SITRANS F M. Transmitter MAG 5000/6000 4/33. Overview

Similar documents
SITRANS F M. Transmitter MAG 5000/6000 4/33

Flow Measurement SITRANS F M

SITRANS F flowmeters SITRANS F M. Transmitter MAG 6000 I/6000 I Ex d e 4/43

Flow Measurement SITRANS F M

Analog Input Module HART Ex i / I.S. Inputs, 8 Channels Type 9461/

Analog Input Module HART Ex n Inputs, 8 Channels Series 9461/15

Capacitance Level Measurement Electronic insert FEC 12

Level Measurement Point level measurement - RF Capacitance switches

Level instruments. Continuous level measurement - Radar transmitters SITRANS LR250. 5/224 Siemens FI Overview.

TFRH Hygienic RTD temperature sensor. CombiTemp TM

Level Measurement Point level measurement RF Capacitance switches

Process Transmitter RMA 422

9 Control Equipment. Measurement- and Control Equipment Series 8214, 8218

Process transmitter RMA422

Level instruments. Point level measurement - Capacitance switches. Pointek CLS100. 5/10 Siemens FI Overview.

Level Measurement Continuous level measurement - Capacitance transmitters

MB 204/MB 204-Ex. Multi-Barrier for PROFIBUS PA and FOUNDATION Fieldbus 10/ EN

TFRN Industrial RTD temperature sensor. CombiTemp TM

TFR5 Room or outdoor temperature sensor. CombiTemp TM

DS 200 P DS 200 P. Electronic Pressure Switch with Flush Process Connection

Inductive sensor. 2-wire, analog output BI8-M18-LI-EXI

DMP 335 DMP 335. Industrial Pressure Transmitter. Welded, Dry Stainless Steel Sensor. Pressure Transmitter. Industrial

Fibre Optic Modem ODW-611

TFRN Industrial RTD temperature sensor. CombiTemp TM

DS 400 P. Intelligent Electronic Pressure Switch in Hygienic Stainless Steel Ball Housing. on hygienic process connections

Wall-/Pipe-Mount Level Indicator

TFR5 Room or outdoor temperature sensor. CombiTemp TM

FLOMAT. Flowmeters. Electromagnetic Insertion Flowmeter: Series. Instrumentation for fluids

DS 210. Electronic Pressure Switch. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure

Inductive sensor BI1.5-EG08K-Y1

Inductive sensor NI3-EG08K-Y1-H1341

Dissolved Oxygen Transmitter

DMP 331 DMP 331. Industrial Pressure Transmitter. for Low Pressure. Stainless Steel Sensor. Pressure Transmitter. Industrial. tri-matic.

DATA SHEET. The Advantys model STBDAI5260 Digital Input Module provides 2 isolated discrete input points that operate on a 115 VAC power source.

MID-EX-E. Figure: MID-EX-E*010* ka052000en; Rev. 2.3 J

Pendant Light Fitting IEC Series 6470/2

Modular Lube Lubrication Systems System Controls

ISOMAG THE FRIENDLY MAG METER. Gas, mists or vapours Zone 0 1G Gas, mists or vapours Zone 1 2G. Gas, mists or vapours Zone 2 3G

FLOMAT Electromagnetic Insertion Flowmeter

Level Measurement. Continuous level measurement Radar transmitters. SITRANS LR250 Flanged Encapsulated Antenna. 4/228 Siemens FI

- special pressure ranges - variety of electrical and mechanical. Characteristics. - other versions on request

Magnecraft General Purpose Relays 750R Series DPDT and 3DPT, 10 A

RIA46. Technical Information. Field meter Digital field meter with control unit for monitoring and visualizing analog measured values

Advantys STB DATA SHEET. Advantys STB Output Module. Description. Features

Magnecraft Power Relays

Dissolved Oxygen Transmitter

Inductive sensor With analog output BI15-M30-LI-EXI

DMP 331i / DMP 333i LMP 331 i

Level Measurement Continuous level measurement Radar transmitters

AF-300 E11 Adjustable Frequency Drive

Ex td A21 IP66 T58 C. max. 3 x 4 mm 2 or 1 x 10 mm 2 with pin cable lug + 1 x 2.5 mm 2 fine wire. max. 4 x M20 metal thread

Fibre Optic Modem ODW-622

LAUREL ELECTRONICS, INC.

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0,35 % FSO. Nominal pressure

ODW-621. RS-232 Point-to-point applications

Single Point Flow Monitor SW118 SW119

PSI-MOS-RS232/FO 850 E Serial to Fiber Converter

CAPACITIVE SENSORS KAS. Series 26. Level Control Food Industry Pharmaceutical Industry Semiconductor Industry Chemical Industry

Nessie Coils for VDH and VDHT valves

DMP 331i DMP 333i. Precision Pressure Transmitter. Stainless Steel Sensor. accuracy according to IEC 60770: 0.1 % FSO

DATA SHEET ML210 ML210_IT_EN_DS017REV12_IS

DMP 339. Pressure Transmitter. with G ¼" flush diaphragm. Pressure Transmitter. Industrial. accuracy according to IEC 60770: 0.

Fibre optic router for TP/FT-10 LRW-112PP

Mechanical flow switch for On/Off control

DMP 335. Industrial Pressure Transmitter. Welded, Dry Stainless Steel Sensor. accuracy according to IEC 60770: 0.5 % FSO.

Level Measurement silometer FMC 420, FMC 423

Datasheet - SRB301LC 24VAC/DC

SQM40/41 Actuators for air and gas dampers

Inductive sensor NI10-M18-Y1X-H1141

ABB. Measure IT Uvisor System MFD Multi Flame Detector - Control Unit. Product Data Sheet (Rev. C)

Level Measurement Continuous level measurement - Radar transmitters

Level Measurement Continuous level measurement - Radar transmitters

LMP 331 LMP 331. Screw-In Transmitter with piezoresistive Stainless Steel Sensor

Datasheet EC-750 Ex DOME Camera

Generator protection relay

Level Measurement Continuous level measurement Radar transmitters

LONWORKS Fibre Optic Converter

Magnecraft General Purpose Relays 782 Power Series SPDT 20 A; DPDT 15 A

Guide for installers. METTLER TOLEDO MultiRange System solution analog Ex1. Hazardous area. Safe area

Contactor 4 kw / 400 V with 3 Main Contacts and max. 4 Auxiliary Contacts Series 8510/122

LONWORKS Fibre Optic Router

Sensor technologies Proximity sensors Series ST4. Brochure

DMP 331Pi. Precision. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm. Pressure Transmitter.

Ordering details. Approval. Classification

DATA SHEET Panorama Rudder Indicator, TRI-2

PROPORTIONAL VALVES Sentronic plus

Water temperature recommended range 5 45 C ( F) UV transmittance 254 nm, 1 cm) min. 45 %

3.3. Ex-Escape sign luminaires. Ex-Lite Metal version with LED technology for Zone 1 and Zone 21 / NEC applications

ELE C TRO MAG N E T I C FLO W M E T E RS

LD OEM/LD PDS/LD PeCo

Datasheet - SRB-E-201ST

RCA3 SOLENOID PILOT VALVES PENTAIR CLEAN AIR SYSTEMS

Magnetic field sensor for pneumatic cylinders BIM-IKT-Y1X

RMA42. Technical Information

Guide for installers. METTLER TOLEDO MultiRange System solution Point Ex.

Photoelectric Sensors E3F2

2000 W, 2450 MHz Microwave Generator GMP 20K SM 56M230 FST 3 IR

1.1. IEC Contactors and Starters. Contents Description Relays and Timers... Miniature Controls... Contactors and Starters. XT IEC Power Control

Liquid Level Sensor Type AKS 4100U

TEK-LCD 7801A. NEMA 4X Loop-Powered Feet and Inches Level Meter. ACCESSORIES. Technology Solutions

Transcription:

Siemens AG 2011 Flow Measurement Overview Design The transmitter is designed as either IP67 NEMA X/6 enclosure for compact or wall mounting or 19" version as a 19 insert as a base to be used in: 19" rack systems Panel mounting IP20/NEMA 1 (prepared for IP65/NEMA 2 display side) Back of panel mounting IP20/NEMA 1 Wall mounting IP66/NEMA X Several options on 19 versions are available such as: Transmitters mounted in safe area for Ex ATEX approved flow sensors (incl. barriers) Transmitters with electrode cleaning unit on request compact version (left) and 19 insert version (right) The MAG 5000 and 6000 are transmitters engineered for high performance, easy installation, commissioning and maintenance. The transmitters evaluate the signals from the sensors type MAG 1100, MAG 1100 F, MAG 3100, MAG 3100 P and MAG 5100 W. Transmitter types: MAG 5000: Max. measuring error ± 0. % ± 1 mm/s (incl. sensor) MAG 6000: Max. measuring error ± 0.2 % ± 1 mm/s (incl. sensor, see also sensor specifications) and with additional features such as: "plug & play" add-on bus modules; integrated batch functions. Benefits Superior signal resolution for optimum turn down ratio Digital signal processing with many possibilities Automatic reading of SENSORPROM data for easy commissioning User configurable operation menu with password protection. 3 lines, 20 characters display in 11 languages. Flow rate in various units Totalizer for forward, reverse and net flow as well as additional information available Multiple functional outputs for process control, minimum configuration with analogue, pulse/frequency and relay output (status, flow direction, limits) Comprehensive self-diagnostic for error indication and error logging (see under diagnostics) Batch control (MAG 6000 only) Custody transfer approval: PTB, OIML R 75, OIML R 117, OIML R 9, MI-001 and PTB K 7.2 for chilled water MAG 6000 with add-on bus modules for HART, FOUNDATION Fieldbus H1, DeviceNet, Modbus RTU/RS85, PROFIBUS PA and DP Application The flowmeters are suitable for measuring the flow of almost all electrically conductive liquids, pastes and slurries. The main applications can be found in: Water and waste water Chemical and pharmaceutical industries Food and beverage industries Power generation and utility Function The MAG 5000/6000 are transmitters with a build-in alphanumeric display in several languages. The transmitters evaluate the signals from the associated electromagnetic sensors and also fulfil the task of a power supply unit which provides the magnet coils with a constant current. Further information on connection, mode of operation and installation can be found in the data sheets for the sensors. Displays and controls Operation of the transmitter can be carried out using: Control and display unit HART communicator PC/laptop and SIMATIC PDM software via HART communication PC/laptop and SIMATIC PDM software using PROFIBUS or Modbus communication Transmitter Operating and display panel HART communication Master HART- Communicator PROFIBUS DP PROFIBUS PA communication + Coupler with power supply Coupling module PC/ laptop RS 232 PROFIBUS PA... min. 230 Ω Transmitter with PROFIBUS PA interface T Bus term. Siemens FI 01 2012 /33

Flow Measurement Siemens AG 2011 Technical specifications Mode of operation and design Measuring principle Empty pipe Excitation frequency Electrode input impedance Input Digital input Activation time Current Output Current output Signal range Load Time constant Digital output Frequency Electromagnetic with pulsed constant field Detection of empty pipe (special cable required in remote mounted installation) Depend on sensor size > 1 x 10 1 Ω 11... 30 V DC, R i =. KΩ 50 ms I 11 V DC = 2.5 ma, I 30 V DC = 7 ma 0... 20 ma or... 20 ma <800 Ω 0.1 30 s, adjustable 0... 10 khz, 50 % duty cycle (uni/bidirectional) Pulse (active) 2 V DC, 30 ma, 1KΩ R i 10 KΩ, short-circuitprotected (power supplied from flowmeter) Pulse (passive) 3 30 V DC, max. 110 ma, 200 Ω R i 10 KΩ (powered from connected equipment) Time constant Relay output Time constant Load Low flow cut off Galvanic isolation Max. measuring error (incl. sensor and zero point) MAG 5000 MAG 6000 Rated operation conditions Ambient temperature Operation Storage Mechanical load (vibration) Compact version 0.1 30 s, adjustable Changeover relay, same as current output 2 V AC/2 A, 2 V DC/1 A 0... 9.9 % of maximum flow All inputs and outputs are galvanically isolated. 0. % ±1 mm/s 0.2 % ±1 mm/s Display version: -20... +60 C (-... +10 F) 1) Blind version: -20... +60 C (-... +10 F) -0...+70 C(-0...+158 F) 18... 1000 Hz, 3.17 g rms, sinusoidal in all directions to IEC 68-2-36 19 insert 1... 800 Hz, 1 g, sinusoidal in all directions to IEC 68-2-36 Degree of protection Compact version IP67/NEMA X/6 to IEC 529 and DIN 0050 (1 mh 2 O 30 min.) 19 insert IP20/NEMA 1 to IEC 529 and DIN 0050 EMC performance IEC/EN 61326-1 (all environments) IEC/EN 61326-2-5 Display and keypad Totalizer Two eight-digit counters for forward, net or reverse flow Display Time constant Design Enclosure material Compact version 19" insert Back of panel Panel mounting Background illumination with alphanumeric text, 3 x 20 characters to indicate flow rate, totalized values, settings and faults; Reverse flow indicated by negative sign Time constant as current output time constant Fiber glass reinforced polyamide; stainless steel AISI 316/1.36 (IP65) Standard 19 insert of aluminum/steel (DIN 19), width: 21 TE, height: 3 HE IP20/NEMA 1; Aluminum IP20/NEMA 1 (prepared for IP65/NEMA 2 display side); ABS plastic IP66/NEMA X; ABS plastic Wall mounting Dimensions Compact version See dimensional drawings 19 insert See dimensional drawings Weight Compact version 0.75 kg (2 lb) 19 insert See dimensional drawings Power supply 115... 230 V AC +10 % -15 %, 50... 60 Hz 11... 30 V DC or 11... 2 V AC Power consumption 230 V AC: 17 VA 2 V AC : 9 VA, I N =380mA, I ST =8A (30ms) 12VDC: 11W, I N =920mA, I ST =A (250ms) I ST = A (250 ms): For solar panel please secure stable current supply Certificates and approvals CE, C-UL general purpose, C-tick; FM Class I, Div 2, CSA Class I, Div 2 Custody transfer approval (MAG 5000/6000 CT) Communication Standard MAG 5000 MAG 6000 Optional (MAG 6000 only) MAG 5000/6000 CT 1) Cold water: MI-001, PTB/OIML R 9 (pattern approval DE/DK) Hot water: PTB and DANAK OIML R 75 (pattern approval DE/DK) (MAG 6000 CT) Chilled water: PTB K 7.2 Other media than water (milk, beer etc.): PTB and DANAK OIML R 117 (pattern approval DE/DK) (MAG 6000 CT) Without serial communication or HART as option Prepared for client-mounted addon modules HART, Modbus RTU/RS85, FOUNDATION Fieldbus H1, DeviceNet, PROFIBUS PA, PROFIBUS DP as add-on modules No communication moduls approved MAG 5000 CT/MAG 6000 CT: -20... +50 C (-... +122 F). /3 Siemens FI 01 2012

Siemens AG 2011 Flow Measurement Safety barrier (e/ia) Application For use with MAG 5000/6000 19 and MAG 1100 Ex ATEX/MAG 3100 Ex ATEX Ex approval MAG 1100 Ex [EEx e ia] IIB ATEX MAG 3100 Ex [EEx e ia] IIC ATEX Cable parameter Group Capacity in μf Inductance in mh Electrode IIC.1 80 IIB 5 87 IIA 5 87 Ambient temperature During operation -20... +50 C (-... +122 F) During storage -20... +70 C (-... +158 F) Enclosure Material Standard 19 insert in aluminum/steel (DIN 19) Width 21 TE (.75 ) Height 3 HE (5.25 ) Rating IP20 / NEMA 1 to EN 60529 Mechanical load 1 g, 1 800 Hz sinusoidal in all directions to EN 60068-2-36 Electrode cleaning unit for MAG 5000 or 6000 in 19" insert version The purpose of electrode cleaning is to remove unwanted deposits on the electrodes in water applications by applying either a DC or AC voltage to the electrodes. AC cleaning is used in waste water applications to remove fatty deposits on the electrodes by warming up the electrode. DC cleaning is used in district heating applications to eliminate electrically conductive deposits. Application for use with transmitters MAG 5000 and 6000 19 to clean the electrodes on sensors MAG 1100 or MAG 3100 Not to be used with intrinsically safe Ex sensors Not to be used with sensors with Hastelloy and Tantalum electrodes Available on request Siemens FI 01 2012 /35

Flow Measurement Siemens AG 2011 Selection and Ordering data Transmitter MAG 5000 Transmitter MAG 5000 Blind for compact and wall mounting; IP67/NEMA X/6, fibre glass reinforced polyamide 11...30VDC / 11... 2 V AC 7ME6910-1AA30-0AA0 115... 230 V AC, 50/60 Hz 7ME6910-1AA10-0AA0 Transmitter MAG 5000 Display for compact and wall mounting; IP67/NEMA X/6, fibre glass reinforced polyamide 11...30VDC / 11... 2 V AC 7ME6910-1AA30-1AA0 115... 230 V AC, 50/60 Hz 7ME6910-1AA10-1AA0 115... 230 V AC, 50/60 Hz, with HART Transmitter MAG 5000 CT for compact and wall mounting, approved for custody transfer; IP67/NEMA X/6, fibre glass reinforced polyamide 11...30VDC / 11... 2 V AC 7ME6910-1AA10-1BA0 7ME6910-1AA30-1AB0 115... 230 V AC, 50/60 Hz 7ME6910-1AA10-1AB0 Transmitter MAG 5000 for 19 rack and wall mounting 11...30VDC/ 11... 2 V AC 115... 230 V AC, 50/60 Hz Short lead time (details in PMD) 7ME6910-2CA30-1AA0 7ME6910-2CA10-1AA0 Accessories for MAG 5000 and MAG 6000 Wall mounting unit for IP67/ NEMA X/6 version, wall bracket, terminal box in polyamide x M20 cable glands FDK-085U1018 x ½ NPT cable glands FDK-085U1053 Sun lid for MAG 5000/6000 A5E0232885 transmitter (Frame and lid) Transmitter MAG 6000 Transmitter MAG 6000 Blind for compact and wall mounting; IP67/NEMA X/6, fibre glass reinforced polyamide 11... 30 V DC / 11... 2 V AC 1AA30-0AA0 115... 230 V AC, 50/60 Hz 1AA10-0AA0 Transmitter MAG 6000 for compact and wall mounting; IP67/NEMA X/6, fibre glass reinforced polyamide 11... 30 V DC / 11... 2 V AC 1AA30-1AA0 115... 230 V AC, 50/60 Hz 1AA10-1AA0 Transmitter MAG 6000 CT for compact and wall mounting, approved for custody transfer (no communication modules possible); IP67/NEMA X/6, fibre glass reinforced polyamide 11... 30 V DC / 11... 2 V AC Transmitter MAG 6000 for compact and wall mounting; IP65/NEMA, stainless steel AISI 316/1.36 (only for sensor with SS terminal box) (for remote installation order SS terminal box separately) 11... 30 V DC / 11... 2 V AC 115... 230 V AC, 50/60 Hz 1QA30-1AA0 1QA10-1AA0 1AA30-1AB0 115... 230 V AC, 50/60 Hz 1AA10-1AB0 Transmitter MAG 6000 SV for compact and wall mounting; special excitation Hz settings for Batch application DN 25/1 IP67/NEMA X/6, fibre glass reinforced polyamide 11... 30 V DC / 11... 2 V AC 115...230VAC, 50/60Hz Transmitter MAG 6000 for 19 rack and wall mounting 11... 30 V DC / 11... 2 V AC 1AB30-1AA0 1AB10-1AA0 2CA30-1AA0 115... 230 V AC, 50/60 Hz 2CA10-1AA0 Transmitter MAG 6000 SV for 19 rack and wall mounting; special excitation Hz settings for Batch application DN 25/1 11... 30 V DC / 11... 2 V AC 115... 230 V AC, 50/60 Hz 2CB30-1AA0 2CB10-1AA0 /36 Siemens FI 01 2012

Siemens AG 2011 Flow Measurement MAG 6000 with IP66/NEMA X enclosure; 115...230VAC, 50/60Hz; cable gland PG13.5 MAG 6000 with safety barrier for Ex-approved sensors, complete mounted with IP66/NEMA X wall mounting enclosure, ATEX, 115...230VAC, 50/60Hz; cable gland PG13.5 For ATEX 2G D sensors MAG 6000 SV, 19 insert, in IP66/NEMA X, ABS plastic enclosure, excitation frequency Hz for Batch application DN 25/1 ; cable gland PG13.5 11 30 V DC, 11 2VAC, 50/60Hz 115 230 V AC, 50/60 Hz Short lead time (details in PMD) 2EA10-1AA0 2MA11-1AA0 2EB30-1AA0 2EB10-1AA0 Operating instructions for MAG 5000/6000 For MAG 5000/6000 IP67 English German Spanish French For MAG 5000/6000 19" English A5E02338368 A5E029982 A5E029995 A5E029990 A5E02082880 This device is shipped with a Quick Start guide and a CD containing further SITRANS F literature. All literature is also available for free at: http://www.siemens.com/flowdocumentation Communication modules for MAG 6000 HART (not for MAG 6000 I) FDK-085U0226 Modbus RTU/RS85 FDK-085U023 PROFIBUS PA Profile 3 FDK-085U0236 PROFIBUS DP Profile 3 FDK-085U0237 DeviceNet FDK-085U0229 FOUNDATION Fieldbus H1 A5E0205250 Operating instructions for SITRANS F add-on modules HART English A5E03089708 PROFIBUS PA/DP English A5E00726137 German A5E0102629 MODBUS English A5E0075397 German A5E03089262 Spanish A5E03089278 French A5E03089265 FOUNDATION Fieldbus English A5E02318728 German A5E0288856 Spanish A5E02512177 French A5E02512169 DeviceNet English A5E03089720 This device is shipped with a Quick Start guide and a CD containing further SITRANS F literature. Accessories for MAG 5000 and MAG 6000 Cable for standard electrode or coil, 3 x 1.5 mm² / 18 gage with shield PVC 10m(33ft) FDK-083F0121 20m(65ft) FDK-083F0210 0 m (130 ft) FDK-083F0211 60 m (200 ft) FDK-083F0212 100 m (330 ft) FDK-083F0213 150 m (500 ft) FDK-083F3052 200 m (650 ft) FDK-083F3053 500 m (1650 ft) FDK-083F305 Electrode cable for empty pipe or low conductivity, double shielded, 3 x 0.25 mm² 10m(33ft) FDK-083F3020 20m(65ft) FDK-083F3095 0 m (130 ft) FDK-083F309 60 m (200 ft) FDK-083F3093 100 m (330 ft) FDK-083F3092 150 m (500 ft) FDK-083F3056 200 m (650 ft) FDK-083F3057 500 m (1650 ft) FDK-083F3058 Low-noise electrode coax cable for low conductivity and high vibration levels of cables, 3 x 0.13 mm 2 2 m (6.6 ft) A5E02272692 5m(16.5ft) A5E02272723 10m(33ft) A5E02272730 Siemens FI 01 2012 /37

Flow Measurement Siemens AG 2011 Cable kit with standard coil cable, 3 x 1.5 mm²/18 gage with shield PVC and electrode cable double shielded, 3 x 0.25 mm² 5m(16.5ft) A5E02296329 D) 10m(33ft) A5E0118167 15m(9ft) A5E022966 D) 20m(65ft) A5E01181656 25m(82ft) A5E0229690 D) 30m(98ft) A5E022969 D) 0 m (130 ft) A5E01181686 50 m (16 ft) A5E0229698 D) 60 m (200 ft) A5E01181689 100m(330ft) A5E01181691 150m(500ft) A5E01181699 200m(650ft) A5E01181703 500 m (1650 ft) A5E01181705 Potting kit for terminal box of FDK-085U0220 MAG sensors for IP68/NEMA 6P (not for Ex sensors) Sun shield for remote MAG 5000/6000 transmitters Sun Shield for compact MAG 5000/6000 transmitters on MAG 3100 (DN 15... 2000 (½"... 78") or MAG 5100 W (DN 150... 1200 (6"... 8") Short lead time (details in PMD) A5E0120996 D) A5E01209500 D) D) Subject to export regulations AL: N, ECCN: EAR99H 19 safety barrier (21 TE) [EEx e ia] IIC for MAG 1100 Ex sensors and MAG 3100 Ex sensors, incl. back plate FDK-083F503 Panel mounting enclosure for 19 insert (21 TE); IP65/NEMA 2 enclosure in ABS plastic for front panel mounting Panel mounting enclosure for 19 insert (2 TE); IP65/NEMA 2 enclosure in ABS plastic for front panel mounting Back of panel mounting enclosure for 19 insert (21 TE); IP20/NEMA 1 enclosure in aluminum FDK-083F5030 FDK-083F5031 FDK-083F5032 Back of panel mounting enclosure for 19 insert (2 TE); IP20/NEMA 1 enclosure in aluminum FDK-083F5033 IP66/NEMA X, wall mounting enclosure for 19 inserts (without back plates) 21TE FDK-083F5037 2TE FDK-083F5038 Front cover (7TE) FDK-083F525 /38 Siemens FI 01 2012

Siemens AG 2011 Flow Measurement Spare parts Connection board (for polyamide terminalbox) 12... 2V 115... 230 V Connection board (for stainless steel terminalbox) 12... 2V A5E02559817 A5E02559816 A5E0260280 Cable glands, for above cable, pcs. M20 ½ NPT Sealing screws for sensor/ transmitter, 2 pcs A5E0082290 A5E00822501 FDK-085U0221 ½ NPT M20 115... 230 V 19 enclosure, 12... 2 V, 115...230V Connection board for standard 19 transmitter A5E0260272 A5E02559809 Terminal box, in polyamide, inclusive lid M20 ½" NPT Terminal box lid, in polyamide FDK-085U1050 FDK-085U1052 FDK-085U1003 Connection board for transmitter ia and safety barrier Connection board for transmitter ia/ib and safety barrier (only for sensors produced before October 2007) Connection board for transmitter and cleaning unit SENSORPROM memory unit (Sensor code and serial numbers must be specified on order) 2kB (for MAG 5000/6000/ MAG 6000 I) 250 B (for MAG 2500/3000) Display unit for MAG 5000/6000 Black neutral front Siemens front A5E02559810 A5E02559811 FDK-083F123 FDK-085U1005 FDK-085U1008 FDK-085U1038 FDK-085U1039 Terminal box, in stainless steel, inclusive lid for MAG 6000 in stainless steel and for all Ex sensors M20 ½" NPT Terminal box (3A) for MAG 1100 F in polyamide, inclusive lid M20 ½" NPT Wall unit enclosure IP66, 12...2V, 115...230V PCB for standard transmitter A5E00836867 A5E00836868 A5E0082278 A5E0082279 A5E02559813 PCB for transmitter ia/e A5E0255981 and safety barrier PCB for transmitter ia/ib A5E02559812 and safety barrier (7ME6130, 7ME6150 and 7ME6330) PCB for transmitter and A5E02559815 cleaning unit Short lead time (details in PMD) Display unit for 19" versions FDK-085U339 Siemens FI 01 2012 /39

Flow Measurement Siemens AG 2011 Dimensional drawings Transmitter IP67/NEMA X/6 compact polyamide 155 (6.10) 6 (0.2) 131 (5.16) 71 (2.80) 178 (7.01) Ø8 (0.31) Ø12 (0.7) Ø8 (0.31) 13 (0.51) 13 (0.51) 71 (2.80) 20 (9.5) 155 (6.10) 102 (.02) 155 (6.10) 30 (1.18) 50 (2.0) 170 (6.70) Transmitter compact mounted, dimensions in mm (inch) Transmitter, 19 IP20/NEMA 1 standard unit Transmitter wall mounted, dimensions in mm (inch) 193 (7.60) 1 (0.6) 106 (.17) 91 (3.58) 175 (6.89) 168 (6.61) 128 (5.0) 122 (.80) 2.8 (0.11) Weight incl. back print 0.8 kg (1.8 lbs) /0 Siemens FI 01 2012

Siemens AG 2011 Flow Measurement Transmitter, wall mounting IP66/NEMA X, 21 TE 13 (0.51) 210 (8.27) 235 (9.25) 90 (3.5) 120 (.72) 16 (5.75) 90 (3.5) 13 (0.51) 22 (0.86) 192 (7.56) Weight excl. transmitter: 2.3 kg (5.0 lbs) Transmitter, wall mounting IP66/NEMA X, 2 TE 13 (0.51) 90 (3.5) 120 (.72) 13 (0.51) 357 (1.05) 339 (13.35) Weight excl. transmitter: 2.9 kg (7.0 lbs) 22 (0.87) 90 (3.5) 16 (5.75) 235 (9.25) Siemens FI 01 2012 /1

Flow Measurement Siemens AG 2011 Transmitter, panel front IP20/NEMA 1, 21 TE 197 (7.75) 18 (7.2) 198 (7.80) 50 (1.97) 10 (5.51) 1 (5.67) Weight excl. transmitter: 1.2 kg (2.7 lbs) 185 (7.28) Transmitter, panel front IP20/NEMA 1, 2 TE 29 (1.57) 1 (5.67) 281 (1.06) 198 (7.80) 50 (1.97) 10 (5.51) Weight excl. transmitter: 1.6 kg (3.5 lbs) 282 (1.10) /2 Siemens FI 01 2012

Siemens AG 2011 Flow Measurement Transmitter, back of panel IP20/NEMA 1, 21 TE 163 (6.2) 1 (5.67) 218 (8.58) 132 (5.20) 57 (2.2) Weight: 0.7 kg (1.6 lbs) Transmitter, back of panel IP20/NEMA 1, 2 TE 269 (10.59) 253 (9.96) 218 (8.58) 132 (5.20) 57 (2.2) Weight: 0.9 kg (2.0 lbs) Siemens FI 01 2012 /3

Flow Measurement Siemens AG 2011 Schematics Electrical connection Grounding PE must be connected due to safety class 1 power supply. Mechanical counters When mounting a mechanical counter to terminals 57 and 58 (active output), a 1000 µf capacitor must be connected to the terminals 56 and 58. Capacitor + is connected to terminal 56 and capacitor - to terminal 58. Output cables If the output cable length is long in noisy environment, we recommend to use shielded cable. Power supply Transmitter 115... 230 V AC L1 L N N Outputs 31 32 Passive output (External powered) 56 57 58 R Active output (Powered from transmitter) 56 57 58 Relay output NO 5 NC 6 Common Digital input 77 78 Current output (Powered from transmitter) R PE 11... 30 V DC 11... 30 V DC/11... 2 V AC 1 2 Digital output Vx 3... 30 V max. 110 ma PLC-Digital input 2 V max. 30 ma Counter or PLC-Digital input Input Relay 2 V DC/1A 2 V AC/2A PE 0/ - 20 ma Load 800 Ω Menu setup Negative: Positive: R = Pull up/down resistor 1... 10 KΩ may be required - depending on Cables/Input resistance 91 92 93 9 95 96 97 Reserved for communication modules Sensor Sensor connection 1) 81 82 0 83 8 Electrode cable 81 82 0 83 8 WA SA & NT (08) 912 6100 (08) 837 7800 hydramet@hydramet.com.au hydrasa@hydramet.com.au 85 86 Coil cable 85 86 VIC & TAS NSW & ACT (03) 9325 3900 (02) 350 8200 hydravic@hydramet.com.au hydransw@hydramet.com.au Shield 1) Note: Special cable with individual wire shields (shown as dotted lines) are only required when using empty pipe function or long cables. QLD (07) 3802 9500 hydraqld@hydramet.com.au www.hydramet.com.au / Siemens FI 01 2012