TriLIN Program Utility TPU_ TriLIN Programming Manual

Similar documents
800 Displaying Series Flowmeter

Triple RTD. On-board Digital Signal Processor. Linearization RTDs 20 Hz averaged outputs 16-bit precision comparator function.

HD-CM HORIZON DIGITAL CABLE METER

REMOTE I/O R30 SERIES. PC CONFIGURATOR SOFTWARE Model: R30CFG Ver Users Manual

Operating instructions Electronic preset counter Type series 717

OPERATION MANUAL JENDPM008 THERMOCOUPLE PANEL THERMOMETER. TIP TEMPerature Products 340 W Broad Street Burlington, NJ 08016

MT03A Electronic converter for flow rate transmitters

FTS84/85 Feature Application Field. Temperature & Humidity / Flow Measuring Specialist. Monitor gas and flow for industrial process

Electronic converter for level transmitters MT03L Instructions manual

ST-4000 SIGNAL LEVEL METER

ST-4000D SIGNAL LEVEL METER

Operating Instructions BTX-1 Series Digital Band Tension Meter

Noise Detector ND-1 Operating Manual

User Manual. for. Panel mounted instrument

Installation and Setting up Instructions for the 990 Signal Conditioning Instrument

Revision 1.2d

TECHNICAL MANUAL DRAW-WIRE DISPLACEMENT TRANSDUCER TYPE DWT

Transmitter Interface Program

Lt DELTA USA, Inc

COMPANY. MX 9000 Process Monitor. Installation, Operating & Maintenance Manual AW-Lake Company. All rights reserved. Doc ID:MXMAN082416

DICKSON ES120/ES120A DICKSON. Electronic Signal Data Logger. Specifications. Applications, Features, & Getting Started. Instructions / Operating

VXF17 QUICKSTART GUIDE. 17" Full HD 3G-SDI/HDMI Rackmount Studio and Field Monitor

Model IQ4-PC User Manual Revision Date:

Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display.

AP-40. AP-40 Series Features Industry s smallest-sensor head Ultra lightweight High-speed response Two-color LED digital pressure display

Aurora Grid-Tie Installation Instructions (Model Number: PVI-3.0-OUTD-US-W) Revision 4.1

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Manual. Analog (U/I) Sendix M3661 / M3681. Sendix M3661R. Sendix M5861. Absolute multiturn encoder. Order code: 8.M36X1.XXXX.XX12

FTS84/85 Feature Application Field. Temperature & Humidity / Flow Measuring Specialist. Monitor gas and flow for industrial process

2 WIRE - LOOP POWERED TRANSMITTER FOR PT100 AND NI100 PROBES

Troubleshooting. 1. Symptom: Status indicator (Red LED) on SSR is constant on. 2. Symptom: Output indicator (Yellow LED) on SSR is flashing.

Electronic converter for flowmeters MT03F Instructions manual

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

Set Point Controller OC557. Operator s Manual

CARLO GAVAZZI Automation Components

ARC HEIGHT CONTROL (ARC VOLTAGE CONTROL) MODEL : HAC-01-A-1M

Auxiliary states devices

Mortara X-Scribe Tango+ Interface Notes

medlab One Channel ECG OEM Module EG 01000

PicoScope 4000 Automotive PC Oscilloscopes

MICROPROCESSOR CALCULATION MODULE - S2000

4.9 BEAM BLANKING AND PULSING OPTIONS

How to Order. Flow rate (l/min) Applicable models PFA710, PFA750. Nil

Ambiflex MF620 USER GUIDE

PulseFlow FP100 Pulse to 4 20mA Flow Converter (Flow Rate Transmitter / Totalizer / Indicator)

ZIRCONIA OXYGEN ANALYZER CONVERTER

Ocean Sensor Systems, Inc. Wave Staff, OSSI F, Water Level Sensor With 0-5V, RS232 & Alarm Outputs, 1 to 20 Meter Staff

Contents: 1 LANsmart Pro Main Unit 4 Remote Unit: ID1, ID2, ID3, ID4

Instruction Manual MODEL RSP SANITARY ELECTRONIC PRESSURE TRANSMITTER

SCALE & WEIGHT DISPLAYS

Table of contents 1. INTRODUCTION GENERAL CONTENTS OF PACKAGE GENERAL USE OF THE EQUFLOW 6100 D/A CONVERTER STORAGE..

Dynatel Advanced Cable Locator 2250

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

L, LTC, LTM, LT are registered trademarks of Linear Technology Corporation. Other product

COLOR TFT LCD MONITOR. Manual

IMPORTANT SAFETY INSTRUCTIONS

DMP 331i DMP 333i. Precision Pressure Transmitter. Stainless Steel Sensor. accuracy according to IEC 60770: 0.1 % FSO

AZ DISPLAYS, INC. COMPLETE LCD SOLUTIONS SPECIFICATIONS FOR 15.0 OPEN FRAME MONITOR

Wall-/Pipe-Mount Level Indicator

PicoScope 2000 Series PC Oscilloscopes

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Operating Instructions ACX-Series Digital Aircraft Cable Tension Meter

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630)

Operating Instructions

FOTS100 User Manual. BIOPAC Systems, Inc. Opsens Inc. 42 Aero Camino, Goleta, CA Tel (805) , Fax (805)

Micro-DCI 53ML5100 Manual Loader

7 3G -SDI Monitor w/ Signal Conversion QUICKSTART GUIDE

Module 28 Scan Tool Advanced

C-net WIND. User s Guide

NS-3 RF Noise Source Operation Manual

Operating Instructions

DH7-DK QUICKSTART GUIDE. DH7 4K Support HDMI On-Camera Field Monitor Deluxe Kit

Dynamic Cable Rating Model Guidelines

BE3-GPR GENERATOR PROTECTIVE RELAY

Panel cutout required: 1.772" x 3.622" (45mm x 92mm) 1.76" (45mm) 2.45" (62mm) 3.20" (81mm) 3.60" (91mm) 0.59" (15mm) Special Features

PicoScope 3000 Series Automotive User guide

FRQM-2 Frequency Counter & RF Multimeter

HDMI 1X4 Distribution Amplifier (Splitter)

Broadcast A/V Division M-LYNX-702 V.3. Dual 7 LCD Display. User Manual

ISOMAG THE FRIENDLY MAG METER. Gas, mists or vapours Zone 0 1G Gas, mists or vapours Zone 1 2G. Gas, mists or vapours Zone 2 3G

Broadcast A / V Division M-LYNX-702 V.3. Dual 7 LCD Display. User Manual

On-screen display signal strength meter Version 1.02

Product Manual MNX10015 / REV C MODEL SB142, SB242. Dual Output Series Switch Boxes

Fits the Leopard Family FREQ./RPM INPUT MODULE WITH 24 V EXC.

TRANSCENSION 6-CHANNEL DMX DIMMER PACK (order code: BOTE40) USER MANUAL

APPLICATIONS. Automatic warehouse. Trasportation lines

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

NI-DAQmx Device Considerations

DS 400 P. Intelligent Electronic Pressure Switch in Hygienic Stainless Steel Ball Housing. on hygienic process connections

Electronic Pulse Counter Combination device

1 Output 1 operation. 3 Pressure unit display. 4 Main display Large 4-character LCD display. 5 Sub-display Small 4-character LCD display.

IRIG-B PTP Clock Converter Output Module Hardware Installation Manual

CURRENT INJECTION RELAY TESTER CR-100. ( Code P60212 (770081) ) INSTRUCTIONS MANUAL ( M / 05A ) (c) CIRCUTOR S.A.

Programming instructions Software Version 5

Displays Open Frame Monitor Model Number: AND-TFT-150Bxx

FOUR CHANNEL USB RECORDER PCRU01. User manual. Table of Contents

LCD Thermometer / Clock S No. 1253

Industriefunkuhren. Technical Manual. IRIG-B Generator-Module for analogue / digital Signals of Type: IRIG-B / IEEE C / AFNOR NF S87-500

Guide for installers. METTLER TOLEDO MultiRange System solution analog Ex1. Hazardous area. Safe area

DS 200 P DS 200 P. Electronic Pressure Switch with Flush Process Connection

Transcription:

Program Utility TPU_2.8.2 TriLIN Programming Manual

1. Programming the electronics a. Tool necessary Power supply 12-32 V DC Computer with appropriate USB port and a Windows Operating System (We recommend Windows 7 or higher) LPU software Programming cable Possibly a voltmeter (Amperemeter) for the new comparison of the analogue outputs and functional test b. Read-out from the electronics Supplying the electronics with voltage, connecting the programming cable and opening the software. "COM Port" tab and then choose an appropriate port from the list Reading out from the electronics with the help of the "Read EE" button. If the data is not read out correctly, the operation must be repeated. TrigasDM GmbH September 2016 2

c. Importing the calibration curve A Trigas calibration file "*.sav" is necessary for the programming Choose the option "Load F/V kcurve" in the "File" menu for temperaturecompensated programming (Frequency/Viscosity vs. K-factor) / Refer to Section 4.3. Choose optional "Load F kcurve" for non-temperature compensated programming (Frequency vs. K-factor). After the data has been reloaded, the serial number must be compared with that of the calibrated flow meter. Optional: Manual entry of the calibration curve is necessary only if the data is not available to you. NOTE The values must be entered with a dot as the decimal point separator. e.g. 15.45 instead of 15,46 d. Adjusting the electronics Cut-off (ms) Maximum waiting time for completing a signal period. If this time gets exceeded, no input signal gets detected and all outputs are set to zero. Update rate (ms) Measurement period before a measurement result is obtained and calculated. Special case: "0" means that the calculation is already taking place after one pulse. Recommendation: Cut Off = 2 Update Rate TrigasDM GmbH September 2016 3

Timebase (sec.) Time constant for conversion from the K-factor and flow units. o 1 = xxx/sec. o 60 = xxx/min. o 3600 = xxx/hour Minimum frequency (Hz) Lowest frequency (Leakage flow rate). Below the frequency, the TriLIN displays zero at all outputs. Crystal factor Correction factor for internal time base. Use Volts Input 0 PT100 Input 1 Analog Input (analog Temp.) TriLIN can use as temperature input PT100, or analog temperature signal 0-10V e.g. 0 10V = -20 120 C Attention! To use the input, please contact the manufacturer. Mode Operating mode of the electronics. Mode Compensation Analogue output Frequency output 0 UVC Volumetric Volumetric 1 None Volumetric Volumetric 2 UVC Mass Volumetric 3 None Mass Volumetric 4 UVC Volumetric Mass 5 None Volumetric Mass 6 UVC Mass Mass 7 None Mass Mass Version Should not be changed. Pt100 K-factor (bits to Ohm) K-factor of the PT100 input. Should be corrected only when determining deviations of temperature input. TrigasDM GmbH September 2016 4

e. Liquid data changes The data pertaining to the media can be corrected whenever desired. To do this, choose the option "Load Fluid" in the "File" menu in order to load the new data. The table can also be edited manually as an option. NOTE The values must be entered with a dot as the decimal point separator. e.g. 15.45 instead of 15,46 f. Scaling the outputs Frequency output You only assign the desired scaling in the fields. You can set the units with "Mode" and "Time Base". e.g.: Mode 0: Frequency output, volumetric Time Base 60 per minute Flow / Frequency 0-100 l/min. = 0-1000 Hz TrigasDM GmbH September 2016 5

Flow, analogue output Mode 0: Analogue output, volumetric Time Base 60 per minute 0-100 l/min. = 0-10 V (possibly 4-20 ma) Volts (bits) A/D value is calculated automatically Temperature, analogue output -50 150 C = 0-10 V (possibly 4-20 ma) g. Programming Press the "Write EE" button to program the electronics. After successful programming, the message appears: Attention! The period of time after pressing the "Write..." and the "Succes" message should be 3sec. If not, the process should be performed again. TrigasDM GmbH September 2016 6

Run Live After programming, the program enables the data to be checked. The Live mode is started by pressing the "Run Live" button. h. Saving data Save the settings in the "File" menu option with "Save All" (*.lin) If you need further information, please contact us. TrigasDM GmbH September 2016 7

i. Printing the data You need Excel 2010 or later to print the data. Open the Excel table enclosed. Confirm the security warning with "Activate contents" Load the saved data with the "Load LIN" button. Enter the serial number of the electronics. Print out the report. TrigasDM GmbH September 2016 8

2 Adjusting the analogue outputs NOTE The adjustment should be carried out only with a calibrated voltmeter or ammeter. Connect the electronics as you do for programming with the data cable and power supply. Connect the voltmeter (or ammeter) to the analogue output to be calibrated. Read out from the electronics with "Read EE". There are arrow keys and scroll bars located to the right side of the scaling fields with the help of which the bit values of the respective analogue outputs can be changed. With the upper scroll bar, you can set the minimum value on the multimeter (e.g. 0 V for 0-10 V), and then with the lower scroll bar you can set the maximum value (e.g. 10 V for 0-10 V). Press STOP at the end -> repeat this for the second output. Press the "Write EE" button to program the electronics. TrigasDM GmbH September 2016 9

3 Adjusting the PT100 input NOTE You should carry out this adjustment only with a calibrated resistance decade Connect the electronics as you do for programming with the data cable and power supply. Connect the decade at the temperature input and set 100 Ω. Read out from the electronics with "Read EE" Enter 100 in the calibration field and press the "Calculate" button The PT100 K- factor is recalculated. Press the "Write EE" button to program the electronics. TrigasDM GmbH September 2016 10

4 TriLIN pin assignments Terminal block assignment: TERMINAL DESCRIPTION TRILIN TRILIN +TEMP. 1 +9-32 V DC supply input 2 0 V 3 Sensor input + 4 Sensor input - 5 Analogue flow output 6 GND 7 Analogue temperature output 8 RTD PT100 supply + 9 RTD PT100 sensor + 10 RTD PT100 sensor - 11 RTD PT100 sensor - 12 0-10 V DC temperature input 13 5 V (TTL pulse) frequency output 14 GND 15 NC 16 NC TrigasDM GmbH September 2016 11

5 Trouble shooting 4.8.1 Electronics has no output signal a. Check supply voltage (9-32V DC) i. Standard current consumption is (depending on the configuration) up to 40mA. - Current > 100mA - board defective. - Current = 0 ma - electronics connected incorrectly. b. Flowmeter functionality check i. Attention! The turbine should not be tested with compressed air - turbine may be damaged. ii. Flow simulation: turbines pickoff evaluation Remove pickoff from the turbine and move an iron element (not stainless steel) under the pickoff back and forth. iii. Pickup Resistance Measuring - between pin 1 and 8 RF Sensor: 10Ω +/- 15% c. Check the condition of the electronics (if available) connect electronics with PC and RUN LIVE Switch Mode Programming - Point 4.4 4.8.2 Baseline value appears to be wrong a. If a problem is noted - RUN LIVE check mode. Compare Analog temperature output versus medium (liquid, gas) temperature. TrigasDM GmbH September 2016 12

Customer service If you should require customer service for your TrigasDM products, please contact our customer service department. All requests for information concerning a specific measuring instrument must include the type and serial number of the measuring instrument. We will provide you with all possible assistance over the phone. If your equipment has to be examined or repaired at our plant, whether within the warranty period or after its expiration, our customer service department will issue you with an authorisation number, which is used to initiate our quick and efficient customer service processing. On receipt at our plant, your equipment will be repaired or replaced without delay and returned to you within the shortest period possible. Please do not return any products without an authorisation number. TrigasDM GmbH Tel.: +49 8165 9999 300 Erdinger Str. 2b Fax: +49 8165 9999 369 Email: info@trigasdm.com 85375 Neufahrn, Germany www.trigasdm.com TrigasDM GmbH September 2016 13