Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Similar documents
Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

FEI FIB Focused Ion Beam

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

Figure 1. MFP-3D software tray

MTL Software. Overview

Topic: Instructional David G. Thomas December 23, 2015

E X P E R I M E N T 1

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Checking the fluids of the cart

Series 1100 ColorTS Servo Manual Registration System

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

RADIO FREQUENCY SYSTEMS

Activity P27: Speed of Sound in Air (Sound Sensor)

PHI 5000 VersaProbe TM Operator s Guide

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Transmitter Interface Program

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

SC24 Magnetic Field Cancelling System

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Fig. 1. The Front Panel (Graphical User Interface)

SC24 Magnetic Field Cancelling System

SC26 Magnetic Field Cancelling System

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

Standard Operating Procedure of nanoir2-s

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

Guide to the FACS Calibur

Operating Instructions

PulseCounter Neutron & Gamma Spectrometry Software Manual

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Three Axis Digital Readout System

1-Touch Vibratory Sieve Shaker SS-10

Approved by Principal Investigator Date: Approved by Super User: Date:

ivw-fd122 Video Wall Controller MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Page i Rev. 1.

Approved by: / / R. Battaglia 12/16/2016

PSC300 Operation Manual

Mini Micro Pulse Lidar System

Simple and highly effective technology to communicate your brand s distinctive character

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

GEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor

TECHNICAL MANUAL DRAW-WIRE DISPLACEMENT TRANSDUCER TYPE DWT

TE 86 MULTI-STATION HIP JOINT SIMULATOR

HRXRD Analysis of Epitaxial Thin Films

INSTALATION PROCEDURE

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

CytoFLEX Flow Cytometer Quick Start Guide

University of Minnesota Minnesota Nano Center Standard Operating Procedure

TOMELLERI ENGINEERING MEASURING SYSTEMS. TUBO Version 7.2 Software Manual rev.0

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

Operation Procedure for Phillips XL30 ESEM

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

12months. on-site warranty. DZE ELECTRONIC PRESSURE SWITCH for detection of overload per EN 81 2 featuring two adjustable switching points

BROTHER, BROTHER & SONS

D-901 PC SOFTWARE Version 3

INSTRUCTION AND MAINTENANCE VIDEO MOTORE PROJECTION SCREEN

PogoStick and Research Pogo App Operator s Guide

180w led gobo moving USER MANUAL

ASE 369 K Measurements and Instrumentation. LAB #9: Impulse-Force Hammer; Vibration of Beams

Casambi App User Guide

Stage Wash 7x 10W LED Moving Head (RGBW)

Import and quantification of a micro titer plate image

Introduction to the FTA2000

Digital Video Recorder

Working with CSWin32 Software

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA

Unicable II Programmer. IDLU-PROG01-OOOOO-OPP Item: Installation & User Guide

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English

INSTALLATION INSTRUCTIONS RGB / RGBW 24V LED TAPE

ST350 Radiation Counter

DektakXT Profilometer. Standard Operating Procedure

DS-7200HVI/HFI-SH Series DVR Quick Operation Guide

EVD-L04/100A1-960, EVD-L08/200A1-960 and. EVD-L16/400A1-960 DVRs. Quick Operation Guide

VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below.

Revision 1.2d

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5)

Noise Detector ND-1 Operating Manual

Figure 1: Device components

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0

Operating Instructions

Vibratory Deck Sieves 15 in. (380 mm)

Quick Operation Guide of LTN7700/7600 Series NVR

EDL8 Race Dash Manual Engine Management Systems

Troubleshooting the CTS 1100

RK-2 ENVIRONMENTAL DATA CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER WITH FLANGED HEAD. Main characteristics

Installation and Setting up Instructions for the 990 Signal Conditioning Instrument

Table of Contents. Versa TILE & Versa DRIVE D2 Quick Start Manual

Overview. Signal Averaged ECG

800 Displaying Series Flowmeter

USER MANUEL. SNIPE 2 Ref R13

Technology Control Technology

HD-CM HORIZON DIGITAL CABLE METER

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE

Transcription:

4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject to change. We hold no responsibility for errors in the protocol or user error that lead to personal or property damage. Please follow all federal, local, and institutional rules and regulations when using this protocol. Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. Potential Hazards Hydraulic testing systems have the potential to do great bodily harm if used improperly. Never have any part of your body below the actuator when the hydraulics are on. Materials Several cotton swabs (to use dowels for practice test) 50lb Sensotec load cell (S/N 997569) External LVDT (050 MHR) Small frame with extra set of springs Jepsen 4pt mouse fixtures Squeeze bottle of 1X Phosphate Buffered Saline (PBS) Glass petri dish Small pair pickups 4x4 gauze Method Equipment Set-Up 1. Set-up small frame with a set of extra springs 2. Mount the load cell and connect to the proper load cell cable 3. Mount the external LVDT and connect to the DTR-451 input cable 4. Make sure the points are aligned properly. See the image below for properly aligned points. a. Set the brass point first. b. While holding one of the lower points, loosen the screw (barely) for that point. If you don t hold the point that you are loosening, it will rotate, which causes alignment issues. Hold the point in position as you tighten it! You want the gap between the two bottom points to be consistent from front to back. c. If you are having issues with positioning, spin the bottom two points around. They are easier to align from one direction. d. Make sure all 4 screws in the image are tight before testing begins. 4pt Bending, Mouse Page 1 of 6 Revision 1

MTS System Set-Up 1. Turn on computer 2. Turn on the Teststar box (Switch is located on the back-top right) 3. Turn on DTR-451 (Digital Transducer Readout) 4. Turn on main hydraulics in pump room (turn to low first, then to high) 5. Log-in to computer (user: ORL user; password: trabeculae) 6. Double Click Station Manager Icon 7. Select a configuration file (ex. External_lvdt.cfg) 8. Select a parameter set (ex. ext_lvdt) 9. Click Open 10. Select MPT icon from left task menu 11. Display Signal Auto Offset by selecting the first button under the Station Controls (this field box contains the input signals for the parameter set you selected and allows you to zero any of the inputs by selecting the button in between the current value and offset boxes) 12. Create a new output folder for the day in the directory C:\MTS_Specimen_Data (every user should create their own folder under their name or the study name and then organize data by date) 13. Display the MPT Specimen Editor by clicking the third button from the left on the MPT control panel (it looks like 3 dumbbells) and set the Specimen Directory Path to the folder for the day that you just created in step 12 (ex. C:\MTS_Specimen_Data\User\StudyName\TodaysDate). 14. Close Specimen Editor 15. Reset hydraulic interlocks on the Station control panel. Select the HSM Low button, then the High button. This can also be done on the Remote Station Controller. 16. Select Enable Manual Contol on the Remote Station Controller and find the center of the actuator s traveling distance (+5 mm). Zero the axial displacement on the Signal Auto Offset panel. 4pt Bending, Mouse Page 2 of 6 Revision 1

17. Move the MTS crosshead to within your working distance and then tighten down all the bolts. 18. Record the PID values of the system. Go into Station Setup, Channels-Axial. The PID values should be in the first tab. System Warm-Up 1. System gets warmed up at the beginning of each test day to get rid of transient changes in temperature, performance, etc. Warm-up consists of running a cyclic program for an hour. 2. Open the warmuplarge Procedure by selecting the open procedure button, which is the first from the left on the MPT control panel, and select the warmuplarge file 3. Create a new specimen called warmup. To create a new specimen, click on the second button from the left on the MPT control panel (it looks like a single dumbbell). No data will be written to this warmup file, but the system needs a specimen to proceed 4. Do not have any fixtures below the actuator during warmup. 5. If Enable Manual Command is selected, deselect it (On the Remote Station Controller). 6. Select the Play button on the MPT Station Control panel and wait an hour. LVDT Calibration 1. While the system is warming up, is the most time efficient to run the LVDT calibration. 2. Select the LVDT in which you want to use (Jepsen lab uses 050 MHR.) 3. Mount the LVDT rod and cylinder into an axial micrometer stage. 4. Disconnect the Teststar cable from the DTR-451 output and connect it to a volmeter. 5. Make sure the LVDT cable is not connected to the DTR-451 and turn the Zero adjustment on the DTR-451 to reach a 0.00 reading on the voltmeter 6. Reconnect the LVDT cable to the DTR-451 and move the LVDT core with the micrometer stage so that the readout is as close to 0.00 as possible on the voltmeter. 7. Zero the micrometer stage display and select the mm increment button to work in metric units. (This is the null position of the LVDT from which the plus and minus displacements are usually measured) 8. Move the LVDT core from the null position to the LVDT plus full scale position in the positive direction (ex. +1mm) using the micrometer stage adjustment. (You should set this range based on your test design.) 9. Turn the Span adjustment to reach a display reading of approximately +10 Volts on the voltmeter. (This should correspond to approximately 10,000 on the DTR-451 display) 10. You may need to adjust the Phase control next to obtain a maximum display reading of +10 Volts, otherwise skip this step. 11. Displace the LVDT core back to the precise null position with the micrometer stage adjustment. If necessary adjust the Zero control until the voltmeter display reads 0.00 Volts. 12. Displace the LVDT core to the negative full scale position (ex. -1mm) using the micrometer stage adjustment. 13. Turn the Span adjustment to reach a display reading of approximately -10 Volts on the voltmeter. (This should correspond to approximately 10,000 on the DTR-451 display) 14. Repeat Steps 8-13 as necessary until the plus and negative full range positions coordinate as closely as possible to the +\- 10 Volt display on the voltmeter. 15. If the directionality of the core appears reverse it is okay as the directionality is based on which direction you have the rod relative to the cylinder. 4pt Bending, Mouse Page 3 of 6 Revision 1

16. Move the micrometer in equal increments over the full range (positive and negative) and record the voltage output on the voltmeter at each increment. Make sure to have at least 10 data points. 17. Use excel to record and graph the data points so that you can calculate the calibration factor in mm/volts (Based on your range you should be able to determine that your calibration factor is close to the ideal calibration factor) For example, if your range was +/-1mm your ideal calibration factor is (2 mm / 20 Volts =.1 mm/volts). 18. Mount the LVDT rod and cylinder into your test set-up. 19. Reconnect DTR-451 output cable to Teststar cable Standard Testing 1. Several small dowels should be tested at the beginning of each day to confirm good data collection and point alignment. 2. Open the SpecialK procedure from the MPT panel. This program ramps down to -3 mm displacement at a rate of 0.05 mm/sec. Data is acquired at a frequency of 25 Hz and is sent to an output file (ex. Specimen.dat) in your directory. Open the Procedure Editor in the MPT panel to confirm these settings haven t changed. 3. Open the Scope window. Ensure that it is displaying load and the external LVDT. 4. Zero the load readout by selecting the Auto Offset button in the Signal Auto Offset window. 5. Move the actuator down (make sure you select Enable Manual Control on the Remote Station Controller) until the upper points overlap the lower points. 6. Create a new specimen in the MPT panel. It is important that you always create a new specimen for each specimen or else data may get combined into one output file. 7. Load the dowel between the points ensuring contact with all 4 points. 8. Slowly move the actuator up/down to get good contact. In rare cases, adjusting the brass point on the top may be necessary. Make sure that the dowel is snug in place, without trying to apply any pre-load. 9. Once the dowel is in position, disable the Manual Control on the Remote Station Controller. 10. Adjust the LVDT height until you are in the proper range. Make sure that you are starting in the correct positive or negative position ensuring that you will stay in range throughout your test. 11. Zero both the axial displacement and external LVDT readings on the Signal Auto Offset window. 12. Select the Play button on the Station Control panel. This begins the data acquisition. 13. A Start button will appear in the bottom of the MPT panel. Pressing this will begin the actuator movement. 14. Repeat for all standards. If anything appears odd in the scope window, make any adjustments as needed. Specimen Testing 1. Open the SpecialK procedure file from the MPT panel 2. Zero the load readout from Signal Auto Offset window. 3. Create a new specimen using the identification numbers of your testing samples. 4. Load your sample a. Place the anterior side of the bone in tension. b. Center the mid-diaphysis of the bone between the upper two points. c. Bring the actuator down slowly so that the bone is in contact with all 4 points. In rare cases, you may need to loosen the brass points at the top. d. The bone should be snug, without having a pre-load on it. 4pt Bending, Mouse Page 4 of 6 Revision 1

e. If having difficulties positioning the sample, remove it, place in PBS, and try again. It is critical to keep the bone hydrated, as a drying bone will have different mechanical properties. f. If there is still a lot of soft tissue on the bone, gently remove it with gauze. Rehydrate before loading the bone into the fixture. 5. Check that the external LVDT is still reading in an appropriate range as the dowels can be different diameters from the bone samples. As the LVDT has a ±1mm range, make sure the DTR- 451 box is reading -9500 at the minimum, so that as loading commences, it will move towards 10000. Place the bone sample back in PBS, and confirm that the LVDT is reading in range. 6. Secure your sample into the test set-up 7. Disable the Manual Control on the Remote Station Controller. 8. Zero both the axial displacement and external LVDT readings on the Signal Auto Offset window. 9. Select the Play button on the Station Control panel. This begins the data acquisition. 10. A Start button will appear in the bottom of the MPT panel. Pressing this will begin the actuator movement. 11. Repeat for all samples 12. Record the position of the break for every sample. If anything outside of the ordinary happens during a test (rotation of a bone sample, slipping, etc) note it next to the test. MPT Procedure Editor Any test can be set-up with the MPT Procedure Editor wizard which is found by selecting the Procedure Editor button on the MPT panel (it is the fourth from the right). The process palette (the button that looks like a painter s palette) associated with this window allows you to drag and drop different types of test commands, data acquisition inputs, events, etc. If you double click on the button associated with the procedure type it will pop up a new window for you to enter the test parameters. Make sure to always include a data limit detector that will interrupt your test command if the limit is reached. Base your limit on the type of load cell you are using. Make sure that if you start with a previously existing procedure and edit it that you save it as a NEW procedure and DO NOT save over the existing procedure unless you are the person who created it. Shutting Down 1. Before turning off the hydraulics, move the actuator up and remove the small testing frame from below the actuator. Then turn off the hydraulics. The actuator drifts after the pressure is turned off, so you want to make sure it doesn t run in to anything. 2. Close the Signal Auto Offset Window and Station Manager. 3. Answer the series of questions as follows: Do you really wish to quit? YES You have changed the station parameters.do you want to save these changes? NO Save the current test parameters? NO 4. Turn off the Teststar box and DTR-451. 5. Log-Off Certification 4pt Bending, Mouse Page 5 of 6 Revision 1

I have read and understand the above SOP. I agree to contact my Supervisor or Lab Manager if I plan to modify this procedure. Name Signature Trainer s signature Date Principal Investigator / Safety Officer Date (signature) Printed name: Designated Trainer Date (signature) Printed name: 4pt Bending, Mouse Page 6 of 6 Revision 1