STANDARD OPERATING PROCEDURE:

Similar documents
TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer

Liquid Chromatography- Mass Spectrometer Manual

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications

UWPR Advancing Proteomics

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

AIM INTRODUCTION SIMPLIFIED WORKFLOW

Performance. Reliability. Productivity. Automated Flash Chromatography Systems

Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

MestReNova Manual for Chem 201/202. October, 2015.

OZONESONDE FLIGHT PREPARATIONS

Processing data with Mestrelab Mnova

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

Guide to the FACS Calibur

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

FEI FIB Focused Ion Beam

CytoFLEX Flow Cytometer Quick Start Guide

ChromQuest 5.0 Chromatography Data System

Use of the cytometry platform - Rules of Procedure. Use of the cytometry platform University of Bordeaux - CNRS UMS INSERM US 005

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

VAD Mobile Wireless. OBD-II User's Manual Version 1.0

Waters Synapt High Definition Mass Spectrometry System Quick Start Guide

ICS-3000 System Installation Qualification

Control Unit CU 2.1 User Manual V6850A HPLC

Magnum GCMS Training and Operating Instructions 02/07/2012 S.V.

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

Click Here To Start Demo

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Instrument Control and Data Acquisition with Chromeleon

Unicable II Programmer. IDLU-PROG01-OOOOO-OPP Item: Installation & User Guide

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

EDL8 Race Dash Manual Engine Management Systems

If you have questions or problems in providing anything described as Customer

PulseCounter Neutron & Gamma Spectrometry Software Manual

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

UNITY 2 TM. Version 1.0. February 2008

Operating Instructions

MSQ Plus Mass Detector. Getting Started Guide

Approved by: / / R. Battaglia 12/16/2016

PAL LSI - Robust and Reliable Sample Handling

Achieving More Efficient Data Review with OpenLAB CDS

University of Minnesota Minnesota Nano Center Standard Operating Procedure

ECT 224: Digital Computer Fundamentals Digital Circuit Simulation & Timing Analysis

PAL LSI - Robust and Reliable Sample Handling

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

University of Minnesota Nano Fabrication Center Standard Operating Procedure

PLASMA MONITOR (PT20 UVVis) USER GUIDE

Operation Procedure for Phillips XL30 ESEM

TE-4HCM-S User Manual. l ::: J TECH SUPPORT. MetraDealer.com

Maintenance Service Bulletin

Logic Controls LV3000 VGA Connected Virtual Pole Display

Physics 105. Spring Handbook of Instructions. M.J. Madsen Wabash College, Crawfordsville, Indiana

Florida State University Thayumanasamy Somasundaram

Gordian. Multifunctional Power Distributor / Conditioner. v1.3

CHEMISTRY SEMESTER ONE

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

Solutions to Embedded System Design Challenges Part II

Integration of the Zetasizer µv and OmniFACE into an SEC system

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below.

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

LSR II Users Guide Stanford Shared FACS Facility

PHI 5000 VersaProbe TM Operator s Guide

Users Manual. Stroboscope. September 2014

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5)

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

GC/LC-MS: data acquisition rate and peak reconstruction

Operating Instructions

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)

Amazon Account 3/14 TM 1

Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual

NMR. picospin. Maintenance Guide

KNX Dimmer RGBW - User Manual

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

AU5800 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU5800 Weekly (Analyzer and ISE) Maintenance

Select source Click the Bio Settings button to modify device settings. Select Fingers Use Ctrl+Left mouse button to select multiple fingers to scan.

Simulation Lab MediaSite Training Workshop

013-RD

Matrox PowerStream Plus

Procedures for conducting User QA on the scanner

Outdoor & Indoor Full Color LED Display Screen. Installation Manual

LM-TV09-4K2K. 4K video wall controller. User. Manual

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

Ultra 4K Tool Box. Version Release Note

VISSIM Tutorial. Starting VISSIM and Opening a File CE 474 8/31/06

ForwardT Plugins. AutoDetect (SCTE-35) Automatic Ad insertion using SCTE-35 cue messages. Revision as of. October 28, User s Guide.

Lab 2: A/D, D/A, and Sampling Theorem

MonitorKey Operation Manual: content/uploads/ MonitorKey Operation Manual.pdf

MATE3 Owner s Manual Addendum

ENGR 1000, Introduction to Engineering Design

My XDS Receiver- Affiliate Scheduler

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

Transcription:

Q-EXACTIVE STANDARD OPERATING PROCEDURE: Purpose of this Instrument: Essential tool for molecular weight identification for small molecules, peptides, proteins, ect. Location: WVU Chemistry Research Laboratory Building Room: 381 Primary Staff Contact: Dr. Huiyuan Li (304) 293-0747 The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. 1) INITIAL CHECK This text is a guide to operating the Q-exactive mass spectroscopy. A more detailed and thorough account of these procedures can be found in the Q-exactive Operating Manual. 1. Log in your Mass Spec Q-exactive session on the CORES. Q-exactive mass spec WVU SRF 1

2. Check log sheet to see the notes of previous users. This will inform you about anything unusual that occurred before, and solvents were used before. 3. Write down your name, date, start/end time, sample type, and ion source information on the log sheet. 4. Check the nitrogen generator reading (It is located on the left of the computer bench). It should be around 100 psi (see generator figure below). If the service light is on, and the reading is lower or higher, do not continue and contact SRF personnel immediately. Do not turn off the nitrogen generator. Keep it on all the time. Figure 1. Nitrogen generator. Keep it on all the time. 5. Check the Q-exactive instrument status. The first 3 indication lights should be always green. If any of them becomes yellow, red, or not on, please contact SRF personnel. The fourth one is system, and yellow light means the instrument is in STANDBY. The fifth one is scanning, which is only on when scanning. Figure 2. Q-exactive mass spec. Normal stauts on standby. 6. Go to computer. Double click the Tune. Check the mass in the standby or off condition, and all the green lights are on. If any light shown yellow or red for the first two lights, do not continue and contact SRF personnel immediately. If the third one is yellow, please continue on calibration. Q-exactive mass spec WVU SRF 2

Figure 3. Tune software interface. Normal stauts on standby. The third yellow lights is on suggesting performance issue (calibration is needed usually) 7. Check the connected ion source on Mass Spec if it is the correct ion source for your experiment. We have HESI and nanospray sources. If not, contact the SRF personnel to switch the ion source for your experiment. Figure 4. HESI ion source and Nanospray ion source. 8. If you are going to use LC column, check if there is enough mobile phase on the back of LC. Contact the SRF personnel to prepare the mobile phase or prepare by yourself if you have been trained. The common solvent A is 0.1% formic acid in H2O and solvent B is 0.1% formic acid in ACN. Check literature for your solvent system. Please use the LC/MS grade solvent to prepare the mobile phase. If you are going to use other types of mobile phase, please contact the SRF personnel for permission. Q-exactive mass spec WVU SRF 3

Figure 5. Check mobile phase solvent level at the back of the autosampler. 2) START-UP (CALIBRATION) Figure 6. Check solvent level for solvent 1 and 2 for needle wash. 1. Make sure the HESI ion source is connected and configuration is set up for HESI ion source. Q-exactive mass spec WVU SRF 4

2. Go to Instrument configuration. Check the components that are connected. Choose the component Add >> to configured device side if needed. Then highlight one by one to configure, then click Done. Figure 7. Instrument configuration window. Configure the component when you make any changes of the instrument. 3. Open Tune file on the desktop, or go to Start> all programs > Thermo Exactive series> Tune. 4. Turn the instrument on from the top left corner. If any of the light showing yellow or red. Put mouse on the top of the light, and read the error message. If the message is not Mass Spec Calibration, then contact SRF staff for assistance. If the message is Mass Spec Calibration, then go ahead and do the calibration. Figure 8. Tune window. The tune is for direct Mass Spec view and control. Q-exactive mass spec WVU SRF 5

5. Turn on the LC Accela 1250 pump direct control. 6. Check the take pump under control. Input flow rate 100 ul/min, inlet C (usually MeOH). Flow solvent C for about 10 min. 7. Prepare for the calibration: Figure 9. LC direct control window. (1) Find the calibration solution in the Chemical storage area cabinet. Figure 10. Calibration solution. Find both positive and negative calibration solutions at the same cabinet. (2) Find the direct injection tubing and syringe Q-exactive mass spec WVU SRF 6

Figure 11. Direct injection line and syringe. Find in the top draw under LC. (3) Stop the LC injection, Release direct control and Standby the mass spec. 8. Remove the LC connection. 9. Connect the direct injection line ( calibration only ) to mass spec. 10. Get 100 ul positive calibration solution with direct injection syringe ( calibration only ). Connect the syringe to the end the direct injection tubing. And then place the syringe on the syringe pump. Figure 12. Connect direct injection line to the ion source. Q-exactive mass spec WVU SRF 7

11. Put the mass spec in the on position from Tune. 12. Load calibration tune file at C:> xcailbur> methods> huiyuan methods> weekly calibration> positive calibration 13. Turn on the syringe control. Input flow rate 5 ul/min. 14. Wait for the expecting signals: 195, 262, 524, 1221.991, 1321.985, 1421.978, 1521.972, 1621.966, and etc. And the TIC variation should be less than 12%. 15. In the left column. Find Tune. Choose TIC. Click Tune. Then the system will tune automatically. Figure 13. Signal tuning window. 16. In the left column. Find Calibrate. Check the MS mass calibration (pos). Then click Calibrate. Choose MS mass calibration (pos). The system will calibrate automatically. After this step is done, a window shows successful!. MS mass calibration (pos) is needed every 4 days. For comprehensive calibration, including Baseline calibration, Isolation Mass and Res (pos), and MS mass calibration (pos), it could be done every month. Figure 14. Signal calibration window. Q-exactive mass spec WVU SRF 8

17. Stop the direct injection. Standby the instrument. 18. Wash the syringe and direct injection line with MeOH. Make sure it is clean. 19. Inject 400 ul MeOH to system to wash the signal out of the system. 20. Stop the injection. Standby the instrument. 21. Do the negative calibration with the same procedure with Negative calibration solution and Tune file. 22. Disconnect the injection line. Put the syringe into the box and direct injection line into the corresponding bag. IMPORTANT: Standby the instrument when you change the inlet or direct injection line. IMPORTANT: Dispose any HPLC chemical waste solvent into the jar located in the chemical fume hood with label HPLC chemical waste. Do not overfill the bottle. If the bottle is near full, please contact BNRF staff to dispose the bottle and get a new one for your chemical waste. 3) SET UP METHOD FOR YOUR EXPERIMENT 1. Use a regular LC/MS/MS experiment as an example. Configure all necessary parts, including orbitrap mass, pump, autosampler, and PDA. Figure 15. Configuration window. Q-exactive mass spec WVU SRF 9

2. Double click Xcalibur on the desktop. 3. Figure 16. xcalibure interface. 4. Click on Instrument setup. Figure 17. Instrument set up interface. Q-exactive mass spec WVU SRF 10

5. Set up all parameter as needed. For example: PDA (3 channels could be monitored at the same time. Choose the UV absorption wavelength that are best for your sample) Accela open AS (could choose from the Template dropdown window: standard injection accela open) Accela pump (set up a gradient run. See figure 18 as an example. You need adjust the starting solvent condition, and gradient run time/final B% of the gradient based on the polarity and complexity of your sample.) Orbitrap MS (a. Global lists: Use when you need specify the signal that to be tested. b. Tune file. This is mass parameter file when you initially tune your sample. c. Scan groups. Full-ms-SIM, when you want to see the entire components in the sample. Only MS will be detected, not MS/MS. AIF: All ion fragmentation Full MS/AIF: Detect in two modes, full mass and all ion fragmentation. Full MS/dd-MS2 (TopN): Detect in two modes, full mass and data dependent MS2. Targeted-SIM: Use it when you have certain targets, list accurate masses of these targets in Global list in inclusion list. Only the targeted/selected ions will be detected. ) Figure 18. An example of HPLC gradient set up Q-exactive mass spec WVU SRF 11

Drag the proper scan groups method into the bar area. Then change the parameter at the right panel, such as run time which should match your LC method, select polarity, scan range, and resolution. Figure 19. Mass spec scanning method set up 6. After the parameters are set up, save the file in the folder C:/xcalibur/methods/YOUR NAME FOLDER. Note: Build your folder under C:/xcalibur/methods and C:/xcalibur/data at the beginning and only save your method and data in your own folders. If you change the file location later, the instrument may not recognize your file. Do not move the location of your files once saved. 7. Go to sequence set up. Q-exactive mass spec WVU SRF 12

Figure 20. Main xcalibur page. Go to sequence setup from here. 8. Build your data folder under C:/xcalibur/data/YOUR NAME FOLDER. Choose unknown sample type, then modify the sample name, where you want to save the data, the instrument method that you just set up, position of your sample from autosampler, and injection volume. 9. Click on the to RUN a sample. Then the sample run will start immediately. You will see the autosampler is picking up sample, inject, and you will see the real time monitor window. 10. If you run several runs in turn, you click to run sequence. Q-exactive mass spec WVU SRF 13

Figure 21. Set up sequence run. List all your sample here. TURN OFF PROCEDURES 1. It is very important that the user never turn off the instrument. Only STANDBY is needed. 2. After sample run is end, make sure the solvent flow is off. Please close the xclibur. Go to Tune file, if the instrument is still on, put the instrument in Standby from the left top corner. Then turn off the TUNE file. 3. Log off the computer. 4. If the PDA is on, turn off the PDA physically from the bottom of PDA. Figure 22. Power off the PDA when the experiment is done. Q-exactive mass spec WVU SRF 14

EMERGENCY OPERATING PROCEDURES If you have any questions, even if you are just slightly unsure, ASK someone who knows and can help. There are no penalties for asking for help but there may be for not reporting damage to the equipment that may delay or prevent others from working. If, at any time, you need to contact someone for help, call or locate the following staff of the Shared Research Facility (SRF): Huiyuan Li Office: CRL 381 Phone: (304) 293-0747 Emergency situation: 1) Instrument failure. Such as POWER light is not on. Wait until data acquisition, if any, is complete. Turn off the flow of sample solution from the LC or direct injection. In the Q-exactive Tune software window, click on the ON/STANDBY button to put the instrument in OFF position. All high voltage are shut off, as are the sheath and auxiliary gas. Place the electrics service switch in the Service Mode position. Put the Mainpower circuit breaker switch of the mass spectrometer in the OFF position. Q-exactive mass spec WVU SRF 15

2) Power outage During your experiment, there is a power outage. No worries. The instrument was protected by uninterrupted power supply (UPS). The short time power break will not affect your sample run. The computer itself is not connected to UPS. You could restart computer after the power outage. If the power outage is longer than the UPS support, then you need turn off the instrument by following steps: Turn off the PDA and PUMP physically from the bottom. Turn off the instrument physically (Electronics and Mainpower) from the left side panel of the Mass Spec. Note on the log book about the emergency situation and your contact information. Figure 23. Mainpower and electronics. Only turn them off during the service or emergency situation. COMPLETE MASS SPEC RESTART PROCEDURE If the power outage happened during night, when you come to the lab and the instrument is not in work condition. Do not start your experiment. Please contact SRF staff to restart the instrument. If you already have been trained on restarting instrument, please follow the exact procedure to restart instrument. Turn on the computer. If the Mainpower and Electronics are off, turn Mainpower and Electronics on in order. Open Tune software. Turn the instrument on from the top left corner of Tune window. From the left column, find Vacuum/bakeout bar, and click Bakeout. It will take about 14 hours to complete the process. Then you could start your experiment. If a dangerous situation is evident (smoke, fire, sparks, etc), ONLY if it is safe to do so, the user should switch off the Electronics and Mainpower to turn OFF the entire Mass spectrometer and switch off the Pump to stop the solvent flow, and notify the proper emergency personnel. In any Q-exactive mass spec WVU SRF 16

case, the user should leave the facility and contact emergency personnel as soon as possible from a safe place. Huiyuan Li Office: CRL 381 Phone: (304) 293-0747 Q-exactive mass spec WVU SRF 17