An Improvised Two-Dimensional Laser Surface Scanner for Diagnosis of Rf Thermionic Electron Gun Problems

Similar documents
Advanced Photon Source - Upgrades and Improvements

These are used for producing a narrow and sharply focus beam of electrons.

INSTRUMENT CATHODE-RAY TUBE

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

The Cathode Ray Tube

Mechanical aspects, FEA validation and geometry optimization

Performance of a DC GaAs photocathode gun for the Jefferson lab FEL

OF THIS DOCUMENT IS W8.MTO ^ SF6

2.2. VIDEO DISPLAY DEVICES

In-process inspection: Inspector technology and concept

Design Studies For The LCLS 120 Hz RF Gun Injector

Characterizing Transverse Beam Dynamics at the APS Storage Ring Using a Dual-Sweep Streak Camera

Development of Multiple Beam Guns for High Power RF Sources for Accelerators and Colliders

Screen investigations for low energetic electron beams at PITZ

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Display Systems. Viewing Images Rochester Institute of Technology

CMPE 466 COMPUTER GRAPHICS

CATHODE-RAY OSCILLOSCOPE (CRO)

INSTRUMENT CATHODE-RAY TUBE

Physics Requirements for the CXI Ion Time-of-Flight

Design and Simulation of High Power RF Modulated Triode Electron Gun. A. Poursaleh

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

CATHODE RAY OSCILLOSCOPE (CRO)

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

This work was supported by FINEP (Research and Projects Financing) under contract

Quick Start Bruker Dimension Icon AFM

Capstone Experiment Setups & Procedures PHYS 1111L/2211L

Overview of Graphics Systems

High QE Photocathodes lifetime and dark current investigation

Computer Graphics Hardware

OSCILLOSCOPE AND DIGITAL MULTIMETER

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

Illumination Challenges in Non- Industrial Vision Applications. Simon Stanley Managing Director ProPhotonix IRL Ltd

ABRS Series. Air-Bearing Rotary Stage. Direct-drive, slotless brushless servomotor. Zero cogging motor for outstanding velocity stability

Computer Graphics: Overview of Graphics Systems

Elements of a Television System

CHAPTER 4 OSCILLOSCOPES

SLAC R&D Program for a Polarized RF Gun

Design, Fabrication and Testing of Gun-Collector Test Module for 6 MW Peak, 24 kw Average Power, S-Band Klystron

Connection for filtered air

Recent APS Storage Ring Instrumentation Developments. Glenn Decker Advanced Photon Source Beam Diagnostics March 1, 2010

TESLA FEL-Report

Lecture Flat Panel Display Devices

THE OPERATION OF A CATHODE RAY TUBE

Operating Experience and Reliability Improvements on the 5 kw CW Klystron at Jefferson Lab

B. TECH. VI SEM. I MID TERM EXAMINATION 2018

Lecture 17 Microwave Tubes: Part I

Introduction. Characteristics (Note 1, 2) Parameters Typical Value Units Conditions

Standard Operating Procedure of nanoir2-s

General Specifications

Linac-Beam Characterizations at 600 MeV Using Optical Transition Radiation Diagnostics *

Part 1: Introduction to computer graphics 1. Describe Each of the following: a. Computer Graphics. b. Computer Graphics API. c. CG s can be used in

An Introduction to Vibration Analysis Theory and Practice

OM2000N INSTALLATION MANUAL

3B SCIENTIFIC PHYSICS

Screens; media that use additive primaries

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE

Characterization and improvement of unpatterned wafer defect review on SEMs

Digital SWIR Scanning Laser Doppler Vibrometer

Summary report on synchronization, diagnostics and instrumentation

Evaluation of Performance, Reliability, and Risk for High Peak Power RF Sources from S-band through X-band for Advanced Accelerator Applications

TITLE PAGE. Title of paper: PUSH-PULL FEL, A NEW ERL CONCEPT Author: Andrew Hutton. Author Affiliation: Jefferson Lab. Requested Proceedings:

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

New Filling Pattern for SLS-FEMTO

Electron Beam Technology

L14 - Video. L14: Spring 2005 Introductory Digital Systems Laboratory

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

THE OPERATION OF A CATHODE RAY TUBE

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

Accelerator Instrumentation RD. Monday, July 14, 2003 Marc Ross

Production of quasi-monochromatic MeV photon in a synchrotron radiation facility

Computer Graphics : Unit - I

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING

Development of an Abort Gap Monitor for High-Energy Proton Rings *

Advancements in Acoustic Micro-Imaging Tuesday October 11th, 2016

EMC-Scanner. HR-series

DEEPFRAME BASIC KIT- USER MANUAL VERSION ORIGINAL USER MANUAL

Digital SWIR Scanning Laser Doppler Vibrometer

SPEAR 3: Operations Update and Impact of Top-Off Injection

TWO BUNCHES WITH NS-SEPARATION WITH LCLS*

LCLS Injector Technical Review

Types of CRT Display Devices. DVST-Direct View Storage Tube

Digital BPMs and Orbit Feedback Systems

PROJECT DESCRIPTION. Longitudinal phase space monitors for the ILC injectors and bunch compressors

Jefferson Lab Experience with Beam Halo, Beam Loss, etc.

TECHNICAL SPECIFICATION Multi-beam S-band Klystron type BT267

PEP-II IR-2 Alignment

Three Axis Digital Readout System

Magnetic Force Microscope (MFM)

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Progress Update FDC Prototype Test Stand Development Upcoming Work

PEP-II Overview & Ramp Down Plan. J. Seeman DOE PEP-II Ramp Down-D&D Review August 6-7, 2007

Display Devices & its Interfacing

Academic and Research Staff. Prof. John G. King Dr. John W. Coleman Dr. Edward H. Jacobsen. Norman D. Wittels

2.4.1 Graphics. Graphics Principles: Example Screen Format IMAGE REPRESNTATION

Figure 1. MFP-3D software tray

F7000NV ROBOT VISION OPERATING MANUAL

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

PHI 5000 VersaProbe TM Operator s Guide

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Transcription:

An Improvised Two-Dimensional Laser Surface Scanner for Diagnosis of Rf Thermionic Electron Gun Problems William G Jansma, Engineering Specialist Advanced Photon Source Argonne National Laboratory Work supported by the U.S. Department of Energy, Office of Science, Office of Basic Energy Sciences, under Contract No. DE-AC02-06CH11357

WELCOME! Thank You For Attending This Morning ARGONNE NATIONAL LABORATORY ADVANCED PHOTON SOURCE BACKGROUND PROBLEMS ARISE MEASUREMENT IMPROVISED 2-D LASER SURFACE SCANNER DISTORTION REVEALED ALTERNATIVES REPAIR OF THE RF GUNS CONCLUSION ACKNOWLEDGMENTS 2

CMSC 2011 July 25 29, 2011 Phoenix Arizona First National Laboratory in the United States, chartered July 1, 1946 Argonne National Laboratory is a direct descendant of the University of Chicago s Metallurgical Laboratory, where, on Dec. 2, 1942, Enrico Fermi and his band of about 50 colleagues created the world s first controlled nuclear chain reaction. The U of C Metallurgical Lab was part of the Manhattan Project. A multi-disciplinary facility operated by UC-Argonne, LLC for the U.S. Department of Energy, the laboratory's current R&D programs focus on basic science research, energy storage and renewable energy, environmental sustainability, and national security. 3

Advanced Photon Source (APS) Third-Generation Synchrotron X-ray Source Premier National User Facility in the Western Hemisphere hosting ~3,500 Scientists Per Year Multi-Disciplinary Laboratory: Accelerator and X-ray R&D Biological Science Materials Science Chemical Science Geoscience Environmental Science Nanoscience (Center for Nanoscale Materials) Pharmaceutical Research Atomic, Molecular and Optical Physics CMSC 2011 July 25 29, 2011 Phoenix Arizona 4

BACKGROUND Electron guns are devices that produce, accelerate and emit electrons. The most common type of electron guns are found in the cathode ray tubes (CRT) of older televisions and computer monitors. Other types of electron guns are used as sources for electron microscopes, particle accelerators and electron beam welders. Radio frequency (rf) thermionic electron guns were developed at SLAC National Accelerator Laboratory in the late 1980s and early 1990s. Various generations of these guns have been used to inject electrons into the APS accelerator system since 1996. 5

BACKGROUND Electrons are produced in rf electron guns by heating a tungsten alloy cathode to 800 o C to 1200 o C, boiling off electrons directly into an rf cavity energized by microwaves with a frequency of 2856 MHz and a 10 cm wavelength. The electrons are accelerated by the microwaves, analogous to the way a surfer is pushed along by waves in the ocean. CMSC 2011 July 25 29, 2011 Phoenix Arizona Electrons emitted from the rf guns are injected into the APS accelerator system, forming a beam of high energy electrons that is further accelerated to velocities approaching the speed of light. When the beam is deflected by magnets in the accelerator system x-rays are emitted. With the bright, penetrating high energy x-rays scientific instruments can see deep into the atomic structure of matter, allowing scientists to carry out cuttingedge research that positively impacts our lives. 6

Problems Arise In 2001 three new rf thermionic electron guns were purchased forthe APS. In 2009, after 8 years of operation, the ageing cathodes were replaced. Problems arose immediately after the cathodes were changed out. More heater power (25 W, as opposed to the normal 15 W) was required, and excess reflected power was observed. Abnormally high reflected power levels subject the cathode surface to excess back bombardment, with toomany of the electrons being accelerated in the reverse direction. Two rf guns failed in mid 2010 due to excess reflected power. Spare guns were installed, however there was much concern that another rf gun failure could limit the capabilities of the machine and, in the worst case, shut down APS operations altogether. An investigation was initiated to discover the cause of the problems observed in the rf guns. Inspection revealed distortion of the cathode cartridges, leading to suspicion that the mating surfaces within the gun bodies werealso distorted. The APS Survey and Alignment section was asked to assist in the diagnosis and measure the mating surfaces of the gun bodies to identify and characterize the suspected distortions. 7

Measurement Cathode mating surface difficult to reach located at a depth of 75-mm within a 60-mm dia. vacuum chamber section Initial measurement attempt using a Faro arm provided useful data for larger features, but resolution was insufficient to profile the mating surface distortion. APS did not have an instrument capable of profiling the surface at the needed resolution. Several types of profiling instruments were available, both contact and optical; none ideal for this problem, expensive.$20k plus. Improvised measurement system devised after much thought. 8

IMPROVISED 2-D LASER SURFACE SCANNER Improvised two-dimensional laser surface scanner set up to measure an rf gun body System employs two Keyence LK-G series CCD laser displacement sensors configured to measure two orthogonal axes. Scanning sensor is mounted to an Indi Square device for vertical translation. Synchronous data output from both sensor heads allowing coordinates to be ascertained by combining the data in a spreadsheet. Keyence LK-G sensors emit a laser signal that is reflected from any surface back on to an integrated CCD array that registers displacement along the beam axis. 9

Laser Displacement Sensors LK-G 152 and LK-G 157 sensors were used for the improvised scanner. Range is 150 mm +/-40 mm at 0.5 micron resolution. Sensors were connected to a single controller unit integrated with a PC or laptop. Measurement data recorded using Keyence LK-Navigator freeware. Calibration was verified by measuring several gage block lengthsover the range of the sensor heads. Sensors may be configured for diffuse or specular surface reflectivity and are capable of measuring through a transparent medium such as glass. 10

Vertical Translation Sensor head mounted to an Indi-Square model 18 for vertical translation. Translation accuracy is 0.0025 mm over 500 mm of travel. Designed to accept a typical dial indicator head. Sensor head weighs 290 grams, not including aluminum mount bracket. Counterweight was needed to balance system. Translation by hand not satisfactory. Crankwheel height gage used to translate sensor head. Despite added weight and unorthodox translation method, the setup was smooth and precise. CMSC 2011 July 25 29, 2011 Phoenix Arizona 11

COMPONENT INTEGRATION Initially a LK-G157 sensor head was mounted vertically to the translation, but the return signal was blocked when the laser spot was transitioned over steps in the surface. The problem was solved by horizontally mounting the sensor head to the translation. The LK-G 157 sensor head projects a 1.7-mm-wide laser spot, recording an average profile over the width of the laser spot. To improve accuracy a LK-G 152 sensor head that projects a 120 micron diameter laser spot was mounted to the translation, allowing a much thinner measurement slice to be recorded. 12

ORIENTATION & MEASUREMENT Rf gun bodies were rested on v-blocks that in turn rested on precision lab jacks. Primary surface to be scanned was placed at the optimum distance, 150 mm, from the sensor head. Gun bodies were set parallel to the translation by adjusting the pitch angle of the gun until the sensor display indicated equal values within +/-10 µm when translated across a reference plane surface within the gun. To obtain limited 3-D information scans were taken across azimuth angles of 0, 45, 90, and 315 degrees. Common reference was plane was used to allow scans across different angles to overlay each other CMSC 2011 July 25 29, 2011 Phoenix Arizona 13

DATA COLLECTION & PROCESSING Data collection was accomplished using the LK-G3001 controller and laptop with LK-Navigator freeware. Data was saved as a.csv text file that opens as a single columnin a spreadsheet that is cut and pasted into a coordinate worksheet. Synchronous controller output allowed the columns to be pasted side by side in the worksheet forming coordinates. Coordinates for each azimuth scan were adjusted to fit a common centerline and plotted over each other to produce an abstract 3-D view in a 2-D graph. 14

ERROR ANALYSIS Accuracy initially estimated to be +/-5 microns, taking into account the sensor resolution of 0.5 microns, translation accuracy of 2.5 microns over 500 mm, and the typical scan translation of only 25 mm. Further analysis indicates the accuracy of measurements taken with the improvised system to be closer to +/-15 microns. Decrease from the initial estimate is attributed to noise produced by the motion of the sensor head during translation. Accuracy might be improved through refinements in motion controlfor the translation. CMSC 2011 July 25 29, 2011 Phoenix Arizona 15

DISTORTION REVEALED Cathode Cartridge Distortion Distortion was caused by over-tightening of eight #4-40 screws that secure the cartridge within the gun body. Damage was inadvertent, as no drawings or procedure were available. Source of problem was discovered, but what now? Plots of the profile scans taken with the improvised 2-D laser scanner revealed distortion of the mating surfaces of both the cathode cartridges and the rf gun bodies. The flat, planar mating surface is intended to seal the rf cavity from leakage and act as a thermal conductor. Arcing can result from rf energy leakage, and improper heat transfer affects thermal stability. 16

REPAIR OF THE RF GUNS Several alternatives were explored in the effort to solve the problem of the damaged rf guns including replacement, modification, and repair. Replacement cost is $70,000, and modification would be costly as well. Decision was made to attempt to repair the damaged guns. Special retraction tool was designed and tested successfully on a mockup. Repair was performed, using the improvised 2-D scanner to monitor the progress of the repair effort. Gun body was ultrasonically cleaned, and a new cathode cartridge was installed using a modified retaining system suggested by the author. The repaired rf gun was placed in the APS injector test stand, activated, and rigorously tested. In March 2011 it was installed in one of the APS injector stations and has been functioning normally. 17

REPAIR OF THE RF GUNS In April 2011 the second rf gun was successfully repaired using the same process and was installed into the other APS injector station in July. The successful repair of two rf thermionic electron guns resulted in significant savings of both time and money for the APS. The expense of re-engineering the gun / cathode assembly was spared, and the purchase of a replacement rf gun was rendered unnecessary. In addition, by improvising a measurement system using equipment on hand, the cost of purchasing a new instrument to identify and characterize the rf gun distortions was saved. 18

CONCLUSION The improvised 2-D laser surface scanner proved to be a valuable instrument, helping to both identify and repair damage in two rf thermionic electron guns. Although the instrument is not a revolutionary development, it shows that problems can occasionally be solved by using imagination and applying the tools on hand in seeking a solution. In this case an improvised application of coordinate metrology tools and methods were used to revive two damaged rf thermionic electron guns, resulting in substantial cost savings for the APS. 19

ACKNOWLEDGMENTS The author would like to thank the following people for their contributions to this endeavor: Emil Trakhtenberg, Nick Sereno, Bran Brajuskovic, StanPasky, Randy Flood, John Hoyt, Wayne Michalek, Terry Smith and Scott Wesling (all APS); Jose Arroyo (talktoanit.com); Jeff Huerta (Keyence Corporation) and HeatWave Labs, Inc. 20