKINS-JOHNSON. Features May 1996 Technical Data WATKINS-JOHNSON HF Tuner WJ-9119 WJ designed the WJ-9119 HF Tuner for applications requiring maximum dynamic range. The tuner specifically interfaces with the Hewlett-Packard E1430A

More information

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications March 2006 Agilent multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical

More information

Fluke /

Fluke / Fluke 190-104/190-204 ScopeMeter 190 Series II Service Manual PN 4822 872 05405 March 2011 2011 Fluke Corporation, All rights reserved. Printed in the Netherlands All product names are trademarks of their

More information

3124mb+ All Discrete 4 Channel Mic/Instrument Preamplifier with Stereo Mixer Operator s Manual

3124mb+ All Discrete 4 Channel Mic/Instrument Preamplifier with Stereo Mixer Operator s Manual 3124mb+ All Discrete 4 Channel Mic/Instrument Preamplifier with Stereo Mixer Operator s Manual Written by Carl J Houde 2015 Table of Contents 1.0 Introduction... 3 2.0 Overview... 4 2.1 3124mb+ Features

More information

University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope

University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope Name:, A. Stolp, 2/2/00 rev, 9/15/03 NOTE: This is a fill-in-the-blanks lab. No notebook is required. You are encouraged

More information

8650A SERIES UNIVERSAL POWER METERS

8650A SERIES UNIVERSAL POWER METERS 8650A SERIES UNIVERSAL METERS 8650A SERIES UNIVERSAL METERS The Capabilities to Test Today s Sophisticated Communications Systems TIME BURST AVERAGE The Giga-tronics 8650A Series Universal Power Meters

More information

JD725A Cable and Antenna Analyzer - Dual Port

JD725A Cable and Antenna Analyzer - Dual Port COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS JD725A Cable and Antenna Analyzer - Dual Port Key Features Portable and lightweight handheld instrument Built-in wireless frequency bands as well as the most

More information

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & 2542

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & 2542 User Manual Digital Storage Oscilloscopes Models 2534, 2540 & 2542 General Safety Summary General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product

More information

ASNT_PRBS20B_1 18Gbps PRBS7/15 Generator Featuring Jitter Insertion, Selectable Sync, and Output Amplitude Control

ASNT_PRBS20B_1 18Gbps PRBS7/15 Generator Featuring Jitter Insertion, Selectable Sync, and Output Amplitude Control ASNT_PRBS20B_1 18Gbps PRBS7/15 Generator Featuring Jitter Insertion, Selectable Sync, and Output Amplitude Control Broadband frequency range from 20Mbps 18.0Gbps Minimal insertion jitter Fast rise and

More information

WVR500 Waveform/Vector Monitor

WVR500 Waveform/Vector Monitor Service Manual WVR500 Waveform/Vector Monitor 070-8897-01 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are

More information

CARLO GAVAZZI Automation Components

CARLO GAVAZZI Automation Components CARLO GAVAZZI Automation Components UDM 35/40 Digital Panel Meter Programming Guide Index Description 2 Programming Fundamentals 3 Access to Programming Mode/Password Protection 4 Programming 5-18 Inputs

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 NATIONAL INSTRUMENTS CALIBRATION SERVICES PENANG (NICSP) NI Malaysia Sdn Bhd 8 Lebuh Batu Maung 1 Bayan Lepas 11960 Penang Malaysia

More information

CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock

CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock Products: CMU200 CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock This application explains the setup and procedure to measure the exact time relationship

More information

R&S ZVA110 Vector Network Analyzer Specifications

R&S ZVA110 Vector Network Analyzer Specifications ZVA110_dat-sw_en_5214-4813-22_cover.indd 1 Data Sheet 04.00 Test & Measurement R&S ZVA110 Vector Network Analyzer Specifications 15.11.2013 14:42:28 CONTENTS Definitions... 3 Specifications... 4 Overview...

More information

Agilent E5100A Network Analyzer

Agilent E5100A Network Analyzer Agilent E5100A Network Analyzer Data Sheet These specifications are the performance standards or limits against which the instrument is tested. When shipped from the factory, the E5100A meets the specifications

More information

18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V

18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V 18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V With its characteristics of power stability whatever the load, very fast response time when pulsed (via external modulated signal), low ripple,

More information

SM02. High Definition Video Encoder and Pattern Generator. User Manual

SM02. High Definition Video Encoder and Pattern Generator. User Manual SM02 High Definition Video Encoder and Pattern Generator User Manual Revision 0.2 20 th May 2016 1 Contents Contents... 2 Tables... 2 Figures... 3 1. Introduction... 4 2. acvi Overview... 6 3. Connecting

More information

VNA Master Model MS203xA

VNA Master Model MS203xA Maintenance Manual VNA Master Model MS203xA Vector Network Analyzer Anritsu Company 490 Jarvis Drive Morgan Hill, CA 95037-2809 USA P/N: 10580-00168 Revision: B Printed: December 2008 Copyright 2008 Anritsu

More information

Calibrating the CMD Output Level for BER

Calibrating the CMD Output Level for BER Ref Lvl -32 dbm -40-50 Marker 1 [T1] -35.50 dbm 5.000000 ms RBW 300 khz RF Att 0 db VBW 3 MHz SWT 5 ms Unit dbm 1 SUMMARY [T1] A RMS AVG -85.86 dbm SGL TRG -60-70 1RM -80 EXT -90-100 -110-120 -130 T1 T2

More information

PROGRAMMABLE DC SOURCE VIEW RECORDERS Programmable DC Source FEATURES

PROGRAMMABLE DC SOURCE VIEW RECORDERS Programmable DC Source FEATURES Programmable DC Source Sink and Source Capability Conventional DC sources usually have the function of power supply (source) only without power absorption (sink) function. In addition to the source function,

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

PM Series Microwave Power Calibration System

PM Series Microwave Power Calibration System PM Series Microwave Power Calibration System Supports Sensors from most major manufactures up to 50 GHz Faster than direct compare method Lowest total uncertainty National Metrology Institute class thermistor

More information

PM Series Microwave Power Calibration System

PM Series Microwave Power Calibration System PM Series Microwave Power Calibration System Supports Sensors from most major manufacturers from 6 khz to 50 GHz Faster than direct compare method Lowest total uncertainty National Metrology Institute

More information

Product Introduction. Duplexer Box MN2555A. Signal Analyzer MS2830A. Ver.1.0

Product Introduction. Duplexer Box MN2555A. Signal Analyzer MS2830A. Ver.1.0 Product Introduction Duplexer Box MN2555A Signal Analyzer MS2830A Ver.1.0 Duplexer Box MN2555A Duplexer Box MN2555A Connecting MN2555A, Signal Analyzer MS2830A, and USB Power Sensor *The shape of the accessory

More information

Digital Storage Oscilloscopes 2550 Series

Digital Storage Oscilloscopes 2550 Series Data Sheet Digital Storage Oscilloscopes 2550 Series The 2550 series digital storage oscilloscopes provide high performance and value in 2-channel and 4-channel configurations. With bandwidth from 70 MHz

More information

SMS3000X Series Spectrum Analyzer

SMS3000X Series Spectrum Analyzer Data Sheet SMS3000X Series Spectrum Analyzer SMS3032X SMS3021X General Description SMS3000X series spectrum analyzer has a frequency range from 9 khz up to 2.1 GHz/3.2 GHz, it is light weight and small

More information

C8491 C8000 1/17. digital audio modular processing system. 3G/HD/SD-SDI DSP 4/8/16 audio channels. features. block diagram

C8491 C8000 1/17. digital audio modular processing system. 3G/HD/SD-SDI DSP 4/8/16 audio channels. features. block diagram features 4 / 8 / 16 channel LevelMagic2 SDI-DSP with level or loudness (ITU-BS.1770-1/ ITU-BS.1770-2, EBU R128) control 16 channel 3G/HD/SD-SDI de-embedder 16 in 16 de-embedder matrix 16 channel 3G/HD/SD-SDI

More information

V6118 EM MICROELECTRONIC - MARIN SA. 2, 4 and 8 Mutiplex LCD Driver

V6118 EM MICROELECTRONIC - MARIN SA. 2, 4 and 8 Mutiplex LCD Driver EM MICROELECTRONIC - MARIN SA 2, 4 and 8 Mutiplex LCD Driver Description The is a universal low multiplex LCD driver. The version 2 drives two ways multiplex (two blackplanes) LCD, the version 4, four

More information

Analog Arts SA985 SA975 SA935 SA915 Product Specifications

Analog Arts SA985 SA975 SA935 SA915 Product Specifications Analog Arts SA985 SA975 SA935 SA915 Product Specifications Oscilloscope/ Spectrum Analyzer/ Data Recorder Model SA985 SA975 SA935 SA915 Oscilloscope (Typical by Design) Bandwidth (Max at probe tip) [1]

More information

PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUGS, ELECTRICAL, COAXIAL RADIO FREQUENCY, (SERIES BNC (CABLED), PIN CONTACT, CLASS 2)

PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUGS, ELECTRICAL, COAXIAL RADIO FREQUENCY, (SERIES BNC (CABLED), PIN CONTACT, CLASS 2) INCH-POUND MIL-PRF-39012/16H 16 November 2006 SUPERSEDING MIL-PRF-39012/16G 26 September 1994 PERFORMANCE SPECIFICATION SHEET CONNECTORS, PLUGS, ELECTRICAL, COAXIAL RADIO FREQUENCY, (SERIES BNC (CABLED),

More information

ATS MS/s 8-Bit PCI Digitizer

ATS MS/s 8-Bit PCI Digitizer 2 channels sampled at 8-bit resolution 50 MS/s simultaneous real-time sampling rate on each input ±20mV to ±20V input range 256 Kilo samples of on-board acquisition memory per channel AlazarDSO Oscilloscope

More information

GFT Channel Digital Delay Generator

GFT Channel Digital Delay Generator Features 20 independent delay Channels 100 ps resolution 25 ps rms jitter 10 second range Output pulse up to 6 V/50 Ω Independent trigger for every channel Fours Triggers Three are repetitive from three

More information

Converters Series Tri-Band Frequency Converters. Back to Multiband. Specifications Upconverters Downconverters Phase Noise Options Rear Panel

Converters Series Tri-Band Frequency Converters. Back to Multiband. Specifications Upconverters Downconverters Phase Noise Options Rear Panel R Back to Multiband Converters 9700 Series Tri-Band Frequency Converters 9700 SERIES TRI-BAND FREQUENCY CONVERTERS The MITEQ frequency converters are designed for advanced satellite communication systems

More information

ScopeMeter 190 Series Specifications

ScopeMeter 190 Series Specifications Seite 1 von 7 ScopeMeter 190 Series Specifications Product Home Features Specifications Models, Options & Accessories Oscilloscope Mode Meter Mode Recorder Mode General Specifications Oscilloscope Mode

More information

SDA 3302 Family. GHz PLL with I 2 C Bus and Four Chip Addresses

SDA 3302 Family. GHz PLL with I 2 C Bus and Four Chip Addresses GHz PLL with I 2 C Bus and Four Chip Addresses Preliminary Data Features 1-chip system for MPU control (I 2 C bus) 4 programmable chip addresses Short pull-in time for quick channel switch-over and optimized

More information

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations DTA-2115B Verification of Specifations APPLICATION NOTE January 2018 Table of Contents 1. Introduction... 3 General Description of the DTA-2115B... 3 Purpose of this Application Note... 3 2. Measurements...

More information

Solid State Logic S O U N D V I S I O N

Solid State Logic S O U N D V I S I O N Solid State Logic S O U N D V I S I O N SUPERANALOGUE X - R A C K Super-Analogue Outboard XR622 X-Rack Master Module User s Guide This documentation package contains the User s Guide for your new X-Rack

More information

BNC-2120 INSTALLATION GUIDE. Connector Accessory for Multifunction DAQ Devices

BNC-2120 INSTALLATION GUIDE. Connector Accessory for Multifunction DAQ Devices INSTALLATION GUIDE BNC-2120 Connector Accessory for Multifunction DAQ Devices This installation guide describes how to install, configure, and use your BNC-2120 accessory. If you have not already installed

More information

PicoScope 6407 Digitizer

PicoScope 6407 Digitizer YE AR PicoScope 6407 Digitizer HIGH PERFORMANCE USB DIGITIZER Programmable and Powerful 1 GHz bandwidth 1 GS buffer size 5 GS/s real-time sampling Advanced digital triggers Built-in function generator

More information

R&S RT-Zxx High-Voltage and Current Probes Specifications

R&S RT-Zxx High-Voltage and Current Probes Specifications R&S RT-Zxx High-Voltage and Current Probes Specifications Test & Measurement Data Sheet 14.00 CONTENTS Definitions... 3 Probe/oscilloscope chart... 4 R&S RT-ZH10/-ZH11 high-voltage probes... 5 R&S RT-ZD01

More information