NOTES ON FACSCalibur USE AND TROUBLESHOOTING PROCEDURES. Department of Immunology Monash University 14/12/07

Similar documents
Guide to the FACS Calibur

LSR II Users Guide Stanford Shared FACS Facility

Analysis and sorting of cells with FACSAria II flow cytometer Tiina Pessa-Morikawa / Revised

CFlow User Guide. Science is hard. Flow cytometry should be easy.

FACSAria I Standard Operation Protocol Basic Operation

Checking the fluids of the cart

CyAn User Manual. Flow & Mass Cytometry Center Biomedical Core facility. Technion Faculty of Medicine

Operation of the ACEA NovoCyte Flow Cytometer

CytoFLEX Flow Cytometer Quick Start Guide

Standard Operating Procedure Aria III - Irchel

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

Standard Operating Procedure Of The CytoFLEX S Flow Cytometer At The Institute Of Virology

Flow Cytometry. ZE5 Cell Analyzer. Superior performance and flexibility to align with your flow cytometry needs.

Confocal Application Notes Vol. 2 May 2004

LSR II and LSR Fortessa User Guide

More results with advanced optics and superior electronics. Gallios Flow Cytometer

Gallios. Flow Cytometry. Powerful Versatile Performance

CyFlow Space Your flexible flow cytometer

Indiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope

Cell Sorter. Operator s Guide. Software Version 1.8. LE-SH800 Series. Chapter 1 Overview. Chapter 2 Preparation. Chapter 3 Basic Operation

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

influx Cell Sorter Operator Manual Version 2.2

BD FACSAria III User s Guide

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Use of the cytometry platform - Rules of Procedure. Use of the cytometry platform University of Bordeaux - CNRS UMS INSERM US 005

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003

PCASP-X2 Module Manual

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Instructions for Use. CytoFLEX. Flow Cytometer. B49006AE November Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA U.S.A.

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

Application Note: Using the Turner Designs Model 10-AU Fluorometer to Perform Flow Measurements in Sanitary Sewers by Dye Dilution

AU5800 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU5800 Weekly (Analyzer and ISE) Maintenance

This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus:

Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual

Quick Start Operating Instructions

SUMMARY OF THE METHOD SAMPLE PREPARATION

PulseCounter Neutron & Gamma Spectrometry Software Manual

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

STANDARD OPERATING PROCEDURE:

User Manual. May 2005 AIN LIGHT

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

SPP-100 Module for use with the FSSP Operator Manual

PHI 5000 VersaProbe TM Operator s Guide

User Manual OVP Raman

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Figure 1. MFP-3D software tray

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

E X P E R I M E N T 1

OZONESONDE FLIGHT PREPARATIONS

Tali Image-Based Cytometer

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

ZYLIA Studio PRO reference manual v1.0.0

Biolyte 2000 Quick User Manual

OBS Studio Installation / Settings

Standard Operating Procedure of nanoir2-s

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

#PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis)

Session 1 Introduction to Data Acquisition and Real-Time Control

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

3.22 Finalize exact specifications of 3D printed parts.

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

UNITY 2 TM. Version 1.0. February 2008

NMR. picospin. Maintenance Guide

LASER TROUBLESHOOTING GUIDE

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

Approved by Principal Investigator Date: Approved by Super User: Date:

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

A6 OPERATING INSTRUCTIONS

SEM- EDS Instruction Manual

DPD80 Infrared Datasheet

Scout 2.0 Software. Introductory Training

Three Axis Digital Readout System

FEI FIB Focused Ion Beam

Using the BHM binaural head microphone

Countess II and Countess II FL Automated Cell Counters

Control Unit CU 2.1 User Manual V6850A HPLC

Mounting a Scintillation Detector

My XDS Receiver- Affiliate Scheduler

Operation Procedure for Phillips XL30 ESEM

PicoScope 3000 Series Automotive User guide

Integration of the Zetasizer µv and OmniFACE into an SEC system

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

Embryo In-Situ. 5) Transfer to fix in scintillation vial for 40 minutes (8ml First fix + 8ml heptane)

MS2540 Current Loop Receiver with RS485 Communication

Procedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement

Performing a Sound Level Measurement

FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091

RIDA GENE Color Compensation Kit I

HV/PHA Adjustment (PB) Part

The digital copy of this thesis is protected by the Copyright Act 1994 (New Zealand).

ab FirePlex Immunoassay Human Key Cytokines 17-plex Protocol Booklet

Dektak Step by Step Instructions:

Fragment Analyzer Genomic DNA Analysis Kit

Fiber Optic Testing. The FOA Reference for Fiber Optics Fiber Optic Testing. Rev. 1/31/17 Page 1 of 12

Transcription:

NOTES ON FACSCalibur USE AND TROUBLESHOOTING PROCEDURES Department of Immunology Monash University 14/12/07 Paul U Cameron paul.cameron@med.monash.edu.au

Page Responsibilities of FACSCalibur users 3 Preparation of Samples 3 Starting up the FACSCalibur 4 Checking Fluidics 4 Pre-run Procedure 4 CellQuest 5 Running Samples 5 Post-run Procedure 6 Time-Delay Calibration when using FL4 detector 6 FACSing of Human and Infectious Samples 7 Setting Detectors/Amps and Compensation Windows 9 Troubleshooting Guide 15 Blockage Prevention 16 Blockage Removal Solutions 16 Contact numbers 16 License documentation 17 2

Responsibilities of all flow cytometer users These guidelines suggest ways in which all users can help to ensure that the FACSCalibur machines remain operational and in excellent condition. This is every user's responsibility so please pay attention to these guidelines for the sake of yourself and all other FACSCalibur users. Users deliberately ignoring the suggestions in these guidelines and thereby creating problems with the flow cytometers may have their licence revoked. The responsibilities of all FACSCalibur users are: 1. 2. 3. 4. 5. the proper use and care of the FACSCaliburs proper sample preparation Make bookings on the online system when using the FACSCaliburs to fill out the logbooks after each use, noting any problems to alert Paul Cameron or Geza to problems or potential problems with the FACSCaliburs (see page 17 for contact numbers) 6. to let the next user know if you have finished early or changes your booking time. For the last booked user of the day (including weekends) to turn off the 7. FACSCalibur and computer. If you are booked to use the FACSCalibur and you change your mind but don't remove your booking from the computer system you must still ensure that the FACSCalibur is turned off - including on weekends. Preparation of Samples Proper sample preparation will lessen the chance of blockages during your FACS run. Blockages are caused by cell clumps and by DNA released by dead cells. Remember that the internal diameter of the probe is only ~160 µm. Hold your samples up to the light. If you can see clumps then: 1. filter the samples through nylon mesh (<100-200 micron), OR 2. add 1-5 mm EDTA to the samples prior to acquisition. If your cell viabilities are low (<70%) add 0.1% DNAse [25-100ug/ml] to your sample prior to acquisition. DNA is very sticky and will block the probe. Preparation of Control Samples In addition to your test samples you will need to prepare a number of control samples on each occasion for compensating the FACSCalibur. These include: 1. cells alone 3

2. Single-stained samples for FL1, FL2, FL3 and FL4 as necessary. Use an antibody that will stain a large proportion of cells but not all, eg. CD4 or CD8 on splenocytes or PBMC The use of these control samples for compensation is described on pages 9-14. Starting up the FACSCalibur Turn on the FACSCalibur first, wait 30 seconds then start up the computer. Check Fluidics 1. Check that the waste tank is empty. 2. Fill the sheath tank to the indicator line with milliq water (3/4 capacity). Take the tank to the carboy containing milliq waternote: For rare event work (ie. 0.5% populations) all buffers including sheath fluid must be filtered to ensure noise reduction. This should not be necessary when using milliq water as it is filtered during preparation. 1. 2. 3. 4. 5. Check that the tubing is not kinked and that the tank lids are all screwed on tight. Pressurise the sheath tank. Check the filter lines for air bubbles and purge as necessary. See Paul or Geza if you need help. Allow the cytometer to warm up for 5 minutes. If using FL4 detector, check the Four Color box at the bottom of the Detectors/Amps window. This will turn on the red-diode laser. Wait 15-20 min for this laser to warm up before proceeding. Pre-run Procedure 1. Open the tube support arm and take off the MilliQ water tube. 2. Press the PRIME fluid control button. Wait for the STANDBY button to light up. Repeat. 3. Press the RUN fluid control button. 4. Install MilliQ water on the probe again. NB. The RUN button should turn green if machine is operating OK. If the button is still orange, there is a pressure problem. 5. RUN MilliQ water at HI flow rate for 2-3 minutes. 4

CellQuest 1. Open CellQuest from the Apple menu. 2. Open Document template from the File menu to expand the size of the document. 3. Open an existing document template or make a new one by creating dot plots or histograms using the plot tools in the tool palette. 4. When creating a new dot plot "Format dot plot/histogram" window in CellQuest allows you to format your plot 5. Choose Connect to Cytometer from the Aquire menu. 6. Use the Aquisition Control and Parameter Description to define the acquisition and the data file location and description. 7. Choose Dectectors/Amps and Compensation Control from the Cytometer menu to set up your instrument settings or choose Instrument Settings from the Cytometer menu to open former settings. 8. use the Aquire menu in CellQuest to access the counter and the acquisition dialogs. Running Samples 1. Make sure your samples have been prepared properly (ie. no clumps/dna). 2. Press the RUN fluid control button before putting on your sample. 3. Secure and remove sample tubes to the probe using an anti-clockwise twisting action to prevent loosening of the probe sleeve. 4. Be careful not to knock the probe while putting sample tubes on and off as this may cause the cell stream and laser to become misaligned. 5. Start with the flow rate on LO and move to MED or HI if there are few cells. The flow rate should be no more than 1500-2000 events/sec (check on counter - found under Aquire menu). 6. Pause a few seconds with the tube support arm open between taking off and putting on sample tubes to allow the probe to be flushed out. 7. Fluid levels in tubes should not be greater than 1-2 ml to prevent fluid getting into the air intake at the top of the probe. NB. When setting up Acquisition and Storage parameters (see page 5) please: (i) set the Resolution to 256 (rather than 1024). This aquires data in only 256 channels instead of 1024 channels, which is adequate for most uses. Furthermore, the files saved will be much smaller. (ii) open 'Parameters Saved' and make sure that all the parameters you are aquiring have their boxes checked. Unchecking boxes for parameters that aren't being used will make files smaller. 5

Post-run Procedure 1. RUN 1% Extran for 3 minutes on HI flow rate. 2. Run 1% hypochlorite* for 3 minutes on HI flow rate (last user of the day: 10 minutes). Keep cap on tubes contaiing 1% hypochlorite as the bleach will oxidize and be useless. 3. RUN MilliQ water for 3 minutes on HI flow rate (last user of the day: 5 minutes). 4. Press the STANDBY button. 5. Depressurise the sheath tank and refill with millq water. Take the tank to the miliq carboy - DON'T refill it by use of a cylinder. 6. Empty the waste tank. 7. Disconnect from cytometer. 8. Quit CELLQuest. 9. Transfer files to your own computer and delete the files from the FACSCalibur hard drive. NB. Data files from CellQuest or CellQuest Pro can be read by any analytical software [winlist, winmdi, flowjo, weasel] but the documents are specific for each program. 10. Last user of the day: Turn off the computer then the FACScalibur. Time-Delay Calibration when using FL4 detector If using the FL4 detector, check the Four Color box at the bottom of the Detectors/Amps window. This will turn on the red-diode laser. Wait 15-20 min for the laser to warm up before proceeding. Time-Delay Calibration electronics synchronizes the FSC signal and the FL4 signal in time. Changes in sheath flow rate might change the number of microseconds it takes a particle to go from the red beam to the blue beam. 1. Open 'Time-Delay Calibration alias' under the Apple menu. 2. Choose 'Connect to Cytometer' from the Aquire menu. 6

3. Choose 'Threshold' from the Cytometer menu. Adjust the FSC threshold to 200. 4. Install a tube containing APC beads on the SIP. The tube is stored in the fridge in a box marked 'Calibration Beads'. 5. Choose 'Dectectors/Amps' from the cytometer menu. Adjust the FSC amp gain to place the mean Peak on the FSC histogem to Channel 400+/-5. 6. Choose 'Counters' under the Aquire menu. Make sure the event rate is above 400 events/second. If the event rate is too low, add more beads to the tube. 7. Adjust the FL4 PMT voltage to place the mean peack in the FL4 histogram to the channel value written on the APC beads card +/-5. 8. Choose 'Time-Delay Calibration' from the Cytometer menu. Click Calibrate. If not successful, see Geza. 9. Return the FSC threshold to 52 and the FSC amp gain values to their previous settings. 10.Remove the tube of APC beads from the SIP. Place back in the fridge. 11.Note the FSC amp gain and FL4 PMT voltage values in the log book. The FACSCalibur is now ready for four-colour work. FACS analysis of Human and Infectious Samples Treat ALL human samples as infectious and follow this protocol. Virus-infected samples should also be treated as follows: 1. Wear gloves, a gown and protective glasses 2. Ensure that there is at least 100 ml neat hypochlorite in the waste tank 3. Place Glad Wrap on the computer key board and mouse 4. On completion of acquisition of files, run 1% hypochlorite for 5 minutes followed by Milli Q water for a further 5 minutes 7

5. Remove Glad Wrap and swab down the bench area and cytometer surfaces with 70% ethanol 6. Remove the waste tank and store for 30 minutes (not days!) 7. Empty the waste tank into the sink and flush away the contents with copious volumes of water 8

Setting detecetors/amps and compensation windows Forward-scattered light (FSC) - proportional to cell-surface area or size Side-scattered light (SSC) - proportional to cell granularity FL1-FL4 detectors - measure fluorescence-emission wavelength bands Fluorochromes FITC PE Emission peak (nm) 525 (green) 575 (orange-red) Detector FL1(and FL2) FL2(and FL3 and FL1) Propidium iodide (PI) 620 (red) FL2 and FL3 - can't be compensated PerCP >670 (red) FL3(and FL4) TRI-COLOR >670 (red) FL3(and FL4) CyChrome >670 (red) FL3(and FL4) APC 660 (red) FL4(and FL3) From Becton Dickinson FACSCalibur System User's Guide 02-61760-02 9

Control samples Cells alone Single-stained samples for FL1, FL2, FL3 and FL4 as necessary. Use an antibody that will stain a large proportion of cells but not all, eg. CD4 or CD8 on splenocytes. This allows the location of both positive and negative cells to be assessed. Adjusting detector sensitivity 1. Open Detectors/Amps and Compensation windows under Cytometer menu. 2. If using the FL4 detector, check the Four Color box at the bottom of the Detectors/Amps window. This will turn on the red-diode laser. Wait 20 min for the laser to warm up before proceeding. 3. Set all Compensation values to 0%. 4. Turn fluid control dial to RUN and place your "cells alone" sample on the probe. 5. Set up a FSC vs SSC dot plot. 6. Adjust FSC and SSC Voltage and Amp Gain as required so that your populations sit clearly within the dot plot and are not scrunched up against an axis: Voltage Amp Gain Mode FSC Increase for 10-fold gain Fine control adjustment Lin SSC Adjust as required Leave on 1.00 Lin 10

7. Place a gate around your population of interest. If you have more than one population of interest you will have to compensate each separately. 8. Set up FL1 vs FL2, FL2 vs FL3 and FL3 vs FL4 dot plots and FL1, FL2, FL3 and FL4 histograms as necessary and set the gates on G1=R1. 9. Set Mode to Log for FL1, FL2, FL3 and FL4 in the Detectors/Amps window. 10. Set Mode to Log for FL1, FL2, FL3 and FL4 in the Detectors/Amps window. 11. Adjust the FL1, FL2, FL3 and FL4 Voltages so the negative population in the "cells alone" sample sits between 100 and 101 on the histograms and in the bottom left hand corner of the dot plots. 12. Place your single-stained FITC control sample on the probe. 13. Check that the negative population still sits between 100 and 101 on the FL1 histogram. 14. Adjust the FL2-%FL1 compensation while viewing the FL1 vs FL2 dot plot. 11

15. Increase FL2-%FL1 to subtract FL1 fluorescence from FL2 detector. Usually set at 15-40%. 16. Often the positive cells will have a wider scatter than the negative cells. Adjust so that the median of each population is the same. There is a risk of overcompensation if you try to get the population within the same quadrant. 17. Place your single-stained PE control sample on the probe. 18. Check that the negative population still sits between 100 and 101 on the FL2 histogram. 19. Adjust the FL1-%FL2 compensation while viewing the FL1 vs FL2 dot plot. 12

20. Increase FL1-%FL2 to subtract FL2 fluorescence from FL1 detector. Usually set at <1%. 21. Adjust the FL3-%FL2 compensation while viewing the FL2 vs FL3 dot plot. 22. Increase FL3-%FL2 to subtract FL2 fluorescence from FL3 detector. Usually set at ~10%. 23. Place your single-stain control sample for the FL3 detector (eg. TRI-COLOR or PERCP) on the probe. 24. Check the negative population still sites between 100 and 101 on the FL3 histogram. 25. Adjust the FL2-%FL3 compensation while viewing the FL2 vs FL3 dot plot and the FL4-%FL3 compensation while viewing the FL3 vs FL4 dot plot, as before. 26. If doing four colour work, place your single-stained APC control sample on the probe. 27. Check the negative population still sits between 100 and 101 on the FL4 histograms. 28. Adjust the FL3-%FL4 compensation while viewing the FL3 vs FL4 dot plot, as before. 13

NOTE: If you have to change the voltage of any detector during compensation you must check the compensation on ALL single-stained tubes again. 29. Mix a small amount of each of your single-stained samples in a new tube and place this on the probe to check that all populations sit correctly against the axes. 30. Uncheck the Setup box in the Acquisition Control window of the Browser and collect singlestained samples and mixed stained samples for your own records. Open the Acquisition and Storage window under the Acquire menu bar and fill in the number of events you wish to collect and in what region (eg. G1=R1). Collect:10,000 events for single-stained samples 20,000 events for double-stained samples 30,000 events for triple-stained samples 31. A Resolution of 256 channels is usually adequate for routine analysis (rather than 1024 channels) and stores less data. For for best quality publication plots acquire at maximum resolution and increase the number of acquired events to twice the above recommendations. 32. Save your instument settings in the InstrSettings window under the Cytometer menu bar. 33. If analysing more than one population, place a gate around this new population, adjust gates on all FL histograms and dot plots to G2=R2 and repeat all steps from step 10 to step 26. 14

TROUBLESHOOTING GUIDE: From: Current Protocols in Immunology (1991). Edited by JE Coligan et al. 15

BLOCKAGE PREVENTION: The best blockage prevention is proper sample preparation. Refer to page 3 of these notes. However, if you are acquiring a large number of samples, wash through 1% Triton and/or 1% hypochlorite followed by MilliQ water and saline at regular intervals. This will help to clear any potential blockages as they occur. BLOCKAGE REMOVAL SOLUTIONS: Run fresh 1% hypochlorite* at HI flow rate for 5-15 minutes. Run 1% Triton X at HI flow rate for 5-15 minutes. Run MilliQ water at HI flow rate for 5 minutes. Run saline at HI flow rate for 5 minutes. *Hypochlorite (bleach) will oxidise if left in the air so keep the cap on the tube when not in use. If you are unsure of how old the tube of bleach is, make up a fresh tube. Make up bleach solution by diluting 1 volume of 10% bleach into 9 volumes of ddh20 If the cytometer is still blocked, seek help from Geza or Paul Cameron. CONTACT NUMBERS: If you have any problems or queries please contact the following people: Name Location Phone Email Geza Paukovics Burnet 9282 2246 paukovic@unimelb.edu.au Paul Cameron ID Unit 92071449 Sarah immunology paul.cameron@med.monash.edu.au Mobile 0404 205 406 sarah.snelgrove@med.monash.edu.au 16

AMREP cell sorting facility Certification of competence for operation of PC2 FACS analysers Staff member... Supervised operation of the FACS analyser in PC2 Analyser identification date hours supervisor date hours supervisor Unsupervised analysis analyser Declaration by FACS user I have read the guidlines and notes for operators of FACS calibur and am familiar with the operation of this machine. I am able to deal with common problems including problems with sample preparation and analysis of sample under PC2 conditions. Signed... Date... Supervisor... Date... Laboratory Manager /FACS operator... Date... 17

Version 1.29 Revision Date 15/02/06 Author Paul Cameron 18