Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK

Similar documents
Liquid Chromatography- Mass Spectrometer Manual

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

STANDARD OPERATING PROCEDURE:

Staying Connected in GC

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

AIM INTRODUCTION SIMPLIFIED WORKFLOW

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

Creating and fully utilizing the retention time standard processing method for a more automated IC workflow

Performance. Reliability. Productivity. Automated Flash Chromatography Systems

EXA PH200/400 and EXA PH202/402 Troubleshooting and Error Code Guide

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

Waters Synapt High Definition Mass Spectrometry System Quick Start Guide

How Quadrupole Size Affects Research

Operating pressure should be 6000 psi or better suitable for both analytical and semi-preparative applications

Overview and Interpretation of D7900/D7169 Merge Analysis

GC/LC-MS: data acquisition rate and peak reconstruction

Achieving More Efficient Data Review with OpenLAB CDS

Online Process Control using Modular Fluid Delivery and Fast Process Gas Chromatography: From the Sample Point to the DCS Connection

Cross Examination of the Criminalist. Felipe Plascencia

Sample Analysis Design. Element2 - Basic Software Concepts (cont d)

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer

UNITY 2 TM. Version 1.0. February 2008

NGC8206 Start-Up Guide

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

NMR. picospin. Maintenance Guide

Model CB 60 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus:

PAL LSI - Robust and Reliable Sample Handling

Design of NIR Chromenylium-Cyanine Fluorophore Library for Switch-ON and Ratiometric Detection of Bio-Active Species in Vivo

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

Real-time ph and Conductivity Data Acquisition with the ACQUITY UPLC H-Class Bio System, Auto Blend Plus, and the GE Healthcare Monitor ph/c-900

Ecoline S series. Precision solutions for quality production. Innovations for a better world.

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Bruce Quimby. Agilent Technologies October 28, For Forensic Use. Page 1

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

Linac3 experience for LHC ion runs

C200H-AD002/DA002 Analog I/O Units Operation Guide

ChromQuest 5.0 Chromatography Data System

A6 OPERATING INSTRUCTIONS

Instrument Control and Data Acquisition with Chromeleon

WELDING CONTROL UNIT: TE 450 USER MANUAL

Detailed Design Report

ST-200Pro. Electrolyte Analyzer OPERATOR S MANUAL

MITSUBISHI ELECTRONICS AMERICA, INC.

Henkel Installation Handbook LINEGUARD 2001

Model CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

Control Unit CU 2.1 User Manual V6850A HPLC

Agilent 7000 Series Triple Quadrupole GC/MS System

PAL LSI - Robust and Reliable Sample Handling

Fragment Analyzer Genomic DNA Analysis Kit

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Limitations of a Load Pull System

SPECIFICATION FIBER OPTIC SPLICE CLOSURE. Spec No : VSS-1007-BS403A-04A/SD. VSS-0107-BS403A-04A/SD R & D Center Manufacturing Division

The Importance of Projector Maintenance.

Model KBF 1020 Constant climate chambers with large temperature / humidity range

ICS-3000 System Installation Qualification

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Operation of CEBAF photoguns at average beam current > 1 ma

SmartMix Manual V5350A

Finnigan PolarisQ User s Guide. PN , Revision D

Ladies and Gentlemen,

PHI 5000 VersaProbe TM Operator s Guide

EG4015. Digital Generator Governor Controller User Manual

Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

TVP-20 / RVP-412 CCTV VIDEO-PLEXER

Remote Control. degraded, causing unreliable operation. The recommended effective distance for remote operation is about 16 feet (5 meters).

Analog Devices Welcomes Hittite Microwave Corporation NO CONTENT ON THE ATTACHED DOCUMENT HAS CHANGED

2. Pin assignment list for EXC-73 Signal conversion

Quick Start Function Summary Instructions for ASHCROFT GC52 Differential Pressure Transmitter Version 6.03 Rev. B

9070 Smart Vibration Meter Instruction Manual

Inserting System K1600

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Developing an AFM-based Automatic Tool for NanoAsperity Quantification

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003

PINMRF. Checkout Quiz - Varian Inova-300 Version

TECHNICAL SPECIFICATION

PanelView 1400e CRT Maintenance

Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018)

Program Risks Risk Analysis Fallback Plans for the. John T. Seeman DOE PEP-II Operations Review April 26, 2006

AccuLoad II TM - STD Operator Guide

Release Note. Product Name: Current Version: Ver.Type Ver No. Date. RN Date: June 2017 Reference: Kramer Product Affected:

Model KBF 115 Constant climate chambers with large temperature / humidity range

Up to 85% higher Service Life due to efficient sealing method.

Operator s Manual. Ultegra. Health Scale. Fairbanks Scales by Fairbanks Scales Inc. All rights reserved. Revision 5 06/07

Operation Manual for the Refractive Index detector VE-3580

UNIGAS Hand-held combustion gas analyzer with 3 sensors

Model KBF P 240 Humidity test chambers with ICH-compliant light source

Fragment Analyzer Genomic DNA Analysis Kit

Troubleshooting. 1. Symptom: Status indicator (Red LED) on SSR is constant on. 2. Symptom: Output indicator (Yellow LED) on SSR is flashing.

Model ED 400 Drying and heating chambers Classic.Line with natural convection

CR7000. CRT Analyzer & Restorer. Easily Test And Restore CRTs With The Most Complete Tests Available For Added Profit And Security.

Fragment Analyzer CRISPR Discovery Gel Kit

Transcription:

Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK Science for a safer world

13/10/2016 2 2

What is a System Suitability Test (SST)? Also referred to as a System Suitability Sample (SSS) A standard or extracted sample (*in extraction solvent) Usually contains internal standard Stored in a stable manner if possible Sample ID Rack Code Vial Position Reagent Blank DWP96 1 Reagent Blank DWP96 2 SST DWP96 3 Reagent Blank DWP96 4 Injected onto your LC-MS before you submit your sample batch 13/10/2016 3

What does it tell us? 1) Retention Time 2) Instrument Response (or Signal:Noise) 3) Peak Shape When combined with a back pressure reading it provides us with a wealth of knowledge.

What type? Generic OR Specific 13/10/2016 5

What do the SST and Back Pressure tell us? 1) The HPLC is OK Mobile Phases Column Flow rate Column Oven Auto-sampler Injection Seals Valves (Carry-over)

What do the SST and Back Pressure tell us? 2) The Mass Spec is OK Ionisation Gas Flow Source Ion transmission Detection

What doesn t the SST and Back Pressure tell us? Sample extraction issues Contaminants from the extraction solvents (Reagent Blank/Control Blank)

Setting Up the Acceptance Criteria 1) Determine SST concentration (LLOQ or match to LLOQ?) 2) Run before every validation batch to obtain a feel for normal Instrument API3200 Back Pressure at t0 Retention Time Intensity S:N P A/S Maintenance Notes DATE Pump A/B Pump C/D 25OHD3 25OHD2 25OHD3 25OHD3 27/03/2014 386 3876 4.68 4.96 2400 38 1.2 27/03/2014 396 4029 4.77 5.15 1900 26 1.6 LLOQ looks poor, QC Low out 28/03/2014 386 3813 4.66 4.93 2100 35 1.3 Fresh Mobile Phases 01/04/2014 383 3804 4.67 4.95 2000 33 1.2 Weekly & Monthly 02/04/2014 372 3632 4.69 4.98 2350 37 1.3 08/04/2014 379 3687 4.69 4.96 2400 34 1.2 Weekly 11/04/2014 385 3235 4.91 5.26 1550 18 1.2 Leak in column oven 11/04/2014 374 3645 4.68 4.92 1950 35 1.2 14/04/2014 385 3636 4.68 4.95 2250 37 1.2.......... 3) From these experiments set the SST acceptance limits 4) You will need to do this for each type of LC-MS instrument 13/10/2016 9

Retention Time: 4.50-4.80 Minimum Intensity 1200cps Minimum Signal:Noise 20:1 Peak Asymmetry (A S ) h A C Time A s = CB/AC B } 10 % of peak height (h) Plate Count (N) t N 5.54 W R 0.5 2

Different machines 25OH Vitamin D SST XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 137 (2014_08-04 post weekly) of SST.wiff (Heated Nebulizer) Max. 3945.0 cps. 4800 4600 4400 4200 4000 3800 3600 3400 3200 25OHD3 RT: 4.64min 4.64 4.90 Older instrument In te n s ity, c p s 3000 2800 2600 2400 2200 2000 1800 1600 1400 1200 1000 Lower Limit 2400cps 3-epi- 25OHD3 800 600 400 200 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 Time, min XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 quant from Sample 28 (2014_08_04) of SST.wiff (Heated Nebulizer) Max. 7.3e4 cps. Newer instrument In te n s ity, c p s 7.3e4 7.0e4 6.5e4 6.0e4 5.5e4 5.0e4 4.5e4 4.0e4 3.5e4 3.0e4 2.5e4 2.0e4 1.5e4 1.0e4 5000.0 Lower Limit 2.0e4 cps 25OHD3 RT: 4.75min 0.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 1 55 108 162 215 269 322 376 430 483 537 590 Time, min 4.75 5.07 3-epi- 25OHD3 5.47 5.83

How much RT shift/ BP change/ Peak symmetry change is still OK? No hard and fast rule batch failures during establishment/validation as a guide If the method is already live or there weren t (m)any failures during method development then. Guess! Try not to make a rod for your own back (not too tight unless you have cause & don t set unnecessarily low LLOQs) Initial criteria are opinions that you then review over time Look at every batch failure as an opportunity to review these criteria to make it a better tool 13/10/2016 12

Retention Time Ensure that the criteria are set so that the peak doesn t shift into a drastic gradient change Monitor IS in batches with altered RT to look for ionsuppression issues Back Pressure Limits for fittings Check traces to see the % change over the gradient (extrapolate for changes at t0*) Peak Shape (A S and N) Known interferences? Integration Assists in judging column lifetimes 13/10/2016 13

Example: adding in known interferences XIC of +MRM (4 pairs): 383.200/257.300 Da ID: 25OHD3 1 from Sample 221 (2016-03-23 TN) of SST.wiff (Heated Nebulizer), Smoothe... Max. 4184.6 cps. 4185 4.75 4000 3800 OH 3600 3400 3200 OH 4.99 3000 2800 2600 25OHD 3 HO In te n s ity, c p s 2400 2200 2000 1800 HO 3epi -25OHD 3 1600 1400 1200 1000 800 600 5.29 400 200 0 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 6.0 Time, min

Example something a little different: Unknown interferences 1 Normetanephrine 2 3 Metanephrine

Something has gone wrong.

I can t see anything Step 1 start simple Correct method being run? Is there anything in the vial? Is the vial in the correct position? Did the auto-sampler pick up the sample? Needle depth? Wash solvents full? Is the LC connected to the MS? Is the back pressure OK? YES Spray in the source? Correct probe? (ESI vs APCI) corona needle? Divert valve correct? Check ion path Signal Minimum Intensity Retention Time Time NO Correct mobile phases? Lines not purged properly Too High Blockage in lines somewhere? Column too old? Too Low Leak somewhere? Fluctuating Leak on single pumphead Pump seals

Signal Signal What is it doing over there? Retention Time Time Retention Time Time No point looking at the Mass Spec Is the back pressure OK? Yes Mobile phase made correctly? Gradual change? (column degradation) Correct wash solvents loaded (peak shape OK?) Injection mode correct? No Mobile phase made correctly? All lines purged correctly? Leak somewhere (early and late peaks) Faulty pump (swap A&B lines and see if RT changes Gradual change? (column degradation) 13/10/2016 18

Right time - but it is a bit small Signal Minimum Intensity Can t rule out the HPLC or MS Back pressure will often be OK but check anyway Time SST sample mixed properly? Mobile phase made correctly? Auto-sampler working properly?.. blockage/ leak/ wash solvent/ injection mode (*can check with manual injector) Leak somewhere after the column? Correct ion probe? Capillary extended in probe? APCI Corona needle positioned correctly? Mass Spec within calibration? Gradual change across all methods? (MS needs cleaning) Sudden drop MS source/ion guide/q1 may need cleaning Vacuum OK? Gas pressures OK? 13/10/2016 19

Where did all that come from? Signal Minimum Intensity Time SST sample mixed properly? Mobile phase made correctly? Right amount injected? Mass Spec set to correct resolution? Correct ion probe? Was MS incorrectly calibrated before? Recent maintenance performed? (Source/Gas/Vacuum) Is it really a problem? Can you just inject less? 13/10/2016 20

It looks ugly SST sample solvent correct? Auto-sampler wash solvents OK? Mobile Phases OK? Pumps OK and delivering gradient? Leak somewhere? Peak tubing cut squarely and fitted correctly? Column degradation? 13/10/2016 21

It s EVERYWHERE! Signal Minimum Intensity Reagent Blank Sample ID Rack Code Vial Position Reagent Blank DWP96 1 Reagent Blank DWP96 2 Signal Reagent Blank SST DWP96 3 Reagent Blank DWP96 4 Signal SST Ghost Peaks are often a sign that your mobile phase or auto-sampler wash contains a contaminant Signal Reagent Blank Time 13/10/2016 22

It wont go away Carry-over peaks point the finger at: Auto-sampler wash not working Internal damage to injection port Signal Signal Minimum Intensity Reagent Blank Reagent Blank Issue with peak not being fully eluted from SST Peak tubing not being cut squarely and fitted correctly Signal SST Signal Reagent Blank Time 13/10/2016 23

Questions? Science for a safer world