Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper

Similar documents
Iterative Direct DPD White Paper

R&S FSV-K40 Phase Noise Measurement Application Specifications

R&S FSW-K160RE 160 MHz Real-Time Measurement Application Specifications

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications

Pre-5G-NR Signal Generation and Analysis Application Note

Mastering Phase Noise Measurements (Part 3)

R&S FSW-K144 5G NR Measurement Application Specifications

R&S FSW-K76/-K77 3GPP TD-SCDMA BS/UE Measurement Applications Specifications

R&S FSV-K73 3G FDD UE (UL) Measurements incl. HSUPA Specifications

R&S Spectrum Rider FPH Handheld spectrum analyzer

R&S FSV-K76 TD-SCDMA BS (DL) Measurements Specifications

Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes

This application note is a simple step-by-step guide that introduces a practical method to perform reliable small cell planning.

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

R&S FPS-K18 Amplifier Measurements Specifications

Multi-port calibration by using a two port calibration unit. Application Note. Products: R&S ZVT R&S ZNB

R&S FS-Z60/75/90/110 Harmonic Mixers for the R&S FSP/FSU/ FSQ/FSUP/FSV

RF amplifier testing from wafer to design-in

R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz

R&S CA210 Signal Analysis Software Offline analysis of recorded signals and wideband signal scenarios

Product Brochure Version HZ-15_16_17_bro_en_ _v0100.indd 1

R&S FSW-K54 EMI Measurement Application Detecting and eliminating electromagnetic

R&S HA-Z24E External Preamplifier 1 GHz to 85 GHz Specifications

R&S ZN-Z103 Calibration Unit Specifications. Data Sheet V02.01

R&S SFD DOCSIS Signal Generator Signal generator for DOCSIS 3.1 downstream and upstream

R&S GX465 Digital Wideband Storage Device Recording and replaying of I/Q data with up to 80 MHz bandwidth

Configuring the R&S BTC for ATSC 3.0 Application Note

R&S FSV-K8 Bluetooth /EDR Measurement Application Specifications

R&S ELEKTRA EMI Test Software Easy to use software for measuring electromagnetic disturbances

Product Brochure Version R&S TSML-CW Radio Network Analyzer Powerful scanner for CW applications

Product Brochure Version R&S RSC Step Attenuator Where precise signal levels count

The Future of EMC Test Laboratory Capabilities. White Paper

R&S HF907DC SHF Downconverter Specifications

DOCSIS 3.1 Full channel loading Maximizing data throughput

Scope of the art Scope Rider Handheld digital oscilloscope

R&S ZN-Z154 Calibration Unit Specifications

Your partner in testing the Internet of Things

R&S CONTEST ITS Test cases and applications

R&S EFL110/EFL210 Cable TV Analyzer and Leakage Detector Detecting interference in cable TV and LTE networks

R&S FPC1000 Spectrum Analyzer Specifications

R&S ZN-Z85 Switch Matrix Specifications

Test and Communications Antennas for the R&S TS8991 OTA Performance Test System Specifications

R&S SMBV-Z1 Reference Frequency Converter Specifications

R&S GX460 Digital Wideband Storage Device Recording and replaying device for I/Q data with up to 40 MHz bandwidth

Product Brochure Version R&S ENV A Four-Line V-Network RFI voltage measurements at high currents

R&S ZN-Z151/-Z152/-Z153 Calibration Unit Specifications

EUTRA/LTE Downlink Specifications

R&S ZNrun Automated Test Software PC-based server platform for automated VNA tests

EUTRA/LTE Measurement Application Specifications

Troubleshooting EMI in Embedded Designs White Paper

R&S ZVA-Zxx Millimeter-Wave Converters Specifications

Test and Communications Antennas for the R&S TS8991 OTA Performance Test System Specifications

Test Port Adapter Rohde & Schwarz Interchangeable Port Connector Application Note

R&S TS-BCAST DVB-H IP Packet Inserter Compact DVB H signal generator with integrated IP packet inserter

Five Reasons to Upgrade from Legacy VNAs to a R&S ZNB Vector Network Analyzer

R&S ZVA-Zxx Millimeter-Wave Converters Specifications

R&S ZN-ZTW Torque Wrench Specifications

R&S ZN-Z32/-Z33 Automatic In-line Calibration Modules Ensuring high accuracy with thermal vacuum testing and multiport measurements

R&S FSQ-K91/K91n/K91ac WLAN a/b/g/j/n/ac Application Firmware Specifications

Oscilloscopes for debugging automotive Ethernet networks

R&S RSC Step Attenuator Where precise signal levels count

R&S MDS-21 Absorbing Clamp Measurement of disturbance power and screening effectiveness on cables

R&S ZV-Z81 Multiport Test Set, models.05/.09/.29 Specifications

R&S WMS32 Wireless Measurement System Software Specifications

Concise NFC Demo Guide using R&S Test Equipment Application Note

Correlated Receiver Diversity Simulations with R&S SFU

Be ahead in 5G. Turn visions into reality.

R&S AVG050 DVB Satellite Receiver Specifications

How To Demonstrate Improved ACLR Dynamic Range With FSU and Noise Correction

Coherence Measurement between two Signals regarding Timing, Phase and Gain Application Note

R&S RT-ZM Modular Probe System

R&S FSW Signal and Spectrum Analyzer Resolving Security Issues When Working in Secure Areas

R&S ADMC8 Multicoupler Active UHF multicoupler for 8-port ATC signal distribution

R&S NESTOR-FOR Crime Scene Investigation

R&S RT-Zxx High-Bandwidth Probes Specifications

R&S RT-Zxx High-Voltage and Current Probes Specifications

Using R&S NRP Series Power Sensors with Android TM Handheld Devices. Application Note. Products: R&S NRP Series. R&S NRP-Zxx Series

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz

Product Brochure Version R&S OSP Open Switch and Control Platform Modular solution for RF switch and control tasks

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Implementation of Real- Time Spectrum Analysis

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz

Product Brochure Version R&S OSP Open Switch and Control Platform Modular solution for RF switch and control tasks

LabVIEW driver history for the R&S RTH Handheld Digital Oscilloscope Driver Documentation

EUTRA/LTE and LTE-Advanced Signal Analysis Transmitter measurements on LTE signals

R&S VSE Vector Signal Explorer Base Software Specifications

Application Note DT-AN DTU-315 Verification of Specifications

R&S TSMx Radio Network Analyzers Powerful scanner family for mobile applications

R&S PSL3 Industrial Controller The powerful industrial controller

WLAN IEEE802.11a/b/g/j/p/n/ac/ax Measurement Application Specifications

R&S VENICE On air. 24/7.

Test Device Containment Module (TCM) Optimal conditions for reliable results

Step-by-step guide Designing for EMI testing (step-by-step guide) Improve your time to market with oscilloscopes

Meeting your needs R&S RTO2000 Digital Oscilloscope

Product Brochure Version R&S AdVISE Visual Inspection Software A new way to eliminate human inattention

R&S GU221 Filter Control Unit Specifications

Test and measurement solutions for electronics manufacturers

R&S NESTOR-FOR Alibi Verification

R&S InstrumentView Release Notes Software Version 1.70

<EDGE Looped based BER and TX measurements using ARB files on the R&S CMW 500> Products: R&S CMW500 R&S CMW- KW200 R&S CMW- KM200. <Application Note>

Transcription:

Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper Products: ı ı R&S FSW R&S FSW-K50 Spurious emission search with spectrum analyzers is one of the most demanding measurements in the design, verification and production of RF and microwave devices. RF designers, especially in the aerospace & defense industry need to detect very low level spurs. Very narrow resolution bandwidths are required to achieve the low noise required for these measurements, which increases measurement time. Even working with very fast spectrum analyzers, a spur search may take several hours or even days. In this paper we will review the basics of spurious measurements and how the parameters used can affect the detection performance. A new technique as used in the R&S FSW-K50 spurious measurement application which makes spurious search faster and easier to configure. Note: Please find the most up-to-date document on our homepage https://www.rohde-schwarz.com/appnote/1ef97 This content was previously presented at EDICON China 2017, Shanghai White Paper Laura Sanchez 5.2017 1EF97_0e

Table of Contents Table of Contents 1 Introduction... 3 2 Spurious emission measurement with spectrum analyzers performance parameters... 4 2.1 Resolution bandwidth vs thermal noise floor considerations for spurious measurements... 4 2.2 Measurement detector choice for spurious measurements... 6 3 Speed up spurious measurement using the optimum RBW... 8 3.1 Residual spurs identification...11 3.2 Directed search at predefined frequencies...11 4 Summary... 12 5 References... 13 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 2

Introduction 1 Introduction Spurious emissions are unwanted emissions inside and outside the necessary signal bandwidth; they can be caused by harmonic emissions, parasitic emissions, intermodulation and frequency conversion products. Spurious emissions decrease the system performance and cause interference with other systems working in adjacent frequency bands or with other modules within the equipment. Figure 1 Changes in I/Q Constellations diagram of a 64 QAM modulated signal with increasing interference Spurious emission search with spectrum analyzers is an essential measurement in the design, verification and production of RF and microwave devices. Manufactures of all kinds of RF and microwave transmitters such as radars, satellite up- and downconverters need to ensure that their equipment spurious emissions are within the specified limits. Some transmitter and receiver devices, especially in aerospace & defense applications need to fulfill very stringent spurious emissions limits. This implies searching for very low level spurs within wide frequency ranges. Thus a very low resolution bandwidth (RBW) is required to make the measurements with low noise floor and high sensitivity. Moreover the location of the spurs in the spectrum is often not known beforehand, i.e. a very large frequency range has to be measured with narrow resolution bandwidths, sometimes of only a few Hz, which increases the measurement time. Even working with fast spectrum analyzers with FFT filters a spur search may take several hours or even days. The new approach with the R&S FSW-K50 spurious measurements application reduces the measurement time and automates the measurements when looking at low level spurs of -120 dbm or below. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 3

Spurious emission measurement with spectrum analyzers performance parameters 2 Spurious emission measurement with spectrum analyzers performance parameters Spurious measurements require measuring large frequency ranges with a very high sensitivity or very low noise floor. In addition to the spectrum analyzer's thermal noise floor, the resolution bandwidth filter and measurement detector are the main factors that determine performance and speed. 2.1 Resolution bandwidth vs thermal noise floor considerations for spurious measurements The resolution bandwidth filter establishes the resolution bandwidth (RBW) of the spectrum analyzer. The lower this resolution bandwidth, the higher the spectral resolution. RBW influences the displayed noise floor and crucially, the sweep speed of the spectrum analyzer. In general use, the RBW is automatically set by the spectrum analyzer somewhere between 1% and 4% of the selected frequency span; this is a good value which is narrow enough to suppress spectral components outside the channel to be measured and wide enough to achieve acceptable measurement speed. The RBW can also be adjusted by the user. If a given RBW is reduced by a factor of 10, the noise floor of the measurement will be lowered by 10 db and vice versa. This is shown graphically in the figure below. For spurious measurement applications the RBW usually needs to be reduced to help lower the noise floor in order to meet the test requirements: Thermal noise floor should typically be about 10 db below the limits that are stated for a spurious emissions test to result in an acceptable signal to noise ratio (SNR) and correspondingly low uncertainty contribution of the noise. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 4

Spurious emission measurement with spectrum analyzers performance parameters Figure 2 Changes in the displayed noise floor for RBW changing between 3 MHz (yellow trace) and 10 Hz (red trace) for 1 GHz span When the resolution bandwidth is not defined in the test requirement, it is possible to calculate the required resolution bandwidth for a given maximum allowed noise level according to the formula below. RBW (Hz) = 10 Maximum Noise Level(dBm) Noise Floor( dbm Hz ) 10 where: ı ı the maximum noise level corresponds to the maximum allowed spur level, with a certain margin to achieve sufficient signal to noise ratio. the thermal noise floor of a spectrum analyzer is determined by design and specified as the Displayed Average Noise Level (DANL) in a given resolution bandwidth (i.e. RBW 1 Hz). A typical value for a high performance spectrum analyzer is about -155 dbm in 1 Hz bandwidth. According to the equation above, an RBW of around 3 khz would be required for a high performance spectrum analyzer with a noise floor of -155dBm/Hz, to achieve a -120 dbm spur measurement, assuming 10 db SNR. The higher the spectrum analyzer's thermal noise floor, the lower the RBW needed to meet the measurement requirements and the longer the sweep time becomes. Therefore it is recommended to use a high performance spectrum analyzer with DANL as low as possible for spurious measurements, especially if the spur limit is -120 dbm or below. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 5

Spurious emission measurement with spectrum analyzers performance parameters The RBW selected not only defines the frequency resolution and the noise floor of the measurement, it is also the main limitation of sweep speed, i.e. the lower the RBW the longer the measurement time required to measure a given frequency range. The required sweep time increases inversely with the square of the resolution bandwidth according to the below formula. Where: Span (Hz) SWT(s) = k RBW(Hz) 2 SWT is the sweep time, Span is the frequency range, and k is the correction factor for the settling time of the resolution filter (k=1 for FFT RBW filter and between 2 and 3 for older, analog spectrum analyzers) It is therefore important to choose a bandwidth that just fits to the spurious level requirement in order to keep the total measurement time acceptable. The instruments noise floor varies along the frequency range; this means that different RBWs will need to be used at different frequencies. In traditional spurious search applications the RBW which is used at each frequency needs to be calculated and entered manually in the analyzer. 2.2 Measurement detector choice for spurious measurements In modern spectrum analyzers the resolution of both the level and the frequency display is limited to the display resolution (number of pixels). Therefore, especially when large spans are displayed, there are many more measurement samples than display pixels in the frequency axis. This means a certain amount of measurement samples have to be combined into one single pixel (point displayed on the screen). The measurement detector choice determines how the measurement samples are weighted to get a single measurement value per display pixel. Spectrum analyzers offer different measurement detectors. The principles of the detectors are shown in Figure below. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 6

Spurious emission measurement with spectrum analyzers performance parameters Figure 3 Selection of sample to be displayed as a function of detector used [1] Spurious emissions measurements for very low level signals are typically performed with a peak detector. The max peak detector displays the maximum value. From the samples allocated to a pixel the one with the highest level is selected and displayed. The advantage of the peak detector compared to other detector types is that even if wide spans are displayed with very narrow resolution bandwidth (span/rbw >> number of pixels on frequency axis), no input signals are lost and the result of the measurement is a worst-case scenario. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 7

Speed up spurious measurement using the optimum RBW 3 Speed up spurious measurement using the optimum RBW The spurious search algorithm within FSW-K50 automates and speeds up spurious measurements when looking at very low level spurs. This new technique requires the user to set only the frequency ranges to be investigated and the maximum allowed spurious level. A 10 db SNR margin between the instrument noise floor and the specified spurious detection threshold is used per default. However, the SNR margin can be also changed by the user along with the other relevant parameters shown in Figure 4. Users don t need to set the RBW for each frequency range like in traditional applications because the algorithm calculates it automatically. Figure 4 Exemplary measurements settings for a wide spurious search Figure 5 Measurement process in the FSW-K50 spurious search algorithm: 3 sweeps /measurement steps. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 8

Speed up spurious measurement using the optimum RBW The FSW-K50 application first performs a fast reference sweep in the defined frequency ranges to obtain an overview of the input signal, the so called spectral overview. The spectral overview allows the application to gather the necessary information about the signal to estimate the noise floor for the user defined frequency span and other settings. Figure 6 Measurement screen of spectral overview and noise floor estimate for a 2 GHz span measurement. The noise floor estimation is required to set the RBW optimally for subsequent measurement steps: due to noise variations across the frequencies, the RBW required to achieve the user-defined spur detection threshold varies. Each frequency range is thus split into smaller segments that use a constant RBW setting for the complete segment span. The frequency span of the segments depends on the signal and noise conditions and on the settings. As a result, a segment table is created. Figure 7 Segment table calculated after the reference sweep and noise floor estimation. Shows the RBW required to measure at each frequency for the spurious limits set by the user. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 9

Speed up spurious measurement using the optimum RBW After the reference sweep the FSW-K50 application performs a second sweep, a so called spur search using the RBW settings calculated in the previous step. The results of this sweep produce a preliminary table containing each detected spur. A third and final sweep, the spot search, is performed over very narrow spans around each spur detected in the previous step. The spot search is required to eliminate noise, refine the spur levels and frequencies, as necessary to meet the user SNR requirements. Thus it determines whether the peak is a real spur, an artifact of noise, or generated internally by the analyzer (residual spurs). The RBW for these spot searches may be reduced further compared to the spurious detection sweep. The resulting "Spurious Detection Table" contains all peaks that still exceed the detection threshold, and are really a spur. Figure 8 Zoom into a spot search: the RBW around the spur is reduced to lower the noise floor and fulfill the SNR requirement set by the user and determine whether the peak is a real spur. The algorithm within the FSW-K50 application automatically sets the optimum RBW: The RBW is as low as necessary to recognize the real spurs at the required frequencies but not unnecessarily low in those frequency ranges where no potential spurs have been found. This technique speeds up spur searches up to 50 times in comparison with traditional spur search methods, when wide frequency ranges and very stringent spur detection limits are applied. The table below a shows comparison of the required measurement time with traditional spurious search applications and the FSW-K50 spurious measurement application. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 10

Speed up spurious measurement using the optimum RBW RBW / Noise (dbm) 1 GHz span R&S FSW spectrum analyzer, FFT Sweep Measurement speed (s) R&S FSW spectrum analyzer, FSW-K50 measurement application Measurement speed (s) 1 Hz / -140 12246 200 2 Hz / -138 3088 84 3 Hz / -135 1384 35 5 Hz / -132 507 12 10 Hz / -130 132 7 20 Hz / -128 36 6 30 Hz / -125 17 5 50 Hz / -122 7.1 4 100 Hz/-120 4.1 3 Table 1 Measurement times comparison with traditional and new spurious search applications 3.1 Residual spurs identification Residuals are spurs that are created by the analyzer itself. These spurs are identified by the FSW-K50 spurious measurements application automatically; they can be displayed or removed from the measured results. 3.2 Directed search at predefined frequencies In some cases potential spur frequencies are known beforehand, like harmonic frequencies for example. For these cases the FSW-K50 application also provides directed search measurements at predefined discrete frequencies with a small span around each frequency. The measurement steps for both wide and directed search measurements are the same; however, because the span is smaller, the directed search saves measurement time and can provide more precise results, using a larger SNR. It s possible to combine a wide search measurement and directed search measurement. A function allows transferring the frequencies of interest of a previous wide search into a directed search, simplifying the setup for a directed spur search. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 11

Summary 4 Summary The R&S FSW-K50 spurious measurements application automates and speeds up the spurious search, especially when looking at very low level spurs over a broad frequency range. The user only needs to enter the frequency range and the desired spur detection level. The application calculates and applies the optimum RBW to measure at each frequency. Very narrow RBWs will be used only in those frequencies where potential spurs have been found. Spurious search becomes up to 50 times faster than traditional spurious search measurements with high performance spectrum analyzers. Figure 9 FSW-K50 Spurious measurement application screen: the spectral overview shows the reference sweep to calculate the necessary RBW. The spurious detection spectrum shows the pass /fail indication according to the spur detection limits. The spurs detection table shows the detected spurs: frequency, power level, RBW used for the measurement, etc. 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 12

References 5 References [1] C. Rauscher, Fundamentals of Spectrum Analysis, Rohde & Schwarz, Munich, ISBN 978-3-939837-01-5 [2] K-U. Sander, "Speed considerations for Spurious Measurements with Spectrum Analyzers", retrieved June 2017 at www.rohde-schwarz.com/appnote/1ef80 [3] R&S FSW-K50 Spurious Measurements User Manual, Rohde & Schwarz, Munich 1EF97_0e Rohde & Schwarz Advanced Techniques for Spurious Measurements with R&S FSW-K50 13

Rohde & Schwarz The Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and measurement, broadcast and media, secure communications, cybersecurity, radiomonitoring and radiolocation. Founded more than 80 years ago, this independent company has an extensive sales and service network and is present in more than 70 countries. The electronics group is among the world market leaders in its established business fields. The company is headquartered in Munich, Germany. It also has regional headquarters in Singapore and Columbia, Maryland, USA, to manage its operations in these regions. Regional contact Europe, Africa, Middle East +49 89 4129 12345 customersupport@rohde-schwarz.com North America 1 888 TEST RSA (1 888 837 87 72) customer.support@rsa.rohde-schwarz.com Latin America +1 410 910 79 88 customersupport.la@rohde-schwarz.com Asia Pacific +65 65 13 04 88 customersupport.asia@rohde-schwarz.com China +86 800 810 82 28 +86 400 650 58 96 customersupport.china@rohde-schwarz.com Sustainable product design ı ı ı Environmental compatibility and eco-footprint Energy efficiency and low emissions Longevity and optimized total cost of ownership This white paper and the supplied programs may only be used subject to the conditions of use set forth in the download area of the Rohde & Schwarz website. R&S is a registered trademark of Rohde & Schwarz GmbH & Co. KG; Trade names are trademarks of the owners. PAD-T-M: 3573.7380.02/02.05/EN/ Rohde & Schwarz GmbH & Co. KG Mühldorfstraße 15 81671 Munich, Germany Phone + 49 89 4129-0 Fax + 49 89 4129 13777 www.rohde-schwarz.com 14