Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Similar documents
Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba

Scope: All CT staff technologist

2012 Computed Tomography

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

2017 Computed Tomography

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

FFDM Quality Control in Canada - a Vendor Neutral Approach

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

Guidelines for Assuring Softcopy Image Quality

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018)

Imaging solutions for any size patient and any size budget.

Performance Evaluation of Industrial Computed Radiography Image Display System

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Guidance for Quality Assurance of PACS Diagnostic Display Devices

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Understanding CT image quality

Ride the Lightning. A Review of Radiographic Physics

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

Essentials of Digital Imaging

JEFFERSON COLLEGE. Image Intensification & Equipment

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

Initial Experience With Automatic Image Transmission to an Intensive Care Unit Using Picture Archiving and Communications System Technology

Breast Imaging Essentials

Collimation and Light Field Brightness

UArm. Series. Analog or Digital U-Arm System

Request for Proposals

The Versatile and Powerful ACLxy. ACLxy

Creating room. for your visions

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management

SEAMLESS INTEGRATION MULTIPLE SOLUTIONS ONE PRODUCT ONE COMPANY

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Rev. 01 February 2014 X- RAY SAFETY MANUAL

The Digital Debate. Competing Technologies. Up and Coming 1/5

Monitor QA Management i model

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

World First Slim Cassette Type Digital Mammo. Upgrade Solution

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.

Coronis 5MP Mammo. The standard of care for digital mammography

Understanding PQR, DMOS, and PSNR Measurements

2

The Diagnosis of Small Solitary Pulmonary Nodule:

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2

Quality Assurance (QA) Guidelines for Medical Imaging Display Systems

Australian/New Zealand Standard

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Radiological Imaging. Software Version 3.6

Intra-oral X-ray unit with high frequency DC generator. Product Data

In recent years, CT technology has undergone profound

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

+ Human method is pattern recognition based upon multiple exposure to known samples.

CLINICALLY PROVEN DESIGNS

Nio. Industry-standard diagnostic display systems

Mammo Tomosynthesis 5MP

Revised Textbook Edition Digital Radiography & PACS

Corrigendum-II. Up to 30 th October 2018 till 17:00 Hrs. 31 st October 2018 till 17:00 Hrs. 2 nd November 2018 till 14:00 Hrs.

-Technical Specifications-

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D.

Dental Line. High frequency intraoral x-ray unit. radiology ahead

The Amadeo M systems wireless or wired, also suitable for field work

VENDOR MANUAL VERSION 2.0. SECTION 8 Quality Assurance Requirements

Performance characteristics and quality assurance considerations for displays used in interventional radiology and cardiac catheterization facilities

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

MR Accreditation Programs - E. Jackson

MQSA Quality Control Manual

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

DRAFT. Proposal to modify International Standard IEC

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date:

CytoFLEX Flow Cytometer Quick Start Guide

Imaging systems and Telemedicine. Dr. Adrian Mondry

DICOM Conformance Statement

Technologist Quality Control Procedures

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life.

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

This talk covers currently available display technology.

APP USE USER MANUAL 2017 VERSION BASED ON WAVE TRACKING TECHNIQUE

BCA 9R. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

BCA 9S. ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400

Enhanced Diagnostics through Ultrasound Imaging

RADIOGRAPHIC PERFORMANCE OF CYGNUS 1 AND THE FEBETRON 705

Image Contrast Enhancement (ICE) The Defining Feature. Author: J Schell, Product Manager DRS Technologies, Network and Imaging Systems Group

Biograph Vision See a whole new world of precision

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

Transcription:

Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks system performance over time, and reveals performance trends. This document outlines the tests that are typically part of a QC program for digital radiography equipment, including portables with a dedicated detector. Contact Imaging Physics for assistance in setting up your program. What are the benefits of a QC program? Performance degradation can be identified leading to preventative action. Patients benefit when equipment performance is maintained at acceptable levels. A QC program is an important element in achieving accreditation. What are the components of a QC program? The QC program is set up by the facility under the guidance of a medical physicist certified by the Canadian College of Physicists in Medicine. The program consists of acceptance testing, on-going quality control, and periodic review of QC data and outcomes. Typically, the routine QC activities are carried out by a technologist while in-depth checks are performed by, or under the guidance of, a medical physicist. A typical QC program includes the following: Acceptance Testing Acceptance testing must be performed by or under the oversight of a medical physicist when a system is installed, relocated (where relevant) or undergoes significant upgrades or maintenance. Acceptance testing verifies vendor specifications and establishes performance baselines. It is the facility s responsibility to make arrangements for acceptance testing by a medical physicist. Daily 1. Inspect system operation and verify operational status. Follow manufacturer s recommendations for equipment warm-up. Check meters and audible and visual indicators for proper function. Inspect equipment for mechanical soundness and smooth motion. 2. For the image acquisition system, inspect operation and verify operational status. 3. Check status of network queue and send images as necessary to PACS. Monthly 1. For CR systems, erase all plates in the inventory. Scan a randomly selected unexposed cassette of each size, and check for obvious non-uniformities and artefacts. 2. Perform an image quality test with a phantom. An image quality test based on a phantom manufactured by CancerCare Manitoba is described below. If you wish to use a manufacturer s QC test, contact Imaging Physics to confirm that the test is appropriate.

3. Perform retake and reject analysis. A reasonable target reject rate for general radiography is 8%, with thresholds for investigation set at 10% and 5%. A low reject rate can mean that recording or analysing rejects is not being done correctly. For paediatrics, the reject rate should be around 5%, with thresholds for investigation at 7% and 3% 1. If these requirements are not met, causes should be identified and documented, along with any corrective action taken, if applicable. In consultation with medical physics, the target reject rates can be modified to fit clinical practice. For sites that do less than 250 patients per month, the reject analysis can be done quarterly 2. 4. Verify performance of modality displays qualitatively by displaying and evaluating an image of the SMPTE pattern or equivalent. Verify visibility of the 5% contrast patches and the absence of distortions or artefacts. Refer to the modality display QC instructions available on the Imaging Physics website. Quarterly 1. For CR systems, clean all imaging plates if recommended by the manufacturer or deemed necessary in consultation with the physicist. 2. Check collimator operation to ensure smooth collimator blade motion and tracking. 3. Check door interlocks if applicable. Annually 1. Annual testing by or under the oversight of a medical physicist to evaluate performance against vendor specifications and baseline levels established at acceptance. 2. Annual equipment QC review by a medical physicist. With regard to the suggested test frequencies, daily refers to each day the equipment is used. QC data should be recorded in a manner that allows monitoring of trends in performance levels. It is recommended that QC data trends be reviewed at least semi-annually. Note that manufacturer-supplied QA software that only provides a pass or fail result is not adequate. Where to go for help? Imaging Physics at CancerCare Manitoba provides physics testing for diagnostic imaging systems, and can assist diagnostic imaging departments in setting up QC programs, training staff to perform QC tests, and identifying appropriate QC tools. You can contact Imaging Physics at Imaging.Physics@cancercare.mb.ca or by calling 204-787-4145. The radiography specialists are Dr. Idris Elbakri (204-787-2856) and Dr. Harry Ingleby (204-787-2126). 1 Ongoing quality control in digital radiography: Report of AAPM Imaging Physics Committee Task Group 151. American Association of Physicists in Medicine Report No. 151. 2015. 2 Quality control in diagnostic radiology: Report of Task Group #12 Diagnostic X-ray Imaging Committee. American Association of Physicists in Medicine Report No. 74. 2002. Page 2 of 9

CCMB Monthly Image Quality Test A phantom consisting of two 10 cm thick blocks of acrylic can be obtained from Medical Physics at cost. One of the blocks has a pattern of low contrast discs as shown in the figure below. This phantom can be used to qualitatively and quantitatively assess image quality. Instructions for Phantom Image Acquisition Stack the two blocks in the centre of the field of view such that the dark spot in the corner of the top block is aligned with the dark spot on the lower block. The block with the discs should be on top with the discs facing the x-ray tube. Collimate the beam to just inside the edges of the phantom. If using CR, it is recommended that the CR cassette be erased and that the same cassette be used every time. Acquire an image using a common clinical protocol (abdomen is appropriate) with AEC. Make note of the AEC configuration, filtration, SID and grid status and use identical conditions every time the test is repeated. Record the kvp, mas, exposure index and system-reported dose. Instructions for Image Scoring Inspect the image visually using the modality display. Use the same display every time. It is best if the same person performs the image scoring every time. Optimize the window and level settings (referred to as brightness and contrast on some systems) for best visibility of the discs in the image. Examine rows A through E. Starting with row A, record the number of rows in which all 4 discs are clearly visible. A disc is clearly visible if its edges are discernible enough to circumscribe its perimeter. Stop scoring when you first encounter a row where less than 4 discs are clearly visible. Now examine the largest disc in each row. Starting with row A and working towards row E, record the number of large discs that are clearly visible. Stop scoring when you encounter the first large disc that is not clearly visible. The same definition of clearly visible applies as before. The images below give an example of how the images are to be scored. Page 3 of 9

This image would be scored as having 3 rows (A, B and C) where all discs are clearly visible. It would have a score of 4 for the number of large discs clearly visible. This image would be scored as having 3 rows (A, B and C) where all discs are clearly visible. It would have a score of 4 for the number of large discs clearly visible. Page 4 of 9

This image would be scored as having 2 rows (A and B) where all discs are clearly visible. It would have a score of 3 for the number of large discs clearly visible. The first time this QC test is performed establishes baseline values against which future measurements are compared. Baselines are established kvp, mas, exposure index and systemreported dose. The baseline values must be established with the help of imaging physicists. It is important that the same exposure and viewing conditions be replicated each time the test is performed. This includes phantom positioning, SID, filtration, anatomical program, AEC configuration, display and window/level settings. Use the spreadsheet developed by Imaging Physics to record and track your results. Performance Criteria If device performance is stable, it is expected that the exposure parameters set automatically by the system, the dose, and the number of visible discs will not change significantly. The system passes if the exposure index remains within ±15%, the number of rows in which all 4 discs are clearly visible does not decrease and the number of visible large discs does not decrease by more than 1 from baseline values. If the test fails, repeat it. If you still get a failure, contact Imaging Physics. Page 5 of 9

Annual Medical Physics Review - Digital Radiography This is a sample form of the annual QC review to be conducted by a medical physicist. This form is required by MANQAP to demonstrate ongoing compliance with the QC requirements. Facility System make/model Date of review Overall QC Program Assessment ( ) ACCEPTABLE ( ) ACCEPTABLE but requires remediation ( ) NOT Acceptable. Immediate action required Department System Location Contact Person Time period of data reviewed (mm/yy to mm/yy) Daily Radiography System Inspection QC Test Status Comments Daily Image Acquisition System Inspection Daily Check of Network Status Monthly Erasure of all CR Plates Monthly Cassette Artefact Check Monthly Image Quality Phantom Test Monthly Modality Display Performance Test Monthly Retake and Reject Analysis Quarterly Cleaning of CR Plates (if recommended by vendor) Quarterly Collimator Operation Check Quarterly Door Interlocks Check Annual Physics Testing (or acceptance if equipment is new or relocated) Overall QC Program Assessment: Recommendations: Review conducted by Signature Date Page 6 of 9

Site Radiographic System Acquisition System Room Year Month Day Radiographic equipment inspection Daily Digital Radiography QC Log Acquisition device inspection Network status Month Day Radiographic equipment inspection Acquisition device inspection Network status Month Day Radiographic equipment inspection Acquisition device inspection Network status

Site Radiographic System Acquisition System Room Digital Radiography Monthly QC Log Year Month CR erasure and spot checks Image quality phantom test Modality display check Page 8 of 9

Site Radiographic System Acquisition System Room Digital Radiography Quarterly QC Log Year Day/Month CR IP cleaning (if applicable) Retake/reject analysis Collimator operation Door interlocks Page 9 of 9