Physics Requirements for the CXI Ion Time-of-Flight

Similar documents
Engineering Specifications for the CXI Ion Time-of-Flight

TWO BUNCHES WITH NS-SEPARATION WITH LCLS*

vacuum analysis surface science plasma diagnostics gas analysis

Design, Fabrication and Testing of Gun-Collector Test Module for 6 MW Peak, 24 kw Average Power, S-Band Klystron

These are used for producing a narrow and sharply focus beam of electrons.

arxiv:hep-ex/ v1 27 Nov 2003

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

The Construction Status of CSNS Linac

Quantar Technology 3300/2400 Series Imaging Detector Systems

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Klystron Lifetime Management System

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

A dedicated data acquisition system for ion velocity measurements of laser produced plasmas

4.9 BEAM BLANKING AND PULSING OPTIONS

Jefferson Lab Experience with Beam Halo, Beam Loss, etc.

Mechanical Design of Recirculating Accelerator Experiments for Heavy-Ion Fusion

EPJ Web of Conferences 95,

TECHNICAL SPECIFICATION Multi-beam S-band Klystron type BT267

4.4 Injector Linear Accelerator

Progress Update FDC Prototype Test Stand Development Upcoming Work

Linac 4 Instrumentation K.Hanke CERN

CATHODE-RAY OSCILLOSCOPE (CRO)

North Damping Ring RF

-Technical Specifications-

IOT OPERATIONAL EXPERIENCE ON ALICE AND EMMA AT DARESBURY LABORATORY

... A COMPUTER SYSTEM FOR MULTIPARAMETER PULSE HEIGHT ANALYSIS AND CONTROL*

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator.

Commissioning and Performance of the ATLAS Transition Radiation Tracker with High Energy Collisions at LHC

Characterizing the Electro-Optic Properties of a Microfabricated Mass Spectrometer

PICOSECOND TIMING USING FAST ANALOG SAMPLING

Display Systems. Viewing Images Rochester Institute of Technology

Neutron Spectrometer Operation Manual

MSX600 Table Top Type X Ray Inspection System

The PEFP 20-MeV Proton Linear Accelerator

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

Detailed Design Report

The TORCH PMT: A close packing, multi-anode, long life MCP-PMT for Cherenkov applications

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

New gas detectors for the PRISMA spectrometer focal plane

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Development of Multiple Beam Guns for High Power RF Sources for Accelerators and Colliders

The Cornell/Purdue TPC

Particle-in-cell simulation study of PCE-gun for different hollow cathode aperture sizes

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Instruments for Surface Science

USA Price List. January or 8-channel LIBSCAN system (100 mj, 1064 nm laser) Alternative LIBSCAN configurations

Lecture 14: Computer Peripherals

In-process inspection: Inspector technology and concept

Review of the CMS muon detector system

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16]

R&D on high performance RPC for the ATLAS Phase-II upgrade

The Use of an Electron Microchannel as a Self-Extracting and Focusing Plasma Cathode Electron Gun

Single-sided CZT strip detectors

Sep 09, APPLICATION NOTE 1193 Electronic Displays Comparison

Photo cathode RF gun -

Pseudospark-sourced Micro-sized Electron Beams for High Frequency klystron Applications

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

APPARATUS FOR GENERATING FUSION REACTIONS

The basic parameters of the pre-injector are listed in the Table below. 100 MeV

GFT channel Time Interval Meter

Nova NanoSEM Superior Imaging and Analytical Performance

RF Power Generation II

CX1725W Liquid Cooled, Hollow Anode Two-Gap Metal/Ceramic Thyratron

A pixel chip for tracking in ALICE and particle identification in LHCb

Reading a GEM with a VLSI pixel ASIC used as a direct charge collecting anode. R.Bellazzini - INFN Pisa. Vienna February

This work was supported by FINEP (Research and Projects Financing) under contract

CAEN Tools for Discovery

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

BEAM DIAGNOSTICS IN THE CNAO INJECTION LINES COMMISSIONING

Universal High Current Implanter for Surface Modifications with ion beams Extensive range of ion species, including refractory metals Magnetic mass

Preparing for remote data collection at NE-CAT

Drift Tubes as Muon Detectors for ILC

Transmissive XBPM developments at PSF/BESSY. Martin R. Fuchs

Spatial Light Modulators XY Series

Updates on the Central TOF System for the CLAS12 detector

A New 4MW LHCD System for EAST

14 GHz, 2.2 kw KLYSTRON GENERATOR GKP 22KP 14GHz WR62 3x400V

Teaching Plasma Nanotechnologies Based on Remote Access

MTL Software. Overview

HV/PHA Adjustment (PB) Part

LCLS RF Reference and Control R. Akre Last Update Sector 0 RF and Timing Systems

GENCOA Key Company Facts. GENCOA is a private limited company (Ltd) Founded 1995 by Dr Dermot Monaghan. Located in Liverpool, UK

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

SuperFRS GEM-TPC Development Status Report

An Improvised Two-Dimensional Laser Surface Scanner for Diagnosis of Rf Thermionic Electron Gun Problems

The field cage for a large TPC prototype

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Diamond detectors in the CMS BCM1F

Introducing Purion H, a Scanned Spot Beam High Current Ion Implanter

Trigger synchronization and phase coherent in high speed multi-channels data acquisition system

Tracking Detector R&D at Cornell University and Purdue University

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

3 cerl. 3-1 cerl Overview. 3-2 High-brightness DC Photocathode Gun and Gun Test Beamline

TIME RESOLVED XAS DATA COLLECTION WITH AN XIA DXP-4T SPECTROMETER

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Scanning For Photonics Applications

Performance and aging of OPERA bakelite RPCs. A. Bertolin, R. Brugnera, F. Dal Corso, S. Dusini, A. Garfagnini, L. Stanco

AREAL- Phase 1. B. Grigoryan on behalf of AREAL team

Tests of Timing Properties of Silicon Photomultipliers

Transcription:

PHYSICS REQUIREMENT DOCUMENT (PRD) Doc. No. SP-391-000-30 R0 LUSI SUB-SYSTEM CXI Physics Requirements for the CXI Ion Time-of-Flight Sébastien Boutet CXI Scientist, Author Paul Montanez CXI Lead Engineer Darren Marsh LCLS Quality Assurance Manager Tom Fornek LUSI System Manager Revision Description of Changes Approved R0 18DEC08 Initial release Verify that this is the latest revision. 1 of 5 Check for change orders or requests

Table of Contents Table of Contents...2 1. Overview...2 2. Performance Requirements...3 3. Size Requirement...3 4. Positioning Requirements...3 5. Vacuum Requirements...3 6. Electrical Requirements...3 7. Controls Requirements...4 8. Data Acquisition Requirements...4 9. Interface Requirements...4 10. Safety Requirements...5 1. Overview Samples illuminated by the LCLS beam in the CXI instrument will get highly ionized and undergo a Coulomb explosion. Detecting the products of this explosion is required in order to provide information about the chemical components of the sample and to provide a diagnostic on whether a particle was hit by the LCLS pulse. This document describes the requirements for an Ion Time-of- Flight (ITOF) mass spectrometer capable of performing these tasks. The coordinate system is defined in Mechanical Design Standards Supplement DS-391-000-36. 2 of 5

2. Performance Requirements 2.1. The ion TOF shall detect the arrival time of charged particles at the ITOF detector plane. 2.2. The ion TOF shall have a mass resolution of 1 atomic mass unit over a range up to 100 atomic mass units, assuming the initial kinetic energy of the created ions is 10 ev or less. 2.3. For more energetic ions, in the range of 1-10 kev that will be produced with the 0.1 micron KB system of the CXI instrument, the resolution of the ion TOF shall be 1 atomic mass unit up to at least a mass of 20 AMU. 2.4. The ion TOF shall be used to veto diffraction images with the ITOF signal used to determine if a particle was hit by a given LCLS pulse. 3. Size Requirement 3.1. The ion TOF mechanical components and electrical feedthroughs shall all be attached to a single flange no larger than 8 O.D. 3.2. The ion TOF shall consist of a drift tube of 40 ±5 mm inner diameter. 3.3. An electrode grid shall be positioned at the entrance of the drift tube. 3.4. The closest point of approach of the ion TOF drift tube from the interaction region shall be 25 ±2 mm. 3.5. The drift tube of the ion TOF shall be at least 200 ±25 mm long. Interface issues with the sample chamber may preclude the drift tube from being this long and 200 mm shall be the design goal. 3.6. A repeller electrode shall be located on the opposite side of the interaction region from the drift tube. This electrode shall be circular and of the same diameter as the drift tube. 3.7. The repeller electrode shall be located 25 ±2 mm away from the interaction region. 4. Positioning Requirements 4.1. The long axis of the ion TOF shall be pointed directly at the interaction region to within ±3 degrees. 4.2. The surface normal of the repeller electrode shall be parallel to the axis of the drift tube to within ±2 degrees. 5. Vacuum Requirements 5.1. The ion TOF shall operate with a 10-7 Torr pressure environment or better under all operating conditions and the appropriate vacuum practice for the design, manufacturing, and installation of the system components shall be implemented. 6. Electrical Requirements 6.1. It shall be possible to apply voltages up to 5 kv (with either polarity) electrode as well as on the drift tube and the entrance grid to the drift tube. on the repeller 3 of 5

6.2. The components inside the Sample Chambers shall be precluded under all circumstances to come into contact with the electrodes of the ion TOF in order to prevent electrical discharges. 6.3. The electric field at the interaction region shall have lines parallel to the axis of the drift tube to within ±2 degrees. 7. Controls Requirements 7.1. The voltage on up to 8 channels shall be controlled remotely via the instrument control system. 7.2. The voltages on up to two of the previously mentioned eight channels shall be pulsed at 120 Hz with synchronization to the LCLS pulse to within 100 picoseconds. 7.3. It shall be possible to delay the voltage pulse with respect to the LCLS by up to 1 µsec with 0.1 nsec resolution. 7.4. The instrument control system shall display all controls and status of the high voltage supplies. 7.5. It shall be possible to lock the settings of the high voltage supplies with password protection. 8. Data Acquisition Requirements 8.1. The analog signal produced by the ions impinging on the TOF detector shall be digitized at a rate of 1 GHz or faster. 8.2. It shall be possible to digitize and record up to 100,000 samples for every LCLS pulse, at 120 Hz, at the digitization rate of Requirement 8.1. 8.3. It shall be possible for the user to set a trigger level on the digitized signal. This trigger shall be used to select which diffraction images will be saved by the system. 8.4. It shall be possible to set the trigger level on a limited window of the digitized waveform as well as on multiple independent windows. 8.5. The trigger shall be used as a particle hit counter by counting the number of trigger events per user defined time window. 8.6. The particle hit counter shall be used as a detector when aligning the particle injector. 8.7. The hit rate averaged over a desired time window shall be displayed at the control console at a refresh rate of 1 second. 9. Interface Requirements 9.1. The ion TOF shall be mounted on either of the CXI Sample Chambers (1 micron Sample Chamber PRD# SP-391-000-20 and 0.1 micron Sample Chamber PRD# SP-391-001-41). 9.2. The drift tube and the repeller electrode, along with their respective supports shall not encroach under any circumstance on the ±45 degree angle from the interaction region to the 2D X-ray Detector (LCLS PRD # 1.6-002) which is mounted inside the CXI Detector Stage (PRD # SP-391-000-28). 4 of 5

9.3. In the case where functionality of the ion TOF is in conflict with the functionality of the imaging system in the CXI Sample Chamber, the functionality of the Sample Chamber shall have precedence and lead to reduced functionality of the Ion TOF. 9.4. The ion TOF shall be designed so it cannot interfere or collide with any of the components inside the Sample Chambers (PRD # SP-391-000-20 and PRD# SP-391-001-41). 10. Safety Requirements 10.1. The vacuum system of the sample chamber on which the ion TOF is attached shall be interlocked with the high voltage power supplies of the ion TOF so that the system is not energized when the pressure is above 10-4 Torr. 10.2. The pressure level at which the interlock is triggered shall be user-modifiable with password protection. 10.3. All SLAC electrical safety requirements associated with the high voltages required for the Ion TOF shall be observed in the design, fabrication and operation of this device. 5 of 5