Application Note #63 Field Analyzers in EMC Radiated Immunity Testing

Similar documents
Application Note #44 RF Field Probe Selection for EMC Testing

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

4 MHz Lock-In Amplifier

Choosing an Oscilloscope

Revision History. SDG2000X Firmware Revision History and Update Instructions

GFT Channel Digital Delay Generator

PCM ENCODING PREPARATION... 2 PCM the PCM ENCODER module... 4

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1

What to look for when choosing an oscilloscope

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

Broadcast Television Measurements

8650A SERIES UNIVERSAL POWER METERS

Portable Performance for Debug and Validation

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series

2 MHz Lock-In Amplifier

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel)

Logic Analysis Basics

Logic Analysis Basics

Datasheet SHF A

Model 4700 Photodiode Characterizer

OMNISTAR GX2. GX2-LM1000E Series 1310 nm Broadcast Transmitter DATA SHEET BENEFITS. 1 GHz bandwidth

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

Expect to Make Waves.

843-R 843-R LASER POWER METER USER MANUAL. NEWPORT CORPORATION

Advanced Test Equipment Rentals ATEC (2832)

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

Technical Datasheet. 8650B Series Universal Power Meters 10 MHz to 50 GHz. 8651B Single Channel 8652B Dual Channel Rev.

SMART Trigger modes like Glitch, Window and Dropout allow you to capture precisely the events of interest.

Identifying Setup and Hold Violations with a Mixed Signal Oscilloscope APPLICATION NOTE

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor

Boonton 4540 Remote Operation Modes

R&S ZVA110 Vector Network Analyzer Specifications

PEP-II longitudinal feedback and the low groupdelay. Dmitry Teytelman

GFT Channel Slave Generator

LeCroy Digital Oscilloscopes

TROUBLESHOOTING DIGITALLY MODULATED SIGNALS, PART 2 By RON HRANAC

PicoScope 6407 Digitizer

Manual of Operation for WaveNode Model WN-2m. Revision 1.0

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

Oscilloscope Guide Tektronix TDS3034B & TDS3052B

OmniStar GX2 Headend Optics Platform GX2 LM1000E Series

ScopeMeter 190 Series Specifications

Agilent MSO and CEBus PL Communications Testing Application Note 1352

Agilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet

Boosting Performance Oscilloscope Versatility, Scalability

GALILEO Timing Receiver

혼합도메인오실로스코프와 USB 스팩트럼분석기를 이용한광대역노이즈최신측정기술소개

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE Composite Distortion Measurements (CSO & CTB)

WVR500 Waveform/Vector Monitor

Troubleshooting EMI in Embedded Designs White Paper

PicoScope 9300 Series migration guide

Modulated Wideband Power Amplifier

Network Line Card Testing using the TDS3000B DPO Application Note. Line Card Testing Example: Throughput = Shippable Dollars

Reference. TDS7000 Series Digital Phosphor Oscilloscopes

DS2460Q QAM Analysis Meter

Benefits of the R&S RTO Oscilloscope's Digital Trigger. <Application Note> Products: R&S RTO Digital Oscilloscope

MTP200B WLAN / BT LE Tester

USB Mini Spectrum Analyzer User s Guide TSA5G35

R&S EDS300 DME/Pulse Analyzer Specifications

PicoScope 6407 Digitizer

R&S ZND Vector Network Analyzer Specifications

G4500. Portable Power Quality Analyser. Energy Efficiency through power quality

SELECTION GUIDE Series of RF and Universal Frequency Counter/Timers

V6118 EM MICROELECTRONIC - MARIN SA. 2, 4 and 8 Mutiplex LCD Driver

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Agilent Technologies 54522A

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Application Note DT-AN DTU-315 Verification of Specifications

Digital Storage Oscilloscopes 2550 Series

MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer

Immunity testing example using Tekbox TEM Cells

In the development of electronic instruments or the analysis

OLP-87/87P. SmartClass Fiber PON Power Meter and Microscope

N5264A. New. PNA-X Measurement Receiver. Jim Puri Applications Specialist March Rev. Jan Page 1

Therefore, HDCVI is an optimal solution for megapixel high definition application, featuring non-latent long-distance transmission at lower cost.

Building Video and Audio Test Systems. NI Technical Symposium 2008

Major Differences Between the DT9847 Series Modules

ENGINEERING COMMITTEE Interface Practices Subcommittee SCTE STANDARD SCTE

DSA-1. The Prism Sound DSA-1 is a hand-held AES/EBU Signal Analyzer and Generator.

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

Beginners How to Test DSO138mini

SPATIAL LIGHT MODULATORS

Viavi T-BERD 5800 CPRI Testing Guide with ALU BBU Emulation

Operating Instructions

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

Bluetooth Tester CBT. Specifications. Specifications. Version January 2006

DA E: Series of Narrowband or Wideband Distribution Amplifiers

R&S NRPM Over-the-Air (OTA) Power Measurement Solution Specifications

R&S DDF200M Digital Direction Finder Specifications

Getting Started with the LabVIEW Sound and Vibration Toolkit

APM CALIBRATION PROCEDURE Rev. A June 3, 2015

UFX-EbNo Series Precision Generators

Troubleshooting Your Design with the TDS3000C Series Oscilloscopes

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Sources of Error in Time Interval Measurements

PicoScope 4000 Automotive PC Oscilloscopes

EMS DATA ACQUISITION AND MANAGEMENT (LVDAM-EMS) MODEL 9062-C

M R X

Transcription:

Application Note #63 Field Analyzers in EMC Radiated Immunity Testing By Jason Galluppi, Supervisor Systems Control Software In radiated immunity testing, it is common practice to utilize a radio frequency (RF) field probe to measure the electric field at the location of the unit under test (UUT). Most RF field probes are designed to measure the root mean square (RMS) of a continuous wave (CW) electric field over a broad range of frequencies and amplitudes. RF field probes are also commonly used to measure modulated fields. In general, this is not a good practice if the RF field probe is diode based. Depending on the analog response time and the sample rate, measuring a modulated field with a diode based RF field probe will be unpredictable and could yield erroneous results. In some modulated field measurements, such as those found in MIL-STD 461, calibration factors can be obtained to relate modulation amplitude to CW amplitude. These calibration factors will only be good for the specific amplitude and modulation scheme that was used to obtain them. In addition, this will only work well with a slow modulated field where the analog response time of the RF field probe is also slow. Unfortunately, most modern diode based RF field probes have a reasonably fast analog response time. This coupled with a slow sample rate will cause measurements to be unstable. In this case, averaging many readings over time can be used to obtain a stable reading. In applications where a narrow pulsed field is to be measured, most diode based RF field probes cannot be used. These RF field probes will see the pulses, but not sample fast enough to consistently measure them. Thermocouple RF field probes offer a solution to some of the above listed problems, however they can only provide an average field measurement. Since the maximum RMS field level is what is typically desired, an additional computation is required. In order to perform this computation, prior knowledge of the modulation is necessary. Additionally, this type of RF field probe does not allow the user to see a visual representation of the modulation envelope. This leaves questions about the validity of the modulation as seen by the UUT. Being able to see the modulation envelope is extremely useful for verification of the waveform or uncovering any distortion that may be caused by items in the signal generation path. For example, distortion can occur with pulsed RF signals that are amplified by an amplifier that is driven into saturation or into a poor load. As a result, overshoot, droop, or ringing can be introduced. To address the problems and inconsistencies listed above, devices called Field Analyzers were developed. Field Analyzers are a relatively new type of RF field probe. These patent pending 160 School House Road Souderton, PA 18964-9990 215-723-8181 www.arworld.us 1 of 5

devices have the ability to measure modulated electric fields in addition to CW electric fields. They allow the user to view and measure the electric field in the time domain using an oscilloscope-type display and interface. Table 1 shows a comparison between Field Analyzers and the above listed conventional RF field probes. Field Probe (Diode) Field Probe (Thermocouple) Field Analyzers (Diode) CW Field (RMS) CW Field (Peak) Calculable from RMS value Calculable from RMS value Calculable from RMS value AM Field (Average RMS) AM Field (Maximum RMS) AM Field (Minimum RMS) AM Field (Peak) Calculable from Maximum RMS value Pulsed Field (Average RMS) Pulsed Field (Maximum RMS) Calculable from Average RMS value* Pulsed Field (Minimum RMS) Pulsed Field (Peak) Calculable from Maximum RMS value* Calculable from Maximum RMS value Pulsed Field (Rise Time) Pulsed Field (Period) Pulsed Field (% Duty) * Only if % duty is known Table 1. Comparison of RF Field Probe Measurement Capabilities The AR designed and manufactured FA7000 series of Field Analyzers (see Figure 1) utilize an isotropic field sensor to sample the composite field and transmit its amplitude digitally over optical fiber to a processor unit. The sample rate of these devices is significantly faster than conventional RF field probes allowing them to accurately measure pulsed electric fields in the micro-second range. Table 2 lists some common modulations per various standards that can be measured with the FA7000 series of Field Analyzers. Field Sensor Field Sensor Processor Unit Figure 1. FA7000 Series of Field Analyzers ( Photos not to scale) EMC Standard IEC 610000-4-3 ISO 11452 ISO 11452 MIL STD-461 MIL STD-461 Modulation 4% Duty 1kHz Pulse 90% Depth Square Wave 1kHz 4uS Pulse 1kHz Repetition Rate 80% Depth 1kHz AM 3uS Pulse 80% Depth 1kHz AM Peak Conservation 50% Duty 1kHz Pulse 50% Duty 400Hz Pulse Table 2. Common Modulations and their Associated EMC Standard 2 of 5

The processor unit is used to buffer the digital samples and host an embedded webpage where the samples can then be viewed in an oscilloscope-type display. The embedded webpage can be viewed on any web-enabled device such as a personal computer, tablet computer, or smart phone. This webpage contains a number of familiar controls that allow the user to adjust the way the modulation envelope is displayed. It also provides feedback about the Field Analyzer system and provides some useful calculations based on the displayed waveform. Like the FL7000 and PL7000 series of field probes designed and manufactured by AR, all of the field sensors in the FA7000 series of Field Analyzers are laser powered. This allows for continuous testing without ever having to recharge or replace batteries. Figure 2. FA7000 Series Embedded Webpage Figure 2 shows a screen shot of the embedded webpage for the FA7000 series of Field Analyzers. At the top of the page is the waveform display in a 10 by 10 grid configuration. Below the waveform display is a set of controls for adjusting how the modulation envelope is displayed. Similar to a standard oscilloscope, the two main controls are scale and time base. In addition, there is a control to select the trigger method. When internal trigger mode is selected, additional controls for threshold level and edge become available. 3 of 5

In free run trigger mode a trigger event is automatically set each time the waveform display refreshes. This is a good initial mode to obtain a quick view of the electric field if one does not know the specifics of the amplitude or modulation of the electric field being measured. Internal trigger mode is essentially a conventional level trigger in which the user can select a threshold field level for which the refresh of the display will occur. The direction for the threshold trigger is also user selectable as either rising or falling edge. The waveform will appear with the threshold crossing event in the center of the waveform display window providing the user with 50% pre-trigger data and 50% post trigger data. An external pulse signal can also be used to create a trigger event. This is particularly useful for single shot events or when synchronization with a signal generator is necessary. Most signal generators can provide a compatible trigger signal that can be connected to the external trigger input found on the rear panel of the processor unit. To correct for deviation of the field sensor s frequency response from a flat response, the user can load a table of correction factors (found in the calibration report) for up to 30 different frequencies. This is done using the Table Loader program found on the AR website. Once correction factors have been loaded, the embedded webpage can use them to correct the displayed waveform based on a user specified carrier frequency. The calculated multiplier correction factor is displayed so the user understands the amount of correction being applied. Linear interpolation is used between frequency points in the table. If the user specified carrier frequency is beyond the end points of the correction factor table, a warning message is displayed. The correction factor table can also be viewed directly from the embedded webpage by using the View Table button. Below the waveform controls are indicators for the minimum, maximum and average field level. These are all calculated based on the waveform as it is displayed. To the right is a set of buttons that control the refresh of the display regardless of trigger events. One button is a simple Run/Stop button while the other allows the user to capture a single event. Both of these controls are useful for closer examination or printing of the waveform. The bottom of the webpage contains useful information about the Field Analyzer system. This includes model, serial number, and firmware revision information along with the status of the processor unit s front panel controls and indicators. Field Analyzers can also be used where automated software such as the AR emcware (formerly the SW1007) controls all of the test equipment in an RF radiated immunity test setup. The minimum, maximum, and average field levels are easily obtained through any of the standard remote ports (USB, GPIB, and Ethernet) found on the rear panel of the processor unit. A set of backward-compatible remote queries are also available for easy substitution with any FL7000 or PL7000 conventional RF field probes. This allows for compatible connectivity with an AR FM7004 Field Monitor. All of the waveform display controls are also accessible remotely which allows the user to setup to measure a specific type of modulation. When a Field Analyzer is used in this way, frequency correction factors must be applied by the automated software controlling the test. 4 of 5

The field sensors in the FA7000 series of Field Analyzers are designed to be isotropic. However, they each have a specified deviation from isotropic. This means that, like conventional RF field probes, the best accuracy is achieved by positioning the field sensor with a single antenna element aligned with the electric field. Each field sensor has indicators showing where the individual antenna elements are located. For example, the stalk style field sensors indicate the tips of each of the three dipole antenna elements. Since Field Analyzers are relatively new devices, all of their potential applications have yet to be discovered. Their fast sampling allows them to provide more information and better accuracy than a conventional RF field probe, which in turn, allows for a better understanding of the amplitude of the electric field in the time domain. This innovative instrument measures modulated electric fields that resemble actual real-world situations, thereby giving the user a more correct and more accurate measurement. 5 of 5