Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards

Similar documents
Oscilloscope Measurement Tools to Help Debug Automotive Serial Buses Faster

Oscilloscope Display Quality Impacts Ability to View Subtle Signal Details

Time-Saving Features in Economy Oscilloscopes Streamline Test

FlexRay Physical Layer Eye-diagram Mask Testing

Memory-Depth Requirements for Serial Data Analysis in a Real-Time Oscilloscope

CAN, LIN and FlexRay Protocol Triggering and Decode for Infiniium 9000A and 9000 H-Series Oscilloscopes

RS-232/UART Triggering and Hardware-Based Decode (N5457A) for Agilent InfiniiVision Oscilloscopes

Agilent 87405C 100 MHz to 18 GHz Preamplifier

Exceptional performance

CAN/LIN Measurements (Option AMS) for Agilent s InfiniiVision Series Oscilloscopes

Keysight Technologies Oscilloscope Display Quality Impacts Ability to View Subtle Signal Details. Application Note

Logic Analysis Fundamentals

Keysight Technologies Using Oscilloscope Segmented Memory for Serial Bus Applications. Application Note

Agilent Series Harmonic Mixers

Agilent Understanding the Agilent 34405A DMM Operation Application Note

Selecting the Right Oscilloscope for Protocol Analysis Applications

Agilent N5183A MXG Microwave Signal Generator

Agilent Technologies N5454A Segmented Memory Acquisition for Agilent InfiniiVision Series Oscilloscopes

Keysight Technologies RS-232/UART Triggering and Hardware-Based Decode (N5457A) for InfiniiVision Oscilloscopes

Keysight Technologies Oscilloscope Measurement Tools to Help Debug Automotive Serial Buses Faster. Application Note

Agilent 86120B, 86120C, 86122B Multi-Wavelength Meters. Data Sheet

Agilent N4876A 28 Gb/s Multiplexer 2:1

Keysight Technologies Oscilloscope Memory Architectures Why All Acquisition Memory is Not Created Equal. Application Note

Keysight Technologies Segmented Memory Acquisition for InfiniiVision Series Oscilloscopes. Data Sheet

Agilent N4876A 28 Gb/s Multiplexer 2:1

Keysight Technologies Achieve High-Quality Compliance Test Results Using A Top-Quality Test Fixture. Application Note

Evaluating Oscilloscopes for Best Signal Visibility

Mixed Analog and Digital Signal Debug and Analysis Using a Mixed-Signal Oscilloscope Wireless LAN Example Application

Keysight Technologies N4974A PRBS Generator 44 Gb/s. Data Sheet

Keysight Technologies Decoding Automotive Key Fob Communication based on Manchester-encoded ASK Modulation

Keysight Technologies Multi-Channel Audio Test using the Keysight U8903A Audio Analyzer

Evaluating Oscilloscope Vertical Noise Characteristics

Debugging Memory Interfaces using Visual Trigger on Tektronix Oscilloscopes

Agilent 87075C 75 Ohm Multiport Test Sets for use with Agilent E5061A ENA-L Network Analyzers

Keysight Technologies ad Integrated RF Test Solution

Agilent ESA Series Spectrum Analyzers

Agilent Migration from 8712/8714 Series to ENA-L Network Analyzers

Identifying Setup and Hold Violations with a Mixed Signal Oscilloscope APPLICATION NOTE

FlexRay Measurements (Option FLX) for Agilent s InfiniiVision Series Oscilloscopes

Analyzing 8b/10b Encoded Signals with a Real-time Oscilloscope Real-time triggering up to 6.25 Gb/s on 8b/10b encoded data streams

Troubleshooting Analog to Digital Converter Offset using a Mixed Signal Oscilloscope APPLICATION NOTE

Automated Limit Testing

Keysight N9355/6 Power Limiters 0.01 to 18, 26.5 and 50 GHz High Performance Power Limiters. Technical Overview

Limit and Mask Test Application Module

Agilent M9362A-D01-F26 PXIe Quad Downconverter

Agilent Technologies Series of RF and Universal Frequency Counter/Timers. Family Overview

Keysight N9355/6 Power Limiters 0.01 to 18, 26.5 and 50 GHz High Performance Power Limiters. Technical Overview

Keysight Technologies High-Power Measurements Using the E5072A ENA Series Network Analyzer. Application Note

Keysight Technologies Evaluating Oscilloscope Vertical Noise Characteristics. Application Note

Network Line Card Testing using the TDS3000B DPO Application Note. Line Card Testing Example: Throughput = Shippable Dollars

Electrical Sampling Modules Datasheet 80E11 80E11X1 80E10B 80E09B 80E08B 80E07B 80E04 80E03 80E03-NV

A Simple, Yet Powerful Method to Characterize Differential Interconnects

Keysight N1085A PAM-4 Measurement Application For 86100D DCA-X Series Oscilloscopes. Data Sheet

Keysight Technologies

Quick Signal Integrity Troubleshooting with Integrated Logic Analyzers & Oscilloscopes

Agilent 83000A Series Microwave System Amplifiers

Logic Analyzer Triggering Techniques to Capture Elusive Problems

Keysight Technologies Understanding and Improving Network Analyzer Dynamic Range. Application Note

E6607A EXT Wireless Communications Test Set. Non-signaling Test Overview. Application Note

Debugging a Mixed Signal Design with a Tektronix Mixed Signal Oscilloscope

SignalCorrect Software and TCS70902 Calibration Source Option SC SignalCorrect software

Keysight Technologies Millimeter Wave Frequency Extenders From Virginia Diodes Inc. for the Keysight X-Series Signal Analyzers. Technical Overview

Keysight Technologies Mixed Analog and Digital Signal Debug and Analysis Using a Mixed-Signal Oscilloscope

40 Gb/s PatternPro Programmable Pattern Generator PPG4001 Datasheet

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool

40 Gb/s PatternPro Programmable Pattern Generator PPG4001 Datasheet

Agilent N4965A Multi-Channel BERT 12.5 Gb/s Data Sheet

Black and Frozen Frame Detection

Using Triggered Video Capture to Improve Picture Quality

Agilent V3500A Handheld RF Power Meter

Meeting Embedded Design Challenges with Mixed Signal Oscilloscopes

Keysight Technologies

Agilent E6701G GSM/GPRS and E6704A EGPRS Lab Applications

PAM4 Transmitter Analysis

Keysight Technologies Intrinsic Contact Noise: A Figure of Merit for Identifying High Resolution AFMs. Application Note

Agilent N7744A 4-Channel Optical Multiport Power Meter N7745A 8-Channel Optical Multiport Power Meter. Fully compliant to LXI Class C specification

Video Reference Timing with Tektronix Signal Generators

Low Cost, High Speed Spectrum Analyzers For RF Manufacturing APPLICATION NOTE

Spearhead Display. How To Guide

46 GBaud Multi-Format Optical Transmitter

Fast Ethernet Consortium Clause 25 PMD-EEE Conformance Test Suite v1.1 Report

Agilent E3238 VHF/UHF Voice Activity Detection System

Timesaving Tips for Digital Debugging with a Logic Analyzer

Accuracy Delta Time Accuracy Resolution Jitter Noise Floor

Memory Interface Electrical Verification and Debug

Keysight 1GC DC GHz Packaged Biasable Integrated Diode Limited

Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes

Electrical Sampling Modules

C-PHY Essentials Transmitter Test Solution TekExpress C-PHY Essentials Tx

Agilent I 2 C Debugging

Portable Performance for Debug and Validation

Agilent N9344C Handheld Spectrum Analyzer (HSA)

The Measurement Tools and What They Do

Optical Sampling Modules 80C01 80C02 80C07B 80C08C 80C10 80C11 80C12

Agilent CSA Spectrum Analyzer N1996A

BFI RESEARCH AND STATISTICS PUBLISHED AUGUST 2016 THE UK FILM MARKET AS A WHOLE. Image: Mr Holmes courtesy of eone Films

The Benefits of External Waveform Monitors in Color Correction for Video. Application Note

Memory Interface Electrical Verification and Debug

How-To Guide. LQV (Luminance Qualified Vector) Measurements with the WFM8200/8300

Agilent E4887A HDMI TMDS Signal Generator Platform

Transcription:

Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards Application Note Introduction Engineers use oscilloscopes to measure and evaluate a variety of signals from a range of sources. Oscilloscopes play a major role in both design and manufacturing presenting an engineer with a visual display of voltage over time. It is crucial to have an oscilloscope that will not only provide accurate visual representations of the signal, but will also provide usability and functions that makes designing and testing signals easier. Every manufacturer must address quality assurance (QA) to ensure products consistently meet standards for design and performance and maintain a low defect rate. QA is a systematic approach to designing and providing processes to ensure confidence in and sustainability of a product s quality. As part of the QA process, you must test and verify product operation many times at each stage in design and validation. For products that generate electronic signals such as integrated chips (ICs) and field programmable logic arrays (FPGAs), you must verify that each signal operates within required tolerances for signal integrity. These tolerances, measured with an oscilloscope, include various signal attributes such as peak-to-peak voltage, frequency, RMS voltage, and delay. The ideal QA process for products that use electronic signals would test signals for high quality standards (low failure rates) utilizing a large number of samples, without impacting test times. Such a QA process would yield three key benefits: 1) high quality products in less time, speeding up time to market, 2) greater customer satisfaction, and 3) allow designers and engineers to spend more time creating new designs rather than fixing old ones. This application note discusses the statistical principles that apply to the QA process and the evolution of Six Sigma efficiency defined as achieving a defect rate of 3.4 per million or less. We will then explore the benefits that mask testing brings to the QA process for electronic signals and its use to achieve Six Sigma quality in as little as 1.1 seconds.

Gaussian Theory To use QA processes effectively we first need to understand statistical theory and the role it plays in modeling a process. QA is not only concerned about testing and validating a product s performance; it is a methodical way of developing techniques that increase the probability your product will function as specified. The best way to verify functionality is to make many observations of your product s characteristics and see how these characteristics vary from unit to unit. To do this, you need to measure the critical characteristics and build a database or histogram of your results. Figure 1: After 50,000 samples, the histogram of turtle shell diagrams acquires a curved shape Building a statistical database begins with taking many samples of a specific measurement. For example, we could measure the shell diameter of a specific species of turtle to the nearest millimeter and then plot the shell sizes (x-axis) against the number of occurrences of that shell size (y-axis). We would find that certain shell sizes occur more frequently than others. We would also find that as we plot the diagrams of more and more turtle shells, the histogram takes on a curved shape, as shown in Figure 1. The plot may take on a variety of shapes depending on whether each sample is independent of the next. To simplify and broaden the application of this example, however, assume that each sample is independent and randomly chosen (in other words, an independent random variable). If these two requirements hold true, the plot will form a curve shaped like a bell (see Figure 1). This bell curve is known as a Gaussian (or normal) distribution. Gaussian distributions are important in the study of statistical probabilities and are commonly observed in nature. The thicknesses of zebra stripes, the height of pine trees, and the height of humans are all naturally influenced measurements that result in Gaussian distributions. The same can be said of the characteristic and shapes of most electronic signals. 2

Gaussian Theory (cont.) From the Gaussian distribution of turtle shell sizes, we can calculate a variety of significant statistical values that describe the shape of the bell curve. The distribution will be centered at the mean value, and will diverge from the mean at a standard deviation value related to the width of the bell. In a Gaussian distribution, 68.27% of all the shells measured (samples) will occur between one standard deviation to the left and right of the mean, as shown in Figure 2. As the number of standard deviations from the mean increases, the percentage of samples occurring between the standard deviation increases. 0.4 0.3 0.2 0.1 0.0 34.1% 34.1% 2.1% 2.1% 0.1% 13.6% 13.6% 0.1% -3σ -2σ µ -1σ 1σ 2σ 3σ Figure 2: In a perfect Gaussian distribution 68.27% of all samples occur between one standard deviation to the left and right of the mean Six Sigma for Electronic Signals In the 1980s, Motorola developed a business management strategy that used the 6σ value and applied it to manufacturing processes and business practices. This practice of achieving Six Sigma quality was very successful and has become a popular technique for increasing efficiency. The accepted standard of achieving Six Sigma is a process that produces no more than 3.4 defects per million occurances (DPMO). This is equivalent to having 3.4 outliers for every 1,000,000 samples taken. How can we apply the Six Sigma process to evaluating electronic signals? Imagine that instead of measuring turtle shell diameters, we measure specific attributes of each waveform of a signal. We could then determine if the attribute we are measuring meets or exceeds specified tolerances. To achieve Six Sigma quality standard for the signal, we would need to measure 1,000,000/3.4= 297,117 independently- acquired waveforms. If we measure 294,117 waveforms and no defects occur, then the product is quality assured to a Six Sigma standard. Measuring and analyzing 294,117 waveforms can be time consuming. Automated test benches using oscilloscopes on the market today can be set up to constantly measure and record data, but acquiring and analyzing 294,117 waveforms can sometimes require many hours or perhaps days of testing time to complete. After the automated tests have evaluated the waveforms you would need to check to see if any waveforms failed the specified tolerances and possibly repeat this procedure. This requires significant time spent testing signals and waiting for results, slowing time to market. 3

Mask Testing Another way to test your signal is with the mask testing feature of an oscilloscope. Instead of checking various parameters such as peak-topeak voltage, frequency, RMS voltage, and delay, what if you could simply compare your signal against a correct waveform? Mask testing allows just that. You can set up a pass/fail mask or tolerance/limit band around your signal, specifying time and voltage regions that your signals can and cannot enter, as Figure 3 illustrates. If a glitch occurs in the signal under test and the waveform enters into the mask region, the oscilloscope alerts the user to this error. You can also set up masks around areas of the signal that should be off-limits. That way you do not need a perfect signal; just an understanding of how a signal looks when it functions correctly, as shown in Figure 4. Figure 3: Mask testing compares your signal with correct signal and alerts you if your signal passes outside specific limits Figure 4: Mask testing lets you specify areas around your signal that should be off limits 4

Mask Testing (cont.) Mask testing allows for flexibility in both R&D and manufacturing test. In R&D, engineers can test signals they are developing through many waveform repetitions. The more waveforms they can test, the more confidence they have that the design functions correctly. In manufacturing, engineers must test signals to ensure customers do not receive defective units. Mask testing provides more reliability than testing individual attributes because the entire signal can be evaluated against a correct signal to find glitches and/or errors. As a result, mask testing saves time and money in design and manufacturing and ensures customers receive higher-quality products sooner. Now let s return to the analogy of measuring turtle shell diameters. Sampling turtle shell diameters is similar to sampling specific signal attributes. Let s simplify this process and instead of measuring specific signal attributes, use a mask test to qualify the signal. Rather than developing a histogram of voltage values, we would only have two outcomesfor the mask test pass or fail. Figure 5 shows a waveform that failed mask testing. Figure 5. Mask tests make it easy to observe when a signal fails the QA process We are no longer using a Gaussian distribution to evaluate results; instead we are implying that mask test failures occur beyond the Six Sigma limit. Recall that to achieve Six Sigma resolution you would need to test 294,117 waveforms. If we reach 294,117 passes without a single failure, then no samples occurred beyond the Six Sigma limits. Our signal is therefore quality assured to Six Sigma! 5

Accelerating Quality Assurance In conventional mask testing solutions, oscilloscopes typically test about 10 to 60 waveforms per second using software-based algorithms. At 60 waveforms per second (wps), the update rate on the screen is fast enough to appear responsive to our eyes but the time required to accumulate enough data to be statistically meaningful will take several minutes, or possibly hours. To acquire enough points for Six Sigma resolution or samples of 294,117 waveforms at 60 wps it would take almost 1.5 hours. With many different electronic signals on ICs and FPGAs, acquiring Six Sigma confidence in all the signals can take a lot of time. For example, if 10 signals needed to be probed and tested for Six Sigma quality, current solutions would take greater than 10.5 hours. If failures were found in the signals, these tests would need to be repeated, significantly multiplying the total time to acquire Six Sigma quality assurance. Although gaining Six Sigma assurance is desirable, most engineers cannot afford the luxury of waiting this long for the outcome. Agilent has studied the current limitations of acquiring Six Sigma quality, and has created a hardware-accelerated mask testing solution that decrease this 1.5 hour test time to only 1.1 seconds. Agilent InfiniiVision Series scopes are also the first scopes in the industry that can report mask test pass/fail statistics in term of sigma quality. The mask testing statistics provide the user with a variety of information including number of tests, number of failures, failure rate, and process sigma and test sigma quality factors. The process sigma is the maximum sigma resolution possible based upon the number of tests run, and the test sigma shows the actual sigma efficiency your test has achieved. Acquiring enough samples for Six Sigma resolution using an InfiniiVision 3000 X-Series scope takes only 1.1 seconds (depending on the timebase setting, this 1.1 second is best case at 10 ns/div). Figure 6 shows a test of 1 million waveforms (in a little under 4 seconds) that provides resolution to 6.2 sigma. Agilent InfiniiVision oscilloscopes provide another benefit; lower inherent vertical noise than competitive scopes. This is important because lower noise impacts the quality and precision of mask testing. Vertical noise can cause random increases (spikes) in all areas of the signal, including trigger timing issues. A consistent trigger point is vital to prevent jitter and drift of the waveforms. If a signal drifts outside the mask limits due to noise and/or trigger jitter, mask testing will analyze the ocilloscope-induced error components as a failure, thereby corrupting the test results. A low noise floor is essential to attain the precision and reliability of Six Sigma testing. Agilent InfiniiVision oscilloscopes are designed for the best signal visibility in the market. With an update rate of up to 1,000,000 waveforms per second, InfiniiVision 3000 X-Series scopes have less dead time and can therefore catch signal anomalies that other scopes miss. Mask testing with InfiniiVision 3000 X-Series scopes can perform hardware-based mask testing at rates up to 280,000 waveform tests per second. Figure 6: Mask testing to 6.2 σ resolution takes only seconds using an InfiniiVision scope with a mask test rate of 270,000 waveforms tests per second 6

Conclusion Not only does Agilent provide the fastest, low-noise mask testing performance in the market, but the mask test application on InfiniiVision scopes also has a variety of features for customizing the test procedure. You can setup the mask test to run for a specified time, until a specific number of tests have occurred, or until a certain process sigma threshold is reached. InfiniiVision oscilloscopes also offer you the option to continue running upon a failure, stop on a failure, or save the data on a failure. For example, if the purpose of the mask test was to ensure Six Sigma efficiency, you can set the scope to run until a process sigma quality factor of six is achieved. This will occur after approximately 294,117 waveforms have been tested. If a failure occurs within these tests the waveform would automatically be stopped or saved for viewing and analysis later. The capability to assure the quality of products to Six Sigma in 1.1 seconds saves time and effort in R&D and manufacturing. In R&D, engineers can test signals they are developing through many waveform repetitions. The more waveforms they can test, the more confidence they have that the design functions correctly. In manufacturing engineers must test signals to ensure customers do not receive defective units. Mask testing lets engineers spend less time in the R&D lab ensuring signal stability, and in manufacturing probing for signal reliability. Your company saves manpower hours and can invest the capital in profitable initiatives, such as new product design. At the same time, customer satisfaction will increase as you produce high-quality products and move them quickly into the market. Related Literature Publication title Publication type Publication number Mask Test for InfiniiVision Series Oscilloscopes Data sheet 5990-3269EN InfiniiVision 2000 X-Series Oscilloscopes Data sheet 5990-6618EN InfiniiVision 3000 X-Series Oscilloscopes Data sheet 5990-6619EN InfiniiVision 7000B Series Oscilloscopes Data sheet 5990-4769EN Evaluating Oscilloscopes for Best Waveform Update Rates Application note 5989-7885EN Agilent Technologies Oscilloscopes Multiple form factors from 20 MHz to >90 GHz Industry leading specs Powerful applications 7

www.agilent.com www.agilent.com/find/masktest Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. www.lxistandard.org LXI is the LAN-based successor to GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent s measurement expertise and product breadth, combined with channel partner convenience. Motorola is a U.S. registered trademark of Motorola Inc. Agilent Advantage Services is committed to your success throughout your equipment s lifetime. We share measurement and service expertise to help you create the products that change our world. To keep you competitive, we continually invest in tools and processes that speed up calibration and repair, reduce your cost of ownership, and move us ahead of your development curve. www.agilent.com/find/advantageservices www.agilent.com/quality For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas Canada (877) 894 4414 Brazil (11) 4197 3500 Mexico 01800 5064 800 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 375 8100 Europe & Middle East Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 /minute Germany 49 (0) 7031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 United Kingdom 44 (0) 118 9276201 For other unlisted Countries: www.agilent.com/find/contactus Revised: October 14, 2010 Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2011 Printed in USA, February 15, 2011 5990-3200EN