SUMMARY OF THE METHOD SAMPLE PREPARATION

Similar documents
Application Note: Using the Turner Designs Model 10-AU Fluorometer to Perform Flow Measurements in Sanitary Sewers by Dye Dilution

Quick Start Operating Instructions

Operation Manual. September 7, 2014 P/N Revision 1.2

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

800 Displaying Series Flowmeter

INSTALATION PROCEDURE

Fluorescence Spectrophotometer

bio bas.1 USER S MANUAL

Quick Start Function Summary Instructions for ASHCROFT GC52 Differential Pressure Transmitter Version 6.03 Rev. B

This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus:

RQflex 10. Manual English

Ambient NOx monitor APNA-370 Operation Manual CODE:GZ D

LAUREL ELECTRONICS, INC.

Analysis and sorting of cells with FACSAria II flow cytometer Tiina Pessa-Morikawa / Revised

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

WQ Series Water Quality Bench top Meters

ZX-44XL Liquid Fuel Analyzer. User s Manual Version 1.2

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

Ambient SO 2 /H 2 S monitor APSA-H370 Instruction Manual

Three Axis Digital Readout System

Performing a Sound Level Measurement

MF-PFT. Programmable differential pressure transmitter for pressure or flow measurement and control. DIL1 Output/Utsignal

TUSKIN Equipment Corporation

RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation

BLUNIK II ACCESSORIES SPORT DRIVE SPORT CALCULATOR... 49

Night Hawk Firing System User s Manual

Checking the fluids of the cart

FREQUENCY COUNTERS TO 18 GHZ USING THE DATUM FREQUENCY STANDARD

ED3. Digital Encoder Display Page 1 of 13. Description. Mechanical Drawing. Features

TABLE OF CONTENTS. Instructions:

User Calibration Software. CM-S20w. Instruction Manual. Make sure to read this before use.

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

Weighing Scale. 10key&12key User manual

Ambient NH 3 monitor APNA-370 Instruction Manual CODE:GZ

Type AKS 4100/4100U Coaxial D14 version

TC-1 Timeclock. Operating and Programming Instructions Thursday, 25 March Lighting Controls for the World we live in

AU5800 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU5800 Weekly (Analyzer and ISE) Maintenance

AH-PMP-1DA & AH-1DA Operator's Manual

1 Ver.mob Brief guide

DATA Count and Fill S 60

SMART II. GRAVIMETRIC CONTROL FOR Gravibelt & Plasticolor USER MANUAL

Blown Film Extrusion. Line Control. 5 Layers CD Series IBC Temperature USER MANUAL

Operating Manual B 300

SMART. GRAVIMETRIC CONTROL FOR BeltColor & GraviColor USER MANUAL. Manual Number: ST105 June, 2011

CARLO GAVAZZI Automation Components

Pin setup: 1 (EXC+) 2 (EXC-) 3 (SIG+) 4 (SIG-) Pin setup: 2 (TXD) 3 (RXD) 5 (GND) others (NC)

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

Smart Hawk Firing System User s Manual

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

BER MEASUREMENT IN THE NOISY CHANNEL

Section 2 Normal Procedures

ASSEMBLY AND CALIBRATION

Centurion. Guided Radar. Manual. A Higher Level of Performance. Sultan Sonar Manual Rev 1.0. CGR Series.

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB

Obtain Power Measurements of a DOCSIS Downstream Signal Using a Spectrum Analyzer

VLC-3 USER'S MANUAL. Light Program Controller. M rev. 04 K rev. 00 & ( ( 5, 352*5$0 1 : $ 2 ' 6(77,1*6 )81&7,216

Radio Thermostat Clock

Advanced Test Equipment Rentals ATEC (2832)

Quick-Start for READ30

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

FRQM-2 Frequency Counter & RF Multimeter

EDL8 Race Dash Manual Engine Management Systems

013-RD

Total Chlorine Sensor

Programming instructions Software Version 5

Manual. Simrad IS80 Heading Repeater HR80. English

127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) (800) (бесплатно на территории России)

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630)

USER MANUAL MODEL 650. TWO-WIRE ph/orp TRANSMITTER. um

Experiment 7: Bit Error Rate (BER) Measurement in the Noisy Channel

Keysight FieldFox Microwave Analyzers

Signal Stability Analyser

Model IQ4-PC User Manual Revision Date:

Short Manual. ZX2 Short Manual.doc Page 1 of 12

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

721N Spectrophotometer. Operating Manual

DEX Fruit & Stem Growth Dendrometer User Manual

COMPANY. MX 9000 Process Monitor. Installation, Operating & Maintenance Manual AW-Lake Company. All rights reserved. Doc ID:MXMAN082416

2009 MFJ ENTERPRISES, INC.

OPTICAL POWER METER WITH SMART DETECTOR HEAD

Approved by Principal Investigator Date: Approved by Super User: Date:

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Four-way Return Optical Receiver Manual. Transmitter OP-OR124RJ. Shenzhen Optostar Optoelectronics Co., Ltd (Version 1)

Sprite TL Quick Start Guide

DA-88 QUICK TIPS (VERSION 4 SOFTWARE)

SMART II. GRAVIMETRIC CONTROL FOR BeltColor & GraviColor USER MANUAL. Manual Number: SM II 102 December 2017

Figure 1: Standard 906 Sensor and Pulser Disc. Figure 2: Standard 906 Sensor and Pulser Wrap

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 3rd Edition

CN Remove the scanner assembly (X476 and X576 models) and all doors/covers.

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

eblot L1 Protein Transfer System Fast Wet Transfer System for Mini Gels

Dektak Step by Step Instructions:

0RGHO.20LFUR&HOO.LW $VVHPEO\DQG&KHFNRXW,QVWUXFWLRQV

ANALOG I/O MODULES AD268 / DA264 / TC218 USER S MANUAL

HulaMixer Sample Mixer

TRIMBLE GPS / 10MHz REFERENCE MONITOR DISPLAY V January 2015

OPERATING INSTRUCTIONS

Introduction...2. Features...2 Safety Precautions...2. Installation...4

STB Front Panel User s Guide

Transcription:

INSTRUMENT SET-UP: Model 10-AU Digital Fluorometer equipped with: 1. 13 mm cuvette holder; 2. Optical Filter Kit PN 10-040 or 10-040R, which includes: PN 10-113 (436 nm) Excitation Filter; PN 10-115 (680 nm) Emission Filter; PN 10-032 (1ND) Reference Filter; the PN 10-089 Blue Lamp (F4T4.5B2 equivalent). SUMMARY OF THE METHOD Conventional fluorescence methods for measuring chlorophyll a require samples to be measured twice; once before acidification and once afterwards. Under the most extreme ratio of chlorophyll a/chlorophyll b likely to occur in nature (1:1 molar), conventional acidification techniques result in approximately a 60% underestimate of true chlorophyll a. Under these conditions, the new method 1 described in these pages yields only a 10% overestimate of true chlorophyll a. In addition, it requires a single fluorescence determination and is sensitive enough for estimates of euphotic zone chlorophyll a in all marine and freshwater ecosystems. Filtration of less than 200 ml of water provides adequate sensitivity even in the most oligotrophic environments. The method requires: 1. Sample preparation. 2. Setting the basic operating level of your fluorometer (this only needs to be done once, prior to running samples for the FIRST time). 3. Calibration of the Fluorometer. 4. Reading samples. SAMPLE PREPARATION Detailed instructions for extracting chlorophyll a and measuring with the Turner Designs analog fluorometer can be found in United States Environmental Protection Agency (EPA) Method 445.0 "In Vitro Determination of Chlorophyll a and Pheophytin a in Marine and Freshwater Phytoplankton by Fluorescence." A copy is enclosed for your convenience. Method 445.0 sets forth the conventional fluorescence procedure, requiring two readings for each sample--before and after acidification. (Method 445.0 can be found in the EPA standard methods book, Methods for the Determination of Chemical Substances in Marine and Estuarine Environmental Samples.) PLEASE NOTE that the procedure described in these instructions is without acidification. The Model 10-AU must be configured with special optical filters and lamp which read chlorophyll a in the presence of chlorophyll b and pheopigments. These optical filters and lamp should be installed according to instructions in your Model 10-AU User's Manual ("Filter and Light Source Replacement" appendix). USING THIS METHOD, you must NOT acidify your samples as set forth in EPA Method 445.0, section 11.2.2; and you do not need to perform any calculations as required by section 12.1 of Method 445.0. 998-9000 Revision A Page 1 of 6

SETTING THE BASIC OPERATING LEVEL Application Note: The basic operating level of your fluorometer is set on screen 3.2 using the Sensitivity Adjustment Knob. Your Model 10-AU User's Manual contains detailed instructions for setting the basic operating level (Appendix 5B of the Model 10-AU-000 Laboratory Fluorometer User's Manual, and Appendix 6B of the Model 10-AU-005 Field Fluorometer User's Manual). Once the basic operating level (sensitivity) is set in this way, you should not have to adjust it unless you want to drastically increase or decrease sensitivity, or unless you change the lamp or filters. Fluorescence readout: XXXX.XXX* PM signal output: XXXX.XXX Cal std val: XXX.XX Blank: XX.XXX FS: XX% of XXXX.XXX (PPB) at HIGH Span level%: 48 <ENT> to next screen #3.2 Take care not to spill any liquids inside the fluorometer's sample chamber. If you do have a spill, turn off the fluorometer immediately and consult the Sample System appendix of your user's manual. Wipe up splashes on the instrument's exterior promptly. Through experience, we have found the following operating level to be appropriate for extracted chlorophyll. Have on hand one of the following: 50, or 75, or 100 g/liter (parts per billion) standard of chlorophyll a in 90% acetone. 1. Turn on the fluorometer and allow to warm up for 10 minutes. 2. Using the keypad, access screen 2.0, Calibration. Press <6> once and then <9> five times to reset calibration defaults. 3. From screen 2.0, press <4>, then <2> to access screen 2.42. Set the concentration range to HIGH. 4. Access screen 3.2, Diagnostic. Loosen (do not remove) the Sensitivity Knob Lock (hex screw to the left of the keypad; see Figure 1 in the user's manual) with the 5/32" Allen wrench (stored behind and to the lower right of the keypad). 5. Fill a clean 13 mm test tube with your standard. Dry the outside of the tube with a lab wipe and insert it in the opening to the sample compartment. Replace the light cap. 6. On screen 3.2, look at the third line from the bottom: FS: XX% of XXX ( g/l) at HIGH. Turn the Sensitivity Knob slowly, pausing between adjustments for the reading to stabilize, until the FS % reads: If you are using 50 ppb standard: FS: 30-50% of 900.000 at HIGH If you are using 75 ppb standard: FS: 50-70% of 900.000 at HIGH If you are using 100 ppb standard: FS: 70-90% of 900.000 at HIGH Turning the Sensitivity Knob clockwise increases the FS %; turning it counter-clockwise decreases FS %. 7. When desired FS % is reached, use the Allen wrench to tighten the Sensitivity Knob Lock. (IMPORTANT!) 8. Remove the standard from the fluorometer and set aside. 9. Before running samples, calibrate your fluorometer (see next section). 998-9000 Revision A Page 2 of 6

CALIBRATION Application Note: Once the basic operating level is set, all you need to do is calibrate the instrument with the following procedure. The calibration should remain stable for some time (at least two weeks), and unless you change your blank or standard or want to change from reading very high levels to very low levels (or vice versa), you do not have to calibrate every time you read a new batch of samples. (You will, of course, need to recalibrate if you reset the basic operating level, or change the lamp or filters.) Table 1 shows a recommended range and concentration of standard to use for reading various concentrations of sample. These are recommendations and not absolute concentrations; find what works for you. Please note, however, that the standard must be within the linear range for accurate readings (according to the EPA Method 445.0, using the 13 mm cuvettes, chlorophyll a is linear to 250 g/l). Resolution on samples of low concentration will be better if you calibrate with a standard not too much more concentrated than the samples. For example, if you want to read samples less than 3 g/l, you will get better resolution calibrating with a 3-10 g/l standard on the LOW or MEDIUM range than with a 75 g/l standard on the HIGH range. You will note, however, that you will probably still obtain good results reading low samples, even if you have calibrated with a high standard. (The fluorometer is designed to give good range-to-range correlation.) To calibrate: Have ready a blank of 90% acetone and your standard of known concentration of pure chlorophyll a in 90% acetone. 1. Turn on the fluorometer and allow it to warm up for 10 minutes. TABLE 1. Recommended Standard Concentration and Fluorometer Range Concentration Standard Fluorometer to be Read Concentration 2 Conc. Range 0-3.75 g/l 2.1-3.75 g/l LOW 0-37.5 g/l 21-37.5 g/l MED 0-100 g/l 75-100 g/l HIGH 0-250 3 g/l 200 g/l HIGH 2. To begin calibration, from the Main Menu, press <2> to access screen 2.0, Calibration. 3. Set the concentration range control to MAN. From screen 2.0, press <4> to bring up screen 2.4, then <3> to bring up screen 2.43 (set conc. range control), and press <ENT> to toggle. 4. Refer to Table 1 for the concentration range (LOW, MED, HIGH). To change to the desired concentration range, from screen 2.4, press <2> to bring up screen 2.42 (change concentration range), and press <ENT> to toggle. Return to screen 2.0. 5. From screen 2.0, press <2> to access screen 2.2 (standard solution concentration). Enter the actual concentration for your standard. I.e., 3, 25, 75, etc. See Table 1. (This number must be less than 1000.) 6. To Run Blank, press <1> to access screen 2.1. Make sure #2 on screen 2.1 reads YES. Then from screen 2.1, press <1> to call up screen 2.11. If you wish to abort the blank run and revert to the former calibration settings, press <ESC> before pressing <0>. This will retain the current settings for Span and blank. Because temperature affects fluorescence, do not allow the blank to remain in the instrument any longer than necessary for a stable reading. Use UP or DOWN arrow to set Blank % to less than 200%! Blank%: TC: 8 (SEC) Range: MED Span: 48% Press <0> after reading is stable! #2.11 998-9000 Revision A Page 3 of 6

Fill a clean 13 mm test tube with the blank of 90% acetone. Put the blank in the sample chamber and replace the light cap. After the Blank % reading is stable ("TC" on screen 2.11 cycles from 1 to 8 sec) and assuming the Blank % is less than 200%, press <0>. When "FINISHED" appears, press <ESC>. If Blank % is not less than 200%, reduce the Span% (fine tuning of sensitivity) by pressing the down arrow until the Blank % is less than 200%. Then press <0>. When "FINISHED" appears, press <ESC>. Remove the blank. Set cuvette aside. 7. Run Standard solution. On screen 2.0, press <3> to bring up screen 2.3. If you wish to abort the standard run and revert to the former calibration settings, then press <ESC> before pressing <*>. This will retain the current settings for the Span and standard. FS is the maximum concentration or relative fluorescence you will be able to read on a particular range, and it is not necessary or likely that FS match the value of the standard. For example, if you are calibrating on the MED range with a 25 g/l standard, you would want to adjust the Span % until the FS for the MED range is approximately 30 (meaning you will be able to read samples up to 30 g/l on the MED range). Read help <?> before running Standard! Use UP or DOWN arrow to adjust Table RANGE: High Medium Low FS (PPB): TC: 8 (SEC) Range: MED Span: 48% Press <*> after reading is stable! #2.3 Because temperature affects fluorescence, do not allow the standard to remain in the instrument any longer than necessary for a stable reading. Fill a clean 13 mm test tube with your pure chlorophyll a standard (See Table 1). Put the standard in the sample chamber and replace the light cap. Adjust the Span % using the UP and DOWN arrows until the FS readings for each range are satisfactory. Pressing the UP arrow increases Span % (sensitivity), but decreases FS. Pressing the DOWN arrow decreases Span %, but increases FS. Wait until readings are stable ("TC" on screen cycles from 1 to 8 sec), then press <*>. When "FINISHED" appears, press <ESC>. If the LOW range reads OVER, with the Span % close to 0, it means the standard concentration is too high for the LOW range. Press <ESC>, change to the MED range on screen 2.42, and return to step 6, Calibration. Similarly, if the MED range reads OVER, with the Span close to 0%, it means the standard concentration is too high for the MED range: press <ESC>, change to the HIGH range on screen 2.42, and repeat steps 6-8, Calibration. Refer to Section 2G, subsection 2d of the user's manual for an explanation of minus (-) and BLK>FS. Remove the standard. Set cuvette aside. 8. Press <HOME> to return to the HOME screen, and you are ready to begin reading samples (see next section). READING SAMPLES Refer to your user's manual, Section 3C, Routine Operation, for additional details. After calibration is complete, you may begin reading your extracted samples on the HOME screen. 998-9000 Revision A Page 4 of 6

For your convenience, the Model 10-AU has a "Discrete Sample Averaging" capability, where the instrument averages a reading over a preset period, allowing you to read samples after they have been in the instrument for the same amount of time. This CONC: MED (MAN) XXX (PPB) removes the guesswork from reading the digital display and Time Const: 2 (SEC) minimizes error due to temperature changes. Defaults for the Model 10-AU are 15 seconds pre-delay for the signal to stabilize, and an averaging period of 10 seconds. (To change 0 499 999 these periods, access screen 1.63. See appendix on Optional Parameters in your user's manual) To use Discrete Sample Averaging, after putting in your sample, from the HOME screen, <?> for help 4:42:05 PM 1/25/93 press <*> and the instrument will countdown a delay period, average the reading, and then display "DONE" in the upper right hand corner of the screen. The averaged reading will be displayed for 10 seconds. Log the reading. Procedure for running samples: 1. It is suggested that you read samples in the manual mode (MAN, set on screen 2.43). For the best resolution, read your sample on the lowest range in which you can get an on-scale reading. (You can see what the full scale values are for each range under the current calibration by pressing the LEFT arrow while on the HOME screen.) 2. Fill a clean cuvette with a sample, wipe the outside of the cuvette dry with a lab wipe, and place in the instrument. Replace the light cap. 3. Wait about 15 seconds for the reading to stabilize, and log the reading. (Remember: Because of temperature effects, for greatest accuracy, read all samples after they have been in the fluorometer for approximately the same length of time.) If the display reads "OVER," and you are in the LOW range, go to screen 2.42 and change to the MED range. If OVER in the MED range, go to screen 2.42 and change to the HIGH range. If you read OVER on the HIGH range, it means the sample is too concentrated to be read at the current calibration (and probably over the linear range anyway). In this case, dilute the sample by 25% (1 part sample to 3 parts blank water), and read it. Multiply the reading by 4 to get the actual concentration. 4. Remove the cuvette and put in the next sample. 5. Repeat steps 2-3 until all samples are read. These readings are the actual concentration of extracted chlorophyll a in the cuvette. To arrive at the environmental chlorophyll a, for each sample you must correct for the volume of water filtered and the volume of 90% acetone used in the extraction. 6. Turn off the fluorometer. NOTE: It won't hurt the fluorometer to leave it on all day. If you are going to be reading samples off-and-on over the course of a few days, it is better to leave the fluorometer on. 998-9000 Revision A Page 5 of 6

NOTES 998-9000 Revision A Page 6 of 6