Report of Test LLIA000536-002 Integrating Sphere Report Performance Summary Voltage Current Power Frequency Power Factor Current THD Total Luminous Flux Efficacy Chromaticity (x,y) (u',v') Duv CCT CRI (Ra) R9 120.0 Vac 0.1289 A 14.52 W 59.97 Hz 0.939 4.5 % 563.7 lm 38.8 lm/w (0.4396, 0.4039) (0.2523, 0.5217) -0.0004 2954 K 97 88 Prepared For: Specialty Lighting Industries, Inc. Awi Salomon 1306 Doris Avenue Ocean, NJ 07712-4041, USA Test date: 01/27/2016 Signed: Page 1 of 6 Report date: 01/28/2016
0.9 0.8 CIE 1931 (x, y) Diagram 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 Page 2 of 6
0.5 CIE 1931 (x, y) Diagram 0.45 0.4 0.35 0.3 0.25 0.3 0.35 0.4 0.45 0.5 Page 3 of 6
Spectral Data Chromaticity CIE 1931 (x, y) (0.4396, 0.4039) Chromaticity CIE 1976 (u', v') (0.2523, 0.5217) Correlated Color Temperature (CCT) 2954 K Color Rendering Index (Ra) 97 R1 99 R2 97 R3 94 R4 96 R5 98 R6 97 R7 97 R8 95 R9 88 R10 93 R11 95 R12 89 R13 99 R14 95 Distance from Planckian Locus (Duv) -0.0004 Scotopic/Photopic Ratio * 1.409 Electrical Data Voltage Current Power Frequency Power Factor 120.0 Vac 0.1289 A 14.52 W 59.97 Hz 0.939 Current THD 4.5 % Page 4 of 6
0.014 380 385 390 395 400 405 410 415 420 425 430 435 440 445 450 455 460 465 470 475 Summary Spectral Power Distribution (wavelength - nm, spectral power - mw) 0.000051 0.000059 0.000079 0.000100 0.000132 0.000191 480 0.002316 0.002701 0.003215 0.003812 0.004377 0.004876 580 0.008335 0.008597 0.008948 0.009345 0.009824 0.010339 680 685 690 695 700 705 0.007428 0.006730 0.006061 0.005528 0.004836 0.004270 0.000291 0.005278 0.010834 710 0.003753 0.000439 0.005621 0.011273 715 0.003305 0.000664 0.005849 0.011635 720 0.002888 0.001028 0.006164 0.011899 725 0.002519 0.001569 0.006487 0.012050 730 0.002186 0.002283 0.006761 0.012063 735 0.001883 0.003121 0.007014 0.011890 740 0.001624 0.004281 0.007238 0.011608 745 0.001395 0.005143 0.007420 0.011203 750 0.001207 0.004409 0.003192 0.002618 0.002305 0.002133 485 490 495 500 505 510 515 520 525 530 535 540 545 550 555 560 565 570 575 0.007604 0.007726 0.007844 0.007967 0.008113 585 590 595 600 605 610 615 620 625 630 635 640 645 650 655 660 665 670 675 0.010704 0.010135 0.009503 0.008817 0.008126 755 760 765 770 775 780 0.001040 0.000897 0.000766 0.000656 0.000561 0.000480 Spectral Power (mw) 0.012 0.01 0.008 0.006 0.004 0.002 0 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660 680 700 720 740 760 780 Wavelength (nm) Page 5 of 6
Test Equipment Configuration: LightLab International Allentown 2m Integrating Sphere Measurements acquired using a Labsphere CDS 2600 spectroradiometer Testing was performed using 4π geometry Test Temperature: Test Procedure: Significance: Notes: 25.3 C Tested in accordance with the applicable sections of: LM-79-08, LM-78-07, LM-58-13, ANSI_ANSLG C78.377-2011, ANSI C82-77-2002 The laboratory has not participated in the selection of samples to be tested. All testing is performed on the understanding that the significance of the report is limited to the extent that the test sample is representative of production units. The measurements and other derived quantities contained in this report are based on the absolute data as measured. Prorating the performance of the sample for the use of other component combinations (such as lamp / LED / Ballast / driver), or for use in different environmental conditions than that tested, may produce erroneous results. This report is free of erasures and corrections This report may contain data that are not covered by the NVLAP accreditation. Quantities marked with * are not covered. Page 6 of 6 Sphere Report Template V2-5