Quartzlock Model A7-MX Close-in Phase Noise Measurement & Ultra Low Noise Allan Variance, Phase/Frequency Comparison

Similar documents
Signal Stability Analyser

Instrumentation Grade RF & Microwave Subsystems

Agilent E5500 Series Phase Noise Measurement Solutions Product Overview

DA MHz Series of Narrowband or Wideband Distribution Amplifiers

DA : Series of Narrowband or Wideband Distribution Amplifiers

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

Technical Data. HF Tuner WJ-9119 WATKINS-JOHNSON. Features

DA E: Series of Narrowband or Wideband Distribution Amplifiers

Sources of Error in Time Interval Measurements

Precision testing methods of Event Timer A032-ET

Keysight Technologies E5500 Series Phase Noise Measurement Solutions

Tech Note: How to measure additive phase noise of amplifiers using the 7000 Series

R-1580A Microwave Downconverter. Product Brochure

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

Techniques for Extending Real-Time Oscilloscope Bandwidth

INTEGRATED ASSEMBLIES MICROWAVE SOLUTIONS FROM TELEDYNE COUGAR

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series

The high-end network analyzers from Rohde & Schwarz now include an option for pulse profile measurements plus, the new R&S ZVA 40 covers the

Datasheet SHF A

Broadcast Television Measurements

LaserPXIe Series. Tunable Laser Source PRELIMINARY SPEC SHEET

SYNTHESIZED COMMUNICATION CONVERTERS

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

CAM Series Channelized Agile Audio/Video Modulators

Synthesized Clock Generator

High-Value 100 Series (Outdoor)

AR SWORD Digital Receiver EXciter (DREX)

Trigger synchronization and phase coherent in high speed multi-channels data acquisition system

VXI RF Measurement Analyzer

Agilent 5345A Universal Counter, 500 MHz

USB-TG124A Tracking Generator User Manual

Application Note DT-AN DTU-315 Verification of Specifications

MULTIBAND 1/3 RACK-MOUNTED

GALILEO Timing Receiver

R-1550A Tempest Wide Range Receiver

Product Guide. WaveAnalyzer High-Resolution Optical Spectral Analysis

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications

N5264A. New. PNA-X Measurement Receiver. Jim Puri Applications Specialist March Rev. Jan Page 1

PXI MICROWAVE. Model PXI-1420

2 MHz Lock-In Amplifier

VXIbus Microwave Downconverter

AFRL-RY-WP-TR

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel)

Accuracy Delta Time Accuracy Resolution Jitter Noise Floor

Clock Measurements Using the TimePod 5330A with TimeLab and Stable32

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor

Wideband Downconverters With Signatec 14-Bit Digitizers

Synthesized Block Up- and Downconverter Indoor / Outdoor

FREQUENCY CONVERTER 1/3 RACK-MOUNTED BLOCK CONVERTER. Narda-MITEQ FEATURES OPTIONS. Unit shown with Option 17. Unit shown without Option 17

Advanced Test Equipment Rentals ATEC (2832)

Dac3 White Paper. These Dac3 goals where to be achieved through the application and use of optimum solutions for:

R3267/3273 Spectrum Analyzers

Features. PFD Output Voltage 2000 mv, Pk - Pk. PFD Gain Gain = Vpp / 2π Rad khz 100 MHz Square Wave Ref.

SMS3000X Series Spectrum Analyzer

GFT Channel Digital Delay Generator

Model RTSA7550 Specification v1.1

The Distortion Magnifier

1/3 RACK-MOUNTED BLOCK CONVERTERS

Advanced Test Equipment Rentals ATEC (2832)

Extension kit for R&S Vector Network Analysers

GHz Sampling Design Challenge

CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock

Clock Jitter Cancelation in Coherent Data Converter Testing

How advances in digitizer technologies improve measurement accuracy

MP7200. RF Recorder / Player.

CHAPTER 3 SEPARATION OF CONDUCTED EMI

PXI MICROWAVE IF DOWNCONVERTER MODULE

Loop Bandwidth Optimization and Jitter Measurement Techniques for Serial HDTV Systems

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X

!Ill ~ 168. Model490 Dual Input, Dual Trace Automatic Peak Power Meter

USB-SA124B Spectrum Analyzer User Manual

The PQConsole Program

RF Level Test System +20 dbm to 130 dbm

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

ECE 5765 Modern Communication Fall 2005, UMD Experiment 10: PRBS Messages, Eye Patterns & Noise Simulation using PRBS

What to look for when choosing an oscilloscope

Asynchronous inputs. 9 - Metastability and Clock Recovery. A simple synchronizer. Only one synchronizer per input

Simplified Block Diagram

PicoScope 9300 Series migration guide

SHF Communication Technologies AG,

MIGRATION TO FULL DIGITAL CHANNEL LOADING ON A CABLE SYSTEM. Marc Ryba Motorola Broadband Communications Sector

International Journal of Engineering Research-Online A Peer Reviewed International Journal

Agilent Modular Instrumentation

Agilent PN Time-Capture Capabilities of the Agilent Series Vector Signal Analyzers Product Note

SSA3000X Series Spectrum Analyzer

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16]

CS311: Data Communication. Transmission of Digital Signal - I

Mastering Phase Noise Measurements (Part 3)

14 GHz, 2.2 kw KLYSTRON GENERATOR GKP 22KP 14GHz WR62 3x400V

ThinkRF D GHz RF Downconverter

Field Programmable Gate Array (FPGA) Based Trigger System for the Klystron Department. Darius Gray

SDA 3302 Family. GHz PLL with I 2 C Bus and Four Chip Addresses

Spectrum Analyzer 1.6 GHz 3 GHz HMS-X

MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Modules

FREQUENCY CONVERTER HIGH-PERFORMANCE OUTDOOR BLOCK UP AND DOWNCONVERTERS. Narda-MITEQ 1 FEATURES OPTIONS

Converters Series Tri-Band Frequency Converters. Back to Multiband. Specifications Upconverters Downconverters Phase Noise Options Rear Panel

e2b calibration 521 Fifth Avenue Chardon, Ohio Phone: Fax: Certificate Of Calibration

SC24 Magnetic Field Cancelling System

Transcription:

Quartzlock Model A7-MX Close-in Phase Noise Measurement & Ultra Low Noise Allan Variance, Phase/Frequency Comparison Measurement of RF & Microwave Sources Cosmo Little and Clive Green Quartzlock (UK) Ltd, Gothic, Plymouth Road, Totnes, Devon, TG9 5LH, UK. Tel: +44 1803 862062 cgreen@quartzlock.com www.quartzlock.com SCOPE This presentation describes the measurement of time domain stability of microwave and RF sources and passive devices between 20MHz and 40GHz. An outline of a suitable measurement instrument is proposed. INTRODUCTION The A7MX frequency/phase difference comparator is a Quartzlock product for measuring a wide range of frequency standards, isolation amplifiers, frequency multipliers and dividers, and passive devices such as cables. The frequency reference can be either 5MHz or 10MHz. The high resolution input operates on 5MHz or 10MHz with only a small deviation from the nominal frequency being permitted. The time resolution on the high resolution input is 0.125fs. The actual achievable resolution is noise limited. The high resolution input is intended for the measurement of frequency standards and atomic clocks with outputs of either 5MHz or 10MHz. The low resolution input operates between 50kHz and 65MHz with a basic time resolution of 12.5ps. This input cannot give an Allen variance noise floor lower than 3E-11 at tau=1second due to the resolution limit. There are three possibilities for extending the instruments frequency range to the high RF and microwave domains. These are: a) Construct a high resolution downconvertor which converts the input frequency to 10MHz for input to the high resolution input of the A7MX. Because of the narrow band of frequencies the high resolution input will accept, the downconvertor local oscillator (LO) would have to have 0.01Hz resolution. This is extremely difficult to do. b) Construct a low resolution downconvertor which converts the input frequency to an IF in the range 10MHz to 65MHz for input to the low resolution input of the A7MX. The step size of the LO can then be equal to the IF bandwidth, ie for LO synthesizer steps of 10MHz, the IF bandwidth would be 10MHz. There is plenty of room in the FXQ measurement range to accommodate the IF bandwidth. The disadvantage of this approach is that it is difficult to

provide a range of measurement bandwidths, important for getting the best possible time resolution. c) Construct a high resolution downconvertor using double conversion. The first IF would be in the range 15 to 25MHz. The first LO would only need 10MHz steps as in approach (b). The second LO would tune over the range 15 to 25MHz and would have a high resolution of sub 1Hz. The final IF could be 100kHz. At this frequency filtering is much easier, and in fact the FXQ could provide the narrow measurement filters. This has already been proposed for the A7000. (lower cost version of the A7MX) The general problem with downconvertors is that noise and phase instabilities in the LO are indistinguishable from noise and phase instabilities on the measured signal. Thus the LO synthesizer needs impeccable performance, both as regards phase noise and Allen variance noise floor. The Allen variance noise floor at taus greater than 10seconds will be largely determined by phase variations due to temperature gradients. Between 0.1 and 10 seconds there will be a contribution from both temperature instabilities and phase noise. Below 0.1 seconds the LO phase noise will dominate. COMPARISON WITH OTHER INSTRUMENTS Before deciding on an ideal specification of a microwave phase and frequency measuring instrument, it would be useful to consider different types of competing instruments. a) Frequency counters There are many types of frequency counter, however what is common to all instruments in this category is that they make time measurements of the zero crossing of the signal to be measured. An older generation reciprocal counter will make two measurements of the zero crossings, at the start and end of the gate period. By also counting the number of zero crossings between time tags, the frequency can be calculated. The total phase advance between time tags can also be calculated. The time resolution cannot be better than that of a single time tag. This is typically 80ps to 1ns. More modern counters make multiple time tag measurements during the gate period, at a rate up to 250ks/s. The intermediate time tags can be used to improve the resolution by root(n), where N is the number of measurements available. Some counters such as the most advanced made by Pendulum can store and output a large number of time tags for external processing. External processing can identify and measure phase modulation on the original signal, and can calculate Allen variance. Counters can be extended to microwave signals by either downconverting or prescaling. Prescalers are now available up 20GHz. Downconverting using a sampling mixer is more complicated, but can extend the frequency up to 50GHz. The advantage of the downconversion approach is that the basic time resolution of the counter is improved by the ratio of the input frequency to the IF frequency that is measured. A prescaler does not affect the time resolution. Counters are almost never specified in terms of Allen variance noise floor (phase stability) or long term drift. Resolution is usually specified, and is dominated by trigger noise, as the input bandwidth is very wide. For this reason counters can never achieve the lowest Allen variance

noise floors. Counters are usually dual channel. They can make time difference measurements between the channels with one channel acting as the start gate, and the other the stop gate. By using a single source with a power splitter, time domain stability of passive devices such as amplifiers can be measured by placing the device in the path between one output of the power splitter and the counter channel. The resolution of this technique is limited to the basic time resolution of the counter. b) Modulation domain analysers Modulation domain analysers are frequency counters with very rapid measurement rates, and a deep measurement memory. Measurement rates go up to 80Ms/s. The frequency resolution inversely depends on the measurement rate ( reciprocal of gate time). The single shot time resolution is usually around 100ps. Modulation domain analysers are useful for measuring phase/frequency modulation and drift on CW or pulsed sources. because of their rapid measurement rate, they can measure very short pulses used in pulse compression radar. c) Signal source analysers These instruments are quite new, and combine a number of measurement types into one instrument. They basically work by downconverting the input signal to either baseband or a lower IF. This is then sampled by a high rate ADC (250Ms/s). A digital phase or frequency lock loop is used to lock the LO at the correct frequency. They usually have 2 channels, each with its own reference oscillator and synthesised LO. By making cross correlation measurements between the channels, the reference and LO noise can be eliminated leaving only the noise from the source. This enables quite low phase noise floors irrespective of the reference noise. They can also make rapid frequency measurements and can replace modulation domain analysers. In fact Agilent have now discontinued all their modulation domain analysers. The frequency resolution is not particularly good, reaching 10-12 for 1 second gate time at 1GHz, but does not increase with gate time in the usual way. The frequency resolution is a complicated function of the down convertor bandwidth and gate time, for more information refer to the data sheet for the E5052B. d) Vector network analysers Network analysers cannot measure sources, but are very good at measuring phase or group delay ( rate of change of phase with frequency, equal to the time delay of the device) in amplifiers or other passive devices. Because they have a narrow resolution bandwidth, they have very good phase resolution. Typical resolution for differential phase measurements is 0.1 degrees, with a temperature stability of 0.2degrees / degc. At 10GHz this is a resolution of 30fs, with a stability of 60fs. A vector network analyser can also measure absolute delay, this is called group delay and is calculated from the rate of change of phase over a known frequency sweep. Group delay resolution can be 10ps over 100MHz sweep range. e) Frequency/phase analysers Instruments such as the A7MX are quite rare, and really require their own category. They are

basically counter based instruments, but the input signal receives considerable modification, including down conversion, before being measured. Most instruments of this type use a dual channel downconvertor with a single LO. The most basic form measures the phase difference between two signals of the same nominal frequency. In this situation the IF frequencies are the same, and time difference measurements may be made between the channels using a frequency counter in its time difference mode. The counter time resolution is multiplied by the downconversion ratio (the ratio of the input signal frequency to the IF frequency). The advantage of the basic form is that to a first approximation the noise and drift of the LO cancels out. The problem with the basic form is that as the common IF is made lower in frequency, with the intention of increasing the time resolution, the available measurement bandwidth gets less. The other problem is that the phase jitter in the required zero crossing detectors increases as the slew rate of the IF decreases. Very narrow bandwidth zero crossing detectors have been developed to avoid this. The Quartzlock A7MX actually has two downconversions and two PLL multipliers in each channel in order to get the required time resolution multiplication without the final IF becoming too low in frequency. The multiplication is either 103 or 105, with a final IF of 100kHz. The channels are subtracted to remove noise on the common LO using an offset mixing scheme before the frequency counter, so only a single channel counter is required. Extension of the basic instrument to measurement of an input frequency substantially different from the reference frequency means that the downconversion LO must be different for the two channnels. This immediately means that the condition for cancellation of the LO phase jitter and drift is lost. The characteristics of the LO will be measured as well as the input signal. If one examines the block diagram of the Picotime, for example, one can see that it is essentially a single channel downconvertor. The 10MHz reference is used to generate two LOs using DDS synthesisers which downconvert the input to an IF of 1kHz. (double conversion). All the drift and spurii of the LOs will be measured along with the input signal. DISCUSSION OF REQUIREMENTS FOR NEW INSTRUMENT The proposed microwave frequency and phase analyser will be used for the following measurement tasks: a) Measurement of time domain stability of sources between 10MHz and 20(40) GHz referred to a reference source of 10MHz (single channel mode) b) Measurement of time domain stability of two identical microwave sources with a relatively small frequency offset. (dual channel mode) c) Measurement of time domain stability of microwave passive devices ( not sources) (dual channel mode) For requirement (a) the main competition is a conventional and relatively low cost microwave counter. The Pendulum CNT-91 offers a fixed single shot time resolution of 50ps, applicable up to 20GHz. It uses prescalers to cover the microwave range, rather than downconvertors, so there is no improvement in resolution at higher frequencies. The frequency resolution claims to be 12 digits in 1sec of measurement time. This is quite a vague specification, but implies resolution of 1ps after averaging. The counter makes multiple time tag measurements and

improves its resolution by linear regression of the time tags to obtain average frequency during the measurement period. As with all counters, there is no specification on Allen variance noise floor or drift.the basic counter will show Allen variance at one gate time, but in order to produce a graph the counter must be used with a PC and the Timeview software. A new instrument based on a downconvertor would certainly be able to get much better resolution than 1ps. An instrument based on approach (b) above at 10GHz would improve the basic resolution of the FXQ ( the phase meter used in the A7MX) by 500 times, which would give a basic time resolution of 25fs. It is very difficult to estimate the usable noise limited resolution, however, as this will depend upon the noise performance of the LO. One might question whether there are any free running sources to be measured at microwave frequencies. Most microwave sources are synthesized, in which case there is hardly any point in making Allen variance measurements at long averaging times as the performance will be equal to the synthesizer reference that is used, and this could be measured using a standard A7MX. Conceivable microwave measurements using a 2 channel instrument might be the phase stability of cables and other passive ( not sources) devices such as amplifiers, phase shifters, and power splitter/isolation devices. The phase performance of these is important for applications such as phased array radars. Unfortunately most labs seriously developing microwave devices will have a vector network analyser, and will use this to evaluate phase stability. In single channel mode, the instrument could measure frequency stability against a 10MHz reference. By using the same sort of averaging techniques as the Pendulum counter ( linear regression of multiple time tags) resolution of 10E-14 in one second of gate time could be achieved. The actual Allen variance floor will be determined by the stability of the local oscillators. OUTLINE PERFORMANCE OF NEW INSTRUMENT The proposed instrument will have the following features: a) Two measurement channels with common local oscillators in order to measure identical frequency sources and passive devices. In this mode substantial cancellation of the local oscillator noise and drift can be expected. b) A basic measurement range of 10MHz to 2GHz c) An extended measurement range of 2GHz to 20GHZ ( possibly 30GHz) d) Measurement bandwidths of 10kHz to 0.1Hz provided by digital filters e) Minimum single shot phase resolution of 125fs at 10MHz rising to 0.625fs at 2GHz. f) A user interface similar to the existing A7MX, with displays of Allen variance and power spectral density.

OUTLINE OF DETAILED DESIGN It is not possible to be too explicit about the design details at this moment. The following comments can be made: a) The instrument will be double conversion on the basic range and triple conversion on the extended range b) On the basic range the first IF will be tunable between 15MHz and 25MHz. The second IF will be 100kHz c) On the extended range the first IF will be tunable between 1GHz and 2GHz ( tbc) d) In order to achieve the best possible phase stability, the design of the first LO synthesiser, which will cover the range 25MHz to 2GHZ in steps of 10MHz, will be optimised for phase stability with temperature and low noise. It is probable that a harmonic locking loop will be used with a sampling phase detector. This type of loop uses no digital dividers. e) The second LO synthesiser covers the range 15MHz to 25MHz in steps of <0.1Hz. A fractional N loop will probably be used covering the range 960MHz to 1.6GHz followed by a digital divide by 64. The divider will reduce spurii and phase noise on the synthesiser by about 36dB. Drift and noise in the divider will probably be the limiting factor. f) The extended range will be accomplished by a dual channel harmonic downconvertor using sampling mixers. The sampler drive will be at about 1GHz. Due to the wide tunable bandwidth of the first IF in this mode (1GHz to 2GHz) only two possible sampler drive frequencies will be needed to cover all the possible input frequencies. Thus the synthesiser used to drive the sampler will be quite simple.