University of Minnesota Nano Fabrication Center Standard Operating Procedure

Similar documents
2.1. Log on to the TUMI system (you cannot proceed further until this is done).

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FEI FIB Focused Ion Beam

Operation Procedure for Phillips XL30 ESEM

FEI Strata Dual-beam FIB

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

University of Minnesota Minnesota Nano Center Standard Operating Procedure

Approved by: / / R. Battaglia 12/16/2016

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

The Quanta 200 3D Simplified Operation Manual

Dektak Step by Step Instructions:

Tender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016)

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

Description of task... 2 Sample preparation... 2 Sample insertion and SEM settings... 3 Eucentricity... 4

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2.

JEM 3010 Manual (2011 Edition)

PHI 5000 VersaProbe TM Operator s Guide

Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018)

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

SC24 Magnetic Field Cancelling System

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000

SC24 Magnetic Field Cancelling System

Based on the discussion, the committee recommends the following amendments to the specification.

CHA EVAPORATOR Short Reference User guidelines

Screen investigations for low energetic electron beams at PITZ

Focused Ion Beam System MI4050

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

Approved by Principal Investigator Date: Approved by Super User: Date:

Operation of CEBAF photoguns at average beam current > 1 ma

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE

Product Data Sheet 4656 EZ

TECHNICAL SPECIFICATION Multi-beam S-band Klystron type BT267

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual

Characterization of Diamond Samples CHESS Run Summer 2013

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

e-vap 3kW Evaporation Sources

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Connection for filtered air

University of MN, Minnesota Nano Center Standard Operating Procedure

Nova NanoSEM Superior Imaging and Analytical Performance

HIGH CURRENT IMPLANTER 200KeV - SERIES 1090 TECHNICAL DESCRIPTION

COLOR TFT LCD MONITOR. Manual

MP 35" Zero-G 100Hz Curved Monitor with AMD FreeSync 2.0

MP Zero-G 27" WQHD 144Hz TN-LED Monitor with AMD FreeSync

Focused-ion-beam fabrication of nanoplasmonic devices

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

Element 78 MPE-200. by Summit Audio. Guide To Operations. for software version 1.23

Mechanical aspects, FEA validation and geometry optimization

THERMIONIC GUN CATHODE-GRID ASSEMBLY TEST PROCEDURE

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator.

The Martin Wood Lecture Theatre. Notes for Speakers on the use of the Audio/Visual equipment

OPERATING INSTRUCTIONS TOM-0431IP

Max. temp. of sensor C. Capillary tube mm. Switching difference K. Sensor cartridge mm

Standard Operating Manual

Camden Technology Center. Smart Classroom Guide A/V Sympodium

Simple all-in-one design style with front stereo speakers and natural ventilation system

DIRECT DRIVE ROTARY TABLES SRT SERIES

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

AFM Standard Operating Procedure

SOFT LED 4 Vari-White INSTRUCTION MANUAL

High Resolution Multicolor Contrast Scanner. Dimensioned drawing

Figure 1. MFP-3D software tray

Improvements in Gridless Ion Source Performance

Preparing for remote data collection at NE-CAT

A. All equipment and materials used shall be standard components that are regularly manufactured and used in the manufacturer s system.

2.2. VIDEO DISPLAY DEVICES

Standard Operating Procedure of nanoir2-s

Quick Start Bruker Dimension Icon AFM

INFORMATION TO USER CAUTION RISK OF ELECTRIC SHOCK, DO NOT OPEN

STANDARD OPERATING PROCEDURE:

ITER Product Catalogue. Total Pressure Measurement V Page 1 / 15

Fabrication of Lithium Niobate nanopillars using Focused Ion Beam (FIB)

Click Here To Start Demo

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

USER MANUAL. 27" 2K QHD LED Monitor L27HAS2K

The BAT WAVE ANALYZER project

PRODUCT NEWS FEI LAUNCHES APREO HIGH- PERFORMANCE SEM RENISHAW OFFERS CONFOCAL RAMAN MICROSCOPE

Backside Circuit Edit on Full-Thickness Silicon Devices

Check what you have received against the component checklist and hardware above.

40-Meter Subsystems: As LIGO-Like as Possible

SHOWLINE SL PUNCHLITE 220 LED LUMINAIRE SPECIFICATIONS.

An Overview of Beam Diagnostic and Control Systems for AREAL Linac

Dell D3218HN. User s Guide. Regulatory model: D3218HNo

SUPER LED F4.7 VW Vari-White INSTRUCTION MANUAL

Mini Micro Pulse Lidar System

INSTALATION PROCEDURE

USING MEDIACAST FOR LIVE EVENTS THERE ARE THREE SEPARATE KEYS NEEDED TO ACCESS THE FRONT AND BACK OF THE VIDEO CART.

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 2nd Edition

USER MANUAL. 22" Class Slim HD Widescreen Monitor L215DS

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

Network Disk Recorder WJ-ND200

22" Touchscreen LED Monitor USER'S GUIDE

Transcription:

Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam system (fitted with both an E-beam and an Ion Beam) comprised of a tungsten electron column and a Ga ion column that uses either high vacuum or environmental SEM (ESEM) technology to section, image, and analyze a wide range of conducting and non-conducting samples. This system offers the capability for in-situ dynamic experiments, 3D imaging and analysis, and transmission electron microscope (TEM) sample preparation for more in-depth analysis. 2 Safety a Must be a qualified user who has taken the short course and the follow-up with an NFC instructor before operating the machine independently. 1

3 Restrictions/Requirements b Must be a qualified user who has taken the short course and the follow-up with an NFC instructor before operating the machine independently. 4 Required Facilities a 240 VAC, 30 amps b 10 psi nitrogen 5 Definitions a Working Distance (WD) physical distance between the sample surface and the pole piece of the e-beam column. b Focused Working Distance (FWD or WD on Data Bar) working distance as determined by the electron optics of the column. This value is displayed on the data bar of the image. This value changes as one adjusts the focus knob. c Eucentric Height the height at which the specimen can be tilted and no projection of the image is seen in the vertical axis of the screen. This is related to the design of the electron optics. This is approximately 15 mm from the Electron Beam s pole piece in the Quanta 200 3D, but may vary slightly based on the acceleration voltage used. This is also the height at which the Ion Beam has been engineered to cross paths with the Electron Beam in the Quanta 200 3D d Beam Coincidence making the Ion Beam coincident with the E-beam. This is usually done using the Ion Beam s beam shift capability with the E-beam s beam shift set to zero. However, the E-beam s beam shift may also be adjusted if necessary. Note that the Ion Beam can only be made coincident with the E- beam at the Eucentric Height. At other z-heights, the distance needed for adjustment will be outside the range of either one or both beam s beam shift limits. 6 Setup a) Log into Coral. b) Enable fib under the E-Beam, SEM, & Mask Making node in Coral. c) Log into the Support PC. This involves logging into the Windows 2000 operating system of the computer which may require going through the Start Menu and selecting the Shut Down option. d) Log into the Quanta PC (also called the Quanta Server). This involves logging into the application xt which is the user interface for operating the FIB instrument. e) Select the Vent icon on the Quanta PC. It will take approximately 3 minutes to vent. The Chamber Pressure Symbol will go from green to orange during the venting process, then finally to black when it is fully vented. 2

Vent Icon (upper right of Beam Page ) Vent Icon Chamber Pressure Icon (lower right of Beam Page ) Chamber Pressure Icon f) Load one of the following three stages 1) Single Stub Holder: Elephant Height Indicator 3

2) Multiple Stub Holder / Large Sample Holder 3) Universal Mounting Block ( UMB ) g) Attach sample to stage. Usually this involves carbon tape, but it might involve Au coating of the sample in Chase 4, painting with carbon or silver paint, and/or taping with Cu tape. h) Using the Elephant height indicator, make certain the top of the sample has a Working Distance (WD) of 15 mm or greater. This assures it will not run into anything when the door is closed. i) Close chamber door. Note you can watch on the CCD camera quadrant of the Quanta PC s screen as the door closes to insure nothing will be damaged when closing the door. j) Select the Pump icon on the Quanta PC. 4

Pump Icon Pump Icon It will take approximately 4 minutes to pump to 2 x 10^-4 or lower. Once this vacuum is achieved, then the sources can be turned on. The chamber icon will go from orange to green during the pump down process. Green indicates desirable vacuum has been achieved. Chamber Pressure Icon (lower right of Beam Page ) Chamber Pressure Icon 7 Operating Instructions a) Turning on the Sources 1) Select Wake Up to initiate warm-up of the E-beam and the I-beam sources. The E-beam will take 5 sec. to warm up, but the I-beam will take 4 to 5 min. 5

Alternatively, one can choose to turn on the E-beam or the I-beam separately. This can be done by selecting the HV symbol for each beam. Wake Up Symbol and HV Symbol Wake Up Icon HV Icon 2) In the E-beam quadrant, un-pause the screen. Adjust focus and magnification accordingly to determine where one is on the sample. Move to the region of interest on the sample to perform Find the Eucentric Height. b) Find the Eucentric Height 1) Place top of sample at "15mm" yellow marker. 2) Focus @ 2000+ X magnification on some horizontal feature in the area of interest. 3) Link Z to FWD. Use Icon. 4) Adjust the Tilt Angle with small, incremental changes (2 o, 5 o, 10 o, 25 o, 52 o ). With each incremental change in Tilt, adjust the Stage Height (Z- Knob) to bring the horizontal feature back to the cross-hair. 5) Repeat this procedure until 0 o (ctrl-e) and 52 o (ctrl-i) remain within a couple of microns of your horizontal feature. Z to FWD 6

6) With the sample normal to the Ion Beam, make the Ion Beam image coincident with the E-beam image by adjusting the Ion Beam's Beam Shift. c) Patterning 1) In the Ion Beam quadrant, locate the section of the sample you would like to pattern; this could be either an etch or a deposition with Pt. 2) Patterns are drawn on the Ion Beam quadrant using the Patterning page. Patterning Page 3) Choose and draw the desired pattern from the list of pattern functions on the Patterning Page. Once drawn, the pattern at hand can be dimensioned by adjusting the boxes for x, y, and z. 4) When choosing an ion current between 1 pa and 20 na for a patterning job, keep in mind a good rule of thumb is to have the job take between 3 and 8 minutes. 7

5) To monitor the progress of the patterning, one can take snapshots of the region of interest using the Snapshot Icon in the either the E-Beam or Ion-Beam quadrant. Most of the time, the E-Beam will be the preferred method of taking snapshots, as it does not impart damage to the sample as the Ion-Beam does. Snapshot Icon d) Shutting Down the System 1) Return the stage to Tilt = 0. 2a) If one wishes to keep the Ion Source warm because another sample is going to be loaded, or another user is about to use the system, one can simply click on Vent. The High Voltage (HV) will be shut down for both sources, and the E- beam source (filament) will be shut off, but the Ga source will stay warm. Upon re-loading and pumping down again, all that will have to be done is clicking on the HV icons for each source. Be sure to shut off the Pt source warmer if it is not to be used again, otherwise it will remain on. 2b) On the other hand, if one no longer wants to use the system, and no other user is waiting to immediately use the system, one can simply click on the Sleep Icon. This will turn of the HV for both sources, turn off the E-Beam source (filament), and allow the Ga source to cool down. In addition, Sleep will also turn off the Pt source warmer. 3) Remove sample stubs from the stage, then remove the sample from the stubs. 4) Close door. 5) Click on Pump icon. Allow to pump down until the Vacuum Status Icon turns green (2 x 10^-4 Torr or lower). 6) Log off the Quanta PC s xt User Interface application (under the File menu heading). 7) Log off the Support PC s OS (Windows 2000 can be found under the Shut Down window). 8) Disable the FIB in Coral. 8

8 Problems/Troubleshooting a Problem: System locks up when stage is moving. Solution: Avoid performing other functions when the stage is moving. Allow stage to come to a complete stop before initiating the other functions. b Problem: System locks up when frequent scans or snapshots are performed. Solution: Avoid scanning a new quadrant or selecting a new snapshot too frequently. Allow the scan or the snapshot in progress to complete before selecting another. Frequent selecting of the scanning or the snapshot function overloads the video card and the video processing software. 9 Online Appendices (see http://www.nfc.umn.edu/equipment/) FIB Appendix 1: TEM Lamella Formation 9