How To Demonstrate Improved ACLR Dynamic Range With FSU and Noise Correction

Similar documents
Tests on 3G-Base Stations to TS with R&S Spectrum Analyzers and SMIQ Vector Signal Generator

Determing Amplifier Nonlinearities by using Real-World Signals European Microwave Week 2003

Generating 3GPP Multi Carrier Signals for Amplifier Tests with R&S SMIQ03HD and WinIQSIM

Improving the accuracy of EMI emissions testing. James Young Rohde & Schwarz

RF Level Test System +20 dbm to 130 dbm

Level Accuracy and Electronic Level Settings of SMIQ

Tests on 3G-Base Stations to TS with FSIQ and SMIQ

Out of Band Spurious Measurement for Bluetooth Modules

NPR - Noise Power Ratio Signal Generation and Measurement

Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper

R&S FSV-K73 3G FDD UE (UL) Measurements incl. HSUPA Specifications

Iterative Direct DPD White Paper

Calibrating the CMD Output Level for BER

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X

R&S FSV-K40 Phase Noise Measurement Application Specifications

ANSI/SCTE 40 Conformance Testing Using the R&S SFU, R&S SFE and R&S SFE100

Measurements on GSM Base Stations According to Rec

DVB-T Bursted Noise Signal Generation

W-CDMA Base Station Transmitter Tests According to TS Rel. 10

Agilent CSA Spectrum Analyzer N1996A

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Application Note DT-AN DTU-315 Verification of Specifications

RF Signal Capture & Playback Simple Operation Guide

Converting MediaFLO Waveform Files to R&S SFU / SFE / SFE100 ARB Format Using IQWIZARD/WinIQSIM TM for R&S SFx-K35 ARB

Transmitter Measurements on Bluetooth TM Modules

R&S FS-Z60/75/90/110 Harmonic Mixers for the R&S FSP/FSU/ FSQ/FSUP/FSV

Spectrum Analyzer 1.6 GHz 3 GHz HMS-X

Digital Pre-Distortion Techniques for RF Power Amplifiers

TEST REPORT FROM RFI GLOBAL SERVICES LTD

R&S SGMA Product Family Compact fast reliable

R&S FPS-K18 Amplifier Measurements Specifications


1xEV-DO Test Solutions

TEST REPORT No

GENERAL PURPOSE Signal generators. R&S SMBV100A vector signal generator allrounder and specialist at the same time

cdma2000 Base Station Test with R&S Equipment

R3267/3273 Spectrum Analyzers

Transmitter Measurements on Bluetooth TM Modules with FSP

EUTRA/LTE Downlink Specifications

The high-end network analyzers from Rohde & Schwarz now include an option for pulse profile measurements plus, the new R&S ZVA 40 covers the

Agilent PSA Series Spectrum Analyzers Data Sheet

Be ahead in 5G. Turn visions into reality.

AMIQ-K2 Program for Transferring Various-Format I/Q Data to AMIQ. Products: AMIQ, SMIQ

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

R&S FSW Signal and Spectrum Analyzer Specifications

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal. Application Note

R&S FSW-K76/-K77 3GPP TD-SCDMA BS/UE Measurement Applications Specifications

RF Test Report: Airspan ib440 to 47CFR SC_TR_150_B

APM CALIBRATION PROCEDURE Rev. A June 3, 2015

Calibrating attenuators using the 9640A RF Reference

LTE-A Base Station Transmitter Tests According to TS Rel. 12. Application Note. Products: R&S SMW200A R&S FSW R&S FSV R&S SMBV100A R&S FSVA

Speeding up Spectrum Analyzer Measurements Application Note

R&S FSMR Measuring Receiver Specifications

R&S FSV-K76 TD-SCDMA BS (DL) Measurements Specifications

NPR Noise Power Ratio Signal Generation and Measurement Application Note

Seamless Level Setting on the R&S SFU

R&S FSU Spectrum Analyzer High-end spectrum analysis

R&S FPS Signal and Spectrum Analyzer Specifications

Tests on CDMA2000 1xEV-DO Access Terminals in Accordance to IS-866-A

Tests On 3GPP WCDMA FDD Node Bs in Accordance with Standard TS

ESU EMI Test Receiver Specifications

R&S FSV-K8 Bluetooth /EDR Measurement Application Specifications

TEST RESULTS FOR A DIGITAL PREDISTORTION SYSTEM FOR 3G CELLULAR TELEPHONY

R&S FPC1000 Spectrum Analyzer Specifications

News from Rohde&Schwarz Number 195 (2008/I)

R&S FSU Spectrum Analyzer Specifications

SMS3000X Series Spectrum Analyzer

Broadcast Television Measurements

<WCDMA Inner Loop Power Control tests, test pattern E&F,G&H using ARB files on the R&S CMW 500>

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications

EUTRA/LTE Measurement Application Specifications

GSM Mobile Tests under Conditions of Fading

R&S Spectrum Rider FPH Handheld spectrum analyzer

Effective Test Procedures for Installing and Maintaining RF Transmitter Sites

LTE-A Base Station Transmitter Tests According to TS Rel. 13. Application Note. Products: R&S SMW200A R&S FSW R&S FSV R&S SMBV100A R&S FSVA

Coherence Measurement between two Signals regarding Timing, Phase and Gain Application Note

Production Tests on ICO User Terminals with SMIQ and FSE/FSIQ

Versatile RF Fading Simulator With R&S FSQ/FSG/FSV and R&S SMU Application Note

PXI UMTS DL Measurement Suite Data Sheet

IMPORTANT NOTICE. Company name - STMicroelectronics NV is replaced with ST-NXP Wireless.

R&S FPS Signal and Spectrum Analyzer Specifications

CONTENTS 1. GENERAL INFORMATION INTRODUCTION PRODUCT INFORMATION DESCRIPTION OF TESTS CHANNEL BANDWIDTH...

APPLICATIONS S240 Real-Time Spectrum Analysis application Numerous third-party analysis applications

Correlated Receiver Diversity Simulations with R&S SFU

R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz

Advanced Test Equipment Rentals ATEC (2832)

R&S FSW-K160RE 160 MHz Real-Time Measurement Application Specifications

RS Pro SPECTRUM ANALYZER SSA3000X SERIES

Pre-5G-NR Signal Generation and Analysis Application Note

Research Results in Mixed Signal IC Design

Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV

Spectrum Analyzers. 2399A 9 khz to 3 GHz Spectrum Analyzer.

WCDMA Base Station Performance Tests according to TS25.141

R&S FSQ Signal Analyzer Specifications

History and release notes for Spectrum Analyzer R&S FSU

WLAN Tests According to Standard a/b/g

DOCSIS 3.1 Full channel loading Maximizing data throughput

R&S FSMR Measuring Receiver All-in-one calibration of signal generators and attenuators

Transcription:

Product: Spectrum Analyzer FSU How To Demonstrate Improved ACLR Dynamic Range With FSU and Noise Correction Application Note This application note provides information about the ACLR measurement with noise correction. The basic requirements and the limiting factors of a spectrum analyzer are explained. Dynamic range improvements by means of a noise correction is presented. Measurement examples show the practical realisation of ACLR measurements. Subject to change Kay-Uwe Sander/1ESP 05/2002-1EF49_0E

Contents 1 Overview...2 2 Dynamic Range Limitations...2 3 Signal Processing - Noise Correction...3 4 Setup Procedures...4 Setting of the the input attenuator...5 Setting of the the reference level...5 5 Performing the Measurements...5 SMIQ setup:...6 FSU setup (IBW method):...6 Activate Noise Correction:...7 FSU setup (Fast ACP method):...7 6 Literature...8 7 Ordering information...8 1 Overview ACLR measurement is one of the most demanding measurement on base stations tranmitter regardless of the technology used. It requires an extremely high dynamic range of the spectrum analyzer. This application note explains the parameters of the spectrum analyzer defining the ACLR dynamic range. Next the use of smart signal processing routines to improve the ACLR dynamic range are explained. Derived from these parameters recommendations for optimum setup of the analyzer are given. The recommended setups are supported by practical measurements on a WCDMA signal in the last section. 2 Dynamic Range Limitations The inherent dynamic available for ACLR measurement is an important factor. It must be well below the limits for the device under test. This assures that the test result is minimally affected by the ACLR floor of the spectrum analyzer. The inherent dynamic range of the FSU when measuring power in the adjacent channels is determined by three factors: The internal noise floor of the spectrum analyzer The phase noise The spectral re-growth due to 3 rd order intermodulation of the analog signal path in the 5-MHz offset channel. From these contributions the optimum level settings can be determined taking the internal structure of the spectrum analyzer into account. The following figure shows the contributions of the three limiting factors to the inherent ACLR dynamic range of the FSU spectrum analyzer: 1EF49_0E 2 Rohde & Schwarz

ACLR /dbc -50-55 -65-70 total ACLR -75 thermal noise spectral regrowth -85 phase noise -20-15 -10-5 0 Mixer level /dbm Fig. 1 Inherent dynamic range for 3 GPP ACLR in the 5-MHz offset channels dependent on the level at the input mixer. 3 Signal Processing - Noise Correction The idea of doing additional signal processing is to compensate for the internal sources of error. There are at least three major error sources: The spectral re-growth due to 3 rd order intermodulation of the analog signal path.this signal component is hard to compensate due to ist complex structure. The phase noise could be measured, but an additional test signal is neccessary to measure the spectrum analyzers contribution to the phase noise at the measurement frequency. The internal noise floor of the spectrum analyzer can easily be measured. Noise correction is a mathematical prodedure to remove the spectrum analyzer inherent noise from the measured signal. The procedure to compensate for the inherent spectrum analyzer noise works in two steps: In a first step, the input signal is removed from the spectrum analyzer input and the input is terminated. This is done within the FSU by means of a switch within the RF Attenuator. In this setup, the inherent noise power at the actual frequencies is measured. In the second step, the input is switched on again and the input signal is measured. The signal processing calculates the input power by subtracting the inherent noise power from the measured values. With this procedure, the inherent noise power of the spectrum analyzer is set to a value very close to Zero. The subtraction of the inherent noise from the input signal might generate negative power values, which would cause negative infinite numbers on the typical used LOG scaling. To avoid this, the FSU does not subtract the full amount of inherent noise power from the measured values. This leads to a partly compensated inherent noise display with no infinite numbers. As a result of this procedure, the noise contribution to the available ACLR dynamic range chart is reduced by 15...20 db. 1EF49_0E 3 Rohde & Schwarz

Fig. 2 ACLR dynamic range with reduced noise due to noise correction It can be derived from the above chart, that the optimum mixer level has moved to a lower value due to the mathematical reduction of the noise power. The optimum mixer level for the FSU spectrum analyzer with active noise correction is about -18 to -13 dbm. Very important for the successful noise correction is a stable power measurement. The FSU support this requirement: The accurate and stable power mesurement is possible by using the RMS detecor. The stability of the power measurement is controlled with the sweep time, since this has a direct influence on the amount of samples available to calculate the average value. A long sweep time will give more stable results and thus leads to better compensation of the inherent noise power. The linearity of the receiver has a great impact on the accuracy of the noise power measurement. The FSU is using digital filtering for RBW settings up to 100kHz, thus providing an excellent linearity. This is important because the inherent noise power measurement must be performed with the same instrument setting which is used for the input signal measurement, but only internal noise is measured. Since the noise compansation measures the inherent noise power of the spectrum analyzer with the actual setting, any change of the spectrum analyzer setting must be avoided (especially settings of Frequency, RF ATT, RBW,...). Otherwise it requires a re-measurement of the inherent noise by reactivating the Noise Compansation. 4 Setup Procedures Two parameters must be considered when setting up a spectrum analyzer to achieve the best dynamic range: The RF attenuation in order not to generate spectral re-growth in the internal mixer stages. The reference level in order not to overdrive the IF path following the internal bandpass filters and the A/D Converter. 1EF49_0E 4 Rohde & Schwarz

5 Performing the Measurements Improved Dynamic Range with Noise Correction Setting of the the input attenuator The input attenuator has to be set so that the signal path is loaded with optimum mixer level (i.e. level at RF input minus RF attenuation). The optimum mixer level for the FSU is about -13 dbm to -8 dbm dependent on the crest factor of the W-CDMA signal. At -13 to -8 dbm level at the input mixer the FSU provides an ACLR dynamic range of about 77 dbc in the 5-MHz offset channel ( see also above figure). Setting of the the reference level The reference level can be set independently from the input attenuator in a certain range. The reference level sets the level at the A/D Converter input. It can handle signals up to 2 db above reference level w/o clipping. For greater signal levels it begins to clip the signal and generates distortion. For 3 GPP W-CDMA signals the reference level (and dynamic range) is optimum when the A/D Converter is driven to full scale to minimize the influence of wideband noise after the IF filters The RF attenuation has to be kept fixed, however. The setting must be done in a way that the overload message of the A/D Converter is just not appearing in the display (Trace about 10..15 db below reference level). To free the user from optimum mixer level calculations the FSU provides automatic routines to establish the optimum RF attenuation and reference level. With the signal from the device under test applied to the FSU the user simply activates the soft key ADJUST REF LEVEL in the ACPR menu of the FSU to optimize the dynamic range of the instrument. This section shows how to set up the instruments for ACLR measurement. For the measurement examples the R&S signal generator SMIQ is used. The output signal is a W-CDMA signal according to TS 25.141, test model 1. In order to measure the limits of the FSU the SMIQ output signal is filtered by a bandpass filter with a center frequency of 380 MHz, a bandwidth of ~ 4 MHz and ~ 10 db insertion loss. Figure 3 Filter for ACLR demos SMIQ Bandpass Filter FSU f c = 380 MHz B ~ 4 MHz Figure 4 Test setup for the measurement examples 1EF49_0E 5 Rohde & Schwarz

SMIQ setup: Improved Dynamic Range with Noise Correction The SMIQ is set to generate the 3GPP ACLR test signal. The following settings have to be performed: Preset the SMIQ Set the output frequency to 380 MHz Set the level to 0 dbm Choose Digital Standards - WCDMA/3GPP Choose TEST MODELS... - TEST1_64 Choose FILTER... - set FILTER MODE to LOW_ACP Go to STATE and switch the WCDMA modulation ON Go to VECTOR DEMOD - IQ FILTER and set to 2.5 MHz FSU setup (IBW method): The FSU is set to measure the ACLR. The following settings have to be performed: Preset the FSU Set the center frequency to 380 MHz Set the reference level to 0 dbm Presse MEAS - CHAN PWR / ACP - CP/ACP STANDARD, Choose WCDMA 3GPP FDD Change the sweep time to 1 s Adjust SMIQ level to get a channel power reading of about -10dBm Press ADJUST REF LEVEL Ref -12.4 dbm *Att 0 db * RBW 30 khz * VBW 300 khz *SWT 1 s Marker 1 [T1 ] -30.31 dbm 380.000000000 MHz 1 RM* CLRWR -20-30 -40-50 cl2-70 -100-110 Center 380 MHz Tx Channel cl2 2.55 MHz/ Span 25.5 MHz PRN W-CDMA 3GPP FWD Bandwidth 3.84 MHz Power -10.15 dbm Adjacent Channel Bandwidth 3.84 MHz Lower -71.22 db Spacing 5 MHz Upper -71.15 db Alternate Channel Bandwidth 3.84 MHz Lower -71.68 db Spacing 10 MHz Upper -71.74 db 1 A Figure 5 Measurement result with Auto level setting. The FSU is now set to optimized level settings and shows the ACP measurement (Attenuator is set to 0 db, reference level is set to ~ -13 dbm). The level setting is done in a way that no overload is displayed under any circumstances. A lower reference level setting will increase the dynamic range furthermore: change the reference level in 1 db steps while observing the channel power, the adjacent channel power and the overload message. A reference level setting of -17 dbm can be reached, while the ACP reading is gaining to about -74dBc. A further decrease will gain more dynamic range, but the overload message will appear. The reference level can be decreased furthermore until the channel power is starting to change. This is a sign of an overloaded IF path. Set the reference level to the lowest possible value without overload. 1EF49_0E 6 Rohde & Schwarz

1 RM * CLRWR Ref -17 dbm -20-30 -40-50 cl2-70 -100-110 Center 380 MHz Tx Channel cl2 * Att 0 db * RBW 30 khz * VBW 300 khz *SWT 1 s 2.55 MHz/ Span 25.5 MHz PRN W-CDMA 3GPP FWD Bandwidth 3.84 MHz Power -10.09 dbm Adjacent Channel Bandwidth 3.84 MHz Lower -73.50 db Spacing 5 MHz Upper -73.32 db Alternate Channel Bandwidth 3.84 MHz Lower -74.15 db Spacing 10 MHz Upper -74.28 db 1 Marker 1 [T1 ] -30.86 dbm 380.000000000 MHz A Figure 6 Measurement result with reduced reference level Activate Noise Correction: Change the SWEEP TIME to 10 s Presse MEAS - CHAN PWR / ACP - NOISE CORR ON The FSU will automatically set the RF Attenuator to 5 db more than the actual value to cover the reduced optimum mixer level. 1 RM* CLRWR Ref -17 dbm -20-30 -40-50 -70 cl2-100 -110 Center 380 MHz Tx Channel cl2 *Att 5 db 2.55 MHz/ * RBW 30 khz * VBW 300 khz * SWT 10 s W-CDMA 3GPP FWD Span 25.5 MHz Bandwidth 3.84 MHz Power -10.09 dbm 1 Marker 1 [T1 ] -30.39 db 380.000000000 MHz A SGL NOR PRN Adjacent Channel Bandwidth 3.84 MHz Lower -84.53 db Spacing 5 MHz Upper -84.22 db Alternate Channel Bandwidth 3.84 MHz Lower -86.41 db Spacing 10 MHz Upper -85.70 db Figure 7 Measurement result with noise coerrection active FSU setup (Fast ACP method): The disadvantage of the long sweep time in the IBW mode can be overcome in the Fast ACP mode. In this case the measurement of ACLR is much faster due to the wider RBW used in the power measurement. The following settings have to be performed (starting from the last measurement): 1EF49_0E 7 Rohde & Schwarz

Preset the FSU Change the sweep time to 500 ms Presse MEAS - CHAN PWR / ACP - FAST ACP ON Press ADJUST REF LEVEL Presse MEAS - CHAN PWR / ACP - NOISE CORR ON In this case the level setting is optimized 1 RM* CLRWR Ref 1 dbm 0-10 -20-30 -40-50 -70 Center 380 MHz Tx Channel 1 *Att 5 db 50 ms/ SWT 500 ms W-CDMA 3GPP REV Marker 1 [T1 ] -103.19 db 100.000000 ms Bandwidth 3.84 MHz Power -9.97 dbm A SGL NOR PRN Adjacent Channel Bandwidth 3.84 MHz Lower -77.24 db Spacing 5 MHz Upper -74.41 db Alternate Channel Bandwidth 3.84 MHz Lower -86.58 db Spacing 10 MHz Upper -86.47 db Figure 8 Measurement result in FAST ACP with noise coerrection 6 Literature [1] Application Note 1EF40_E, Measurement of Adjacent Channel Power on Wideband CDMA Signals [2] Application Note 1EF41_E, Measurement of Adjacent Channel Leakage Power on 3GPP W-CDMA Signals with the FSP [3] Application Note 1EF45_E, Spurious Emission Measurement on 3GPP Base Station Transmitters 7 Ordering information Type of instrument Order number FSU3 20 Hz to 3.6 GHz 1129.9003.03 FSU8 20 Hz to 8 GHz 1129.9003.08 FSU26 20 Hz to 26.5 GHz 1129.9003.26 SMIQ03B 300 khz to 3.3 GHz 1125.5555.03 SMIQ03HD 300 khz to 3.3 GHz 1125.5555.33 SMIQB57 High ACLR for WCDMA 3GPP (2110 to 2170 MHz) 1105.1831.02 ROHDE & SCHWARZ GmbH & Co. KG. Mühldorfstraße 15. D-81671 München. P.O.B 80 14 69. D-81614 München. Telephone +49 89 4129-0. Fax +49 89 4129-13777. Internet: http://www.rohde-schwarz.com This application note and the supplied programs may only be used subject to the conditions of use set forth in the download area of the Rohde & Schwarz website. 1EF49_0E 8 Rohde & Schwarz