HITACHI S FAST SPIN ECHO TECHNOLOGY

Similar documents
RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc

EPI. Thanks to Samantha Holdsworth!

Image quality in non-gated versus gated reconstruction of tongue motion using Magnetic Resonance Imaging:

M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia

WHY CHOOSE HITACHI? * Based on Hitachi's factory shipment records, as of end of March, Others 743 EUROPE.

Abstract. Learning Objectives 8/1/2017

Open Your Vision Make a Smart Choice

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

Light. Engineered for Performance- Esaote MRI, designed to make a difference.

Simple motion control implementation

Magnetic resonance imaging phase encoding:

The Widest Wide Bore in the Industry

S M A R T U LT R A S O U N D

Premium 1.5T MRI System

Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1

Understanding Compression Technologies for HD and Megapixel Surveillance

High Performance TFT LCD Driver ICs for Large-Size Displays

Open up to Extremity MRI

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Breast MR Imaging and Quality Control

Introduction to Data Conversion and Processing

The Diagnosis of Small Solitary Pulmonary Nodule:

Peacefully quiet. Remarkably fast.

COMPRESSION OF DICOM IMAGES BASED ON WAVELETS AND SPIHT FOR TELEMEDICINE APPLICATIONS

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

PATIENT POSITION IMAGING PARAMETERS

Peacefully quiet. Remarkably fast.

Initial Experience With Automatic Image Transmission to an Intensive Care Unit Using Picture Archiving and Communications System Technology

Chapter 7. Scanner Controls

CARESTREAM DIRECTVIEW Elite CR System

Using the MAX3656 Laser Driver to Transmit Serial Digital Video with Pathological Patterns

CARESTREAM DIRECTVIEW Elite CR System

Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper

A HIGHLY INTERACTIVE SYSTEM FOR PROCESSING LARGE VOLUMES OF ULTRASONIC TESTING DATA. H. L. Grothues, R. H. Peterson, D. R. Hamlin, K. s.

CAEN Tools for Discovery

Nuclear Associates and

Using the new psychoacoustic tonality analyses Tonality (Hearing Model) 1

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

Quartzlock Model A7-MX Close-in Phase Noise Measurement & Ultra Low Noise Allan Variance, Phase/Frequency Comparison

Operating Instructions

Spatio-temporal inaccuracies of video-based ultrasound images of the tongue

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING

Route optimization using Hungarian method combined with Dijkstra's in home health care services

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality.

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS

Coherent Receiver for L-band

Online Control System Migration of Industrial Centrifuge Project Description

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

Color Image Compression Using Colorization Based On Coding Technique

Quick reference guide

CAPTURE CAPTURE. VERSiON 1.2. Specialists in Medical. Digital Imaging Solutions

DISPLAY WEEK 2015 REVIEW AND METROLOGY ISSUE

2. AN INTROSPECTION OF THE MORPHING PROCESS

MSK Imaging Fundamentals

Development of Simple-Matrix LCD Module for Motion Picture

CHAPTER 2. Black and White Television Systems

Experiences in ACR MRI Accreditation Vendor Nuances That Every Clinical MRI Physicist Should Know

Spectrum Analyser Basics

Nuclear Instruments and Methods in Physics Research A

Artifacts in Abdominopelvic MR A Pictorial Review

Practical considerations of accelerometer noise. Endevco technical paper 324

Application Note. PICOLO Alert, Sharing the Digitizing Power

Spartan-II Development System

RECOMMENDATION ITU-R BT.1203 *

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Agilent PN Time-Capture Capabilities of the Agilent Series Vector Signal Analyzers Product Note

1ms Column Parallel Vision System and It's Application of High Speed Target Tracking

New Medical Light Source using NTT s Communication Laser Technology


*Please note that although this product has been approved in Japan, its launch in other countries has not yet been confirmed.

Achieving 10 Terabytes/Cartridge by 2011

Hitachi Europe Ltd. ISSUE : app084/1.0 APPLICATION NOTE DATE : 28/04/99

Design Trade-offs in a Code Division Multiplexing Multiping Multibeam. Echo-Sounder

What's the SPO technology?

Design of Fault Coverage Test Pattern Generator Using LFSR

Data Converters and DSPs Getting Closer to Sensors

What to consider when choosing a mammography display

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Scanning System S-2100

LCD Motion Blur Reduced Using Subgradient Projection Algorithm

Research & Development of Surface-Discharge Color Plasma Display Technologies. Tsutae Shinoda

True comfort and flexibility with the power of 3T.

ITS-I. Test station for evaluation of image quality of image intensifier tubes. Fig. 1. Photo of the ITS-I test station: a)photo, b)block diagram

Broadcast Sound Mixing System

Switching Solutions for Multi-Channel High Speed Serial Port Testing

(TCRS) Radial Choke Coil Inductors

All-in-one solutions For applications with imaging challenges, we offer a wide range of TDI solutions.

AM-OLED pixel circuits suitable for TFT array testing. Research Division Almaden - Austin - Beijing - Haifa - India - T. J. Watson - Tokyo - Zurich

United States Patent (19) Starkweather et al.

NEWS239 SILVER C M Y BK 12/17 12/20 00/00 op :tobi

MATHEMATICAL APPROACH FOR RECOVERING ENCRYPTION KEY OF STREAM CIPHER SYSTEM

Advanced Sensor Technologies

Beyond the Resolution: How to Achieve 4K Standards

New Spill Structure Analysis Tools for the VME Based Data Acquisition System ABLASS at GSI

Thyratrons. High Energy Switches. Features. Description

Transcription:

primefse TECHNOLOGY Yosuke Hitata RT Makoto Sasaki MD Kunio Esashika RT Hiroshi Gakumazawa RT HITACHI S FAST SPIN ECHO TECHNOLOGY Efficacies in Improving Image Quality & Usability Hitachi Medical Systems America Inc.

Hitachi s Prime Fast Spin Echo Technology: Efficacies in Improving Image Quality and Usability Yosuke Hitata RT 1 Makoto Sasaki MD 2 Kunio Esashika RT 1 Hiroshi Gakumazawa RT 1 1 Center for Radiological Sciences, Iwate Medical University School of Medicine 2 Advanced Medical Research Center, Iwate Medical University School of Medicine We evaluated the efficacies of the prime fast spin echo (FSE) technique, which enables the optimization of the echo shift scheme and direct assignment of the effective echo time, as compared with those of the conventional FSE technique. We could easily obtain images with arbitrary contrast by using the prime FSE technique. In addition, in this technique, truncation artifacts were diminished, and blurring artifacts were decreased without prolongation of the acquisition time. Prime FSE is considered to be a promising technique for improving the image quality of FSE imaging. Key Words: MRI, Prime FSE, k-space, Echo Shift, Truncation Artifact 1. Introduction Recent modern pulse sequences on MRI including a fast spin echo (FSE) technique contribute to improving the quality of diagnostic imaging. The FSE technique is widely used for a majority of routine imaging; however, a scheme for filling echo signals in the k-space varies across MRI systems, and this issue occasionally causes considerable artifacts or leads to complications in the parameter settings. This paper reports the features and efficacies of prime FSE, a novel technique implemented in the 1.5T Echelon as an advanced FSE sequence, by focusing on the area of the central nervous system. 2. Description of the Echelon in our institute The Echelon was installed as the sixth MRI unit in our institute in March 2007. Due to its compact design, after the removal of a superconductive 0.5T system of another manufacturer, the system could be installed in a small space without any expansion of the scanning room. In the Echelon, all the sequences and capabilities commonly used at 1.5T are implemented, including rapid acquisition through a parallel imaging design (RAPID) as a parallel imaging technique, radial acquisition regime (RADAR) as a hybrid radial scan technique, balanced steady-state acquisition with rewound gradient echo (BASG) as a true fast imaging with steady-state procession (FISP) technique and MR Spectroscopy. Of these, the clinical MR spectroscopy (MRS) package is one of the unique advantages of the Echelon. Workflow of the MRS is greatly simplified in the Echelon as compared with that in systems produced by other manufacturers. Positioning, pre-scanning, data acquisition, data analysis, and filming can be performed on a console with simple procedures involving minimal clicks. On the other hand, the conventional FSE technique in Echelon, similar to MR imagers of other manufacture, has some limitations. Optimization of the scan parameters to minimize FSE-related artifacts while maintaining image contrast and acquisition time is often a laborious task. In addition, the quality of dual-contrast FSE was not optimal; therefore, T2-weighted images (T2WIs) and proton density-weighted images (PDWIs) had to be obtained separately, consequently prolonging

the examination time. In order to overcome these limitations, Hitachi Medical Corporation recently released prime FSE, a sequence that enables automated optimization of the parameter settings and image quality by improving the scheme to fill the k-space with signals. From April 2007, we have implemented this technique in our institute. 3. Features of prime FSE 3.1 Improvement in the echo shift scheme In the FSE technique, the effective echo time (TE) is determined by the echo that is allocated at the low-frequency area of the k-space 1,2. In the conventional FSE technique, first, median, and last echoes can be assigned at the center of the k-space by using centric, sequential, and reverse centric orders, respectively. Other echoes also can be set as the effective TE by using an echo shift technique that reorders allocations of multiple echo signals in k-space. However, echo shift can cause significant truncation artifacts due to increased discreteness in the signal intensity at the boundary of the reordered neighboring echoes (Fig. 1A). On the other hand, in prime FSE, echoes are scrolled while maintaining relationships with the adjacent echoes of the centric order. Hence, difference in the signal intensity between the neighboring echoes is maintained at a minimum level regardless of the amount of echo shift, in order to maintain truncation artifacts at a minimum level 1,3 (Fig. 1B). Further, it is possible to arrange echo shifts independently in the PDWIs and T2WIs of dualcontrast FSE imaging. 3.2 Improvement in parameter settings In the conventional FSE technique on this system, direct assignment of the effective TE is not possible. Hence, combinations of the echo number, echo interval, k-space order, and echo shift are manually adjusted to obtain the intended effective TE. This procedure can be complicated when parameters must change, sometimes causing combinations of parameters that do not optimize image quality. In prime FSE, the effective TE values are directly entered, just as in conventional spin-echo (SE). Echo shift is automatically executed to allocate the effective TE as one of the multiples of the echo interval proximate to the entered values, while maintaining values of the echo interval, echo number, and readout bandwidth. For example, with the following conditions of the echo interval of 12 ms, echo number of 10, and bandwidth of 32 khz, effective TE can be set at 12, 24, 36, 48, 60, 72, 84, 96, 108, or 120 ms maintaining image quality, signal-tonoise ratio, and acquisition time (Fig. 2). Figure 1. Echo shift in conventional FSE and prime FSE A: Conventional FSE, B: Prime FSE (both sequences are in the centric order with an echo number of 5). Echo shift by conventional FSE produces large differences in signal intensity between neighboring echoes (A, arrows) that generate significant truncation artifacts. Echo shift by prime FSE maintains a minimum difference in signal intensity to suppress artifacts (B). A C E Figure 2. Assignment of effective TEs in prime FSE A:Effective TE, 12ms; B:36 ms; C :60 ms ; D :84 ms ; E :96 m s ; F:108 ms (TR, 2800 ms; echo interval, 12 ms; echo number, 11; bandwidth, 32 khz; FOV, 220 mm; matrix size, 256 256; and number of excitations, 2). Prime FSE technique can assign arbitrary effective TEs as multiples of the echo interval while keeping other parameters, including the echo interval, echo number, and bandwidth, constant and maintaining FSE-related artifacts at a minimal level. B D F

4. Improved image quality by prime FSE Prime FSE enables us to obtain excellent PDWIs and T2WIs simultaneously with relatively short acquisition times by using a dual-contrast mode. (Fig. 3A, D). In the conventional FSE technique, PDWIs and T2WIs should be scanned separately because echo assignments in the first and last halves cannot be independently controlled in the conventional dual-echo mode. PDWIs equivalent to images in prime FSE can be obtained by setting the echo number to half of that in dual-echo prime FSE, but the scan time is doubled (Fig. 3B). On the other hand, images with the same echo number can avoid prolongation of the scan time, but they may worsen artifacts such as blurring and pseudo-edge enhancement (Fig. 3C). Blurring and/or truncation artifacts in T2WIs are remarkable both in the centric and sequential orders because of the use of the echo shift and/or the prolongation of the echo interval (Fig. 3E, 3F). In addition, prime FSE enables us to obtain high-quality short inversion-time inversion recovery (STIR) images that have high contrast resolution between gray and white matters. Truncation artifacts were lower in STIR with prime FSE as compared with conventional FSE (Fig. 4). A C Figure 4. Short inversion-time inversion recovery (STIR) images in prime FSE and conventional FSE A, C: Prime FSE; B, D: conventional FSE (TR, 4000 ms; effective TE, 36 ms; inversion time (TI), 140 ms; echo interval, 12 ms; echo number, 8; bandwidth, 35 khz; FOV, 200 mm; matrix, 256 256; number of excitations, 2). Truncation artifacts that are prominent in STIR images by conventional FSE (D, arrows) are substantially reduced in those by prime FSE, and the internal structures of the hippocampus are readily visualized (C). B D A B C D E F Figure 3. Dual-contrast images by prime FSE and conventional FSE A-C: proton density-weighted image (PDWI); D-F: T2-weighted image (T2WI); A, D: dual-contrast prime FSE (echo number, 10 (5 + 5)); B-C and E-F: conventional FSE (B: centric order with an echo number of 5, C: centric order with an echo number of 10, E: centric order with an echo number of 9, and F: sequential order with an echo number of 10); (TR, 2800 ms; echo interval, 12 ms; FOV, 220 mm; matrix; 256 256; number of excitation, 2). Satisfactory PDWIs and T2WIs are obtained within 5 min by using the dual-contrast prime FSE that can assign arbitrary effective TEs with independent optimization of the reordering in 2 kinds of k-spaces (A, D). In conventional FSE images, prolongation of the acquisition time (B) and/or remarkable artifacts such as pseudoedge enhancement (C), blurring, and truncation artifacts (E, F) are unavoidable.

5. Conclusion Prime FSE is an important new scanning feature that permits easy operation, increase in latitude of selecting effective TEs, optimization of numerous sequence parameters, shortening of the acquisition time, and suppression of the blurring and truncation artifacts by a new algorithm for k-space ordering. Therefore, it is considered to be a promising technique for improving image quality in MR imaging of the central nervous system and other regions. *Echelon is a registered trademark of Hitachi Medical Corporation in Japan. Reference 1) Listerud J, Einstein S, Outwater E, et al: First principles of fast spin echo. Magn Reson Quarterly 1992; 8:199-244 2) Costable RT, Anderson AW, Zhong J, et al: Factors influencing contrast in fast spin-echo imaging. Magn Reson Imaging 1992; 10:497-511 3) Melki PS, Mulkern RV, Panych LP, et al: Comparing the FAISE method with conventional dual-echo sequences. J Magn Reson Imaging 1991; 1:319-326

Hitachi Medical Systems America Inc. 1959 Summit Commerce Park Twinsburg, Ohio 44087 Tel: 330.425.1313 800.800.3106 FFax: 330.963.0749 www.hitachimed.com Hitachi Medical Corporation Akihabara UDX 4-14-1 Soto-Kanda Chiyoda-ku, Tokyo 101-0021, JAPAN Tel: 81.3.3526.8407 Fax: 81.3.3526.8418 2008 Hitachi Medical Systems America Inc. 1008/DM#56091 v1 All rights reserved.