EXA PH200/400 and EXA PH202/402 Troubleshooting and Error Code Guide

Similar documents
Operation Manual TX20 1/8 DIN Microcomputer Based ph/orp Controller with relays

This document courtesy of:

USER MANUAL MODEL 650. TWO-WIRE ph/orp TRANSMITTER. um

Instruction Manual Aquamon 2000 plus 2-Wire ph/orp Transmitter

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630)

Sensopress LCD Special English

OPERATING AND SAFETY INSTRUCTIONS for DIGITAL TEMPERATURE CONTROLS (PLSM SERIES)

Sartorius Professional Meter

Instruction Manual Alpha ph 190 1/8 DIN ph/orp Controller Transmitter ph / mv / ºC

TX2000 Intelligent ph/orp Transmitter

ISC202 Quick Start Guide

Registers type ILR7XX and ILR7XXT

PACSystems* RX3i. Isolated Thermocouple Input Module, 6 Channels, IC695ALG306-EB Isolated Thermocouple Input Module, 12 Channels, IC695ALG312-EB

THE ASTRO LINE SERIES GEMINI 5200 INSTRUCTION MANUAL

Signet Chlorine Transmitter

Noise Detector ND-1 Operating Manual

TRANSCENSION 6-CHANNEL DMX DIMMER PACK (order code: BOTE40) USER MANUAL

ANALOG I/O MODULES AD268 / DA264 / TC218 USER S MANUAL

PACSystems* RX3i Thermocouple Input Module, 12 Channels, IC695ALG412-CB

PanelView 1400e CRT Maintenance

Data Acquisition Networks. Installing and Configuring the DM01 Hardware

Instruction Manual MODEL RSP SANITARY ELECTRONIC PRESSURE TRANSMITTER

802DN Series A DeviceNet Limit Switch Parameter List

SONARtrac Technical Bulletin

MT03A Electronic converter for flow rate transmitters

Electronic converter for level transmitters MT03L Instructions manual

WELDING CONTROL UNIT: TE 450 USER MANUAL

Good Laboratory Practices (GLP) Notepad. Automatic Datalogging. Automatic Stability Function. P.I.N. Code Security Access.

MG-XV operating instruction. Measuring of norm signals, 4-8-digit. Panel instrument type MG-BV Construction instrument type MG-AV

FX-2DA SPECIAL FUNCTION BLOCK USER'S GUIDE

LX3V-4AD User manual Website: Technical Support: Skype: Phone: QQ Group: Technical forum:

N3ZI Digital Dial Manual For kit with Serial LCD Rev 3.04 Aug 2012

3214NXT. Service Manual. IMPORTANT: Fill in Pertinent Information on Page 3 for Future Reference

DPM 942-FPSI 4-20mA Loop Meter with Programmable Backlighting

OPERATING AND SAFETY INSTRUCTIONS for DIGITAL TEMPERATURE CONTROLS (PLSM SERIES)

TVAC20000 User manual

Aeroforce FAQ. 2. Before I purchase, how do I know what parameters will be supported on my particular vehicle?

Sigma 1 - Axis Servo Motor and Cables - Troubleshooting Guide

BASIC PH METER BASIC PH METER KIT

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

Innovation in Magnetic Measuring Instruments. Operation Manual for. Mag648 and Mag649 Low Power Three-Axis Magnetic Field Sensors.

INSTALLATION MANUAL. ST-CVTMD420-WPIR-W Covert Motion Detection Color Camera. v1.3 8/11/11 1

MDA-2-HV. 1x2 Moblie Distribution Amplifier. Instruction Manual

LAUREL ELECTRONICS, INC.

Metal Electrode Meter

CIRCULATOR & ISOLATOR APPLICATION NOTES: IM CONTROL PANEL APPLICATION NOTES: CAVITY RESONATOR APPLICATION NOTES:

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

WQ Series Water Quality Bench top Meters

CARLO GAVAZZI Automation Components

ABB Instrumentation Operating Instructions Series Transmitters. Second Retransmission Output Supplement All Models.

Troubleshooting CS800/LC900 Bikes

Spectra Batten (Order code: LEDJ95)

STX Stairs lighting controller.

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

Preventing Fieldbus Physical Layer Problems

User Manual CC DC 24 V 5A. Universal Control Unit UC-1-E. General Information SET. Universal Control Unit UC-1 Of Central Lubrication PAUSE CONTACT

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E

Wired Troubleshooting Manual

INST No.CP CP Computing unit. Instruction Manual. Important Notice. Keep this manual with due care after understanding operation.

USER MANUAL Model: A5

PCR AUTOMATIC HEAT TREATMENT CONSOLE

FX-4AD-TC SPECIAL FUNCTION BLOCK USER'S GUIDE

INSTALLATION MANUAL. ST-CVTSD520-WSD-W Smoke Detector Covert Camera. v1.2 8/11/11 1

Sprite TL Quick Start Guide

Model IQ4-PC User Manual Revision Date:

TABLE OF CONTENTS. Instructions:

POET-1 P.O.E. TEST PORT MEASUREMENT TOOL INSTRUCTION BOOK

AES-402 Automatic Digital Audio Switcher/DA/Digital to Analog Converter

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

INT-X-P and INT-X-L Programming instructions

XCOM1002JE (8602JE) Optical Receiver Manual

Advanced Digital Melting Point Apparatus

MITSUBISHI ELECTRONICS AMERICA, INC.

OPERATION MANUAL MODEL DF3000 DIGITAL WEIGHT INDICATOR REVISION A-1

Owner s Manual. Backup Monitor System. LCD Monitor & CCD Color Camera

Operating Instructions

Global Water Instrumentation, Inc.

Triple RTD. On-board Digital Signal Processor. Linearization RTDs 20 Hz averaged outputs 16-bit precision comparator function.

Monitor QA Management i model

TABLE OF CONTENTS. 1) Introduction 2. 2) Unpacking the Ares 2. 3) Installing the Ares in your system 3. 4) Setting the Operational Parameters 4

DGC-1000A CLAMP-ON GROUND RESISTANCE TESTER

ph GENERAL INFORMATION

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1

Operating Instructions BTX-1 Series Digital Band Tension Meter

N3ZI Digital Dial Manual For kit with Backlit LCD Rev 4.00 Jan 2013 PCB

Instruction Manual ph/ion 510 Bench ph/ion/mv Meter

Operation/Users Manual

OPERATIONS MANUAL. ba76140e01 02/2013 CONDUCTIVITY CELL FOR ULTRAPURE WATER

ROUGH DRAFT. Guide. Installation. Signal Booster. Wilson. AG Pro 75 Smart Technology In-Building Wireless 800/1900 Signal Booster.

This book provides the information you need to quickly set up and use your instrument. It is a great reference guide.

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

ULTRA-TRAC APL INSTRUCTION MANUAL. Read and understand instructions before use. Patented. 851 Transport Drive Valparaiso, IN

Revision 1.2d

Troubleshooting Guide for the ESP MC controller Feb Royce Mason

INSTRUCTION MANUAL COMMANDER BDH MIG

SI-510 DUAL TRIP AMPLIFIER

Amplifier for fiber optics. Dimensioned drawing

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS

Transcription:

EXA PH200/400 and EXA PH202/402 Troubleshooting and Error Code Guide Introduction The EXA Series of Instruments (EXA PH200, PH400, PH202, PH402) provide much more than just a measurement. They are also a tool, providing both on-line and off-line diagnostics of the components making up the measurement loop (sensors, cabling and Analyzer or Transmitter). These diagnostics serve two key functions: 1) Status Information: In the DISP (display) section of the Message Display, information concerning the status of key components is available. As an example, the Impedance value of the Reference Electrode can be checked by using the YES/NO keys and stepping through the DISP section until the RZ value is reached. 2) Failure Alarm: The diagnostic program continually checks key functions and components of the measurement loop for problems and provides information to the operator via Error Codes on the dual LCD display. These Error Codes are designed to lead the operator to the source of the problem as quickly as possible and offer suggestions for correction. To properly utilize the information provided by these Error Codes the user should first determine the following: 1) Under what circumstances the Fault occurred: Was it at start-up, or had the unit been on-line for awhile? This will help determine if the problem is related to a new installation or whether an electrode has been on-line long enough that it may need replacement. 2) What Sensor mounting assembly and specific electrodes are being used: Is the best electrode for the application being used? Often an electrode that works fine in one application, is moved to an entirely different one for which it was not designed and consequently does not perform well. The misapplication is the problem, but the equipment is incorrectly blamed and often replaced unnecessarily. 3) The Process Parameters, including temperature, chemical make up and whether it is a batch or continuous operation: Sometimes the actual process is changed. When this happens problems often begin to occur with ph equipment. By redefining the process, it is usually a simple matter to discover the reason for the current problems. By changing to different electrodes, modifying the installation or moving to a different measuring point will resolve the problem.

Start-Up Problems The majority of problems experienced by new EXA users are a result of not taking enough time during the initial installation to go through the Instruction Manual completely before setting up their unit and important steps may be omitted causing Error Codes which are mistakenly attributed to defective equipment. As an example, if an EXA user does not make sure the correct Temperature Electrode is selected (programmed) an E7 or E8 code may be displayed. For ph users, if the Measuring and Reference Electrode wiring are accidentally switched, an E4 and an E6 code will be displayed simultaneously. Therefore, the diagnostics (and subsequent Error Codes) can assist a new user in determining that the unit has not been properly set up and in which area the problem can be found. Routine Maintenance Error Codes The EXA give users two key maintenance tools. First, they provide status information about the health of the Temperature or Measuring electrodes. This information is found under DISP in the Second Display area. The EXA display s the Asymmetry Potential (Reference Electrode) and Slope value (Glass Electrode) measured at the last 2- point calibration. In addition, the Impedance of the Glass ( 202/402 only) and the Reference Electrode Junction (dynamic value) is displayed. Utilizing this Diagnostic information allows the user to change an aging electrode before it actually fails, or adjust the maintenance routine (cleaning) to assure an accurate measurement. The following pages list the "Probable Causes" for each Error Code in order from most common possibility for the Error to the least probable cause. Note in all but a few cases the error codes used by the EXA 200 and 400 are the same as used by the EXA202 and the 402. However since the newer EXA202 and 402 has enhanced sensor diagnostics, there are some Error Codes used by the EXA202 and 402 which are different from the EXA 200 and 400. These unique to the EXA202 and 402 Error Codes are so identified to remove any confusion for the reader. E0: Temperature of Buffer is outside the range of 0º - 50 C - (Calibration Error) This code occurs only when an Auto Cal is being done. The EXA compares the measured temperature of the buffer being used to the acceptable limits of 0-50 C (this is the range of the Buffer Tables used for Auto Cal). If the buffer temperature is outside these limits an error code of EO is displayed in the second display area and the calibration will not be accepted. 1) Buffer temperature outside acceptable range. 2) EXA set up for wrong Temperature Electrode (set for Pt 100 but is actually a Pt 1000). 3) Improper wiring or Temperature Electrode Cable defective. 4) Temperature Electrode defective.

1) Correct Buffer Temperature (cool or heat buffer to acceptable value). 2) Go to Service Section (Code 03) and enter code for correct Temperature Electrode. 3) Check wiring connections and replace Electrode Cable if needed. 4) Replace Temperature Electrode. E1: Measurement has not stabilized during calibration - (Calibration Error) This code occurs after an Auto Cal or Man Cal has been attempted. The EXA compares the programmable values for (1) allowable stabilization time (Factory Default is 5 seconds ) and (2) allowable variance of the measured value (Factory Default is.02 ph) to those of the electrodes being calibrated. If the buffer ph value does not stabilized within the acceptable range within the time allotted, an error code E1 will be displayed in the second display area and the calibration will not be accepted. 1) Contaminated or old buffer, or dirty electrodes. 2) Programmed Stabilization is too short, or Programmed Range is too narrow. As electrodes age they become sluggish and slower to respond to ph changes. If an older electrode is being used, the EXA will need to be reprogrammed with a longer allowable response time or an E1 error will continue to occur. 3) Measuring Electrode (Glass) is too old and slow to respond. 4) Improper wiring or one or more Electrode Cables are faulty causing noise on the measurement signal. 1) Get fresh buffer and clean ph electrodes in 5-10% HCl solution or Muratic Acid. 2) Reprogram Stabilization Time ONLY if an aged electrode must be used. Otherwise change the aged electrode for new faster responding one. 3) Replace Measuring Electrode. 4) Check wiring connections and replace Electrode Cable if needed. E2: Calculated Asymmetry Potential is outside acceptable range: -120 to +120 mv - (Calibration Error) This code occurs after an Auto Cal or Man Cal has been attempted. The EXA compares the calculated Asymmetry value against the acceptable limits of - 120 mv to +120 mv. Asymmetry is the mv offset value required to get back to 0 mv. The closer to 0 mv a Reference Electrode is, the better it is. New electrodes have values of ± 15 mv. As they age and are affected by the

process this value will move further away from zero in either a positive or negative direction. As this occurs, response becomes slower and more frequent calibration will be required. If the electrode is outside the designated -120 mv to +120 mv range, an error code E2 is displayed in the second display area and the calibration will not be accepted. 1) Bad Buffer 7 (zero buffer). 2) Reference Electrode is too old and out of acceptable limits. 3) Reference Electrode is poisoned or damaged. 4) Bad wiring or Electrode Cable is faulty (bad isolation). 5) Solution Ground Electrode (Post) is coated, dirty or connection is faulty. 1) Clean ph electrodes in 5-10% HCl solution or Muratic Acid. 2) Replace Reference Electrode. 3) Replace Reference Electrode 4) Check wiring connections and replace Electrode Cable if needed. 5) Clean Solution Ground, check wiring connection and replace if faulty. E3: Calculated Slope is outside acceptable theoretical range: 70% to 110% - (Calibration Error) This code occurs when either Auto Cal or Man Cal is being done. The EXA compares the calculated Slope value against the acceptable theoretical Slope values of 70% to 110%. New Measuring Electrodes have a Slope value between 95 and 99%. As they age this value decreases and the response of the electrode becomes slower. An error code E3 is displayed in the second display area and the calibration will not be accepted if the electrode is outside the designated 70 to 110% range. If no Error code is displayed, but the electrode is sluggish and slow to respond, checking the SL value in the Second Display after a 2-point calibration should indicate a value below 90%. 1) Coated or dirty Measuring Electrode. 2) Measuring Electrode (Glass) is too old and slow to respond. 3) Measuring Electrode (Glass) is damaged. 4) Bad wiring or Electrode Cable is faulty (bad isolation). 5) Solution Ground Electrode (Post) is coated, dirty or connection is faulty.

1) Clean ph electrodes in 5-10% HCl solution or Muratic Acid. 2) Replace Measuring Electrode. 3) Replace Measuring Electrode. 4) Check wiring connections and replace Electrode Cable if needed. 4) Clean Solution Ground Post, check wiring and replace if faulty. Programming and Operating Error Codes E4: Impedance of Glass Electrode is too LOW E4.1 Impedance of Glass Electrode is too LOW (PH202 and PH402 only) The Impedance of a Glass Electrode will increase as it ages. Therefore, if an Error Code E4 or E4.1occurs, it is not due to an aging electrode, but to an electrode failure. The EXA compares the measured impedance value of the electrode to an acceptable minimum limit. If that limit is exceeded, an E4 or E4.1occurs will be displayed in the second display area. When this Error Code appears, the ph value shown on the Main Display is usually near 7 ph. This is because a shorted Glass Electrode has a 0 mv potential which is the millivolt value generated at 7 ph. 1) Cracked Measuring Electrode. 2) Bad wiring or Electrode Cable is faulty (bad isolation). 1) Replace Measuring Electrode. 2) Check wiring connections and replace Electrode Cable if needed. E4.2 Impedance of Reference Electrode is too LOW (PH202 and PH402 only) The Impedance of a Reference Electrode can decrease from damage or deterioration Therefore, if an Error Code E4.2 occurs, it is due to failure. The EXA compares the measured impedance value of the electrode to an acceptable minimum limit. If that limit is exceeded, an E4.2 will be displayed in the second display area. 1) Cracked Reference Electrode. 2) Reference junction is loose or missing 3) Bad wiring or Electrode Cable is faulty (bad isolation).

1) Replace Reference Electrode. 2) Replace Reference Electrode 3) Check wiring connections and replace Electrode Cable if needed. E5: Impedance of Glass Electrode is too HIGH E5.1: Impedance of Glass Electrode is too HIGH (PH202 and PH402 only) The Impedance of a New Glass Electrode varies depending on the type of electrode (SM21-AG4 = 50-100 megohms; SM21 -AL6 = 600-900 megohms). As an electrode ages, or is coated by the process, its impedance value will increase and its responsiveness will decrease. The EXA compares the measured impedance value of the electrode to an acceptable maximum limit. If that limit is exceeded, an E5 or E5.1 will be displayed in the second display area. 1) Coated or dirty Measuring Electrode. 2) Measuring Electrode (Glass) is too old and slow to respond. 3) Bad wiring or Electrode Cable is faulty (bad isolation). 4) Solution Ground Electrode (Post) is coated, dirty or connection is faulty. 5) Electrodes not fully immersed in the process or buffer. 1) Clean ph electrodes in 5-10% HCl solution or Muratic Acid. 2) Replace Measuring Electrode. 3) Check wiring connections and replace Electrode Cable if needed. 4) Clean Solution Ground Post, check wiring and replace if faulty. 5) Assure Electrodes are properly immersed in solution. E5.2: Impedance of the Reference Electrode is too HIGH (PH202 and PH402 only) E6: Impedance of the Reference Electrode is too HIGH The Impedance of a New Reference Electrode is usually be 5K ohms or less. If the process coats the electrode, the impedance value will increase. The EXA compares the measured impedance value of the reference electrode to an user programmable (50K to 9999K ohms) minimum limit. The factory default value is 100K ohms. If that limit is exceeded E5.2 or E6 will be displayed in the second display area.

1) Dirty Reference Electrode, or plugged Reference Junction. 2) Faulty Reference Electrode. 3) Bad wiring or Electrode Cable is faulty (bad isolation). 4) Dirty or coated Solution Ground Electrode. 5) Electrodes not fully immersed in the process or buffer. 1) Clean Electrodes with 5-10% HCl or Muratic Acid. 2) Replace Reference Electrode. 3) Check wiring connections and replace Electrode Cable if needed. 4) Clean Electrodes with 5-10% HCl. 5) Assure Electrodes are properly immersed in solution. E7: Measured Temperature is too HIGH (above 130 C) E8: Measured Temperature is too LOW (above -10 C) Either of these Error Codes (E7 or E8) may be accompanied by E9 as an error in the Temperature circuit can cause the measured value to be out of range. The EXA compares the resistance of the Temperature Electrode to acceptable values. If these values are exceeded (130 C) E7, or (-10 C) E8 will be displayed in the second display. 1) Wrong Temperature Electrode programmed. 2) Bad wiring or Electrode Cable is faulty (bad isolation). 3) Temperature Electrode is faulty. 4) Process Temperature out of range. 1) Reprogram for Temperature Electrode being used. 2) Check wiring connections and replace Electrode Cable if needed. 3) Replace faulty Temperature Electrode. 4) Correct Process Temperature. E9: Measured ph value is outside acceptable range: -1 ph to +15 ph The EXA has a measurement range of -1 ph to +15 ph. If the instrument sees a value outside this range Error Code E9 will be displayed in the second display.

1) Solution Ground Electrode disconnected or faulty cable. 2) Measuring Electrode is faulty. 3) Wrong Temperature Electrode programmed or Electrode is faulty. 4) Incorrect wiring or faulty Electrode Cable. 5) Process ph value is outside acceptable measurement range. 1) Check Solution Ground connection and replace wire if needed. 2) Replace Measuring Electrode. 3) Check programmed value and replace Temperature Electrode if faulty. 4) Check wiring connections and replace Electrode Cable if needed. 5) Correct process. E10: Programming efforts to the Analyzer's EEPROM failed If efforts to program the EXA are not accepted by the EEPROM, Error Code E10 will be displayed in the second display. 1) Power surge or interruption during programming. 2) EEPROM is damaged or internal wiring is faulty. 1) Reprogram instrument. 2) CONTACT NEAREST YOKOGAWA SERVICE CENTER. E11: User programmed values for Response Time Check for electrode Wash Cycle has not been achieved. (Does not apply to EXA200 or EXA202) After a Wash Cycle has been completed, if the measurement does not return to acceptable values within the programmed Response Time, Error Code E11 will be displayed in the second display. 1) Wash was not effective, electrodes are still dirty. 3) Measuring Electrode is aged or faulty. 1) Check volume of cleaning solution and manually initiate Wash again 2) Replace Measuring Electrode.

E12: Measured ORP/rH value is outside acceptable range: -1500mV to +1500mV (ORP) and 0 to 55 rh(rh). (PH202 and PH402 only) The EXA202/402 has a measurement range of: -1500mV to +1500mV (ORP) and 0 to 55 rh(rh). If the instrument sees a value outside this range Error Code E12 will be displayed in the second display. 1) Solution Ground Electrode disconnected or faulty cable. 2) ORP or Glass Electrode is faulty. 3) Incorrect wiring or faulty Electrode Cable. 4) Process ORP/rH value is outside acceptable measurement range. 1) Check Solution Ground connection and replace wire if needed. 2) Replace ORP or Measuring Electrode. 3) Check wiring connections, replace Electrode Cable if needed. 4) Correct process. E14: No valid calibration data. (PH202 and PH402 only) If the EXA202/402 instrument is changed from ph to ORP(or rh) or is changed from ORP(or rh) Error Code E14 will be displayed in the second display if the calibration values are lost 1) Data has been lost 1) Perform a new calibration E15: Cable resistance of Temperature Electrode exceeds acceptable limit of 5K ohms The EXA compensates for Temperature Measurement errors resulting from cable resistance up to 5K ohms. If the cable resistance exceeds this limit, Error Code E15 will be displayed in the second display.

1) Temperature cable is too long, wrong gage wire, corroded or faulty. 2) Temperature Electrode is faulty or connector is corroded. 1) Check cable and replace if required. 2) Check connection and Replace Temperature Electrode if required. E16: Call for maintenance interval has been exceeded. (PH202 and PH402 only) The EXA202/402 instrument can be programmed (up to a 250 day interval) to call for a calibration. If the watchdog reminder alarm occurs, then the interval has been exceeded without being reset and a Error Code E16 will be displayed in the second display. 1) A calibration was not performed in the interval time set in Access Code 55. 1) A calibration needs to be performed. Consider resetting the interval time set to a longer duration in Access Code 55. E17: Incorrect Programming. Entered unacceptable values. E19: Incorrect Programming. Entered unacceptable values. If values that are incorrect or outside acceptable limits of the EXA, Error Code E17 or E19 will be displayed in the second display. 1) E17: Values outside minimum acceptable span values were programmed. 2) E19: programmed values are not accepted as valid. 1) Check for acceptable limits and reenter values. 2) Follow prompts and reenter values.

E20: All Programmed values are lost. If all programmed values are lost, Error Code E20 will be displayed in the second display. This instrument will have to be returned to the factory for repair and/or re-initialization. 1) Severe EMF interference. 2) Failure of electronics. 1) CONTACT NEAREST YOKOGAWA SERVICE CENTER. E21: Checksum error. (PH202 and PH402 only) If a internal software failure occurs in the Check Sum function a Error Code E20 will be displayed in the second display. This instrument will have to be returned to the factory for repair. 1) Failure of electronics. 1) CONTACT NEAREST YOKOGAWA SERVICE CENTER. E22: Alarm activation time exceeded. (PH202 and PH402 only) If the EXA s process control function is being used and the user s programmed time interval for effective control action has expired a Error Code E22 will be displayed in the second display. 1) Too short a time interval has been programmed in Access Code 47 2) A failure of some nature has occurred in the control system and reagent addition has not occurred.

1) Program a longer time interval in Access Code 47 before calling for assistance due to ineffective control. 2) Find the failure in the control system and fix the problem Example: Refill the reagent addition tank. E23: Calculated Asymmetry Potential is outside acceptable range: -120 to +120 mv - (Calibration Error for PH202 and PH402 only) This code occurs after an single point calibration has been attempted. The EXA compares the calculated Asymmetry value against the acceptable limits of -120 mv to +120 mv. Asymmetry is the mv offset value required to get back to 0 mv. The closer to 0 mv a Reference Electrode is, the better it is. New electrodes have values of ± 15 mv. As they age and are affected by the process this value will move further away from zero in either a positive or negative direction. As this occurs, response becomes slower and more frequent calibration will be required. If the electrode is outside the designated -120 mv to +120 mv range, an error code E23 is displayed in the second display area and the calibration will not be accepted. 1) Bad Buffer 7 (zero buffer). 2) Reference Electrode is too old and out of acceptable limits. 3) Reference Electrode is poisoned or damaged. 4) Bad wiring or Electrode Cable is faulty (bad isolation). 5) Solution Ground Electrode (Post) is coated, dirty or connection is faulty. 1) Clean ph electrodes in 5-10% HCl solution or Muratic Acid. 2) Replace Reference Electrode. 3) Replace Reference Electrode. 4) Check wiring connections and replace Electrode Cable if needed. 5) Clean Solution Ground, check wiring connection and replace if faulty.