Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel)

Similar documents
Digital Delay/Pulse Generator

Digital Delay/Pulse Generator

Digital Delay/Pulse Generator

2 MHz Lock-In Amplifier

Synthesized Clock Generator

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

GFT Channel Slave Generator

GFT Channel Digital Delay Generator

Agilent 5345A Universal Counter, 500 MHz

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display

2 MHz Lock-In Amplifier

Data Pattern Generator


Advanced Test Equipment Rentals ATEC (2832)

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System

4 MHz Lock-In Amplifier

GFT channel Time Interval Meter

WAVEJET 300 SERIES OSCILLOSCOPES. Unmatched Performance, Portability, and Value

WAVEJET 300 SERIES OSCILLOSCOPES. New Cover to Come. Unmatched Performance, Portability, and Value

PM 6680B / PM 6681 / PM 6681R

Precision testing methods of Event Timer A032-ET

Sources of Error in Time Interval Measurements

Advanced Test Equipment Rentals ATEC (2832)

Data Pattern Generator DG2020A Data Sheet

Portable Performance for Debug and Validation

Oscilloscope Guide Tektronix TDS3034B & TDS3052B

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Expect to Make Waves.

SELECTION GUIDE Series of RF and Universal Frequency Counter/Timers

PicoScope 6407 Digitizer

Data Pattern Generator

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

Frequency. Range. HP 83751A/B: 2 GHz to 20 GHz HP 83752A/B: 10 MHz to 20 GHz. Standard 10 MHz timebase: f10 ppm

Digital Storage Oscilloscopes 2550 Series

PicoScope 6407 Digitizer

8650A SERIES UNIVERSAL POWER METERS

Datasheet SHF A

Agilent Technologies 54522A

Microwave Counter, Power Meter and DVM in One Portable Package

RIGOL. Data Sheet. DS1000B Series Digital Oscilloscopes DS1074B, DS1104B, DS1204B. Product Overview. Easy to Use Design. Applications.

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor

WaveSurfer MXs. The Specs You Want The Features You Need 200 MHz to 1 GHz

Full-featured CW Microwave Counters for Field, Factory or Lab

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix

Fluke PM6690. Frequency Counter/Timer/Analyzer

Signal Conditioners. Highlights. Battery powered. Line powered. Multi-purpose. Modular-style. Multi-channel. Charge & impedance converters

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

CDMA2000 1xRTT / 1xEV-DO Measurement of time relationship between CDMA RF signal and PP2S clock

FREQUENCY COUNTERS TO 18 GHZ USING THE DATUM FREQUENCY STANDARD

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

4.9 BEAM BLANKING AND PULSING OPTIONS

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

DIGITAL TRANSMISSION MEASURING INSTRUMENTS

MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer

Bluetooth Tester CBT. Specifications. Specifications. Version January 2006

SMART Trigger modes like Glitch, Window and Dropout allow you to capture precisely the events of interest.

2016 RIGOL TECHNOLOGIES, INC.

!Ill ~ 168. Model490 Dual Input, Dual Trace Automatic Peak Power Meter

Assembly Level Service Guide

MP212 Principles of Audio Technology II

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

Advanced Test Equipment Rentals ATEC (2832)

Analog Storage Oscilloscope TS-81000/ Newly developed CCD(Charge-coupled device) scan converter tube

Technical Datasheet. 8650B Series Universal Power Meters 10 MHz to 50 GHz. 8651B Single Channel 8652B Dual Channel Rev.

Experiment # 4 Counters and Logic Analyzer

Microwave/Counter/Analyzer & Integrated Power Meter

High Speed Data Acquisition Cards

Analog Discovery Scope and Waveform Generator Edited 11/15/2016 by Eric Scotti & DGH

ScopeMeter 190 Series Specifications

PROGRAMMABLE DC SOURCE VIEW RECORDERS Programmable DC Source FEATURES

40 Gb/s PatternPro Programmable Pattern Generator PPG4001 Datasheet

Signal Stability Analyser

R&S RT-Zxx High-Bandwidth Probes Specifications

RX40_V1_0 Measurement Report F.Faccio

PULSE & DELAY GENERATOR. Output Amplitude HIGH SPEED DIGITIZER. BW / max. Sample Rate STREAK CAMERA

Introduction to the oscilloscope and digital data acquisition

R3267/3273 Spectrum Analyzers

Generation of Novel Waveforms Using PSPL Pulse Generators

WAVECREST Corporation DTS User s Guide and Reference Manual REV B

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

Microwave/Counter/Analyzer with Integrated Power Meter MCA3000 Series Datasheet

Digital Sampling Oscilloscope

EM1. Transmissive Optical Encoder Module Page 1 of 9. Description. Features

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I

DIGITAL OSCILLOSCOPES & DIGITAL SCOPES

DIGITAL ELECTRONICS: LOGIC AND CLOCKS

ATS MS/s 8-Bit PCI Digitizer

Agilent 8560 E-Series Spectrum Analyzers

DA : Series of Narrowband or Wideband Distribution Amplifiers

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A

Agilent E4887A HDMI TMDS Signal Generator Platform

WAVEEXPERT SERIES OSCILLOSCOPES WE 9000 NRO 9000 SDA 100G. The World s Fastest Oscilloscope

Advanced Test Equipment Rentals ATEC (2832)

TSG 90 PATHFINDER NTSC Signal Generator

V6118 EM MICROELECTRONIC - MARIN SA. 2, 4 and 8 Mutiplex LCD Driver

WAVESURFER XS SERIES. The Everyday Oscilloscope that s Easy to Use

Field Programmable Gate Array (FPGA) Based Trigger System for the Klystron Department. Darius Gray

EM1. Transmissive Optical Encoder Module Page 1 of 8. Description. Features

Transcription:

Digital Delay / Pulse Generator Digital delay and pulse generator (4-channel) Digital Delay/Pulse Generator Four independent delay channels Two fully defined pulse channels 5 ps delay resolution 50 ps rms jitter Adjustable amplitude and offset Delays up to 1000 seconds 1 MHz maximum trigger rate Standard GPIB interface Optional ±32 V outputs The Digital Delay/Pulse Generator provides four precisely timed logic transitions or two independent pulse outputs. The delay resolution on all channels is 5 ps and the channel-to-channel jitter is less than 50 ps. Front-panel BNC outputs deliver TTL, ECL, NIM or variable level ( 3 to +4 V) pulses into 50 Ω or high impedance loads. The high accuracy, low jitter and wide delay range make the ideal for laser timing systems, automated testing and precision pulse applications. Delay Outputs There are four delay output channels: A, B, C and D. The logic transitions of these outputs can be delayed from an internal or external trigger by up to 1000 seconds in 5 ps increments. The T0 pulse, which marks the beginning of a timing cycle, is generated by the trigger signal. The insertion delay between an external trigger and the T0 pulse is about 85 ns. Delays for each channel may be "linked" to T0 or any of the other delay channels. For instance, you can specify the delays of the four channels as: A = T0 +.00125000 B = A +.00000005 C = T0 +.10000000 D = C +.00100000

Digital Delay/Pulse Generator In this case, when the A delay is changed, the B output will move with it. This is useful, for instance, when A and B specify a pulse and you want the pulse width to remain constant as the delay of the pulse is changed. Regardless of how the delay is specified, each delay output will stay asserted until 800 ns after all delays have timed out. The delays will then become unasserted and the unit will be ready to begin a new timing cycle. Pulse Outputs In addition to the four delay outputs there are four pulse output channels: AB, AB, CD and CD. The leading edge of the AB pulse coincides with the leading edge of A, and the falling edge of the AB pulse coincides with the leading edge of B. For instance, in the previous example, a 50 ns pulse would appear at the AB output and a 1 ms pulse at CD. Pulses as short as 4 ns (FWHM) can be generated in this manner. The complementary outputs ( AB and CD) provide a pulse with identical timing and inverted amplitude. Output Amplitude Control Each delay and pulse output has an independently adjustable offset and amplitude which can be set between 3 V and +4 V with 10 mv resolution. The maximum transition for each output is limited to 4 V. In addition, you can also separately select 50 Ω or high impedance termination for each output. For convenience, preset levels corresponding to standard logic families can also be selected. TTL, NIM and ECL levels can each be set with a single key press. Triggering The can be triggered internally from 1 mhz to 1 MHz with 4-digit frequency resolution. External, single-shot and burst mode triggers are also supported. For power control applications, the can be synchronized to the AC line. An optional trigger inhibit input allows you to enable or disable triggering with a TTL level input signal. ±32 Volt Outputs For applications requiring higher voltages, a rear-panel high voltage (±32 V) option is available. This option provides five rear-panel BNCs which output amplified, 1 µs pulses at the transition times of the front-panel T0, A, B, C and D outputs. The high voltage option does not affect the function or the timing of the front-panel outputs. The amplitude of the rearpanel outputs is approximately 8 the corresponding frontpanel output, and the outputs are designed to drive 50 Ω loads. Since they can only drive an average current of 0.8 ma, charging and discharging the cable capacitance may be the most important current limiting factor to consider when A timing cycle is initiated by an internal or external trigger. T0 is asserted approx. 85 ns after an external trigger. Outputs A, B, C and D are asserted relative to T0 after their programmed delays. All of the outputs return low about 800 ns after the longest delay. The pulse outputs, AB and CD, go high for the time interval between their corresponding delay channels. t trig t cycle t ID t BUSY t A,B,C,D >5 ns >1 µs + longest delay <85 ns <800 ns + longest delay 0 to 999.999 999 999 995 s t CYCLE t TRIG TRIG T0 t ID t BUSY A B t A t B t C C D td AB AB CD CD timing diagram

Digital Delay/Pulse Generator using them (assuming a high impedance load). In this case, the average current is: I = 2Vtf/Z, where V is the pulse step size, t is the length of the cable in time (5 ns per meter for RG-58), f is the pulse repetition rate, and Z is the cable's characteristic impedance (50 Ω for RG-58). Internal or External Timebase Both internal and external references may be used as the timebase for the. The internal timebase can be either the standard 25 ppm crystal oscillator timebase, or the optional 1 ppm temperature-compensated crystal oscillator (TCXO). The internal timebase is available as a 1 Vpp square wave on a rear-panel BNC. This output is capable of driving a 50 Ω load and can be used to provide a master timebase to other delay generators. Any external 10.0 MHz reference signal with a 1 Vpp amplitude can also be used as an external timebase. individual digits. The 20-character backlit LCD display makes it easy to view delay settings in all lighting conditions. The comes standard with a GPIB (IEEE-488) interface. All instrument functions can be queried and set via the interface. You can even display the characters the has received over the interface on the front-panel LCD display. This can be valuable when debugging programs which send commands to the instrument. Fast Rise and Fall Time Modules External in-line modules are available to reduce the rise or fall time of the outputs to 100 ps. These modules use step Ordering Information Option 02 Option 03 Option 06 O4A O4B O4C O5 Delay/pulse generator w/ GPIB ±32 V rear panel outputs 1 ppm TCXO timebase Trigger inhibit input 100 ps rise time module 100 ps fall time module Bias Tee (for 02 & O4A or O4B) Dual rack mount tray recovery diodes to speed up the rise time (option O4A) or the fall time (option O4B). A bias tee (option O4C) allows these modules to be used with the optional rear-panel outputs to produce steps up to 15 V. For step amplitudes of less than 2.0 V, the fast transition time units should be attached directly to the front panel of the. Easy to Use, Easy to Program All instrument functions can be accessed through a simple, intuitive, menu-based interface. Delays can be entered with the numeric keypad in either fixed-point or exponential notation, or by using the cursor keys to select and change rear panel (with opt. 02)

Applications ATE Applications Laser Timing Applications ATE Application: Setup Time Measurement GPIB T0 A B C D Device Under Test D PR CK CLR Q Q Oscilloscope CH A CH B TRIG Laser Timing Example Pulsed Laser Experiment Q-Switch Trigger A B C D Flashlamp Trigger T o Detector SR250 Gated Integrator Sig Trig The 's versatility, precision and accuracy make it ideal for a wide variety of test and measurement tasks. In this example, the is used to measure the setup times for the data, preset and clear inputs to a flip-flop. The measurements may be made with picosecond resolution. The logic thresholds for the device under test may be measured using the 's adjustable output levels. All measurements may be controlled from the front panel or by a computer via the GPIB interface. Precision Time Control Application The 's four independent outputs make it ideal for laser timing applications. In this example the T0 output of the fires the flashlamp of a pulsed laser. Its internal rate generator controls the repetition rate of the laser and the overall experimental repetition rate. The A delay output controls the firing of the laser Q-switch. The B delay output can be used to synchronize some aspect of the experiment to the laser pulse, e.g. the application of a voltage pulse, or the triggering of a discharge. Finally, the C delay is used to trigger the gated integrator looking at the detector output. Note that both the B and C delays can be specified relative to the A delay. In this way, as the laser pulse is moved by changing the A to T0 delay, the experimental trigger and the gated integrator trigger will stay fixed relative to the laser pulse. Precision Time Sequencing GPIB 10.000 MHz TRIGGER EXT EXT EXT T0 A B C D T0 A B C D T0 A B C D Delay Outputs To Control System Timing A single can provide four transitions for precise system timing. Several s may be used if more channels are needed. The 10 MHz reference may be daisy-chained between units so that each in an experiment uses the same timebase. All of the units may be controlled over the same GPIB bus. The flexible output levels and simple architecture of the pulse/delay generators make it simple and easy to rapidly reconfigure test systems.

Specifications Delays Channels Range Resolution Timebase RMS jitter Trigger delay (typ.) External Trigger Rate Threshold Resolution Slope Impedance Internal Rate Generator Four independent delay outputs 0 to 999.999,999,999,995 seconds 5 ps 1500 ps + timebase error delay Standard: 25 ppm crystal oscillator Optional: 1 ppm TCXO (opt. 03) External: 10.0 MHz reference input <50 ps + 10 8 delay (T0 to any output) <60 ps + 10 8 delay (ext. trigger to any output) 85 ns (ext. trigger to T0 output) DC to 1/(1 µs + longest delay) ±2.56 VDC 10 mv Trigger on rising or falling edge 1 MΩ + 40 pf or 50 Ω Rate Single hot, 0.001 Hz to 1.000 MHz, or line Resolution Four digits, 0.001 Hz below 10 Hz Same as timebase Jitter 1:10,000 Settling <2 seconds for any rate change Burst mode 2 to 32766 pulses per burst at integer multiples (4 to 32767) of the trigger period Outputs Load Rise time Slew rate Overshoot Levels 50 Ω or high impedance 2 to 3 ns (typ.) 1 V/ns <100 mv + 10 % of pulse amplitude TTL: 0 to 4 VDC (normal or inverted) ECL: 1.8 to 0.8 VDC (normal or inverted) NIM: 0.8 to 0.0 VDC (normal or inverted) VAR: Adjustable offset and amplitude between 3 and +4 VDC with 10 mv resolution. 4 V maximum transition. Option 02 Fast Rise Time (opt. O4A) Output amplitude Output offset Transition time Rise (20/80 %) Fall (20/80 %) Pulse aberrations Foot Ring Fast Fall Time (opt. O4B) Output amplitude Output offset Transition time Rise (20/80 %) Fall (20/80 %) Pulse aberrations Foot Ring General Display Computer interface Dimensions Weight Power Warranty ±(50 mv + 3 % of pulse amplitude) Rear panel 1 µs pulses corresponding to T0, A, B, C, D outputs with nominal amplitude of 8 the frontpanel outputs (1 khz rep. rate). Output level is reduced by 2V/mA of additional average output current. +0.5 to 2.0 VDC 0.8 VDC (typ.) 100 ps (max.) 2000 ps (max.) 4 % (typ.) ±5 % (typ.) 0.5 to 2.0 VDC +0.8 VDC (typ.) 2500 ps (max.) 100 ps (max.) 4 % (typ.) ±5 % (typ.) 20-character backlit LCD GPIB (IEEE-488). All instrument functions and settings may be controlled over the interface bus. Interface queue can be viewed from the front panel. 8.5" 4.75" 14" (WHD) 10 lbs. 70 W, 100/120/220/240 VAC, 50/60 Hz One year parts and labor on defects in materials and workmanship

Digital Delay/Pulse Generator, Jitter, and Drift, jitter, and drift are three terms often used when discussing delay generators and time measurement equipment. Here we'll discuss what these terms mean, and how they relate to the performance of the. Jitter Various noise sources in the modulate the time delay for the outputs causing jitter. Some of these noise sources are common to all channels, others are independent. The distribution of the pulses around the desired time can be approximated by a Gaussian distribution: where: pt () = 1 σ 2π ( e t T 2 2 ) / 2σ p(t) = probability of pulse occurring at time t T = time delay for the output (mean value) σ = standard deviation of the distribution The rms jitter of the output is defined as σ, the standard deviation of the pulse delay distribution. In the, the rms jitter is a function of the delay setting. For delays less than 100 µs, the rms jitter is 50 ps. At greater delays the jitter is about 10 8 of the delay setting. Note that on an oscilloscope display you will not see the rms jitter but will instead observe the peak-to-peak jitter which is about 4 times worse. The diagram below illustrates the relation between the rms jitter and the apparent peak to peak jitter. RMS JITTER 1µs 100ns 10ns 1ns 100ps 10ps RMS Jitter vs. Time Delay Tektronix 2465 oscilloscope using int. timebase using ideal ext. timebase 1ps 10ns 100ns 1µs 10µs 100µs 1ms 10ms 100ms 1s 10s 100s 1000s DELAY SETTING Another factor to consider when trying to observe jitter on an oscilloscope is the jitter in the scope's timebase. For a good 350 MHz scope, the timebase jitter is typically 25 ps rms + 10 ppm of the timebase setting. To show how this affects the apparent jitter seen on a scope, the diagram above plots the rms jitter vs. delay for a Tektronix 2465 oscilloscope, the with its standard internal timebase, and the with an ideal, jitter-free timebase. Note that for delays above Apparent Jitter 200ps Apparent Jitter 200ps RMS Jitter 50ps 500ps/div Scope Display Gaussian Distribution

Digital Delay/Pulse Generator 1 µs, the 's jitter is swamped by that of the scope timebase. To summarize, it is difficult to accurately measure small amounts of jitter with an oscilloscope. The only accurate way to measure jitter is with a time interval counter, such as the SR620. The accuracy of a delay generator is defined as the difference between the mean value of the pulse probability distribution and the nominal front-panel delay setting. In the, the maximum error in the time delay between any two outputs can be expressed as: Error = 1.5 ns + Timebase Error t where t is the time delay between the two outputs. This error is exclusive of time shifts due to slew rates at the outputs. In other words, to accurately measure the error between two outputs they should be set to the same output levels and be driving the same load impedance. The timebase error term depends on which timebase is being used: Timebase Error A time delay of 1.0 ms implies an absolute error of ±25 ns, ±1.5 ns and ±0.5 ns respectively for the standard, optional and external timebases (assuming a 0.01 ppm external source specification). If A=100.000 µs and B=100.010 µs, the error with respect to T0 could be as large as 4.0 ns with the standard timebase, however the error of A with respect to B will be less than 1.5 ns. A graph showing the time error as a function of time delay is shown below. The four curves show the time error for the standard, optional, 0.01 ppm external, and a perfect timebase. The excess error for time delays longer than one second on the ideal external source curve is due to drift in the analog jitter compensation circuits. Drift The drift of the timebase over several hours is substantially less (10 to 100 ) than the absolute timebase error. The major factor in the timebase drift is the instrument's temperature: after the instrument is warm, the timebase drift is about 0.5 ppm/ C for the standard timebase, and about 0.05 ppm/ C for the optional timebase. The drift between several delay generators in the same experiment may be eliminated by daisy-chaining the reference output from one to the reference input on the other unit. Standard <25 ppm, 0 to 50 C Option 03 <1 ppm, 0 to 50 C External Source spec + 0.0002 ppm