THE IBA SUPERCONDUCTING SYNCHROCYCLOTRON PROJECT S2C2

Similar documents
Recent developments in cyclotrons for proton therapy at IBA

NEW CYCLOTRON DEVELOPMENTS AT IBA

III. Proton-therapytherapy. Rome SB - 3/5 1

Particle Therapy with the Varian / ACCEL 250 MeV S.C. Proton Cyclotron

Hall-B Beamline Commissioning Plan for CLAS12

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

EPJ Web of Conferences 95,

A HIGH POWER LONG PULSE HIGH EFFICIENCY MULTI BEAM KLYSTRON

PEP II Design Outline

Particle Beam Production - A Synchrotron-Based System - Prof. Dr. Thomas Haberer Scientific-technical Director Heidelberg Iontherapy Center

INSTRUMENT CATHODE-RAY TUBE

2008 JINST 3 S LHC Machine THE CERN LARGE HADRON COLLIDER: ACCELERATOR AND EXPERIMENTS. Lyndon Evans 1 and Philip Bryant (editors) 2

DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC*

The PEFP 20-MeV Proton Linear Accelerator

45 MW, 22.8 GHz Second-Harmonic Multiplier for High-Gradient Tests*

PRESENT STATUS OF J-PARC

Mechanical aspects, FEA validation and geometry optimization

TITLE PAGE. Title of paper: PUSH-PULL FEL, A NEW ERL CONCEPT Author: Andrew Hutton. Author Affiliation: Jefferson Lab. Requested Proceedings:

HIGH POWER BEAM DUMP AND TARGET / ACCELERATOR INTERFACE PROCEDURES *

Development of Multiple Beam Guns for High Power RF Sources for Accelerators and Colliders

Hadron Therapy Technologies

Implementing a Proton Beam Scanning System within an Operating Clinical Facility

OPERATIONAL EXPERIENCE AT J-PARC

Design and Simulation of High Power RF Modulated Triode Electron Gun. A. Poursaleh

Commissioning of Accelerators. Dr. Marc Munoz (with the help of R. Miyamoto, C. Plostinar and M. Eshraqi)

The ESS Accelerator. For Norwegian Industry and Research. Oslo, 24 Sept Håkan Danared Deputy Head Accelerator Division Group Leader Beam Physics

INSTRUMENT CATHODE-RAY TUBE

Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy

Performance of a DC GaAs photocathode gun for the Jefferson lab FEL

Status of BESSY II and berlinpro. Wolfgang Anders. Helmholtz-Zentrum Berlin for Materials and Energy (HZB) 20th ESLS-RF Meeting

Experience with the Cornell ERL Injector SRF Cryomodule during High Beam Current Operation

Cyclotron Institute upgrade project. H. L. Clark, F. Abegglen, G. Chubarian, G. Derrig, G. Kim, D. May, B. Roeder and G. Tabacaru

Proton Engineering Frontier Project

ILC Damping Ring Lattice Status Report. Louis Emery and Aimin Xiao Argonne National Laboratory Presented at KEK workshop Dec 18th, 2007

4.4 Injector Linear Accelerator

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Digital BPMs and Orbit Feedback Systems

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM

LHC Beam Instrumentation Further Discussion

100 MeV H - CYCLOTRON DEVELOPMENT AND 800 MeV PROTON CYCLOTRON PROPOSAL*

Development of high power gyrotron and EC technologies for ITER

Cyclotron Institute upgrade project. H. L. Clark, F. Abegglen, G. Chubarian, G. Derrig, G. Kim, D. May, and G. Tabacaru

New Filling Pattern for SLS-FEMTO

3 cerl. 3-1 cerl Overview. 3-2 High-brightness DC Photocathode Gun and Gun Test Beamline

OF THIS DOCUMENT IS W8.MTO ^ SF6

2 Work Package and Work Unit descriptions. 2.8 WP8: RF Systems (R. Ruber, Uppsala)

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

Pseudospark-sourced Micro-sized Electron Beams for High Frequency klystron Applications

Periodic Seasonal Variation of Magnets Level of the STB ring

Development of High Power Vacuum Tubes for Accelerators and Plasma Heating

30 GHz Power Production / Beam Line

COMMISSIONING SCENARIOS FOR THE J-PARC ACCELERATOR COMPLEX

K800 RF AMPLIFIER TUBE UPGRADE

Detailed Design Report

Investigation of Radio Frequency Breakdown in Fusion Experiments

DESIGN AND PERFORMANCE OF L-BAND AND S-BAND MULTI BEAM KLYSTRONS

1.0 Abstract : 2.0 The drift tube Linac for LEHIPA

PBS Products from Pyramid

High Power Cyclotrons

Design Studies For The LCLS 120 Hz RF Gun Injector

AARHUS UNIVERSITET November, ASTRID2 Status. Heine Dølrath Thomsen on behalf of the ASTRID2 Team

TOSHIBA Industrial Magnetron E3328

5 Project Costs and Schedule

The Elettra Storage Ring and Top-Up Operation

Program to improve the ion beam formation and transmission at JYFL

An Overview of Beam Diagnostic and Control Systems for AREAL Linac

Screen investigations for low energetic electron beams at PITZ

Status of SOLARIS Arkadiusz Kisiel

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

FINAL DESIGN OF ILC RTML EXTRACTION LINE FOR SINGLE STAGE BUNCH COMPRESSOR

COMMISSIONING OF THE ALBA FAST ORBIT FEEDBACK SYSTEM

DESIGN OF 1.2-GEV SCL AS NEW INJECTOR FOR THE BNL AGS*

PEP-I1 RF Feedback System Simulation

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Oak Ridge Spallation Neutron Source Proton Power Upgrade Project and Second Target Station Project

INFN School on Electron Accelerators. RF Power Sources and Distribution

CHAPTER 4 OSCILLOSCOPES

A NOVEL GANTRY FOR PROTON THERAPY AT THE PAUL SCHERRER INSTITUTE

STATUS OF THE SWISSFEL C-BAND LINEAR ACCELERATOR

Proceedings of the 1997 Workshop on RF Superconductivity, Abano Terme (Padova), Italy

The Construction Status of CSNS Linac

18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V

THE NEW AXIAL BUNCHER AT INFN-LNS

North Damping Ring RF

CEBAF Accelerator Update. Michael Tiefenback CASA Accelerator Physics Experimental Liaison June 14, 2017

RF Design of the LCLS Gun C.Limborg, Z.Li, L.Xiao, J.F. Schmerge, D.Dowell, S.Gierman, E.Bong, S.Gilevich February 9, 2005

Next Linear Collider. The 8-Pack Project. 8-Pack Project. Four 50 MW XL4 X-band klystrons installed on the 8-Pack

Present Status and Future Upgrade of KEKB Injector Linac

Studies on an S-band bunching system with hybrid buncher

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

Quadrupoles have become the most widely used

LEP OPERATION AND PERFORMANCE WITH ELECTRON-POSITRON COLLISIONS AT 209 GEV

SPEAR 3: Operations Update and Impact of Top-Off Injection

Recent ITER-Relevant Gyrotron Tests

AREAL- Phase 1. B. Grigoryan on behalf of AREAL team

Lecture 17 Microwave Tubes: Part I

DC ELV Accelerators: Development and Application

Low-Energy Electron Linacs and Their Applications in Cargo Inspection

P. Emma, et al. LCLS Operations Lectures

Transcription:

THE IBA SUPERCONDUCTING SYNCHROCYCLOTRON PROJECT S2C2 W. Kleeven, M. Abs, E. Forton, S. Henrotin, Y. Jongen, V. Nuttens, Y. Paradis, E. Pearson, S. Quets, J. Van de Walle, P. Verbruggen, S. Zaremba, IBA, Louvain-La-Neuve, Belgium M. Conjat, J. Mandrillon, P. Mandrillon, AIMA Development, Nice, France Abstract In 2009 IBA started developing a compact superconducting synchrocyclotron as part of the small footprint proton therapy system ProteusOne. The cyclotron has been completely designed and constructed and is currently under commissioning at the IBA factory. Its design and commissioning results are presented. INTRODUCTION The ProteusOne is an innovative single treatment room solution for protontherapy. It consists of the S2C2 superconducting proton cyclotron [1], the new IBA compact gantry [2] and a state-of the-art patient treatment room facility; it is designed for lower cost and compactness to make proton therapy more widely accessible. In the new gantry, the scanning magnets are placed upstream of the last bending magnet. Figure 1 shows the layout of this new gantry. This configuration offers the combined advantage of compactness and pencil beam scanning with reasonable SAD (source to gantry axis distance). The energy selection system is included in the straight inclined part of the gantry. This also gives a considerable reduction of the facility footprint. Several presentations on the S2C2 were given at the 2012 European Cyclotron Progress Meeting (ECPM) at PSI [3]. GENERAL CONSIDERATIONS The average field in an isochronous cyclotron is limited to about 2.5 Tesla. Above this value, the flutter quickly becomes too small to provide sufficient vertical focusing and simultaneously constant orbit frequency. Much higher fields can be used in the superconducting synchrocyclotron where the requirement of isochronism is unnecessary and weak focusing is obtained from the negative gradient of the rotationally symmetric magnetic field. Other important differences exist: Figure 1: The ProteusOne compact gantry. Figure 2: The assembled cyclotron placed in the shielded beam-vault. It can be opened in the median plane as well as at the top of the cryostat. Also visible are the rotcoshield (left), the cryocooler-shield (middle) and the vacuum station (right). i) the RF frequency is periodically modulated and the beam is pulsed, ii) longitudinal dynamics becomes a major aspect of the beam physics with energy-phase oscillations bound by a separatrix, iii) beam is captured at injection only during a limited time window, iv) regenerative extraction is needed to recover the beam which has a very small turn separation at extraction, v) the extracted beam has a relatively low intensity typically in the order of namps (this is perfectly sufficient for proton therapy applications), vi) the central region is strongly reduced in size compared to an equivalent isochronous cyclotron. Figure 3: OPERA3D model of the cyclotron. 115

Table 1: General Features of the S2C2 Maximum Energy 230/250 MeV Size yoke/pole radius 1.25 m/0.50 m weight 50 tons Coil NbTi - wire in channel ramp up rate / time 2-3A/min / 4 hours windings/coil 3145 stored energy 12 MJ Magnetic field central/extraction 5.7 T/5.0 T Cryo cooling conductive 4 cryocoolers 1.5 W initial cooldown 12 days recovery after quench less than 1 day Beam pulse rate/length 1000 Hz/7 µsec RF system self-oscillating frequency 93-63 MHz voltage 10 kv Extraction Passive regenerative Ion source PIG cold cathode Central region removable module New computational tools where developed and existing tools improved, to be able to simulate all aspects of beam dynamics in this new accelerator [3]. Some main features and parameters of the cyclotron are listed in Table 1. The new mapping system that has been developed for this cyclotron is described in a separate paper [4]. Figure 2 shows the cyclotron in the beam-vault. MAGNETIC DESIGN The optimization of the magnetic circuit has been a long process carried out by the three main contributors to this development: IBA, AIMA and ASG (manufacturing company used for the coil and the cryostat). Many aspects have to be considered: i) optimization of the pole-gap profile, ii) definition of the pole radius and the 230 MeV extraction radius iii) optimization of coil current density and dimensions to assure a margin with respect to the critical surface and to allow operation up to 250 MeV iv) dimensioning of the yoke to reasonably balance the outside stray fields, v) dimensioning and placement of all horizontal and vertical yoke penetrations (ports are needed for RF, ion-source, vacuum, beam exit, cryocoolers, 3 horizontal and 2x3 vertical tierods) vi) the optimization of the extraction system vii) the shielding required for external systems such as the rotco and the cryo-coolers viii) the influence of the external iron systems on the accelerated beam ix) the influence of the fringe field on the external beam line x) median plane errors introduced by the vertical asymmetry in the magnetic design and compensation of these errors, xi) magnetic forces acting on the return yoke, the coils, the extraction system elements, external components etc. xii) design 116 Figure 4: Upper => OPERA2D model showing the particular pole-gap shape that is used for obtaining the required focusing properties and the field distribution on the coil at nominal current. Middle => separate contributions of the iron and the coil to the average field; for comparison the field-profile at one fixed azimuth as produced by the regenerator is also shown. Lower => magnetization curve with the seperate contribution of the iron; the maximum field on the coil is also shown. of the harmonic coils, xiii) compensation of first harmonic field errors, xiv) the influence of the cyclotron feet and the yoke lifting system xv) and many others. For the design of the superconducting coil, the transient behavior of the magnet with eddy currents and AC-losses and also the quenchbehavior need to be studied in detail. To make this optimization, magnetic finite element models were made in OPERA2D, OPERA3D and CST. Especially in OPERA3D, very detailed models were made as illustrated in Figure 3. Some main magnetic field properties as obtained with OPERA2D and shown in Figure 4.

CENTERING OF THE MAIN COIL To accelerate the beam up to extraction during 40000 turns, correct horizontal and vertical centering of the coil is essential. An innovative method (patent pending EP13170532) permits to do this with millimeter to submillimeter precision. It is based on measurement of the vertical and radial field profiles between the two main coils, where the field is coil- dominated. Measurements are made in three radial ports (vacuum, rotco and source). For horizontal centering, the B z -profiles are compared in the radial region where the radial fall-off is at maximum. For vertical centering, the B r -profiles are measured in a region where B z 0 so that the tilt-error of the Hall-probe is not important. Results are shown in The upper of Figure 5 shows the deduced horizontal coil position (as a function of the radial position of the probe) in the initial configuration and in the final configuration where the coil is centered to better than 0.5 mm. The lower figure shows the vertical off-centering measured in each of the three ports. We estimate this positioning error to be about 1 mm or less, depending on the port. The centering of the main coil was done hand-in-hand with measurements of the magnetic forces acting on the cold mass. These are obtained from strain gauges mounted in the tierod-assemblies. A special effort was made to reduce the total vertical force on the cold-mass by proper coil positioning and by placing iron shims on top of the return yoke. OPERA2D calculations indicate that the median plane error is reduced when this vertical force is compensated. Figure 6 shows the measured forces as function of the ramp-current. The vertical force in the final configuration (after shimming) is substantially lower than in the Figure 6: Measured forces acting on the cold mass. initial configuration. The horizontal forces are larger. This is due to the extraction system and the asymmetry of the yoke. During ramp-up the coil moves about 1.5 mm, to arrive at a centered position at nominal current. BEAM EXTRACTION Regenerative extraction based on on 2Q h = 2-resonance is used. The regenerator (lower Figure 3) creates a strong bump (middle Figure 4) which locally increases the radial focusing and locks Q h to 1. It is essential to avoid the Walkinshaw resonance (upper Figure 7). Extraction sets in at this condition and a displacement of the beam towards the extraction channel steadily builds up (lower Figure 7). Correctors are used to compensate the undershoots of the regenerator and the extraction channel. A 3-bar corrector guides the beam through the fringe field. Figure 5: Measured coil positions as a function of the radial position of the Hall probes. Figure 7: Extraction is based on the 2Q h = 2 resonance. 117

RF SYSTEM The RF resonator (Figure 8) operates as a half-wave transmission line terminated on one side by the 180 dee and on the opposite side by the rotco. The rotco is an innovative patented design having 8-fold symmetry. This allows excellent mechanical stability and very good reproducibility of the RF pulse. It rotates at 7500 rpm giving a 1 khz repetition rate. The measured RF-frequency curve (Figure 9) agrees very well with the CST-model that was developed. The structure is coupled to a triode tube and operates in self-oscillating mode. The dee and the centerdummy dee are biased at 1 kvolt, to suppress the multipactor. Two side-stubs provide fine-tuning of df/dt during capture. To avoid eddy currents, the rotco and the triode are placed in a shielded volume, outside of the yoke. Figure 8: The fully assembled RF-system mounted on a test-bench. Detail of the rotco modeling in CST (top). Figure 9: The measured RF-frequency curve. ION SOURCE Fast and precise pencil beam scanning (PBS) requires a high dynamic range (a factor 100) in the charge per pulse and also good pulse repeatability. A cold-cathode PIGsource is used because of its fast response, long cathode 118 Figure 10: The very good ion source arc-stability as measured at 5 Tesla in the S2C2 also promises good beam pulse reproducibility. life-time and good pulse stability. The source is pulsed during 50 µsec, in synchrony with the RF. The source was initially characterized in a 1 Tesla test-stand [3], where a DC extracted beam current up to 6 ma was demonstrated (much more than required). Subsequent tests in the S2C2 at 5 Tesla confirm a very good arc-stability as a function of arc-current (Figure 10). This promises the good beampulse reproducibility as required. CENTRAL REGION The central region (CR) is extremely compact with a first turn radius less than 2.5 mm and the first 100 turns within a radius of about 3 cm. The vertical gap tapers in a cone to enable better vertical focusing and transit time factors during the first turns. The CR and ion source are removable as one sub-system for easy maintenance and precise alignment (Figure 11). Detailed modeling of the full accelerating structure was done in OPERA3D as well as CST. Particle tracking in 3D was done, to optimize the geometry and to predict the accelerated beam quality and properties as needed for the simulation of the extraction process. Good internal beam quality is predicted, as shown in Figure 12. EXTRACTED BEAM LINE A permanent magnet quadrupole in the return yoke matches the cyclotron to the beam line. The beam is focused by a quadrupole doublet on a 1x1 mm 2 (1-sigma) beam spot on the degrader placed at 2 meters from the yoke exit (see Figure 13). In between the two quads is an adjustable horizontal collimator that cuts the horizontal beam divergence at the degrader, to provide constant optical conditions independent of the gantry rotation angle. CURRENT STATUS The cyclotron is now fully assembled and installed in a shielded bunker for beam testing. The magnet cools down correctly. The magnetic field has been mapped and found to be in very good agreement with OPERA3Dsimulations [4]. The coil has been centered correctly and

Figure 13: Extracted beam line and degrader placed 2 meters from the yoke. Figure 14: The extracted beam observed by a radiation sensitive foil. REFERENCES [1] E. Pearson, W. Kleeven, J. Van de Walle and S. Zaremba, The New IBA Superconducting Synchrocyclotron (S2C2): from Modeling to Reality, Accepted for publication in the proceedings of the 11th Int. Topical Meeting on Nucl. Appl. of Accelerators, 5-8 August, Bruges, Belgium. Figure 12: Radial and vertical beam emittances at 2 MeV as obtained by particle tracking. [2] E. Pearson et.al., Magnet Developments and Commissioning for the IBA Compact Gantry, Accepted for publication in the proceedings of the 23rd Int. Conf. on Magnet Technology, Boston, 2013. the vertical forces on the cold mass have been substantially reduced by asymmetric shims placed on top of the yoke. The RF-system, the ion source and the central region have all been tested and are performing well. The external beam line up to the degrader has been fully designed and is under construction. At the end of August 2013 (a few days [3] European Cyclotron Progess Meeting 2012, PSI, Switzerland, http://indico.psi.ch/conferencedisplay.py? confid=1146 [4] J. Van de Walle et. al., Mapping of the new IBA Superconducting Synchrocyclotron (S2C2) for Proton Therapy, In Proceedings of Cyclotrons2013. 119 Figure 11: Upper: the CR as well as the full dee-structure are modelled in detail in OPERA3D, allowing precise tracking of the accelerated beam. Lower: the ion source and CR can be removed as one sub-system for easy maintanance and alignment. after the beginning of beam tests) we have observed the beam extracted from the cyclotron by the colouring of a thin dose-sensitive foil placed at the end of the beam tube (Figure 14). This was obtained without any re-adjustment of subsystems in the cyclotron. Currently internal measurements are in progress to characterize the beam quality and to fine-tune the central region and the extraction system.