Screen investigations for low energetic electron beams at PITZ

Similar documents
Photoinjector Laser Operation and Cathode Performance

Design Studies For The LCLS 120 Hz RF Gun Injector

Summary of recent photocathode studies

An Overview of Beam Diagnostic and Control Systems for AREAL Linac

The Elettra Storage Ring and Top-Up Operation

TESLA FEL-Report

Beam Instrumentation for CTF3 and CLIC

Activities on FEL Development and Application at Kyoto University

Transmissive XBPM developments at PSF/BESSY. Martin R. Fuchs

EUROFEL-Report-2007-DS EUROPEAN FEL Design Study

Non-Invasive Energy Spread Monitoring for the JLAB Experimental Program via Synchrotron Light Interferometers

Summary of the 1 st Beam Line Review Meeting Injector ( )

The basic parameters of the pre-injector are listed in the Table below. 100 MeV

Recent APS Storage Ring Instrumentation Developments. Glenn Decker Advanced Photon Source Beam Diagnostics March 1, 2010

P. Emma, et al. LCLS Operations Lectures

Present Status and Future Upgrade of KEKB Injector Linac

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Results of recent photocathode studies at FLASH. S. Lederer, S. Schreiber DESY. L. Monaco, D. Sertore, P. Michelato INFN Milano LASA

RUNNING EXPERIENCE OF FZD SRF PHOTOINJECTOR

Mechanical aspects, FEA validation and geometry optimization

First Results and Future of the Photo Injector Test Facility at DESY Zeuthen PITZ. introduction first measurements future schedule

Dark current and multipacting trajectories simulations for the RF Photo Gun at PITZ

Photo cathode RF gun -

Jefferson Lab Experience with Beam Halo, Beam Loss, etc.

Experience with the Cornell ERL Injector SRF Cryomodule during High Beam Current Operation

High Rep Rate Guns: FZD Superconducting RF Photogun

Beam Losses During LCLS Injector Phase-1 1 Operation

Diamond RF Status (RF Activities at Daresbury) Mike Dykes

New Filling Pattern for SLS-FEMTO

IOT OPERATIONAL EXPERIENCE ON ALICE AND EMMA AT DARESBURY LABORATORY

SRF-gun Development Overview. J. Sekutowicz 17 th September, 2015 SRF15, Whistler, Canada

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

LCLS Injector Technical Review

4.4 Injector Linear Accelerator

COMMISSIONING AND FIRST RESULTS OF THE ELECTRON BEAM PROFILER IN THE MAIN INJECTOR AT FERMILAB*

High Brightness Injector Development and ERL Planning at Cornell. Charlie Sinclair Cornell University Laboratory for Elementary-Particle Physics

Technology Challenges for SRF Guns as ERL Source in View of BNL Work

INSTALLATION STATUS OF THE ELECTRON BEAM PROFILER FOR THE FERMILAB MAIN INJECTOR*

PEP II Design Outline

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

Performance of a DC GaAs photocathode gun for the Jefferson lab FEL

High QE Photocathodes lifetime and dark current investigation

Operation of CEBAF photoguns at average beam current > 1 ma

Future Performance of the LCLS

Cathode Studies at FLASH: CW and Pulsed QE measurements

LHC Beam Instrumentation Further Discussion

POLARIZED LIGHT SOURCES FOR PHOTOCATHODE ELECTRON GUNS AT SLAC?

Linac-Beam Characterizations at 600 MeV Using Optical Transition Radiation Diagnostics *

Detailed Design Report

Drift Tubes as Muon Detectors for ILC

DPD80 Visible Datasheet

Photocathodes FLASH: Quantum Efficiency (QE)

LCLS-II Injector Tuning Procedures. F. Zhou (SLAC) & F. Sannibale (LBNL) 2/6/2017

Linac 4 Instrumentation K.Hanke CERN

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

The FLASH objective: SASE between 60 and 13 nm

DPD80 Infrared Datasheet

Requirements for the Beam Abort Magnet and Dump

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

A HIGH POWER LONG PULSE HIGH EFFICIENCY MULTI BEAM KLYSTRON

Digital BPMs and Orbit Feedback Systems

Status of the Jefferson Lab Polarized Beam Physics Program and Preparations for Upcoming Parity Experiments

Current status of XFEL/SPring-8 project and SCSS test accelerator

CBF500 High resolution Streak camera

Beam Instrumentation for X-ray FELs

BUNCH-COMPRESSOR TRANSVERSE PROFILE MONITORS OF THE SwissFEL INJECTOR TEST FACILITY

RF considerations for SwissFEL

STATUS AND COMMISSIONING RESULTS OF THE R&D ERL AT BNL*

BEAM DYNAMICS AND EXPERIMENT OF CPHS LINAC *

STATUS OF THE SWISSFEL C-BAND LINEAR ACCELERATOR

COMMISSIONING RESULTS OF BEAM DIAGNOSTICS FOR THE PETRA III LIGHT SOURCE

G0 Laser Status Parity Controls Injector Diagnostics

Status of Elettra, top-up and other upgrades

Test of ScannerMAX Saturn 1 with 600Hz Sine-wave input, having an optical scan angle of 40 optical degrees peak to peak.

A New 4MW LHCD System for EAST

Digital SWIR Scanning Laser Doppler Vibrometer

Intensity based laser distance measurement system using 2D electromagnetic scanning micromirror

RF Design of the LCLS Gun C.Limborg, Z.Li, L.Xiao, J.F. Schmerge, D.Dowell, S.Gierman, E.Bong, S.Gilevich February 9, 2005

Production of quasi-monochromatic MeV photon in a synchrotron radiation facility

The hybrid photon detectors for the LHCb-RICH counters

KARA and FLUTE RF Overview/status

Pulsed Klystrons for Next Generation Neutron Sources Edward L. Eisen - CPI, Inc. Palo Alto, CA, USA

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

JLab 10kW FEL Driver Beam Diagnostics

Connection for filtered air

MEMS Mirror: A8L AU-TINY48.4

Applications Keypad Backlighting Symbol Backlighting Status Indication Front Panel Indicator

Lt DELTA USA, Inc

An Operational Diagnostic Complement for Positrons at CEBAF/JLab

ILC Damping Ring Lattice Status Report. Louis Emery and Aimin Xiao Argonne National Laboratory Presented at KEK workshop Dec 18th, 2007

Photocathode activity for. P. Michelato

Semiconductors Displays Semiconductor Manufacturing and Inspection Equipment Scientific Instruments

Design, Fabrication and Testing of Gun-Collector Test Module for 6 MW Peak, 24 kw Average Power, S-Band Klystron

DARK CURRENT IN SUPERCONDUCTING RF PHOTOINJECTORS MEASUREMENTS AND MITIGATION

Supplementary Figure 1. OLEDs/polymer thin film before and after peeled off from silicon substrate. (a) OLEDs/polymer film fabricated on the Si

Position Resolution of Optical Fibre-Based Beam Loss Monitors using long electron pulses

TWO BUNCHES WITH NS-SEPARATION WITH LCLS*

The HSMB-C110 is a right-angle package with the universally accepted dimensions of 3.2 x 1.0 x 1.5 mm. This part is ideal for

Status of RF Power and Acceleration of the MAX IV - LINAC

SURFACE MOUNT LED LAMP STANDARD BRIGHT 1206 (Reverse Mount)

Transcription:

1 Screen investigations for low energetic electron beams at PITZ S. Rimjaem, J. Bähr, H.J. Grabosch, M. Groß Contents Review of PITZ setup Screens and beam profile monitors at PITZ Test results Summary

Optimization of L- band Photo Cathode RF gun @ PITZ 2 Photo cathode (Cs 2 Te) QE~0.5-5% 1.6 cell RF un NC (copper) Coaxial RF coupler Cathode laser Bucking solenoid Main solenoid, Bz_peak~0.2T Electron bunch Mirror in vacuum Parameter Value Max. RF repetition rate 10 Hz Max. RF power 6 MW peak power Max. RF pulse length 800 s No. of pulses / train 1 800 Bunch spacing 0.2 1 s Max. bunch charge a few nc

Current PITZ Setup (PITZ1.8) high energy section (p z ~24.8 MeV/c) low energy section (p z ~6.7 MeV/c) 3 booster RF gun Trans. projected emittance & phase space @ Emittance Measurement SYstem (EMSY) Trans. phase space reconstruction @ Tomography module

Measurements of Transverse Projected Emittance & Phase Space 4 Single slit scan technique EMSY stations consist of horizontal / vertical actuators with / OTR screens 10 / 50 m slits Beam size is measured @ EMSY position Beam RMS sizes: 0.2 2.5 mm Beam divergence is estimated from beamlet sizes @ observation screen Minimum beamlet RMS sizes: ~50 m Phase space reconstruction tomography module consists of 3 FODO cells 4 screen station Phase advance of 45 o Beam size is measured with a / OTR screen Minimum beam RMS sizes: 120 m Observation screen 2.64 m EMSY1 (z = 5.74 m)

Ce-doped Yttrium Aluminum Garnet () Powder Coating Screen 5 Chemical formula Y 3 Al 2,5 Ga 2,5 O 12 :Ce Thickness of powder layer 5-20 m Density of power 5.1 g/cm 3 Thickness of silicon substrate 100, 275, 380 m Density of silicon substrate 2.33 g/cm 3 Incident angle 90 o Resolution of beam image Beamlet image from 10 m slit (03.06.2009) Beam momentum ~14.7 MeV/c Measured @ 1.76 m from slit position Detail structure of image ~50 m Vertical RMS size <70 m Wavelength of peak emission 510 nm screen viewed by 12 bits camera ~ 60 m 3 mm 3 mm 14.6 mm 14.6 mm ( measured by microscope)

Optical Transition Radiation (OTR) Screen 6 OTR screen 45 o e-beam Mirror Lens CCD camera Material Thickness of silicon wafer Incident angle Consider wavelength range Si-wafer with Al coating 100 or 275 m 45 o 400 750 nm Beam energy increase from ~15 to ~25 MeV (energy) ~10 MeV (intensity) ~40% intensity (a.u.) 1 0.8 0.6 0.4 0.2 0 Optical aperture of 0.2 rad (~11.5 o ) 0 20 40 60 80 100 electron energy (MeV)

Wire Scanner (WS) 7 Wire material Wire size Average step size used in measurements tungsten 20 m 100 m Measured beamlet Y-profile from 50 m slit Measured with two straight wires (10 mm distance) 0.25 Voltage signal (a. u.) 0.2 0.15 0.1 0.05 0 0 5 10 Y-position (mm) Reference: Grabosch, FEL2007 beam x-movement y-movement straight line was used in measurements

Chemical Vapour Deposition (CVD) Diamond Screen 8 thickness diameter Incident angle High thermal conductivity Emission wavelengths 100 m 30 mm 45 o 5 times higher than Cu 415 478 nm Reference: M. Degenhard, CVD Diamond Screens for Beamline Diagnosis at PETRA III, not yet publish

Comparison of screen and wire scanner (WS) 9 WS (z = 9.47 m) H1.S4 (z = 8.34 m) H1.S5 (z = 8.92 m)

screen and wire scanner (WS): beam size & profile 10 rms beam size (mm) 1.2 0.9 0.6 0.3 0.0 H1.S4 H1.S5 WS 0 200 400 600 800 1000 rms beam size (mm) 1.4 1.1 0.8 0.5 0.2 H1.S4 H1.S5 WS 0.55 m 0.58 m 8 8.5 9 9.5 10 z-position (m) (pc) rms beam size: bunch charge (pc) xy x y X-profiles for Q~500 pc Fixed parameters: Momentum ~24.5 MeV/c 1 bunch per train Focusing (fixed main solenoid current) Camera gain: 2 db for H1.S4, 7 db for H1.S5 Varied parameters: Bunch charge ~ 0.1, 0.2, 0.4, 0.6, 0.8, 1 nc

Comparison of and OTR Screens 11 H1.S5 (z = 8.92 m) H1.S4 (z = 8.34 m) comparison of beam size between and OTR @ PST.Scr1-4 * use 12 bit camera and adjusted camera gain and no. of pulses to have intensity a bit below saturation

and OTR screens: comparison of RMS beam size 12 RMS beam size: xy x xy : geometrical mean RMS size x : horizontal RMS size y : vertical RMS size OTR (%) 100% OTR y Included correction of different optical path length for and OTR - Not yet corrected for depth-of-field from 45 o mounting configuration Fixed parameters - Momentum ~24.8 MeV/c - No. of bunch per train: 1 - Camera gain - Focusing (solenoid & quadrupole) < x > ~ 16.8%, < y > ~ 14.9%, < xy > ~ 15.8% Screen station screen OTR screen x (mm) y (mm) x (mm) y (mm) PST.Scr1 0.342 0.280 0.304 0.244 PST.Scr2 0.246 0.229 0.210 0.201 PST.Scr3 0.245 0.266 0.204 0.226 PST.Scr4 0.290 0.257 0.246 0.228

and OTR screens: comparison sensitivity 13 intensity per bunch (%) 1000 100 10 1 0.1 factor of ~25 OTR 0 300 600 900 1200 charge (pc) intensity per bunch (%) 1000 OTR 100 10 1 0.1 0 5 10 15 20 camera gain intensity per bunch = intensity per bunch per charge Intensity is linearly proportional to bunch charge Measured @ H1.S4 Fixed parameters: Momentum ~24.8 MeV/c Camera gain: 2 db Varied parameters: No. of bunches per train Focusing (adjusted solenoid current to have the same beam area) Intensity is linearly proportional to camera gain Measured @ H1.S5 Fixed parameters: Momentum ~24.8 MeV/c Bunch charge: (200 pc), OTR (1 nc) Varied parameters: Camera gain No. of bunches per train Focusing

and OTR screens: dependence on momentum 14 rms beam size (mm) 1.6 1.4 1.2 1 0.8 0.6 OTR intensity 1.E+05 1.E+04 1.E+03 1.E+02 1.E+01 1.E+00 OTR 14 16 18 20 22 24 26 mean momentum (MeV/c) 14 16 18 20 22 24 26 mean momentum (MeV/c) RMS beam size: xy x y Intensity = integrating intensity per bunch per beam area per charge RMS beam size and intensity vs. momentum @ H1.S5 Included correction of beam size due to different optical path length Varied parameters: No. of bunches per train Bunch charge per train: (200 pc), OTR (1 nc) Camera gain Focusing (solenoid current)

and OTR screens: intensity distribution & projection profiles 15 wire scanner profiles OTR Signal ADC (a.u.) 6000 4500 3000 1500 0-6 -4-2 0 2 4 6 x-position (mm) screen shows more detail structure of the beam image & profile (OTR: smoothing image and profiles) Fixed parameters: Momentum ~24.5 MeV/c Bunch charge: 1 nc Focusing Signal ADC (a.u.) 6000 4500 3000 1500 0 Camera gain: 1 db -6-4 -2 0 2 4 6 Varied parameters: y-position (mm) No. of bunches per train : 1 bunch OTR: 24 bunches

Comparison of and CVD diamond Screens 16 H2.H4 (z = 19.83 m) Optical system is not yet optimized: measured beam size value was not yet reliable

intensity per bunch (%) 1000 100 10 1 and CVD diamond screens: sensitivity 0 300 600 900 1200 charge (pc) Integrating intensity vs. bunch charge @ H2.S4 Fixed parameters: Momentum ~24.8 MeV/c Camera gain: 20 db Varied parameters: No. of bunches per train Focusing CVD diamond factor of ~5 CVDdiamond 36%? CVDdiamond intensity per pulse (%) rms beam size (mm) 1000 100 10 1 0.1 0.6 0.55 0.5 0.45 0.4 factor of ~5 CVD diamond 0 5 10 15 20 camera gain CVD diamond 0 5 10 15 20 camera gain 17

and CVD diamond screens: dependence on momentum 18 rms beam size (mm) 1.4 1.2 CVD diamond 1 0.8 0.6 0.4 14 16 18 20 22 24 26 mean momentum (MeV/c) RMS beam size vs. beam momentum @ H2.S4 Fixed parameters: Momentum: ~24.8 MeV/c Bunch charge: 1 nc Camera gain: 20 db Varied parameters: No. of bunches per train: (1 bunch), CVD diamond (5 bunches) Focusing (solenoid) 8.E+06 6.E+06 CVD diamond CVD diamond intensity 4.E+06 2.E+06 0.E+00 14 16 18 20 22 24 26 mean momentum (MeV/c) Intensity = integrating intensity per bunch per beam area CVD diamond screen has more smoothing image and profiles

Summary & Outlooks 19 Properties of powder coating, OTR and CVD screens @ PITZ were summarized Test results of different screen type were presented Sensitivity: > CVD diamond > OTR OTR provided smaller RMS beam size than Screen Sensitivity RMS beam size 100 % 100 % OTR 4 % 84 % CVD diamond 20 % to be investigated Experimental tests to compare RMS beam sizes and beam profiles from wire scanner and from screen at two locations were performed with different bunch charges Results showed good agreement for both beam size and projection profiles

Acknowledgements 20 G. Asova for data taken at the tomography module Y. Ivanisenko and G. Vaschenko for information and discussions about optical systems PITZ members