CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

Similar documents
CARESTREAM DIRECTVIEW Elite CR System

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

CARESTREAM DIRECTVIEW Elite CR System

Basic Vortex Mixer Standard Vortex Mixer Advanced Vortex Mixer Pulsing Vortex Mixer

Cosmos2 The Economical X-ray Solution with the Best Image Quality and High Reliability

2013, 2014 Hewlett-Packard Development Company, L.P.

UltraTwist. Product Description. Instruction Guide

Motor Operated Solar Shade with Valance Installation and Care Instructions Complete Video Instructions Available Online at

EN - English Washington Street Melrose, MA Phone Toll Free Revision 4 20/06/17

Essentials of Digital Imaging

UArm. Series. Analog or Digital U-Arm System

KODAK INDUSTREX HPX-1 Hardware

High Performance (Gold Plus) Spliceable Tape Feeder Part Number: Part Number: Revision 3 Jun 2008 No.

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

HP EliteDisplay LED Backlit Monitors. User Guide

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer

K Service Source. Apple High-Res Monochrome Monitor

QubeVu Operator s Guide

USER MANUAL. 28" 4K Ultra HD Monitor L28TN4K

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA

K Service Source. Apple High-Res Monochrome Monitor

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:

WID-DL74 WID-DL74 BLP WID. Designed for. Installation guide for workitdesk interactive table for. BrightLink Pro

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400

Imaging solutions for any size patient and any size budget.

3B SCIENTIFIC PHYSICS

SMART CINEMAHORIZONTAL. User Guide VPSP Projector side. model. Notice SmartCrystal Cinema MUV V1R0

Radiological Imaging. Software Version 3.6

Photovoltaic Module Installation Manual (IEC)

ivw-fd122 Video Wall Controller MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Page i Rev. 1.

World s Thinnest Flat Panel Mounting Solutions

1.0 DESCRIPTION. This specification covers roll-up signs to be used in temporary traffic control zones.

PRJTPFL inch PRJTPFL inch PRJTPFL inch. Fixed Wall Mount Projector Screen. Universal Home/Office Projector Viewing Display

2017 ezframe Series Product Brochure The ezframe series offers several screen materials models

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

Dental Line. High frequency intraoral x-ray unit. radiology ahead

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

Tripod Portable Projection Screen Elite Tripod Series

Photovoltaic Module Installation Manual (IEC)

Corrigendum-II. Up to 30 th October 2018 till 17:00 Hrs. 31 st October 2018 till 17:00 Hrs. 2 nd November 2018 till 14:00 Hrs.

PLL1920M LED LCD Monitor

3M Fiber Optic Wall Mount Enclosure 8430 Series

Reference Manual. Notes 9/16 Series H

RHP-IS DESKTOP IMAGE STABILIZER

BROTHER, BROTHER & SONS

DISTRIBUTION AMPLIFIER

CM-S38901SV TVL IR Long Range camera

DH5e-V2. Delta 5 On-Camera 4K HDMI Monitor with 3D LUTs. Quick Start Guide. What s Included

SmartCrystal Cinema Neo

FD Trinitron Colour Television

In-Ceiling Electric Motorized Front Projection Screen Evanesce Series. User s Guide

ezframe Series Section 1: Screen Design Section 2: Product Features PRODUCT WHITE PAPER What is it for?

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS

3M Coupling Tray FDTC-10S-COUPLINGTRAY

MedicaLight SURGICAL AND EXAMINATION LIGHTING

CHECK LINE. Model LS-36-LED. Stationary Stroboscope. Operating Manual BY ELECTROMATIC

KODAK Video Monitor CFH-V10

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

ULTRA-TRAC APL INSTRUCTION MANUAL. Read and understand instructions before use. Patented. 851 Transport Drive Valparaiso, IN

INSTRUCTION & INSTALLATION GUIDE

SIM 5 series Lamp Kit

High Performance DL-60 (Gold Plus) (7 in - 13 in) Dual Lane Spliceable Tape Feeder Part Number: Revision 3 Sep No.

The Versatile and Powerful ACLxy. ACLxy

SCREEN WINCH SYSTEM INSTALLATION MANUAL FOR SCREENS UP TO 300 cm. of width

With 3 models available, you can choose the one that meets your needs.

Watkiss PowerSquare CREATIVITY ACCURACY EFFICIENCY. Watkiss Print Finishing Watkiss PowerSquare. Watkiss Vario Collating and Finishing System

USER MANUAL. 27" 2K QHD LED Monitor L27HAS2K

Intra-oral X-ray unit with high frequency DC generator. Product Data

CM-S23349SV. Vari-Focal IR Bullet Camera

Creating room. for your visions

INSTRUMENT CATHODE-RAY TUBE

SCREEN WINCH SYSTEM INSTALLATION MANUAL FOR SCREENS FROM 300 cm. UP TO 450 cm. of width

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System

Check what you have received against the component checklist and hardware above.

USER MANUAL. GOLDMUND LOGOS 1N-2N SPEAKER SYSTEM Active Speaker

Interactive Satellite Terminal Installation / Validation Manual

SLiC Fiber Aerial Closure System

TV Lift System Model CL-65 Installation Instructions

Website: Tel: ADDRESS: 6475 Las Positas Rd. Livermore, CA Item No. E5B/E5S Installation Guide

PHI 5000 VersaProbe TM Operator s Guide

SHOWLINE SL BEAM 100 LED LUMINAIRE SPECIFICATIONS.

Orbit TM DIGITAL SHAKERS

Operating instructions Retro-reflective sensor. OJ50xx laser / / 2010

Outdoor IR Audio Camera

Dragonfly Quad. User Manual V1.4. Order code: EQLED101

SAFETY WARNINGS AND GUIDELINES... 3 INTRODUCTION... 4 CUSTOMER SERVICE... 4 PACKAGE CONTENTS... 4 RECOMMENDED TOOLS... 6 CONTROL PANEL OVERVIEW...

USER MANUAL. 22" Class Slim HD Widescreen Monitor L215DS

Introduction. Characteristics (Note 1, 2) Parameters Typical Value Units Conditions

Contents. 1. General Information. 2. Contents. 3. Operating Instruction. 4. Program update. 5. Trouble Shooting. 6. Specifications

DD E SIG N E D & E N GIN EE R E. Holman Garden Lights App Instruction Manual. Android Android manual

Physical Processing Guidelines for Shelf- Ready Chinese Approval Materials

DEEPFRAME BASIC KIT- USER MANUAL VERSION ORIGINAL USER MANUAL

HP Engage One 10.1 Display (Black) HP Engage One 10.1 Display (White)

ivw-fd133 Video Wall Controller MODEL: ivw-fd133 Video Wall Controller Supports 3 x 3 and 2 x 2 Video Wall Array User Manual Page i Rev. 1.

PL2410W LCD Monitor USER'S GUIDE.

USER MANUEL. SNIPE 2 Ref R13

Quick Setup 1 Unpack the projector

INSTRUMENT CATHODE-RAY TUBE

Transcription:

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

Use of the Guide Carestream CR Systems are designed to meet international safety and performance standards. Personnel operating the unit must have a thorough understanding of the proper operation of the system. This guide has been prepared to aid medical and technical personnel to understand and operate the system. Do not operate the system before reading this manual and gaining a clear understanding of the operation of the system. If any part of this manual is not clear, please contact your Carestream Health representative for clarification. Authorized European Representative Carestream Health France 1, rue Galilee 93192 NOISY-LE-GRAND CEDEX France Carestream Health, Inc. Rochester, N.Y., 14608 CARESTREAM is a trademark of Carestream Health. Carestream Health, Inc., 2010 8H8437-01 ii

Table of Contents 1 Introduction Document Conventions...1-1 Indications for Use...1-1 2 Safety and Related Information Medical Equipment Classification...2-3 Conforming Standards...2-3 Cassette Disposal...2-4 3 LLI System Components Overview System Components...3-5 Long-Length Cassette...3-5 Portable Grid...3-6 Wall-Mounting the Portable Grid...3-7 Vertical Cassette Holder...3-11 4 Processing a Long-Length Image Recommended X-Ray Settings...4-16 Using Beam Attenuation Filters...4-16 5 Scanning an LLI Cassette Entering Patient Data...5-19 Processing the LLI Cassette...5-20 6 LLI Troubleshooting Errors, Causes, and Corrective Actions...6-23 i

-ii

1 Introduction Document Conventions The information contained herein is based on the experience and knowledge relating to the subject matter gained by Carestream Health, Inc. prior to publication. No patent license is granted by this information. Carestream Health reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information. Carestream Health shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by Carestream Health s negligence or other fault. NOTE: From this point on, CARESTREAM VITA/VITA LE/VITA SE CR LONG LENGTH IMAGING System will referred to as LLI or LLI system. NOTE: Notes provide additional information, such as expanded explanations, hints, or reminders. IMPORTANT: Important highlights critical policy information that affects how you use this manual and this product CAUTION: Caution points out procedures that you must follow precisely to avoid damage to the system or any of its components, yourself or others, loss of data, or corruption of files in software applications. Indications for Use The LLI System generates and controls images in long-bone regions. 8H8437-01 1-1

Introduction 1-2 8H8437-01

2 Safety and Related Information Medical Equipment Classification Conforming Standards Mechanical Equipment This equipment cannot be classified for electrical safety because it does not use or generate electrical energy. Provides ordinary protection against harmful ingress of water (IPXO). Suitable for continuous operation. Not suitable for use in the presence of flammable materials. The following standards apply to the Long-Length Vertical Cassette Holder only: UL 60601-1:2003 - Medical Electrical Equipment, First Edition CAN/CSA-C22.2 No. 601.1-M90 (R2001) - Medical Electrical Equipment CAN/CSA-C22.2 No. 601.1S1-94 (R1999) Supplement No. 1-94 to CAN/CSA-C22.2 No. 601.1-M90 CAN/CSA C22.2 No. 601.1B-90 (R2002) - Amendment 2 to CAN/CSA-C22.2 No. 601.1-M90 IEC 60601-1:1998, + Amendment 1:1991, + Amendment 2:1995 - Medical Electrical Equipment EN 60601-1:1990, + Amendment 1:1993, + Amendment 2:1995, +Amendment 13:1996 - Medical Electrical Equipment IEC / EN 60601-2-32:1994 - Safety of Associated Equipment of X-ray Equipment 8H8437-01 2-3

Safety and Related Information Cassette Disposal CAUTION: CARESTREAM VITA CR Cassettes contain lead. Disposal of lead may be regulated due to environmental conditions. For disposal or recycling information, contact your local authorities. For further information concerning these products, inside the United States, call Carestream Health, Inc. at 1-800-328-2910 or if outside the United States, write to Carestream Health in your country. 2-4 8H8437-01

LLI System Components Overview 3 LLI System Components Overview System Components CARESTREAM VITA CR LONG LENGTH IMAGING System Long-Length Cassette Long-Length Phosphor Screen (84 x 34 cm) Long-Length Portable Grid Long-Length Vertical Cassette Holder with Grid Long-Length Cassette Cassette Specifications Width Length Thickness Maximum load Weight 35 cm (14 in.) 86 cm (33 in.) 15 mm (0.591 in.) 158.76 kg (350 lb) 4.5 kg (10 lb) Cassette Labels Bar Code Labels located in a product label recess on the non-tube side panel of the cassette. Process Indicator Labels a large number 1 and a large number 2 provide a reminder to process both ends of the cassette. NOTE: The image area is approximately 1.5 cm (inside the cassette s length and 1.3 cm inside the width. Position the area of interest within these dimensions. Operation To use the cassette, expose the patient once using the length of the cassette. Scan both ends of the cassette. The software stitches the image together automatically. See Processing the LLI Cassette on page 5-20. 8H8437-01 3-5

LLI System Components Overview Portable Grid The Portable Grid has been designed to accommodate the Long-Length Cassettes. It gives you greater flexibility to take long-length images outside the radiology department. ➀ ➁ ➂ H193_0506GC Number Feature Function ➀ Handles Make it easy to transport. You can also use the handle to hang the grid on the wall. ➁ Markers Identify the tube side for exposure. Each marker provides an approximate boundary for the phosphor screens inside. When transporting or storing the grid, make sure that the markers are facing out. ➂ Rollers Make it easy to transport. For safety reasons, the rollers rotate in only one direction. See Moving and Storing the Portable Grid on page 3-7. 3-6 8H8437-01

LLI System Components Overview Wall-Mounting the Portable Grid You can mount the Portable Grid on the wall using brackets or by hanging the handle on a hook (brackets and hook not supplied). H193_0526GC Wall-Mounting Long-Length Portable Grid Grid Specification Specification Description Ratio 8:1 Source to Image Distance (SID) 102 183 cm (40 72 in.) Focal Distance 150 mm Density 103 lines per inch (40 lines per cm) Dimensions 15 x 34 in. Moving and Storing the Portable Grid To move or store the Portable Grid, grasp the handle and pull the grid so that the screen markers are facing up. The wheeled grid makes moving the portable system easy to do. For best results, do not push the grid. 8H8437-01 3-7

LLI System Components Overview H193_0516GC CAUTION: The rollers can turn if the grid is stored upright with the processing labels facing out, and the unit could fall. Make sure you store the grid with the markers facing out as shown. H193_0510GC CAUTION: Handle the grid with care. Dropping or bumping the grid causes serious internal damage that produces permanent, unacceptable imaging artifacts. 3-8 8H8437-01

Point-of-Care Long Length Imaging System 14" x 33" Kodak LLI System Components Overview Loading and Unloading the Portable Grid To load the cassette into the grid from an upright position: 1. Hold the grid upright. 2. Hold the portable cassette at a slight (15 ) angle so that the lower right corner enters first. Let the weight of the cassette press down the release latch. 3. Slide the cassette into the grid. Point-of-Care Long Length Imaging System CR Cassette with Screen 14" x 33" Kodak CR Cassette with Screen 8H8437-01 3-9

CR Cassette with Screen Point-of-Care Long Length Imaging System 14" x 33" Kodak CR Cassette with Screen 14" x 33" Kodak Point-of-Care Long Length Imaging System LLI System Components Overview To load the cassette into the grid from a horizontal position: 1. Turn the grid so that the open back side is facing you. 2. Place the grid on a table or floor. 3. Hold the cassette flat with one corner leading into the grid. 4. Push the cassette into the grid. Let the weight of the cassette push down on the release latch. CAUTION: Do not load or unload the cassette while the patient is in contact with the grid. H193_0547BC 3-10 8H8437-01

LLI System Components Overview Vertical Cassette Holder The Vertical Cassette Holder with Grid contains an X-ray grid for optimizing image quality. The cassette holder lets you expose long-length images on a single cassette in a vertical position. You can move the cassette holder up or down to position the cassette for any long-length exam, in the upper region or lower region, as shown: Upper and Lower Regions H193_1075AC H193_1074AC Insert the cassette in the upper region for spine studies or the low region for leg studies. Grid Specification Specification Description Ratio 8:1 (Source to Image Distance (SID) 102 183 cm (40 72 in.) Density 103 lines per inch (40 lines per cm) Dimensions 15 x 34 in. 8H8437-01 3-11

LLI System Components Overview ➂ ➀ ➃ ➁ ➂ H193_1076GC Vertical Cassette Holder 1. Handle Lock Pin pull to unlock the cassette holder s position, release to lock. 2. Grid Cabinet holds the cassette. 3. Attachment Bar use to attach the cassette holder to the wall. 4. Handle push up or down to change the position of the cassette holder. 3-12 8H8437-01

LLI System Components Overview Loading a Cassette To load a cassette into the vertical cassette holder: 1. Select a long-length cassette. Slide the cassette into the cassette holder. Loading a Cassette H193_1077GC lock pin H193_1078GC Moving the Cassette Holder Into Position 2. Pull and hold the lock pin, and adjust the height of the grid cabinet. Release the pin to lock it into a pre-selected opening. 8H8437-01 3-13

LLI System Components Overview 3-14 8H8437-01

Processing a Long-Length Image 4 Processing a Long-Length Image 1. Insert the Long-Length CR Cassette into the cassette holder with the tube side toward the X-ray machine. 2. Position the patient. 3. Set the Source to Image Distance (SID). Vertical Casette Holder: Enter a value from 102 cm to 183 cm (40 to 72 in.) ± 1 cm. Portable Grid: Enter a value from 102 cm to 183 cm (40 to 72 in.) ± 1 cm. Verify that the entire length of the cassette is illuminated by the collimator. 4. Set the X-Ray equipment. See Recommended X-Ray Settings on page 4-16. 5. Set the Beam Attenuation Filter (BAF). See Using Beam Attenuation Filters on page 4-16. 183 cm (72 in.) recommended 102 cm (40 in.) minimum X-ray Source 84 cm (33 in.) H193_0034BC 8H8437-01 4-15

Processing a Long-Length Image X-ray Source 102 cm (40 in.) minimum 183 cm (72 in.) maximum 150 cm (59 in.) recommended H193_0524GC Recommended X-Ray Settings Using Beam Attenuation Filters The settings in the table below are general guidelines. Optimal settings may vary according to patient size and site equipment.. Organ and Position Kv mas Filter model Full Spine AP child 85 50 57-405 Full Spine Lateral child 96 125 57-407 Full Spine AP 96 65 57-405 Full Spine Lateral 110 160 57-407 Beam Attenuation Filters (BAF) help ensure uniform exposure by compensating for varying degrees of thickness on the body. Two filters are recommended: Filter Type Transparent Transparent Image Full spine AP / Full leg Lateral Full Spine Place the BAF holder (model 57-426) onto the x-ray collimator as shown in 4-16 8H8437-01

Processing a Long-Length Image the figure. BAF Holder Transparent acrylic BAF H220_0023AC Beam Attenuation Filter Placed in BAF Holder Position the BAF according to the patient anatomy shown in the figures (AP and Lateral) and according to Recommended X-Ray Settings. Note that the transparent BAF shadow on the patient's body is visible. For full spine procedure, adjust the lower edge shadow under the xiphoid part in the AP projection and on the xiphoid part on the lateral projection. For full leg images, position the lower border shadow of the BAF on the lowest part of the anatomy visible (ankle area) since this is the thinnest area in the image. Top border of BAF shadow Top edge of BAF shadow Top edge of Bottom border of BAF shadow H220_0024BCA H220_0024BC A B C Bottom edge of BAF shadow under Bottom edge of BAF shadow under the xiphoid part 8H8437-01 4-17

Processing a Long-Length Image Proper placement of BAF shadows, AP projection Top edge of BAF shadow Top edge of Bottom edge Bottom edge of BAF shadow under the xiphoid part H220_0025BCA H220_0024BCA H220_0025BC Proper Placement of BAF Shadows, Lateral Projection Perform the exposure. Wait 2-3 minutes before starting the scan process to allow the energy on screen to stabilize. 4-18 8H8437-01

5 Scanning an LLI Cassette Entering Patient Data 1. Start the CARESTREAM VITA CR System Software. 2. Enter your username and password. 3. Select Patient in the lower left corner of the screen. Entering Patient Data NOTE: See the "CARESTREAM VITA CR System Software User s Guide" for information about entering patient data. 4. Search for an existing patient or enter a new patient record. 5. Select Acquire. 6. Select a body part from one of the following: Full leg lateral Full leg AP Full spine lateral 8H8437-01 5-19

Scanning an LLI Cassette Full spine AP Processing the LLI Cassette 1. Extend the support arm on the front of the CR System. NOTE: When extended, the support arm aligns the long-length cassette properly and keeps the cassette level while scanning. After scanning, push the arm all the way back in. 2. Insert the end of the cassette labeled 1 into the CR System. Inserting the Cassette into the System H232_0002AC NOTE: If auto-scan is selected, scanning begins automatically. 3. Select Acquire. The reader scans the first half of the plate. 5-20 8H8437-01

Scanning an LLI Cassette First Half of LLI Exposure The CR System extracts the phosphor screen and draws it into the drum where one side is read and erased. The screen re-enters the cassette. 4. Remove the cassette from the processor, turn it 180 degrees, and insert side two into the processor (end of cassette labeled "2"). 5. Select the second thumbnail. Processing the Other End of the Cassette. H232_0003AC 8H8437-01 5-21

Scanning an LLI Cassette After the second long-length image is scanned, the screen is erased and returns to the cassette. The images are stitched together automatically. The stitched image appears as a third thumbnail image. Stitched Image Displayed NOTE: You can view each image using by selecting Full Screen viewing. See the CARESTREAM VITA CR System Software Guide for other viewer options. IMPORTANT: You must select Accept or Reject after scanning is complete before you can exit the screen. 6. Select Accept to send the stitched image to DICOM Workstations, PACs, or printer destinations. Select Accept All to send the stitched image and its components to destinations. Select Reject to send the the stitched image to the database as "Rejected." The resulting image will not be sent to PACs. 5-22 8H8437-01

6 LLI Troubleshooting Errors, Causes, and Corrective Actions Error Message Cause Action The cassette Bar Code ID does not match the top side. Operation is cancelled. Can t choose an LLI sub-organ without a license. Operation is cancelled. Can t choose LLI sub-organ after regular sub organ was chosen. Operation is cancelled. Can t choose bottom before top is scanned, or scan top side twice. Operation is cancelled. Can t choose a sub-organ which is not for LLI during LLI study. Operation is cancelled. The cassette is not an LLI cassette. Please either replace the cassette or re-select the sub-organ. Can t choose a sub-organ during LLI procedure. Press the Cassette Size button and select a cassette. The cassette does not match the selected sub-organ. Please either replace the cassette or reselect the sub-organ. Can t choose a sub-organ during scanning. Operation is cancelled. A wrong cassette was used by mistake. A license may not be present. A non-lli study was started, but an LLI body parts was selected. Side two was scanned first. or side one was scanned twice. A non-lli body part was selected during an LLI study. A standard cassette was scanned as a long-length cassette. The cassette type is not compatible with the sub-organ. The user attempted to select a new body part between scans one and two. The bar code label fails or the cassette bar code is not recognized. The cassette type is not compatible with the sub-organ. The user attempted to select a new body part between scans one and two. Check that the barcode label does match (bad cassette if not). Or, make sure the first and second sides belong to the same cassette. Verify presense of a license, by going to the About tab. Go to the patient list, reselect the patient, and select non-lli body part, or select a new patient for an LLI study. Scan side one first, then scan side two. See Processing the LLI Cassette on page 5-20. Once you begin an LLI procedure, you must select an LLI sub-organ. Select a long-length cassette or select a standard scan. To choose a different sub-organ after LLI scanning, go to the patient list and select the patient again without the LLI procedure. Make sure the bar code label is readable. Allow scan to continue, and choose LLI1 or LLI2. Select a long-length cassette, or select a standard scan. No thumbnail is added. See Processing the LLI Cassette on page 5-20. 8H8437-01 6-23

LLI Troubleshooting 6-24 8H8437-01

Publication History Publication Date Publication Number Changed Pages Notes 15JULY2010 8H8437 New Publication 08NOV2010 8H8437-01 1,2 Updated Title, changed Authorized European Agent information 0086

Carestream Health, Inc. Rochester, N.Y. 14608 Carestream Health, Inc., 2010 8H8437-01