Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

Similar documents
Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent 1260 Infinity Clinical ed. LC Site Preparation Checklist

If you have questions or problems in providing anything described as Customer

Model KBF 115 Constant climate chambers with large temperature / humidity range

Liquid Chromatography- Mass Spectrometer Manual

Model KBF 1020 Constant climate chambers with large temperature / humidity range

Model KT 115 Cooling incubators with thermoelectric cooling

Model KT 53 Refrigerated incubators with thermoelectric refrigeration

Model KT 53 Cooling incubators with thermoelectric cooling

Model FP 240 Drying and heating chambers Classic.Line with forced convection and program functions

Model KBF P 240 Humidity test chambers with ICH-compliant light source

ICS-3000 System Installation Qualification

INSTALLATION REQUIREMENTS - THERMAL ANALYSIS & RHEOLOGY SYSTEMS

Model CB 60 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

Model CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications

Agilent High Capacity RV Pumps & Roots Pumping Systems

Model ED 400 Drying and heating chambers Classic.Line with natural convection

Model KBF 720 Constant climate chambers with large temperature / humidity range

Model ED 23 Drying and heating chambers Classic.Line with natural convection

HIGH VACUUM PUMPS. Hybrid Turbomolecular Pumps. ATH series. Adixen by Alcatel Vacuum Technology

CTC Series. Compact Temperature Calibrator

Model KT 170 Cooling incubators with thermoelectric cooling

Control Unit CU 2.1 User Manual V6850A HPLC

ACP 15 / 28 / 40 SERIES 2 COMPACT DRY ROUGHING PUMPS

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

Staying Connected in GC

Operating pressure should be 6000 psi or better suitable for both analytical and semi-preparative applications

Optical Cryostat - Omniplex TM

Flash-Point with Closed Cup - Abel Method - ABA 4 (automatic)

Performance. Reliability. Productivity. Automated Flash Chromatography Systems

A3H series. Energy and cost efficient solutions for the most demanding applications. Multi-stage. Roots pumps

InSight Pro. Valuable real-time process insight

UNIGAS Hand-held combustion gas analyzer with 3 sensors

Industrial Diode Laser (IDL) System IDL Series

Model KMF 240 Constant climate chambers with expanded temperature / humidity range

Submittal. Downflow Two Stage Condensing Gas Fired Furnace 120,000 BTUH S9V2D120D5-SUB-1-EN. March 2016 S9V2D120D5PSAA TAG:

THE ULTIMATE DRY CHOICE

OTHER FEATURES UNIQUE FEATURES

EVF 300 series. Controllers for electric bread and pizza ovens, with touch-keys, in split version and which can be integrated into the unit.

ZN-PD. Smallest Air Particle Sensor in the Industry for In-line Measurement. Air Particle Sensor. Features

Detailed Design Report

Basic Vortex Mixer Standard Vortex Mixer Advanced Vortex Mixer Pulsing Vortex Mixer

Product Data Sheet 8506 VW-691

User s Manual. LG Industrial Systems G3F AT4A. LG Programmable Logic Controller

Upflow/ Horizontal Left/Right, Downflow Two Stage Condensing Gas Fired Furnace

ih Series Drypumps (505) idealvac.com idealvac.com

Infra-Red Vortex-Evaporator. User manual Version 3.4

FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091

Product Data Sheet 4656 EZ

Product Data Sheet 4656 ZW

AGILENT ROUGH VACUUM PUMPS

Optical Cryostat - Economy

Duo 65 C, 3-phase motor, 3TF, 230/400 V, 50 Hz 265/460 V, 60 Hz

SmartMix Manual V5350A

PAL LSI - Robust and Reliable Sample Handling

The extremely compact laser head is approximately 480 mm long and can

TABLE OF CONTENTS Important Safety Instructions Package Content Setting Up the Display Trouble shooting Specifications Product Dimensions

Dry Compressing Vacuum Pumps. Dry Com. CLAWVAC Claw Vacuum and Overpressure Pumps

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

SOGEVAC. Rotary Vane Vacuum Pumps single-stage, oil-sealed, m 3 x h -1 ( cfm)

Operation Manual for the Refractive Index detector VE-3580

TECHNICAL GUIDE 95.5% AFUE SINGLE STAGE RESIDENTIAL GAS FURNACES MULTI-POSITION MODELS: TG9S

Cooling systems for ORBIMAT

IKA Analytical line Calorimeters C 2000

STANDARD OPERATING PROCEDURE:

TEK-LCD 7801A. NEMA 4X Loop-Powered Feet and Inches Level Meter. ACCESSORIES. Technology Solutions

Duo 65 MC, 3-phase motor, 3TF, 230/400 V, 50 Hz 265/460 V, 60 Hz

General Specifications

Thermo Scientific Merlin Recirculating Chillers

Thermo Scientific NESLAB ThermoFlex Recirculating Chillers

Modular Lube Lubrication Systems System Controls

A PASSION FOR PERFECTION. OktaLine. The Power Pump Portfolio. Roots Pumps for Every Low and Medium Vacuum Application.

USER MANUAL. WARNING Read the instructions before using the machine. EN (Original Instruction) / 1704

Section C Recorders. DPR mm Digital Strip Chart Recorder

ROTARY EVAPORATOR BOECO ROTARY EVAPORATOR RVO 400 SD

Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer

Sentronic PLUS Electronic Pressure Regulator. Installation Manual

Forced Convection Laboratory Ovens

ABB MEASUREMENT & ANALYTICS. MB3600 The most reliable FT-NIR laboratory analyzer designed for QA/QC

PROPORTIONAL VALVES Sentronic plus

TWIN-PUMP. Dual line system. Designed to work all the day, every day in extreme condition and difficult environments

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

Cylindrical Photoelectric Sensor CY-100 SERIES

Application Note [AN-007] LCD Backlighting Technologies and Configurations

SMALL OIL SEALED PUMPS

Küba comfort DP. Draught-free ventilation and extremely quiet operation

GCxGC Comprehensive Two-Dimensional Gas Chromatography

PAL LSI - Robust and Reliable Sample Handling

DH551C/DH550C/DL550C Double Sided Display User Manual

Label Applicator HERMA 400

AP-40. AP-40 Series Features Industry s smallest-sensor head Ultra lightweight High-speed response Two-color LED digital pressure display

Tekna evo. innovation technology future 03

New Eppendorf Freezers Sample security meets the latest in energy efficiency. Stay Cool. CryoCube ULT Freezers: Efficiency has a new name

SPECIFICATION NO Model 207 Automatic GTAW Welding System

Prisma Optical Networks Ancillary Modules

Okta 1000, Roots pump without motor

Duo 65 M, 3-phase motor, 3TF, 230/400 V, 50 Hz; 265/460 V, 60 Hz

Lite USER GUIDE. 3 Lite. For more information visit

MICROPROCESSOR CALCULATION MODULE - S2000

Transcription:

1290 Site Preparation Infinity LC/1260 Checklist Infinity LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment for your site. For additional information about our solutions, please visit our web site at http://www.chem.agilent.com/en-us/pages/homepage.aspx Customer Responsibilities Make sure your site meets the following prior to the installation date using the checklist below. For details, see specific sections within this document, including: The necessary laboratory or bench space is available. The environmental conditions for the lab as well as laboratory gases, tubing. The power requirements related to the product (e.g. number & location of electrical outlets). The required operating supplies necessary for the product and installation. Please consult Other/Special Requirements section below for other product-specific information. If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Important Customer Information 1 If you have questions or problems in providing anything described as Customer Responsibilities above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or its partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory. 2 Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to re- arrange any services that have been purchased. 3 Other optional services such as additional training, operational qualification (OQ) and consultation for user- specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately. Page 1 of 21

Module List Module G1170A G1310B G1311B G1311C G1312B G1312C G1314B G1314C G1314E G1314F G1315C G1315D G1316A G1316C G1321B G1321C G1322A G1329B G1330B G1361A G1362A G1364B G1364C G1364D G1365C G1365D G1367E G1376A G1377A Instrument Description 1290 Infinity Valve Drive and Valve Heads 1260 Infinity Isocratic Pump 1260 Infinity Quatenary Pump 1260 Infinity Quaternary Pump VL 1260 Infinity Binary Pump 1260 Infinity Binary Pump VL 1260 Infinity Variable Wavelength Detector VL 1260 Infinity Variable Wavelength Detector VL+ 1290 Infinity Variable Wavelength Detector 1260 Infinity Variable Wavelength Detector 1260 Infinity Diode Array Detector VL+ 1260 Infinity Diode Array Detector VL 1260 Infinity Thermostatted Column Compartment 1290 Infinity Thermostatted Column Compartment 1260 Infinity Fluorescence Detector 1260 Infinity Fluorescence Detector 1260 Infinity Standard Degasser 1260 Infinity Standard Autosampler 1290 Infinity Thermostat 1260 Infinity Preparative Pump 1260 Infinity Refractive Index Detector 1260 Infinity Preperative-scale Fraction Collector 1260 Infinity Analytical-scale Fraction Collector 1260 Infinity Micro-scale Fraction Collector 1260 Infinity Multiple Wavelength Detector 1260 Infinity Multiple Wavelength Detector VL 1260 Infinity High Performance Autosampler 1260 Infinity Capillary Pump 1260 Infinity High Performance Micro Autosampler Page 2 of 21

Module G1379B G1390B G1968F G2226A G2257A G2258A G2260A G4204A G4208A G4212A G4212B G4220A G4220B G4225A G4226A G4227A G4240A G4260A G4260B G4261A G4261B G4277A G4278A G5611A G5664A G5667A Instrument Description 1260 Infinity Micro Degasser 1200 Infinity Series Universal Interface Box Active Splitter 1260 Infinity Nanoflow Pump 1260 Infinity Sample Capacity Extension 1260 Infinity Dual-Loop Autosampler 1260 Infinity Preparative Autosampler 1290 Infinity Quaternary Pump 1200 Infinity Series Instant Pilot 1290 Infinity Diode Array Detector 1260 Infinity Diode Array Detector 1290 Infinity Binary Pump 1290 Infinity Binary Pump 1260 Infinity High Performance Degasser Module 1290 Infinity Autosampler 1290 Infinity Flexible Cube 1260 Infinity Chip Cube MS Interface Agilent 380-ELSD 1260 Infinity Evaporative Light Scattering Detector Agilent 385-ELSD 1290 Infinity Evaporative Light Scattering Detector 1290 Infinity LC Injector HTS 1290 Infinity LC Injector HTC 1260 Infinity Bio-inert Quaternary Pump 1260 Infinity Bio-inert Analytical-scale Fraction Collector 1260 Infinity Bio-inert High-Performance Autosampler Page 3 of 21

Dimensions and Weight Identify the laboratory bench space before your system arrives based on the table below. Pay special attention to the total height and total weight requirements for all system components you have ordered and avoid bench space with overhanging shelves. Also pay special attention to the total weight of the modules you have ordered to ensure your laboratory bench can support this weight. Special Notes 1 The thermostatted version of all samplers and the fraction collector include the G1330B thermostat module. The thermostat module must be placed directly under the sampler or the fraction collector to be thermostatted. It is recommended that the thermostat module is positioned as the bottom module of the stack, directly on the laboratory bench. Any stack containing a G1330B thermostat module needs at least 25 cm (10 inches) of space on either side to guarantee proper ventilation. Instrument Description Weight Height Depth Width kg lbs cm in cm in cm in G1322A, G1379B 7.5 16.5 8 3 43.5 17 34.5 13.5 G4225A 5 11 8 3.1 43.5 17 34.5 13.5 G1310B 11 25 18 7 43.5 17 34.5 13.5 G1311B/C, G5611A 14.5 32 18 7 43.5 17 34.5 13.5 G1312B/C 15.5 34 18 7 43.5 17 34.5 13.5 G1376A, G2226A 17 39 18 7 43.5 17 34.5 13.5 G1361A 15 32.9 20 8 43.5 17 34.5 13.5 G4220A/B 21.8 48 24 9.3 43.5 17 34.5 13.5 G4204A 15.2 33.4 20 8 43.5 17 34.5 13.5 G2258A 14 29.8 20 8 43.5 17 34.5 13.5 G1329B, G2260A 14.2 31.3 20 8 43.5 17 34.5 13.5 G1367E, G1377A, G4226A, G5667A 15.5 34.2 20 8 43.5 17 34.5 13.5 G4277A 10 22.8 64.8 25.5 38.5 15.1 82.8 32.6 G4278A 8 18.2 64.8 25.5 38.5 15.1 54.4 21.5 G1364B/C/D, G5664A 17 38 18 7 43.5 17 34.5 13.5 G4240A 14 31 36 14.1 30 11.7 35 13.7 Page 4 of 21

Instrument Description Weight Height Depth Width G1330B 18.5 40.7 14.4 5.5 43.5 17 34.5 13.5 G4227A 8.2 18.1 14.4 5.5 43.5 17 34.5 13.5 G1316A/C 11.2 22.5 14 5.5 43.5 17 41 16 G1314B/C/E/F 11 25 14 5.5 43.5 17 34.5 13.5 G1315C/D, G1365C/D 11.5 26 14 5.5 43.5 17 34.5 13.5 G1321B/C 11.5 25.4 14 5.5 43.5 17 34.5 13.5 G4212A/B 11.5 26 14 5.5 43.5 17 34.5 13.5 G1362A 17 38 18 7 43.5 17 34.5 13.5 G4260A/B 11 24.2 41.5 16.3 45 17.7 20 7.9 G4261A/B 13 28.6 41.5 16.3 45 17.7 20 7.9 G1390B 0.9 2 16.5 6.5 5.5 2.2 13.5 5.3 G1968F 2.1 4.6 14.1 5.6 19.1 7.5 7.8 3.1 Page 5 of 21

Environmental Conditions Operating your instrument within the recommended temperature ranges insures optimum instrument performance and lifetime. Special Notes 1 Performance can be affected by sources of heat and cold, e.g. direct sunlight, heating/cooling from air conditioning outlets, drafts and/or vibrations. 2 The site s ambient temperature conditions must be stable for optimum performance. 3 The following table summarizes some key physical specifications. For the complete set of physical specifications, please refer to the corresponding module manual. Instrument Description G1314B/C/E/F, G1315C/D, G1316A/C, G1322A, G1365C/D, G1362A, G4220A, G4225A, G4226A Operating temp range C (F) Operating humidity range (%) 0 55 C (32 131 F), constant temperature. < 95 %, non-condensing G1321B/C 0 40 C (32 104 F), constant temperature. < 95 %, non-condensing G4212A/B 4 40 C (39 104 F), constant temperature. < 80 %, non-condensing G1379B 0 45 C (32 113 F) < 95 %, non-condensing G1330B, G1361A, G1364B/C/D, G5664A, G2258A, G4212B G4260A/B, G4261A/B 4 40 C (39 104 F) < 95 %, non-condensing 10 35 C (50 95 F) < 95 %, non-condensing G4240A 5 40 C (41 104 F) < 80 %, non-condensing G4277A, G4278A 4 40 C (39 104 F) < 75 %, non-condensing G1968F 0 40 C (32 104 F), constant temperature 80 % r.h. up to 31 C, decreasing to 50 % at 40 C, non-condensing All other modules 4 55 C (39 131 F), constant temperature < 95 % r.h. at 40 C, non-condensing Page 6 of 21

Power Consumption Special Notes: 1 If a computer system is supplied with your instrument, be sure to account for those electrical outlets. 2 The heat dissipation can be calculated from the the active power, using the following equation: 1 W = 3.413 BTU/h Instrument Description Line Voltage & Frequency (V, Hz) Maximum Power Consumption (VA) Maximum Power Consumption (W) G1379B, G1322A, G4225A 100 240 V(AC), 50 or 60 Hz 30 VA 30 W G1310B 100 240 V(AC), 50 or 60 Hz 180 VA 55 W G1311B/C, G5611A 100 240 V(AC), 50 or 60 Hz 180 VA 110 W G1312B/C 100 240 V(AC), 50 or 60 Hz 220 VA 74 W G1376A, G2226A 100 240 V(AC), 50 or 60 Hz 220 VA 75 W G1361A 100 240 V(AC), 50 or 60 Hz 250 VA 210 W G4220A/B 100 240 V(AC), 50 or 60 Hz 350 VA 270 W G4204A 100 240 V(AC), 50 or 60 Hz 220 VA 180 W G1329B, G2260A, G1367E, G1377A, G4226A, G5667A 100 240 V(AC), 50 or 60 Hz 300 VA 200 W G2258A 100 240 V(AC), 50 or 60 Hz 260 VA 210 W G4277A, G4278A 100 240 V(AC), 50 60 Hz N/A 120 W G1364D 100 240 V(AC), 50 or 60 Hz 200 VA 180 W G5664A 100 240 V(AC), 50 or 60 Hz 180 VA 180 W G4240A 100 240 V(AC), 50 or 60 Hz N/A 80 W G1330B 100 240 V(AC), 50 or 60 Hz 260 VA 210 W G4227A 100 240 V(AC), 50 or 60 Hz 80 VA 45 W G1316A/C 100 240 V(AC), 50 or 60 Hz 320 VA 150 W G1314B/C/E/F 100 240 V(AC), 50 or 60 Hz 220 VA 85 W G1315C/D, G1365C/D 100 240 V(AC), 50 or 60 Hz 160 VA 160 W G4212A/B 100 240 V(AC), 50 or 60 Hz 160 VA 130 W Page 7 of 21

Instrument Description Line Voltage & Frequency (V, Hz) Maximum Power Consumption (VA) G1362A 100 240 V(AC), 50 or 60 Hz 160 VA 65 W G1321B/C 100 240 V(AC), 50 or 60 Hz 180 VA 70 W G4260A/B, G4261A/B 100 120 V(AC) or 220 240 V, 50 or 60 Hz N/A 150 W G1390B 100 240 V(AC), 50 or 60 Hz 140 VA 65 W G1968F 100 240 V(AC), 50 or 60 Hz 93 111 VA N/A Maximum Power Consumption (W) Page 8 of 21

Required Operating Supplies by Customer Special Notes: For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-us/products/consumables/pages/default.aspx Page 9 of 21

Other/Special Requirements Gas requirements G4260B, G4261B A supply of inert gas (typically nitrogen) is required to operate the detector. The gas supply needs to be free of oil, humidity and particles, as such contaminations will create background noise in the chromatograms and may damage the built- in pressure sensor. In case of such noise for example for newly installed gas lines, flush the gas lines for sufficient time (might take days) and use additional filters of 0.5 µm or less. The typical gas pressure is 4 bar (60 psi) and must be set by an external pressure regulator. Pure gas is not required as the gas is only used as a carrier for the solid sample particles. The gas inlets of the detector have an outer diameter of 4 mm (0.157 inches). The lab installation must therefore allow the installation of a tubing with 4 mm (0.157 inches) outer diameter. Gas consumption is typically 0.9 SLM to 3.25 SLM, depending on the detector settings. Item description, (including dimensions etc) G4260A/B, G4261A/B ELSD Gas Nitrogen (typical) Vendor/Part Number (if applicable) N/A Recommended quantity N/A Solvent requirements Customer should have available HPLC grade Acetonitrile and water with a dry residue below 1 ppm or MS grade solvents. Exhaust venting and drain requirements The exhaust from the detector must be directed into a fume hood or exhaust vent. If a vacuum is used, it should be moderate so as to avoid turbulence in the glass cell siphon. The potentially hazardous exhaust of evaporated solvent and sample must not be allowed to enter the laboratory atmosphere and any appropriate accessory like solvent filters should be disposed according to local environmental requirements. If the extraction tube provided with the instrument is to be extended it is recommended that the diameter of the extension is increased to at least 50 mm (2 in) diameter tubing so the extraction quality is not inhibited. NOTE Do not connect the exhaust vent directly to the detector. This might cause either positive pressure or negative back pressure, both of which will impact the quality of your measurement results. Page 10 of 21

The drain tube must be directed to a waste container supplied with the instrument. The user is responsible for decontamination or recycling of any residue, regarding to local environmental requirements. Further requirements The 1290 Infinity ELSD (G4261B) can be controlled either via RS232 or via LAN. If the RS232 interface is used for control, the ELSD must be installed close to the control PC unless special data transmission systems are used. The length of the straight female/female RS 232 cable supplied with both detectors is 2.9 m. Page 11 of 21

Gas Requirements G4240A 1260 Chip LC For HPLC Chip Cube installation, high purity air is required. The high purity air is used to reduce background ions in the spray chamber and to enhance spray stability. More details regarding the use of the Background Reduction Kit can be found in the installation manual for that kit. Please see Table 1 on page 12 for a summary of the air for Chip Cube requirements. Table 1 Air for Chip Cube requirements Source Purity Gas Pressure Flow High pressure air source1 for 6520A/B, 6530, 6538, 6540 High pressure air source 1 for 6550 99.99 % pure or better and hydrocarbon free2 1 Air source must be able to deliver air at a contstant pressure. 5.5 6.9 bar 2 (80 120 psi) Up to 4 L/min (240 L/h) 99.99 % pure or better and 5.5 6.9 bar 2 (80 120 psi) Up to 10 L/min (600 L/h) hydrocarbon free 2 2 Inlet gas pressure for air must be slightly higher than the pressure used for the nitrogen gas supply pressure. Exhaust Venting Please refer to the Site Preparation Guide of your Agilent mass spectrometer for detailed specifications of the spray chamber exhaust vent. Interface Please plan for 15 cm (6 inch) free space between the LC stack and the HPLC- Chip Cube Interface. It is required for lifting the Chip Cube off the mass spectrometer. The LC and the mass spectrometer should be installed on the same table. This allows to remove the Chip Cube quickly without disconnecting capillaries for routine maintenance tasks. Page 12 of 21

Module Stacking Figure 1 Recommended single stack configuration for HPLC-Chip MS Page 13 of 21

Figure 2 Recommended 2-stack configuration for HPLC-Chip MS Page 14 of 21

Module Stacking G4277A 1290 HTS, G4278A 1290 HTC The G4277A and G4278A have to be installed on free bench space next to the 1290 Infinity LC stack. To avoid excessive delay volumes from the pump to the injection valve and from the injection valve to the LC columns: Install the LC Injector HTS/HTC as close as possible to the 1290 Infinity LC stack. Move the injection valve on the x- rail to the side, where the LC stack is installed. When using a two- stack configuration, the LC Injector HTS/HTC should be placed next to the stack that includes the 1290 Infinity Pump and the 1290 Infinity Thermostatted Column Compartment. Two 600 mm stainless steel capillaries are delivered with the LC Injector HTS/HTC. These should be used to connect the injection valve to the pump outlet and LC column. Peltier cooled stacks require at least 2.5 cm (1 in) clearance between the back of the stack and the wall for air circulation. Figure 3 Recommended stack configuration for 1290 Infinity LC Injectors HTS/HTC Page 15 of 21

Stack Configurations Figure 4 Recommended 1-Stack Configuration for 1260 Infinity (without Thermostat, analytical scale) Page 16 of 21

Figure 5 Recommended 2-Stack Configuration for 1260 Infinity (with Thermostat, analytical scale) Page 17 of 21

Figure 6 Recommended stack configuration for 1290 Infinity (front view) Page 18 of 21

Figure 7 Recommended two stack configuration for 1290 Infinity (front view) Page 19 of 21

Figure 8 Recommended 2-Stack configuration with fraction collector (preparative scale system) Page 20 of 21

Figure 9 Recommended 2-Stack configuration with fraction collector (analytical scale system) NOTE The Agilent 1260 Infinity Purification Solution requires special stack configurations. For details, refer to the Agilent 1260 Infinity Purification Solution - System User Guide and the Agilent 1260 Infinity Purification Solution - Installation Checklist. Separate recommendations may be required for ELSD and Method Development Solution. Page 21 of 21