Operation Manual TX20 1/8 DIN Microcomputer Based ph/orp Controller with relays

Similar documents
OPERATION MANUAL JENDPM008 THERMOCOUPLE PANEL THERMOMETER. TIP TEMPerature Products 340 W Broad Street Burlington, NJ 08016

JENCO ELECTRONICS, LTD.

Global Water Instrumentation, Inc.

Instruction Manual Alpha ph 190 1/8 DIN ph/orp Controller Transmitter ph / mv / ºC

BASIC PH METER BASIC PH METER KIT

Instruction Manual Aquamon 2000 plus 2-Wire ph/orp Transmitter

SCALE & WEIGHT DISPLAYS

EXA PH200/400 and EXA PH202/402 Troubleshooting and Error Code Guide

TX2000 Intelligent ph/orp Transmitter

This document courtesy of:

Flat-Bed Module Recorders

AE-341 SERIES PROCESS INDICATORS 2 TO 6 DIGITS 0-20 MA, 4-20 MA, 0-5V, OR 0-10V SCALED INPUT

Model /29S RF Splitter

Set Point Controller OC557. Operator s Manual

Assembly Level Service Guide

LAUREL ELECTRONICS, INC.

STD-525T PRECISION TIMER

OPERATING AND SAFETY INSTRUCTIONS for DIGITAL TEMPERATURE CONTROLS (PLSM SERIES)

INSTRUCTION MANUAL MODEL 2710 SUBCARRIER DEMODULATOR

Model Number Structure

UDC100 Universal Digital Controller. Specification. Overview. Features. Features, continued /99 Page 1 of 4

Instruction Manual Model # Block Upconverter

MG-XV operating instruction. Measuring of norm signals, 4-8-digit. Panel instrument type MG-BV Construction instrument type MG-AV

DCP100 Digital Control Programmer Specifications

Operating Instructions BTX-1 Series Digital Band Tension Meter

NS-3 RF Noise Source Operation Manual

263 Series LED Bargraph Indicator and Controllers

Large Display ORP Pen

AEXX-200F SERIES 4-DIGIT LINE FREQUENCY MONITORS

AES-402 Automatic Digital Audio Switcher/DA/Digital to Analog Converter

MX-206 Stereo Microphone Mixer. Operating Manual

Instruction Manual Model BlockUpconverter

AES-404 Digital Audio Switcher/DA/Digital to Analog Converter

Electronic Panel Meters DIGEM Preference Program Process control, automation & laboratory uses Class 0.01 to 1 Current, Voltage, Frequency,

LEVEL ADJUST POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

TC3005(LED/ELED/LASER) User's Manual

On-screen display signal strength meter Version 1.02

User s Manual of Signal Level Meter TABLE OF CONTENTS

Weekly Timer. Mounting track 50 cm (1.64 ft) length PFP-50N 1 m (3.28 ft) length PFP-100N

OPERATOR MANUAL OSD351 FIBRE OPTIC CCTV TRANSMITTER CARD

USER INSTRUCTIONS MODEL CSI-200 COAXIAL SYSTEM INTERFACE

OPERATOR MANUAL OSD553 TRIPLE VIDEO FIBRE OPTIC RECEIVER

Utility Amplifier GA6A Model

TC Mbps - 622Mbps FIBER OPTIC MODE CONVERTER/REPEATER (Rev A0.1) User's Manual

LEVEL ADJUST POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

UNIVERSAL DIGITAL METER DC Volts and Amps AC RMS Volts and Amps Thermocouples and RTDs Process Signals Strain Gauge and Load Cell

Sartorius Professional Meter

MONITOR POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

WQ Series Water Quality Bench top Meters

Metal Electrode Meter

AK-PVE4 Operating Instructions. Measuring of norm signals in wall-type units. Performance:

SV-LCD50. Installation and User Guide. Thin-Film Transistor (TFT) Liquid Crystal Display (LCD) Color Rear Vision Monitor. Version 1.

Sphinx II. Owner s Manual. Tube Hybrid Integrated Power Amplifier. Rogue Audio, Inc. 3 Marian Lane Brodheadsville, PA Issue date: 08/01/16

TECHNICAL MANUAL DRAW-WIRE DISPLACEMENT TRANSDUCER TYPE DWT

Model R177M and R177S Baseband Switch

Instruction Manual ph/ion 510 Bench ph/ion/mv Meter

Product Identification

Triple RTD. On-board Digital Signal Processor. Linearization RTDs 20 Hz averaged outputs 16-bit precision comparator function.

Data Acquisition Networks. Installing and Configuring the DM01 Hardware

TC3301. User's Manual

Electronic converter for level transmitters MT03L Instructions manual

Omega.com TM. DPS3100 & DPS3200 series Programmable Scanners

INSTALLATION MANUAL FT-FOTR-1VDE-ST-S

SX7 e l e c t r o n i c c r o s s o v e r

Panel cutout required: 1.772" x 3.622" (45mm x 92mm) 1.76" (45mm) 2.45" (62mm) 3.20" (81mm) 3.60" (91mm) 0.59" (15mm) Special Features

GP-500 SERIES - Low Voltage

DP1 DYNAMIC PROCESSOR MODULE OPERATING INSTRUCTIONS

Noise Detector ND-1 Operating Manual

E8Y. Micropressure Sensor with Easy-to-Read Digital Display. Differential Pressure Sensor. Ordering Information. Sensors

Quality. Accurate, reliable results in a durable easy-to-use design

OPERATOR MANUAL OSD8865 DIGITAL TRIPLE VIDEO FIBER OPTIC RECEIVER

clipping; yellow LED lights when limiting action occurs. Input Section Features

WC-2100 WORD CLOCK DISTRIBUTION AMPLIFIER INSTRUCTION MANUAL

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Signal Conditioners. Highlights. Battery powered. Line powered. Multi-purpose. Modular-style. Multi-channel. Charge & impedance converters

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

6170 Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

ANALOG I/O MODULES AD268 / DA264 / TC218 USER S MANUAL

Connevans.info. DeafEquipment.co.uk. This product may be purchased from Connevans Limited secure online store at

DUAL/QUAD DISPLAY CONTROLLER Operation Manual

Electronic Panel Meters DIGEM Preference Program Process control, automation & laboratory uses Class 0.01 to 1 Current, Voltage, Frequency,

Standard signal metering in wall-mounting case

MT03A Electronic converter for flow rate transmitters

32 Channel CPCI Board User Manual

Weekly Time Switch. Rated time Time setting range Time division 24 hrs x 7 days 00:00 to 23:59 1min

VectorVGA Tempest User Manual

Vinmetrica SC-200 User Manual

VK-P10SE WARRANTY REGISTRATION FORM

MS2540 Current Loop Receiver with RS485 Communication

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630)

OPERATING INSTRUCTIONS. Single Function Timers Model No. 740-T, 740-TJ & 740-TSeries ELECTRICAL TABLE OF CONTENTS

BC104 BATCH CONTROLLER

OPERATING INSTRUCTIONS. Multi-Function Computers Model No. 840-C, 840-CJ & 840-CF Series ELECTRICAL TABLE OF CONTENTS

PIX/2 Operation Manual

SDC-500 Digital Optical Chopper Operating Instructions

48x24. Options ORDER NUMBER OF TYPE PFL B

DPM OWNERS MANUAL Website: Phone: (800)

Multifunction Digital Timer

Veeder-Root brand Series C628 Dual Preset Counter (C628-8XXX) Technical Manual

Panel-mounting thermostats, type series EM

Transcription:

Operation Manual TX20 1/8 DIN Microcomputer Based /ORP Controller with relays 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1

CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE SENSOREX TX20...3 A. Mounting procedure...3 B. Front panel...4 C. LCD screen...5 D. Rear connectors...6 E. Measure mode...7 F. Setting mode...7 G. calibration mode...10 H. RmV calibration mode...12 I. Controlling the relays...12 ERROR DISPLAY AND TROUBLESHOOTING.....13 BUFFERS.....14 SPECIFICATIONS.....14 WARRANTY.....15 pg. 2

GENERAL INTRODUCTION Thank you for selecting the SENSOREX TX20 /ORP Controller, a rugged microprocessor based instrument assembled in a watertight 1/8 DIN case, designed for use in laboratories and process control applications. The system displays or ORP or Temperature and relay status in one large LCD screen. The TX20 s microprocessor performs a self-diagnostic routine every time you turn on the unit providing you with basic information on the stability of the instrument. The model TX20 is equipped with 2 control relays. All control relays are programmable and hysteresis driven. INITIAL INSPECTION Carefully unpack the unit and accessories. Inspect for damages made in shipment. If any damage is found, contact Sensorex or supplier immediately. All packing materials should be saved until satisfactory operation is confirmed. USING THE SENSOREX TX20 /ORP TRANSMITTER ERVIEW A. Mounting Procedure 1. Make a cutout on any panel, with a thickness of 1/16 inch (1.5mm) to 3/8 inch (9.5mm). Refer to FIG 1 Figure 1 2. Remove the mounting bracket from the controller and insert the controller into the cutout. Refer to FIG 2. Figure 2 Mounting Brackets pg. 3

3. Replace the mounting bracket assembly onto the controller and secure the controller to the mounting panel. Refer to FIG 3. Panel Mounting brackets Figure 3 B. Front Panel The front panel consists of a 4-digit LCD display and 4 keys. 1. [ MODE ] key: 1a. In the Measure mode, this key will switch the display in sequence from, Temperature, ORP absolute mv, ORP relative mv and back to again. 1b. In the Calibration/Setting mode, pressing this key for three seconds will move you back to the previous parameter in the case when recalibration / resetting is required. 2. [ UP ] key: 2a. In the Calibration mode, pressing this key will show the next possible option. In the Setting mode, pressing this key will show the next possible option and increases the numeral increment. 2b. In the Measure mode, pressing this key and [] key at the same time, the unit will enter the Calibration mode. 3. [ DOWN ] key: 3a. In the Calibration mode, pressing this key will show the next possible option. In the Setting mode, pressing this key will show the next possible option and decreases the numeral increment. 3b. In the Measure mode, pressing this key and [] key at the same time, the unit will enter the Setting mode. pg. 4

4. [ ] key: In any mode where the user can change the settings, pressing this key will save the new settings. If no change has been made then pressing this key will just move the user to the next setting. C. LCD screen 1 2 3 4 5 6 HI1 HI2 LOW1 LOW2 C STAND SLOPE RmV 6.86 7.00 ATC MAN 7 8 9 10 11 12 13 1. Major LCD display. 2. This icon will be displayed if the meter is in the Calibration/Setting mode. 3. HI1 & LOW1 These icons, when displayed, indicate relay action and relay number. 4. HI2 & LOW2 These icons, when displayed, indicate relay action and relay number. 5. Temperature and unit display. 6. Unit indicator. 7. STAND This icon will blink before Buffer 1 calibration. The icon will stay on while Buffer 1 is being calibrated. 8. SLOPE This icon will blink before Buffer 2 calibration. The icon will stay on while Buffer 2 is being calibrated. 9. 6.86 The 6.86 buffer group: 6.86, 4.00, 9.18. 10. 7.00 The 7.00 buffer group: 7.00, 4.01, 10.01. 11. ATC This icon will be displayed when a temperature probe is connected. 12. MAN This icon will be displayed when a temperature probe is not connected. 13. RmV Unit indicator. pg. 5

D. Rear connectors INPUT RFF +V -V GND 10k TH Pt 1000 L N E 100~230VAC Rel 1 Rel2 1. Connect the AC line to the rear of the instrument. The model TX20 can be used with 100~240V AC at 50/60 HZ. Make sure the EARTH connector is connected to the earth lead of the AC power line. 2. Connect the proper load to the output relays. Make sure that the load does not exceed the relay rating, 5 Amp at 115VAC and 2.5 Amp at 230 VAC. 3. V+ (5VDC) and V- (5VDC) output to provide excitation voltage for /ORP pre-amplifier only. Note : (1) Make sure that the power is unplugged before wiring your probes, relay etc. (2) Make sure you connect the AC power cord to the correct AC terminals. Connecting incorrectly may damage the unit permanently. pg. 6

Measure mode Turning on the unit will always display the Measure mode. This instrument is designed to provide 4 distinct measurements: MODE C MODE 7.00 ATC ATC mv MODE MODE RmV 1. The degree of acidity or alkalinity of the solution. 2. Temperature Current temperature of the solution. 3. ORP mv A measurement of absolute ORP mv. 4. ORP RmV A measurement of relative ORP mv. The offset value at the RmV calibration will be added to the ORP absolute value to display the ORP relative value. Note : ORP relative value range: ORP absolute value - 1000mV to ORP absolute value + 1000mV Pressing [MODE] key in the Measure mode will cycle the display from the four modes above. E. Settings mode Pressing [DOWN] key and [] key at the same time, the meter will enter into the Setting mode. 1. Temperature compensation select: + ATC ATC buffer 1 select MAN Pressing [UP] key or [DOWN] key in this screen will cycle the display from 01 (Thermistor: 10k ohm), 02 (Resistor: PT1000), 03 (Manual) modes above. Select the preferred temperature compensation mode, press [] key to save, and enter the next setting screen. pg. 7

2. buffer 1 select: Pressing [UP] key or [DOWN] key in this screen will cycle the display from 7.00, 6.86 buffer above. Select the preferred buffer, press [] key to save, and enter the next setting screen. Temperature compensation select 7.00 6.86 buffer 2 select Buffer 1 7.00 Buffer 2 6.86 7.00 7.00 6.86 6.86 Relay control parameters select Relay control parameters select Pressing [UP] key or [DOWN] key in this screen will cycle the display from 4.01, 10.01 (or 4.00, 9.18) buffer above. Select the preferred buffer, press [] key to save, and enter the next setting screen. Note : The buffer 2 is either 4.01 or 10.01 if select 7.00 buffer at buffer 1 select screen. The buffer 2 is either 4.00 or 9.18 if select 6.86 buffer at buffer 1 select screen. pg. 8

3. Relay control parameters select: Buffer 2 select mv RmV Relay 1 HI/LOW select Pressing [UP] key or [DOWN] key in this screen will cycle the display from, ORP mv ORP RmV modes above. Select the preferred mode, press [] key to save, and enter the next setting screen. Note : The relay will work with the choice of parameters. Relay control paramelers select HI1 LOW1 Relay 1 set point value 4. Pressing [UP] key or [DOWN] key in this screen will cycle the display from HI LOW modes above. Select the preferred mode, press [] key to save, and enter the next setting screen. 5. Note : See I. Controlling the relays page. 6. Relay 1 set point value: Pressing [UP] key or [DOWN] key in this screen to adjust the value, press [] key to save, and enter the next setting screen. pg. 9

Relay 1 HI/LOW select Relay 1 hysteresis value 7. Relay 1 hysteresis value: Pressing [UP] key or [DOWN] key in this screen to adjust the value, press [] key to save, and enter the next setting screen. Relay 1 set point value Relay 2 HI/LOW select 8. Relay 2 High / Low select: Same as Relay 1 work way select 9. Relay 2 set point value: Same as Relay 1 set point value 10. Relay 2 hysteresis value: Same as Relay 1 hysteresis value G. Calibration mode The TX20 uses 2-point calibration for. The first point must be 6.86/7.00, and the second point can either be 4.00/4.01 or 9.18/ 10.01. In the Measure mode, pressing [UP] key and [] key at the same time to allow the meter to go to the Calibration mode. pg. 10

1. Buffer 1 (STAND) calibration: HI2 + STAND SLOPE 7.00 STAND 7.00 STAND 7.00 Buffer 2 calibration Enter into the calibration mode, the STAND icon will flash, the unit is ready to be standardized at the first buffer. Rinse the and ATC/Temp probes in distilled water and immerse them in the first buffer solution (either 7.00 or 6.86). Allow temperature reading to stabilize, then press key to calibrate. The icon will flash until the unit detects a stable reading. Once the unit calibrates the first point, the SLOPE icon will flash. The unit is ready to be sloped at the second buffer. 2. Buffer 2 (SLOPE) calibration: Buffer 1 calibration Measure mode STAND SLOPE 7.00 STAND SLOPE 7.00 Rinse the and ATC/Temp probe in distilled water and immerse them in the second buffer solution (either 4.00/4.01 or 9.18/10.01). Allow temperature reading to stabilize, then press key to calibrate. The icon will flash until the unit detects a stable reading. Once the unit calibrates the second point and the unit will automatically exit the calibration mode and goes to the Measure mode. Dual point calibration is complete. Note : In the Setting mode (1. Temperature compensation select), select 03 (Manual temperature compensation mode) if no temperature probe is being used. Press the [UP] key or [DOWN] key in the Manual temperature compensation mode to adjust the value to that of the test solution temperature. Then calibrate buffer 1 and buffer 2. pg. 11

F. RmV Calibration mode + or RmV RmV RmV RmV normal mode The model TX20 uses 1-point calibration for RmV. In the RmV Measure mode, pressing [UP] key and [] key at the same time, the meter will enter into RmV calibration mode. Rinse the ORP probe in distilled water and immerse it in the ORP standard solution, then press [UP] or [DOWN] key to adjust the ORP value to that the ORP standard. Press [] key to save. The unit beeps to indicate a successful calibration. Calibration is now complete and the unit will automatically switch to the ORP relative mv Measure mode. Note : When the ORP absolute mv reading is off, recalibrate RmV value. G. Controlling the relays 1. Isolation voltage: The maximum isolation voltage of the relay output contacts is 1500 VDC. The voltage differential between the relay output contacts and the load should not exceed 1500 VDC. 2. Output load: The current through the relay output contacts should not exceed 5 Amp at 115 VAC and 2.5 Amp at 230 VAC in order not to cause permanent damage to the relay contacts. This rating is specified for resistive loads only. 3. Relay action, relay set point and hysteresis value: Relay Effective RELAY-ON Set Effective RELAY-OFF Set Action Point Point HI S.P. + (1/2 H.V) S.P. (1/2 H.V) LOW S.P. (1/2 H.V) S.P. + (1/2 H.V.) S.P. = Relay Set point H.V. = Hysteresis value (Dead Band) If the relay action is set to HIGH, the relay will turn ON at (Set Point +1/2 Hysteresis ), and will turn OFF at (Set Point -1/2 Hysteresis ). If the relay action is set to LOW, the relay will turn ON at (Set Point -1/2 Hysteresis ), and will turn OFF at (Set Point +1/2 Hysteresis ). There are two Independent relays the user can bind to the, ABSOLUTE mv or RELATIVE mv mode. The user can only bind the two relays to one reading mode at a time. The user can change this anytime by changing option at the setting mode. Note : 1. The ideal set point range for is 0.00 to 16.00. 2. The ideal set point range for absolute mv is 2000 to 2000 mv. 3. The ideal set point range of for relative mv is -3000 to 3000 mv. pg. 12

ERROR DISPLAY AND TROUBLESHOOTING ERVIEW /ORP Temperature Possible cause(s) Display Mode Display Display [Action(s)] "OvEr" -10.0~120.0 C measure mode > 16.00 [Recalibrate] Undr -10.0~120.0 C measure mode < -2.00 [Recalibrate] "OvEr" "OvEr" measure mode a. Temperature > 120.0 C. [Bring buffer solution to lower temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to - 10~120 C.] "OvEr" Undr measure mode a. Temperature < -10.0 C. [Bring buffer solution to higher temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to - 10~120 C.] "OvEr" Temperature reading Absolute ORP mv or relative ORP Absolute ORP mv > +1999 mv [Bring solution to a lower ORP reading] "Undr" Temperature reading Absolute ORP mv or relative ORP Absolute ORP mv < -1999 mv [Bring solution to a higher ORP reading] ORP reading "OvEr" Absolute ORP mv or relative ORP a. Temperature > 120.0 C. [Bring buffer solution to lower temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to - 10~120 C.] ORP reading "Uder" Absolute ORP mv or relative ORP a. Temperature < -10.0 C. [Bring buffer solution to higher temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to - 10~120 C.] "OvEr" 0~60 C a. calibration mode-buffer 1 (STAND)-7.00 b. calibration mode-buffer 1 (STAND)-6.86 c. calibration mode-buffer 2 (SLOPE) a. Offset < -100 or > 100 mv b. Offset < -91.7 or > 108.3 mv c. Slope > ideal mv by 30% [Use a new buffer solution.] [Replace electrode.] pg. 13

BUFFERS The temperature characteristics of calibration buffers 4.00, 4.01, 6.86, 7.00, 9.18 & 10.01 are stored inside the instrument. The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values. C 4.00 6.86 9.18 4.01 7.00 10.01 0 4.01 6.98 9.46 4.01 7.11 10.32 5 4.00 6.95 9.39 4.01 7.08 10.25 10 4.00 6.92 9.33 4.00 7.06 10.18 15 4.00 6.90 9.28 4.00 7.03 10.12 20 4.00 6.88 9.23 4.00 7.01 10.06 25 4.00 6.86 9.18 4.01 7.00 10.01 30 4.01 6.85 9.14 4.01 6.98 9.97 35 4.02 6.84 9.10 4.02 6.98 9.93 40 4.03 6.84 9.07 4.03 6.97 9.89 45 4.04 6.83 9.04 4.04 6.97 9.86 50 4.06 6.83 9.02 4.06 6.97 9.83 55 4.07 6.83 8.99 4.08 6.97 9.80 60 4.09 6.84 8.97 4.10 6.98 9.78 Note : The actual reading of the instrument can differ from the values shown by ±0.01. SPECIFICATIONS Mode Range Resolution Accuracy -2.00 to 16.00 0.01 ±0.1% ± 1 digit ORP Absolute mv -1999 to 1999 mv 1 mv ±0.1% ± 1 digit ORP Relative mv -2999 to 2999 mv 1 mv ±0.1% ± 1 digit Temperature -10.0 to 120.0 C 0.1 C ±0.3 C : Recognized buffers US (4.01, 7.00, 10.01) or NIST (4.00, 6..86, 9.18) Temperature compensation Manual/Auto -10.0 C to 120.0 C Buffer Temperature range 0.0 C to 60.0 C Electrode Offset recognition +/-100 mv at 7.00 +108.3 mv/-91.7 mv at 6.86 Electrode Slope recognition +/- 30% at 4.00, 4.01, 9.18, 10.01 Input impedance Calibration end point sensing >10 12 Ω Yes pg. 14

Temperature: Temperature sensor Thermistor: 10k ohm at 25 C, (User selectable) Resistor ( PT1000) or Manual Controller: Control type Relay output(resistive load only) Two ON/OFF control 5A at 115VAC or 2.5A at 220VAC General: Keys Power: Audio feedback in all keys 100VAC to 240VAC, 50/60Hz Ambient Temperature range 0.0 to 50.0 C Case Weight IP65, 1/8DIN case, depth 90mm 290 g WARRANTY Sensorex warrants this product to be free from significant deviations in material and workmanship for a period of 1 year from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse, within the year period, please return-freight-prepaid and the correction of the defect will be made free of charge. If you purchased the item from our Sensorex distributors and it is under warranty, please contact them to notify us of the situation. Sensorex Service Department alone will determine if the product problem is due to deviations or customer misuse. Out-of-warranty products will be repaired on a charge basis. RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason. When applying for authorization, have the model and serial number handy, including data regarding the reason for return. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Sensoex will not be responsible for damage resulting from careless or insufficient packing. A fee will be charged on all authorized returns. Note : Sensorex reserves the right to make improvements in design, construction and appearance of our products without notice. pg. 15