Analysis of WFS Measurements from first half of 2004

Similar documents
Characterisation of the far field pattern for plastic optical fibres

The Effect of Plate Deformable Mirror Actuator Grid Misalignment on the Compensation of Kolmogorov Turbulence

End-to-end simulations of a near-infrared pyramid sensor on Keck II

Renishaw Ballbar Test - Plot Interpretation - Mills

Characterization and improvement of unpatterned wafer defect review on SEMs

ZONE PLATE SIGNALS 525 Lines Standard M/NTSC

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

Topic: Instructional David G. Thomas December 23, 2015

Update on Antenna Elevation Pattern Estimation from Rain Forest Data

E X P E R I M E N T 1

THE OPERATION OF A CATHODE RAY TUBE

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

THE OPERATION OF A CATHODE RAY TUBE

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

Measurement of overtone frequencies of a toy piano and perception of its pitch

GPA for DigitalMicrograph

Common assumptions in color characterization of projectors

Department of Electrical & Electronic Engineering Imperial College of Science, Technology and Medicine. Project: Real-Time Speech Enhancement

CHAPTER 4 OSCILLOSCOPES

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK

Example the number 21 has the following pairs of squares and numbers that produce this sum.

Reducing tilt errors in moiré linear encoders using phase-modulated grating

RECOMMENDATION ITU-R BT (Questions ITU-R 25/11, ITU-R 60/11 and ITU-R 61/11)

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Using Multiple DMs for Increased Spatial Frequency Response

KRAMER ELECTRONICS LTD. USER MANUAL

ISCEV SINGLE CHANNEL ERG PROTOCOL DESIGN

Pole Zero Correction using OBSPY and PSN Data

Mechanical aspects, FEA validation and geometry optimization

Chapter 27. Inferences for Regression. Remembering Regression. An Example: Body Fat and Waist Size. Remembering Regression (cont.)

Lab P-6: Synthesis of Sinusoidal Signals A Music Illusion. A k cos.! k t C k / (1)

Monitoring Of Drag Anchor Embedment Parameters. Roderick Ruinen, Vryhof Anchors BV

Selected Problems of Display and Projection Color Measurement

Comparison of SONY ILX511B CCD and Hamamatsu S10420 BT-CCD for VIS Spectroscopy

Overview. Teacher s Manual and reproductions of student worksheets to support the following lesson objective:

RST INSTRUMENTS LTD.

Colour Reproduction Performance of JPEG and JPEG2000 Codecs

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Beam test of the QMB6 calibration board and HBU0 prototype

ENGINEERING COMMITTEE

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Video Signals and Circuits Part 2

most often asked questions about mixers

Recommended Operations

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

Tutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program

Dektak Step by Step Instructions:

Functions Modeling Change A Preparation for Calculus Third Edition

Investigation of Digital Signal Processing of High-speed DACs Signals for Settling Time Testing

ZX-44XL Liquid Fuel Analyzer. User s Manual Version 1.2

HIgh speed Photo-POlarimeter (HIPPO) User's manual Version November 2008 Stephen Potter

Status of Pulse Tube Cryocooler Development at Sunpower, Inc.

CARLO GAVAZZI Automation Components

Quick Start Bruker Dimension Icon AFM

RECOMMENDATION ITU-R BT Studio encoding parameters of digital television for standard 4:3 and wide-screen 16:9 aspect ratios

DISPLAY WEEK 2015 REVIEW AND METROLOGY ISSUE

SAM User Guide. 1 How to use this guide 2. 2 SAM at a glance 3

Python Quick-Look Utilities for Ground WFC3 Images

Removing the Pattern Noise from all STIS Side-2 CCD data

DEX Fruit & Stem Growth Dendrometer User Manual

INSTRUMENT CATHODE-RAY TUBE

The Measurement Tools and What They Do

Import and quantification of a micro titer plate image

Optical Technologies Micro Motion Absolute, Technology Overview & Programming

A High-Resolution Flash Time-to-Digital Converter Taking Into Account Process Variability. Nikolaos Minas David Kinniment Keith Heron Gordon Russell

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

Improving Color Text Sharpness in Images with Reduced Chromatic Bandwidth

A review on the design and improvement techniques of comb filters

Mathematics 5 SN SINUSOIDAL GRAPHS AND WORD PROBLEMS

Delivery test of the ALFOSC camera with E2V CCD Ser. no

Algebra I Module 2 Lessons 1 19

United States Patent: 4,789,893. ( 1 of 1 ) United States Patent 4,789,893 Weston December 6, Interpolating lines of video signals

Lecture 2 Video Formation and Representation

GCSE MARKING SCHEME AUTUMN 2017 GCSE MATHEMATICS NUMERACY UNIT 1 - INTERMEDIATE TIER 3310U30-1. WJEC CBAC Ltd.

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn

TROUBLESHOOTING DIGITALLY MODULATED SIGNALS, PART 2 By RON HRANAC

Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs. By: Jeff Smoot, CUI Inc

The Inframetrics 760 airborne thermal infrared data

Blueline, Linefree, Accuracy Ratio, & Moving Absolute Mean Ratio Charts

Reducing GMOS Spectroscopic data

Ultrasonic Testing adapts to meet the needs of the Automotive Tube Industry

Bias, Auto-Bias And getting the most from Your Trifid Camera.

SPECIAL REPORT OF THE SUBCOMMITTEE ON POLARITY STANDARDS 1

RST INSTRUMENTS LTD.

Summary of recent photocathode studies

4. ANALOG TV SIGNALS MEASUREMENT

Understanding solar spectral variability: A six-year comparison between SIM observations and the SATIRE model

INSTALATION PROCEDURE

HIGH QUALITY GEOMETRY DISTORTION TOOL FOR USE WITH LCD AND DLP PROJECTORS

Chapter 6. Normal Distributions

User Manual VM700T Video Measurement Set Option 30 Component Measurements

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

Results of the June 2000 NICMOS+NCS EMI Test

Environmental Controls Laboratory

DIFFERENTIATE SOMETHING AT THE VERY BEGINNING THE COURSE I'LL ADD YOU QUESTIONS USING THEM. BUT PARTICULAR QUESTIONS AS YOU'LL SEE

AskDrCallahan Calculus 1 Teacher s Guide

ISOMET. Compensation look-up-table (LUT) and Scan Uniformity

Transcription:

Analysis of WFS Measurements from first half of 24 (Report4) Graham Cox August 19, 24 1 Abstract Described in this report is the results of wavefront sensor measurements taken during the first seven months of 24. It includes details of some instrumental problems encountered, which produced some unexpected features and the method of evaluating and resolving these problems. A final estimate of the true telescope aberrations versus zenith angle is given with proposed functions calculated for the correction of the most significant term, namely astigmatism. Also checked for were the possible effects of hysteresis of the aberrations due to altitude motion and telescope rotator position influence. 2 Introduction On the 25 March 24 a few hours of WFS measurements were obtain during the period of about 12.3 to 4.3. Before this time the seeing was very variable sometimes been worse than 2.5 arcsecs. During the measurements the seeing was typically better than 1.2 arcsecs, and.7 arcsecs was measured from a short integration on a focus star. It was only possible to get data for zenith angles from 1 to 42 degrees. Follow-up observations were made on 6th April where only a few specific pointing were made to try and verify some of the findings from the 25th March data. This is the first set of measurements taken with ALFOSC since the new camera and CCD have been installed, though because the pixel scale has been kept the same the reduction software is still valid. Due to some unexpected and inconsistent results from the data of 25 March 24 and 6 April 24, extra measurements were made on 26 May 24 to try and clarify the situation. The measurement made on 26 May and more on 13 July 24 finally show some consistency and from them it has been possible to determine satisfactory fits to the data that if implemented by the TCS should reduce the effects of the most significant aberration, astigmatism, by a substantial amount. From the 26 May data it is possible to calibrate the TCS Zenned amplitudes but the Zernike to rotator angles relationship still needs to be determined. 1

2 3 RESULTS OF 25 MARCH AND 6 APRIL 24 DATA 3 Results of 25 March and 6 April 24 data On 25 March it was possible to obtain a series of WFS measurements by tracking a single star for a few hours resulting in data for zenith angles from 1 to 45 degrees. The results of these observations are shown in figure 3. Table 1 shows the mean values subtracted from the March data, the mean of ten consecutive measurements made eight months earlier, at a zenith angle of approximately 2 degrees, after the large astigmatism was removed but before the new camera and CCD were installed, and also the means of the April data, though this latter data is of poor quality. From the data, as seen before, the variations in the higher order aberrations, namely tri-coma and quad-astigmatism, are relatively small compare to changes in astigmatism and focus. Also the mean values are in the same order of magnitude to previously measured results, been dominated by astigmatism, coma, spherical and focus, though spherical is higher than previously measured. One interesting feature that has come from these results is that the slope of the astigmatism amplitude against zenith angle has appeared to have changes sign compared to measurements from 22 and 23 [1], though the range is about the same as before and the mean value is similar to the last measurements. It is known the CCD in the new AL- FOSC camera it up side down with respect to the old detector, though since astigmatism is symmetrical about 18 o you would expect to get the same result. A check has been made against old data, taken on 11 April 23, by just looking at the same zenith range, zenith angle decreasing from 1 to 45 degrees, and the slope is opposite with respect to 25 March 24. The plot of astigmatism angle against zenith angle is consistent with previous results. The only explanation for this change must be because of the applied extra forces to M1, to correct for the astigmatism, causing different deformation in the mirror when tilted. Though there are large apparent mean measured values for astigmatism, coma, spherical and focus none of these aberrations have been reported in any science observations, and in fact after the large clearly seen astigmatism was corrected for and check from point source observations to have been removed, the WFS analysis still gives a mean value of over 1. Also the apparent large coma has never been reported which would appear as elongated images with a comet type profile. For spherical it would be very difficult to separate this from poor focus or seeing affects so would not be expected to be seen. To try and verify if the change of slope of astigmatism against zenith angle is real, a very few measurements where made on 6 April, these included taking two sets of measurements at the same zenith angle but for opposite rotator positions, namely and 18 degrees. Also measurements for a range of zenith angles were made with the TCS astigmatism set to.8 and for the previous setting of.2, again to see if this influences the astigmatism behaviour. Unfortunately the results are not very good, presumably due to poor seeing and limited measurements made. From the measurements for the two rotator positions a significant change in the magnitude of astigmatism results, for ROT-POS = o the astigmatism is 22 but for ROT-POS = 18 o it is 121 (both the mean of three measurements). The large value from ROT-POS zero is consistent with the mean value obtained for measurements for the same rotator position and for different zenith angles. The measurements of zenith angle for the applied TCS astigmatism of.2 give a mean value of 96 and a change of slope with respect to the TCS astigmatism of.8. The slope for TCS astigmatism of.2 is consistent with data from the previous years 22 and 23, before the applied TCS astigmatism was made. A unexplained curiosity of the stellar wavefront mask images taken on 25 March and 6 April is that they all appear vertically elongated but with the centre hole obstruction elongated horizontal. This shape is seen for the TCS astigmatism set to.2 or.8. A typical WFS mask image can be seem in figure 1. Calibration images taken from the WFS

3 Figure 1: This is a typical WFS mask stellar image taken on night of 25 March 24. Note the vertical elongation of the image with a perpendicular elongation of the centre obscuration. Figure 2: Typical WFS calibration mask image. lamp in the aperture wheel look normal, an annulus, as in figure 2. Aberrations 2/8/3 25/3/4 6/4/4 Amp. Ang. Amp. Ang. Amp. Ang. Tilt (arcsec) 2.5 35.17-14.2.2 12 Astigmatism 17 9 154.5-49.9 211-25 Coma 14-69 91.5-54.2 84 7 Tri-coma 43-9 31.6 6.8 35 32 Quad-ast. 15 35 18.7-1.4 3 1 Spherical -17 - -79.1 - -96 - Focus 145-83. - -129 - Table 1: The mean aberration amplitudes () and angles (degrees) including a comparison of the mean of 1 consecutive measurements made August 23. The 6 April data is for rotator position and TCS astig. =.8 4 Results of 26 May 24 The unexpected and inconsistent results from the data of 25 March 24 and 6 April 24, described above, compared to previous measurements, prompted a series of new tests to be devised to try and clarify the situation and data was taken on 26 May 24. From the seeing monitor the seeing was very variable with a mean or around 1.2 arcsec, even though only four frames were lost (out of over one hundred taken) due to poor image quality. The most important item was to first get round WFS-mask images and this was achieve (with advice from Michael Andersen) by setting the position of the WFS-mask more precisely aligned to the Lyot-stop. Also the calibration lamp was position accurately on the

4 5 RESULTS OF 13 JULY 24 COMPARED WITH 26 MAY DATA optical axis. This then produced the desired circular images confirming that the cause was vignetting of the primary mirror image by the WFS-mask Lyot-stop combination. The first set of measurements on the sky where to see if the change in zenith dependence (shape) of the astigmatism, with respect to post aluminisation data, is real, an artifact of the poor WFS-mask images or due to the change in the TCS astigmatism Zernike. A set of measurements were made for the TCS astigmatism Zernike set to.8 (the current value for minimum aberration) and the previously used.2. Comparing the results of 26 May data shown in figure 4 with those of 25 March data of figure 3, it can be seen that the shape for the astigmatism amplitude for TCS Zernike =.8 is approximately the same for zenith angles greater than 15 degrees, namely as the zenith angle increases the astigmatism goes down with a minimum for zenith angle 3-35 degrees and then starts to increase. There does appear to be a difference for small zenith angles where the new results show the astigmatism amplitude continuing to increase as you approach zenith where as from 25 March data this aberration appears to fall off for small zenith angles. Another feature of the astigmatism, seen in 26 May data of figure 4, is it seems to rotate with respect to the zenith angle with an approximately linear relationship. A check was made of the apparent rotator position affecting the amplitude of the aberrations. It was found that there is no relationship between rotator angle of o and 18 o and aberration amplitude, as you would expect. The zenith angle dependence was determined my making a series of measurements starting at angle 6 degrees, moving up to zenith in steps of 1 degrees and then back down to zenith angle 6 degrees. For each position three measurements were made and the mean of these three used for the value of each position. By plotting the up and down halves together, shown in figure 5, it was possible to see that there is negligible hysteresis in the aberration changes with respect to altitude. Again only the high order aberrations, astigmatism and focus are significant. Large mean values for coma and spherical are found but these may simply been offsets within the software rather than true aberrations since there has be no reported asymmetry in images, which you would expect from coma and astigmatism, from observers. 5 Results of 13 July 24 compared with 26 May data A repeat of the 26 May measurements was taken to confirm the results, in particular the astigmatism trend. The procedure for taking the data was again to make three measurements at intervals of approximately 1 degrees for zenith angles from 6 up to 5 degrees then back down to 6 degrees. Also three sets of measurements were taken for three different focus positions, at a small zenith angle, to calibrate the focus in focus units. Looking at the plots in figure 6 it can be seen that the 13 July data is consistent with that from 26 May, showing that the high order aberrations have only a small effect on the overall distortions. Looking particularly at the astigmatism result it can be clearly seen that the amplitude has a well defined and repeatable zenith dependency, though the angle does appear to have significant discrepancies especially near zenith. The May data gives a roughly linear angle of rotation with zenith angle. It was possible to check the July data at small zenith angles using the measurements take to determine the focus calibration. From the mean of these measurements an amplitude value of about 5 was found at an angle of 6 degrees (after correcting for the different rotator angle used) which is consistent with the general data. With the focus plot now given in focus units it is clear to see that the typical focus versus zenith angle is reasonable flat with the polynomial fit giving a maximum of ±1 units and

5 with no very obvious zenith dependency. Coma shows a very flat curve for angles up to zenith angle 6 degrees then with a possible increase at larger angles. This has been seen in previous measurements taken in 22 and 23 before the aluminisation and is probably due to incorrect correction of displacement of the secondary, by the so-called.6mm mechanism, at these low altitudes. The spherical aberration does show a clear zenith dependency which if necessary could be compensated for though it does have a small amplitude. The mean astigmatism values for both the May and July measurements are in agreement, at approximately 1, though actually interpreting exactly what the significant of the mean is is more difficult. The observations measure the sine and cosine terms of the respective aberrations and the plots show the amplitude and phase of these parameters. Amplitude can never be negative, except for spherical that has no phase, because this would simply represent a phase shift. e.g. for astigmatism a sign change will be the same as a rotation of 9 degrees, and in fact the astigmatism does rotate about zero symmetrically so if this was represented as a sign change in the amplitude the mean would be approximately zero. For both the May and July data the mean of all the angles are approximately zero and the fluctuations are due to measurement errors because of the small amplitudes of the respective aberrations. Looking at the raw Zernike terms for astigmatism Z4 and Z5 we see a very interesting feature, the cosine and sine terms are straight lines as shown in figure 9 with the following fits. AstigmatismCosine = 187 + 5.285z (1) AstigmatismSine = 29.4326z (2) Z1: Tilt X = -.586 +.382z Z2: Tilt Y =.577 -.991z Z3: Focus = 58 +.216z Z4: Ast Cos = -187 + 5.2845z Z5: Ast Sin = -29 -.4326z Z6: Coma Cos = 82 +.8296z Z7: Coma Sin = -12 +.445z Z8: Tri-Coma Cos = -8 +.393z Z9: Tri-Coma Sin = -38 +.5151z Z1: Spherical = -31 -.335z Z11: Quad Ast Cos = -3.66 +.816z Z12: Quad Ast Sin = 7.1 -.2226z Table 2: Fit to WFS Zernikes in terms of zenith angle (z) 6 Zenith Dependency Correction It can be seen from the 26 May and 13 July results that astigmatism has a strong zenith angle dependency and to remove this the TCS must apply a suitable correction. Using the IDL polynomial fitting function polynomials were determined for astigmatism amplitude

67 DETERMINING THE RELATIONSHIP BETWEEN THE TCS AND WFS VALUES and angle for the combined data from both months. It was found that a satisfactory fit to the angle is a straight line though the amplitude needs a third order polynomial. The fits to the astigmatism amplitude and angle are given in equations 3 and 4 respectively, for zenith angles (z) from 5 to 65 degrees. Also by trial and error a cosine function has been made to fit the data and is given in equation 5, for zenith angles in degrees. For both astigmatism amplitude equations the mean of 1, subtracted from the plotted and fitted data, has been included giving a minimum at zenith angle 33 degrees of approximately 55. AstigmatismAmplitude = 238 12.2z +.23z 2.9z 3 (3) AstigmatismAngle = 7 2.1z (4) AstigmatismAmplitude = 255 2cos(4.4πz/36 π/2.4) (5) Subtracting the polynomial fit from the original data it is found that the residuals of the astigmatism amplitude have be reduced to standard deviation of approximately 21, as seen in figure 7. Doing the same for the spherical aberration a reduction in the standard deviation by 3 can be achieved, essentially making the slight curve in the zenith dependency straight, which is considered insignificant and will not be implemented, though the mean of approximately -45 should be removed. It is possible to use the 26 May data to determine the relationship between the so called TCS astigmatism amplitude parameter and the astigmatism amplitude given by the AL- FOSC WFS, since on this night two values for the TCS astigmatism amplitude were used. From the astigmatism amplitude data at zenith angle 33 degrees, where the astigmatism angle is approximately equal to zero, the difference between the two measured values is 14 and the TCS astigmatism amplitude change is -.6. So in round numbers 1 is equivalent to -.43 on the TCS. To determine the TCS astigmatism angle extra measurements are need. From the observations the TCS astigmatism amplitude and angle were adjusted to give minimum aberrations at a zenith angle of approximately 3 degrees, this position corresponds to a change in astigmatism amplitude wrt the astigmatism at zenith angle (z) equal to zero of 185 or.787 in TCS units. Since all the measurements were made with a TCS astigmatism set to.8 this also has to been added to the result giving a final value of 1.587, this gives a TCS astigmatism minimum value of.8 at zenith angle 3 degrees. Therefore for the TCS the equation to apply to remove the astigmatism, taking into account there is a sign change within the TCS, is given as: T CS Ast Amp = 1.587 52.46x1 3 z + 9.89x1 4 z 2 3.87x.1 6 z 3 (6) Similarly for the estimated cosine fit the equation becomes: T CS Ast Amp = 1.66.86cos((4.4πz)/36 (π/2.4)) (7) Plots of equations 6 and 7 are given in figure 8. 7 Determining the relationship between the TCS and WFS values As stated above a value for the relationship between the TCS astigmatism amplitude and the measured ALFOSC WFS value has been determined as is give as 1 measured is

7 equivalent to -.43 on the TCS. To try and get the angles a series of measurements were made for TCS positions from zero to 135 degrees in steps of 45 degrees at a zenith angle of approximately 2 degrees. The results of these measurements are given in figures 12 and 13. Figure 12 shows the relationship between TCS astigmatism angle and the Z4 and Z5 (cosine and sine terms respectively) Zernikes coefficients and figure 13 the amplitude and angle relationship. The error bars for all the plots are ±1 sigma. From the amplitude and angle graphs for the applied TCS astigmatism amplitude, figure 11, they show that the applied TCS astigatism amplitudes of.2 and 1.2 for the given zenith angle of about 2 degrees and TCS angle of 1 degrees, is at 9 degrees with respect to TCS amplitude.8, which is can be seen from the measured change of angle as well as amplitude. This is what you would expect since a negatively applied astigmatism is equivelant to a 9 degree rotation. The affect of the rotation of the applied TCS astigmatism angle, seen in the two graphs for TCS applied astigmatism angle, figure 13, and shows the default applied amplitude of.8 at 1 degrees is at a minimum and adds to the actual mirror astigmatism after a 9 degree rotation, which is what you would expect. The results shown in 11 and 13 of the applied TCS astigmatism amplitude and angle agree exactly with each other and suggest the rotator angle and the TCS astigmatism angles are the same in magnitude though it is impossible to say if the sign is correct. 8 Conclusion Looking at the data presented in this report and comparing it with the data taken before [1] the aluminisation and problems with the lateral supports, it is clear that after the primary mirror has been removed from the telescope it is necessary to make a sequence of wavefront sensor measurements to determine the aberration zenith angle dependences. From the zenith angle dependency measurements it is clear that astigmatism has the most significant changes and should be compensated for. Two equations have been determined that should implement the TCS astigmatism correction. Spherical also shows a clear zenith angle dependency and could be corrected for, though this is of a much small amplitude than the astigmatism. The proposed measurement procedure of taking a few WFS measurements at roughly every 1 degrees zenith angle seem to be sufficient to obtain data that demonstrates the affects of telescope altitude on mirror aberrations, and that satisfactory results can be obtained for seeing as poor as 1.2 arcsec, though better the seeing better the data. Also it is important to get the optical elements aligned correctly within ALFOSC to ensure we get good and correct results. A procedure has been produced and placed on the web [2] describing the correct method to get reliable WFS data from ALFOSC, and which was used for both the May and July observations. References [1] www.not.iac.es/telescope/, Analysis of WFS data, March 22 - April 23 [2] www.not.iac.es/telescope/, WFS Measuring Procedure

8 REFERENCES.1 Tilt Amplitude 2 Tilt Angle.5 1 arcsec. -.5-1 -.1 1 15 2 25 3 35 4-2 1 15 2 25 3 35 4 Astigmatism Amplitude Astigmatism Angle 6 4 4 2 2-2 -2-4 -6-4 1 15 2 25 3 35 4 1 15 2 25 3 35 4 Coma Amplitude Coma Angle 6 4 4 2 2-2 -2-4 -6-4 1 15 2 25 3 35 4 1 15 2 25 3 35 4 Tri-coma Amplitude Tri-coma Angle 6 4 4 2 2-2 -2-4 -6-4 1 15 2 25 3 35 4 1 15 2 25 3 35 4 Quad-astigmatism Amplitude Quad-Astigmatism Angle 6 4 4 2 2-2 -2-4 -6-4 1 15 2 25 3 35 4 1 15 2 25 3 35 4 Spherical Amplitude Focus Amplitude 6 6 4 4 2 2-2 -2-4 -4-6 -6 1 15 2 25 3 35 4 1 15 2 25 3 35 4 Figure 3: Aberration evolution against zenith angle on 25 March, the dotted line is a 5th order polynomial fit. Table 1 gives the mean offset subtracted from the data.

REFERENCES 9 Tilt Amplitude Tilt Angle.6.4 mean =.65 arcsec mean =.56 arcsec 5.2. -.2 -.4-5 -.6 2 1 2 3 4 5 6 Astigmatism Amplitude 1 2 3 4 5 6 Astigmatism Angle 1 mean = 14 mean = 255 5-1 -5-2 2 1 2 3 4 5 6 Coma Amplitude 1 2 3 4 5 6 Coma Angle 1 mean = 9 mean = 98 5-1 -5-2 2 1 2 3 4 5 6 Tri-coma Amplitude 1 2 3 4 5 6 Tri-coma Angle 1 mean = 41 mean = 4 5-1 -5-2 2 1 2 3 4 5 6 Quad-astigmatism Amplitude 1 2 3 4 5 6 Quad-astigmatism Angle 1 mean = 13 mean = 24 5-1 -5-2 1 2 3 4 5 6 1 2 3 4 5 6 2 Spherical Amplitude 2 Focus Amplitude 1 mean = 5 mean = 46 1 mean = 96 mean = 153-1 -1-2 1 2 3 4 5 6-2 1 2 3 4 5 6 Figure 4: Aberration evolution against zenith angle on 26 May. Shown solid line is the TCS AST Zernike =.8 and dash line for the old TCS AST Zernike =.2.

1 REFERENCES Tilt Amplitude Tilt Angle.5 5. -.5-5 15 1 2 3 4 5 6 Astigmatism Amplitude 1 2 3 4 5 6 Astigmatism Angle 1 5 5-5 -1-5 -15 15 1 2 3 4 5 6 Coma Amplitude 1 2 3 4 5 6 Coma Angle 1 5 5-5 -1-5 -15 15 1 2 3 4 5 6 Tri-coma Amplitude 1 2 3 4 5 6 Tri-coma Angle 1 5 5-5 -1-5 -15 15 1 2 3 4 5 Quad-astigmatism Amplitude 1 2 3 4 5 Quad-astigmatism Angle 1 5 5-5 -1-5 -15 1 2 3 4 5 1 2 3 4 5 15 Spherical Amplitude 15 Focus Amplitude 1 1 5 5-5 -5-1 -1-15 1 2 3 4 5-15 1 2 3 4 5 Figure 5: Aberration hysteresis against zenith angle on 26 May. Shown solid line for decreasing zenith angle and dash line for increasing zenith angle.

REFERENCES 11 Tilt Amplitude Tilt Angle.6.4 5.2. -.2 -.4-5 -.6 1 2 3 4 5 6 Astigmatism Amplitude 1 2 3 4 5 6 Astigmatism Angle 5 5-5 -5 1 2 3 4 5 6 Coma Amplitude 1 2 3 4 5 6 Coma Angle 5 5-5 -5 1 2 3 4 5 6 Tri-coma Amplitude 1 2 3 4 5 6 Tri-coma Angle 5 5-5 -5 1 2 3 4 5 6 Quad-astigmatism Amplitude 1 2 3 4 5 6 Quad-astigmatism Angle 5 5-5 -5 1 2 3 4 5 6 Spherical Amplitude 4 1 2 3 4 5 6 Focus Amplitude 5 2 Focus Units -2-5 1 2 3 4 5 6-4 1 2 3 4 5 6 Figure 6: Aberration evolution against zenith angle on 26 May (dashed line) and 13 July (solid line). Included is a polynomial fit of both sets of data combined. (dotted)

12 REFERENCES Astigmatism Amplitude Residuals Astigmatism Angle Residuals 4 6 4 2 2-2 -2-4 -4-6 1 2 3 4 5 6 1 2 3 4 5 6 Figure 7: Residuals of the astigmatism amplitude and angle after the above polynomial fits, of the combined data from both dates, has been subtracted. 2. Estimates of Astigmatism Amplitude 1.8 1.6 TCS Units 1.4 1.2 1..8 2 4 6 8 () Figure 8: Two estimates of the Astigmatism Amplitude versus. The solid line is a polynomial fit and the dotted a best guess Cosine fit.

REFERENCES 13 2 Astigmatism Cosine (Z4) 2 Astigmatism Sine (Z5) 1 1-1 -1-2 1 2 3 4 5 6-2 1 2 3 4 5 6 Figure 9: Zernike terms (Z4 and Z5) of the astigmatism with straight line fits. Measured Astigmatism Cosine () 1-1 -2 TCS Astigmatism Amplitude versus Measured Astigmatism Cosine.2.4.6.8 1. 1.2 TCS Astigmatism Amplitude Measured Astigmatism Amplitude () 25 2 15 1 TCS Astigmatism Amplitude versus Measured Astigmatism Amplitude.2.4.6.8 1. 1.2 TCS Astigmatism Amplitude TCS Astigmatism Amplitude versus Measured Astigmatism Sine TCS Astigmatism Amplitude versus Measured Astigmatism Angle Measured Astigmatism Sine () 1 5-5 -1-15 Measured Astigmatism Angle () 5-5.2.4.6.8 1. 1.2 TCS Astigmatism Amplitude.2.4.6.8 1. 1.2 TCS Astigmatism Amplitude Figure 1: Relationship between applied TCS astigmatism amplitude versus measured Zernikes Z4 and Z5, Cosine and Sine respectively. Figure 11: Relationship between applied TCS astigmatism amplitude versus measured amplitude. Measured Astigmatism Cosine 4 2-2 TCS Astigmatism Angle versus Measured Astigmatism Cosine 2 4 6 8 1 12 14 TCS Astigmatism Angle () Measured Astigmatism Angle () 1 5-5 -1 TCS Astigmatism Angle versus Measured Astigmatism Angle 2 4 6 8 1 12 14 TCS Astigmatism Angle () Measured Astigmatism Sine 2-2 -4 TCS Astigmatism Angle versus Measured Astigmatism Sine 2 4 6 8 1 12 14 TCS Astigmatism Angle () Measured Astigmatism Amplitude () 6 5 4 3 2 1 TCS Astigmatism Angle versus Measured Astigmatism Amplitude 2 4 6 8 1 12 14 TCS Astigmatism Angle () Figure 12: Relationship between applied TCS astigmatism angle versus measured Zernikes Z4 and Z5, Cosine and Sine respectively. Figure 13: Relationship between applied TCS astigmatism angle versus measured angle.