HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

Similar documents
Liquid Chromatography- Mass Spectrometer Manual

ChromQuest 5.0 Chromatography Data System

Technical Note. High-speed Agilent 1100 Series diode-array detector SL for optimization of resolution, sensitivity, spectral sensitivity and linearity

Operating pressure should be 6000 psi or better suitable for both analytical and semi-preparative applications

ICS-3000 System Installation Qualification

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

Corporate Headquarters Dionex Corporation 1228 Titan Way P.O. Box 3603 Sunnyvale, CA Tel: (408) Fax: (408)

Achieving More Efficient Data Review with OpenLAB CDS

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK

SmartMix Manual V5350A

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer

Instrument Control and Data Acquisition with Chromeleon

AIM INTRODUCTION SIMPLIFIED WORKFLOW

PAL LSI - Robust and Reliable Sample Handling

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Operation Manual for the Refractive Index detector VE-3580

PAL LSI - Robust and Reliable Sample Handling

GC/LC-MS: data acquisition rate and peak reconstruction

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003

Staying Connected in GC

LC-4000 Series. Integrated HPLC and UHPLC systems

Creating and fully utilizing the retention time standard processing method for a more automated IC workflow

GCxGC Comprehensive Two-Dimensional Gas Chromatography

GCxGC Comprehensive Two-Dimensional Gas Chromatography

NMR. picospin. Maintenance Guide

This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus:

Real-time ph and Conductivity Data Acquisition with the ACQUITY UPLC H-Class Bio System, Auto Blend Plus, and the GE Healthcare Monitor ph/c-900

Supporting Information for Effectiveness of Global, Low-Degree Polynomial Transformations for GCxGC Data Alignment

Performance. Reliability. Productivity. Automated Flash Chromatography Systems

Control Unit CU 2.1 User Manual V6850A HPLC

STANDARD OPERATING PROCEDURE:

BR2517B. 168 Diode Array & 166 Programmable Detector Modules. Packaging Processing. Bid on Equipment

Integration of the Zetasizer µv and OmniFACE into an SEC system

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

If you have questions or problems in providing anything described as Customer

Customer Responsibilities. Important Customer Information. Agilent 1260 Infinity Clinical ed. LC Site Preparation Checklist

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0

Fluorescence Spectrophotometer

Sample Analysis Design. Element2 - Basic Software Concepts (cont d)

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

TESLA FEL-Report

Karl Heinz Feller. Arbeitsgruppe Instrumentelle Analytik FB Medizintechnik und Biotechnologie Ernst-Abbe-Fachhochschule Jena.

Advanced Sensor Technologies

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

Printed: File: RI2012 V1.02 english doc Last modified: Makowski

Practicality and Performance

Measurement of Microdisplays at NPL

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16]

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

Application Note: Using the Turner Designs Model 10-AU Fluorometer to Perform Flow Measurements in Sanitary Sewers by Dye Dilution

721N Spectrophotometer. Operating Manual

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK

Design of NIR Chromenylium-Cyanine Fluorophore Library for Switch-ON and Ratiometric Detection of Bio-Active Species in Vivo

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Comparison of SONY ILX511B CCD and Hamamatsu S10420 BT-CCD for VIS Spectroscopy

Standard Test Method for Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography

Agilent 7000 Series Triple Quadrupole GC/MS System

1. Add Capture Antibody diluted in coating buffer. 4. Add standards, controls and samples to the plate in duplicate.

Quick Start Operating Instructions

MSQ Plus Mass Detector. Getting Started Guide

Detailed Design Report

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

PACS. Dark Current of Ge:Ga detectors from FM-ILT. J. Schreiber 1, U. Klaas 1, H. Dannerbauer 1, M. Nielbock 1, J. Bouwman 1.

UNITY 2 TM. Version 1.0. February 2008

LaserPXIe Series. Tunable Laser Source PRELIMINARY SPEC SHEET

New Filling Pattern for SLS-FEMTO

SPEAR 3: Operations Update and Impact of Top-Off Injection

1 PROTOCOL FOR FLUORESCENCE IMAGING SYSTEM. General Switch (1), (2) DG4 Lamp switch (3) DG4 Main Switch (4) LAMBDA 10-2 (5) Camera (6) Computer (7)

MTS/T-BERD 8000 Platform

Finnigan PolarisQ User s Guide. PN , Revision D

SPATIAL LIGHT MODULATORS

50 mw green DMX laser

MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Modules

Online Process Control using Modular Fluid Delivery and Fast Process Gas Chromatography: From the Sample Point to the DCS Connection

eblot L1 Protein Transfer System Fast Wet Transfer System for Mini Gels

specification hyperion colorimeter

LHC Beam Instrumentation Further Discussion

SC26 Magnetic Field Cancelling System

UNIT V 8051 Microcontroller based Systems Design

LEGEND MAX Human α-synuclein ELISA Kit with Pre-coated Plate Catalog No.: (Previously Covance Cat. No. SIG-38974)

specification hyperion colorimeter

Digital BPMs and Orbit Feedback Systems

013-RD

HV/PHA Adjustment (PB) Part

DATA SHEET. Two (2) fibers Detachable HDMI 2.0 Extender,

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Assay Creosote Extraction of Selected Posts from the 1958 Cooperative Test After 50 Years of Exposure as a Ground Contact Preservative

AP-40. AP-40 Series Features Industry s smallest-sensor head Ultra lightweight High-speed response Two-color LED digital pressure display

40G SWDM4 MSA Technical Specifications Optical Specifications

WELDING CONTROL UNIT: TE 450 USER MANUAL

Module 11 : Link Design

Bruce Quimby. Agilent Technologies October 28, For Forensic Use. Page 1

Smart Lighting Demos

111 Highland Drive Putnam, CT USA PHONE (860) FAX (860) SM32Pro SDK

SC24 Magnetic Field Cancelling System

Transcription:

HPLC Tips and Tricks Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI) The world leader in serving science

Overview Thermo Scientific UltiMate 3000 UHPLC system Thermo Scientific Vanquish UHPLC system 2

Common Recommendations Mobile Phase 3

Common Problems: Mobile Phase Solvent compatibility Try to use pre-mixed solvents Add 5-10% of organic eluent to the aqueous eluent Add 5-10% aqueous eluent to the organic eluent Avoids local crystallization in the pump (with buffers) Eluents with salt buffers Change eluents with salt buffers regularly Filtrate buffers Use water with 18,2M Ohm AND <5ppb TOC 4

Solvent Quality TN140: Solvent Quality The quality of the eluent is very important to keep the noise as low as possible Make sure that the eluents are good by running them without injection (sample type "Blank ) For MS and Thermo Scientific Corona charged aerosol detector only use volatile buffers UV-spectra at 200 250 nm of two methanol samples (both LC/MS grade) 5

Common Recommendations: Mobile Phase Symptoms Causes Peaks shift Loss of resolution Mobile Phase ph changes Prevention Use correct buffer for ph range Control ph of mobile phase Maintain buffer strength in aqueous phase Control temperature ph 5.3 ph 5.4 6

Common Problems: Mobile Phase 60 mm The two chromatograms illustrate the effect of a small reduction (3 mm) in buffer concentration on the separation of six food additives: Symptoms 57 mm Causes Peaks shift Loss of resolution Mobile Phase buffer strength changes Prevention Maintain buffer strength in aqueous phase. Control temperature. Filter solvents rather than using vacuum degassing. 7

Common Recommendations The Pump 8

Pump Types LPG pump type Eluent composition at the low pressure side => Before the pump head Eluent is composed through proportioning valves Eluent segments pass working and equilibration cylinder Up to four solvents can be mixed HPG pump type Eluent composition at the high pressure side => Behind the pump head Pure solvents pass working and equilibration cylinder Eluent mixture is prepared with two pump blocks => Binary eluent mixture only 9

10 Peaks in 10 Seconds with a Ballistic Gradient 40.0 1 2 3 Overlay of 6 consecutive runs 4 5 6 7 95.0 40.0 8 9 0 6 12 Seconds 10 Test conditions Column: Eluents: Flow: C18, 30 x 2.1 mm, 1.8 µm A: Water B: Acetonitrile Temperature: 100 C Inj. volume: 1 µl Test mixture: 3.70 ml/min @ 725 bar (10,500 psi) Uracil and 9 alkylphenones Resolution (Critical Peak Pair): 1.7 Peak number 1 2 3 4 5 6 7 8 9 10 Retention time RSD [%] 0.76 0.41 0.29 0.14 0.15 0.14 0.11 0.09 0.08 0.05 Retention time SD [ms] 8.54 8.81 8.61 9.14 17.00 18.34 16.26 15.25 11.09 10.97 HPG-3200RS 10

The Comprehensive SpinFlow Mixer Portfolio Range of static mixers suitable from low-gdv LC-MS application to high-sensitive TFA application 11

Effects of Different Mixers These samples were run with the same Vanquish Horizon system, the only change was the size of the mixer Flow 0,250µl/min A: (0.1% TFA, 50 mm NaCl):MeCN 95:5. B: (0.1% TFA, 50 mm NaCl):MeCN 30:70 Black 10µl mixer Blue 350µl mixer 12

Easy Tunable for Optimum Application Performance For fast separations where the mixing ripple does not interfere with the detection (e.g., Corona CAD or MS detectors), use the low mixer volumes (35 µl, 100 µl). Use the medium sized mixers (200 µl, 400 µl) as the best balance between fast separation and low mixing ripple in UV detection. For highest sensitivity and when mixing ripples interfere with the detection (e.g., due to use of UV-absorbing solvents), use a larger mixer volume (400 µl, 800 µl). For UV-absorbing solvent additives that amplify the mixing ripples by interaction with the stationary phase (e.g., TFA application), use for highest sensitivity the largest mixer volumes (800 µl, 1550 µl). 13

Prime and Degas Air bubble stuck in the pump Prime the pump Degas the eluents 5 min in an ultrasonic water bath 160 bar p_ 140 6 7 120 5 100 3 4 80 2 60 1 40 20 min 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.45 14

Practical Use of a HPLC Pump Use eluent filters if necessary Use a degasser It might be a good idea to degas in an ultrasonic water bath for 5 min Seal wash helps the pump seal to survive If the pump is out of wash solution it should not start If the pump starts to leak Thermo Scientific Chromeleon chromatography data system (CDS) will give a warning 15

Common Recommendations Autosampler 16

Autosamplers Before you run samples flush the syringe Even a tiny air bubble ruins the performance. Autosamplers need a transport liquid. Usually eluent A is connected to the sampler and the pump. It is possible to use a separate bottle of transport liquid. The transport liquid is used to wash the needle. 17

Important Parameters DrawSpeed Defines the speed at which the sample is drawn by the syringe In analytical range (5-100 µl) a draw cycle should normally take 3-4 seconds. At lower volumes, a draw cycle should take approximately 10 times longer. Examples (normal HPLC eluents and samples dissolved in starting eluent) => 10 µl injection volume -> recommended draw speed: 2-3 µl/sec => 2 µl injection volume -> recommended draw speed: 0.2-0.3 µl/sec Draw speed has to be adapted in case of samples and eluents with higher viscosity. Incorrect setting is frequently responsible for area precision problems. 18

Sample Solvent A. Matching injection volume with solvent strength Ideally the sample solvent will not affect chromatographic separation. If a stronger sample solvent is required, injection volumes should be kept to a minimum. In a strong solvent, the sample moves more quickly through the mobile phase and often split or distorted peak shapes are observed as in chromatogram A. The syringe is usually connected to eluent A. 19

Vials and Septa Do not overfill the vial The septum is used to clean the needle. Fill 2/3 of the volume. If you use inserts make sure there are no airbubbles in them. Autosamplers with PEEK (Victrex Plc.) needle must have caps with slit septa. Never shake the vial. If you do there is going to be sample on the underside of the septum which gives carry over. Adjust the needle height according to the vial dimensions. Rubber septas get dry and can block the injector.... and use the vial and septum only once! 20

Common Recommendations Column Compartment 21

Pre-column Heater Serpentine-shaped capillary embedded in aluminum block Different types available 2µL, 7µL, 11µL, 0.12mm ID 0.18mm ID 0.25mm ID 22

Pre-column Heater Column oven 40 C 60 C 60 C Without pre-heating: Poor peak resolution Peak broadening Peak splitting Extended analysis time Column oven 60 C 60 C 23

Pre-column Heater 1 Methyl-Paraben 1 2 - Hydrocortisone With pre-column heater 2 Without pre-column heater 3 Propyl-Paraben 3 0 1 2 3 4 5 6 7 8 9 10 Minutes 24

Common Recommendations UV-Detectors 25

Operating Principle Variable Wavelength Detector (VWD) Forward optics design Only the selected wavelength passes the flow cell and is detected by the sample diode. A part of the light beam is redirected to the reference diode. Reference signal is not influenced by the content of the flow cell. True reference Light source Dispersion device Flow cell Sample diode 26

Recommended Parameters: Reference Wavelength DAD Reversed optics design Light beam passes the flow cell before being diffracted No true reference signal can be obtained Instead -and with limitations-, any diode or bunch of diodes can be selected as a reference If selected reference and acquisition wavelength are the same, the resulting signal would be zero (0) As a consequence either don t use a reference (preferred) or select a reference wavelength in a quiet area of the spectrum (where little absorbance occurs) Reference wavelength range should not interfere with absorbance range of any compound of interest Light source Flow cell Dispersion device Diode array 27

Flow Cell Volume Flow cell volume depends on peak volume. Rule: Flow cell volume should not exceed 10% of the peak volume. Micro column peak volumes Micro column peak volumes Flow Cell volume Peak Volume Flow cell ok Flow cell too big Flow cell ok Note: Besides lamp age the light intensity is highly dependent on the installed flow cell. Smaller cell volume Less light is passing through the flow cell. 28

Diode Array Detector Bandwidth and Slit width Slit width Slit defines optical resolution and therefore minimal physically meaningful bandwidth. VWD detector: Bandwidth is defined by entrance slit. Slit width Baseline noise Spectral resolution Bandwidth Bandwidth S/N ratio Spectral resolution 29

Setting Bandwidth Bandwidth 4 nm s/n = 5 Narrow bandwidth gives fewer stray light errors but lower sensitivity. Bandwidth 30 nm s/n = 25 4 x light = < 0.5 noise 30

Recommended Parameters: Data Acquisition 250 250 Data rate: 5 Hz Data rate: 100 Hz Absorbance [mau] 200 150 100 Absorbance [mau] 200 150 100 50 50 0 0,100 0,105 0,110 0,115 0,120 t [min] 0 0,100 0,105 0,110 0,115 0,120 t [min] Too few data points effect peak form, reproducibility and area precision. A minimum of 20, ideally 30-40 data points/peak are required. 31

Time Constant The parameter is a measure of how quickly the detector responds to a change in signal. Defined as the time it takes the detectors output signal to rise from 10% of its final value to 90% The rise time (Response time) is closely releated to the time constant: Rise time = 2,2 x Time constant Noise is much more influenced by time constant (TC) than by data collection rate (DCR). 32

Recommended Parameters: Data Acquisition The Program Wizard of Chromeleon has a dedicated step for setting the correct Data Collection Rate and Time Constant. The internal calculation is based on the peak width at half peak height of the slimmest peak in the chromatogram. 33

Sampling and Rise Time The same instrument, back pressure loop, eluent and sample. The area is the same the peakshape is very different. 2,5 Hz 2s response time 10 Hz 0,5s response time 34

THANK YOU! Technical Support for Chromatography Columns and Consumables www.thermoscientific.com/chromexpert Thermo Scientific Library of Analytical Applications www.thermoscientific.com/appslab 35

Thank You Very Much for Your Attention! Questions? Do you have additional questions or do you want to talk to an expert from Thermo Fisher Scientific? Please send an E-Mail to analyze.eu@thermofisher.com and we will get back to you. 36