Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2

Similar documents
Understanding CT image quality

Scope: All CT staff technologist

2017 Computed Tomography

2012 Computed Tomography

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Abstract. Learning Objectives 8/1/2017

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

2

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

In recent years, CT technology has undergone profound

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Request for Proposals

MR Accreditation Programs - E. Jackson

Mammo Tomosynthesis 5MP

Magnetic resonance imaging phase encoding:

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY

Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

FFDM Quality Control in Canada - a Vendor Neutral Approach

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D.

Brilliance CT 64-channel configuration

Troubleshooting Guide. Prep, Scan Errors, and Artifacts

Role of Color in Telemedicine Applications. Elizabeth A. Krupinski, PhD

Quality Control. Chapter 23

Nio. Industry-standard diagnostic display systems

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors

This talk covers currently available display technology.

A method for calculating the dose length product from CT DICOM images

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

Breast Imaging Essentials

What to consider when choosing a mammography display

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Monitor QA Management i model

M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

CLINICALLY PROVEN DESIGNS

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

World First Slim Cassette Type Digital Mammo. Upgrade Solution

A Comparison of DPM and VMC++ Mont Carlo codes applied to heterogeneous media

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications

Procedure Manual for MRI of the Brain

GUARD BAGAŽO TIKRINIMO VARTAI

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018)

Imaging diagnostico in Sanità Stato attuale e prospettive

Performance Evaluation of Industrial Computed Radiography Image Display System

Australian/New Zealand Standard

Guidelines for Assuring Softcopy Image Quality

Guidance for Quality Assurance of PACS Diagnostic Display Devices

Technologist Quality Control Procedures

Dental Line. High frequency intraoral x-ray unit. radiology ahead

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System

CT Numbers: Think of a number, double it, add 20, divide by 4. Jane Edwards Royal Free Hospital, London

+ Human method is pattern recognition based upon multiple exposure to known samples.

The Digital Debate. Competing Technologies. Up and Coming 1/5

Light. Engineered for Performance- Esaote MRI, designed to make a difference.

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM

SRS and ERLP developments. Andrew moss

Breast MR Imaging and Quality Control

CASE HISTORY#3 COOLING TOWER GEARBOX BEARING FAULT. Barry T. Cease Cease Industrial Consulting

Leona Cummins Speaker! Motivator! Educator! Humorist!

CLINICAL SOLUTIONS THE TECHNOLOGICAL ADVANTAGE FOR DENTAL CLINICS AND UNIVERSITIES

SPECIFICATION NO NOTE

Clarity and confidence. Raising the bar in diagnostic imaging with Barco radiology and mammography displays

S M A R T U LT R A S O U N D

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Broken Wires Diagnosis Method Numerical Simulation Based on Smart Cable Structure

Imaging solutions for any size patient and any size budget.

Preface. system has put emphasis on neuroscience, both in studies and in the treatment of tinnitus.

MPKOMNP. kéï=~ë=çñw== d^ifiblp. j~áåíéå~ååé=fåëíêìåíáçåë=d^ifiblp=`çãñçêí=mirp. båöäáëü. Cover page

Implementing a Proton Beam Scanning System within an Operating Clinical Facility

Dear Mama Project. Ricardo Martínez. Monday, 03 July Centro Nacional de Microelectrónica CSIC. Pisa

Precision, power, and productivity

Name Date. Listen to the conversations. Check the correct sentences. #1 Vera is a new patient not a new patient

Newsletter. A Discovery in Africa

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management

Evaluation of Performance, Reliability, and Risk for High Peak Power RF Sources from S-band through X-band for Advanced Accelerator Applications

MITSUBISHI ELECTRONICS AMERICA, INC.

Performance characteristics and quality assurance considerations for displays used in interventional radiology and cardiac catheterization facilities

X-ray Inspection. Series.

Case 10A EMI AND THE CT SCANNER [A] *

Coronis 5MP Mammo. The standard of care for digital mammography

Enhanced Diagnostics through Ultrasound Imaging

The Diagnosis of Small Solitary Pulmonary Nodule:

Mosby's Exam Review For Computed Tomography, 2e Ebooks

Detectability of breast cancer and reading time on screening mammography: Comparison between 5-MP and 8-MP LCD monitors.

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

GE Healthcare. CT/e 0459 * Operator Manual Revision 3. Copyright by General Electric Company

YXLON Cougar EVO PLUS

INSTRUCTION & OPERATION MANUAL. MODEL MMW-05 (5kW) MUROMACHI MICROWAVE FIXATION SYSTEM

Coronis Fusion multi-modality displays. The ultimate in diagnostic flexibility

MQSA Quality Control Manual

Transcription:

James M. Kofler, Ph.D. Mayo Clinic Rochester AAPM Annual Meeting 2017 1 Case 1 Ring Artifact? 2 3 1

Star Pattern caused by metal earrings! 4 Same star artifact commonly caused by Dental Amalgam Metal Implants Metal Objects (wires, syringes, bullets, etc.) 5 Case 1 Star pattern Obvious, and interferes with diagnostic content Typically easy to determine cause Cause Metal (high atten.) in FOV Remedy Tilt gantry, avoid if possible Increase kv, mas (diminishing returns) Metal Artifact Reduction algorithms WE-G-209-2 Artifacts: CT 6 2

Case 1: Important Points Look at the localizer radiograph WE-G-209-2 Artifacts: CT 7 Case 2 One or more concentric rings in image Subtle to obvious Cause typically straightforward 8 Ring Artifact 9 3

Ring Artifact Service was called 10 Bad Detector Module 11 Ring Artifacts from Photon Starvation 12 4

Case 2 One or more concentric rings in image Subtle to obvious Cause typically straightforward Cause Detector(s) imbalance/malfunction or blocked Photon starvation Remedy Service Increase technique, if possible 13 Case 2: Important Points Ring artifacts common Usually requires Service Scrolling can help visualize Check centering Patient not always centered but rings are 14 Case 3 Some shape superimposed on images Can be subtle but usually obvious Usually doesn t mimic pathology Not intermittent 15 5

Cushion in FOV during morning calibration 16 Case 3 Some shape superimposed on images Can be subtle but usually obvious Usually doesn t mimic pathology Not intermittent Cause Object scanned during calibrations Remedy Re-calibrate 17 Case 3: Important Points Don t overlook the simple things Object could be cushion, pillow, phantom, etc. 18 6

Case 4 Dark blotches on head scan Not too subtle but mimic critical pathology Not intermittent 19 Normal (prior from previous day) Diffuse right hemispheric abnormalities -Very serious 20 QA Phantom from Morning QC WW: 400, WL: 0 Typical abd settings 21 7

QA Phantom from Morning QC WW: 100, WL: 0 ACR settings 22 QA Phantom from Morning QC WW: 40, WL: 0 Better settings? 23 Case 4 Dark blotches on head scan Not too subtle but mimic critical pathology Not random Cause Contrast material on gantry window Remedy Wipe off gantry Note: Make sure not calibrated into system 24 8

Case 4: Important Points Morning QAs must be reviewed carefully using appropriate ww/wl Suspected artifacts must be reported Inspect gantry between every patient for contrast spillage, if needed. Clean with water and tissue/cloth (no soap/disinfectants) Known spills should be cleaned immediately 25 Case 5 Irregular dark bands Very subtle (2-3 HU) and mimics pathology Intermittent, very infrequent This is the most challenging, and most dangerous, type of artifact 26 Suspected cerebral edema -Very serious -Patient transferred by ambulance 45 miles from remote site Happened with 2 different patients within 24 hours 27 9

Morning QA images - 2 of 12 showed very subtle artifact Both at WW: 40, WL: 0 Service was called 28 Case 5 Irregular dark bands Very subtle (2-3 HU) and mimics pathology Intermittent, very infrequent Cause Air bubbles in tube cooling system Remedy Repair by Service 29 Case 5: Important Points WW / WL very important 400 / 40 100 / 0 40 / 0 Alert staff of intermittent issue 30 10

Case 6 (last one!) Partial rings not centered over isocenter Very obvious, does not mimic pathology Intermittent, very infrequent 31 32 33 11

Case 6 (last one!) Partial rings not centered over isocenter Very obvious, does not mimic pathology Intermittent, very infrequent Cause Moving air bubbles in patient Remedy No remedy just identify See it once and recognize forever! 34 Clinical Image Artifacts Be familiar with common artifacts How to recognize How to address Trouble shooting Start with the simple sources Communicate concerns Call Service when necessary Provide sample cases, if possible Let staff know of any potential problems Technologists should be diligent 35 References: Hsieh J., Computed Tomography: Principles, Design, Artifacts, and Recent Advances (Chapter 7), SPIE Press, Bellingham, WA, ISBN 0-8194-4425-1. Computed Tomography Quality Control Manual, 2012, American College of Radiology. http://www.acr.org/education/education- Catalog/Products/8336734 Barrett JF, Keat N, Artifactis in CT: Recognition and Avoidance, Radiographics 2004; 24:1679-1691. Hedrick WR, Markovic MA, Short JA, Vera CD, Computed Tomography Artifact Created by Air in the X-ray Tube Oil, JCAT, 40(1) 2016. Liu F, Cuevas C, Moss AA, Kolokythas O, Dubinsky TJ, Kinahan PE, Gas Bubble Motion Artifact in MDCT, AJR: 190, Feb 2008. 36 12

Artifact from Tube Arcing 37 13