The Application of X-Rays in Radiology: From Difficult and Dangerous to Simple and Safe

Similar documents
Performance of a DC GaAs photocathode gun for the Jefferson lab FEL

CATHODE RAY OSCILLOSCOPE (CRO)

THE X-RAY ADVANTAGE Pros and cons X-ray and Gamma

3B SCIENTIFIC PHYSICS

Sep 09, APPLICATION NOTE 1193 Electronic Displays Comparison

J.J. Thomson, Cathode Rays and the Electron

Design, Fabrication and Testing of Gun-Collector Test Module for 6 MW Peak, 24 kw Average Power, S-Band Klystron

In the tube collection there are several sensors designed for applications in some kinds of physics measurements or detection.

These are used for producing a narrow and sharply focus beam of electrons.

Monitor QA Management i model

TECHNICAL SPECIFICATION Multi-beam S-band Klystron type BT267

GA A26497 SOLID-STATE HIGH-VOLTAGE CROWBAR UTILIZING SERIES-CONNECTED THYRISTORS

CATHODE-RAY OSCILLOSCOPE (CRO)

Performance Characteristics of Steady-State MPD Thrusters with Permanent Magnets and Multi Hollow Cathodes for Manned Mars Exploration

3B SCIENTIFIC PHYSICS

High QE Photocathodes lifetime and dark current investigation

Principles of Electrostatic Chucks 6 Rf Chuck Edge Design

E2V Technologies CX2668A, CX2668AX Air-Cooled, Hollow Anode, Two-Gap Metal/Ceramic Thyratrons

The Knowledge Bank at The Ohio State University. Ohio State Engineer

Standby...For the Truth

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

About LED Lighting. White Paper: Operating Characteristics. Low Power LEDs

High-Current Hollow Cathode Development *

EE 230 Lecture 9. Amplifiers and Feedback

High Brightness Injector Development and ERL Planning at Cornell. Charlie Sinclair Cornell University Laboratory for Elementary-Particle Physics

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

IONIVAC Transmitter ITR 90

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

CX1725W Liquid Cooled, Hollow Anode Two-Gap Metal/Ceramic Thyratron

MICROFOCUS X-RAY SOURCE PROJECT*

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

Design and Simulation of High Power RF Modulated Triode Electron Gun. A. Poursaleh

P-224: Damage-Free Cathode Coating Process for OLEDs

About vacuum power tubes.

Nuclear Instruments and Methods in Physics Research A

CHAPTER 9. Actives Devices: Diodes, Transistors,Tubes

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

TOSHIBA Industrial Magnetron E3328

Development of OLED Lighting Applications Using Phosphorescent Emission System

Application Note [AN-007] LCD Backlighting Technologies and Configurations

Lecture 17 Microwave Tubes: Part I

DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC*

Correlation of Hollow Cathode Assembly and Plasma Contactor Data from Ground Testing and In-Space Operation on the International Space Station *

The Use of an Electron Microchannel as a Self-Extracting and Focusing Plasma Cathode Electron Gun

TWO BUNCHES WITH NS-SEPARATION WITH LCLS*

SLAC R&D Program for a Polarized RF Gun

TEST WIRE FOR HIGH VOLTAGE POWER SUPPLY CROWBAR SYSTEM

Tutorial Cathode Rays Year 12 Physics - Module 9.3 Motors and Generators

Design Studies For The LCLS 120 Hz RF Gun Injector

RADIOGRAPHIC PERFORMANCE OF CYGNUS 1 AND THE FEBETRON 705

Reading 21 ELECTRON TUBES

Basically we are fooling our brains into seeing still images at a fast enough rate so that we think its a moving image.

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM

Thyratrons. High Energy Switches. Features. Description

Episode 103: Currents and Charge Carriers

Electrical & Electronic Measurements: Class Notes (15EE36) Module-5. Display Devices

Pseudospark-sourced Micro-sized Electron Beams for High Frequency klystron Applications

KLYSTRON GUN ARCING AND MODULATOR PROTECTION

Slides on color vision for ee299 lecture. Prof. M. R. Gupta January 2008

Operation of CEBAF photoguns at average beam current > 1 ma

Improvements to Siemens Eclipse PET Cyclotron Penning Ion Source

ELECTROMAGNETIC FIELDS AND PUBLIC HEALTH

CONTENTS. Table of Figures

EG4015. Digital Generator Governor Controller User Manual

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16]

G. Pittá(*), S. Braccini TERA Foundation, Novara, Italy (*) Corresponding author.

Program Risks Risk Analysis Fallback Plans for the. John T. Seeman DOE PEP-II Operations Review April 26, 2006

Spectroscopy on Thick HgI 2 Detectors: A Comparison Between Planar and Pixelated Electrodes

Duke University. Plasma Display Panel. A vanished technique

An RF Excited Plasma Cathode Electron Beam Gun Design

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date:

Uniformity of Plasma Density and Film Thickness of Coatings Deposited Inside a Cylindrical Tube by Radio Frequency Sputtering

Approved by: / / R. Battaglia 12/16/2016

Physics Requirements for the CXI Ion Time-of-Flight

Particle-in-cell simulation study of PCE-gun for different hollow cathode aperture sizes

OPTIMIZED LIGHT-EMITTING DIODE (LED) DEVICES THAT HAVE A HIGH COLOR RENDERING INDEX (CRI) FOR LIGHTING APPLICATIONS

Mechanical aspects, FEA validation and geometry optimization

Valves Artzt circuit (a.k.a. the SRPP and the µ-follower)

CLINICAL SOLUTIONS THE TECHNOLOGICAL ADVANTAGE FOR DENTAL CLINICS AND UNIVERSITIES

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400

Figure 1. High Efficiency T8 LED Tube. Figure 3. Full View of Light High Efficiency T8 LED Tube

Detailed Design Report

Klaran HC Series UVC LEDs

Improvements in Gridless Ion Source Performance

CNT FIELD EMISSION CATHODE CATALOG. XinRay Systems Inc. April 2014

GENCOA Key Company Facts. GENCOA is a private limited company (Ltd) Founded 1995 by Dr Dermot Monaghan. Located in Liverpool, UK

This work was supported by FINEP (Research and Projects Financing) under contract

How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices -

HSME-C400. Data Sheet. Side-Fire Mono-Color Surface-Mount ChipLED. Features. Description. Applications

Screen investigations for low energetic electron beams at PITZ

SLD266ZS. Octa-Beam AlGaAs Laser Diode. Description. Features. Applications. Recommended Operating Optical Power Output

OSCILLOSCOPE AND DIGITAL MULTIMETER

Defense Technical Information Center Compilation Part Notice

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

Development of OLED Lighting Panel with World-class Practical Performance

Teltron Delection Tube D

Display Technologies CMSC 435. Slides based on Dr. Luebke s slides

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life.

Transcription:

Medical Physics and Informatics Opinion Kemerink et al. X-Rays in Radiology Medical Physics and Informatics Opinion FOCUS ON: Martijn Kemerink 1,2 Tom J. Dierichs 1 Julien Dierichs 1 Hubert Huynen 1 Joachim E. Wildberger 1 Jos M. A. van Engelshoven 1 Gerrit J. Kemerink 1 Kemerink M, Dierichs TJ, Dierichs J, et al. Keywords: gas x-ray tubes, history of radiology, occupational radiation exposure DOI:10.2214/AJR.11.7844 Received September 8, 2011; accepted after revision October 4, 2011. J. E. Wildberger has received payment for lectures and scientific courses from Bayer Schering Pharma, Boston Scientific, and GE Healthcare. 1 Departments of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands. 2 Present address: Department of Applied Physics, Eindhoven University of Technology, PO Box 513, 5600 MB Eindhoven, The Netherlands. Address correspondence to G. J. Kemerink (gerrit.kemerink@mumc.nl). AJR 2012; 198:754 759 0361 803X/12/1984 754 American Roentgen Ray Society The Application of X-Rays in Radiology: From Difficult and Dangerous to Simple and Safe OBJECTIVE. This article will provide an assessment of the application of x-rays in the early days of radiology, which is an excellent way to come to value the convenience and safety of modern x-ray systems. CONCLUSION. The gas tubes that were originally applied for x-ray production were very unstable because of variations in the tube s vacuum. In an effort to understand some of the problems of these tubes and the high occupational exposure that was indirectly caused by the tubes erratic behavior, we measured x-ray output rates as a function of the gas pressure inside the tube. The pressure range for the optimal production of x-rays, using an original Ruhmkorff inductor as a high-voltage generator, was found to be narrow. With the vacuum changing over time, this might explain the many photographs from the first years of radiology with operators watching their unshielded tube, either with bare eyes or with a fluoroscope, and their own hand as a test object. This practice often led to severe damage of the hands and to many early deaths due to cancer. Today, after a century of technologic development of x- ray tubes and associated equipment, the total average effective dose of workers in radiology can be close to natural background levels. R ecently we had the opportunity to repeat x-ray imaging experiments with equipment and procedures from January 1896 [1]. The original experiments by Hoffmans and van Kleef had been performed within 1 month after Röntgen s first publication on x- rays [2]. While performing these experiments, we not only recreated original images but also experienced the cumber someness of the operation of the gas tubes used for the generation of x-rays in the first days of radiology. Here we describe more detailed measurements on these gas tubes and highlight the immense progress that has been made in the way x-rays are generated and the concomitant protection of radiologic workers. In addition, by performing a systematic study of the gas pressure dependence of the tube performance we believe that we provide some hitherto unavailable quantitative data. The gas tube was the standard x-ray source in the first two decades after Roentgen s discovery of the new type of radiation. It was an only partially evacuated glass bulb with two metal electrodes at opposite ends. By applying a high voltage across the electrodes, a discharge ionized the residual gas. The posi- tive gas ions would strike on the (negative) cathode and liberate electrons, which, after being accelerated by the positive high-voltage applied to the anode, produced x-rays in any surface they hit. In the initial experiments, the x-rays were generated at the bottom of a glass bulb, but very early in 1896, the electrons were focused onto a metal target [3, 4] (Fig. 1). More information on radiology in its infancy is available in an extended body of contemporary books [5 13], all available in digital form or in facsimile. Most of this information is of a very practical and qualitative nature, however, and less scientific in character than, for instance, Roentgen s original article [14]. Gas tubes were notorious for their unpredictable behavior because of a highly variable gas pressure. To cite H. Hulst [15] from 1902: Nature seems capricious to the savage mind; tubes do to the x-ray worker. He rests, pets, punishes, smashes tubes as the fetishist his idols. He learns to know them all by their name, their temper, and their capabilities. He has a number of them the more the better. He has trained tubes, trick tubes, high-spirited, highbred tubes as well as gentle steady tubes.a flashy tube is worse than a hysterical woman, 754 AJR:198, April 2012

X-Rays in Radiology it is incurably useless. A more moderate evaluation of gas tubes, with essentially the same conclusion, is given by Coolidge [16]. The problems caused by the varying gas pressure are related to a few basic facts. First, when the vacuum was too high for a given high voltage to cause a discharge, no x-rays were generated. Second, when the vacuum was low, the discharge current could be so strong that it loaded the (unstabilized) power supply so heavily that the high voltage dropped to a substantial degree. In the latter case, the electrons were also likely to lose part of their kinetic energy in collisions with the relatively abundant residual gas. Voltage drop and electron collisions resulted in a lowering of the energy of the electrons, and the x-ray radiation hardness intended might not be obtainable. These considerations illustrate fundamental limitations of the gas tube: Any radiation quality required a rather unique vacuum, and, for a given tube, the x- ray output was generally not independently adjustable from the radiation hardness. The problematic behavior of gas tubes had unfortunate side effects. Users of these tubes would often want to check how well their tube was functioning and make electrical adjustments when possible. The functioning could partly be judged from the visual aspects of A C the discharge in the tube and the greenish fluorescence of the glass, but a more conclusive way was to observe one s own hand using a fluoroscope [3, 10, 17] (Fig. 2). In the long term, this practice gave high doses to the hands and the body as well [17]. When the tube was considered not working properly or not giving radiation of the required hardness for some body part, it would have to be reconditioned or replaced. These requirements may have been instrumental in delaying the use of shielding, generally until about 1908 [17], many years after the risks of x-ray exposure had become evident. Fig. 2 Photograph shows that while radiograph of hand is taken, operator checks unshielded tube by looking at image of his own hand using handheld fluoroscope [10]. Fig. 1 Photographs of three types of early gas tubes in order of increasing sophistication. All tubes are part of historical collection owned by city of Maastricht. A, Functioning evacuable Crookes tube number 9. Intensity of fluorescence is indication of strength of x-ray emission. B, Tube with focusing cathode and metal anticathode. Anticathode was effectively anode. C, Same type tube as in B but with device for regulation of gas pressure in tube. As already indicated, our recent repetition of the 1896 experiments with tubes originally used by Hoffmans and van Kleef was accompanied by problems [1]. Crookes tube number 1 only gave off x-rays after a fortunate incident, and Crookes tube number 9 would only produce x-rays when it was powered with reverse high voltage (Fig. 3). Note the difference in fluorescence with the normally functioning tube at the top of Figure 1. Here we will concentrate on the more subtle problem with Crookes tube number 9. Initially, we did not realize that we had driven Crookes tube number 9 with reverse B AJR:198, April 2012 755

Kemerink et al. polarity because we had simply wired it the way it gave x-rays [18]. However, we had noticed the uncommon fluorescence pattern in the tube [1] and attributed it to gas pressure in the tube being too high. The tube must have been evacuated and sealed before 1896. Unfortunately, there is no easy way to investigate the pressure in a closed tube. Because the Maastricht historical collection of instruments also included a slightly simpler version of Crookes tube number 9 with a ground-glass joint for connecting to a vacuum pump (Fig. 4), we decided to investigate x-ray production by this tube as a function of the gas pressure. Although many publications gave (widely varying) values for the gas pressure needed for x-ray generation, we found no articles with primary data or references to such data, with one exception, a communication by Röntgen [19]. Today, x-ray tubes are high-vacuum devices with a heated tungsten cathode that emits electrons because some electrons have thermal energy that is sufficiently high for escape from the cathode wire or foil. This socalled thermionic electron emission can be controlled by the temperature of the cathode, making the electron current that generates x- rays adjustable over a wide range. The hardness of the radiation is independently controlled by a combination of high voltage, the x-ray beam filter, and, to some extent, the anode material. The hot cathode tube was introduced around 1913 by Coolidge [16], providing radiology with an easily adjustable, stable, and very reliable source of x-rays that replaced the gas tube in the next decade [4]. This facilitated a major change in the operation of x-ray equipment. We report on the results of x-ray output measurements on some 19th-century gas tubes, and we illustrate the immense progress radiology Fig. 3 Photograph shows Hoffmans original Crookes tube number 9 as retrieved from warehouse. Reverse polarity was required to generate x-rays. As in Figure 1A, brightness of fluorescence is indication of intensity of emitted x-rays. has made with respect to the source of x-rays and the associated improvement in the protection of the radiologic worker. Experiments The open version of Crookes tube number 9 and the one originally used by Hoffmans are similar in size, with the lengths of the glass pear 22.5 and 24 cm, respectively. The open tube was connected to a vacuum pump system to obtain a variable pressure in the tube. The pump system (PT 50 KIT, Oerlikon Leybold Vacuum) (Fig. 4) reached a lowest pressure of about 4 10 6 mbar when connected to the tube. Higher pressures, with pressure readings varying by less than a few percent, could be obtained by largely closing the valve above the pump and letting in air using an adjustable needle valve. The accuracy of the Penning/ Pirani vacuum meter is specified as better than 30% over the pressure range 100 to 1 10 8 mbar and the reproducibility as better than 5% over the same range. A pulsed high voltage was generated using a Ruhmkorff inductor (F. Ernecke) and a 6-V car battery with the spark gap set at 11 cm, which limited the peak high voltage to about 80 kv. More details on this setup and its workings are given in our earlier article [1]. We measured the gas-pressure-dependent x-ray output rate of the system at a distance of 10 cm in front of the bottom of the pear (Fig. 1). X- ray dose rates were derived from readings of thermoluminescence dosimeters (TLDs) consisting of three TLD-100 chips sealed in thin plastic. For the gas pressure giving the highest x-ray intensity, the dose rate on the surface of the glass was measured because Hoffmans and van Kleef and others had reported incidentally to have positioned the tube as close as possible to the object (e.g., the hand of the daughter of van Kleef). The radiation half-value layer in Al was determined from the entrance and exit dose on stacks of Al sheets of varying thickness. For this purpose, we again used TLD-100 dosimeters. Hoffmans original Crookes tube number 9 was reinvestigated. Dose rates were remeasured in the state in which we had retrieved the tube from the warehouse for normal and reverse polarity, and the visual appearance of the gas discharge was compared with the discharges observed for the evacuable tube. After it had become apparent that the pressure probably was just too high for x- ray production using normal polarity, we conditioned Hoffmans tube by running it the normal way for about 2 hours. Because we were concerned about causing damage to the tube, we would run it for periods of 5 minutes only and then wait, taking care that the glass did not become more than hand-warm ( 40 C). The reason for this conditioning is that the Fig. 4 Photograph shows evacuable Crookes tube number 9 and vacuum pump system. 756 AJR:198, April 2012

X-Rays in Radiology Dose Rate (mgy/min) 5.0 4.0 3.0 2.0 1.0 0.0 0.00 0.002 0.004 0.006 0.008 0.010 Air Pressure (mbar) vacuum in a gas tube is known to improve on prolonged normal use with an electron current that is not too high, even to the extent that x- ray production will finally stop [10, 11, 16]. One important mechanism of this evacuation is the implantation of gas ions into the cathode and the glass wall (e.g., for 50 kev nitrogen ions, the projected range in soda lime glass is about 138 nm; the projected range straggle is 45 nm [20]). Another mechanism is gettering by sputtered metal from the electrodes [4]. To increase the gas pressure again after prolonged use, the tube might be baked in an oven because, using this procedure, gas is again driven out of the glass and the Al cathode [3, 10, 11]. Very soon, vacuum regulators came into use [3, 4] (Fig. 1), but these could not prevent the tube s function from deteriorating over time. Figure 5 shows the x-ray dose rate at 10 cm in front of the evacuable tube as a function of the air pressure in the tube; for pressures higher than 0.004 mbar, discharges within the 11-cm spark gap of the Ruhmkorff inductor disappeared. The two top rows of Figure 6 show the appearance of the corresponding gas discharges. Clearly, both dose rate and appearance strongly depend on pressure. Comparison of gas discharges on photographs from Hoffmans Crookes tube number 9 with those from the evacuable tube at various pressures (Fig. 6) suggested that Hoffmans tube as retrieved had a pressure of about 0.01 mbar, which is just at the highpressure limit for x-ray production (Fig. 5). After conditioning, Hoffmans tube also Fig. 6 Crookes tube number 9. Top images show evacuable Crookes tube number 9 at various pressures in mbar (1 mbar = 100 Pa). Bottom row shows Hoffmans Crookes tube number 9 in original condition not generating x-rays (left) and after 2 hours of conditioning (right), now also generating x-rays. All tubes conventionally powered. Intensities not to same scale. Fig. 5 Graph shows dose rate at 10 cm in front of tube as function of gas pressure in evacuable Crookes tube number 9 using high voltage of normal polarity. gave off x-rays when driven with conventional polarity, and the discharge pattern had become different from when we started experimenting (Fig. 6). Table 1 shows the dose rate at 10 cm in front of the evacuable tube with a gas pressure of 0.003 mbar as well as that for Hoffmans Crookes tube number 9 after conditioning. For comparison, the dose rates reported in our previous study [1] are reproduced, including halfvalue layers in Al. In Table 2, dose rates on the hottest places on the glass surface are shown for the evacuable tube and the original Crookes tube number 9. Results in Perspective For a given high voltage, x-ray production occurred within a rather narrow pressure window (Fig. 5). This certainly explains part of the problems with gas tubes in the past. All early tubes would regularly need conditioning to give them the right vacuum for a particular task, and this conditioning would only work for a limited number of cycles [3, 10, 11]. Furthermore, the appearance of the gas discharge turned out to be very sensitive to the gas pressure (Fig. 6). This allowed us to estimate the approximate pressure in Hoffmans original Crookes tube number 9. It also makes understandable why early x-ray workers preferred unshielded tubes, because it allowed the assessment of the tube s functioning by observing the discharge and fluorescence with their bare eyes and the x-ray intensity and hardness by using a fluoroscope and a hand as a test object [3, 10, 17]. The fact that Hoffmans tube started generating x-rays after having run for some time under normal polarity is consistent with the hypothesis that gas pressure that was initially too high. We did not proceed with conditioning beyond 2 hours because we did not want to damage the equipment, which is fragile. The cause of x-ray production under reverse polarity in the as-retrieved Hoffmans Crookes tube number 9, notwithstanding the initial poor vacuum, must have been the different geometry in reverse mode. It is important to notice that the electrons receive most of their energy when still close to the cathode. In reverse geometry, the distances from the cathode to where the electrons hit a sur- AJR:198, April 2012 757

Kemerink et al. TABLE 1: X-Ray Dose Rate at 10-cm Distance From Various Gas Tubes Tube Polarity of Power Dose Rate (mgy/min at 10 cm) Half-Value Layer (mm Al) Hoffmans Crookes tube number 9 as retrieved [1] Reverse 0.86 0.56 Hofmanns Crookes tube number 1 [1] Normal 0.46 0.56 Hoffmans Crookes tube number 9 after conditioning Normal 0.27 Not measured Evacuable Crookes tube number 9 at 0.003 mbar Normal 3.90 0.80 TABLE 2: X-Ray Dose Rate on Surface of Hoffmans and Evacuable Crookes Tube Number 9 Tube Polarity of Power Dose Rate (mgy/min) Hoffmans tube as retrieved Surface glass stub around anode (Fig. 3) Reverse 50 Surface glass opposite anode (Fig. 3) Reverse 5.8 Surface bottom pear (Fig. 6) Normal < 0.005 Evacuable tube with 0.004 mbar air Surface bottom pear (Fig. 1) Normal 107 Hoffmans tube after 2-h conditioning Surface bottom pear (Fig. 6) Normal 10 face and can generate x-rays are much smaller than in the normal situation. A mean electron-free path that is small compared with the length of the tube (from cathode to the front of the pear, conventional polarity, resulting in many collisions and no x-rays) can be large compared with the distance from the side electrode to the local wall of the glass stub (reverse polarity, no or few collisions, and x- ray generation). This illustrates that the pressure for which x-ray generation takes place depends on the geometry and size of the tube. Röntgen stated in his third communication that he succeeded in observing x-rays for gas pressures as high as 3.1 mm Hg in a narrow tube (which is 4.1 mbar because 1 mm Hg = 1.33 mbar). The lowest pressure for which he observed x-rays was 0.0002 mm Hg, but he gave no further details [19]. Hoffmans Crookes tube number 9 in reverse mode was also an effective source of x- rays, as can be seen in Table 2. However, the evacuable tube under optimal conditions produced about 4 5 times more x-rays. It is noteworthy that for the same electron current, gas tubes with a Pt anticathode generated many more x-rays than Crookes tube numbers 1 and 9 because the efficiency of x-ray production scales with the atomic number, which is 78 for Pt, whereas for the main components of glass, silicon, and oxygen, the atomic numbers are only 14 and 8, respectively. A tube with a focusing cathode and a Pt target had already been constructed by Crookes in 1879 to show that cathode rays could heat the anticathode to red hot [21], and in early 1896 such tubes were applied [3, 4, 10, 11]. The attitude of x-ray workers with respect to not taking safety measures to prevent high exposures to themselves and their patients has been discussed by several authors [11, 14, 17, 22]. The consequences of the lack of shielding and the way of working are well known: serious damage to the hands of workers, frequently leading to amputations of fingers and eventually hands and arms, followed by premature death due to the spread of cancer [17]. On the memorial for x-ray martyrs in the garden of St. George s Hospital, Hamburg, Germany, were 169 names inscribed in 1936; in 1959 their number had increased to 359 [23]. Currently, x-ray production is a very wellcontrolled process thanks to high-tech instruments that resulted after innumerable developments in technology. A modern tube does not need maintenance and can be placed in well-shielded permanent housing. Modern electronics allow remote control, with, at least for diagnostic imaging, the x-ray operator at a safe position. Nevertheless, ghosts of the past linger because even modern tubes still need seasoning or conditioning, at least after manufacturing, and some systems even at the start of each working day. The purpose is to prevent x-ray beam instabilities by removal of any discharge paths on isolators and burning of surface irregularities and loose particles that could create field emission sites causing spurious discharges. The gaseous remains of this electrical cleaning and other gas freed during loading of the tube will be ionized by the electron beam and implanted or bound by a zirconium-graphite based getter. For a tube, the final vacuum is on the order of 10 9 mbar when cold and 10 4 mbar when fully loaded (Behling R, Philips Healthcare Hamburg, personal communication). What are the consequences of the evolution of tubes and associated equipment for radiologic workers? For purely diagnostic radiology, the occupational exposure is nearly negligible, but even the average dose of all workers in a department of radiology was found to be similar to the average residential dose [24]. Thus, in slightly more than a century, the risk of x-ray imaging has been lowered from a situation in which x-ray martyrs were made to a situation where many workers may receive a dose not higher than at home. Summary The unpredictable and time-dependent behavior of x-ray tubes in the first x-ray imaging systems was critically related to a varying tube vacuum. The resulting constant need for surveillance and system readjustment certainly contributed to the high exposures of early x-ray workers. At the same time, it is clear that a modern, stable, and well-shielded x-ray tube that is remotely controlled can be operated with minimal occupational exposure. References 1. Kemerink M, Dierichs TJ, Dierichs J, et al. Characteristics of a first generation x-ray system. Radiology 2011; 259:534 539 2. Hoffmans HJ. Proefnemingen met Röntgen sche Stralen in het laboratorium der Hoogere Burgerschool te Maastricht. Maastricht, The Netherlands, Leiter-Nypels, 1896 3. Mould RF. The early history of x-ray diagnosis with emphasis on the contributions of physics 1895 1915. Phys Med Biol 1995; 40:1741 1787 4. Rønne P, Nielsen ABW. Development of the ion x-ray tube: with a description of the collection in 758 AJR:198, April 2012

X-Rays in Radiology the Medical Historical Museum, University of Copenhagen (Acta Historica Scientiarum Naturalium Et Medicinalium, vol. 35). Copenhagen, Denmark: CA Reitzel Publishers, 1986 5. Addyman FT. Practical x ray work. London, United Kingdom: Scott, Greenwood & Co, 1901 6. Albers-Schönberg HE. Die Röntgentechnik. Hamburg, Germany: Lucas Gräfe & Sillem, 1903 7. Kassabian MK. Röntgen rays and electro-therapeutics with chapters on radium and phototherapy. Philadelphia, PA: JB Lippincott Company, 1907 8. Kaye CWC. An introduction to the study of röntgen rays. London, United Kingdom: Longman, Green and Co, 1914 9. Meadowcraft WMH. The ABC of the x rays. New York, NY: The American Technical Book Co, 1896 10. Morton WJ, Hammer EW. The x-ray. New York, NY: American Technical Book Co, 1896 FOR YOUR INFORMATION Mark your calendar for the following ARRS annual meetings: April 29 May 4, 2012 Vancouver Convention Center, Vancouver, BC, Canada April 14 19, 2013 Marriott Wardman Park Hotel, Washington, DC May 4 9, 2014 Manchester Grand Hyatt San Diego, San Diego, CA April 19 24, 2015 Toronto Convention Centre, Toronto, ON, Canada 11. Rollins W. Notes on x-light. Cambridge, United Kingdom: University Press, 1904 12. Snowden Ward H. Practical radiography. London, United Kingdom: Dawbarn & Ward, 1896 13. Thompson EP. Roentgen rays and phenomena of the anode and the cathode. New York, NY: D. Van Nostrand Company, 1896 14. Serwer DP. The rise of radiation protection: science, medicine and technology in society. BNL 22279 Informal Report Upton, NY: Brookhaven National Laboratory Associated Universities, 1976:chapter 1 15. Hulst H. Joys and sorrows of an x-ray worker. Phys Surg 1902; 24:492 498 16. Coolidge WD. A powerful Röntgen ray tube with a pure electron discharge. Phys Rev 1913; 2:409 430 17. Meggitt G. Taming the rays: a history of radiation and protection. Raleigh, NC: Lulu.com, 2008: chapter 1 18. Kemerink M, Dierichs TJ, Dierichs J, et al. First-generation x-ray system. Radiology 2011; 260: 611 612 19. Röntgen WC. Weitere Beobachtungen über die Eigenschaften der X-Strahlen. Sitzungsberichte der königlich preußischen Akademie der Wissenschaften zu Berlin. 1897; 26:576 592 20. James F. Ziegler Website. SRIM: the stopping and range of ions in matter. www.srim.org. Accessed June 27, 2011 21. Crookes W. Radiant matter. Lecture delivered before the British Association for the Advancement of Science at Sheffield. Friday, August 22, 1879 22. Herzig R. Suffering, sacrifice, and the formation of American roentgenology. Am Q 2001; 53:563 589 23. Vogel H. Das Ehrenmal der Radiologie in Hamburg. Ein Beitrag zur Geschichte der Röntgenstrahlen. Fortschr Röntgenstr 2006; 178:753 756 24. Kemerink GJ, Frantzen MJ, de Jong P, Wildberger JE. Less radiation in a radiology department than at home. Insights Imaging 2011; 2:275 280 AJR:198, April 2012 759