1 Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Dektak 3030 Stylus Profilometer Version 2011 May 16 I. Purpose This Standard Operating Procedure (SOP) outlines requirements to be considered by an authorized user of the Dektak 3030 Stylus Profilometer as well as describes the normal operation of the profilometer and any hazards that may be encountered during normal operation. Finally, the SOP explains how to minimize any hazards and how to respond in an emergency situation. This document is to be reviewed one year from the date of approval or as conditions warrant, whichever is the shorter time period. The Dektak 3030 Surface Profiler is an advanced surface profile measuring system, which accurately measures vertical features ranging in height from 131 micron to 50 angstroms on a wide variety of substrate surfaces. II. Personnel A. Authorized Personnel: The Dektak 3030 may be operated only by authorized personnel who are fully cognizant of all safety issues involved in the operation of such a device. These personnel are to ensure that the profilometer is only operated in the manner laid out in this document. To become an authorized user, one must: 1. Complete Environment, Health & Safety (EH&S) training class. 2. Take the baseline BNC Safety Orientation class 3. Read and fully understand the SOP 4. Receive training on the profilometer by an authorized Super User and User. 5. Sign the authorized user sheet to affirm that the above steps have been completed.

2 B. Unauthorized personnel: No unauthorized personnel may enter the BNC clean room facility unless accompanied by an authorized user. All visitors must be briefed on proper safety protocol and must wear appropriate protective eyewear located on the premises. III. Hazards A. Electrical Hazards: Use normal precautions with external house (110VAC) connections behind the chassis. B. Chemical: N/A. C. Pressure Hazards: N/A D. Other: N/A IV. Hazard Controls A. Electrical 1. Only qualified personnel may perform all internal maintenance to the profilometer and more than one user must be present when performing said maintenance. 2. Every portion of the electrical system, including the printed circuit cards, should be assumed to be at dangerous voltage level. C. Chemical and Pressure I. N/A D. Other 1. Proper eye protection must be worn at all times in the clean room. Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 2

3 V. Technical Specifications Stylus: 12.5 μm radius (Diamond) Scan Range: x: 50 μm 50 mm z: 100 Å 131 μm z Resolution: 1 Å/65 KÅ 10 Å/655 KÅ 0 Å/1310 KÅ Force: 1 40 mg Zoom: 35x 200x Scan Speed Ranges: Low, Med, High Leveling: Manual, two-point programmable or cursor leveling Stylus Tracking Force: Programmable, 1-40 mg ( milli Newtons) Maximum Sample Thickness: 45 mm (1.75 inches) Maximum Sample Weight: 0.5 Kg (1 lb) Sample Stage Diameter: 165 mm (6.5 inches) Sample Stage Translation: X Axis, ± 76 mm (± 3 inches) (from center) Y Axis, -76mm (3.0 inches) Sample Stage Rotation: Theta, 360º Power Requirements: 100/115/220Vac ± 10%, 50-60Hz, 200VA Warm-up Time: 15 minutes recommended for maximum stability Operating Temperature: 21º C ± 3º C (70º F ± 5º F) Specification Summary Scientific* Metric* English* VERTICAL DISPLAY RANGE 100 Å to 1310 KÅ 10 nm to 131 μm 0.4 μin to 5200 μin VERTICAL RESOLUTION In a Selected Range 1Å/65 KÅ 10 Å/655 KÅ 20 Å/1310 KÅ 0.1 nm/6.5 μm 1 nm/65.5 μm 2 nm/131 μm μin/255 μin μin/2574 μin μin/5148 μin DATA POINTS PER MICRON 0.04 to to to 40 SCAN LENGTH RANGE 50 μm to 50 mm 50 μm to 50 mm 2000 μin to mils Å = Angstroms μm = micrometer (micron) * Switch selectable on top panel of Main Electronics Unit. Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 3

4 VI. Operating the System 1. Switch ON red power switch on the back/top of Power and Control Box. If the system does not turn on, check the power cord to the system. 2. The Dektak 3030 Logo should appear on the screen display. If the sample or the stage image appears in the background, Press VID until the sign-on message shows against an uniform gray background. 3. Rotate brightness knob left/right to adjust screen brightness. 4. Raise the optics tower until the stylus is approximately one inch off the stage surface. 5. Press PRGM. The Dektak 3030 Main Menu will appear. The menu should be set at the Scan Program Menu. If not, use the Arrow keys to move the prompt. Press ENTR. 6. Now you are ready to set the scanning parameters. 7. The screen should look like the following. Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 4

5 8. Move the prompt to the Scan Program Menu using the arrow keys. Press ENTR key. The Scan Program Menu will appear on the CRT. 9. Using arrow keys move the prompt to the Scan Length parameter. Determine the appropriate Scan Length. Type in the length using the numeric key pad, and press ENTR. 10. Enter No for the Auto Leveling and Smoothing parameters. 11. Enter Medium for the speed. 12. The Profile parameter should be (hills and valleys). 13. Select the appropriate vertical Measurement Range and press the ENTR key. i. 65KA Range will provide 1A Vertical Resolution ii. 655KA Range will provide 10A Vertical Resolution iii. 1,310KA Range will provide 20A Vertical Resolution 14. Enter Auto for the Display Range. 15. The rest of the parameters should not be of concern. However, the stylus force may need to be altered. 16. The Dektak 3030 is now ready for scanning. 17. Sample Positioning is achieve thru turning the thumbwheels located on the front of the sample stage. Warning: DO NOT FORCE ANYTHING IF IT DOES NOT MOVE. Please note that this is strictly a manual stage. The theta adjustment is accomplished by the stage rotation. The X-axis (side to side) is accomplished by the thumbwheel located slightly to the left of center. The Y-axis (front to back) is accomplished by the thumbwheel located slightly to the right of center. And the manual leveling of the stage is accomplished by the base (largest) thumbwheel. 18. Place the calibration standard or your sample near the center rear of the stage. 19. Use the X and Y thumbwheels to move the stage until the sample is below the stylus. 20. Press VID twice. This is the video/graphics overlay mode. Press FCTN, 5. The software reticle and the video image are displayed simultaneously. 21. Press AV (Arrow up and Arrow down symbol). The stylus will lower onto the sample. Adjust the optics until the stylus tip is positioned on the X-axis of the reticle. Adjust the Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 5

6 illumination to obtain the required contrast by using the small knob on the front of the instrument panel. 22. Orient the sample until the video display shows the cross hairs, sample, and the hourglass image of the stylus in alignment. Note: when scanning is performed, the sample moves from back to front. However, the screen will show the sample moving from right to left since the image is optically rotated 90 degrees. 23. Press AV (Arrow up and Arrow down symbol) again to raise the stylus. The instrument is now ready to make a scan. 24. You can now make a scan by pressing SCAN. 25. The stylus will lower and an initial plot will be plotted on the screen as the scan commences. When the scan has completed, the sample stage will automatically return to the home position. The DEKTAK 3030 will immediately replot the original scan using the autoranging feature programmed into the Scan Program Menu. 26. If the initial trace touches either the top or the bottom of the screen before the completion of the scan, the scan has been found to be more than 327KA out of level. The stage needs to be manually leveled. i. Manual leveling is achieved by adjusting the leveling thumbwheel, located on the scanning stage, while the stage is tracking in a horizontal line. ii. Press SCAN Key. As the stage is moving and a trace is being generated on the screen, turn the leveling thumbwheel until the trace is tracking in a horizontal line. Clockwise rotation raises the trace and counterclockwise rotation lowers the trace. iii. Press SCAN key again. The profile must appear totally within the graphic boundaries to achieve the minimum acceptable manual leveling. If not, repeat the manual leveling procedure. In a worst case, stop and get help. 27. Assuming that you were able to scan successfully. 28. If the plot was not leveled, you will need to CURSOR level the data by using the REF and MEAS cursors. Both cursors are positioned with the arrow keys on the horizontal trace. Then pressing LVL. The data and plot will be leveled with respect to the REF and MEAS vertical positions. The screen will plot with both R and M Cursor intercepts and positioned at the horizontal zero grid line with the trace leveled. Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 6

7 29. You are now ready to make a measurement. 30. Once the trace has been adequately leveled, an accurate measurement can be obtained. 31. Press MEAS. Using the left ARROW key, move the M Cursor to the middle of the top step of the trace. 32. Press REF. Using the right ARROW key, move the R Cursor to the base of the step of the trace. The difference between the R cursor and the M cursor intercepts is the step height. This height is automatically displayed in the upper right corner of the monitor labeled Vert. 33. You will need to ZERO the reference. Any point along the trace can be considered the zero point. Press REF. Using the right ARROW key, move the R Cursor to a position on the step and press ZERO. The screen will replot with the trace in the original position but the vertical scale will be rescaled in the negative direction. If you move the R cursor to the base of the step and press ZERO, the R cursor will be the base value or zero vertical position. This is considered the reference point of the measurement. 34. You can now magnify the trace. 35. In order to magnify an area of interest, press LEFT and use the right ARROW key to bring the left boundary near the leading edge of the step. Repeat the procedure using the RT, BOT, and TOP keys along with the appropriate ARROW keys until the annotated screen box looks smaller. You are declaring the boundaries for the areas of interest. 36. Press RPLT. This will replot the profile and rescale the profile of interest to allow an accurate measurement. VII. Shut Down Procedures 1. Rotate Illuminator knob [fully clockwise to turn off the light. 2. Rotate Zoom/Focus knob fully clockwise to lift stylus off the surface. 3. Remove sample. 4. Switch off RED power switch. VIII. Emergency Procedures A. DEKTEK profilometer accidents: Notify lab management and PI immediately. B. Power outage: If there is a power outage, turn off the profilometer per the shut down procedure (VII) to avoid a hazardous situation when power is restored. If there is an emergency, leave laboratory immediately and either return after emergency to shut down profilometer or contact lab management. Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 7

8 Authorized Users: I have read and understood the Standard Operating Procedures for DEKTAK 3030 Stylus Profilometer Name (print) Signature Date PI or Super User Initial Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 8

9 Appendix A In case of medical emergencies, consult lab safety protocol or lab safety plan. In the event of a profilometer accident, follow the procedure below: 1. Ensure that the profilometer is shut off per Section VII in Shut Down Procedures. 2. Provide for the safety of the personnel (first aid, evacuation, etc.) as needed. 3. Obtain medical assistance for anyone who may be injured. UC Optometry Clinic (Normal Hours) UC Optometry Clinic (24 Hour Emergencies) University Health Services (Emergency) Ambulance (urgent medical care) If there is a fire, pull the alarm, and contact the fire department by calling Do not fight the fire unless it is very small and you have been trained in fire fighting techniques. 5. Inform the Office of Environment Health, & Safety (EH&S) as soon as possible. 6. During normal working hours, call the following: EH&S Office BNC Safety Officer ( Paul Lum) EH&S Health & Safety Manager After normal working hours, call to contact the UC Police Department who can contact the above using their emergency call list. 7. Inform (PI NAME) and the BNC safety officer (BNC Management) as soon as possible. If there is an injury, (PI NAME) will need to submit a report of injury to the Worker s Compensation Office. 8 After the incident, do not resume use of the profilometer system until the BNC lab manager and EH&S has reviewed the incident and approved the resumption of research. Standard Operating Procedure BNC DEKTAK 3030 (Stylus Profilometer 2) 9

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Veeco Dektak 6M Profilometer

Veeco Dektak 6M Profilometer Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

K Service Source. Apple High-Res Monochrome Monitor

K Service Source. Apple High-Res Monochrome Monitor K Service Source Apple High-Res Monochrome Monitor K Service Source Specifications Apple High-Resolution Monochrome Monitor Specifications Characteristics - 1 Characteristics Picture Tube 12-in. diagonal

More information

K Service Source. Apple High-Res Monochrome Monitor

K Service Source. Apple High-Res Monochrome Monitor K Service Source Apple High-Res Monochrome Monitor K Service Source Specifications Apple High-Resolution Monochrome Monitor Specifications Characteristics - 1 Characteristics Picture Tube 12-in. diagonal

More information

WVR500 Waveform/Vector Monitor

WVR500 Waveform/Vector Monitor Service Manual WVR500 Waveform/Vector Monitor 070-8897-01 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are

More information

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

Model Camera System (CCTV) User Manual

Model Camera System (CCTV) User Manual Model 4330 Camera System (CCTV) User Manual ETS-Lindgren L.P. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained

More information

A3U5 Replacement 1705A Spectrum Monitor

A3U5 Replacement 1705A Spectrum Monitor Instructions 050-3621-00 A3U5 Replacement 1705A Spectrum Monitor 075-0876-00 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing

More information

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision

More information

Operating Instructions

Operating Instructions Operating Instructions VIA-170 R Video Image Marker- Measurement System Accuracy by Design P/N 700051 ii VIA-170 Video Image Marker-Measurement System User's Manual R iii Copyright 1994-1998 by Boeckeler

More information

Contents. Instruction Manual T-Rex Page 2 of 16 Release 1.01

Contents. Instruction Manual T-Rex Page 2 of 16 Release 1.01 Contents 1 Safety Precautions... 3 2 Introduction:... 5 3 Theory of Operation... 7 4 Unpacking Procedure... 8 5 Operating TR-Mark III with T-Rex... 9 6 Operating a TR-Mark II with a T-Rex... 13 7 Technical

More information

Flat-Bed Module Recorders

Flat-Bed Module Recorders Flat-Bed Module Recorders Model No. 08376-50 08376-55 08376-60 0115-0192 4/28/00 Table of Contents Introduction...3 Power Requirements...3 Chart Paper Installation...3 Pen Installation...5 Grounding...5

More information

Metal Electrode Meter

Metal Electrode Meter Metal Electrode Meter INSTRUCTION MANUAL FOR Metal Electrode Meter MODEL 2900 Serial # Date PO Box 850 Carlsborg, WA 98324 U.S.A. 360-683-8300 800-426-1306 FAX: 360-683-3525 http://www.a-msystems.com Version

More information

ULTRA-TRAC APL INSTRUCTION MANUAL. Read and understand instructions before use. Patented. 851 Transport Drive Valparaiso, IN

ULTRA-TRAC APL INSTRUCTION MANUAL. Read and understand instructions before use. Patented. 851 Transport Drive Valparaiso, IN ULTRA-TRAC APL A C O U S T I C P I P E L O C A T O R INSTRUCTION MANUAL Read and understand instructions before use. Patented MADE IN USA 851 Transport Drive Valparaiso, IN 46383-8432 Phone: 888 4SENSIT

More information

19 / 20.1 / 22 WIDE SCREEN TFT-LCD MONITOR

19 / 20.1 / 22 WIDE SCREEN TFT-LCD MONITOR 19 / 20.1 / 22 WIDE SCREEN TFT-LCD MONITOR V193/ V220 Series V202 Series USER MANUAL www.viewera.com Rev. 2.0 Table of Contents EMC Compliance......1 Important Precautions...2 1. Package contents....3

More information

Activity P32: Variation of Light Intensity (Light Sensor)

Activity P32: Variation of Light Intensity (Light Sensor) Activity P32: Variation of Light Intensity (Light Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Illuminance P32 Vary Light.DS P54 Light Bulb Intensity P54_BULB.SWS Equipment Needed

More information

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD POL-200 Semiautomatic Polarimeter Instruction Manual BANTE INSTRUMENTS CO., LTD POL-200 Semiautomatic Polarimeter 1 Introduction Thank you for selecting the POL-200 semiautomatic polarimeter. This manual

More information

Specification of Main Parts LCD Size 0.76-inch (D10) Pixel Number 2,304,000 dots (1920 x 1200) x 3 Native Resolution WUXGA(16:10) Projection Lens Type Power zoom / Power focus / Power shift F-Number/ Focal

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

DISTRIBUTION AMPLIFIER

DISTRIBUTION AMPLIFIER MANUAL PART NUMBER: 400-0045-005 DA1907SX 1-IN, 2-OUT VGA/SVGA/XGA/UXGA DISTRIBUTION AMPLIFIER USER S GUIDE TABLE OF CONTENTS Page PRECAUTIONS / SAFETY WARNINGS... 2 GENERAL...2 GUIDELINES FOR RACK-MOUNTING...2

More information

CM-S38901SV TVL IR Long Range camera

CM-S38901SV TVL IR Long Range camera 5 40 TVL IR Long Range camera User s Guide CM-S38901SV SAFETY PRECAUTIONS WARNING 1. Be sure to use only the standard adapter that is specified in the specification sheet. Using any other adapter could

More information

Tube Rotator. User Guide. Version 1.2

Tube Rotator. User Guide. Version 1.2 Tube Rotator User Guide Version 1.2 Figure 1: Fixed Speed Model Tube holder spindle Tilt adjustment wheel IEC power inlet socket (at rear) Power on/off switch Figure 2: Variable Speed Model Tube holder

More information

SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS.

SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS. SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. 1.) The luminaire shall be a motorized colour mixing luminaire employing seven (7) homogenized red, green, blue, and white LED

More information

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject

More information

RHP-IS DESKTOP IMAGE STABILIZER

RHP-IS DESKTOP IMAGE STABILIZER RHP-IS DESKTOP IMAGE STABILIZER Instruction Manual Firmware release 2.1 and above 2005-06-22 Rock2000.com P.O. BOX 4242, Middletown, New York, 10941, USA www.rock2000.com support@rock2000.com 2 Contents

More information

LM-WPS41 HD Caption Adder. User. Manual

LM-WPS41 HD Caption Adder. User. Manual LM-WPS41 HD Caption Adder User Manual 1 Table of Contents 1. Installation Notes 3 2.Product Introduction 5 3. Machine installation 7 4. Software debugging 8 5. Remote Control Description 12 Chapter One

More information

The Bloomberg 15-inch Compact Display Terminal Technical Information

The Bloomberg 15-inch Compact Display Terminal Technical Information The Bloomberg 15-inch Compact Display Terminal Technical Information Description The Bloomberg 15-inch Compact Display Terminal is a high-quality display device that is versatile and ergonomically designed.

More information

LVP602S LED Video Processor USER S MANUAL

LVP602S LED Video Processor USER S MANUAL LVP602S LED Video Processor USER S MANUAL TABLE OF CONTENTS I. Safety precautions 3 II. Connections of hardware 1.Rear view 4 2. Port description 4 III. Frontal panel operations 1. Diagram of frontal panel

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

QuickSpecs. Models RB146AA#ABA Standard Configuration RB146AT#ABA Promotional Part Number (SmartBuy) HP L5006tm 15-inch LCD Touchscreen Monitor

QuickSpecs. Models RB146AA#ABA Standard Configuration RB146AT#ABA Promotional Part Number (SmartBuy) HP L5006tm 15-inch LCD Touchscreen Monitor Overview 1. 2. 3. 4. 5. Power: Turns the unit on and off. Select: Selects the adjustment items from the on-screen display (OSD) menus. Arrow down: Enter brightness adjustment, decrease value of the adjustment

More information

Copyright 2018 Xi an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced, extracted or transmitted in any

Copyright 2018 Xi an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced, extracted or transmitted in any MCTRL4K Independent Controller Product Version: Document Number: V1.0.3 NS110100428 Copyright 2018 Xi an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced,

More information

R-1580A Microwave Downconverter. Product Brochure

R-1580A Microwave Downconverter. Product Brochure R-1580A Microwave Downconverter Product Brochure Jan 2018 Highlights The DSII Model R-1580A Microwave Downconverter extends the coverage of the R-1550A, or other DSII wide range receivers, to 22 GHz. The

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam

More information

User Manual. Innovative LCD Display Solutions AP-20 Series DP-20 Series OP-20 Series NAP-20 Series

User Manual. Innovative LCD Display Solutions AP-20 Series DP-20 Series OP-20 Series NAP-20 Series Innovative LCD Display Solutions AP-20 Series DP-20 Series OP-20 Series NAP-20 Series 1. Table of Content 1. Table of Content P.1 2. Introduction 3. Installation A) Overview P.2 B) Features P.2 C) Dimension

More information

Intelligent Security and Fire Ltd

Intelligent Security and Fire Ltd User Manual Product ranges covered by this manual Vi-P14 Vi-P14A Document Reference Date Firmware Vi-Q4C1 Viq601a.doc 26/11/2009 From Viq001a21 Videoswitch Telephone 01252-851510 Ocean House, Redfields

More information

Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer

Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer Table of Contents Package Contents............ 1 Warranty............ 1 Installation............

More information

SERVICE AND OPERATION MANUAL

SERVICE AND OPERATION MANUAL MTG-XX02XT/XN Publication A. Issue 1 SERVICE AND OPERATION MANUAL MTG- XX02XT/XN OPEN FRAME XVGA COLOR MONITORS MTG-1702XT/XN : 17INCH, FST MTG-1902XT/XN : 19INCH, FST Information in this publication current

More information

MS2540 Current Loop Receiver with RS485 Communication

MS2540 Current Loop Receiver with RS485 Communication MS2540 Current Loop Receiver with RS485 Communication User Manual Metal Samples Company A Division of Alabama Specialty Products, Inc. 152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358 4202 Fax:

More information

Operations. BCU Operator Display BMTW-SVU02C-EN

Operations. BCU Operator Display BMTW-SVU02C-EN Operations BCU Operator Display BMTW-SVU02C-EN Operations BCU Operator Display Tracer Summit BMTW-SVU02C-EN June 2006 BCU Operator Display Operations This guide and the information in it are the property

More information

SDC-500 Digital Optical Chopper Operating Instructions

SDC-500 Digital Optical Chopper Operating Instructions SDC-500 Digital Optical Chopper Operating Instructions Gentec-EO USA, Inc. 5825 Jean Road Lake Oswego, OR 97035 Phone: 503-697-1870 Fax: 503-697-0633 ddooley@gentec-eo.com www.gentec-eo.com SDC-500 Manual

More information

Introduction...2. Features...2 Safety Precautions...2. Installation...4

Introduction...2. Features...2 Safety Precautions...2. Installation...4 PE1900 Contents Introduction...2 Features...2 Safety Precautions...2 Installation...4 Unpacking the Display...4 Locations and Functions of Controls...4 Connections...5 Using Your Display...7 Turning the

More information

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE ? 8 LCD Screen? 10.4 LCD Screen LCD Zoom Microscope Instruction Manual Please read the Instruction Manual carefully before installation and keep it for future

More information

CM-S23349SV. Vari-Focal IR Bullet Camera

CM-S23349SV. Vari-Focal IR Bullet Camera Vari-Focal IR Bullet Camera User s Guide CM-S23349SV SAFETY PRECAUTIONS WARNING 1. Be sure to use only the standard adapter that is specified in the specification sheet. Using any other adapter could cause

More information

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications Technical Reference TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications 070-8990-04 Please check for change information at the rear of this manual. Fifth

More information

D-ILA PROJECTOR DLA-G15 DLA-S15

D-ILA PROJECTOR DLA-G15 DLA-S15 D-ILA PROJECTOR Outstanding Projection Im Breakthrough D-ILA projector offers high-contrast 350:1, 1500 ANSI lumen brightness and S-XGA resolution Large-size projection images with all the sharpness and

More information

Tablet Oscilloscope Quick Guide

Tablet Oscilloscope Quick Guide Tablet Oscilloscope Quick Guide For tbook Series Shenzhen Micsig Instruments Co., Ltd. Copyright Copyright Shenzhen Micsig Instruments Co., Ltd. All Rights Reserved. Version Version: MKX2014-001; Product

More information

MG-XV operating instruction. Measuring of norm signals, 4-8-digit. Panel instrument type MG-BV Construction instrument type MG-AV

MG-XV operating instruction. Measuring of norm signals, 4-8-digit. Panel instrument type MG-BV Construction instrument type MG-AV MG-XV operating instruction Measuring of norm signals, 4-8-digit Panel instrument type MG-BV Construction instrument type MG-AV Contents 1. Brief description... 3 2. Safety instructions... 3 2.1. Proper

More information

SQM40/41 Actuators for air and gas dampers

SQM40/41 Actuators for air and gas dampers SQM40/41 Actuators for air and gas dampers Description SQM40/41 actuators are used for the positioning of flow control valves, butterfly valves, dampers or any application requiring rotary motion. The

More information

SHOWLINE SL BAR 640 LINEAR WASH LUMINAIRE SPECIFICATIONS.

SHOWLINE SL BAR 640 LINEAR WASH LUMINAIRE SPECIFICATIONS. GENERAL. A.) Overview. SHOWLINE SL BAR 640 LINEAR WASH LUMINAIRE SPECIFICATIONS. 1.) The luminaire shall be a color mixing luminaire employing twenty-four (24) red, green, blue, and white LED engines.

More information

ACCESSORIES MANUAL PART NUMBER: PRODUCT REVISION: 1 TNP100. Tilt N Plug Interconnect Box USER'S GUIDE

ACCESSORIES MANUAL PART NUMBER: PRODUCT REVISION: 1 TNP100. Tilt N Plug Interconnect Box USER'S GUIDE MANUAL PART NUMBER: 400-0091-001 PRODUCT REVISION: 1 TNP100 Tilt N Plug Interconnect Box USER'S GUIDE INTRODUCTION Your purchase of the TNP100 Tilt N Plug Interconnect Box is greatly appreciated. We are

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

Specifications. Independent Controller MCTRL R5. Rev1.0.0 NS

Specifications. Independent Controller MCTRL R5. Rev1.0.0 NS Specifications Independent Controller MCTRL R5 Rev1.0.0 NS1601000126 Overview MCTRL R5 is an independent master controller developed by NovaStar with an epoch-making significance. The loading capacity

More information

PROFESSIONAL D-ILA PROJECTOR DLA-G11

PROFESSIONAL D-ILA PROJECTOR DLA-G11 PROFESSIONAL D-ILA PROJECTOR DLA-G11 A new digital projector that projects true S-XGA images with breakthrough D-ILA technology Large-size projection images with all the sharpness and clarity of a small-screen

More information

KVANT. Spectrum 1,6W RGB laser system. Product datasheet

KVANT. Spectrum 1,6W RGB laser system. Product datasheet KVANT Spectrum 1,6W RGB laser system Product datasheet SPECTRUM series professional full colour laser system 1. DESCRIPTION Kvant Spectrum is professional laser system with wide range of use, mainly for

More information

PROFESSIONAL D-ILA PROJECTOR DLA-G11

PROFESSIONAL D-ILA PROJECTOR DLA-G11 PROFESSIONAL D-ILA PROJECTOR DLA-G11 A new digital projector that projects true S-XGA images with breakthrough D-ILA technology Large-size projection images with all the sharpness and clarity of a small-screen

More information

Operating Instructions

Operating Instructions Operating Instructions LCDRV700 Digital LCD Color Monitor Please read this manual thoroughly before operating the unit, and keep it for future reference. V1.0 Contents 1. Precautions 2. Features 1 3 3.

More information

Operation Manual VMS 3.0 Video System

Operation Manual VMS 3.0 Video System Operation Manual VMS 3.0 Video System for the AlterG Anti-Gravity Treadmill 1 This manual covers operation procedures for the following AlterG products: AlterG Video System model VMS 3.0 NOTE: The following

More information

D-ILA PROJECTOR DLA-G15 DLA-S15

D-ILA PROJECTOR DLA-G15 DLA-S15 D-ILA PROJECTOR DLA-G15 Outstanding Projection Im Breakthrough D-ILA projector offers high-contrast 350:1, 1500 ANSI lumen brightness and S-XGA resolution Large-size projection images with all the sharpness

More information

OFI Optical Fiber Identifier User s Guide

OFI Optical Fiber Identifier User s Guide OFI Optical Fiber Identifier User s Guide OFI Optical Fiber Identifier User s Guide 2002-2008, AFL Telecommunications, all rights reserved. OFI1-11-1000 Revision A, 05.02.08 Specifications are subject

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

TR6102HD HDTV/DVD/COMPONENT VIDEO TO RGBHV TRANSCODER USER S GUIDE

TR6102HD HDTV/DVD/COMPONENT VIDEO TO RGBHV TRANSCODER USER S GUIDE MANUAL PART NUMBER: 400-0031-003 PRODUCT REVISION: 1 HDTV/DVD/COMPONENT VIDEO TO RGBHV TRANSCODER USER S GUIDE INTRODUCTION Thank you for your purchase of the Transcoder. We are certain that you will find

More information

Model 1476-C SuperQuad HR

Model 1476-C SuperQuad HR Model 1476-C SuperQuad HR Installation and Operating Instructions Table of Contents Page Table of Content... 2 System Description... 3 Features... 3 Installation... 4 Internal Setups... 4 Connections...

More information

LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras

LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras Eng Installation Instructions F D E NL I IMPORTANT SAFEGUARDS 1. Read Instructions All the safety and operating instructions should be read before

More information

K Service Source. Macintosh Color Display

K Service Source. Macintosh Color Display K Service Source Macintosh Color Display K Service Source Specifications Macintosh Color Display Specifications Characteristics - 1 Characteristics Picture Tube 14-in. diagonal (11.5-in. viewable image)

More information

Connection for filtered air

Connection for filtered air BeamWatch Non-contact, Focus Spot Size and Position monitor for high power YAG, Diode and Fiber lasers Instantly measure focus spot size Dynamically measure focal plane location during start-up From 1kW

More information

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303)

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303) Spectroscopy Module Vescent Photonics, Inc. www.vescentphotonics.com 4865 E. 41 st Ave Denver, CO 80216 Phone: (303)-296-6766 Fax: (303)-296-6783 General Warnings and Cautions The following general warnings

More information

SERVICE AND OPERATION MANUAL

SERVICE AND OPERATION MANUAL MTG-2907TN Publication A. Issue 1 SERVICE AND OPERATION MANUAL MTG- 2907TN OPEN FRAME VGA COLOR MONITORS MTG-2907TN : 29INCH, FST Information in this publication current as of Jun, 2003. Information subject

More information

AUTO-LIT. vs advertisers. Catalogue of OLUMINESCENT COMMERCIAL PURPOSE.

AUTO-LIT. vs advertisers. Catalogue of OLUMINESCENT COMMERCIAL PURPOSE. www.vsadvertisers.com vs advertisers AUTO-LIT OLUMINESCENT P U U AND COMMERCIAL PURPOSE Catalogue of S ADD: Plot No-24, Street No-7, Second Floor, Surat Nagar Industrial Area Near Classic Gym, Under Daultabad

More information

Micro-DCI 53ML5100 Manual Loader

Micro-DCI 53ML5100 Manual Loader Micro-DCI 53ML5100 Manual Loader Two process variable inputs Two manually controlled current outputs Multiple Display Formats: Dual Channel Manual Loader, Single Channel Manual Loader, Manual Loader with

More information

1928L Multifunction 19-inch Desktop Touchmonitor

1928L Multifunction 19-inch Desktop Touchmonitor 1928L Multifunction 19-inch Desktop Touchmonitor The fully featured 1928L touchmonitor offers best in class optical performance mated to the leading touch technologies surface acoustical wave and five-wire

More information

EN - English Washington Street Melrose, MA Phone Toll Free Revision 4 20/06/17

EN - English Washington Street Melrose, MA Phone Toll Free Revision 4 20/06/17 - English... 1 Instruction Manual Vortex Mixer, Mini Fix Speed, VXMNFS Vortex Mixer, Mini Analog, VXMNAL Vortex Mixer, Mini Digital, VXMNDG Vortex Mixer, Mini Pulsing, VXMNPS 99 Washington Street Melrose,

More information

instruction manual x-y indicator model 620D

instruction manual x-y indicator model 620D instruction manual x-y indicator model 620D s/n boulder, colorado WARNING This equipment generates, uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction

More information

Tube Roller Shakers. User Guide. Version 1.2

Tube Roller Shakers. User Guide. Version 1.2 Tube Roller Shakers User Guide Version 1.2 Control panel Rollers Side retaining panels Analog models LED display Drip tray (not visible) Digital models Power On/Off and control dial Roller retaining panel

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

OPERATOR MANUAL OSD8865 DIGITAL TRIPLE VIDEO FIBER OPTIC RECEIVER

OPERATOR MANUAL OSD8865 DIGITAL TRIPLE VIDEO FIBER OPTIC RECEIVER OPERATOR MANUAL OSD8865 DIGITAL TRIPLE VIDEO FIBER OPTIC RECEIVER INDEX 1 1 TECHNICAL SUMMARY... 4 1.1 BRIEF DESCRIPTION... 4 1.1.1 OVERVIEW... 4 1.1.2 APPLICATIONS... 4 1.1.3 FEATURES AND BENEFITS...

More information

Standard Operating Procedure: Dicing Saw

Standard Operating Procedure: Dicing Saw Contents Hardware Description and Principle of Operation... 1 Procedure... 2 Emergency Stop... 12 Allowed Activities... 13 Disallowed Activities... 13 What to watch out for during operation... 13 Common

More information

DA1909 COMPUTER VIDEO LINE DRIVER WITH EQUALIZATION USER S GUIDE

DA1909 COMPUTER VIDEO LINE DRIVER WITH EQUALIZATION USER S GUIDE MANUAL PART NUMBER: 400-0108-002 PRODUCT REVISION: 1 COMPUTER VIDEO LINE DRIVER WITH EQUALIZATION USER S GUIDE INTRODUCTION Altinex appreciates your purchase of the Line Driver. We are sure you will find

More information

High Resolution Multicolor Contrast Scanner. Dimensioned drawing

High Resolution Multicolor Contrast Scanner. Dimensioned drawing Specifications and description KRTM 20 High Resolution Multicolor Contrast Scanner Dimensioned drawing en 01-2011/06 50116669 12mm 20mm 50mm 12-30 V DC 50 / 25 khz We reserve the right to make changes

More information

Outdoor IR Audio Camera

Outdoor IR Audio Camera Outdoor IR Audio Camera User s Guide CM-S22326BW-AD SAFETY PRECAUTIONS WARNING 1. Be sure to use only the standard adapter that is specified in the specification sheet. Using any other adapter could cause

More information

Industrial Diode Laser (IDL) System IDL Series

Industrial Diode Laser (IDL) System IDL Series COMMERCIAL LASERS Industrial Diode Laser (IDL) System IDL Series Key Features Round, top-hat beam profile for uniform power distribution Warranted for full rated power in either pulsed or continuous wave

More information

PT-DW750W/DW750LW PT-DW750B/DW750LB

PT-DW750W/DW750LW PT-DW750B/DW750LB S P E C F I L E Product Number : Product Name : DLP Projectors PT-DW750W/DW750LW PT-DW750B/DW750LB. Specifications and appearance are subject to change without notice. 1/18 Specifications Main unit Power

More information

IVS. Integrated Video-Assist System Instruction Manual

IVS. Integrated Video-Assist System Instruction Manual IVS Integrated Video-Assist System Instruction Manual Mini monitor connector Switch ON/OFF/Keyboard locked White balance (Indoor/Outdoor/Automatic) Manual gain control (ON) Mini monitor ( normal video

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

User Guide. HDMI Fiber Optic Extender. DVI-7350a

User Guide. HDMI Fiber Optic Extender. DVI-7350a User Guide HDMI Fiber Optic Extender DVI-7350a Table of Contents Section Page Product Safety.................................... 1 1.0 Introduction...2 2.0 Specifications...3 3.0 Package Contents...3 4.0

More information

HIGH DEFINITION MONITOR HDM 5049 PLUS USER S MANUAL

HIGH DEFINITION MONITOR HDM 5049 PLUS USER S MANUAL BARCO N.V. Communication Systems Th. Sevenslaan 106 B-8500 Kortrijk (Belgium) Tel.: +32.56.233.211 Fax: + 32.56.233.461 E-mail: support.bcs@barco.com Web site: http://www.barco.com HIGH DEFINITION MONITOR

More information

Processing data with Mestrelab Mnova

Processing data with Mestrelab Mnova Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

Installation and Operation Manual Next Generation On-Camera Prompter Displays

Installation and Operation Manual Next Generation On-Camera Prompter Displays Installation and Operation Manual Next Generation On-Camera Prompter Displays Model CSF10 Prompter Contents 1 CueScript CSF10 Prompter... 3 2 CueScript Prompter Features... 3 3 Display Technical Specifications...

More information

USER S MANUAL CCTV LED MONITOR MODEL: ADE-117N1 ADE-119N1 ADE-118W1 ADE-121W1 ADE-124W Atherton Electronics Corp. All rights reserved.

USER S MANUAL CCTV LED MONITOR MODEL: ADE-117N1 ADE-119N1 ADE-118W1 ADE-121W1 ADE-124W Atherton Electronics Corp. All rights reserved. USER S MANUAL CCTV LED MONITOR MODEL: ADE-117N1 ADE-119N1 ADE-118W1 ADE-121W1 ADE-124W1 2015 Atherton Electronics Corp. All rights reserved. TABLE OF CONTENTS FCC information -------------------------------------------------------------------

More information

HIGH RESOLUTION MONITOR

HIGH RESOLUTION MONITOR HIGH RESOLUTION MONITOR 29" Pure Flat 15K-40Khz Installation Instruction WARNING Primary side and Secondary side The monitor's circuit which is divided into the primary side and secondary side, is insulated.

More information

QuickSpecs. Models RB146AA#ABA Standard Configuration RB146AT#ABA Promotional Part Number (SmartBuy)

QuickSpecs. Models RB146AA#ABA Standard Configuration RB146AT#ABA Promotional Part Number (SmartBuy) Overview 1. 2. 3. 4. 5. Power: Turns the unit on and off. Select: Selects the adjustment items from the on-screen display (OSD) menus. Arrow down: Enter brightness adjustment, decrease value of the adjustment

More information

Contacts: English Department Office 238 Moreland Media Services 109 Kidder

Contacts: English Department Office 238 Moreland Media Services 109 Kidder Contacts: English Department Office 238 Moreland 7-3244 Media Services 109 Kidder September 2006 Welcome...3 Starting Out...5 Unlocking the Station...5 Touch Panel Operation...5 Projector...6 Selecting

More information

BROTHER, BROTHER & SONS

BROTHER, BROTHER & SONS AREA SOFT BROTHER, BROTHER & SONS ApS INTRO Brother, Brother & Sons ApS The information in this manual is subject to change without notice. Go to www.brothers-sons.dk for the latest version. Brother, Brother

More information

SK2002DA SIDEKICKER 1-IN, 2-OUT VGA-UXGA DISTRIBUTION AMPLIFIER CABLE USER S GUIDE DISTRIBUTION AMPLIFIERS

SK2002DA SIDEKICKER 1-IN, 2-OUT VGA-UXGA DISTRIBUTION AMPLIFIER CABLE USER S GUIDE DISTRIBUTION AMPLIFIERS MANUAL PART NUMBER: 400-0152-001 PRODUCT REVISION: 0 SK2002DA SIDEKICKER 1-IN, 2-OUT VGA-UXGA DISTRIBUTION AMPLIFIER CABLE USER S GUIDE TABLE OF CONTENTS Page PRECAUTIONS / SAFETY WARNINGS...2 GENERAL...2

More information

-Technical Specifications-

-Technical Specifications- Annex I to Contract 108733 NL-Petten: the delivery, installation, warranty and maintenance of one (1) X-ray computed tomography system at the JRC-IET -Technical Specifications- INTRODUCTION In the 7th

More information