Approved by: / / R. Battaglia 12/16/2016

Similar documents
University of Minnesota Nano Fabrication Center Standard Operating Procedure

Standard Operating Manual

Operation Procedure for Phillips XL30 ESEM

University of Minnesota Minnesota Nano Center Standard Operating Procedure

Exercise 4-2. Counting of Actuator Cycles EXERCISE OBJECTIVE & & &

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

THERMIONIC GUN CATHODE-GRID ASSEMBLY TEST PROCEDURE

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000

HIGH CURRENT IMPLANTER 200KeV - SERIES 1090 TECHNICAL DESCRIPTION

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Flat-Bed Module Recorders

CHA EVAPORATOR Short Reference User guidelines

Figure 1. MFP-3D software tray

Quick Start Bruker Dimension Icon AFM

Dektak Step by Step Instructions:

e-vap 3kW Evaporation Sources

PHI 5000 VersaProbe TM Operator s Guide

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Detailed Design Report

EG4015. Digital Generator Governor Controller User Manual

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

Click Here To Start Demo

MICROFOCUS X-RAY SOURCE PROJECT*

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Performance of a DC GaAs photocathode gun for the Jefferson lab FEL

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

Instruction Manual MANLM101070C REV. D0202

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5)

Noise Detector ND-1 Operating Manual

LITTON IOT CIRCUIT ASSEMBLY MANUAL

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

SceneStyle2 User Guide

PLASMA PROCESS GROUP, INC. RFN Radio Frequency Neutralizer Manual

J R Sky, Inc. tel: fax:

Sentinel I24 Digital Input and Output Configuration

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

HITACHI. VisionCube. Service Manual

SM-1100/SM-1500/SM-1600 ROTARY ACTUATORS

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

Reciprocating Machine Protection

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

User Guide Stand-Alone Metering for OptiPlant

Approved by Principal Investigator Date: Approved by Super User: Date:

4.9 BEAM BLANKING AND PULSING OPTIONS

ArcPro Mach4 Plasma Screen User Guide

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

Application Note No. 157

SATRI AMPLIFIER AMP-51R. Owner s Manual

SonoruS Audio. ATR10 Analog Tape Reproducer. Operating Manual

Operation Manual for. SCU1 Signal Conditioning Unit

User Manual CC DC 24 V 5A. Universal Control Unit UC-1-E. General Information SET. Universal Control Unit UC-1 Of Central Lubrication PAUSE CONTACT

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

EDL8 Race Dash Manual Engine Management Systems

CMPE 466 COMPUTER GRAPHICS

3214NXT. Service Manual. IMPORTANT: Fill in Pertinent Information on Page 3 for Future Reference

FEI FIB Focused Ion Beam

3B SCIENTIFIC PHYSICS

1. General principles for injection of beam into the LHC

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1

imso-104 Manual Revised August 5, 2011

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

RG NDT INTERNATIONAL INC. Drill Pipe and Tubing Inspection System

Universal High Current Implanter for Surface Modifications with ion beams Extensive range of ion species, including refractory metals Magnetic mass

MS2540 Current Loop Receiver with RS485 Communication

CURRENT INJECTION RELAY TESTER CR-100. ( Code P60212 (770081) ) INSTRUCTIONS MANUAL ( M / 05A ) (c) CIRCUTOR S.A.

DektakXT Profilometer. Standard Operating Procedure

PC-250. SMD Taped Parts Counter Operator s Manual. ISO 9001:2008 Certified. V-TEK, Incorporated 751 Summit Avenue Mankato, MN USA

Reduction of Device Damage During Dry Etching of Advanced MMIC Devices Using Optical Emission Spectroscopy

STANDARD OPERATING PROCEDURE:

Interface Practices Subcommittee SCTE STANDARD SCTE Composite Distortion Measurements (CSO & CTB)

Topic: Instructional David G. Thomas December 23, 2015

AREAL- Phase 1. B. Grigoryan on behalf of AREAL team

IEFIS G3 Inputs, outputs and Alarms

Full Disclosure Monitoring

Quadrupoles have become the most widely used

Troubleshooting CS800/LC900 Bikes

Operation of CEBAF photoguns at average beam current > 1 ma

High Performance (Gold Plus) Spliceable Tape Feeder Part Number: Part Number: Revision 3 Jun 2008 No.

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE Composite Distortion Measurements (CSO & CTB)

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0

INSTRUCTION MANUAL COMMANDER BDH MIG

Operating manual incl. Declaration of conformity

USER S GUIDE. 1 Description PROGRAMMABLE 3-RELAY LOGIC MODULE

18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V

Introducing Purion H, a Scanned Spot Beam High Current Ion Implanter

Spectra/Chrom Chart Recorder from Spectrum Chromatography

SPECIFICATION NO Model 207 Automatic GTAW Welding System

Professional 4-channel Thermocouple Thermometer plus Infrared Thermometer

Blown Film Extrusion. Line Control. 5 Layers CD Series IBC Temperature USER MANUAL

Provides an activation of Relay 1 triggered by Input 1. The function also provides an option for reverse-logic on the activation of Input 1.

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

OSCILLOSCOPE AND DIGITAL MULTIMETER

Using an oscilloscope - The Hameg 203-6

3200NT System 14. Service Manual. IMPORTANT: Fill in Pertinent Information on Page 3 for Future Reference

Version 1.10 CRANE SONG LTD East 5th Street Superior, WI USA tel: fax:

The Cathode Ray Tube

Transcription:

Fabrication Laboratory Revision: H Rev Date: 12/16/16 Approved by: Process Engineer / / R. Battaglia 12/16/2016 Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Varian 350D Ion Implanter. All users are expected to have read and understood this document. It is not a substitute for inperson training on the system and is not sufficient to qualify a user on the system. Failure to follow guidelines in this document may result in loss of privileges. 2 REFERENCE DOCUMENTS AND PEOPLE Varian 350D Implant manual Ion Implant lecture and video, L.F. Fuller and I.R. Turkman Steve Stefanski, Kodak Equipment Engineer Scott Blondell, RIT Facilities Manager Richard Battaglia, RIT Equipment Engineer 3 DEFINITIONS 3.1 General Description The Varian 350D Ion Implanter is capable of implanting Boron (B11 or BF2) or Phosphorous (P31) in 4 or 6 wafers with doses of 1E12 to 5E15. Implant energy levels can be adjusted from 10KeV and 200 KeV. NOTE: The REMOTE ELECTRONICS RACK contains switches and joysticks to adjust the beam setup, focus and steering parameters. The operator can also turn the adjustment dials manually on the MAIN IMPLANTER CONSOLE. Several of the remote switches and motor drives are inoperative and must be adjusted from the main console. This instruction set identifies the exact controls to use to avoid confusion. RIT SMFL Page 1 of 10

4 TOOLS AND MATERIALS 4.1 The Varian 350D Ion Implanter has dedicated Teflon wafer boats for the 4 and 6 end stations. An attached label indicating VARIAN 350D identifies them. 5 SAFETY PRECAUTIONS 5.1 Refer to the following MSDS sheets: - Boron Trifluoride - Phosphine/Hydrogen - Argon - Fomblin Vacuum Pump Fluid 5.2 The following stored energies are present in this tool and represent potential hazards to operators and service personnel: - High voltage DC supplies - High voltage: 208VAC 3 phase - Line voltage: 120 VAC - Compressed Air: 80-100 psi - Compressed Nitrogen: 20 psi - Pressurized Water Cooling Lines: 40-60 psi 5.3 The following represent physical hazards: - Pneumatic Cylinders - Wafer Pick Arms - Robotic End Station Cassette Handlers - Load Lock Components 5.4 The following represent potential health hazards: - Source pressurized gases - Contaminated sources and source components - Vacuum pump fluids RIT SMFL Page 2 of 10

6 INSTRUCTIONS 6.1 System Start Up 6.1.0 Activate the card swipe interlock by logging in at CARD SWIPE ACCESS 3 located in cleanroom hallway. Panel 7 POWER key switch is interlocked. 6.1.05 Verify that NITROGEN for the Varian 350 Implanter is turned ON at the NITROGEN MANIFOLD 2765 located in Service Chase 2765. The manifold regulator gauge should read 20 psi. The gauge on Implanter located below the endstation controller should read approx. 9 psi. * NOTE Turn ON the Appropriate END STATION MOTOR POWER SWITCH located on the ENDSTATION OPERATOR CONTROL PANEL. Switches are a lift and select style. Do not force them if they do not move. 6.1.1 Check the levels of the three VACUUM GAUGE CONTROLLERS located on panels 3, 5 and 8. All should read <1E-6 Torr on the LOG scale. 6.1.2 Use the POWER key to turn ON the POWER switch (left side of panel 7 - remote rack). 6.1.3 Remove POWER key from panel 7. Return the MODE key to panel 6. 6.1.4 Use the MODE key to select the SET UP mode (right side of panel 6 remote rack). 6.1.6 Select ES 1 for 6 or ES 2 for 4 (End station 1 or 2) on the BEAM FOCUS AND SCAN portion of panel 6. 6.1.7 Press the TERM ON button (left side of panel 7). Ensure that gauges behind the viewing window inside the HIGH VOLTAGE TERMINAL illuminate. RIT SMFL Page 3 of 10

6.1.8 Turn on the VIEWING WINDOW LIGHT to fully illuminate the meters inside the HIGH VOLTAGE TERMINAL. 6.1.9 Set the BEAM MONITOR to I/X (panel 2). 6.2 Beam Set Up 6.2.1 Ensure that ALL dopant gases are OFF the B, P and CAR ON lights are OFF. If any buttons are lit, press them once to turn OFF that gas function. 6.2.2 For a source warm-up, select the desired gas by pressing the B(oron) or P(hosphorous) button (panel 7 DOPANT). 6.2.3 Verify the SOURCE VACUUM pressure is in the range of 5 E-6 Torr to 5 E-5 Torr, as indicated by the GREEN markers (panel 5). If the vacuum needs to be adjusted, turn the NITROGEN gas adjust pot located on PANEL 11 CW to increase vacuum or CCW to decrease the source vacuum. 6.2.4 Verify that the ARC-VOLTS meter visible through the TERMINAL VIEWING WINDOW is reading between 60-110 volts (gas dependent). B~110V P~60V 6.2.5 Verify that the FILAMENT-AMPS meter visible through the VIEWING WINDOW is approx. 0 amps. 6.2.6 Verify that the EXTRACTION VOLTS meter visible through the VIEWING WINDOW is approx. 33KeV. 6.2.6 Adjust the SOURCE MAGNET dial on PANEL 11 to approx 200. 6.2.7 Set the ARC manual dial to 050 (PANEL 11). 6.2.8 Slowly increase the FILAMENT dial on PANEL 11 while observing the FILAMENT-AMPS meter through the VIEWING WINDOW. As you approach 200 AMPS check the ARC-AMPS meter for any RIT SMFL Page 4 of 10

deflection from its zero position. This indicates plasma has been ignited. NOTE: If NO ARC AMPS are indicated then decrease the FILAMENT current to 50 AMPS and increase the SOURCE MAGNET dial to 400. Slowly increase the FILAMENT current again, while observing the ARC-AMPS meter. This procedure will need to be repeated until the plasma is created, as indicated by an ARC-AMPS reading. If necessary, decrease the FILAMENT to 50 AMPS, increasing the SOURCE MAGNET by another 200 and then increase the FILAMENT current towards 150 AMPS. If no plasma is established after several attempts, contact a Technician for guidance (Rich @ 478.3834 or Scott @ 738.4073). 6.2.9 Pause at the point of ARC-AMPS deflection to allow the source to stabilize for 60 seconds. 6.2.10 Adjust the ARC dial to give approx. 0.1-0.3 AMPS on the ARC- AMPS meter in the VIEWING WINDOW. Allow the source to stabilize at these settings for 5 minutes. 6.2.11 Readjust the ARC dial on panel 11 to give approx. 0.2-0.5 ARC- AMPS. Stable plasma is more desirable than a specific arc current. 6.2.12 Turn ON the high voltage (If Needed) by depressing the HV button located on the left side of panel 7. 6.2.13 Turn up the black BEAM ENERGY ADJUST KNOB located on the right side of PANEL 13, located just above the countertop. Watch PANEL 6 for ENERGY readout in KeV. 6.2.14 Turn ON the chart recorder LINE POWER and SERVO power. Load chart paper and turn ON the CHART HOLD. Uncap the recorder pen. 6.2.15 Locate the desired species peak by monitoring the chart recorder for peaks while adjusting the ANALYZER dial on panel 11: RIT SMFL Page 5 of 10

B11 ~ 290 P31 ~ 485 BF2 ~ 625 F ~ 385 It may be beneficial to plot the entire spectrum to verify the correct peak. 6.2.16 Adjust the ANALYZER dial slightly in both directions to obtain the peak beam current as displayed on the panel 4 DOSE PROCESSOR meter. Select the appropriate scale to display the current. Monitor the chart recorder during this step to stay on the correct peak. 6.2.17 Adjust the SOURCE STEERING joystick on panel 7 in both X directions to maximize the beam current on the panel 4 DOSE PROCESSOR meter. The Y-axis is disabled. 6.2.18 While observing the M image on the panel 2 scope, adjust the focus and center the beam using the TRIM, BALANCE and OFFSET- SHIFT pots on panel 12. Make adjustments for I/X and I/Y (select on panel 2). 6.2.19 On the BEAM SCAN CONTROLLER (panel 10) adjust the H AMP and V AMP pots for proper tail lengths on the I/X and I/Y M s. Adjust the V STEER pot for Y centering. 6.2.20 Set the desired implant dose on the panel 4 DOSE PROCESSOR. 6.2.21 Select the scanned implant area by choosing the desired end station: - ES # 1 for 6 wafers = 196 cm2 - ES # 2 for 4 wafers = 91.5 cm2 6.2.23 Switch to X/Y on the Beam Monitor Display (panel 2). This value closely represents the actual implant current. 6.2.24 Calculate the implant time and adjust the PANEL 11 ARC pot or SOURCE MAGNET pot as necessary to yield an implant time > 20 seconds. RIT SMFL Page 6 of 10

6.3 Doing the Implant Time=(Dose x Area x q) / I where: area =91.5 cm 2 for 4 or 196 cm 2 for 6 q=1.6e-19 NOTE: On the computer desktop double-click on the Beam Current Calculator shortcut and enter the Dose, Area, and Desired Beam Current to calculate the Implant Time. * NOTE Verify the appropriate END STATION MOTOR POWER SWITCH located on the ENDSTATION OPERATOR CONTROL PANEL has been turned ON as per step 6.1.05. 6.3.1 Load the wafers in the marked Teflon cassette provided. DO NOT use any other cassette. 6.3.2 Standard orientation is wafers with flats up, device side facing into the Implanter, starting in the boat location and slot closest to the operator. 6.3.3 On the End Station Operator Control Panel, set the number of wafers to 00 on the thumb wheels. 6.3.4 Set Implanter mode to IMPLANT (key switch on panel 6). 6.3.5 Set the end station to ES2 for 4 or ES1 for 6 wafers (IMPLANT MODE controls on right side of panel 6). 6.3.6 Press the appropriate START button on the END STATION controller. * NOTE It is strongly recommended that the ABORT button be used in the event of a wafer mishandle event. DO NOT continue with a START command. THIS MAY BREAK YOUR WAFER AND/OR DAMAGE THE TOOL. Contact a technician to troubleshoot the problem. 6.3.7 The implanter will look at all 25 wafer slots and return to the LOAD position and an alarm will sound indicating lot completion. On the End Station Operator Control Panel, push the COMP-MUTE and then the ERROR-CLEAR buttons. RIT SMFL Page 7 of 10

6.3.8 ERROR 21 - If the DOSE PROCESSOR returns an ERROR 21 after pressing the START button, this is an indication that the beam current is too high for the dose you have selected. In order to ensure a uniform implant the dose processor calculates the implant time (T) before starting the actual implant. If T< ~ 15 seconds, it will cause an error. Go to SETUP, lower the beam current, return to IMPLANT and press START on the ENDSTATION. 6.4 Shut Down/ Standby * NOTE Turn OFF the appropriate ENDSTATION MOTOR POWER SWITCH located on the ENDSTATION OPERATOR CONTROL PANEL. 6.4.1 Set the Key switch to SET UP mode (panel 6). 6.4.2 On Panel 2, switch the Beam Monitor to I/X. 6.4.3 Set the SOURCE STEERING (panel 7) to maximize current displayed on the BEAM CURRENT meter (panel 4). 6.4.3 Turn down the Beam Energy Adjust knob located on panel 13 to < 35 KeV as shown on panel 6. 6.4.4 Turn down FILAMENT dial to Zero (panel 11). Check meter in viewing window to make sure it is as close to zero as possible. Dial at zero won t always get meter to zero. 6.4.6 Set ARC to 050 on dial (panel 11). 6.4.7 Turn OFF the Dopant Gas button on remote electronics rack (panel 7). 6.4.8 Turn OFF the High Voltage (HV OFF) (panel 7). 6.4.9 Turn OFF the Terminal Voltage (TERM OFF) (panel 7). 6.4.10 Turn OFF the POWER key button by pushing in the button (panel 7). RIT SMFL Page 8 of 10

6.4.11 Turn OFF the VIEWING WINDOW LIGHT on the HIGH VOLTAGE TERMINAL. 6.4.12 Turn OFF Varian nitrogen on N2 Manifold 2765 located in chase 2765. 6.4.13 Turn OFF End Station Motor on the End Station Operator Control Panel. 6.4.14 Deactivate the Card Swipe Interlock by logging off at CARD SWIPE ACCESS 3. 6.5 Errors During Run 6.5.1 Consult the Error Code chart posted on the Implanter for corrective action and contact a technician. RIT SMFL Page 9 of 10

7 ATTACHMENTS REVISION RECORD Summary of Changes Originator Rev/Date Original Issue R. Battaglia A-12/13/02 Updated all instructions Scott Blondell B - 03/17/03 Added N2 instructions and ERROR CODES Scott Blondell C - 03/24/03 Updated all instructions Scott Blondell D - 08/07/07 Updated last page to include Recorder shut off Mike M. E 6/25/07 Updated instructions using written edits from Rev E. Kelly Johnson F 10/19/10 Updated all instructions; eliminated N2 carrier gas warmup. Scott Blondell G 11/08/11 Updated all instructions using written edits in Rev G R. Battaglia H 12/16/2016 RIT SMFL Page 10 of 10