The SLAC Polarized Electron Source *

Similar documents
Performance of a DC GaAs photocathode gun for the Jefferson lab FEL

SLAC R&D Program for a Polarized RF Gun

POLARIZED LIGHT SOURCES FOR PHOTOCATHODE ELECTRON GUNS AT SLAC?

THE SLAC POLARIZED ELECTRON SOURCE * Abstract

Development of Multiple Beam Guns for High Power RF Sources for Accelerators and Colliders

FEL Gun Test Stand (GTS) from construction to beam operations

KLYSTRON GUN ARCING AND MODULATOR PROTECTION

Design Studies For The LCLS 120 Hz RF Gun Injector

High Brightness Injector Development and ERL Planning at Cornell. Charlie Sinclair Cornell University Laboratory for Elementary-Particle Physics

Cathode Studies at FLASH: CW and Pulsed QE measurements

Industrial Diode Laser (IDL) System IDL Series

Results of recent photocathode studies at FLASH. S. Lederer, S. Schreiber DESY. L. Monaco, D. Sertore, P. Michelato INFN Milano LASA

Next Linear Collider. The 8-Pack Project. 8-Pack Project. Four 50 MW XL4 X-band klystrons installed on the 8-Pack

DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC*

RUNNING EXPERIENCE OF FZD SRF PHOTOINJECTOR

High QE Photocathodes lifetime and dark current investigation

VERY HIGH VOLTAGE PHOTOEMISSION ELECTRON GUNS*

IOT OPERATIONAL EXPERIENCE ON ALICE AND EMMA AT DARESBURY LABORATORY

Detailed Design Report

Photoinjector Laser Operation and Cathode Performance

TWO BUNCHES WITH NS-SEPARATION WITH LCLS*

Design and Simulation of High Power RF Modulated Triode Electron Gun. A. Poursaleh

RECENT PROGRESS IN UPGRADE OF THE HIGH INTENSITY THzzz zz-fel AT OzSAKzA UNIVERSITYzzzz

LCLS Injector Technical Review

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Photocathodes FLASH: Quantum Efficiency (QE)

CHAPTER 9. Actives Devices: Diodes, Transistors,Tubes

THERMIONIC GUN CATHODE-GRID ASSEMBLY TEST PROCEDURE

Photo cathode RF gun -

AREAL- Phase 1. B. Grigoryan on behalf of AREAL team

Summary of recent photocathode studies

The temperature management of photo cathodes at MAMI and MESA

Solid State Modulators for X-Band Accelerators

Jefferson Lab Experience with Beam Halo, Beam Loss, etc.

Experience with the Cornell ERL Injector SRF Cryomodule during High Beam Current Operation

3 cerl. 3-1 cerl Overview. 3-2 High-brightness DC Photocathode Gun and Gun Test Beamline

RF Power Generation II

Improvements to Siemens Eclipse PET Cyclotron Penning Ion Source

Report on the LCLS Injector Technical Review

Diamond RF Status (RF Activities at Daresbury) Mike Dykes

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

Design, Fabrication and Testing of Gun-Collector Test Module for 6 MW Peak, 24 kw Average Power, S-Band Klystron

DARK CURRENT IN SUPERCONDUCTING RF PHOTOINJECTORS MEASUREMENTS AND MITIGATION

Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720

Lecture 17 Microwave Tubes: Part I

INFN School on Electron Accelerators. RF Power Sources and Distribution

Screen investigations for low energetic electron beams at PITZ

Towards an X-Band Power Source at CERN and a European Structure Test Facility

Status of the Jefferson Lab Polarized Beam Physics Program and Preparations for Upcoming Parity Experiments

Pulsed Klystrons for Next Generation Neutron Sources Edward L. Eisen - CPI, Inc. Palo Alto, CA, USA

Operation of CEBAF photoguns at average beam current > 1 ma

EPJ Web of Conferences 95,

This work was supported by FINEP (Research and Projects Financing) under contract

A HIGH POWER LONG PULSE HIGH EFFICIENCY MULTI BEAM KLYSTRON

RF considerations for SwissFEL

Program Risks Risk Analysis Fallback Plans for the. John T. Seeman DOE PEP-II Operations Review April 26, 2006

CX1725W Liquid Cooled, Hollow Anode Two-Gap Metal/Ceramic Thyratron

Linac 4 Instrumentation K.Hanke CERN

Operating Experience and Reliability Improvements on the 5 kw CW Klystron at Jefferson Lab

25W 9xxnm Uncooled Multimode Laser Diode Module

INFN Milano LASA News on Cathode Development

Closed Cycle Cryogenic Probe Station

Evaluation of Performance, Reliability, and Risk for High Peak Power RF Sources from S-band through X-band for Advanced Accelerator Applications

Fiber-Coupled Acoustic-Optics Modulator (AOM) Module

DEVELOPMENT OF X-BAND KLYSTRON TECHNOLOGY AT SLAC

4.9 BEAM BLANKING AND PULSING OPTIONS

RF Solutions for Science.

The FLASH objective: SASE between 60 and 13 nm

Drift Tubes as Muon Detectors for ILC

CLIC Feasibility Demonstration at CTF3

P. Emma, et al. LCLS Operations Lectures

E2V Technologies CX2668A, CX2668AX Air-Cooled, Hollow Anode, Two-Gap Metal/Ceramic Thyratrons

Critical Benefits of Cooled DFB Lasers for RF over Fiber Optics Transmission Provided by OPTICAL ZONU CORPORATION

RADIOGRAPHIC PERFORMANCE OF CYGNUS 1 AND THE FEBETRON 705

Features. Gain Variation Over Temperature db/ C

First operation of cesium telluride photocathodes in the TTF injector RF gun

ITS-I. Test station for evaluation of image quality of image intensifier tubes. Fig. 1. Photo of the ITS-I test station: a)photo, b)block diagram

High Rep Rate Guns: FZD Superconducting RF Photogun

A quick and dirty magnet design for the magnetized beam LDRD proposal Jay Benesch 10/16/2015

Technology Challenges for SRF Guns as ERL Sources in View of Rossendorf work

Activities on FEL Development and Application at Kyoto University

Tuesday, July 27, Testing and Characterization Plans

Agilent High Capacity RV Pumps & Roots Pumping Systems

High-Current Hollow Cathode Development *

X-Band Klystron Development at

2 Work Package and Work Unit descriptions. 2.8 WP8: RF Systems (R. Ruber, Uppsala)

INSTRUMENT CATHODE-RAY TUBE

Upgrading LHC Luminosity

Current status of XFEL/SPring-8 project and SCSS test accelerator

November 5,1999. The NLC Injector UCRL-JC

LED Driver Linear / area dimming. Driver LCAI 38/75W TW lp Tunable White

CATHODE RAY OSCILLOSCOPE (CRO)

THE NEXT LINEAR COLLIDER TEST ACCELERATOR: STATUS AND RESULTS * Abstract

The Use of an Electron Microchannel as a Self-Extracting and Focusing Plasma Cathode Electron Gun

Quality produced by means of electron beam

Ion Pumps. Technical Information. Specifications/Ordering Information. Introduction Ion Pump Guide

GENCOA Key Company Facts. GENCOA is a private limited company (Ltd) Founded 1995 by Dr Dermot Monaghan. Located in Liverpool, UK

Etching Part 2. Saroj Kumar Patra. TFE4180 Semiconductor Manufacturing Technology. Norwegian University of Science and Technology ( NTNU )

18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V

IOT RF Power Sources for Pulsed and CW Linacs

Transcription:

SLAC-PUB-9509 October 2002 The SLAC Polarized Electron Source * J. E. Clendenin, A. Brachmann, T. Galetto, D.-A. Luh, T. Maruyama, J. Sodja, and J. L. Turner Stanford Linear Accelerator Center, 2575 Sand Hill Rd., Menlo Park, CA 94025 Abstract The SLAC PES, developed in the early 1990s for the SLC, has been in continuous use since 1992, during which time it has undergone numerous upgrades. The upgrades include improved cathodes with their matching laser systems, modified activation techniques and better diagnostics. The source itself and its performance with these upgrades will be described with special attention given to recent high-intensity long-pulse operation for the E-158 fixed-target parity-violating experiment. Presented at the Workshop on Polarized Electron Sources and Polarimeters (PESP 2002), September 4-6, 2002, Danvers, MA. * Work supported by Department of Energy contract DE AC03 76SF00515. 1

The SLAC Polarized Electron Source * J. E. Clendenin, A. Brachmann, T. Galetto, D.-A. Luh, T. Maruyama, J. Sodja, and J. L. Turner Stanford Linear Accelerator Center, 2575 Sand Hill Rd., Menlo Park, CA 94025 Abstract. The SLAC PES, developed in the early 1990s for the SLC, has been in continuous use since 1992, during which time it has undergone numerous upgrades. The upgrades include improved cathodes with their matching laser systems, modified activation techniques and better diagnostics. The source itself and its performance with these upgrades will be described with special attention given to recent high-intensity long-pulse operation for the E-158 fixedtarget parity-violating experiment. 1 INTRODUCTION The polarized electron source (PES) [1] was initially commissioned for the SLC in early 1992. The present post-slc configuration is shown in Fig. 1. The all solidstate Nd:YLF-pumped Ti:sapphire short-pulse laser is used for filling PEP-II and for other experiments not requiring polarized electrons. The flash lamp-pumped longpulse Ti:sapphire laser 1 has been recently improved for a fixed-target experiment 2 that requires a high-charge polarized beam. The characteristics of the PES are listed in Table 1. Which of the two laser beams is delivered to the cathode is chosen on a pulse-to-pulse basis. Improvements to the source are discussed in Section 2 and its present performance in Section 3. 2 IMPROVEMENTS Given the uncertainties of operating a major accelerator with a polarized electron source, four guns were built for the following purposes: one operating, one on standby, one being repaired, and one for experiments in the off-line laboratory consisting of a duplicate of the first few meters of the 3-km linac. However, the addition of a load-lock 3 in 1993 improved the reliability of the PES to such an extent * Work supported by Department of Energy contract DE AC03 76SF00515. 1 A. Brachmann et al., this Workshop. 2 P. Mastromarino et al., this Workshop. 3 Before biasing the cathode, most of the load-lock apparatus must be removed and a corona shield installed around the highvoltage (HV) electrode. Because of space constraints in the linac housing, the HV insulator and corona shield is surrounded by dry gas contained within a sealed shroud. 2

that certainly no more than 3 guns are now needed. All activations 4 are done in the load-lock with the gun isolated. Re-application of Cs to the cathode is done periodically in the gun chamber under computer control (operator initiated) with the bias voltage reduced to 1 kv. Cesium channels from SAES are used exclusively. Additional NF 3 is not added inbetween activations. To monitor the cathode quantum efficiency (QE), a low-power cw diode laser is mounted on a side window of the gun with a view of the cathode. The laser is modulated at a few hertz to minimize the cw beam in the linac. A fiber-optically coupled nanoammeter installed at the gun HV electrode support flange supplies a modulated current signal to a lock-in amplifier as shown in Fig. 2. The control chassis also blanks off the diode laser during the linac rf accelerating pulse to prevent any acceleration of photoelectrons generated by this source. 5 FIGURE 1. The SLAC polarized electron source in its present configuration. The addition of the load-lock required the cathode to be mounted on a detachable puck that is supported by a 2.5-cm diameter emitter tube made from a 8-cm long Mo cup brazed to a Kovar tube, which is welded to a 304L stainless-steel (SST) tube. During operation, cold N 2 is circulated inside the tube to lower the cathode temperature to about 0 C. When activating, a resistive heater is installed. To bring the GaAs crystal to the desired 600 C, even with a heat shield on the vacuum side of 4 Activations consist of heat cleaning for 1-hour at 600 C followed by application of Cs and florine in the form of NF 3. 5 The low intensity of this laser precludes a radiation issue. 3

the crystal, the puck end of the emitter tube reaches about 800 C, which formerly resulted in frequent vacuum failures of the Mo-Kovar brazed joint. Ni-plating the Kovar prior to brazing helped. 6 Eventually an ANSYS 5.1 study of the temperature response was used to determine the optimum length of the Mo cup to keep the temperature of the brazed joint below 250 C [2]. A new Mo-length of 31 cm was chosen. Computer controlled temperature-cycling tests resulted in no failures after 70 cycles. To reduce hazards, the gas mixture in the shroud can now be readily switched from SF 6 to dry air or to a mixture of both. Generally SF 6 is used only in the first couple days of operation following closure of the shroud. TABLE 1. Source Characteristics Component Value Cathode diameter 20 mm (emitting area) Cathode holder Mo puck with Ta clip Electrodes 317L SST with low carbon content, low inclusion density Bias voltage alumina insulator 39.5-cm long by 19.0-cm ID, OD fluted Design/nominal bias voltage -150/-120 kv dc Dark current at nominal voltage 30 na typical Vacuum pumping 120 lps DI and 200 lps NEG pumps Vacuum 10-11 Torr dominated by H 2 Ti:sapphire laser systems: Nd:YLF pumped 2-ns pulse for PEP II, FFTB Flashlamp pumped Polarized 100-400 ns pulse for ESA The major improvement is in the cathode itself. The progress toward higher polarization is illustrated in Table 2. There has been only limited success in improving the cathode preparation technique. A vacuum transfer vessel allows the cathode to be activated in the laboratory and then transferred to the load-lock system at the gun under vacuum. The vessel has a small battery-operated IP, but nonetheless a final re-activation in the load-lock is generally necessary. A glove-box for performing the initial cathode cleaning in an N 2 atmosphere has also been constructed, but so far not used. Presently H* cleaning at 400 C is being studied. 7 3 PERFORMANCE Since 1992 the polarized electron source has provided ~80% of the electron beams for the SLAC linac, accumulating nearly 40,000 h of operation with availability >>95%. Except for 1992 (when there was no load-lock) the pattern has been as follows, independent of the type of cathode used (listed in Table 2). After initial activation in the load-lock, the cathode is inserted into the gun. Neither the cathode bias HV nor the presence of the electron beam are observed to have any 6 Suggested by G. Collet and R. Kirby (SLAC). 7 D.-A. Luh et al., this Workshop. 4

FIGURE 2. Use of nanoammeter (A) cathode bias for continuous monitoring of yield. TABLE 2. Cathode History Cathode P e (%) Experiment Bulk GaAs 25 1992 SLC Bulk AlGaAs 35 1992 E142 300-nm strained GaAs 65 1993 SLC 100-nm strained GaAs 80 1993 143 1994-98 SLC 1995 E154 1996 E155 1999 E155X Gradient-doped strained GaAs 80 2001-present effect on the QE. When the QE decreases to the point where the required current can no longer be obtained because of limited laser energy and/or because of the surface charge limit the cathode is re-cesiated in the gun with the cathode bias reduced to 1 kv. This typically occurs every 3-5 days and takes ~20 min., the time dominated by cycling of the bias voltage. About 0.5 ML Cs is estimated to be added per recesiation. No NF 3 is added. The re-cesiation completely restores the original QE. The rate of decrease of the QE (QE lifetime) changes over long periods (generally to shorter lifetimes), but not between re-cesiations. Re-cesiations are repeated up to 50 times over the period of a year, limited only by operational considerations to preemptively change the cathode during the accelerator scheduled major mainte- 2.5 10 12 2 10 12 electrons per pulse 1.5 10 12 1 10 12 5 10 11 FIGURE 3. Recent gun-chamber RGA mass spectrum. The intensity is in units of 10-13 Torr. The m=2 (offscale) and m=4 peaks are ~10-11 and ~10-13 Torr respectively. 0 0 50 100 150 200 250 microjoules FIGURE 4. Comparison of emission from gradient-doped cathode (top curve, squares) using 100-ns laser pulse with medium-doped (SLC type) cathode (bottom curve, circles) and 300-ns laser pulse. 5

nance periods. An important element in the cathode performance is the gun vacuum. An RGA scan for the present gun, which has not been back-filled for several years, is shown in Fig. 3. The RGA is operated continuously. The high charge that can be generated with the new gradient-doped strained-layer GaAsP-GaAs cathodes [3] is illustrated in Fig. 4. Note the absence of saturation effects in the upper curve, which is generated using a 100-ns laser pulse. Due to the longer pulse length of 300-ns, the current density in the lower curve is much less than the upper one for the same laser energy. For both curves, the laser illuminates a 20-mm diameter spot on the cathode. The parameters achieved for the currently running E158 using this new cathode are compared with those required for NLC in Table 3. Table 3. Comparison of Parameter Achieved in E158 with NLC Requirements Source Parameter E158 NLC Electrons per macropulse 8 10 11 e - 27 10 11 e - Repetition rate 120 Hz 120 Hz Macropulse length 270 ns 266 ns Micropulse spacing DC 1.4 ns Polarization ~80% 80% Intensity jitter <0.5% 0.5% Transverse jitter 5% of spot size 22% σ x, 50% σ y 4 CONCLUSIONS The SLAC polarized electron source has proven to be an outstanding performer since its initial commissioning in 1992. The source produces a charge for the presently running E158 that is close to that required for an NLC macropulse. Even higher charge has been demonstrated. Work continues at SLAC to develop a reliable low-temperature activation technique to provide more flexibility for the gradientdoped cathodes. Higher polarization is being explored, principally using superlattice semiconductor structures. 5 REFERENCES 1. R. Alley et al., The Stanford linear accelerator polarized electron source, Nucl. Instrum. and Meth. A 365 (1995) 1. 2. G. Mulhollan and G. Kraft (SLAC), private communication. 3. T. Maruyama et al., at this Workshop and A very high charge, high polarization gradient-doped strained GaAsP photocathode, Nucl. Instrum. and Meth. A 492 (2002) 199. 6