ITER Product Catalogue V20170405 Page 1 / 15
Content TPG 300 Passive Gauge Controller... 3 Pirani/Cold Cathode Measurement Board... 5 Interface and Relay Boards... 6 Pirani Gauge TPR 018... 7 Cold Cathode Gauge IKR 085... 9 Cables... 11 Cable Connectors Controller Side... 14 Cable Connectors Gauge Head Side... 14 Conversion Tables... 15 Page 2 / 15
TPG 300 Passive Gauge Controller Ideal controller for gauges under extreme conditions such as radiation and high temperature bake out. Modular plug-in circuit boards for measurement and communication. DIMENSIONS Page 3 / 15
SPECIFICATION Display Range 1000 1x10-11 mbar (depends on used gauge and boards) Measured value display 15 mm high numbers / 7 segment LED Unit of measure mbar, Torr, Pa Set point / relays 8 parameters - lower and upper threshold / 4 relays Error display / relay Red LED / 1 relay Measurement rate 100 Hz Display rate 5 Hz Temperature: Storage -40 to +65 C Operating +5 to +50 C Radiation resistant no Max. vertical magnetically field 21 mt (3 db), 15 mt (6 db) tbd Max. horizontal magnetically field 11 mt (3 db), 8 mt (6 db) tbd Main requirement: Voltage (range) 90 264 VAC ±10% Frequency (range) 47 63 Hz Power consumption <55 VA Overvoltage category II Relative humidity 80% at temperatures up to +31 C decreasing to 50% at +40 C Use Indoors only, height up to 2000 m Safety IEC384, class 1,VDE 0411, part 2.80 Filter time constant 16 ms, 160 ms (standard), 16.6 s Weight (without boards) 1.35 kg Controller Unit TPG 300 IO546900 Delivery time 2 weeks ITER Status Existing TPG300 is qualified, except for SIL-2 and magnetically shielding. Shielding tests ahead in May. New controller under and in concept phase (SIL-2 in clarification) ACCESSORIES Mains cable, German plug, 2.5 m 303-350 Mains cable, U.S. plug, 2.5 m 303-353 Mains cable, U.K. plug, 2.5 m 303-354 Mains cable, Japan 303-352 Mains cable, Korea plug, 2,5 m 303-351 Rack 19 for 3 TPG300 in IT491211 CABLES Please see page 11 and following Note: New controller expected to be available in 2019 Page 4 / 15
Pirani/Cold Cathode Measurement Board TPG 300 Measurement Unit Measurement and control down to 10-9 mbar SPECIFICATION Measurement Boards Number of measuring circuits Gauge Measurement range max. Measurement range min. Temperature: Operating Radiation resistant Max. applicable magnetic field Sensor cable length max. Reaction time of the output signal: Rise Weight CP 300 C9 Pirani / Cold Cathode 1 Pirani and 1 Cold Cathode ITER TPR018 and IKR085 1000 mbar 5x10-9 mbar 5 to 50 C no See TPG300 100 m 10 ms (Sudden pressure changes: Rise (10-90% from <10-9 to <10-3 mbar) 0.21 kg Measurement Board IO441000 Delivery time 2 weeks ITER Status Qualified ACCESSORIES Pirani connector Amphenol C91 Cold Cathode connector Lab plug 2mm LS205 Red isolation KT205 Black isolation KT205 IT4722796CA under IT47111152MA IT4711152NZ IT4711152NZ CABLES Please see page 11 and following Page 5 / 15
Interface and Relay Boards TPG 300 Measuring systems ProfiBus communication with control system and Hard wired set point relay outputs. SPECIFICATION Relay Connector Relay Numbers Max. applicable magnetic field Max. applicable magnetic field Relay: Switching voltage Relay: Switching current Relay: Switching power Relay: Resistance (with connector) Interface Interface: Connection IF 300 P D-sub 15 pin, male 4 relay for set point, 1 error relay all with changeover contacts, potential free See TPG300 See TPG300 30 V DC / 50 V AC 1.5 A 45W, 75VA 125 mohm ProfiBus-DP D-sub-connector, 9-pin, male Interface: Baud rates 1200, 2400, 4800, 9600, 19200 Interface: Data format Interface: Cable length max. Weight Interface and Relay Board Delivery time ITER Status ProfiBus protocols 1200 m 0.16 kg IO441395 8 weeks after final test In Qualification (magnetically shielding) ACCESSORIES Mating connector, D-Sub, male, 9 poles Mating connector, D-Sub, female, 15 poles IT941145 IT441129 Page 6 / 15
Pirani Gauge TPR 018 for TPG 300 measuring system Bake able and radiation resistant DIMENSIONS Page 7 / 15
SPECIFICATION TPR 018 TPR 018 Flange DN 16 ISO-KF DN 40 CF-F Display range Measurement range 5x10-4... 1000 mbar 1x10-2 100 mbar Operating temperature, high temperature sensor cable 0 to 120 C Operating temperature, standard sensor cable 0 to 80 C Bake out temperature (with high temperature cable or without cable) 250 C Mechanical Integrity 300 C * Radiation resistance Pressure deviation due to external magnetic field @ 10-2 mbar @ 1 mbar @ 10 2 mbar 1x10 7 Gy Up to 480 mt: < 5% Up to 300 mt: < 2.5% Always positive Accuracy: at room temp. and cable length <20m =±10% of reading in the range of 1x10-2 100 mbar Accuracy: at 0 +70 C and within the entire range of specified cable length =±20% of reading in the range of 1x10-2 100 mbar Accuracy: within the entire specified range of temperatures and cable length =±35% of reading in the range of 1x10-2 100 mbar Repeatability with air =±5% of reading in the range of 1x10-2 100 mbar Materials: Internal seal Al, Ni90 Overpressure Insulator Filament / holder Al 2 O 3, NiFeCo W/Ni Chamber wall, Housing feed through St. Steel 1.4435, 1.4306 9 bar (ltd. to inert gases) Protection category IP 40 Mounting orientation We recommend mounting the gauge aligned with the magnetic field. Weight 0.6 kg 0.85 kg Limited measurement and switching accuracy at pressures above 100 and below 10-3 mbar Gauge IOG15020 IOG15024 Delivery time 6 weeks 6 weeks ITER Status Qualified Qualified CABLES Please see page 11 and following no accidental cold cathode activation *Mechanical integrity vacuum tightness Page 8 / 15
Cold Cathode Gauge IKR 085 for TPG 300 measuring system Bake able and radiation resistant Magnetically fully shielded Lemo Triax P&P connector DIMENSIONS Page 9 / 15
SPECIFICATION IKR 085 IKR 085 Flange DN 40 ISO-KF DN 40 CF-F Display range Measurement range Operating temperature, with normal cable Operating temp., high temp. cable and Magnet 1x10-9 1x10-2 mbar 1 x 10-8 5 x10-3 mbar +5 C +80 C +5 C +250 C Bake out temperature with -Magnet 250 C Mechanical Integrity Radiation resistance 300 C* 1x10 7 Gy Accuracy in measurement range Approx. ± 30% Max. external magnetical field, axial direction (-3 db) Max. external magnetical field, axial direction (-6 db) Max. external magnetical field, radial direction (-3 db) Max. external magnetical field, radial direction (-6 db) Fully reversible recovery from external magnetic field, any orientation 180 mt (<10% deviation), 280 mt (< factor 2 dev.) 120 mt (<10% deviation), 200 mt (< factor 2 dev.) 150 mt (<10% deviation), 180 mt (< factor 2 dev.) 100 mt (<10% deviation), 120 mt (< factor 2 dev.) Tested up to 600 mt Materials: Vacuum connection St. steel (1.4306) Measuring chamber St. steel (1.4306) Feedthrough isolation Ceramic (Al 2 O 3 ) Internal seal Anode 316 L, Ag plating, Ni90 Ignition aid St. steel (1.4330) Magnet Sm Co 2:17 Electrical connection (Connector/Type) Overpressure Mo Triaxial connector female/ Lemo 9 bar, for inert gases and temp. <55 C only Protection category IP 40 Mounting orientation We recommend mounting the gauge aligned with the magnetic field. xxxxxxx Weight 0.6 kg 0.85 kg Gauge IOG18772 IOG18773 Delivery time 8 weeks after final test 8 weeks after final test ITER Status In Qualification In Qualification CABLES Please see page 13 only with SHV and Triaxial connector cable ITC240xxx to use. Tested at room temperature (~25 C) *Mechanical integrity vacuum tightness Page 10 / 15
Cables Connecting TPG300 with Gauges Pirani Cable TPR018 Max. bake out temperature 80 C 250 C Mechanical Integrity 80 C 250 C Radiation resistant to 1x10 3 Gy 1x10 6 Gy Max. measuring cable length Reduced flame propagation (according IEC 60332 or NF 32070 C1) Flame retardant (acc. IEC 60332-1 flame propagation on single cables) Low smoke (according to IEC 61034) Zero Halogen (according to IEC 60754-1) Non toxicity (according to IEC 60754-2) Max. measuring cable length 100 100 RoHS X X Material Amphenol C91 and Lemo (standard plugs on each side) Sensor cable TPR, 3 m, 80 C Sensor cable TPR, 10 m, 80 C Sensor cable TPR, 15 m, 80 C Sensor cable TPR, 20 m, 80 C Sensor cable TPR, 25 m, 80 C Sensor cable TPR, 30 m, 80 C Sensor cable TPR, 35 m, 80 C Sensor cable TPR, 40 m, 80 C Sensor cable TPR, 45 m, 80 C Sensor cable TPR, 50 m, 80 C Other cable length on request ITC548308 ITC548456 ITC548457 ITC548458 ITC548459 ITC548460 ITC548461 ITC548462 ITC548463 ITC548464 ITC154xxx High temperature sensor cable TPR, 3 m, 250 C High temperature sensor cable TPR, 5 m, 250 C High temperature sensor cable TPR, 10 m, 250 C High temperature sensor cable TPR, 15 m, 250 C High temperature sensor cable TPR, 20 m, 250 C High temperature sensor cable TPR, 25 m, 250 C High temperature sensor cable TPR, 30 m, 250 C High temperature sensor cable TPR, 35 m, 250 C High temperature sensor cable TPR, 40 m, 250 C High temperature sensor cable TPR, 45 m, 250 C High temperature sensor cable TPR, 50 m, 250 C Other cable length on request ITC548414 ITC548465 ITC448047 ITC448043 ITC448044 ITC120025 ITC120030 ITC120035 ITC120040 ITC120045 ITC120050 ITC120xxx Delivery time 4 to 6 weeks 4 to 6 weeks ITER Status none None Page 11 / 15
Cold Cathode Cable Max. bake out temperature 250 C Mechanical Integrity 250 C Radiation resistant to 1x10 6 Gy Max. measuring cable length 100 Reduced flame propagation (according IEC 60332 or NF 32070 C1) X Flame retardant (acc. IEC 60332-1 flame propagation on single cables) X Low smoke (according to IEC 61034) Zero Halogen (according to IEC 60754-1) X Non toxicity (according to IEC 60754-2) RoHS X Material Peek With SHV and Triax Lemo Sensor cable CC, 3 m Sensor cable CC, 5 m Sensor cable CC, 10 m Sensor cable CC, 15 m Sensor cable CC, 20 m Sensor cable CC, 25 m Sensor cable CC, 30 m Sensor cable CC, 35 m Sensor cable CC, 40 m Sensor cable CC, 45 m Sensor cable CC, 50 m Other cable length On request Delivery time ITER Status ITC230003 ITC230005 ITC230010 ITC230015 ITC230020 ITC230025 ITC230030 ITC230035 ITC230040 ITC230045 ITC230050 ITC230xxx under under Page 12 / 15
ITER irani Cable TPR018 ITER Pirani cable Max. bake out temperature 250 C 200 C Mechanical integrity 250 C 320 C ITER CC Cable Radiation resistant to 1x10 7 Gy 5x10 7 Gy Max. measuring cable length 100 100 Reduced flame propagation (according IEC 60332 or NF 32070 C1) X X Flame retardant (acc. IEC 60332-1 flame propagation on single cables) X X Low smoke (according to IEC61034) Zero Halogen (according to IEC 60754-1) X X Non toxicity (according to IEC 60754-2) X X RoHS X X Material PI Capton Special cable (Standard plugs controller side, gauge side P&P special) ITER Sensor cable TPR, 3 m, standard connector ITER Sensor cable TPR, 5 m, standard connector ITER Sensor cable TPR, 10 m, standard connector ITER Sensor cable TPR, 15 m, standard connector ITER Sensor cable TPR, 20 m, standard connector ITER Sensor cable TPR, 25 m, standard connector ITER Sensor cable TPR, 30 m, standard connector ITER Sensor cable TPR, 35 m, standard connector ITER Sensor cable TPR, 40 m, standard connector ITER Sensor cable TPR, 45 m, standard connector ITER Sensor cable TPR, 50 m, standard connector Other cable length On request X ITC130003 ITC130005 ITC130010 ITC130015 ITC130020 ITC130025 ITC130030 ITC130035 ITC130040 ITC130045 ITC130050 ITC130xxx ITER Cold Cathode Cable ITER Sensor cable CC, 3 m, SHV and triaxial connector ITER Sensor cable CC, 5 m, SHV and triaxial connector ITER Sensor cable CC, 10 m, SHV and triaxial connector ITER Sensor cable CC, 15 m, SHV and triaxial connector ITER Sensor cable CC, 20 m, SHV and triaxial connector ITER Sensor cable CC, 25 m, SHV and triaxial connector ITER Sensor cable CC, 30 m, SHV and triaxial connector ITER Sensor cable CC, 35 m, SHV and triaxial connector ITER Sensor cable CC, 40 m, SHV and triaxial connector ITER Sensor cable CC, 45 m, SHV and triaxial connector ITER Sensor cable CC, 50 m, SHV and triaxial connector Other cable length On request Delivery time ITER Status Under ITC240003 ITC240005 ITC240010 ITC240015 ITC240020 ITC240025 ITC240030 ITC240035 ITC240040 ITC240045 ITC240050 ITC240xxx Under Page 13 / 15
Cable Connectors Controller Side Cable Connectors Controller Side Pirani Connector at Controller - Amphenol C91 Pirani Connector at Controller Crimp contact C91 plug connector IT4722125 IT4722841 CC Plug at Controller 11_SHV-50- Peek CC Plug at Controller 11_SHV-50- Capton IT4726605 IT4728606 Delivery time ITER Status Under Under Cable Connectors Gauge Head Side Cable Connector Gauge Head Side Pirani Gauge Standard Plug Lemo 4.3 5.2 mm Pirani Gauge Standard Plug Lemo 5.3 6.2 mm IT4722062 IT4722063 CC Push Pull Triax Connector CC Push Pull Triax Connector New IT1001030 IT1001031 Delivery time ITER Status Under Under Page 14 / 15
Conversion Tables Mass kg lb slug oz cwt sh cwt kg 1 2.205 68.522 10-3 35.274 19.684 10-3 22.046 10-3 lb 0.454 1 31.081 10-3 16 8.929 10-3 10 10-3 slug 14.594 32.174 1 514.785 0.287 0.322 oz 28.349 10-3 62.5 10-3 1.943 10-3 1 0.588 10-3 0.625 10-3 cwt 50.802 112 3.481 1.792 10 3 1 1.12 sh cwt 45.359 100 3.108 1.6 10 3 0.893 1 Pressure N/m 2, Pa bar mbar Torr at lbf/in 2, psi N/m 2, Pa 1 10 10-6 10 10-3 7.5 10-3 9.869 10-6 0.145 10-3 bar 100 10 3 1 10 3 750.062 0.987 14.504 mbar 100 10-3 1 750.062 10-3 0.987 10-3 14.504 10-3 Torr 133.322 1.333 10-3 1.333 1 1.316 10-3 19.337 10-3 at 101.325 10 3 1.013 1.013 10 3 760 1 14.696 lbf/in 2, psi 6.895 10 3 68.948 10-3 68.948 51.715 68.046 10-3 1 Pressure Units used in the Vacuum Industry mbar Pascal Torr mmws psi inch of merc. mbar 1 100 750.062 10-3 10.2 14.504 10-3 2.95 10-2 Pascal 10 10-3 1 7.5 10-3 0.102 0.145 10-3 2.95 10-4 Torr 1.333 133.332 1 13.595 19.337 10-3 3.937 10-2 mmws 9.81 10-2 9.81 7.356 10-2 1 1.422 10-3 2.896 10-3 psi 68.948 6.895 10 3 51.715 703 1 2.036 inch of merc. 33.86 3.386 10 3 25.4 345 0.491 1 Dimension mm m inch ft mm 1 10-3 39.37 10-3 3.281 10-3 m 10 3 1 39.37 3.281 inch 25.4 25.4 10-3 1 8.333 10-2 ft 304.8 0.305 12 1 Temperature Kelvin Celsius Rankine Fahrenheit Kelvin 1 C+273.15 R 5/9 ( F+459.67) 5/9 Celsius K-273.15 1 ( R-491.69) 5/9 5/9 F-17.778 Rankine K 9/5 ( C 9/5)+491.69 1 F+459.67 Fahrenheit 9/5 K-459.67 9/5 ( C+17.778) R-459.67 1 Page 15 / 15