Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV

Similar documents
Testing Speech Quality of Mobile Phones in a Live Network Application Note

Measurement of Acoustic Properties of Rooms Using Audio Analyzer R&S UPV Application Note

Generating the Noise Field for Ambient Noise Rejection Tests Application Note

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Effective Test Procedures for Installing and Maintaining RF Transmitter Sites

AMIQ-K2 Program for Transferring Various-Format I/Q Data to AMIQ. Products: AMIQ, SMIQ

Getting Started with the LabVIEW Sound and Vibration Toolkit

Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal. Application Note

Using the BHM binaural head microphone

Collection of Setups for Measurements with the R&S UPV and R&S UPP Audio Analyzers. Application Note. Products:

How To Demonstrate Improved ACLR Dynamic Range With FSU and Noise Correction

Test Automation Tool for POLQA and PESQ Speech Quality Tests Application Note

ANSI/SCTE 40 Conformance Testing Using the R&S SFU, R&S SFE and R&S SFE100

EMI Automotive Band Evaluation

Using R&S NRP Series Power Sensors with Android TM Handheld Devices. Application Note. Products: R&S NRP Series. R&S NRP-Zxx Series

DVB-T Bursted Noise Signal Generation

Pre-5G-NR Signal Generation and Analysis Application Note

Correlated Receiver Diversity Simulations with R&S SFU

Level Accuracy and Electronic Level Settings of SMIQ

Features/Specifications

Noise Detector ND-1 Operating Manual

Quick Reference Manual

Easy Analysis and Precise Optimization of ATSC or ATSC Mobile DTV Single-Frequency Networks Application Note

Design and Use of a DTV Monitoring System consisting of DVQ(M), DVMD/DVRM and DVRG

SigPlay User s Guide

Out of Band Spurious Measurement for Bluetooth Modules

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix

RFEX V Release Notes

USER S GUIDE DSR-1 DE-ESSER. Plug-in for Mackie Digital Mixers

Coherence Measurement between two Signals regarding Timing, Phase and Gain Application Note

USB Mini Spectrum Analyzer User s Guide TSA5G35

NPR - Noise Power Ratio Signal Generation and Measurement

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

Multi-port calibration by using a two port calibration unit. Application Note. Products: R&S ZVT R&S ZNB

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

What is the minimum sound pressure level iphone or ipad can measure? What is the maximum sound pressure level iphone or ipad can measure?

Concise NFC Demo Guide using R&S Test Equipment Application Note

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1

T80-K1 V2.22 Release Note

LabWindows/CVI, VXIpnp driver history for the R&S Vector Network Analyzers Driver Documentation

Binaural Measurement, Analysis and Playback

Iterative Direct DPD White Paper

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

Standard Operating Procedure of nanoir2-s

The BAT WAVE ANALYZER project

R&S FSV-K40 Phase Noise Measurement Application Specifications

Analyzing and Saving a Signal

SC24 Magnetic Field Cancelling System

Grid Interactive Advanced Features Self-Supply Mode

Sound Insulation Reporter

SC24 Magnetic Field Cancelling System

System Requirements SA0314 Spectrum analyzer:

LabWindows/CVI, VXIpnp driver history for the R&S Directional Power Sensors

Diamond Cut Productions / Application Notes AN-2

A Matlab toolbox for. Characterisation Of Recorded Underwater Sound (CHORUS) USER S GUIDE

Multiband Noise Reduction Component for PurePath Studio Portable Audio Devices

Tests on CDMA2000 1xEV-DO Access Terminals in Accordance to IS-866-A

Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators. Technical Brief

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200

AcoustiSoft RPlusD ver

<WCDMA Inner Loop Power Control tests, test pattern E&F,G&H using ARB files on the R&S CMW 500>

R&S BCDRIVE R&S ETC-K930 Broadcast Drive Test Manual

SR-D8-M, SR-D8-S. (Ver ) SOFTWARE INSTRUCTIONS

Manual for the sound card oscilloscope V1.41 C. Zeitnitz english translation by P. van Gemmeren, K. Grady and C. Zeitnitz

<EDGE Looped based BER and TX measurements using ARB files on the R&S CMW 500> Products: R&S CMW500 R&S CMW- KW200 R&S CMW- KM200. <Application Note>

EASERA SysTune. Developed by. AFMG Ahnert Feistel Media Group. The creators of EASE and EASERA.

BER MEASUREMENT IN THE NOISY CHANNEL

APM CALIBRATION PROCEDURE Rev. A June 3, 2015

Application Note #63 Field Analyzers in EMC Radiated Immunity Testing

Transmitter Measurements on Bluetooth TM Modules with FSP

LedSet User s Manual V Official website: 1 /

Mastering Phase Noise Measurements (Part 3)

LabVIEW driver history for the R&S RTH Handheld Digital Oscilloscope Driver Documentation

Transmitter Measurements on Bluetooth TM Modules

Quick Start for TrueRTA (v3.5) on Windows XP (and earlier)

AFMG SysTune. Developed by. AFMG Ahnert Feistel Media Group. The creators of EASE and EASERA

SC26 Magnetic Field Cancelling System

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

Field Test 2. Installation and operation manual OPDAQ Installation and operation manual

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017

LabVIEW driver history for the R&S HMP Power Supplies Family

E X P E R I M E N T 1

USER MANUAL FOR DDT 2D. Introduction. Installation. Getting Started. Danley Design Tool 2d. Welcome to the Danley Design Tool 2D program.

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5

Tests on 3G-Base Stations to TS with FSIQ and SMIQ

Getting started with Spike Recorder on PC/Mac/Linux

Linrad On-Screen Controls K1JT

PulseCounter Neutron & Gamma Spectrometry Software Manual

USER GUIDE FOR NETmc MARINE X-Ops

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X

Troubleshooting EMI in Embedded Designs White Paper

Lab experience 1: Introduction to LabView

Next Generation Software Solution for Sound Engineering

Cisco Spectrum Expert Software Overview

FOUR CHANNEL USB RECORDER PCRU01. User manual. Table of Contents

R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz

Speeding up Spectrum Analyzer Measurements Application Note

RF Solution for LED Display Screen

Manual Version Ver 1.0

Transcription:

Product: R&S UPV Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV Application Note 1GA47_0E This application note explains how to use acoustic calibrations in order to be able to enter or read sound pressure level directly in the Audio Analyzer R&S UPV. An application program is described which performs the necessary measurements for the calibration of microphones and loudspeakers, manages the calibration values and enters them into any appropriate user defined setup. Subject to change 02.2007 1GA47_0E

Contents 1 Overview... 3 2 Introduction... 3 3 Use of Reference Values in the UPV... 4 4 Installation of the Application Program... 6 5 Operation of the Application Program... 7 Microphone Calibration... 7 Loudspeaker Calibration... 8 Using Calibration Values... 9 6 Hardware and Software Requirements... 9 7 Appendix: Using the Quick Launch Buttons... 9 8 Ordering information... 10 1GA47-0E 2 Rohde & Schwarz

1 Overview This application note explains how to use acoustic calibrations in order to be able to enter or read sound pressure level directly in the Audio Analyzer R&S UPV. An application program is described which performs the necessary measurements for the calibration of microphones and loudspeakers, manages the calibration values and enters them into any appropriate user defined setup. The Audio Analyzer R&S UPV is abbreviated to UPV for the remainder of this Application Note. 2 Introduction Acoustic measurements involve the measurement of sound pressure level or the generation of a sound field, or both. Typical examples are noise measurements, loudspeaker measurements, microphone measurements and measurements on systems like hearing aids and mobile phones. A sound level meter, for example, is supposed to directly display the sound pressure level in dbspl (db referred to 20 µpa sound pressure). If an audio analyzer like UPV, which measures voltage, is used for this purpose, the sensitivity of the microphone s ~ VM = p M ~ has to be determined, wherein V ~ M is the rms value of the microphone output voltage and ~ p is the rms value of the sound pressure which produced the output voltage. The value of the sound pressure is obtained by dividing the measured voltage by this sensitivity value. The measurement with which the value of the microphone sensitivity is obtained is commonly called microphone calibration. The sound pressure level for the microphone calibration is generated with a so-called acoustic calibrator which generates a sinewave signal with defined frequency (commonly 1 khz) and defined sound pressure (usually 1 Pa or 10 Pa). As measurement microphones are small and have a well-defined mechanical structure, the sensitivity is frequency-independent within a certain frequency range. Therefore calibration at a single frequency is sufficient. Certain measurements require the generation of a defined sound pressure at a certain point. In order to be able to set a desired sound pressure level in the generator of the UPV, the sensitivity of the loudspeaker ~ p = L s L ~ VG has to be determined, wherein p~ L is the sound pressure generated by the loudspeaker at the pre-defined measurement point, and generator output voltage which produced the sound pressure. V ~ G is the 1GA47-0E 3 Rohde & Schwarz

As a loudspeaker has a far more complex mechanical structure than a measuring microphone, and radiation effects additionally influence the loudspeaker sensitivity depending on the frequency and on the location of the measurement point, the sensitivity of a loudspeaker usually is frequency dependent. For this reason the calibration of a loudspeaker consists of two steps: 1. Measurement of the absolute sensitivity at one frequency 2. Measurement of the frequency response relative to this frequency. In order to generate a defined sound pressure at the measurement point, the generator output voltage has to be set to the desired sound pressure divided by the loudspeaker sensitivity, and to be corrected by the inverse frequency response (equalization). It has to be observed that the loudspeaker calibration is only valid for the point of the calibration, due to the propagation properties of the sound wave. Note that an equalization in the sense of a frequency dependent amplitude correction is only possible with signals which are defined in the frequency domain, like (swept) sinewave and multisine signals. For the equalization of a complex signal defined in the time domain like speech, a filter is required. 3 Use of Reference Values in the UPV For numerical function results, the microphone sensitivity can be entered in the Function Config Panel as Reference Value: If the value entered as reference value is the microphone output produced by 1 Pa sound pressure, the resulting display value is dbpa (db referred to 1 Pa). If it is desired to display dbspl (db referenced to 20 µpa), the microphone output voltage for 20 µpa has to be entered as reference value. It is obtained by subtracting 94 db from the voltage level measured at 1 Pa, or by dividing this voltage by 50000. 1GA47-0E 4 Rohde & Schwarz

For a level display, the unit has to be set to dbr. For display of sound pressure the unit has to be set to V/Vr. The labeling in the display can be adapted to the desired unit if the checkbox Unit Auto is unchecked (see example above). For sweeps, the reference value can be entered in the Sweep Graph Config Panel. Alternatively, Reference in the sweep graph config panel can be set to Meas Panel. This copies the reference value entered in the Function Config Panel to the sweep. The same applies respectively to FFT spectra (FFT Graph Config Panel) and 1/n octave spectra (Bar Graph Config Panel). For the input monitor and the level monitor the reference can be set according to the numeric function result in the Input Config Panel or Level Mon Config Panel, respectively. The only difference is that there is one common reference value for both channels. On the generator side the sensitivity is entered as reference value in the Generator Config Panel. The loudspeaker frequency response can be stored from the Sweep Graph Config Panel as equalization list: The checkbox Invert has to be ticked in order to invert the measured frequency response for the equalization. Norm Freq has to be set to the frequency at which the absolute sensitivity had been measured. The stored equalization list can then be opened as equalization file in frequency-defined generator signals: 1GA47-0E 5 Rohde & Schwarz

For each frequency the amplitude is individually corrected by the factor given in the equalization file, whereby frequencies are interpolated between those present in the file. 4 Installation of the Application Program The use of the application program requires option UPV-K1 to be installed. The application program can be installed as executable program or opened as a project in the IDE of Visual Basic.NET for modification and new compilation. For just using the executable, copy the installer files CalibrationTool.msi, Config.ini, dotnetfx.exe, Settings.ini and Setup.exe to the hard disk of your UPV and start Setup.exe by double-clicking. If the DotNet runtime framework is already present on the UPV (e.g. when the VB.NET IDE is installed or other DotNet application programs had been previously installed), it is sufficient to copy and start CalibrationTool.msi. The program is stored in the program files folder under Rohde&Schwarz\CalibrationTool\CalibrationTool.exe. The installer also creates a shortcut on the desktop of the UPV. Furthermore, the installer copies the necessary setup files to the program folder, and the program installs a folder D:\UPV\Calibrations automatically at its first startup. 1GA47-0E 6 Rohde & Schwarz

If the program is compiled and started from the IDE, the folder "Calibrations" in D:\UPV has to be created in advance, and the two setups MicCal.set and SourceCal.set have to be copied to it. For editing and compiling the source code, unzip the project folder and copy it to the hard disk of the UPV (e.g. in D:\Visual Studio Projects ). Click Open Project on the start page of the Visual Basic IDE, browse to the folder and open AcousticCalibration1.sln. 5 Operation of the Application Program The use of the application program requires the respective measurement setup to be prepared for the entry and application of the reference values. For details please see section 3 above. The application note comprises also an example setup SweepExample.set for calibration of a loudspeaker and two different microphones. In particular, channels for which a microphone calibration is to be entered by the tool, have to be active in the setup. The references of graphs should be set to Meas Panel. In order to enable proper loudspeaker calibration, Equalizer should be checked in the Generator Function Panel. Furthermore dbr amplitude values in the Generator Function Panel should be set in the setup sufficiently low not to overrange the Generator. Any illegal operation condition caused by the entry of the calibration values produces an error window and closes the application program. Starting AcousticCalibration1.exe opens the following entry window on the UPV: The window is divided into three panels: One for the loudspeaker calibration, one for the microphone calibration and one for updating stored calibration values to the UPV setup. Microphone Calibration The microphone calibration is available for each of the two input channels. Mic Calibration starts the calibration routine. Once the calibration routine has successfully finished, a file can be specified to store the results. This file name is simultaneously entered into the Calibration File text box which specifies the file from which the calibration value to be entered into the UPV setup is taken. Calibrate and Update automatically enters the value in the 1GA47-0E 7 Rohde & Schwarz

Function Config Panel for the respective channel. To select a calibration value from a file without perfoming the calibration routine, use Open File. To calibrate a microphone, connect it to the desired analyzer input. Insert the microphone into an acoustic calibrator and switch on the calibrator. After clicking the Mic Calibration button of the respective channel a window opens which queries the SPL produced by the calibrator. If Selective is activated, the calibration measurement is performed with a fast 3 rd octave band pass filter. Therefore in this case also the frequency of the signal produced by the calibrator has to be entered. After that a series of voltage measurements is started. If the single results are sufficiently consistent, the sensitivity of the microphone is displayed, and a file selector box is opened for entering the file name to store the value. Loudspeaker Calibration Loudspeaker calibration requires a valid microphone calibration to exist for the analyzer channel used for the speaker calibration measurement. The used channel has to be selected in Analyzer Channel. Lower Limit, Upper Limit and Point Count specify the parameters for the frequency response sweep done during the calibration measurement. Selective toggles between wideband and selective loudspeaker measurement. To calibrate a loudspeaker, connect it to one of the UPV outputs. If required, a power amplifier can be inserted. Connect a microphone to one of the analyzer inputs and calibrate the microphone or load a valid calibration for this microphone from a file. Select the correct analyzer channel for the loudspeaker calibration. Place the microphone at the point in front of the loudspeaker for which the calibration should be valid. After clicking Speaker Calibration, the SPL for the calibration has to be entered. Due to nonlinearities of loudspeakers there may be deviations in sensitivity and frequency response between calibrations executed at different levels. It is also possible to perform calibrations for a set of different levels and store them under different file names. After the level has been confirmed, the calibration starts with first determining the absolute sensitivity at 1 khz, then measuring the relative frequency response. In a second run the obtained equalization is checked and corrected before a third sweep is done to verify the result of the calibration. After completion of the measurement, a file name has to be specified for storing the calibration values. Subsequently the UPV is returned to local control and the calibration tool window is minimized in order to show the sweeps: 1GA47-0E 8 Rohde & Schwarz

After the Source Sensitivity window has been closed by clicking ok, the calibration tool window re-appears and the previous setup of the UPV is restored. It is essential that the test configuration (gain of the power amplifier, positioning of loudspeaker and device to be measured) is not altered between the calibration and the measurement in which the calibration is to be used. Using Calibration Values Update Calibrations enters the calibration values from the files specified in the three text boxes as reference values for the ticked generator and analyzer channels. An update to input and monitor display is only possible in single channel mode. It can be chosen whether the dbr values in the setup and displays should be interpreted as dbspl (relative to 20 upa) or as dbpa (relative to 1 Pa). Load Setup and Update Calibrations allows to load a UPV setup and automatically enter the activated calibration values as reference values into the loaded setup. The application program returns the UPV firmware to local control after termination of each action. Therefore it can be kept open during the use of the UPV. However, it is not possible to run another remote control application while the calibration tool window is open. The window is minimized after loading and updating the setup and can be retrieved from the Windows task bar. 6 Hardware and Software Requirements The Audio Analyzer UPV needs firmware version 1.4.1 or higher installed. For the use of the application program, option UPV-K1 (Sequence Control) is required. 7 Appendix: Using the Quick Launch Buttons The UPV offers 8 quick launch buttons which can be utilized either for loading pre-defined setups or for starting pre-defined macros. The assignment of the buttons can be defined in Utilities Quick Launch Config : Fig. 1 Starting the Quick Launch Config Panel This menu opens the following configuration panel: 1GA47-0E 9 Rohde & Schwarz

Fig. 2 Quick Launch Config Panel Use the buttons on the right side to browse to the setup file or exe file which should be called by the respective button. The buttons on the left side can be utilized to specify icons on the buttons. Text entered in the text box on the left side will be displayed as tool tip text for the respective button when the mouse cursor is staying on the area of the button. 8 Ordering information UPV Audio Analyzer DC 250 khz 1146.2003.02 UPV-K1 Sequence Controller 1401.7009.02 For additional information about audio analyzers and other measurement equipment, see the Rohde & Schwarz website www.rohde-schwarz.com. ROHDE & SCHWARZ GmbH & Co. KG. Mühldorfstraße 15. D-81671 München. Postfach 80 14 69. D-81614 München. Tel (089) 4129-0. Fax (089) 4129-13777. Internet: http://www.rohde-schwarz.com This application note and the supplied programs may only be used subject to the conditions of use set forth in the download area of the Rohde & Schwarz website. 1GA47-0E 10 Rohde & Schwarz