Scope: All CT staff technologist

Similar documents
2012 Computed Tomography

2017 Computed Tomography

Technologist Quality Control Procedures

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Procedures for conducting User QA on the scanner

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Request for Proposals

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Guidelines for Assuring Softcopy Image Quality

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System

2

-Technical Specifications-

Procedure Manual for MRI of the Brain

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications

Troubleshooting Guide. Prep, Scan Errors, and Artifacts

Understanding CT image quality

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2

CytoFLEX Flow Cytometer Quick Start Guide

FFDM Quality Control in Canada - a Vendor Neutral Approach

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Non-Uniformity Analysis for a Spatial Light Modulator

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality.

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018)

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors

Performance Evaluation of Industrial Computed Radiography Image Display System

Guidance for Quality Assurance of PACS Diagnostic Display Devices

U SER S G UIDE. TS2002A Fiber Optic Test Kit

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

Coronis 5MP Mammo. The standard of care for digital mammography

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

MR Accreditation Programs - E. Jackson

GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

Breast Imaging Essentials

MRI Training for Scanning at the VA

Ambient Weather WS-01 Intelligent Color Changing Temperature Night Light with Ambient Backlight User Manual

Mammo Tomosynthesis 5MP

True comfort and flexibility with the power of 3T.

V9A01 Solution Specification V0.1

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

Initial Experience With Automatic Image Transmission to an Intensive Care Unit Using Picture Archiving and Communications System Technology

Collimation and Light Field Brightness

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

User Guide. HDMI Fiber Optic Extender. DVI-7350a

Safety Light Curtain C 4000

Nio. Industry-standard diagnostic display systems

RQflex 10. Manual English

This talk covers currently available display technology.

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Monitor QA Management i model

Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1

DICOM Conformance Statement

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

+ Human method is pattern recognition based upon multiple exposure to known samples.

MQSA Quality Control Manual

LITE-ON TECHNOLOGY CORPORATION

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System

If you have questions or problems in providing anything described as Customer

Real-time QC in HCHP seismic acquisition Ning Hongxiao, Wei Guowei and Wang Qiucheng, BGP, CNPC

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

Figure 1. MFP-3D software tray

In recent years, CT technology has undergone profound

Essentials of Digital Imaging

An Introduction to Vibration Analysis Theory and Practice

PanelView 1400e CRT Maintenance

Mechanical aspects, FEA validation and geometry optimization

SECTION 5900 TRAFFIC SIGNALS CITY OF LEE S SUMMIT, MISSOURI DESIGN CRITERIA

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Operating Procedures for RECO1 & RECO2

Global Trade Medical Supplies

SINAMICS G130. dv/dt filter plus Voltage Peak Limiter. Operating Instructions 03/2013 SINAMICS

DISTRIBUTION AMPLIFIER

Quality Assurance (QA) Guidelines for Medical Imaging Display Systems

From Autonomous Subsystems to Integrated System

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

Radiological Imaging. Software Version 3.6

Rev. 01 February 2014 X- RAY SAFETY MANUAL

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management

Operating instructions Retro-reflective sensor. OJ50xx laser / / 2010

BN-6 CHIN UP ATTACHMENT OWNER S MANUAL

NOTIFICATION OF A PROPOSAL TO ISSUE A CERTIFICATION MEMORANDUM

ASSEMBLY AND CALIBRATION

White Paper. Diagnostic Color Displays. What s inside?

Brilliance CT 64-channel configuration

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Transcription:

APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist, or supervising physician. The equipment QC program is designed to ensure CT equipment is operating within the specifications guidelines set by the manufacturer, Department Environmental Quality the American College of Radiology (ACR). Scope: All CT staff technologist Policy: The CT Technologist will perform log quality control procedures based on the following criteria: Procedure: 1. Once every 24 hours (scanner used 24/7) or before the first patient is scanned 2. After major repairs. Major repairs include replacement or repair of components such as an x ray tube or detector assembly. 3. Any out of range quality control values or other problems will be immediately brought to the attention of the supervisor or manager. The QC program consists of the following procedures. PROCEDURE MINIMUM FREQUENCY APPROXIMATE TIME IN MINUTES Water CT Number Stard Deviation Daily 5 Artifact Evaluation Daily 5 (or less) Visual Checklist Monthly 5 Display Monitor Quality Control Monthly 5

APPROVED BY: Radiology Technical Director Page 2 of 6 Water CT Numbers Stard Deviation (Noise) is performed to ensure the relative calibration of all CT numbers to water remains within acceptable limits quantum noise electronic system noise do not increase. Too much image noise degrades low contrast detectability. Frequency: Daily Required Use the water phantom provided by the scanner manufacturer. Equipment: Record: are recorded on the data form for daily CT Equipment Quality Control. 1. Warm up the scanner s x ray tube according to manufacturer recommendations. Select Procedure: daily prep on the GE operators console or check up from the setup tab on the Siemens. 2. Perform calibration scans (often called air calibration scans) according to scanner manufacturer recommendations. Perform fast cals on the GE calibration on the Siemens. 3. Place the QC phantom on the holder device provided. Center the phantom at the isocenter of the scanner using the laser alignment lights of the scanner the alignment marks on the phantom surface. 4. For GE equipment, set up a scan of the QC phantom using the scanner s ACR daily QC scan parameter settings. Complete recon 1 2. Plastic, Air Water mean stard deviation values should be monitored for both axial helical scan modes. 5. For Siemens equipment, follow the manufacturer s guidelines. See the attached letter from Thomas Larry of KLS Physics Group. GE Equipment 1. Select the first image of the axial group of images obtained, to analyze. Place a region of interest (ROI) in the plastic, the water in air on the image. If the size of this ROI is not specified by the scanner manufacturer, use an area around 400 mm 2. Record the values reported for the plastic, air water mean stard deviation on the data form. 2. Repeat the above measurements for the first image of the helical group.

APPROVED BY: Radiology Technical Director Page 3 of 6 Record the values reported for the plastic, air water mean stard deviation on the data form. 3. For Siemens equipment, follow manufacturer s suggestions. See the attached letter from Thomas Larry of KLS Physics Group. 4. The mean values ranges for plastic, air water should correspond with the specific unit. CT Unit Plastic Air Water the data form. 3. For Siemens equipment, follow manufacturer s suggestions. See the attached letter from Thomas Larry of KLS Physics Group. 4. The mean values ranges for plastic, air water should correspond with the specific unit. CT Unit Plastic Air Water Stard Deviation LS 16 (2nd fl) 16 to 126 970 to 10005 5 to 5 2.7 to 4.7 VCT 113 to 123 1001 to 991 5 to 5 2.7 to 4.7 LS 16 (FW) 121 to 131 995 to 985 5 to 5 2.6 to 4.6 Siemens AS 5 to 5 11 to 16 5. If either the mean CT number or the noise (stard deviation) is not within the criteria limit for 3 days in a row or 3 times within a 7 day period, corrective action should be taken by reporting the problem to service representatives the manager. Artifact Evaluation is performed to identify correct artifacts in images of a uniform test phantom before they become severe enough to be detected in patient images. Frequency: Daily Required Use the water phantom provided by the scanner manufacturer are recorded on the Equipment: Form for Daily CT Equipment Quality Control. Record: 1. Use the thin reconstructed images acquired during the daily QC 2. Visually assess the Procedure: images acquired using a window width/level setting of 100/0. Look for rings, streaks, lines, or anything that should not be present in the image. This process can be accomplished by quickly cycling through all artifact images in a stack. Record findings on the data form. Indicate if there are artifacts on the image by writing yes or no in the appropriate place on the data form. 1. Ring artifacts typically indicate detector or data channel imbalance. Repeat the air calibration procedure then repeat artifact images. If the ring artifacts are not corrected after performing repeat air calibration procedures, immediately call service to correct issue. Until service is completed, the QC team should decide if the scanner can be used for patient exams (perhaps on a limited basis depending on the type severity of the artifact). 2. If the difference between the artifact mean CT number (within the ring or circle) the unaffected background mean is greater than 3 HU, service should be called to address the artifact. 3. All images with an artifact should be saved to a library of artifact images of the water phantom. These images should be reviewed periodically to ensure that: a. Subtle artifacts that can be removed by repair are properly addressed b. Subtle artifacts that are present when the scanner is functioning properly

APPROVED BY: Radiology Technical Director Page 4 of 6 are ignored to avoid unnecessary downtime expense. Visual Checklist is performed to ensure the CT system s patient bed transport, alignment system indicator lights, intercom, the emergency cart, room safety lights, signage, monitors are present, working properly, are mechanically electrically stable. Frequency: Required Equipment: Record: Procedure: Monthly This visual checklist must be completed on the first working day of the month. The checklist includes the following: 1. Table height indicator 2. Table position indicator 3. Angulation indicator 4. Laser localization light a. inside light b. outside light 5. Table motion smooth 6. X ray on indicator 7. Exposure switch 8. Display window width/level 9. Panel switches/lights/meters 10. Warning labels present 11. Intercom system functioning 12. Postings present a. X ray on light b. Do not enter x ray in progress 13. Pregnancy notification 14. Service records present Each of the items listed in the visual checklist should pass or receive a check mark. Items that do not pass the visual checklist should be replaced or corrected immediately. Items missing from the room should be replaced immediately. Malfunctioning equipment should be reported to the CT service engineer for repair or replacement as soon as possible Gray Level Performance of CT Scanner Acquisition Display Monitors is performed to ensure that images on the monitors of the CT scanner display the entire range of gray shades produced by the CT scanner. Frequency: Required Equipment: Monthly or whenever a significant change is made to the imager s display monitors. SMPTE Pattern or Equivalent

APPROVED BY: Radiology Technical Director Page 5 of 6 Record: Procedure: 1. Display the test pattern on the imaging console. Set the display window width/level to the manufacturer specified values for the SMPTE pattern. Do not set the window width/level by eye; doing so invalidates this procedure. 2. Examine the pattern to confirm that the gray level display on the imaging console is subjectively correct. The visual impression should indicate an even progression of gray levels around the ring of gray level patches. Verify the following: a. The 5% patch can be distinguished in the 0/5% patch; b. The 95% patch can be distinguished in the 95/100% patch; c. All the gray level steps around the ring of gray levels are distinct from adjacent steps. If these conditions are not met, do not adjust the display window width/level in an effort to correct the problem. action for the monitor is needed. The monitor should be positioned so that there is no glare from room lighting. Most often, the problem is caused by incorrect adjustment of the monitor s brightness contrast. Excessive ambient lighting can cause the problem. Occasionally, components of the display may need recalibration, repair, or replacement. Perform the manufacturer s recommended procedure for monitor contrast brightness adjustment. If there is any doubt about the correct procedure, or if the brightness contrast controls are not accessible, have the qualified medical physicist or service engineer make the adjustments. Note: As part of the periodic maintenance program of the CT scanner, the display monitors of the CT scanner should be checked recalibrated at least annually. Daily QA Procedure 1. The daily QA must be ordered in RIS, use following as a guide. 2. Scheduling 3. Enter Edit 4. Locate Patient (see list below) The following demographics have been set up in RIS for each CT unit: CT Unit Name MRN CT 2 nd floor CT, PHANTOMCTSP SHV13000000 CT64 CT, PHANTOMCT64 SHV90000000

APPROVED BY: Radiology Technical Director Page 6 of 6 CTER CT, PHANTOMCTER SHV11000000 CTFW CT, PHANTOMCTFW SHV12000000 ARTIFACT CT, PHANTOMARTIFACT SHV14000000 Accession # New Exam CWQC Pat Status I Location SHV RAD CAT Modifier COMP History Daily QA Provider Gonzales Toledo Requester Gonzales Toledo Save data 1. Go to technologist work list assign order to appropriate scanner. 2. Select ACR QC protocol scan phantom. 3. Perform QA as required a. Send required images to PACS mark them Reference only b. Save screen axial image with measurements c. Save screen helical image with measurements 4. Track images to complete 5. All data will be recorded in a CT QC logbook. University Health Shreveport s Radiology Department is committed to ensuring that every technologist will execute this quality control program on a daily basis. To that end, enforcement of this policy will be accomplished through education progressive disciplinary action steps: Step 1 Verbal counseling education as determined by the section manager. Step 2 Written reprim Step 3 Written reprim with referral to Human Resources for disciplinary action. Violations of guidelines will be tracked within a 12 month period from the first infraction