Technical Note. High-speed Agilent 1100 Series diode-array detector SL for optimization of resolution, sensitivity, spectral sensitivity and linearity

Similar documents
HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

Achieving More Efficient Data Review with OpenLAB CDS

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

Liquid Chromatography- Mass Spectrometer Manual

GCxGC Comprehensive Two-Dimensional Gas Chromatography

PAL LSI - Robust and Reliable Sample Handling

GCxGC Comprehensive Two-Dimensional Gas Chromatography

LC-4000 Series. Integrated HPLC and UHPLC systems

ICS-3000 System Installation Qualification

PAL LSI - Robust and Reliable Sample Handling

GC/LC-MS: data acquisition rate and peak reconstruction

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications

ABB MEASUREMENT & ANALYTICS. MB3600 The most reliable FT-NIR laboratory analyzer designed for QA/QC

Operating pressure should be 6000 psi or better suitable for both analytical and semi-preparative applications

ChromQuest 5.0 Chromatography Data System

BR2517B. 168 Diode Array & 166 Programmable Detector Modules. Packaging Processing. Bid on Equipment

Instrument Control and Data Acquisition with Chromeleon

Performance. Reliability. Productivity. Automated Flash Chromatography Systems

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer

ABB MEASUREMENT & ANALYTICS. MB3000 The most reliable FT-IR laboratory analyzer

OPTICAL POWER METER WITH SMART DETECTOR HEAD

SMT Encoder for High Performance, High Volume Designs Small Size High Resolution Low Cost ChipEncoder Reflective Surface Mount Encoder Features

If you have questions or problems in providing anything described as Customer

Customer Responsibilities. Important Customer Information. Agilent 1260 Infinity Clinical ed. LC Site Preparation Checklist

More reliable zeta potential measurements. without dilution. Agilent Particle Analysis Solutions. Agilent 7020 ZetaProbe Zeta Potential Analyzer

Fluorescence Spectrophotometer

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK

AIM INTRODUCTION SIMPLIFIED WORKFLOW

Practicality and Performance

Agilent 83437A Broadband Light Source Agilent 83438A Erbium ASE Source

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Staying Connected in GC

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

SmartMix Manual V5350A

111 Highland Drive Putnam, CT USA PHONE (860) FAX (860) SM32Pro SDK

Agilent N7744A 4-Channel Optical Multiport Power Meter N7745A 8-Channel Optical Multiport Power Meter. Fully compliant to LXI Class C specification

CBF500 High resolution Streak camera

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

specification MSE series MSE and MSE+ colorimeter

Creating and fully utilizing the retention time standard processing method for a more automated IC workflow

Agilent High Capacity RV Pumps & Roots Pumping Systems

Flash-Point with Closed Cup - Abel Method - ABA 4 (automatic)

013-RD

Operation Manual for the Refractive Index detector VE-3580

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

OpticalProducts. Illuminating Your Network. Testing the World s Networks

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Integration of the Zetasizer µv and OmniFACE into an SEC system

specification hyperion colorimeter

specification hyperion colorimeter

InSight Pro. Valuable real-time process insight

Ultimate quality control. No compromise. PROCESS / QUALITY CONTROL

Corporate Headquarters Dionex Corporation 1228 Titan Way P.O. Box 3603 Sunnyvale, CA Tel: (408) Fax: (408)

ET-5050x-BF1W Datasheet

GFT Channel Digital Delay Generator

Real-time ph and Conductivity Data Acquisition with the ACQUITY UPLC H-Class Bio System, Auto Blend Plus, and the GE Healthcare Monitor ph/c-900

CAEN Tools for Discovery

Critical Benefits of Cooled DFB Lasers for RF over Fiber Optics Transmission Provided by OPTICAL ZONU CORPORATION

IS 140 IGA 140 IS 140-PB IGA 140-PB IS 140-PN IGA 140-PN IS 140-ET IGA 140-ET

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

1995 Metric CSJ SPECIAL SPECIFICATION ITEM 6031 SINGLE MODE FIBER OPTIC VIDEO TRANSMISSION EQUIPMENT

FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091

MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer

LABSONIC M The ultrasonic homogenizer for every application. Benefits

Agilent 7000 Series Triple Quadrupole GC/MS System

SPECIAL SPECIFICATION 6911 Fiber Optic Video Data Transmission Equipment

The extremely compact laser head is approximately 480 mm long and can

High Power ARNS/IFF Limiter Module: Ultra Low Flat Leakage & Fast Recovery Time

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003

NMR. picospin. Maintenance Guide

SPECIAL SPECIFICATION 1987 Single Mode Fiber Optic Video Transmission Equipment

ECE 4220 Real Time Embedded Systems Final Project Spectrum Analyzer

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Quartzlock Model A7-MX Close-in Phase Noise Measurement & Ultra Low Noise Allan Variance, Phase/Frequency Comparison

MTS/T-BERD Platforms WDMPMD Module

SPECIAL SPECIFICATION 1291 Fiber Optic Video Data Transmission Equipment

Control Unit CU 2.1 User Manual V6850A HPLC

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

PulseCounter Neutron & Gamma Spectrometry Software Manual

specification display applications cronus spectrometer & colorimeter

Agilent 81600B Tunable Laser Source Family

Kingbright. L-7104YD-12V T-1 (3mm) Solid State Lamp DESCRIPTIONS PACKAGE DIMENSIONS FEATURES APPLICATIONS ATTENTION SELECTION GUIDE

Real-Time PCR System TC

Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper

Compact live fiber identifier with integrated optical power meter

Advanced Test Equipment Rentals ATEC (2832)

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

Fast Particle Size Distribution Measurements with SMPS utilizing a Novel Classifier

Philips CoralCare FAQ

OPTICAL MEASURING INSTRUMENTS. MS9710C 600 to 1750 nm OPTICAL SPECTRUM ANALYZER GPIB. High Performance for DWDM Optical Communications

SPEAR 3: Operations Update and Impact of Top-Off Injection

2 MHz Lock-In Amplifier

Agilent 81600B Tunable Laser Source Family Technical Specifications August New model: nm, low SSE output!

R&S FSW-K54 EMI Measurement Application Detecting and eliminating electromagnetic

4 MHz Lock-In Amplifier

Gamma instabus. Technical product information

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X

Transcription:

High-speed Agilent Series diode-array detector SL for optimization of resolution, sensitivity, spectral sensitivity and linearity Technical Note Abstract This application has been verified using an Agilent Series LC system, and showed comparable or even better performance. Application Area Fast screening in drug development, food and environmental applications. Standard UV detectors with a maximum of Hz ( data points per second) are not fast enough to ensure the collection of enough data points (at least data points per peak) to provide reliable results for peaks with a peak width at half height as small as.3 s. This Technical Note describes the new Agilent Series diode-array detector SL (DAD SL), which addresses the need for fast and ultra-fast LC. Besides 8 Hz data acquisition for highest resolution in ultra-fast LC the Agilent Series DAD SL provides high precision, linearity and sensitivity. In addition, a built-in data recovery card ensures data never lost insurance. RFID tags for cells and UV lamps ensure unambiguous data traceability. Electronic temperature control provides highest baseline stability and practical sensitivity under fluctuating temperature and humidity conditions. The new standard flow cell increases practical sensitivity with imized dispersion and RI influence.

Introduction The demands of high sample throughput in condensed time frames have given rise to high efficiency, fast LC. From the processing of hundreds of samples in overnight runs, to efficient and timely screening of metabolic studies, to rapid method development, and to reducing solvent disposal costs, fast chromatography has become a necessity in the chromatography lab. Using this new methodology, results can be reported in a few hours, rather than a day or even later. Complete validated results in such a short time mean that manufactuered goods can be released the same day that they are produced. The end result is greater productivity for customers and greater cost efficiency. This Technical Note illustrates the accurate quantitation and spectral confirmation of impurities at trace level under ultra-fast HPLC conditions with the Series HPLC system. To provide a simple structure for discussion, we have categorized fast LC into the following areas: Conventional LC means cycle times > 5 utes Fast LC means cycle times < 5 utes Ultra fast LC means cycle times < utes The later results in a gradient time of..5 utes, cycle times of.5.5 utes and 5 % peak width of.. seconds. Standard UV detectors with a maximum of Hz ( data points per second) are not fast enough to ensure the collection of enough data points (at least data points per peak) to provide reliable results in ultra-fast HPLC. The Agilent Series diode-array detector SL addresses the need for fast and ultra-fast LC by offering the following new features and benefits: 8 Hz data acquisition of up to 8 signals - for more ultra-fast, high-resolution quantitative LC 8 Hz full spectral data acquisition - for ultra-fast peak purity analysis and spectral confirmation Built-in data recovery card - for a data never lost insurance RFID tags for all flow cells and UV lamps - for unambiguous data traceability Improved diode-array front-end electronics - for imized noise (typical < ± 6 µau ASTM) LAN on board - eliates the need for additional LAN interface Built-in web-server, USB, PCM- CIA (WLAN, Bluetooth) - for a future-proof design Electronic temperature control ETC - for maximum baseline stability and practical sensitivity under fluctuating ambient temperature and humidity conditions New standard flow cell - for imized dispersion and RI-influence Micromechanical slit - for automated slit width changes during method Dual lamp design - for optimal light intensity and thus maximum sensitivity In the following examples we will describe the influence and bene- fits of the new features on fast and ultra-fast applications. Hints and tips are provided for setting detector parameters according to the specific needs of ultra-fast applications. Detector performance regarding fast data acquisition, noise, linearity, spectral data acquisition and peak purity data at trace levels are evaluated. Experimental The instrument used was an Agilent Series high-throughput (HT) LC system, equipped with an Agilent Series well-plate sampler with cooling option, an Agilent Series binary pump with optional degasser, an Agilent Series column compartment and the Agilent Series diode-array detector SL. The columns used were short ZORBAX SB C-8, packed with.8-µm particles. The Agilent Series high-throughput LC system is modified versus the standard system for the shortest flow connection and the standard mixer is replaced by a mixer with 8-µL (frit) volume. When the Agilent Series DAD is switched on for the first time or when selecting the default method, default values for all modules are set automatically. These default values are a good starting point for standard applications with a run time > 5 utes. If fast (<5 utes) and ultra-fast runs (< utes) have to be performed the Agilent Series DAD parameters have to be modified to provide the best performance for high-speed applications with optimal resolution, optimum signal sensitivity, best spectral sensitivity or best spectral resolution.

Chromatographic conditions: Column: 3 x 4.6 mm ZORBAX SB C8,.8 µm Instrument optimized for lowest delay volume: 8-µL (frit volume) mixer, short flow capillaries Mobile phases: water (A) and acetonitrile (B) Flow rate: 5 ml/ Gradient: at 5 %B, at.3 %B, at.5 %B, at.6 5 %B Column temp.: 5 C Injection vol.: 3 µl Detector: Sample wavelength 45/ nm, reference wavelength 36/8 nm, data rate from 5 to 8 Hz, 5-µL volume detector cell with 6-mm path length, optical slit width 4 nm 5 5 5 5 5 5 Resolution peak 5 =.5 at 8Hz...3.4.5 Resolution.5 at 4Hz...3.4.5 Resolution.7 at Hz...3.4.5 Optimization of data rate for highest resolution in high-speed applications For LC/UV analysis with cycle times around ute the data rate of the detector can become a limiting factor resulting in peak broadening and reduced peak resolution. In figure an example is given how significantly the data rate influences the peak performance and consequently the quality of results. 5 5 5 5 Resolution.7 at Hz...3.4.5 Resolution.67 at 5Hz...3.4.5 Set of 9 compounds, ng/μl each, dissolved in ACN. Acetanilide. Acetophenone 3. Propiophenone 4. Butyrophenone 5. Benzophenone 6. Valerophenone 7. Hexanophenone 8. Heptanophenone 9. Octanophenone This example demonstrates that a UV detector with only 5, or Hz data rate is not suitable for a demanding application. At and Hz the peak width is increased 4 %, respectively % compared to the peak width obtained for a data rate setting of 8 Hz. Peak width directly influences resolution and peak capacity (table ). At a data rate of 8 Hz the peak width is about.3 s at half height. This means an increase in peak capacity of 4 % versus a data rate of Hz. The resolution is improved by 3 %, which results in an improvement in the column efficiency of 7 %. Compared to a data rate of Hz the improvement is even more Figure Influence of data rate on resolution. Data rate Peak width Resolution Peak capacity 8 Hz.3.5 6 4 Hz.39.5 56 Hz.46.7 55 Hz.666.7 8 5 Hz.36.67 6 Table Influence of data rate on peak width, resolution and peak capacity. apparent. Peak capacity is increased by %, resolution by 9 % and the column efficiency by 6 %. These results clearly demonstrate that data rates above Hz are needed to take full advantage of fast and ultra-fast LC. 3

Setting the Agilent Series DAD SL parameters can be done either through the ChemStation software or with the handheld controller. In the ChemStation the parameters are set in the Agilent Series DAD setup screen (figure ). The data rate is selected in the Peak width (Response time) field. The correlation-todata rates in Hz units is given in table. Optimization of noise The analysis of trace compounds in the presence of a main compound is a common application problem for pharmaceutical samples. The goal here is to be able to simultaneously quantify a compound and its byproducts at a level of.5 %. To achieve this goal the UV detector needs to provide a imum linear range from to m AU. One very important parameter for the precise analysis of trace compounds is the signal to noise ratio. Figure Set up screen for DAD parameters. Peak width at half height sec Response time (sec) Date rate (Hz) <.5.5 8 >.5 >.5.5 8 >.5 >.3. 4 >. >.6. >.3 >..5 >.5 >3. 5 >. >6..5 >. > 4..5 >.4 >4 8..6 >.85 >5 6..3 Table Correlation between peak widths, response time and data rate in Hz. 4

Chromatographic conditions: Eluent: water/acn = 7/3 Flow rate: ml/ Column: 4.6 x 3 mm ZORBAX SB C8,.8 µm Temperature: C DAD: 54 nm, 6 nm, ref 36, 8 nm PW: >. (.5 Hz, s RT), standard 3-µL detector cell with -mm path length This ensures that the detector noise is as low as possible. Figure 3 and table 3 illustrate the influence of the data rate and optical slit on the short-term noise of the Agilent Series DAD SL. At a data rate of.5 Hz the influence of the slit width on the ASTM noise was evaluated. The ASTM noise specification is µau peak-to-peak (± µau). Under the given chromatographic conditions this is fulfilled for all slit width settings. In table 3 the influence of data rate and slit width on ASTM noise is combined, showing that the lowest noise level is obtained using low data rate settings and an optical slit width of 6 nm. However, setting the data rate at 8 Hz and using a 4-nm slit would produce a noise level of 4 µau which would still be sufficient to quantify by-products and impurities at a level of.5 %.. Under actual ultra-fast LC gradient conditions, however, baseline noise may increase to a certain extent. In this case 8-nm or 6-nm slit and/or 4-Hz data rate can be chosen to reduce noise and achieve highly demanding quantitative limits while accepting a certain trade-off in peak resolution and spectral quality. The degree of reduced resolution and spectral accuracy depends on analytical conditions and the natural bandwidth of compound spectra. -. -. -.3 -.4 -. -.4 -.6 -.8 -. -. -.4 -.6 Figure 3 Influence of optical slit width on noise. 6-nm slit width Noise < ±. μau 5 5 8-nm slit width Noise < ±.7 μau 5 5 4-nm slit width Noise < ± 3.7 μau 5 5 4-nm slit 8-nm slit 6-nm slit Peak-to-peak noise in Peak-to-peak noise in Peak-to-peak noise in µau µau µau 8 Hz 4 3 3 4 Hz 3 6 Hz 6 Hz 5 8..5 Hz 7.4 5.4 4. Table 3 Influence of optical slit width and data rate on peak-to-peak noise..5-hz data rate is not sufficient for fast LC, as shown in figure. In this case 8 or 4 Hz with a 6-nm slit width is the best compromise. 5

Chromatographic conditions: Column:. x 3 mm ZORBAX SB C8,.8 µm Flow rate: ml/ Mobile phase: water/acetonitrile = /8 Column temp.: 5 C Injection vol.: µl DAD: 5 nm, 4 nm, ref 36, 8 Hz, slit 4 nm, 5 µl detector cell with 6-mm path length A practical example for the evaluation of limits of detection (LOD) at different data rates is given in figure 4. pg/µl of anthracene were injected at,, 4 and 8 Hz. Using these ultra fast chromatographic parameters the LOD at Hz is about.67 pg, which is the lowest achievable level under these conditions but the resolution is far better at, 4 or 8 Hz. In this case a good compromise would be to use the 4 Hz setting with less sensitivity but with better resolution and more reliable integration for improved quantitation. Evaluation of linearity Measuring high and low concentrations in one run is only possible if the detector offers a wide dynamic range. In figure 5 the linearity was tested using caffeine as a test compound. Two experiments were done, one with only the UV lamp on and the second one with both UV and Vis lamps on. 4 3.5 3.5.5.5 Anthracene pg injected in μl 8 Hz, LOD =.3 pg, PW =.876 s 4 Hz, LOD =.48 pg, PW =.87 s Hz, LOD =.3 pg, PW =.864 s Hz, LOD =.67 pg, PW =.98 s...3.4.5.6.7.8.9 Figure 4 LOD of anthracene at different data rate settings. % Deviation from Linear Value 5.%.% 95.% 9.% 85.% 8.% Vis Lamp off Vis Lamp on 5 5 5 3 Absorbance / Figure 5 Deviation of linearity of the Agilent Series DAD SL, caffeine at 67 nm. The result is that the Agilent Series DAD SL provides a linear dynamic range for the tested caffeine sample up to.5 AU with a deviation of 5 % at.5 AU. At AU the deviation is about %. A common specification for a UV detector is that at. AU the deviation should be in the 5 % range. These experiments also demonstrate that the tungsten lamp could be switched off to maximize the linear range if the analytical wavelength is chosen in the UV range. However, there is a twofold trade-off when switching the visible lamp off. First, the positive effect of using a reference wavelength to increase practical sensitivity by reducing baseline drift and wander is sacrificed due to increased noise in the visible range, where the reference wavelength is typically chosen. Second, qualitative analytical results for peak purity analysis, library searches and spectral confirmation are sacrificed, because spectral quality decreases when the visible lamp is switched off, especially at trace levels (compare figures 4 and 5). 6

Chromatographic conditions: Sample: Phenone test mix Column: 4.6 x 3 mm, 3.5 µm ZORBAX SB C8 Gradient: 5- % ACN in.3 Flow rate: 5 ml/ Temperature: 4 C Data rate: 4 Hz Flow cell: 5 µl Evaluation of precision for areas and retention times for ultra-fast LC For analysis times of 4 s, total cycle time of 55 s (figure 6) and demanding gradients of.3 utes the precision of retention times was detered to be between.7 and. % RSD. The precision for the areas was between.5 and.3 % RSD. The exact data are shown in table 4. 8 6 4 Overlay of 6 runs.5..5..5.3.35 Set of 9 compounds, ng/μl each, dissolved in ACN. Acetanilide. Acetophenone 3: Propiophenone 4. Butyrophenone 5. Benzophenone 6. Valerophenone 7. Hexanophenone 8. Heptanophenone 9. Octanophenone Figure 6 4 second analysis of a phenone test mix. This precision data is typically useful for screening experiments and semi-quantitative work. Lower % RSD values, however, may be needed to comply with strict regulatory performance requirements. One way to achieve this goal is to decrease gradient speed. Figure 7 shows that when using a less demanding gradient and higher concentrations the precision for retention times is between. and.5 % RSD and the precision for areas is between. and.5 %. The more detailed results are summarized in table 5. Both tests show that even under highly demanding analytical conditions with run times below ute very good precision can be achieved. 8 6 4 Overlay of 6 runs...3.4.5.6.7 Set of 9 compounds, ng/μl each, dissolved in ACN. Acetanilide. Acetophenone 3. Propiophenone 4. Butyrophenone 5. Benzophenone 6. Valerophenone 7. Hexanophenone 8. Heptanophenone 9. Octanophenone Figure 7 48 second analysis of phenone mix Peak RSD RT (%) RSD Area.7.3.58.7 3.43.9 4.34.3 5.3.9 6.9.9 7.6.5 8.4.7 9.. Peak RSD RT (%) RSD Area.47.47.44.9 3.3.3 4.7.7 5.5.48 6.3.4 7..33 8.7.35 9.5.7 Table 4 % RSD for retention times and areas for the phenone test mix with a highly demanding gradient. Table 5 % RSD for retention times and areas for the phenone test mix with a less demanding gradient and at a higher concentration. 7

Chromatographic conditions: Column: 4.6 x 5 mm ZORBAX SB C8 RRHT.8 µm Solvents: A=water (. % FA), B=ACN (. % FA), Water MilliQ, ACN: Merck Grad., FA: 96 % Flow: 4. ml/ Temperature: 6 ºC Gradient: 5 7 % B in.85, 7 % B for.5, 5 % B for.5, total.5 UV-Detection: UV=45 nm (), ref. 46 nm (8), range 9 5 nm store all spectra, 8 Hz Injection: 5. µl injections of a mixture of nifedipin and nimodipin in the following ratios: :5, :, : and :4 Agilent Series well-plate sampler with imized carry over, overlapped injection, automatic delay volume reduction, injections 6 4 8 6 4 Nifedipin 3. at.% level.5..5..5 -.5.5.35.45 DMSO Nimodipin..4.6.8 Application example: Analysis of nimodipin as the main compound and nifedipin as the by-product To mimic the analysis of impurities beside a main compound as is common in process control or stability analyses the structurally related compounds nifedipin and nimodipin were mixed in different ratios (figure 8). Our sample of nimodipin as the main compound and nifedipin as the trace compound was analyzed using ultrafast chromatographic conditions with run times of ute. Precision for both compounds was evaluated as well as spectral performance and peak purity. Nifedipin was also injected in a ca. 5-fold concentration (approx. 6 µg/ml) compared to the highest concentration (. µg/ml in the :5 mixtures) in the subsequent experiments. Nimodipin was injected in the same concentration as in the subsequent experiments (approx. 55 µg/ml) and in a /th dissolution (approx. 55 µg/ml). It was discovered that Figure 8 Analysis of nimodipin in the presence of trace levels of nifedipin and some degradation products. 8 6 4 5 4 3 Nifedipin.339.343.346 Nifedipin, degradation products.3.35.4.45.5.55 Nifedipin Nimodipin, degradation products Nimodipin Nimodipin, degradation products..4.6.8 Figure 9 Chromatogram of :5 mixture at 45 nm using different instrument set ups with different delay volumes. 8

nifedipin as well as nimodipin showed a distinct degradation in the DMSO stock solution. The identity of the nifedipin peak was established by comparison to a freshly prepared solution and by LC/MS analysis. Figure 9 shows the influence of different configurations on the peak resolution. The configuration providing the lowest delay volume was chosen (figure ). After selecting the optimum configuration the performance of the analysis was evaluated. In figure the precision for retention times and areas of the main compound are shown. Both precision of retention times and areas for the main compound are very good and quantitation is reliable and robust. In figure the precision of retention time for the trace compound is evaluated. The relative standard deviation for nifedipin is below. %. The precision for areas is < 5 % relative standard deviation for the.5 % level. Chromatographic conditions: Gradient: 5 7 % B in.85 Column: 4.6 x 5, ZORBAX SB C8,.8 µm Injection: 5 µl of 55 µg/ml nimodipin Flow rate: 4 ml/ Flow cell: 3 µl Data rate: 8 Hz Slit: 8 nm Wavelength: 45 nm Binary pump WPS TCC 3-μl heater 4.6 mm x 5 mm RRHT ZORBAX SB C8 column DAD SL (3-μl drill cell) 3 μl Figure Optimized instrument set up for ultra fast LC. 75 5 5 75 5 5 55 μg/ml Nimodipin = % level Data rate 4 Hz Peak width.7 sec.7 (.7 ) x 9-mm capillary.7 (.7 ) x 5-mm capillary).7 (.7 ) x 9-mm capillary 95 9 85 8.67.67.67.67.67.67.67.67.68.68.6.6.63.64.65.66.67 Figure Precision of retention time and area of the main compound. Standard assembly without standard mixer and with 8 μl frit filter.7 (.7 ) x 4-mm capillary.7 (.7 ) x 5-mm capillary.7 (.7 ) x appr. 8-mm capillary (std. inlet capillary of 3-μl drill cell) Overlay of analyses at 45 nm RT Precision:.67 % RSD Area Precision:.3 % RSD 75.65.66.67.68.69 Chromatographic conditions: Nifedipin: Nimodipin = :5 Column: 4.6 x 5, ZORBAX SB C8,.8 µm Gradient: 5-7 % B in.85 Injection: 5 µl Flow rate: 4 ml/ Flow cell: 3 µl Data rate: 8 Hz Slit: 8 nm Wavelength: 45 nm 6 5 4 3 Overlay of analyses..5.3.35.4 Figure Optimized instrument set up for ultra-fast LC. Nifedipin A =.5 (. % level) RT Precision =.9 % RSD Nifedipin degradation product A =.5 (ca..3 % level) RT Precision =.3 % RSD.45.5.55 9

4 8 6.69.6.8 The limit of detection for the trace level impurity nifedipin is far below the.5 % level (figure 3). This means even under ultra-fast LC conditions the Agilent Series DAD SL allows accurate quantitation of impurities and by-products at levels less than.5 % of the main compound(s). Accurate 8 Hz spectra collection allows peak purity data to be obtained with high reliability using ultra fast chromatographic conditions. In figure 4 the spectra of nifedipin at a. % level and the overlay of the spectra of the nifedipin (A) and nimodipin (B) at a high concentration level are shown. Even at the. % level the identification in the low range is possible. Based on the high spectral acquisition rate, peak purity of the trace compound nifedipin could be evaluated, showing that the peak is pure and no other compound is co-eluting (figure 5). To provide accurate data it is important to select the Calculate Threshold option. This ensures that even at trace levels calculations are based on correct threshold values. Norm.75.5.5.75.5.5 A Norm Overlay of 8 Hz Reference and Apex Spectrum.75 Nifedipin Ref: 8.5 (high concentration level).5 Nifedipin Apex:.8 (.% level) Match Factor = 963.75.5.5 3 4 5 6 7 8 9 nm Figure 4 Spectra of nifedipin and nimodipin at different levels..69 Fl.75.5.5.75.5.5 -.5 Figure 5 Peak purity analysis of the nifedipin peak at a. % level. B Overlay of 8 Hz Reference and Apex Spectrum Nimodipin Ref: 8 (high concentration level) Nifedipin Apex:.8 (. % level) Match Factor = 99 3 4 5 6 7 8 9 nm Overlay of extracted Nifedipin spectra at trace level 3 4 5 6 7 8 9 nm.5 Nifedipin at trace levels Peak width =.63 sec Noise = 4 μau =. % level S/N = 5 DAD SL: 54,4 / No REF DAD SL: 54,4 / 36, =.5 % level S/N = 5 ~ μau/ C.5 DAD B: 54,4 / 36, DAD B: 54,4 No REF.5 =.5 % level S/N =.3.3.3.33 Figure 3 Signal height to noise ratio for different trace levels. Figure 6 Ambient Rejection: Comparison between Agilent Series DAD SL and Agilent Series DAD ( B -model). Conditions: Relative humidity = 6 % RH = const; Temp = 5 C +/- C; 4 x h cycles. Note: By keeping RH=const, the absolute humidity is strongly modulated due to temperature variations (worst case).

More features and benefits of the Agilent Series DAD SL Data recovery card DRC This card is situated at the back of the instrument and offers: All signals, spectra and meta data are buffered on a highcapacity, embedded 56 MB compact flash card. Prevents any data loss in case of communication breakdown between instrument and PC. Automatic run recovery in case of temporary communication failures. Manual run recovery in case of permanent communication failures after software, PC, and/or instrument re-boot. Radio frequency identification tags For more compliance the lamp and cell now also have identification tags: RFID tags records all relevant data necessary to recall instrument conditions under which a run has been executed. Minimizes the risk of false data interpretation, because measurement conditions are documented. Meta data stored on RFID tags are saved with each raw data file for unambiguous answers to (auditor) questions like: - Which type of flow cell was used to generate this chromatogram what was the path length and volume? - Did the accumulated burntime of the lamp exceed the pre-defined limit? For the flow cell the following data are stored as part of the method: path length volume maximum pressure date last test passed product number serial number production date For the lamp the following data are stored: accumulated on-time actual on-time number of ignitions date last test passed product number serial number production date All listed parameters can be printed as part of reports. Electronic temperature control ETC This new feature ensures more baseline stability for demanding environments, providing: compensation for changes in ambient conditions (temperature and humidity) reduced baseline wander for improved practical sensitivity and reproducibility under harsh environmental conditions In figure 6 a comparison is made between the Agilent Series 5-5 - -5 - -5 RI test gradient New standard flow cell provides 3-4x lower RI-sensitivity 4 DAD/MWD with old standard flow cell DAD SL/MWD SL with new standard flow cell DAD B and the Agilent Series DAD SL regarding sensitivity versus ambient relative humidity and ambient temperature. The improved ambient rejection on the Agilent Series DAD SL ensures that drift and wander is reduced to < 3 / C versus ~ / C for the Agilent Series DAD B version. New 3-µL standard flow cell design The new flow cell is based on drilled flow path with improved flow characteristics and in addition a ceramic ring was installed for thermal decoupling. This provides: reduced RI-sensitivity (figure 7) reduced peak dispersion imized noise in high-flow, high-temperature applications Several cells are available for the Agilent Series DAD SL covering a wide range of applications. This ensures that the Agilent Series DAD SL is compatible with conventional LC, capillary LC, nano LC as well as with preparative LC systems. 9 36 3 4 5 6 7 Figure 7 Influence of the cell design on refractive index behaviour.

The following cell types are available: Standard: 3 µl, -mm path length, bar Semi-Micro: 5 µl, 6-mm path length, bar Micro:.7 µl, 6-mm path length, 4 bar Semi-Nano: 5 nl, -mm Nano: path length, 5 bar 8 nl, 6-mm path length, 5 bar Preparative: 3 mm, bar Preparative:.3 mm, bar Preparative:.6 mm, bar For fast and ultra-fast LC applications 3 cell types are recommended: 3-µL standard flow cell: for highest sensitivity highly demanding quantitative work, e.g. analytical method development, QA/QC 4.6 3-mm id columns.7-µl micro flow cell: For highest selectivity Ultra-fast semi-quantitative work, e.g. screening experiments, HT LC/MS/UV. -mm id columns 5-µL micro flow cell: Best compromise for sensitivity and selectivity For good quantitative and qualitative results, e.g. screening, HT LC/MS/UV, early formulation studies 4.6 -mm id columns In table 6 an overview is given about the influence of different cell types on resolution and sensitivity for 4.6-mm and.-mm internal diameter columns. The three cell types are stainless steel cells, which can be used from ph up to ph. The.7-µL cell offers the best performance if resolution is the leading requirement. This cell is also recommended if the UV and MS are serially connected and resolution should be as good as possible. For optimum sensitivity the 3-µL with a -mm path length is the best choice. A good compromise is the 5-µL cell with a 6-mm path length for applications where resolution and sensitivity is of equal importance. Flow cell Sensitivity Resolution pathlength & linearity 3 µl/ mm +++ + 5 µl/6 mm ++ ++.7 µl/6 mm + +++ Table 6 Influence of cell types on sensitivity and resolution Kits for optimizing the Agilent Series LC system for high throughput analysis are offered to reduce system delay volume to an appropriate level for different column dimensions and corresponding flow rates. Three kits are available: Fast LC modifications for Agilent Series instrument with DAD, 4.6-mm id columns (Agilent p/n 588-534) Ultra-fast LC modifications for Agilent Series instrument with DAD/MS,.-mm id columns (Agilent p/n 588-538) Ultra-fast LC modifications for Agilent Series instrument with VW UV detector, 4.6-mm id columns (Agilent p/n 588-533) Conclusion The Agilent Series DAD SL provides more accurate results faster. A significantly higher peak capacity and better data security using the 8 Hz data acquisition enables highest resolution in ultrafast LC. The examples have proven the high precision, excellent linearity and high sensitivity. This prevents compromising data quality under ultra-fast LC conditions and ensures compliance with strict regulatory performance requirements. The proven dual lamp design allows for spectral analysis at trace levels (DAD SL only). The high quality of the rugged Series HPLC, the ZORBAX RR-HT column and method stability enables robust 4 x 7 operation. Angelika Gratzfeld-Huesgen and Michael Frank are Application Chemists, Stefan Schuette is Product Manager at Agilent Technologies, Waldbronn, Germany. www.agilent.com/chem/ 5-7 Agilent Technologies Inc. Published April, 7 Publication Number 5989-37EN