XRD-DSC Sample Alignment (BB) Part

Similar documents
HV/PHA Adjustment (PB) Part

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

HRXRD Analysis of Epitaxial Thin Films

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Dektak Step by Step Instructions:

Transient Stability Events & Actions

Florida State University Thayumanasamy Somasundaram

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

PHI 5000 VersaProbe TM Operator s Guide

Using the Agilent for Single Crystal Work

DektakXT Profilometer. Standard Operating Procedure

Standard Operating Procedure of nanoir2-s

SE-SMD0603-PW65 SMD 0603 purweiss, wasserklar

TOMELLERI ENGINEERING MEASURING SYSTEMS. TUBO Version 7.2 Software Manual rev.0

MestReNova Manual for Chem 201/202. October, 2015.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

Data Collection Step Size Selection HOW TO DECIDE STEP SIZE, TIME PER STEP, AND COLLECTION TIME ON PANALYTICAL X-PERT PRO POWDER

SURFACE MOUNT LED LAMP 0603 (0.8 mm Height)

Preparing for remote data collection at NE-CAT

Dynamic Cable Rating Model Guidelines

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:

SURFACE MOUNT LED LAMP STANDARD BRIGHT 0606

SURFACE MOUNT LED LAMP STANDARD BRIGHT 1210

SURFACE MOUNT LED LAMP STANDARD BRIGHT PLCC-2

SEM- EDS Instruction Manual

SURFACE MOUNT LED LAMP FULL COLOR 1210

potentiostat/galvanostat

Fiber Optic Extender User Manual. Model: CV-F01-TX

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

Sound Creation Tool FWCTRL Operation Manual

8000 Plus Series Safety Light Curtain Installation Sheet ( CD206A/ CD206B )

Syntor X Flash Memory Module Revision C

How to Optimize Ad-Detective

User's Guide and Reference

R&S RT-Zxx High-Bandwidth Probes Specifications

SURFACE MOUNT LED LAMP STANDARD BRIGHT 1206 (Reverse Mount)

Introduction. Notes on Reading This Manual

The HSMB-C110 is a right-angle package with the universally accepted dimensions of 3.2 x 1.0 x 1.5 mm. This part is ideal for

S195AVGC-2BM 1.6x0.8mm, Red & Yellow Green LED Surface Mount Bi-Color Chip LED Indicator

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

Contents. 1. System Description 3. Overview 3 Part Names 3 Operating Conditions 7 Start-up Procedure 7. 2.

This guide gives a brief description of the ims4 functions, how to use this GUI and concludes with a number of examples.

TT AFM LongBeach Procedures and Protocols V2.1

Figure 1. MFP-3D software tray

Do not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company.

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

CNC872 Series CNC System Application for Cutting Machines

Import and quantification of a micro titer plate image

INSTRUMENT CATHODE-RAY TUBE

1.10mm Height 1210 Package. Bi-Color (Multi-Color) Chip LED. Technical Data Sheet. Part No: S155VBC-V12B-B41B

HUIYUAN OPTO-ELECTRONIC CO.,LTD

PLCC Series EMC 3030 Single color

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0

Case study: how to create a 3D potential scan Nyquist plot?

3D QUALITY CONTROL SOFTWARE MPES-3DQC1 USER S GUIDE. [English] 1st Edition (Revised 1) Software Version 1.10

OM2000N INSTALLATION MANUAL

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

Renishaw Ballbar Test - Plot Interpretation - Mills

JAMAR TRAX RD Detector Package Power Requirements Installation Setting Up The Unit

Differential Scanning Calorimetry (DSC) Of Powder Coating Materials

R5 RIC Quickstart R5 RIC. R5 RIC Quickstart. Saab TransponderTech AB. Appendices. Project designation. Document title. Page 1 (25)

Software manual. ChipScan-Scanner 3.0

The parameters can be changed according to the needs of the operator. Do not change the settings of the Maintenance screen.

LCD MODULE SPECIFICATION

Mini Micro Pulse Lidar System

Installation Guide AT-HDR-EX-70-2PS

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

3mm Round White LED T-1. Technical Data Sheet. Part No: LL-304WC2E-W2-3TC

Features: Descriptions: Applications:

EM1. Transmissive Optical Encoder Module Page 1 of 9. Description. Features

Characterization and improvement of unpatterned wafer defect review on SEMs

Chapter 4: Entering Bridge Geometry Using DCBRIDGE By Karl Hanson, S.E., P.E. July 2006

Matrox PowerStream Plus

Madell Automatic Pick and Place Machines User s Manual

OPERATION MANUAL. TECHNOS Co.,Ltd. Total Reflection X-ray Fluorescence Spectrometer

Quick Start Bruker Dimension Icon AFM

LCOS-SLM Liquid Crystal Based Spatial Light Modulator OPERATIONAL MANUAL

Scout 2.0 Software. Introductory Training

F7000NV ROBOT VISION OPERATING MANUAL

Design Studies For The LCLS 120 Hz RF Gun Injector

What is new in UNIFIT 2018?

AN-822 APPLICATION NOTE

Luckylight Package Pure Green Chip LED. Technical Data Sheet. Part No.: S150PGC-G5-1B

3mm Round Blue LED T-1 Technical Data Sheet

User Manual OVP Raman

YJ-RGB5050L-G01 RGB LED

TUTORIAL IGBT Loss Calculation in the Thermal Module

YJ-BC-3030-G01 High CRI LED

Luckylight Package Warm White Chip LED. Technical Data Sheet. Part No.: S150W-W6-1E

Luckylight. 1.10mm Height 0805 Package. Warm White Chip LED. Technical Data Sheet. Part No.: S170W-W6-1E

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

GVD-120 Galvano Controller

Mechanical aspects, FEA validation and geometry optimization

There are many ham radio related activities

Operating Procedures for RECO1 & RECO2

De-embedding Gigaprobes Using Time Domain Gating with the LeCroy SPARQ

СВЕТОДИОДЫ BEELED - ТЕХНИЧЕСКОЕ ОПИСАНИЕ

INSTRUMENT CATHODE-RAY TUBE

Matrox PowerStream Plus

Transcription:

XRD-DSC Sample Alignment (BB) Part

Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 5 2. Sample alignment sequence...9 XRD-DSC Sample Alignment (BB) Part i

1.1 Setting conditions 1. How to set Part conditions In this chapter, how to set the XRD-DSC Sample Alignment (BB) Part conditions is described. 1.1 Setting conditions Set the basic conditions in the XRD-DSC Sample Alignment (BB) dialog box. The scan conditions and slit conditions of the sample alignment are determined based on the basic conditions. The scan conditions and slit conditions can also be customized. Fig. 1.1.1 XRD-DSC Sample Alignment (BB) dialog box Flat sample Performs the sample alignment according to the sequence described in Chapter 2. Run recommended sequence Performs the sample alignment using the recommended sequence. Customize conditions Customize Performs the sample alignment under the conditions specified in the Customize dialog box. When the Customize conditions radio button is selected, click the Customize button, then set the scan conditions and slit conditions. 1.2 Customizing scan conditions and slit conditions Tip: It takes about 10 minutes for the sample alignment to complete. XRD-DSC Sample Alignment (BB) Part 1

1. How to set Part conditions Put a sample when the sample alignment starts. If the sample is mounted (or changed) before the sample alignment starts, check the Put a sample when the sample alignment box. If the sample alignment is performed using the currently-mounted sample, uncheck the Put a sample when the sample alignment box. Put a sample every time the sample alignment starts in a repeated measurement. If the sample is changed every time the sample alignment starts in a repeated measurement using the loop Parts or XY mapping Parts, check the Put a sample every time the sample alignment starts in a repeated measurement box. If the sample is mounted during the first cycle only, uncheck the Put a sample every time the sample alignment starts in a repeated measurement box. Print alignment result Check the Print alignment result box to print out the alignment results with the format shown below after alignment. 2 XRD-DSC Sample Alignment (BB) Part

1.1 Setting conditions Execute Executes the sample alignment under the conditions specified in the XRD-DSC Sample Alignment (BB) dialog box. CAUTION: Clicking the Cancel button after executing the sample alignment does not cancel the specified conditions. Tip: The sample alignment is executed with the XRD-DSC Sample Alignment (BB) dialog box open. While the sample alignment is running, the Part conditions cannot be changed. They can be changed again after the alignment has been completed. The setting of invalid. (Show confirmation messages) on the flow bar becomes Import Loads the saved Part conditions. Clicking the Import button opens the Import dialog box. Select the folder including the file you want to import from the Folder tree view. In the Condition file list, select the condition file you want to import and click the OK button to load the Part conditions. XRD-DSC Sample Alignment (BB) Part 3

1. How to set Part conditions Export Saves the specified Part conditions in a file. Clicking the Export button opens the Export dialog box. From the Folder tree view, select a destination folder to save the conditions file then enter a file name in the File name box. Enter comments in the Memo box, if needed. After entering them, click the OK button. OK Sets the conditions and closes the dialog box. Cancel CAUTION: When selecting another Package measurement or switching the task to the Manual Control task, etc. the specified conditions will be cancelled. To save the specified conditions in a file, click the Export button and save the conditions. Does not set the conditions and closes the dialog box. Tip: Clicking the Cancel button also cancels the conditions specified in the Customize dialog box.? Opens the online help of this Part. 4 XRD-DSC Sample Alignment (BB) Part

1.2 Customizing scan conditions and slit conditions 1.2 Customizing scan conditions and slit conditions If you want to customize the scan conditions and slit conditions of the sample alignment, set the conditions in the Customize dialog box. Tip: Refer to 2. Sample alignment sequence to set the scan conditions and slit conditions. Fig.1.2.1 Customize dialog box XRD-DSC Sample Alignment (BB) Part 5

1. How to set Part conditions Start (mm) Stop (mm) Step (mm) Speed (mm/min) Start (deg) Stop (deg) Step (deg) Speed (deg/min) Enter the start position of the first and second Z scans. The value entered here will be used only when Curved sample (Z scan only) is selected. Enter the stop position of the first and second Z scans. The value entered here will be used only when Curved sample (Z scan only) is selected. Enter the step width of the first and second Z scans. The value entered here will be used only when Curved sample (Z scan only) is selected. Enter the scan speed of the first and second Z scans. The value entered here will be used only when Curved sample (Z scan only) is selected. Enter the start angle of the first omega scan. Enter the stop angle of the first omega scan. Enter the step width of the first omega scan. Enter the scan speed of the first omega scan. 6 XRD-DSC Sample Alignment (BB) Part

1.2 Customizing scan conditions and slit conditions Start (mm) Stop (mm) Step (mm) Speed (mm/min) Enter the relative distance of the start position of the final Z scan from the direct-beam-half-cut position determined by the second Z scan. Enter the relative distance of the stop position of the final Z scan from the direct-beam-half-cut position determined by the second Z scan. Enter the step width of the final Z scan. Enter the scan speed of the final Z scan. Peak position +/- FWHM X times Determine the start and stop angles of the final omega scan based on the peak position and FWHM obtained by the first omega scan. Select the X value from 0.5, 1.0, 1.5, 2.0, 3.0, or 5.0. Stop (deg) Speed (deg/min) Enter the step width of the final omega scan. Enter the scan speed of the final omega scan. XRD-DSC Sample Alignment (BB) Part 7

1. How to set Part conditions Set recommended values Sets the conditions in the Direct beam half cut alignment measurement conditions and Direct beam half cut alignment measurement conditions (final conditions) sections to the recommended values. The recommended values are shown in Table 1.2.1. Table 1.2.1 Recommended values of direct beam half cut alignment measurement conditions and its final conditions Scan Direct beam half cut Range Start Stop Step Speed axis alignment measurement Z Absolute -4 mm 1 mm 0.04 mm 3 mm/min conditions Omega Absolute -1.5 deg 1.5 deg 0.01 deg 10 deg/min Direct beam half cut alignment measurement conditions (final conditions) Omega Relative Scan axis Range Start Stop Step Speed Z Relative -0.3 mm 0.3 mm 0.004 mm 1 mm/min Peak position +/ 0.004 deg 1 deg/min FWHM x 1 Close Closes the Customize dialog box. 8 XRD-DSC Sample Alignment (BB) Part

2. Sample alignment sequence The sample alignment is performed automatically. However, the sample and reference sample must be placed (or removed) manually as instructed by messages displayed on the screen. Described below is the general optics alignment sequence. (1) Place the sample and reference sample as instructed by the messages displayed on the screen. (2) Drive the omega axis to the predetermined position for the current optics. Adjust each slit width as shown in Fig. 2.1. Ts θs (predeterminied position) Zs Zd Attenuator : ATT θd (predeterminied position) Incident slit : IS Receiving slit # 1 : RS1 (0.1 mm) (7 mm) Receiving slit # 2 : RS2 (0.15 mm) Fig. 2.1 Setting of the initial positions of the goniometer axes (3) Set the count mode of the counter to integral, the HV value to that registered in the system parameter management database, the PHA baseline to 500 mv, and the PHA window to 1000 mv. (4) Scan the Z axis, then drive the Z axis to the position at which the incident x-ray beam is blocked and reduced to 50% intensity by the sample. Ts Zs Zd ATT θs θd IS Z scan RS1 RS2 Fig. 2.2 Alignment based on Z scan (5) Scan the omega axis, then drive the omega axis to the profile peak position. Ts Zs Zd ATT θs θd IS ω scan RS1 RS2 Fig. 2.3 Alignment based on omega scan XRD-DSC Sample Alignment (BB) Part 9

2. Sample alignment sequence (6) Repeat steps (4) and (5). (7) Scan the Z axis again, then drive the Z axis to the position at which the incident x-ray beam is blocked and reduced to 50% intensity by the sample. Ts Zs ATT θs θs θd IS Z scan RS1 RS2 Fig. 2.5 Alignment based on Z scan (8) The position of the omega axis determined in step (6) is the position at which the sample surface is parallel to the incident x-ray beam. Set the omega zero offset value so that the position is 0. The position of the Z axis determined in step (7) is the direct-beam-half-cut position of the sample. (9) Set the count mode of the counter to differential. Restore the values of the HV, PHA baseline, and PHA window to those registered in the system parameter management database. 10 XRD-DSC Sample Alignment (BB) Part