Spatial Response of Photon Detectors used in the Focusing DIRC prototype

Similar documents
Study of Timing and Efficiency Properties of Multi-Anode Photomultipliers

Focusing DIRC R&D. J. Va vra, SLAC

PID summary J. Va vra

Photon detectors. J. Va vra SLAC

SLAC Cosmic Ray Telescope Facility

Table. J. Va vra,

Performance of the MCP-PMT for the Belle II TOP counter

Large photocathode 20-inch PMT testing methods for the JUNO experiment

Lifetime of MCP-PMTs

HAPD and Electronics Updates

Front End Electronics

MCP Signal Extraction and Timing Studies. Kurtis Nishimura University of Hawaii LAPPD Collaboration Meeting June 11, 2010

Psec-Resolution Time-of-Flight Detectors T979

Front End Electronics

Photo Multipliers Tubes characterization for WA105 experiment. Chiara Lastoria TAE Benasque 07/09/2016

TORCH a large-area detector for high resolution time-of-flight

The hybrid photon detectors for the LHCb-RICH counters

Commissioning and Initial Performance of the Belle II itop PID Subdetector

Scintillation Tile Hodoscope for the PANDA Barrel Time-Of-Flight Detector

Updates on the Central TOF System for the CLAS12 detector

MCP Upgrade: Transmission Line and Pore Importance

Status of the CUORE Electronics and the LHCb RICH Upgrade photodetector chain

Frontend Electronics for high-precision single photo-electron timing

A prototype of fine granularity lead-scintillating fiber calorimeter with imaging read-out

Imaging TOP (itop), Cosmic Ray Test Stand & PID Readout Update

arxiv: v1 [physics.ins-det] 1 Nov 2015

PoS(PhotoDet 2012)018

UVscope an instrument for calibration support

THE TIMING COUNTER OF THE MEG EXPERIMENT: DESIGN AND COMMISSIONING (OR HOW TO BUILD YOUR OWN HIGH TIMING RESOLUTION DETECTOR )

Tests of Timing Properties of Silicon Photomultipliers

Review of High Quantum Efficiency Large Area Photomultiplier Tubes

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND. Doug Roberts U of Maryland, College Park

The 20 inch MCP-PMT R&D in China

Large Area, High Speed Photo-detectors Readout

CAEN Tools for Discovery

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

Photodetector Testing Facilities at Nevis Labs & Barnard College. Reshmi Mukherjee Barnard College, Columbia University

The TORCH PMT: A close packing, multi-anode, long life MCP-PMT for Cherenkov applications

A new Scintillating Fibre Tracker for LHCb experiment

Solid State Photon-Counters

First evaluation of the prototype 19-modules camera for the Large Size Telescope of the CTA

In-process inspection: Inspector technology and concept

SPATIAL LIGHT MODULATORS

Lifetime of MCP-PMTs

ELECTRON OPTICS OF ST-X, ST-Y SERIES OF STREAK & FRAMING CAMERA TUBES

SciFi A Large Scintillating Fibre Tracker for LHCb

li, o p a f th ed lv o v ti, N sca reb g s In tio, F, Z stitu e tests o e O v o d a eters sin u i P r th e d est sezio tefa ectro lity stem l su

Eric Oberla Univ. of Chicago 15-Dec 2015

TitleLarge strip RPCs for the LEPS2 TOF. Author(s) Chu, M.-L.; Chang, W.-C.; Chen, J.- Equipment (2014), 766:

Application of Hamamatsu MPPC to T2K near neutrino detectors

Hamamatsu R1584 PMT Modifications

Single Photoelectron timing resolution of SiPM

Progress Update FDC Prototype Test Stand Development Upcoming Work

The Alice Silicon Pixel Detector (SPD) Peter Chochula for the Alice Pixel Collaboration

SLAC National Accelerator Laboratory, CA, USA. University of Hawaii, USA. CEA/Irfu Saclay, France *

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Digital BPMs and Orbit Feedback Systems

An extreme high resolution Timing Counter for the MEG Upgrade

Comparison of SONY ILX511B CCD and Hamamatsu S10420 BT-CCD for VIS Spectroscopy

Beam test of the QMB6 calibration board and HBU0 prototype

DPD80 Infrared Datasheet

The Scintillating Fibre Tracker for the LHCb Upgrade. DESY Joint Instrumentation Seminar

Measurement of Microdisplays at NPL

Standard Operating Procedure of nanoir2-s

arxiv: v2 [physics.ins-det] 24 Mar 2015

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

OPTICAL POWER METER WITH SMART DETECTOR HEAD

Advanced Sensor Technologies

SPE analysis of high efficiency PMTs for the DEAP-3600 dark matter detector

Concept and operation of the high resolution gaseous micro-pixel detector Gossip

Reading a GEM with a VLSI pixel ASIC used as a direct charge collecting anode. R.Bellazzini - INFN Pisa. Vienna February

Production and Development status of MPPC

Characterizing Transverse Beam Dynamics at the APS Storage Ring Using a Dual-Sweep Streak Camera

LCOS-SLM (Liquid Crystal on Silicon - Spatial Light Modulator)

High Efficiency White OLEDs for Lighting

Imaging diagnostico in Sanità Stato attuale e prospettive

MPPC (multi-pixel photon counter)

High ResolutionCross Strip Anodes for Photon Counting detectors

PMT Gain & Resolution Measurements in High Magnetic Fields

Hamamatsu Silicon Sensors

DPD80 Visible Datasheet

Results on 0.7% X0 thick Pixel Modules for the ATLAS Detector.

Lensed Fibers & Tapered Ends Description:

Development of an Abort Gap Monitor for High-Energy Proton Rings *

Liquid Xenon Scintillation Detector with UV-SiPM Readout for MEG Upgrade

Application note. Materials. Introduction. Authors. Travis Burt, Huang ChuanXu*, Andy Jiang* Agilent Technologies Mulgrave, Victoria, Australia

CGEM-IT project update

Very High QE bialkali PMTs

LHC Beam Instrumentation Further Discussion

APPLICATION NOTE. Fiber Alignment Now Achievable with Commercial Software

Prospect and Plan for IRS3B Readout

ADVANCED OPTICAL FIBER SOLUTIONS

A flexible FPGA based QDC and TDC for the HADES and the CBM calorimeters TWEPP 2016, Karlsruhe HADES CBM

Glast beam test at CERN

NONDESTRUCTIVE INSPECTION OF A COMPOSITE MATERIAL SAMPLE USING A LASER ULTRASONICS SYSTEM WITH A BEAM HOMOGENIZER

Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720

A Review of Tracking Sessions

Wavelength selective electro-optic flip-flop

Spatial Light Modulators XY Series

Transcription:

Spatial Response of Photon Detectors used in the Focusing DIRC prototype C. Field, T. Hadig, David W.G.S. Leith, G. Mazaheri, B. Ratcliff, J. Schwiening, J. Uher, J. Va vra SLAC 11/26/04 Presented by J. Va'vra, RICH 2004 1

Focusing DIRC prototype concept Calibration Fiber Fused Silica bar Focal plane Detector Mirror Filled with KamLand mineral oil 3D imaging: x,y, and TOP (TOP = timeof-propagation of photon in the bar) TOP is measured to σ ~100ps, which allows to correct out the chromatic error contribution to the Cherenkov angle error, and to suppress the background. Spherical mirror removes a bar thickness from the resolution consideration. The size of the prototype designed to yield a similar angular resolution as the present BaBar DIRC, given the pad size of the chosen photon detectors. At present, we do not have a solution for a detector working in the magnetic field. 11/26/04 Presented by J. Va'vra, RICH 2004 2

Expected performance and angular errors Focusing DIRC estimate assumes ~6mm 2 pixel size, completely corrected chromatic error, optics to remove the bar thickness, no loss of photons in the photon detectors. Ultimate DIRC estimate assumes, in addition, an infinitely precise photon detector. 11/26/04 Presented by J. Va'vra, RICH 2004 3

What matching liquid to use in the box? KamLand experiment mineral oil is good match to Fused silica refraction index. However, its transmission is worse than that of pure water, and it limits the performance of the Focusing DIRC prototype. No purification of the mineral oil attempted yet at SLAC. The mineral oil is a temporary solution, as the final mirror of the Focusing DIRC would be made of solid Fused silica, probably. In BaBar DIRC, it is the EPOTEK-301-2 optical glue which limits the bandwidth. 11/26/04 Presented by J. Va'vra, RICH 2004 4

Hamamatsu H-8500 Flat panel MaPMT Hamamatsu Co. data sheet 11/26/04 Presented by J. Va'vra, RICH 2004 5

Burle 85011 MCP-PMT Burle Co. data sheet 11/26/04 Presented by J. Va'vra, RICH 2004 6

Scanning setup to measure the PMT spatial response x&y stage for the fiber final focus : Stepper motor moves the end of the fiber equipped with a lens, resulting in the spot size of ~150 µm. The linear motor is set typically to: x-step ~ 100µm & y-step ~ 1mm. Light source: - PiLas laser diode operating in single photoelectron mode. - 635 & 430nm (on loan from T. Sumyioshi). - Fiber is 63µm dia. multi-mode fiber, equipped with lenses at both ends. Analysis: - A hit is accepted into the efficiency definition if it is within a time window, and it is on the same pad as the laser head is pointing to. - To get a relative efficiency we normalize to the 2 inch dia. Photonis XP 2262B PMT ( or the DIRC PMT, ETL 9125FLB17). - DAQ trigger rate: 20kHz.. 11/26/04 Presented by J. Va'vra, RICH 2004 7

Resolution of the scanning system Hamamatsu Flat Panel H8500 MaPMT #2: Micro-structure of the dynode electrodes: Resolution: Clearly see the details of the dynode electrode structure. Spatial resolution of the system is less than 150 µm, for a step size of 25µm. Electronics chain used in this test: Final SLAC amplifier, LeCroy 4413 discriminators with 100mV threshold, LeCroy 3377 TDCs with 0.5ns/count. 11/26/04 Presented by J. Va'vra, RICH 2004 8

An example of the relative response along a line scan across 8 pads Hamamatsu Flat Panel H8500 PMT #2: Burle 85011-501 MCP-PMT #3: The Hamamatsu MaPMT uniformity is ~1:2.5 and the Burle MCP-PMT uniformity is ~1:1.5, in this example. Electronics chain used in this test: Final SLAC amplifier, LeCroy 4413 discriminators with 100mV threshold, LeCroy 3377 TDCs with 0.5ns/count. 11/26/04 Presented by J. Va'vra, RICH 2004 9

Burle MCP-PMT #8 relative detection efficiency (Normalized to the Photonis XP 2262B PMT) 635nm: At 635nm, which is close to the end of the Bialkali Q.E. range, the relative efficiency scaling to the Photonis PMT is not very reliable. 430nm: At 430nm, the relative efficiency is 50-60% relative to the Photonis PMT, if we include the late arrivals. This is approximately expected based on the MCP design (to be compared with the geometrical MCP collection efficiency (cathode-to-top MCP) of 60-65%, shown on page 6). Electronics chain used in this test: Final SLAC amplifier, LeCroy 4413 discriminators with 100mV threshold, LeCroy 3377 TDCs with 0.5ns/count Light source: PiLas laser diodes operating in the single photoelectron mode (635nm & 430nm). 11/26/04 Presented by J. Va'vra, RICH 2004 10

Burle MCP-PMT #3 Example #1 of a few scans Burle MCP-PMT #8 Burle MCP-PMT #4 Burle MCP-PMT #10 Typical relative efficiency is 50-60% of the 2 inch dia. Photonis XP 2262B PMT at 430nm. The efficiency drops to 30-50% around the edges at 430nm. 11/26/04 Presented by J. Va'vra, RICH 2004 11

Burle MCP-PMT #10 Example #2 of a few scans Burle MCP-PMT #14 Burle MCP-PMT #11 Burle MCP-PMT #15 Typical relative efficiency is 50-60% of the 2 inch dia. Photonis XP 2262B PMT at 430nm. The efficiency drops to 30-50% around the edges at 430nm. 11/26/04 Presented by J. Va'vra, RICH 2004 12

Example #3 of a few scans Burle MCP-PMT #16 Hamamatsu MaPMT #1 Hamamatsu MaPMT #2 Hamamatsu MaPMT #4 MCP-PMT #16 has large inefficiency around edges. However, this is still new experimental tube, and therefore it should not be taken as general observation (it has the MCP-to-cathode distance of 0.75 mm). Hamamatsu Flat Panel MaPMT relative efficiency is 50-70% of the Photonis XP 2262B PMT at 430nm. The efficiency drops to 30-50% around the edges at 430nm. 11/26/04 Presented by J. Va'vra, RICH 2004 13

Exercise in an estimate of the BaBar DIRC and Focusing DIRC prototype relative efficiencies BaBar DIRC: ETL PMT 9125B (Bialkali) N o : 31 cm -1 & N pe /ring : 28 @θ track = 90 o. BaBar DIRC: Focusing DIRC: Burle MCP-PMT (Bialkali) - new tube (see parameters of the new tube on page 6) N o : 29 cm -1 & N pe /ring : 27 @θ track = 90 o. Main Degradation factors: Mineral oil + MCP-PMT losses. Focusing DIRC prototype: The loss due to the mineral oil and the MCP-PMT must be compensated in other factors inherent in the BaBar DIRC, such as: - BaBar DIRC loses ~40% of photons at quartz/water interface due to mismatch of refraction indices. - Overall packing fraction loss in BaBar SOB is ~48%, due to sector gaps, light catchers, PMT packing. See next page for more details. 11/26/04 Presented by J. Va'vra, RICH 2004 14

Relative efficiency of the BaBar DIRC and the Focusing DIRC prototype Both designs have almost the same final efficiency, assuming that one has a new MCP - see page 6. Therefore, at least on a paper, the Focusing DIRC as described in this paper, is possible. However, one cannot tolerate any new losses! For example, any oil transmission deterioration, or a loss by a factor of two as per design such as the MCP85011-501 tube (see see page 6). Clearly, further improvements in detector efficiency is needed. Similarly, if one can make the final focusing element out of the solid fused silica, one would gain in efficiency. 11/26/04 Presented by J. Va'vra, RICH 2004 15

Conclusions We have studied spatial resolution of a number of MCP-PMTs and MaPMTs with a scanning setup with a spatial resolution of ~150µm. Many detailed microstructures are revealed, especially in case of Flat Panel MaPMTs. Typical relative measured efficiency at 430nm, normalized to the Photonis PMT, is typically 50-60% for the Burle MCP-PMTs, and 60-70% in good regions for the Hamamatsu Flat panel MaPMT. Both tubes have edge effects where the efficiency drops by ~10-20%. The performance of a detector based on the Focusing DIRC prototype, would be actually comparable to the present BaBar DIRC, provided that one would not create further inefficiencies. However, before the detectors mentioned in this talk can be considered for a real application, one has to do many things, such as (a) investigate aging issues, (b) develop more compact integrated electronics, etc. The Focusing DIRC prototype is in the test beam at SLAC. 11/26/04 Presented by J. Va'vra, RICH 2004 16