Research Article. The new Invisio DUR-D chip-on-the-tip digital flexible ureterorenoscopes: evaluation of changes in physical properties over time

Similar documents
TP /2017-E POWER LED 300. Luminous. Efficient. Durable.

Power. Flexibility. Usability.

How to Chose an Ideal High Definition Endoscopic Camera System

Litho. Taking care of people, our masterpieces. Surgical Laser System. Surgery

Illumination Challenges in Non- Industrial Vision Applications. Simon Stanley Managing Director ProPhotonix IRL Ltd

Just the Key Points, Please

KARL STORZ GI SILVER SCOPE Series

INSTRUCTIONS FOR AUTHORS

AUTHOR DECLARATION FORM

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

Guide to contributors. 1. Aims and Scope

Publishing India Group

INSTRUCTIONS FOR AUTHORS

Durata. Defibrillation Lead

Fiber-coupled light sources

DUR -Digital Invisio Flexible Ureteroscope/Choledochoscope

The Official Journal of ASPIRE Fertility & Reproduction. Instructions to Authors (offline submission)

Mainz COLOVIEW System

Images for life. Nexxis for video integration in the operating room

CHAPTER 8 CONCLUSION AND FUTURE SCOPE

How to Manage Color in Telemedicine

Barco Smart Laser - High performance cinema projection

HyPrecision Waterjet. The world's most profitable waterjets

VISERA 4K UHD GET CLOSER. GET CLOSER Four Times the Resolution of Full HD.

SURGICAL FIBER OPTIC HEADLIGHT

Tele-Medicine Focusing on the future. Image and data ar Seize the

Barco surgical displays. High-accuracy visualization solutions for surgery and endoscopy

BFW LIGHTING SOLUTIONS GUIDE. Hospitals are switching to BFW

LANGAUGE AND LITERATURE EUROPEAN LANDMARKS OF IDENTITY (ELI) GENERAL PRESENTATION OF ELI EDITORIAL POLICY

- CORRIGENDUM V

Just plug and go. Practical Features. Valuable Benefits

TP /2018-E TIPCAM 1 S 3D ORL. A Dimension Ahead

Information for authors

How to Talk with Your Doctor About Music During Surgery (or other medical or dental procedures)

Aqua Turf International, Inc.

INCREASE the. Performance Level. of Surgical Lighting. Luxtec MLX. Xenon Lightsource

Manuscript Submission Guidelines

Overview of All Pixel Circuits for Active Matrix Organic Light Emitting Diode (AMOLED)

How to write an article for a Journal? 1

Bahrain Medical Bulletin

Power Consumption Trends in Digital TVs produced since 2003

Operating room monitor Medical-grade 4K UHD monitor

Characterization and improvement of unpatterned wafer defect review on SEMs

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System

Vicon Valerus Performance Guide

DIFFERENTIAL CONDITIONAL CAPTURING FLIP-FLOP TECHNIQUE USED FOR LOW POWER CONSUMPTION IN CLOCKING SCHEME

Future Medicine Author Guidelines

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:

Mirth Solutions. Powering Healthcare Transformation.

*Please note that although this product has been approved in Japan, its launch in other countries has not yet been confirmed.

Bravo AV s Structured or Whole-House Wiring Approach

PROGRAMMING GUIDE. Accurate Targeting. Precise Control.

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

E3X-DA-S. Digital Fiber Sensors. Perfection Transcended! A Wealth of Advanced Functions for Easy and Reliable Application

Instructions to Authors

KARL STORZ SILVER SCOPE Series

MOST. Getting the. BMW Assist. Climate. Settings

OverView D. Barco DLP projection series

JPEN Author Guidelines

Guide for Authors. Before you begin

PlexBright LED Optogenetic Stimulation System

Introduction to Fibre Optics

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

DUB SkinScanner. in medicine and cosmetic. high frequency ultrasound since 1978

Instructions for Use: Video Inspection Scope with Display

An Introduction to Cochrane Collaboration and its Impact on Medical Practices

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

THE GLOBAL LEADER IN SCREENING TECHNOLOGY

VALHALLA 2. Reference Cables

Optical Perfection. HIGHYAG Beam Delivery Products

Msquare Innotech Solutions Pvt. Ltd. Complete integration of business solution. About Us: Mission:

Color Reproduction Complex

Understanding the True Cost of Cable Cuts

Downlight Guide Featuring premium LED downlights from the UK marketplace and the latest technology. Helping you to make an informed purchase.

Stretch More Out of Your Data Centre s Multimode Cabling System

Invisio ICN Flexible CystoNephroscope

Most advanced, portable, high-power 532nm Diode-Pumped Solid-State Photocoagulator

10YR. Advanced LED lighting. Ultra efficient, high performance LED battens. a company

Information for Authors

APPLICATION NOTE. Figure 1. Typical Wire-OR Configuration. 1 Publication Order Number: AN1650/D

Coronis 5MP Mammo. The standard of care for digital mammography

March, Solutions for Thermo-Ablation with RF medical technologies

Tipping the balancetotal Cost of Ownership

LIGHTS SO ADVANCED, THEY RE BRILLIANT EVEN BEFORE THEY RE TURNED ON.

Laser phosphor cinema projectors. Reduced total cost of ownership, superior image quality

Laying of cables and lines in electrical installations and data networks

PAPER SUBMISSION HUPE JOURNAL

What to consider when choosing a mammography display

Peer Review Process in Medical Journals

ALL NEW TRANSISTOR ELECTRONIC DATA PROCESSING SYSTEM

Linear Array with Intensity Adjustment

Figure 1. High Efficiency T8 LED Tube. Figure 3. Full View of Light High Efficiency T8 LED Tube

International Journal of Emerging Technologies in Computational and Applied Sciences (IJETCAS)

The Telegärtner Group

Astralite Minor Surgical and Examination Lights

Low Power VLSI CMOS Design An Image Processing Chip for RGB to HSI Conversion

INSTRUCTIONS TO THE AUTHORS FOR PUBLICATION IN BJ KINES-NATIONAL JOURNAL OF BASIC & APPLIED SCIENCE

Therapeutic Delivery Author guidelines

Vortex Mixers. Mini Vortex Mixers Heavy-Duty Vortex Mixers Microplate Vortex Mixers Multi-Tube Vortex Mixers. Asia Pacific

Lensed Fibers & Tapered Ends Description:

Transcription:

Research Article The new Invisio DUR-D chip-on-the-tip digital flexible ureterorenoscopes: evaluation of changes in physical properties over time Background & purpose: Flexible ureterorenoscopes (furss) have become an essential tool in endourological practice. In 2006, a new generation of digital furss (Invisio DUR-D, ACMI, USA) with new optical technology was introduced. We have compared both light intensity and deflection angles, with and without indwelling instruments, between identical scopes of different number of uses to evaluate changes in physical properties over time. Materials & methods: On three identical Invisio DUR-D digital flexible ureteroscopes of different usages we measured and compared the light intensity. Measurements were taken three times each with the ureteroscopes in straight position and maximum upwards and downwards deflection, respectively. The first scope was brand new; the second had been used 30 times while the third had been used 44 times. We also measured and compared the deflection angles without and with widely used accessory instruments inside the working channel in four identical scopes after 0, 10, 30 and 44 uses, respectively. Results: Light intensity for each scope showed slight differences in relation to deflection angles, but it remained stable after many uses. By contrast, the deflection angle is shown to deteriorate with instruments inside the working channel. This is relatively independent of the type of microinstrument used, but the decrease is directly proportional with advanced age of the scope in terms of uses. Conclusions: As an advantage over earlier generation scopes, light output remains constant due to the absence of the fragile low-resolution fiberoptics. Similarly to the earlier generation scopes, there remains a decrease in deflection capability over time with the new Invisio DUR-D digital furss. Keywords: deflection angle digital flexible ureteroscopes light intensity Tamer El-Husseiny 1, Junaid Masood 1, Since the initial description of passive diagnostic flexible ureteroscopy in 1960, the technology has advanced significantly [1]. Flexible ureterorenoscopy is standard for diagnosing and treating upper urinary tract diseases such as urolithiasis, transitional cell carcinoma and ureteral strictures [2]. This has expanded the usefulness and clinical applications of flexible ureteroscopes and thus a passive diagnostic tool has evolved to an active therapeutic instrument [3 7]. Over the last few years, technology has allowed miniaturization of flexible ureteroscopes, laser fibers and wire devices raising important questions regarding fragility, breakage, repair costs and proper utilization. Despite these new developments with increased deflection and smaller ureterorenoscope calibers, durability has not been compromised. Recent reports have even shown that the latest generation of flexible ureterorenoscopes (furss) were more durable than the previous generation [8,9]. The new Invisio DUR-D digital furss (Gyrus, ACMI, USA) was the first chip-onthe-tip flexible ureteroscope to be marketed. It was introduced into urological practice in late 2006. It was claimed that not only should this ureteroscope provide a superior quality image, but also make the instruments more durable in the whole. Physical properties of flexible ureteroscopes are known to change over time. Therefore, the aim of this study was to compare both light intensity and deflection angles, with and without indwelling instruments, between identical scopes of different number of uses. Materials & methods We measured some of the physical properties that are widely considered as good characteristics of durability of flexible ureteroscopes on identical Invisio DUR-D digital flexible ureteroscopes of different usages. The study was carried out in an in vitro setting, where the scopes were fixed on a straight workbench and the deflection angles of empty and loaded scopes were measured using a protractor. The light intensity measurements were taken at a given setting of the light source using an Ocean Optics optical spectrometer at a 20-mm fixed distance from the tip of the scope. To minimize measurement errors, all measurements were taken three times and mean values were calculated. Dimitrios Karamanolakis 2, Malcolm Birch 3 & Noor Buchholz 1 1 Endourology & Stone Services, Barts and The London NHS Trust, London, EC1A 7BE, UK 2 Kings College Hospital, London, UK 3 Medical Physics Department, Barts and The London NHS Trust, London, UK Author for correspondence: Tel.: +44 2078 2418 8706 Fax: +44 207 601 7844 d_tamer@hotmail.com 10.2217/THY.10.35 2010 Future Medicine Ltd Therapy (2010) 7(4), 423 427 ISSN 1475-0708 423

Research Article El-Husseiny, Masood, Karamanolakis, Birch & Buchholz A B Figure 1. (A) Deflection angle without guidewire and (B) deterioration in deflection angle after introduction of a 0.035 Guidewire (Cook Medical). Therefore, we measured and compared the deflection angles without and with instruments a 0.035 Guidewire (Cook Medical), a disposable 200 µm laser fiber and a 3F Zero-Tip Nitinol Basket (Boston Scientific ) inside the working channel of four identical different scopes after 0, 10, 30 and 44 uses, respectively, as the above instruments are some of the widely used instruments in routine endourological practice (Figure 1). In addition, we measured and compared the light intensity at a given setting of the light source. Measurements were taken three times each with the ureteroscopes in straight position and upwards/downwards active deflection, respectively. The first scope was brand new; the Figure 2. The ureteroscope in deflection, placed at a distance of 20 mm from an Ocean Optics optical spectrometer. second had been used 30 times while the third had been used 44 times (Figure 2). Results Statistical analysis was carried out on the results using Analyse-it statistical software program (Version 2.21) and the appropriate statistical test was used according to the type of data. The level of statistical significance was defined as p < 0.05. Light intensity output for each scope showed a slight difference in relation to deflection angles, which was within the experimental error and was not statistically significant (Table 1). On the other hand, all scopes showed more deflection angle with an empty working channel, while the deflection angle is shown to deteriorate with instruments inside the working channel. Results from this study showed that this is relatively independent of the type of microinstrument used, but the decrease is directly proportional with advanced age of the scope. This deterioration was statistically significant in both upward deflection (p = <0.0001) and downward deflection (p = 0.0100) (Table 2). Expert commentary Flexible ureteroscopy has become the standard and most frequent upper tract endourological procedure for the diagnosis and the treatment of numerous urinary tract conditions such as stones, transitional cell carcinoma, ureteric strictures and extrinsic compression. The fact that the flexible scopes can virtually reach every part of the collecting system ensures that patients usually have a successful diagnostic or therapeutic procedure. As it is a minimally invasive procedure, patients can usually be discharged the same day and hence the technique has gained popularity with surgeons and patients. Over the last decade several innovations have improved the performance of flexible ureteroscopes; for example their miniaturization, improved manoeuvrability and optimization of the accessory instrumentation. The ongoing attempts to improve the optical characteristics of the furss resulted in a significant advancement when in 2006 the first digital furs was introduced. The Invisio DUR-D was the first commercial digital flexible ureteroscope that dramatically improved image quality and is considered the current gold standard image quality [10]. The Invisio DUR-D digital furs integrates the endoscope, digital camera and light source in a simple plug-and-play device. It has a 3.6F 424 Therapy (2010) 7(4)

The new Invisio DUR-D chip-on-the-tip digital flexible ureterorenoscopes Research Article Table 1. Light intensity in different deflection (unit: Lux 10 3 ) Uses Straight Upwards Downwards 0 5.5 4.36 4.84 30 6.19 6.30 6.25 44 5.05 5.25 5.15 working channel and is capable of dual 250 of deflection. The shaft s diameter is 9.3F, while the distal tip s diameter is 8.7F; the distal tip houses dual light-emitting diode (LED)-driven light carriers. At the tip, there is also a tiny 1-mm digital camera chip-on-the-tip (Figure 3), which eliminates the need for fragile, low-resolution fiberoptics. These two technological advances reduce the weight of the ureteroscope and eliminate the risk of fires since there are no external cameras or light source [11]. The Invisio DUR-D weighs 505 g compared with the conventional fiberoptic ureteroscope (e.g., DUR-8 Elite) with external camera and light cord that weighs 1012 g. This represents a 50% reduction in weight [11]. Deflection angle Most current new-generation ureteroscopes allow active deflection up to 250. The Invisio DUR-D was the first to offer dual active primary and passive secondary tip deflection that totals 250, offering access to almost every point in the collecting system. The loss of deflection capacity once a tool is used in the working channel is a critical point that might limit the surgeon s ability to reach certain points within the collecting system. Our results show that there is no significant difference in loss of deflection between the use of guidewire, laser fiber or stone basket; there is no significant difference between the upwards or downwards deflection of the scope but the deflection angles are directly proportional to the number of uses of the scope. Hence, in this regard, these new digital ureteroscopes have no real advantage over the standard fiberoptic ureteroscopes as both lose some deflection with increased use. This loss appears proportional to the number of uses of the ureteroscope. In our opinion, combining primary active deflection and strategic passive secondary deflection provides unrestricted access to virtually all calyces in the vast majority of patients. The added cost and complexity of these dual-deflection ureteroscopes might not justify their routine inclusion in the ureteroscopic armamentarium in all urological institutions as yet, although they will play an important role in the often complex case mix found in tertiary referral centers such as our unit. The experienced endourologist can often overcome the limitations of the older potentially inferior fiberoptic ureteroscopes and hence the added cost of these new digital flexible instruments may not be routinely justified. Light intensity The Invisio DUR-D digital furs provides excellent image quality, zoom capability and digital image enhancement, thus achieving a remarkable improvement in comparison with fiberoptic technology. The instrument requires no separate light source; illumination is provided by two LEDs instead of the conventional fiberoptics. This LED light lasts up to 10,000 h, which is 10 20-times longer than the expensive xenon lights [11]. Consequently, there is a decrease in weight and fragility of the endoscope, materialized in the lower repair costs and increased lifespan [12]. Results from this study have shown that the light intensity output from the Invisio DUR-D furss remains constant over time, which can be attributed to the elimination of use of the Table 2. Deflection angles with different instruments. Uses No instruments 0.035 guidewire 200-µm laser fiber 3F nitinol basket 0 upwards 218º 196º 205º 204º 10 upwards 216º 194º 203º 203º 30 upwards 197º 173º 174º 174º 44 upwards 163º 148º 155º 152º 0 downwards 216º 194º 209º 191º 10 downwards 199º 191º 192º 190º 30 downwards 230º 139º 151º 147º 44 downwards 170º 123º 141º 158º www.futuremedicine.com 425

Research Article El-Husseiny, Masood, Karamanolakis, Birch & Buchholz and depending on older fiberoptic technology. We believe that the ideal future flexible ureteroscope should incorporate chip-on-the-tip technology, dual channels and, above all, have a smaller diameter. Figure 3. The tip of the Invisio DUR-D houses a dual light-emitting diode, a 12 o clock digital camera and a 6 o clock 3.6F working channel. fragile low-resolution fiberoptics and the digital enhancement achieved by the instrument. This constancy in one of the most important physical properties of flexible ureteroscopes is a great advantage over earlier generation scopes. Future perspective Flexible ureteroscopes have evolved from purely diagnostic tools to key instruments in the management of stone disease and other pathology of the upper urinary tract. Despite these advances, problems related to visibility and manoeuvrability can hinder the goals of surgical intervention. The ongoing efforts and innovations have resulted in the first digital furs, the Invisio DUR-D. The Cobra Dual-Channel Flexible Ureteroscope has dual channels, which allow for better flow of irrigation while simultaneously using microinstruments, but this is at the expense of the scope being larger in diameter Conclusion The Invisio DUR-D furs has an advantage over earlier generation scopes in that the light output remains constant. However, as with earlier generation scopes, there remains a decrease in deflection capability over time. Only a few studies have been conducted up to now relating to the overall durability and manoeuvrability of this instrument after use for clinical purposes, which is probably due to its recent introduction to the market. This study was carried out in a cross-sectional manner, where measurements and calculations were performed on identical scopes of different number of uses at a certain point of time. Although performing this study in a longitudinal-sectional manner might be more bias free, this was not applicable at our unit owing to instrumental availabilities. Financial & competing interests disclosure The authors have no relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript. This includes employment, consultancies, honoraria, stock ownership or options, expert testimony, grants or patents received or pending, or royalties. No writing assistance was utilized in the production of this manuscript. Executive summary Flexible ureterorenoscopy is now the standard for diagnosing and treating upper urinary tract diseases such as urolithiasis, transitional cell carcinoma and ureteral strictures. The new Invisio DUR-D digital flexible ureterorenoscopes (Gyrus, ACMI, USA) was the first chip-on-the-tip flexible ureterorenoscope to be marketed in late 2006. We compared both light intensity and deflection angles, with and without indwelling instruments, between identical scopes of different number of uses to evaluate changes in physical properties over time. As an advantage over earlier generation scopes, light output remains constant. Similarly to the earlier generation scopes, there remains a decrease in deflection capability over time. Bibliography Papers of special note have been highlighted as: of interest of considerable interest 1 Marshall V: Fibre optics in urology. J. Urol. 91, 110 114 (1964). 2 Basillote JB, Lee DI, Eichel L, Clayman RV: Ureteroscopes: flexible, rigid, and semirigid. Urol. Clin. North Am. 31, 21 32 (2004). 3 Afane JS, Olweny EO, Bercowsky E et al.: Flexible ureteroscopes: a single center evaluation of the durability and function of the new endoscopes smaller than 9Fr. J. Urol. 164, 1164 1168 (2000). Well designed study, systematically evaluating the functional durability of flexible ureterorenoscopes (furss) smaller than 9F in clinical setting. 4 Grasso M, Bagley D: Small diameter, actively deflectable, flexible ureteropyeloscopy. J. Urol. 160, 1648 1654 (1998). A study evaluating the treatment outcome of the small diameter, actively deflectable furs in clinical setting. 5 Elashry OM, Elbahnasy AM, Rao GS, Nakada SY, Clayman RV: Flexible ureteroscopy: Washington University experience with the 9.3F and 7.5F 426 Therapy (2010) 7(4)

The new Invisio DUR-D chip-on-the-tip digital flexible ureterorenoscopes Research Article flexible ureteroscopes. J. Urol. 157, 2074 2080 (1997). A study evaluating roles and indications of smaller caliber instruments in practice. 6 Segura JW, Preminger GM, Assimos DG et al.: Ureteral stones Clinical Guidelines Panel summary report on the management of ureteral calculi. The American Urological Association. J. Urol. 158, 1915 1921 (1997). One of the earliest fundamental American Urological Association guidelines on management of ureteral stones. 7 Monga M, Beeman WW: Advanced intrarenal ureteroscopic procedures. Urol. Clin. North Am. 31, 129 135 (2004). Interesting, well designed study evaluating the strength of the shaft and deflection mechanism of several furss. 8 Afane JS, Olweny EO, Bercowsky E et al.: Flexible ureteroscopes: a single center evaluation of the durability and function of the new endoscopes smaller than 9Fr. J. Urol. 164, 1164 1168 (2000). 9 Traxer O, Dubosq F, Jamali K, Gattegno B, Thibault P: New generation flexible ureterorenoscopes are more durable than previous ones. Urology 68, 276 281 (2006). A study evaluating the deterioration in the maximal active deflection and flow capabilities of furss in clinical setting. 10 Traxer O, Thibault F, Beley S, Sebe P, Gattegno B: Evaluation of optical properties of the new digital flexible URS by comparison with 3 standard flexible ureterorenoscopes. Presented at: the 22th European Association of Urology Congress. Berlin, Germany, 21 24 March 2007 (Abstract 983). 11 Andonian S, Okeke Z, Smith A: New technologies in endourology digital ureteroscopy: the next step. J. Endourol. 22, 603 605 (2008). One of the earliest publications to highlight the physical and optical properties of the latest generation of digital furss and their future role in endourology. 12 Mitchell S, Havranek E, Patel A: First digital flexible ureterorenoscope: initial experience. J. Endourol. 22, 47 49 (2008). A study evaluating the image quality of the first-generation digital furss. www.futuremedicine.com 427