Calibrating attenuators using the 9640A RF Reference

Similar documents
Advanced Test Equipment Rentals ATEC (2832)


RF Level Test System +20 dbm to 130 dbm

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Application Note DT-AN DTU-315 Verification of Specifications

R&S FS-Z60/75/90/110 Harmonic Mixers for the R&S FSP/FSU/ FSQ/FSUP/FSV

Calibration Certificate

How To Demonstrate Improved ACLR Dynamic Range With FSU and Noise Correction

Sources of Error in Time Interval Measurements

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

UFX-EbNo Series Precision Generators

Full-featured CW Microwave Counters for Field, Factory or Lab

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor

R&S ZVA-Zxx Millimeter-Wave Converters Specifications

JD725A Cable and Antenna Analyzer - Dual Port

R&S HA-Z24E External Preamplifier 1 GHz to 85 GHz Specifications

R&S ZN-Z32/-Z33 Automatic In-line Calibration Modules Ensuring high accuracy with thermal vacuum testing and multiport measurements

R&S FSV-K40 Phase Noise Measurement Application Specifications

R&S ZVA-Zxx Millimeter-Wave Converters Specifications

R&S ZN-Z154 Calibration Unit Specifications

Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal. Application Note

MPI Cable Selection Guide

USB-TG124A Tracking Generator User Manual

Critical Benefits of Cooled DFB Lasers for RF over Fiber Optics Transmission Provided by OPTICAL ZONU CORPORATION

RF Characterization Report

Agilent N9355/6 Power Limiters 0.01 to 18, 26.5 and 50 GHz

Product Introduction. Duplexer Box MN2555A. Signal Analyzer MS2830A. Ver.1.0

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

LadyBug Technologies LLC Manual PowerSensor+ Field Certification Procedure

Identifying and avoiding common errors in RF calibration.

R&S ZN-Z151/-Z152/-Z153 Calibration Unit Specifications

R-1580A Microwave Downconverter. Product Brochure

Hewlett Packard 3577A 5Hz MHz Network Analyzer Specifications SOURCE

PM Series Microwave Power Calibration System

PM Series Microwave Power Calibration System

Orbital 694XA Series. Ka BAND EXTERNAL REFERENCE LNB with rear anchor posts. Wide range of Frequencies and Bandwidths LNB 1855R 1000 XA-WN60

R&S SMBV-Z1 Reference Frequency Converter Specifications

e2b calibration 521 Fifth Avenue Chardon, Ohio Phone: Fax: Certificate Of Calibration

R&S ZN-Z103 Calibration Unit Specifications. Data Sheet V02.01

R&S ZN-Z85 Switch Matrix Specifications

R&S HF907DC SHF Downconverter Specifications

Keysight Technologies High Power Ampliier Measurements Using Nonlinear Vector Network Analyzer. Application Note

Extension kit for R&S Vector Network Analysers

HMC7056. Block Upconverters / HPA's. Typical Applications. General Description. Features. Functional Block Diagram

OBSOLETE HMC7056. Block Upconverters / HPA's. Typical Applications. General Description. Features. Functional Block Diagram

ThinkRF D GHz RF Downconverter

R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications

R&S ZV-Z81 Multiport Test Set, models.05/.09/.29 Specifications

Test and Communications Antennas for the R&S TS8991 OTA Performance Test System Specifications

R&S NRPM Over-the-Air (OTA) Power Measurement Solution Specifications

Test and Communications Antennas for the R&S TS8991 OTA Performance Test System Specifications

Spectrum Analyzers. 2399A 9 khz to 3 GHz Spectrum Analyzer.

18 GHz, 2.2 kw KLYSTRON GENERATOR GKP 24KP 18GHz WR62 3x400V

-Preliminary- Ku-band PLL LNB - Internal Reference (L.O. Stability: ±1 ppm) - MODEL No. NJR2835U/36U/37U/39U series

R&S FPS-K18 Amplifier Measurements Specifications

Keysight Technologies High-Power Measurements Using the E5072A ENA Series Network Analyzer. Application Note

Modulated Wideband Power Amplifier

R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz

Prisma Optical Networks Ancillary Modules

INTEGRATED ASSEMBLIES MICROWAVE SOLUTIONS FROM TELEDYNE COUGAR

Agilent CSA Spectrum Analyzer N1996A

Transmitter Calibration

Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change.

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz

Agilent 87075C Multiport Test Set Product Overview

8650A SERIES UNIVERSAL POWER METERS

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

Transcom Instruments. Product Brochure TRANSCOM INSTRUMENTS. Product Brochure. 1

Software. Documentation. Test Equipment

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

High-Value 100 Series (Outdoor)

Agilent 8560 E-Series Spectrum Analyzers

Microwave Counter, Power Meter and DVM in One Portable Package

In-Line or 75 Ohm In-Line

Model P/03P Upconverters

DA MHz Series of Narrowband or Wideband Distribution Amplifiers

Power Amplifier 0.5 W 2.4 GHz AM TR Features. Functional Schematic. Description. Pin Configuration 1. Ordering Information

DA : Series of Narrowband or Wideband Distribution Amplifiers

MAXTECH, Inc. BRC-1000 Series. C-Band Redundant LNB Systems. Technology for Communications. System Block Diagrams

14 GHz, 2.2 kw KLYSTRON GENERATOR GKP 22KP 14GHz WR62 3x400V

4400X Series. Orbital. X-Band Ext Ref LNB. MIL Spec MIL-STD A, 45 db internal filter LNB630S-500X-WN60-G12. Orbital Features: Orbital Specs:

Agilent E5500 Series Phase Noise Measurement Solutions Product Overview

Measurement Accuracy of the ZVK Vector Network Analyzer

R&S ZNL Vector Network Analyzer Specifications

Orbital Ka-ISO. Ext Ref Ka LNB with integrated isolator. Orbital Research Ltd Marine Drive, White Rock, BC. Canada V4B 1A9

R&S SGMA Product Family Compact fast reliable

RS Pro SPECTRUM ANALYZER SSA3000X SERIES

Technical Datasheet. 8650B Series Universal Power Meters 10 MHz to 50 GHz. 8651B Single Channel 8652B Dual Channel Rev.

MULTIBAND 1/3 RACK-MOUNTED

FREQUENCY CONVERTER 1/3 RACK-MOUNTED BLOCK CONVERTER. Narda-MITEQ FEATURES OPTIONS. Unit shown with Option 17. Unit shown without Option 17

R&S NRP Power Sensor Family Specifications

R&S FSMR Measuring Receiver Specifications

Spectrum Master. Compact Handheld Spectrum Analyzer. Technical Data Sheet

Specifications. FTS-4335 Series

Agilent 81600B Tunable Laser Source Family

CHAPTER 3 SEPARATION OF CONDUCTED EMI

Transcom Instruments. Product Brochure TRANSCOM INSTRUMENTS. Product Brochure

Spectrum Analyzer 1.6 GHz 3 GHz HMS-X

Oscilloscopes for field applications

Transcription:

Calibrating attenuators using the 9640A RF Reference Application Note The precision, continuously variable attenuator within the 9640A can be used as a reference in the calibration of other attenuators, or in the calibration of gain or loss of other RF components or paths. This application note will concentrate upon the calibration of fixed and switched attenuators; the method for other components or paths would be similar. The calibration method to be described compares the unknown attenuator with a nominally equal Reference value within the 9640A and uses the 9640A unique Offset capability to determine in a direct readout either a value for the unknown attenuator or its error from nominal value. << Interface Adapters >> N (f) - 2.92 (m) 2.92 (f-f) 9640A-50 Leveling Head 30 db DUT Attenuator Power Sensor DUT Interface Plane Basic method Using the example of a 30 db fixed attenuator, for which we will assume a different connector series to illustrate the application of inter-series adapters. Our device under test (DUT) will be a female-tomale attenuator in 2.92 mm connectors. 1. In this case we need to adapt the 9640A output to 2.92 mm using an N (f) to 2.92 (m) adapter. Connect the unknown attenuator in series, and its output to a power measuring instrument e.g. a power sensor, again adapting interfaces as necessary. 2. Set an output level of say +16 dbm at the desired test frequency, very commonly 50 MHz. 3. Turn the 9640A output on and note the resulting power reading, in this case about -14 dbm (~40 µw). The next step that you take depends upon the desired result format... To determine attenuation value: 1. To find the unknown attenuation value, use the 9640A Offset feature to reduce the output level by -30 dbm, as shown here. From the Fluke Digital Library @ www.fluke.com/library

2. Remove the unknown attenuator (only) leaving both adapters in place (dotted connection above). Note here that it might important to reduce 9640A output (as above) before removing the attenuator to avoid overloading the measuring sensor. 3. We now adjust the 9640A offset to achieve exactly the same measured power level. The 9640A offset value then matches the insertion loss of the unknown attenuator; in this case -30.158 db. 2. Remove the unknown attenuator (only) leaving any adapters in place. Note here that it might important to reduce 9640A output before removing the attenuator to avoid overloading the measuring sensor. We now adjust the 9640A Offset to achieve exactly the same measured power level (in this case -0.158 db). To determine attenuation error from nominal: 1. To find the unknown attenuation error from nominal, first select the Level Offset function, leaving the Offset value at 0 db. Then use the 9640A Level control to reduce the output level by the nominal attenuator value -30 db, as shown here. The 9640A Step Level edit mode may be valuable here. In this example a 5dB step has been set, 10dB steps could also be used. This feature is particularly valuable when a range of equi-spaced values needs to be calibrated (e.g. a switched attenuator or a set of attenuators). 3. By now pressing the Offset as UUT error key, we conveniently display the error of the insertion loss from the nominal attenuator value (+0.158 db or, if we wish, +3.705 %). Multiple attenuators in a set, or a switched attenuator can be calibrated quickly, efficiently and accurately using the above method, which is much less prone to error than traditional methods. The 9640A Step Edit and Offset features significantly simplify and speed the task, providing, as they do, a direct readout of the unknown attenuator value or error from nominal. 2 Fluke Calibration Calibrating attenuators using the 9640A RF Reference

Coping with large value attenuators Use of a power sensor as the comparison measurement instrument will limit the largest attenuation value that can be determined. Typically repeatable measurement using a thermal power sensor will only be possible down to about -10 dbm. Assuming the same 16 dbm start point the largest measurable attenuation value would be around 26 db. A single path diode sensor might support comparison readings down to around -50 dbm or attenuation values to 66 db. A multipath diode sensor might gain another 20 db of useful range. Beyond these we must use the tuned measurement of a spectrum analyzer or a measuring receiver. With the best of these we can make comparisons down to -130 dbm or lower and can potentially resolve attenuation values of 120 db or more. Note that for very large attenuation values isolation of input and output to prevent signal leakage around the DUT is vital. High integrity components, sensor/receiver and correct connector torque must be used. Consideration of test signal spectral content We do not have to consider the spectral content of our comparison measurement so long as the measurement is at a fixed comparison level and the source spectral content does not substantially change throughout the test. The 9640A being of very low and consistent spectral content at all output levels, negates the need for further consideration. The danger of using high input power It will not be lost on readers that the use of a higher input power will extend the usable attenuation measurement range of any given comparison sensor and the 9640A could deliver up to +24 dbm to the attenuator input. There are however two dangers here: 1. That with low attenuator values the comparison power at the sensor may be too high and may damage the sensor. 2. That self heating of the DUT may cause its attenuation value to change. This will depend upon the temperature rise ( C/W) of the attenuator resistive elements and the Temperature Coefficient of the total attenuation value; which may or may not be the same under self heating as under ambient heating. Attenuators are generally inadequately specified in this respect and we often cannot establish a good value for the error that may occur. It is likely that high power attenuators will be less susceptible to this form of non-linearity than lower power or smaller devices. Fluke recommends that for a 1 W attenuator (+30 dbm max) no more than 16 dbm should be used at Calibration. Optimizing accuracy There are three key opportunities to optimize attenuation measurement accuracy: 1). Optimize match conditions. The attenuator must be driven from and loaded with accurate and stable match to minimize mismatch errors. The 9640A is well specified and certified in this respect, but its output match does change with level range switching and a small additional uncertainty results. Best practice would be to apply a small value attenuator or pad between the 9640A and the DUT to reduce this mismatch switching uncertainty and to move the match defining resistor closer to the DUT interface plane. Whilst not generally necessary when measuring with a typically well matched power sensor, best practice when comparing at lower levels would be to also insert a small value pad between the DUT attenuator and a less well matched receiver or spectrum analyzer. Clearly pads reduce measured signal level and may equally be impacted by self heating if we were to raise source level to compensate. An ideal match pad would theoretically improve match (reflection coefficient) by twice its attenuation value, however there is a sharply diminishing return for increasing pad values as their own match error limits the improvement. 9640A source mismatch is specified at VSWR of 1.1 : 1 at frequencies <500 MHz. Attenuation is generally measured in this frequency band so we will use this value in an example evaluation of mismatch uncertainty. A good power sensor may be specified at 1.05 : 1 at frequencies <1 GHz. If we were to pad the source and the sensor in this case, optimum results would be achieved if the source were padded more heavily (twice as heavily) than the sensor. However, the best attenuator match is unlikely to be better than 1.02 : 1 to 100 MHz. In fact it is likely that we would have to measure the attenuator to be certain that it was this good, such tight specification is rarely found. The extent to which the pad can improve match at the DUT attenuator plane is thus limited. << Interface Adapters >> N (f) - 2.92 (m) 2.92 (f-f) 9640A-50 Leveling Head In Pad DUT Attenuator Out Pad Power Sensor DUT Interface Plane 3 Fluke Calibration Calibrating attenuators using the 9640A RF Reference

The table below evaluates uncertainty due to source and load mismatch when measuring a 10 db (nominal) attenuator with hi quality Input and Output pads of various attenuation values and at a match of 1.02 : 1. Two examples are given, that for a DUT with ideal match and a more realistic case, a match of 1.05 : 1. In comparison with the specified attenuation accuracy of the 9640A the additional uncertainty is relatively small in all cases. It is unlikely that anymore than a 3 db source pad would be deemed valuable to uncertainty reduction. In practice, a reflection measurement would be likely to show the power sensor match, inclusive of any good quality adapter fitted, to be somewhat lower than specified and to be repeatable with connection matings. Further reducing the need for, or improvement that could be gained by using an output matching pad. A further consideration in the presence of such small benefits in uncertainty would be the increase in uncertainty that might result from a comparison measurement at lower level. In this case -24 dbm would be a small signal for most thermal power sensors. Nominal Output Pad Nominal Output Uncertainty due to Mismatch 9640A Attenuator Value Input Pad Value Value Level DUT Match Ideal DUT Match 1.05 : 1 Specification 10 db No Pad No Pad 6 dbm ± 0.0035 db ± 0.0110 db ± 0.02 db 10 db 1 db No Pad 5 dbm ± 0.0030 db ± 0.0090 db ± 0.02 db 10 db 3 db No Pad 3 dbm ± 0.0017 db ± 0.0075 db ± 0.02 db 10 db 6 db 3 db -3 dbm ± 0.0006 db ± 0.0035 db ± 0.02 db 10 db 10 db 6 db -10 dbm ± 0.00035 db ± 0.0030 db ± 0.02 db 10 db 20 db 10 db -24 dbm ± 0.00025 db ± 0.0025 db ± 0.02 db The tables above and below are derived from Monte Carlo statistical simulation of the suggested mismatch specifications and representative transmission line lengths between them. All specifications and results are assumed to be of 99 % confidence. All results are consistent up to a frequency of 80 MHz, heavier padding of the source (the longest line length in the measurement) raises the frequency to which these values apply up to 200 MHz (6 db) and 400 MHz (20 db). The uncertainties are relatively independent of DUT attenuator value but do improve at lower values, as shown below. Nominal Source Pad Nominal Output Uncertainty due to Mismatch 9640A Attenuator Value Value Load Pad Value Level DUT Match Ideal DUT Match 1.05:1 Specification 1 db 6 db 3 db 6 dbm ± 0.00015 db ± 0.00015 db ± 0.02 db 3 db 6 db 3 db 4 dbm ± 0.0004 db ± 0.0035 db ± 0.02 db 10 db 6 db 3 db -3 dbm ± 0.0006 db ± 0.0035 db ± 0.02 db 60 db 6 db 3 db -53 dbm ± 0.0006 db ± 0.0008 db ± 0.03 db 2) Optimize comparison measurement noise and repeatability. Assuming fixed input level (+16 dbm) the only benefit that can be made is to reduce measurement noise and drift in the comparison measurement. Good practices such as averaging or filtering the measurement, maintaining fixed and quiet environment, careful handling and correct torque on connections, and minimizing the handling of thermal sensors in particular will all help. It is also important to select the best available measurement for the comparison level. The table on the right indicates typical repeatability of measurement that can be achieved against comparison level for thermal and diode power sensors and high end measuring receivers. These results are measured values in a controlled temperature production calibration environment and include contributions such as connector repeatability. Typically Achievable Measurement Repeatability 10 MHz** to 300 MHz, inclusive noise, stability and connection repeatability Measuring Receiver/ Spectrum Analyzer Comparison Level Thermal Sensor e.g. R&S NRP-Z51 Diode Sensor e.g. R&S NRP-Z4 e.g. R&S FSU26 with pre-amp option 10 dbm to 20 dbm 0.005 db 0.005 db 0.009 db 0 dbm to <10 dbm 0.005 db 0.005 db 0.009 db -10 dbm to <0 dbm 0.005 db 0.005 db 0.009 db -20 dbm to <-10 dbm 0.005 db 0.01 db -30 dbm to <-20 dbm 0.005 db 0.01 db -40 dbm to <-30 dbm 0.006 db 0.01 db -48 dbm to <-40 dbm 0.008 db 0.01 db -57 dbm to <-47 dbm 0.012 db -67 dbm to <-57 dbm 0.02 db -77 dbm to <-67 dbm 0.02 db -87 dbm to <-77 dbm 0.045 db -97 dbm to <-87 dbm 0.04 db (use pre-amp) -107 dbm to <-97 dbm 0.06 db (use pre-amp) -117 dbm to <-107 dbm 0.08 db (use pre-amp) -124 dbm to <-117 dbm 0.15 db (use pre-amp) ** At lower levels it may be necessary to offset measurement from 10 MHz and harmonic multiples to reduce interference from distributed 10 MHz reference clock often found in a calibration lab. +50 khz offset is often used. 4 Fluke Calibration Calibrating attenuators using the 9640A RF Reference

3). Optimize application of 9640A specification. The 9640A specification is optimized in the frequency band 10 MHz to 128 MHz and for an initial level of +16 dbm as shown below. Other Reference levels can be used, but relative to calibration standards, an uncertainty for both the DUT and Reference measurements must be included in the uncertainty summation, as indicated in the lower section of the 9640A Attenuation specification below. Attenuation 50 Ω Output Attenuation Relative to +16 dbm ouput 10 Hz (1) to 128 MHz 0 db to 56 db 56 db to 64 db 64 db to 74 db 74 db to 100 db 100 db to 116 db ± 0.02 db ± 0.03 db ± 0.05 db ± 0.07 db ± 0.15 db Cumulative and incremental attentuation To determine the attenuation specification between any two output levels, apply RSS (2) summation of the db values listed for each output level. Relative to any level between +16 dbm and -100 dbm, 10 Hz to 128 MHz +16 db to -40 db -40 db to -48 db -48 db to -58 db -58 db to -84 db -84 db to -100 db ± 0.02 db ± 0.03 db ± 0.05 db ± 0.07 db ± 0.15 db Total uncertainty in an attenuation measurement derives from the combination of 9640A specification, mismatch uncertainty at the DUT interface and repeatability in the comparison measurement. Using an example DUT of 10 db at a match of 1.05 : 1, and the use of a 3 db source pad, at 99 % confidence we have: (converting to and combining as power uncertainty and converting back to db) (+16 dbm reference) Mismatch uncertainty ± 0.0075 db ± 0.17% pwr Comparison uncertainty at 3 dbm ± 0.005 db ± 0.12 % pwr (thermal or diode sensor) RSS total ± 0.0219 db ± 0.505 % pwr Had we used a +10 dbm reference instead (+10 dbm reference level) (+0 dbm attenuated level) Mismatch uncertainty ± 0.0075 db ± 0.17 % pwr Comparison uncertainty at -3 dbm ± 0.005 db ± 0.12 % pwr (thermal or diode sensor) RSS total ± 0.0296 db ± 0.683 % pwr Conclusion The 9640A RF Reference simplifies and streamlines the calibration of attenuators and, by providing direct readout of either attenuation value or error from nominal value, reduces errors in the recording and certification process. 9640A attenuation specification supports low uncertainty measurements and additional uncertainties due to mismatch or comparison measurement repeatability can readily be kept small or insignificant. In practice, given the mix and magnitude of contributing uncertainties, the benefits of padding the measurements are minimal and in practice only rarely would padding greater than an input pad of 3 db be justifiable. Fluke Calibration. Precision, performance, confidence. 5 Fluke Calibration Calibrating attenuators using the 9640A RF Reference Fluke Calibration PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In the U.S.A. (877) 355-3225 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com 2011 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 4/2011 4015975A A-EN-N Pub-ID 11791-eng Modification of this document is not permitted without written permission from Fluke Corporation.